Escolar Documentos
Profissional Documentos
Cultura Documentos
Olson 2
Figure 1: Isodose Distribution of an AP/PA Plan with Equal Weighting using 6 MV
Olson 3
Plan 2: AP/PA Plan with Equal Weighting using 15 MV
A. A change in beam energy from 6 MV to 15 MV pulled the isodose lines back towards the
surface, decreasing the overall hot spot (Figure 2). In addition, there is a more pronounced
build-up region below the skin surface when using 15 MV, due to its greater Dmax. This
build-up region does not pose any problems as the PTV is located quite deep within the
patient. As stated by Bentel,1 an increase in beam energy can be beneficial when treating
areas with a larger separation. An increased energy allows for greater penetrating ability,
ultimately decreasing the amount of power needed to deliver the prescribed dose to the PTV.
B. Increasing each fields energy from 6 MV to 15 MV resulted in a decrease in the overall hot
spot from 133.2% to 116.8% respectively (Figure 2). Despite the reduction in overall hot
spot, their location is still the same: at the anterior and posterior surfaces of the patient. As
was observed in the 6 MV plan, the larger hot spot of 116.8 is located on the anterior surface
of the patient in soft tissue and bone. Even though 15 MV provides a greater penetrating
ability, this plan is still equally weighted, which requires the anterior field to use more power
to reach the isocenter. The interacting photons more energy, but still interact with soft tissue
and bonejust not as much as previously seen with 6 MV. A neat way to visualize this
difference was an equation provided by my physicist, Daniel Lewis, MS (oral
communication, March 2016). Using a percent depth dose table at a depth of 10 cm: 6MV
delivers approximately 70% of prescription dose while 15 MV delivers approximately 80%.
1
6 MV :
=1.4 hotspot relative prescription dose
0.7
15 MV :
1
=1.2 hotspot relative prescription dose
0.8
Olson 4
Olson 5
Olson 6
Figure 3: Isodose Distribution of an AP/PA Plan with Unequal Weighting using 15 MV
Plan 1
Plan 2
Plan 3
Maximum Dose
Pixel for PTV
113.2%
107.0%
107.0%
Minimum Dose
Pixel for PTV
94.8%
96.1%
96.4%
Mean Dose
Pixel for PTV
107.1%
103.2%
102.9%
Olson 7
Plan 4: Three-Field Lung Plan using 15 MV
A. When adding a third beam into the plan using 15 MV, it is best to position the beam at a
slight oblique to avoid the cord and minimize the dose to the healthy lung tissue.1 To do this,
I tilted the gantry to 310 (A50R). Another option could be 130 or P50L. In this case, it
would be best to utilize the A50R because of concern for table collision when considering
P50L. By doing so, the cord is avoided, and the amount of lung tissue involved is
minimized. To further decrease the amount of normal tissue involved, I also decreased the
margin along the side of the spinal cord and decreased my X2 jaw size from 4.1 cm to 3.8
cm. These things in combination also decreased the hot spot area to 104.8% at the posterior
aspect of the patient. The one concern I have is that while the PTV is receiving a mean dose
of 102.0%, a large portion of healthy lung is also receiving the same amount of prescription
dose. For this reason, it may be beneficial to adjusting the beam weighting or consider a
wedge.
B. Altering the weighting provided some excellent benefitsmore specifically to the
surrounding healthy tissue. Keeping the radiosensitivity of lung tissue in mind, I adjusted the
weighting contribution to AP with 28%, PA with 37.6%, and A50R with 34.4%. This helped
to drastically reduce dose to healthy tissue in the anterior portion of the lung, to only 50%
(Figure 4). Reducing dose to healthy lung tissue is of great importance in order to minimize
lung complications for curative or palliative treatments. Finally, other aspects of the plan
remained unchanged; the hot spot remained at 105% and the minimum dose to the PTV was
96%.
C. Addition of wedges was not as helpful as I had hoped. I was aiming to reducing some of the
prescription dose in the anterior area where the A50R and AP beam converged. I tried
different combinations of wedging with for both fields with the heel towards the anterior on
the A50R and heel towards the patients right on the AP. The wedges did not reduce the
triangle of healthy lung tissue receiving prescription dose, but rather increased the hot spot
posterior to the PTV and reduced the mean coverage to the PTV. Instead of using two
wedges, I tried using a single wedge on the AP with the heel towards the right. Since the AP
contribution was relatively low, I used a 30-degree wedge. This wedge helped to push the
prescription isodose line more medially, where I was lacking coverage. In addition, the hot
spot decreased to only 103.6%. Because my hot spot was relatively low, I finally increased
my normalization value from 293.0 to 289.5 to improve my overall coverage. My end result
achieved was 97.6% of the PTV receiving 100% of prescription dose, with a hotspot of 105%
(Figure 5).
Olson 8
Figure 4. Adjusted Weighting on Plan 4 using 3-Fields and 15 MV
Olson 9
Overall Impression
The best treatment plan of all created would be Plan 4. This plan is conformed tightly to the
tumorreducing dose to normal lung tissue, and has a minimal hot spot of 105%. The area of
hot spot is located just posterior to and to the right of the PTV. As stated before, the coverage of
Plan 4 to the PTV achieved 97.6% to receive 100% of the dose.
The only constraint that was not achieved by this plan was the mean total lung dose (Table 2).
Because I have room for adjustment, I could increase my normalization value to lower the mean
dose. I could also manipulate the wedge or weighting contributions for other fields to lower lung
dose as well.
This planning assignment was very helpful in learning different ways to plan a lung treatment via
the creation of 4 different treatment plans. The lung area is such a radiosensitive structure and
care must be taken to minimize dose to the surrounding healthy tissue while targeting the lesion.
While my center plans most lung treatments using Rapid Arc, I look forward to having the
opportunity to plan the next 3D lung treatment when the occasion arises. This assignment was
extremely helpful in getting to see the different changes that occur with the addition of wedges,
energies, and altered weighting. Due to many time constraints in the real world of planning, this
assignment allowed me to take my time and fully evaluate the results of each action requested
within this assignment.
Something I would try to do differently next time is to find a way to minimize prescription dose
to the healthy lung tissue. As shown in Figure 5, I have a large diamond shape of lung tissue
healthy and tumorousreceiving full prescription dose. The portion of lung anterior to the PTV
is of most concern. I could add in right lateral field with minimal contribution to help reduce the
peaked areas of full prescription. A right lateral field would increase the dose to the esophagus
spinal cord, and total lung but based on my achieved objectives in Table 2, there may be room
for adjustment.
Desired Objective
Max < 3600 cGy
Max < 3000 cGy
Achieved Objective
Max: 251.9 cGy
Max: 5778.7 cGy
4.1% > 3000 cGy
Mean: 782.8 cGya
V20: 15.1%
Mean: 36.6 cGy
Max: 1.0 cGy
Mean: 42.7 cGy
V35: 0%
Olson 10
References
1. Bentel GC. Radiation Therapy Planning. 2nd ed. New York, NY: McGraw-Hill; 1996.
2. Kahn FM, Gibbons JP. Kahns The Physics of Radiation Therapy. 5th ed. Philadelphia, PA:
Lippincott Williams & Wilkins; 2014.