Reference Manual

Section 5: Operation and Maintenance

00809-0100-4100, Rev AA

March 2013

suply
failur

suply
low

radio
failur

confg
failur

Operation and maintenance

Diagnostic Button Screen 7.1: the


terminal voltage has dropped
below level of operating limit.
Replace the Power Module
(Part Number: 701PGNKF)

Diagnostic Button Screen 7.2: the


terminal voltage is below the
recommended operating range the Power Module should be
replaced

Diagnostic Button Screen 8: the


device may not be able to
communicate with the radio or
the radio has an internal error. In
this state the device may still be
operational and publishing HART
data

Diagnostic Button Screen 9.1:


configuration of the transmitter is
invalid such that critical operation
of the device may be affected check the extended configuration
status to identify which
configuration item(s) need to be
corrected