Escolar Documentos
Profissional Documentos
Cultura Documentos
c o m
FIRST IN CL E A N AIR
USERS MANUAL
FOR THE
ONGUARD 3000 (OG3)
REAL-TIME E N V I R O N M E N TA L M O N I TO R I N G
(800)-222-6367
www.purafil.com
TABLE OF CONTENTS
Preface ........................................................................................................................................1 OnGuard Technology ....................................................................................................................2 How The OnGuard Works ............................................................................................................3 Upacking and Inspection ..............................................................................................................5 Installing and Powering the OnGuard 3000 ..................................................................................6 Communication............................................................................................................................8 Configuration ..............................................................................................................................9 Calibrate Temperature & Humidity ..............................................................................................10 Logging and Viewing Data ........................................................................................................11 Replacing Corrosion Sensors ......................................................................................................12 Current Outputs ........................................................................................................................12 OnGuard 3000 LCD Menus ........................................................................................................13 Purafil OG3 Connect PC Software Block Diagram........................................................................14 OnGuard PC Software (OG3) Connect ........................................................................................15 Information on Setting Corrosion Classifications ........................................................................17 OnGuard 3000 (OG3) 4-20 mA Calibration ................................................................................18 Relative Humidity and Temperature Specifications ......................................................................20
i
2654 Weaver Way Doraville, Georgia, 30340, U.S.A. tel: (770) 662-8545 (800)-222-6367 www.purafil.com
Purafil 2010 Users Manual - OG_3000 - 01
PREFACE
Rev 2.2
Airborne gaseous contaminants, as well as fluctuations in temperature and relative humidity, have always posed a threat to everything from modern electronic equipment to centuries-old artwork and archives. In recent years attention has been focused on how we can monitor and control these variables in our environment, creating a new key term: Indoor Air Quality (IAQ). Purafil, Inc., the industry leader in air filtration systems, anticipated the need to merge technology with IAQ and created the OnGuard atmospheric corrosion monitor. After thorough testing and development, Purafil introduced the patented OnGuard Atmospheric Corrosion Monitor. For OnGuards development, Purafil received the 1992 R&D 100 Award, recognizing the top 100 technologically significant products that year. OnGuard has been installed to protect the environment in worldwide facilities such as: Santa Maria delle Grazie (housing da Vincis Last Supper), Italy; the English Channel Tunnel, U.K. and France; Haindl Papier, Germany; British Nuclear Fuels, U.K.; Honeywell, Yokogawa, and Georgia Pacific, USA. Purafil thanks you for your purchase and we know you will be as pleased with your OnGuard unit as we are with its development. For additional information or questions on the product contact:
Purafil, Inc. 2654 Weaver Way Doraville, Georgia 30340 USA Phone: (800) 222-6367 or (770) 662-8545 E-mail: onguard@purafil.com Web: www.purafil.com and www.purafilonguard.com
1
2654 Weaver Way Doraville, Georgia, 30340, U.S.A. tel: (770) 662-8545 (800)-222-6367 www.purafil.com
Purafil 2010 Users Manual - OG_3000 - 01
2
2654 Weaver Way Doraville, Georgia, 30340, U.S.A. tel: (770) 662-8545 (800)-222-6367 www.purafil.com
Purafil 2010 Users Manual - OG_3000 - 01
3
2654 Weaver Way Doraville, Georgia, 30340, U.S.A. tel: (770) 662-8545 (800)-222-6367 www.purafil.com
Purafil 2010 Users Manual - OG_3000 - 01
INCREMENTAL CORROSION Incremental Corrosion is the amount of corrosion that has accumulated on a metal surface over a specific time period. Incremental Corrosion can also be referred to as the rate of corrosion buildup. The OnGuard reports Incremental Corrosion in terms of Angstroms per twenty-four hours (/24 hours). The Incremental Corrosion is an indication of the severity, or lack of severity, of a corrosion-causing event. In the OnGuard, the Incremental Corrosion is recalculated every 15 minutes for the previous 24-hour period. Because Incremental Corrosion is updated every 15 minutes, significant changes in the rate of corrosion can often be determined within a 15-minute time period. Less significant changes in the rate of corrosion may take longer to detect. In certain applications, Incremental Corrosion of less than 10 /24 hours has been determined to be within safety limits. If Incremental Corrosion of greater than 10 /24 hours is detected, alarms within the OnGuard can be activated. Incremental Corrosion is logged within the OnGuard memory module, and this information can be recalled for review and analysis. Reviewing the Incremental Corrosion is especially useful in determining the time that corrosion-causing events begin or end. TEMPERATURE EFFECTS ON CORROSION Increases in Temperature can accelerate the rate of corrosion by increasing chemical reactions. Using the OnGuard to monitor and record Temperature, increases or decreases in Cumulative Corrosion and Incremental Corrosion can be correlated to Temperature effects. The Temperature sensor is read continuously, and the data is averaged over a one-minute time period. This information can be logged within the OnGuard unit and reviewed for analysis to determine if increased Temperature was a factor in a corrosion event. RELATIVE HUMIDITY EFFECTS ON CORROSION Varying levels of Relative Humidity can significantly affect the rate of corrosion. Changes in the Relative Humidity can result in condensation on surfaces within a space, and should corrosive gases be present, they can dissolve into this water layer. Many corrosive gases are often referred to as acid gases in that when they are dissolved in water, the resulting mixture is an acid. For instance, chlorine (Cl2) gas dissolved in water creates hydrochloric acid (HCl). Therefore, Relative Humidity often needs to be carefully monitored and controlled in computer control rooms. The OnGuard monitors and records Relative Humidity for correlation to events that are associated with increased rates of Cumulative Corrosion and Incremental Corrosion. The Relative Humidity sensor is read continuously and the data is averaged over a one-minute time period. This information can be logged within the OnGuard unit and reviewed to determine if Relative Humidity was a factor in a corrosion event. STANDARDS Standards can be used to define the limits of acceptable performance. These limits may be formally set by a committee acting on behalf of an organization or informally set by an individual based upon a particular application. The OnGuard is capable of containing one preloaded formal corrosion standard, which the OnGuard unit will default to on initial start-up. The OnGuard units are also capable of having a user-defined corrosion standard input. Similarly, the acceptable limits for Temperature and Relative Humidity can be selected from the formal, default standard, or an informal standard more applicable to a particular application. Initially, the OnGuard has been set up with industrial standards for corrosion written by the Instrument Society of America (ISA). These standards are ISA S71.04-1985: Environmental Conditions for Process Measurement and Control Systems: Airborne Contaminants and ISA S71.01-1985: Environmental Conditions for Process Measurement and Control Systems: Temperature and Humidity. They were written by a panel of experts in the corrosion field, ranging from manufacturers of sophisticated electronics and gas-phase filtration to professional engineers and end users in the industrial marketplace. The standards have been globally accepted and used in areas where sensitive electronics are housed in industrial complexes. 4
2654 Weaver Way Doraville, Georgia, 30340, U.S.A. tel: (770) 662-8545 (800)-222-6367 www.purafil.com
Purafil 2010 Users Manual - OG_3000 - 01
5
2654 Weaver Way Doraville, Georgia, 30340, U.S.A. tel: (770) 662-8545 (800)-222-6367 www.purafil.com
Purafil 2010 Users Manual - OG_3000 - 01
BATTERY OPERATION
OPENING THE ONGUARD 3000 FOR BATTERY INSTALLATION Remove the OnGuard 3000 from the protective bag it was shipped in. Remove the two screws that hold the front half and the back half of the case together. These are located in the middle on each side of the unit. With these screws removed the case will come apart, providing access to the battery holders. IMPORTANT! The batteries must be installed correctly! In both battery holders, the outline image of a battery is molded into each location where a battery is to be installed. Please orient the batteries as indicated. INSTALLING THE CLOCK BATTERY A standard lithium battery (CR123) must be installed for proper clock operation. Install this battery, noting the proper polarity, in the small battery holder inside the OnGuard 3000. The time displayed will be invalid until it is set via the OG3 Connect program. INSTALLING THE MAIN BATTERIES The OnGuard is designed to operate on four standard AA alkaline batteries. The unit will begin operation when the last of the four batteries is installed. For longest operating life, use high capacity batteries like the Duracell Ultra or Energizer Max. When the batteries are installed reassemble the case by installing the two screws that were removed to open the case.
Battery Life Log Interval (Minutes) 1 2 5 10 15 30 60 120 X/Hr 60 30 12 6 4 2 1 0.5 % On 96.7% 48.3% 19.3% 9.7% 6.4% 3.2% 1.6% 0.8% mAHr 18.38 9.39 4.00 2.20 1.60 1.00 0.70 0.55 WHr 0.0607 0.0310 0.0132 0.0073 0.0053 0.0033 0.0023 0.0018 Hours 158 310 728 1324 1820 2911 4158 5291 Days 6.6 12.9 30.3 55.1 75.8 121.3 173.2 220.5
6
2654 Weaver Way Doraville, Georgia, 30340, U.S.A. tel: (770) 662-8545 (800)-222-6367 www.purafil.com
Purafil 2010 Users Manual - OG_3000 - 01
EXTERNAL POWER
There are two ways to provide external power to the OnGuard 3000. 1. EXTERNAL DC POWER SUPPLY The OnGuard 3000 can be powered by an externally supplied DC voltage between 9VDC and 30VDC. The power source should be able to supply at least 300mA of current. This voltage is applied to pins 9 and 10 of the 10-pin terminal block on the bottom of the OnGuard. Pin 1 of the terminal block is on the left when viewed from the front, pin 10 on the right. The negative lead of the power source is connected to pin 9, and the positive lead is connected to pin 10. When operating from an external power supply the OnGuard will operate continuously and will not power down into sleep mode as it does when powered from the batteries. If the external power supply fails or is turned off the OnGuard 3000 will continue to operate from the internal batteries if they are installed. This allows continuous operation even during power outages. 2. POWERED FROM A PC USB PORT The OnGuard 3000 can also be powered by connecting the OnGuard 3000 to a PC via the USB port on the bottom of the unit. All functions are available and the OnGuard 3000 will operate the same as if it is powered from an external power supply.
7
2654 Weaver Way Doraville, Georgia, 30340, U.S.A. tel: (770) 662-8545 (800)-222-6367 www.purafil.com
Purafil 2010 Users Manual - OG_3000 - 01
COMMUNICATION
The OnGuard 3000 is designed to communicate with a Windows-based PC via a USB port, using the OG3 Connect software. The OG3 Connect software application provides a convenient way to configure the OnGuard 3000 and extract measured data. The most reliable results have been obtained with computers containing a USB 2.0 compatible USB port and Windows XP or XP Pro operating system. The USB cable must have a standard type A plug on one end and a mini-B plug on the other. A travel-type cable is supplied with the OnGuard 3000. Here is the procedure to establish communication between the OnGuard and the PC: 1. Plug the USB cable into the PCs USB port. 2. Start the OG3 Connect software on the PC. 3. Apply power to the OnGuard. If operating on batteries, wake the unit by pressing the ENTER key on the OnGuard 3000. 4. Plug the USB cable into the small USB connector on the bottom of the OnGuard 3000. Be sure to orient the con nector correctly and do not force a connection that does not appear to fit. If your PC sound is not muted and the volume adjusted so it can be heard, you should hear the familiar USB connect beeps from the PC indicating a connection has been made. 5. In the lower left corner of your PC screen you should see the word Connected after a few seconds. You can now interact with the OnGuard 3000.
NOTES:
Always unplug the USB cable after closing the OG3 Connect program. Wait at least 20 seconds after last data communication (configuration or data download) before closing the OG3 Connect program. If the OG3 Connect program is connected to the OnGuard 3000 and the USB cable is accidentally unplugged, just close the program. The program may take up to 30 seconds to close.
8
2654 Weaver Way Doraville, Georgia, 30340, U.S.A. tel: (770) 662-8545 (800)-222-6367 www.purafil.com
Purafil 2010 Users Manual - OG_3000 - 01
CONFIGURATION
The OnGuard 3000 may need to be configured to operate properly in your location, and provide the data desired. The configuration includes setting up the data logger, the date and time, corrosion sensors, and temperature and relative humidity limits. The OG3 Connect software is used for configuration. On the main menu bar click on Device, then Configure. A summary of the available functions is listed below: DATA LOGGER TAB The data logger can be enabled and disabled from this screen. You can also set the logger interval from 1 to 120 minutes. You can maximize battery life by setting the logger interval to a higher value (15+ minutes) as the OnGuard 3000 uses more power during readings than between readings, when it sleeps. The most power is used when interfacing with an operator via the keypad and LCD display. Batteries can be made to last several months with proper setup and minimal operator intervention. The OnGuard 3000 can store over 21,000 readings in its logger memory. LOGGER MEMORY CAPACITY Log Interval 1 Minute 2 Minutes 5 Minutes 15 Minutes 30 Minutes 60 Minutes 120 Minutes Hours 350 700 1750 5250 10500 21000 42000 Days 14 29 72 218 432 864 1728
CLOCK/CALENDAR TAB The date and time must be set. Make sure the OnGuard is connected to the PC and communication is established. From the configuration screen, click on the Clock/Calendar tab. Enter the time and date in the spaces provided and click the Apply Changes button. SENSORS TAB The OnGuard is shipped with copper and silver sensors installed. Copper is on top and silver is on the bottom. From the configuration menu, click on the Sensors tab. Click on the Read Device button to see the current setting for the sensors. By default the Classification Level will be set for Std (Fixed). The table of values selected by this choice corresponds to the ISA standards. Custom levels may be entered by setting the Range to Custom. You can then enter the upper value for each range. The lower value is automatically set. See the appendix for additional information on the Classification Levels. TEMP & HUMIDITY TAB The OnGuard 3000 allows you to set high limit, low limit, and rate-of-change limits for temperature and relative humidity. When limits are exceeded, the red LED will flash.
9
2654 Weaver Way Doraville, Georgia, 30340, U.S.A. tel: (770) 662-8545 (800)-222-6367 www.purafil.com
Purafil 2010 Users Manual - OG_3000 - 01
10
2654 Weaver Way Doraville, Georgia, 30340, U.S.A. tel: (770) 662-8545 (800)-222-6367 www.purafil.com
Purafil 2010 Users Manual - OG_3000 - 01
When viewing the graphed data the Zoom+ can be used in conjunction with the lower scroll bar to enlarge any area of interest. The Zoom- key can be used to zoom back out. The left vertical axis of the graph is the scale for temperature and relative humidity. The temperature can be scaled to 0-100 degrees C or 0-100 degrees F. The humidity scale is 0-100%.
11
2654 Weaver Way Doraville, Georgia, 30340, U.S.A. tel: (770) 662-8545 (800)-222-6367 www.purafil.com
Purafil 2010 Users Manual - OG_3000 - 01
CURRENT OUTPUTS
The OnGuard 3000 has four 4-20mA current outputs that correspond to Copper Cumulative, Silver Cumulative, Temperature, and Relative Humidity. The current outputs are accessible in the large 10-pin terminal block on the bottom of the OnGuard 3000. Pin 1 and Channel 1 is located on the left end of the terminal block as viewed from the front. There is no LCD screen on the OnGuard 3000 that deals with the Current Loop Outputs. These outputs are disabled when the OnGuard 3000 is battery operated. Channel 1, Copper Cumulative, Pin 1 (-), Pin 2 (+), 4-20mA = 0-4000 Angstroms Channel 2, Silver Cumulative, Pin 3 (-), Pin 4 (+), 4-20mA = 0-4000 Angstroms Channel 3,Temperature, Pin 5 (-), Pin 6 (+), 4-20mA = 0-100C Channel 4, Relative Humidity, Pin 7 (-), Pin 8 (+), 4-20mA = 0-100% OnGuard External Power, Negative Input, Pin 9 OnGuard External Power (+12-30VDC), Positive Input, Pin 10
12
2654 Weaver Way Doraville, Georgia, 30340, U.S.A. tel: (770) 662-8545 (800)-222-6367 www.purafil.com
Purafil 2010 Users Manual - OG_3000 - 01
MAIN MENU When the OnGuard 3000 is powered up, the Main Menu Screen appears as shown above. This screen displays the measured results from the on-board sensors, which include copper and silver corrosion sensors, temperature, and relative humidity. The first line shows the screen title and the remaining battery life. The second line shows the copper sensor classification (G1) and the current cumulative value (3). The third line shows the silver sensor classification (P1) and the current cumulative value (2). The fourth line shows the current temperature in Celsius and Fahrenheit. The fifth line shows the current value for Relative Humidity. The sixth line shows the date and time in 24 hour format. ALARM MENU The Alarm Menu shows the status of the OnGuard 3000 sensors and the main power batteries. The status of each will be shown as OK unless there is a problem with one of them. When an alarm condition occurs the red LED will light. The temperature (TEMP) and relative humidity (RELH) sensors will show an L for low, H for high, or R for rate of change limit exceeded. The SENS will show a 1 for copper, a 2 for silver, or 1/2 for both if any sensor has failed or reached end-of-life. The BATT entry shows FAULT when the battery level falls below 10%. LOGGER MENU The Logger Menu shows the log interval. The value may be changed by pressing the UP or DOWN arrows on the keypad. Cycling through the interval values will present options for enabling and disabling the logger, and clearing the logger memory. When the desired option is displayed, pressing the ENTER key will implement that option.
VERSION MENU The Version Menu displays the current revision of the software in the OnGuard 3000.
13
2654 Weaver Way Doraville, Georgia, 30340, U.S.A. tel: (770) 662-8545 (800)-222-6367 www.purafil.com
Purafil 2010 Users Manual - OG_3000 - 01
14
2654 Weaver Way Doraville, Georgia, 30340, U.S.A. tel: (770) 662-8545 (800)-222-6367 www.purafil.com
Purafil 2010 Users Manual - OG_3000 - 01
15
2654 Weaver Way Doraville, Georgia, 30340, U.S.A. tel: (770) 662-8545 (800)-222-6367 www.purafil.com
Purafil 2010 Users Manual - OG_3000 - 01
DEVICE > CONFIGURE This function is used to configure the OnGuard 3000. This includes the data logger, the date and time, corrosion sensors, and temperature and relative humidity limits. A summary of the available functions is listed below: Logger Tab Logger On/Off Logger Interval (1, 2, 5, 10, 15, 30, 60, 120 minutes) Buttons: Read Device, Apply Changes Clock/Calendar Tab Time: HH:MM:SS Date: Month, Day, Year Buttons: Read Device, Apply Changes Sensors Tab Sensor Type Range: Std (Fixed) or Custom Levels (G1, G2, G3, GX, P1, P2, P3, PX) Buttons: Read Device, Apply Changes Temp & Humidity Tab Temp: Low, High, Rate, (Scale: Celsius or Fahrenheit) Humidity: Low, High, Rate Buttons: Read Device, Apply Changes DEVICE > CALIBRATE TEMPERATURE This function is used to calibrate the temperature sensor in the OnGuard 3000. Set the scale to degF or degC, enter the current temperature and click OK DEVICE > CALIBRATE HUMIDITY This function is used to calibrate the relative humidity sensor in the OnGuard 3000. Enter the current relative humidity and click OK. WINDOW This function shows what OG3 Connect program modules are currently active. ABOUT This function provides the version of the OG3 Connect software.
16
2654 Weaver Way Doraville, Georgia, 30340, U.S.A. tel: (770) 662-8545 (800)-222-6367 www.purafil.com
Purafil 2010 Users Manual - OG_3000 - 01
17
2654 Weaver Way Doraville, Georgia, 30340, U.S.A. tel: (770) 662-8545 (800)-222-6367 www.purafil.com
Purafil 2010 Users Manual - OG_3000 - 01
5. In the Device Configuration window, select the far right tab labeled 4-20 Calibration. At this point, the current displayed on the ammeter will represent the output channel (1-4) and level (min or max) as selected on the 4-20 calibration window. The OG3 will hold the selected level, bypassing normal 4-20 functionality, for a period of 60 seconds every time an adjustment selection from this window is chosen.
18
2654 Weaver Way Doraville, Georgia, 30340, U.S.A. tel: (770) 662-8545 (800)-222-6367 www.purafil.com
Purafil 2010 Users Manual - OG_3000 - 01
OPERATION
OUTPUT Selecting any of the output (Channel 1 through Channel 4) will force the currently selected level (Minimum or Maximum) to be output on the selected channel. LEVEL Selecting either option (Minimum or Maximum) will force the currently selected channel to its minimum or maximum level. ARROW KEYS Selecting these will adjust the current on the selected channel up or down by an amount corresponding to the level shown in the drop-box immediately below. STEP AMOUNT This defines the step size affected by the arrow keys. The steps are roughly as follows: X-Fine 0.005 mA Fine 0.05 mA Medium 0.15 mA Course 0.50 mA UNDO Reset will undo any adjustments made since the last time Store Changes was selected for the selected channel and selected level. For example, if Channel 1 and Minimum are selected, clicking Reset will clear changes made to channel 1 minimum. Channel 1 Maximum and other channels will not be affected. STORE CHANGES This stores the changes for the currently selected channel and level. For example, if Channel 1 and Maximum are selected, clicking Store Changes will store the changes made to the maximum level of channel 1, channel ones minimum level and the other channel calibrations will not be stored. This allows the user to change any single attribute of all channels minimum and maximum levels; however, Store Changes must be selected for each channel and that is level intended to be calibrated before exiting the calibration window.
PROCEDURE: 1. Follow the setup steps listed in the Preparation section above. 2. Familiarize yourself with the operation of the calibration window controls as shown above. 3. For each channel: a. Connect a suitable ammeter to the channels output pins b. Select the desired channel in the calibration window. c. Select Minimum (4 mA) d. Verify that the ammeter reads 4.00 mA. Adjust up or down using the arrow keys until the desired level is obtained while using the step amount control to achieve an appropriate adjustment size. e. Select the Store Changes button to store the calibration adjustment. f. Select Maximum (20 mA) g. Verify that the ammeter reads 20.00 mA. Adjust up or down using the arrow keys until the desired level is obtained while using the step amount control to achieve an appropriate adjustment size. h. Select the Store Changes button to store the calibration adjustment. 4. Exit the calibration window and disconnect the ammeter. The OG3 will continue normal 4-20 circuit operation 60 seconds after the most recent (calibration window) button-press. 19
2654 Weaver Way Doraville, Georgia, 30340, U.S.A. tel: (770) 662-8545 (800)-222-6367 www.purafil.com
Purafil 2010 Users Manual - OG_3000 - 01
20
2654 Weaver Way Doraville, Georgia, 30340, U.S.A. tel: (770) 662-8545 (800)-222-6367 www.purafil.com
Purafil 2010 Users Manual - OG_3000 - 01