Flow Measurement
Range: Dependent upon molbloc size and pressure calibration range (see molbloc-S Ranges table) Required Back Pressure Ratio: 0.4 to 0.9 depending on Reynolds number Calibration Gases: Standard: Dry air Optional: N2 (to replace standard air calibration or in addition), Ar, H2, He, N2O. Consult factory regarding availability of availability of these gases for specific ranges, or for other gases. *Dry air calibration is required for humid air operation molbloc-S Pressure Calibration Ranges: Low Pressure (LP): 20 to 200 kPa absolute (3 to 30 psia) upstream Standard Pressure (SP):
<< Back
50 to 500 kPa absolute (7 to 70 psia) upstream High Pressure (HP): 200 to 2 MPa absolute (30 to 300 psia) upstream * HP calibration usable with molbox1+S terminal only * Flow measurement specifications valid from 10% to 100% of molbloc full scale Resolution: 0.0015 % of FS 0.01 % of reading Precision: 0.06 % of reading 0.30 % of reading Predicted Stability (One Year): 0.05 % of reading 0.2 % of reading Measurement Uncertainty 0.2 % of reading Premium calibration: 0.125 % of reading 0.5 % of reading from 50 to 500 kPa