Service

Service Instructions

Service

SM-30-4043-A1-en

Service
Service

Issue:
Replaces issue:

08.07.2003
20.09.2002

page

of 39

MS - TS 1

Contents
Page

Contents .................................................................................................................... 2
Update information ................................................................................................... 4
General information.................................................................................................. 5
Cleaning Care and Servicing of the unit ........................................................................ 5

Tools and test devices ............................................................................................. 5


VISUCAM lite - Unit..................................................................................................... 6
Unpacking / Packing ........................................................................................................ 6
Cable connections ........................................................................................................... 7
Connector panel provided on the display........................................................................ 7
lite
Connections provided on VISUCAM ........................................................................... 7
Connections provided on isolating transformer and additional units ............................... 7
System overview .............................................................................................................. 8
Design............................................................................................................................... 8
VISUCAM lite connector panel ......................................................................................... 8
Overview of functions ..................................................................................................... 9
Help for troubleshooting ............................................................................................... 10
Questions about mechanical problems......................................................................... 10
Questions about electrical problems............................................................................. 10
lite

VISUCAM Components Service ............................................................................. 11


Replacing the blinking diode of the fixation light ........................................................... 11
Opening and closing the camera head ......................................................................... 11
Replacing the halogen bulb .......................................................................................... 11
Separating the camera head from the instrument base ................................................ 12
lite
Replacing the VISUCAM control cable...................................................................... 12
Replacing the fan cooling the lamp housing ................................................................. 13
Replacing the temperature sensor................................................................................ 13
Replacing the light guide .............................................................................................. 14
Exciter filter drive .......................................................................................................... 15
Replacing the stepper motor of the exciter filter ........................................................ 15
Replacing the opto-coupler of the exciter filter .......................................................... 15
Barrier filter drive .......................................................................................................... 15
Replacing the stepper motor of the barrier filter ........................................................ 15
Replacing the opto-coupler of the barrier filter .......................................................... 15
Finder camera .............................................................................................................. 16
Replacing the finder camera ..................................................................................... 16
Positioning the finder camera.................................................................................... 16
Adjusting the focus of the finder camera ................................................................... 16
Checking and adjusting the image position of the finder camera .............................. 17
Replacing the complete lamp housing .......................................................................... 18
Camera module............................................................................................................ 18
Replacing the cameras ............................................................................................. 18
Adjusting the cameras .............................................................................................. 18
Replacing the VISUCAM lite control PCB....................................................................... 19
Opening and closing the control base........................................................................... 20
Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

of 39

Replacing the headrest................................................................................................. 21


Components of the control base................................................................................... 22
Replacing the hard disk............................................................................................. 22
Connections on the GIF instrument interface ............................................................... 23
Connections on the motherboard ................................................................................. 23
Connections on the power supply unit .......................................................................... 23
System settings and work with internal programs...................................................... 24
Activating the system settings ...................................................................................... 24
Mouse ....................................................................................................................... 24
Keyboard .................................................................................................................. 25
Printer ....................................................................................................................... 25
Network..................................................................................................................... 26
Date / Time ............................................................................................................... 28
Regional settings ...................................................................................................... 28
Opening the Windows Explorer program ...................................................................... 29
Closing an active window ............................................................................................. 29
Opening watch files ...................................................................................................... 29
Enter BIOS ................................................................................................................... 29
Password for BIOS ................................................................................................... 29
Installing, uninstalling CD-writer.................................................................................... 29
Opening the Test Tools program .................................................................................. 30
Overview Test Tools program ................................................................................... 30
Service work................................................................................................................... 31
Software ....................................................................................................................... 31
Database .................................................................................................................. 31
Database Images...................................................................................................... 31
Database Sequences................................................................................................ 31
Database TNIs......................................................................................................... 31
VISUCAM Save and Update folder ........................................................................... 32
VISUCAM Installs folder............................................................................................ 32
VISUCAM HP Driver folder ....................................................................................... 32
Data setup .................................................................................................................... 33
Stepper Motor control exciter filter ................................................................................ 34
Stepper Motor control barrier filter ................................................................................ 34
Stepper Motor control camera shift............................................................................... 34
Testing focus drive ....................................................................................................... 35
Testing grabber function and quality of the sensors ..................................................... 35
Measuring the operating voltages................................................................................. 35
Measuring the operating voltages................................................................................. 36
Voltages present at Interface GIF PCB ..................................................................... 36
Voltages present at the VISUCAM lite control PCB .................................................... 37
Voltages present at the VISUCAM lite lamp Control PCB ........................................... 37
Checking the electrical functions .................................................................................. 37
System check ................................................................................................................. 38
Mechanical check ......................................................................................................... 38
Electrical function check ............................................................................................... 38
Adjustment Circle .......................................................................................................... 39

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

of 39

Update information
Issue

Remarks

20.09.2002

Issue 1

08.07.2003

Issue 2: Software transfer or update now in SM 4043-A2

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

of 39

General information
Use the User Manual, SAP No.: 1165-216, as additional literature on the VISUCAM lite fundus
camera. You must be familiar with the contents of the user manual to be able to work with the
service instructions.
The service manual mainly covers the defined spare parts and the description of how the most
important components are replaced.
The replacement modules have been chosen in such a way that replacing them should be
possible without a special training. If in-depth work on the unit is necessary, you should attend
the appropriate service training.
lite
The VISUCAM fundus camera does not have to be installed by service staff. For this reason,
no Service Report is necessary.
No fastening components are allocated to and supplied with the spare parts. For this reason,
the fastening components provided in the unit must be re-used when modules are replaced.
Caution:

After the unit has been opened, hazardous voltages are accessible. Take the
necessary precautions against the coming into contact with hazardous voltages and
comply with the applicable work instructions.

Cleaning Care and Servicing of the unit


Also see the Care and Maintenance chapter in the user manual.
The unit requires almost no servicing. We recommend having the unit serviced once a year on
the basis of an appropriate support contract.
The points listed in the "System check" chapter represent a work plan which must be completed
when a unit is serviced.
The optical surfaces can be cleaned using an ether-methylated spirits mixture (9:1), a special
optics cleaning mixture (SM-30-8005-32) or methanol.

Tools and test devices


Material number

Benennung deutsch
lite
1239-573 Servicekit VISUCAM

English description
Service kit VISUCAM lite

1.

1212-348

Testtool Fokusebene

Test tool for focus plane

2.

1194-369

Schlssel fr Sucherkamera

Key for Finder Camera adjust

3.

0118-751

Sechskantschlssel Zoll 3/32


(Netzteildemontage)

Hex key, inch 3/32


(removal of power supply unit)

4.

1242-737

Recovery CD VISUCAM lite R1.02

Recovery CD VISUCAM lite R1.02

5.

1240-531

Justierfarbtafel

Color test picture

6.

0449-685

Graufilter 2,5%

Grey filter 2,5%

7.

1046-391

Halter fr Demoauge

Holder for Demo Eye

1-7 ist Teil des Servicekits

1-7 is part of the service kit

1177-617 Demoauge mit Halter

Demo eye with clamp

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

of 39

VISUCAM lite - Unit


Unpacking / Packing

Open the packaging.


Open and remove the extra
package containing the
accessories.

lite
Lift out VISUCAM in
protective packaging.

Remove the fastening tape.

Remove the fixation lamp


from its mount.

Remove the plug and the


control cable from their
mounts.

Remove the molded


support at the top.
Remove the molded
support at the bottom.

Remove the front half cover


Remove the molded support at
the top.
Place the support under the
lite
VISUCAM to prop it up.

Remove the complete


lite
VISUCAM from the
packaging and place it on a
table.

Remove the plug and the


control cable from their mounts.

Remove the 2 transport supports on the left and


right.

Remove the protective plate.

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

of 39

Cable connections
Connector panel provided on the display
The display connector panel is also used as a USB
HUB.
All USB connections must be connected here.
Exception: The HP Photosmart printer, Series 100,
lite
must be connected to the VISUCAM
USB port(status 7/02).
Line power
lite
Connection to VISUCAM USB port
Mouse
Keyboard
External CD-ROM burner
Printer
lite
Display connection from VISUCAM monitor outlet
Connections provided on VISUCAM

lite

lite

The VISUCAM is not connected via the isolating


transformer.
lite
The VISUCAM is connected separately to the
power supply.
Power switch
Line power
Display connection, monitor outlet
lite
Connection to VISUCAM USB port
Second USB port e. g. for HP Photosmart Series 100
lite
Connection cable for VISUCAM camera head
Connections provided on isolating transformer and additional units
The isolating transformer is needed to reduce the earth leakage currents produced by
additional units.
Printer
Line power connection, isolating transformer
Power switch, isolating transformer
Isolating transformer
Connection for printer
Connection for display
Connection for CD-ROM writer
CD-ROM writer
Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

of 39

Notes on servicing:
Always disconnect the power plug before opening the unit.
After servicing, check all functions and, if necessary, the electrical safety of the unit.

System overview
Fixation light
Headrest

iInternal fixation target

Camera
head

Instrument
base
Display

Control
base
Keyboard
Mouse

Design
The base, mechanical stage, instrument head and headrest form one functional unit.
The connector panel on the base provides the following connections:
lite
The VISUCAM is switched on/off using power switch (1).
The power cord is connected to power inlet (2).
The display is connected to VGA connector (3). (The display has its own power supply and is
not switched on/off using the power switch.)
The camera head is connected to DEVICE connector (4).
BNC VIDEO socket (5) is provided for external video recordings.
A printer can be connected to LPT1 parallel port (6).
COM1 serial port (7) is provided for the connection of serial units.
lite
NET network connector (8) for integration of the VISUCAM in a network.
The keyboard and the mouse are connected to USB port (9). Another USB port is available
for the connection of additional USB units.
VISUCAM

lite

connector panel

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

of 39

Overview of functions
Camera head
Instrument base
+5V

X3
X7

Brightness

X4

X20

Release

X8

X50 Programming
Halogen
bulb

M Stepper
Exciter filter drive
Opto-coupler

X5

Temperatur
sensor

Fan

M Stepper
Barrier filter drive
Opto-coupler

Control of
VISUCAM lite

+12V

X12

X10
X42

X45

X46

X11

X43

X9

+12V La

X6

X41

+5V

+12V

X16
X18
X17

+5V
+12V

Insul.
Transf.

Line

115/230V
AC
115VA

X141

Signale

+12V

+12V

MUX

X1

X2

X12

X14

X102

X112

X114

X101

+12V color

Fixation

X9
X66 Camera
X11 control
Signals
Y/C

X131

Black/white
sensor
Color
sensor

Signals
BAS 2

X132

Mouse

DISPLAY

Headrest

Instr.
Table

Signals b/w
+12V

Line

Line
Power
inlet
100...240V
AC

Finder camera

M Stepper
Camera change
Opto-coupler
Camera module

Signals

Lamp control

Illumination

Keyboard

Printer
Line

USB-HUB

CD-ROM
writer

115/230V,
AC
DEVICE
X7
+5V, +12V

Power
supply

X5
X6

GIF
instrument
interface
X4

Connector panel
VGA
X1

VGA

VBS
USB1

X21

USB2

NET

COM1 LPT1 TV OUT

Computer PCB

X22

Primary IDE
+5V

Fan

CD-ROM

Position sensor
left - right

Hard disk
lite

VISUCAM
Computer base

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

of 39

Help for troubleshooting


Questions about mechanical problems
Can the camera head be swiveled all the way to the right and left?

Can the VISUCAM

lite

fundus camera be easily moved and locked on the base within the

entire working range?

Is the ophthalmoscope lens clean?

Questions about electrical problems


Have all connections been made and all plugs properly inserted?

Does the computer start up after the power switch has been switched on?

Can the operating software be started?

Can the brightness of the halogen lamp be adjusted?

Does the drive for the exciter filter initialize?

Does the drive for the barrier filter initialize?

Does the illumination of the finder camera function properly?

Does the finder camera function properly and is the image provided by it displayed on the
monitor?

Can the cameras be selected?

Does the focusing system function properly?

Does the black/white camera function properly and can images be captured?

Does the color camera function properly and can images be captured?

Does external archiving function properly?

Can you print with the printer connected?

Have the screen settings been adjusted as required?

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

10

of 39

VISUCAM lite Components Service


Replacing the blinking diode of the fixation
light

Screw out the blinking diode (1) from its


socket(2) .
Screw the new blinking diode into the socket.
Check whether it is properly functioning.

Opening and closing the camera head


lite
Disconnect the VISUCM from line power.
Remove the lamp housing cover.
(Snap lock)
Remove Focus Drive left side and right side.
Loosen 4 x hex socket head screws (2.5 mm) in the left-hand
cover.
Remove the left-hand cover.
Loosen 4 x hex socket head screws (2.5 mm) in the righthand cover.
Remove the right-hand cover.
Note:
When removing and attaching the covers, make sure
that the finder camera cover plate is in the grooves and
not damaged.

Assemble in reverse order.

Note:

When pushing the covers together, take care not to squeeze any cables.

Replacing the halogen bulb


Caution: The halogen bulb may be hot. Risk of burns!
Allow the halogen bulb to cool down.
Open the camera head
(see: Opening and closing the camera head).
Remove the halogen bulb including socket
from the lamp mount.
Take care not to bend the mount!
Remove the defective halogen bulb from the
socket.
Insert the new halogen bulb into the socket.
Insert the halogen bulb including socket into
the mount.
Close the camera head
(see: Opening and closing the camera head).
lite
Switch on VISUCAM and check its functions.
Note:
The lamp need not be readjusted
after bulb replacement.

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

1 Halogen bulb
2 socket with cable
3 lamp mount

Page

11

of 39

Separating the camera head from the instrument base

Open the camera head


(see: Opening and closing the camera
head).
Disconnect the cable for the camera head
from the connector panel.
Disconnect the 2 cables leading to the
lite
instrument base from the VISUCAM
control PCB. Remove the cable ties.

Push the cables through the connecting holes.

Securing ring

Remove the securing ring.


Close the camera head (see: Opening and closing the camera head).
Pull the camera head from the pin and lay it aside.
Mount the new camera head accordingly.
lite
Switch on VISUCAM and check its functions.
lite

Replacing the VISUCAM control cable


Open the camera head
(see: Opening and closing the camera head).
Disconnect the control cable from the
connector panel.
Disconnect the control cable from the
lite
VISUCAM control PCB and the camera
control PCB.
Remove cable ties and cable clamps from the
cable.
lite
Route the new VISUCAM control cable
as per wiring diagram and connect it.
Close the camera head
(see: Opening and closing the camera head).

Switch on VISUCAM
functions.

lite

and check its

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

12

of 39

Replacing the fan cooling the lamp housing


Note: Fan blows outward. Observe markings on the fan!

Open the camera head (see: Opening


and closing the camera head).
Disconnect the fan cable from X45 on the
lamp control PCB.
Screw off the defective fan.
Mount the new fan and connect it to X45
on the lamp control PCB.
Close the camera head
(see: Opening and closing the camera head).
lite
Switch on VISUCAM and check its functions.

Replacing the temperature sensor


Disconnect VISUCAM lite from line power.
Open the camera head (open left half only)
(see: Opening and closing the camera
head).
Disconnect the cable of the temperature
sensor
from X46 on the lamp control PCB.
Screw off the defective temperature sensor.
Mount the new temperature sensor and
connect
it to X46 on the lamp control PCB.
Close the camera head
(see: Opening and closing the camera
head).
lite
Switch on VISUCAM and check its
functions.

Halogen lamp
X42
Fan
X45
Temperature sensor
X46

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

13

of 39

Securing screw

Replacing the light guide

Open the camera head (see: Opening


and closing the camera head).
Loosen the upper mount of the light guide
and push the light guide through the hole.
Press out the light guide from the clip.
Loosen the lower mount of the light guide
and remove the light guide.
Insert the new light guide accordingly,
push it all the way into the mounts and
lock it in position.
Close the camera head (see: Opening
and closing the camera head).
lite
Switch on VISUCAM and check
its functions.

Mount

Hole

Clamp screw

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

14

of 39

Exciter filter drive


Replacing the stepper motor of the exciter
filter
Replacing the opto-coupler of the exciter filter
Open the camera head (see: Opening
and closing the camera head).
Disconnect the stepper motor cable and
the opto-coupler cable from the control
PCB and remove the cable mounts.
Screw out 2 fastening screws and remove
the complete module.
Remove the defective stepper motor or
opto-coupler and replace it by a new
component.
Re-insert the module and secure it in
position.
Re-route and connect the cables for the opto-coupler and the stepper motor.
Close the camera head (see: Opening and closing the camera head).
lite
Switch on VISUCAM and check its functions.
Barrier filter drive
Replacing the stepper motor of the barrier
filter
Replacing the opto-coupler of the barrier filter
Open the camera head (see: Opening
and closing the camera head).
Loosen the lamp housing and move it
slightly aside. (Also see: Replacing the
lamp housing).
Disconnect the stepper motor cable and
the opto-coupler cable from the control
PCB and remove the cable mounts.
Screw out 2 fastening screws and remove
the complete module.
Remove the defective stepper motor or
opto-coupler and replace it by a new
component.
Re-insert the module and secure it in position.
Re-route and connect the cables for the opto-coupler and the stepper
motor.
Install the lamp housing.
Close the camera head (see: Opening
and closing the camera head).
lite
Switch on VISUCAM and check its functions.

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

15

of 39

Finder camera
Replacing the finder camera
Open the camera head (see: Opening
and closing the camera head).
Unplug the connections of the finder
camera from the control PCB.
Remove the cable mounts.
Remove the angled bracket including the
finder camera.
Mount the new finder camera, connect the
cables and secure them in position.
Positioning the finder camera
Mount one of the two covers.
Move the bracket including the finder
camera in such a way that it rests slightly
against the inner side of the housing.
Tighten fastening screws.
Adjusting the focus of the finder camera
lite
Switch on VISUCAM and start the
program (also see the user manual).
Create a test patient file and start the
image capture mode.

The finder camera's field of view is


visible on the bottom left of the screen.
Attach a piece of paper displaying centering circles to
the headrest.
Adjust brightness as required for observation. Center
and focus the illuminated ring on the centering
circles. You can see then structures of fiber optic.

The rings must be displayed sharply defined. If the


rings are blurred, turn the focusing ring on the
finder camera until you have obtained a sharply
defined image.
This completes the mechanical alignment.
Close the camera head (see: Opening
and closing the camera head).

Focusing ring

Camera Field is max. 768 x 576 pics.


Maximum possible settings are
a+c 768
b+d 576
b (y)
Finder camera

d
a (x)

c
Illuminated ring

b > image
a > image
d > image
c > image

b < image
a < image
d < image
c < image

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

16

of 39

Checking and adjusting the image position of the finder camera


The rings should be centered on the field of view. The shadow (visible at the top of the picture)
should not be larger.
If you detect any deviations, the position of the rings can be adjusted in the x and y directions.
The height and width of the image can be adjusted as required.
Open the Windows Explorer (see: Opening the Windows Explorer).
lite
Click the VISUCAM software window to bring it to the front of the screen.
lite
Using the ALT + F4 key combination, close the VISUCAM program.

lite
Open the VISUCAM directory.
Open the VECtrl.ini file.
By changing the values of the marked positions (arrows), you can change both the position and
the size of an image.
Tip: changes in the image do not become visible until you change the values by 10 to 20.

lite
Start VISUCAM program by clicking the main.exe file.
Check the position of the finder camera picture.
If it is necessary, repeat the procedure again.

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

17

of 39

Replacing the complete lamp housing

Open the camera head (see: Opening


and closing the camera head).
Remove the light guide from the top.
lite
Disconnect 2 cables to VISUCAM
control PCB.
Remove various cable mounts on the lamp
housing.
Remove fastening screws.
Remove the lamp housing including PCB.
Install the new lamp housing. Observe the
guide slots inside both covers.
Connect the cables.
Secure the light guide in position at the
top.
Re-attach the mounts for the other cables.
Close the camera head (see: Opening
and closing the camera head).
lite
Switch on VISUCAM and check its functions.
X42
X45
X46

Securing screw
Light guide, top

Halogen lamp
Fan
Temperature

X41

X43

Signal lines to
X11 on
VISUCAM
control PCB

+12V lamp from X141


DEVICE X1
control cable

Camera module
Replacing the cameras
Open the camera head (see: Opening
and closing the camera head)
Remove the cable connections.
Screw out 3 fastening screws and remove
the camera module.
Insert the new module and align it with the
right-hand edge of the housing.
Make electrical connections.
Adjusting the cameras
Start the Service program.
Perform the electrical adjustment and
adjust the camera positions. (See the
description of the
lite
VISUCAM service program)

Close the camera head (see: Opening


and closing the camera head).
lite
Switch on VISUCAM and check its
functions.

X132

bottom of PCB
view

not
used

Device x1
control cable
X9

X131

X66
+12V B/W
X18 on VISUCAM
Control PCB

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

X11
B/W signals
X16 on VISUCAM
Control PCB

+12V color
X17 on
VISUCAM
Control PCB

Page

18

of 39

Replacing the VISUCAM

lite

control PCB

Open the camera head (see: Opening


and closing the camera head).
Unplug all cable connections.
Screw out 4 fastening screws and remove
the PCB.
lite
Insert the new VISUCAM control PCB
and secure it in position.
Re-connect the cables.
Close the camera head (see: Opening
and closing the camera head).
lite
Switch on VISUCAM and check its
functions.

Exciter filter
stepper motor
Exciter filter
sensor

Barrier filter
stepper motor
Barrier filter
sensor

Camera
stepper motor
Camera
sensor
Instrument base pot.

X14 (X114)
X7
X4

X8
X5

X11
X6

X50
X10

X12
(X112)

Finder camera
illumination
DEVICE
(X1 interface)
control cable

Instrument base release

X20

X2 (X102)
X1 (X101)

X43 Lamp control

X3

X9

X18

Programming
X17
X16

Black/white
camera pcb
x9

+12V color
camera pcb
x11

X18 +12V
b/w camera pcb
x66

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

19

of 39

Opening and closing the control base

Removing the camera head including


instrument base.
- Disconnect the cable for the camera
head.
- Loosen the screws in the toothed rack.
(2.5mm hex socket head)
- Push out the camera head including
instrument base to one side.
- Place the camera head including
instrument base on the table.

Opening the control base


- Put the base including headrest to one
side.
- Remove 8 x 3 mm hex socket head
screws.

- Remove 4 x Phillips screws from


the connector panel.
- Remove the base cover and lay it aside.

Caution: Take care not to tear off the fixation


lamp cable. (see also replacing the headrest!).

Close the base in reverse order.

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

20

of 39

Replacing the headrest

Remove the fixation lamp including the blinking


diode.
Loosen the locking screws for the headrest.
Open the base (see: Opening and closing the
control base).
Disconnect the cable for the fixation lamp from X7
on the instrument interface.
Pull the headrest from the clamping mounts.
Push the new headrest into the clamping mounts.
Secure the headrest in position by tightening the
locking screws.
Plug the cable for the fixation lamp into X7 on the
interface.
Close the base (see: Opening and closing the
control base).

Clamping mount

Cable of
fixation lamp

X7
fixation lamp

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

21

of 39

Components of the control base


Med PIII motherboard
GIF instrument interface
Electrical cutoff
20GB hard disk
CD-ROM drive
Line module
+5V, +/-12V power supply unit
Inductive approximation switch (sensor for right left identification)
12V axial fan

CD-ROM drive

20GB hard disk


Inductive approximation switch

Med PIII motherboard

+5V, +12V power


supply

GIF instrument interface

12V axial fan

Con. panel

Line module

Notes on the removal and the installation of the base components:


All interchangeable components can be removed and installed without difficulty.
Before the motherboard can be changed, the GIF instrument interface must be removed.
Replacing the hard disk
Save the patient data (if possible).
Note:
The saving of patient data is the responsibility of the customer!
Both a hard disk with an operating system and a VISUPAC program and a blank hard
disk are available.

Open the base (see: Opening and closing the control base).
Remove the defective hard disk.
Install the new hard disk.
Close the base.
Restore the operating system using the recovery CD-ROM.
Load patient data provided.

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

22

of 39

Connections on the GIF instrument interface

X1

Camera head

Fixation lamp, headrest


Motherboard

X21

X7
X22

Fan, base

X6

Motherboard

X2
X4

X3

X5

Sensor, right-left
movement

Power supply

Connections on the motherboard

Battery

Note: Do not fix the motherboard with this screw!

Hard disk
CD-ROM

Power
supply

User interface
X21

VGA display
con. panel

User interface
X22

TV OUT
VBS, VIDEO

Connections on the power supply unit

Line IN

Red +5V
Black Ground
Yellow +12V

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

23

of 39

System settings and work with internal programs


Note: The system settings are only accessible to authorized service staff and may only be
changed by this staff. The access to the system settings is protected by a password.
All system settings have been adjusted ex factory as required in the country of
destination.
Activating the system settings

VISUPAC lite is on and in the patient or image


capture mode.
lite
Switch on VISUCAM .
Click File in the program bar.
The overview window opens.

Select and click System.

The password window opens.


Enter password iris and click OK.
Note: This Password should be used by
introduced person.

The system selection window opens.


Select and click on the function required.

Mouse
3 menus are available for defining the settings for the mouse.

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

24

of 39

Keyboard
2 menus are available for defining the settings for the keyboard.
In the Speed menu, you can change the mode of operation.
In the Language menu, you can adapt the regional settings to the keyboard used.

Printer
Via this selection, you can
access the printer menu. Here,
a printer can be selected as a
Default Printer. You can also
install new printers here.

Selecting printer features


Click File and then
Properties to open the
settings window for the
printer selected.
You can find the data on the
driver and the port
assignment (e.g. USB) in
the Details register.
The changes you will make in
the properties are archived.
Note:

The driver for


installing the Deskjet
940c is included in
the path:
c:\ProgramFiles
Hewlett-Packard
HPZ
Glue

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

25

of 39

Network
In the Network menu, you can adjust the settings for the network connection.
The Configuration window shows the protocols used.
In the Identification window, you can enter the computer name assigned by the network
administrator (e. g. ZO01Q73J) and the workgroup or domain used (e. g. DEOKO). You can
lite
also enter a name for the unit connected (e.g. VISUCAM ) in the Computer Description box.

If a special IP address must be used, open the TCP/IP protocol and select Properties. You can
specify an IP address in he IP Address box.

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

26

of 39

Logon to a domain it can be setup in network configuration window.

Click Client for Microsoft


Networks.
Click Properties.

Properties General
opens.
Mark Logon to Windows
NT domain.
Type in domain name (in
this case zeiss).
Select Network logon
options.
Finish with OK.

If you have selected the


logon to a domain, after
restart the system asks you
for your login data of the
domain server (zeiss). This
data you get from the domain
administrator.

Note:
If the program asks for new
drivers you can find drivers in
windows directory
c:\WIN98SE\WinInst.

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

27

of 39

Date / Time
In the Date/Time Properties menu you can
change the time currently set.

The time zone can be adapted to the


current location.

Regional settings
In the Regional Settings menu, you can set
the language zones.

In the Date menu you can set the date.

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

28

of 39

Opening the Windows Explorer program

VISUPAC lite is on and in the patient or image


capture mode.
Click the ? in the program bar.
The window with Help Topics and About

opens .
Click About

The About windows opens.


This window contains important data on
lite
VISUCAM including the serial number.
Press the ??? +? key combination.
Note: This key combination is only available to
service staff and is provided during service
training.

The Windows Explorer opens,


providing access to the system settings.

Closing an active window


Use the ALT + F4 key combination to close an active window.
Opening watch files
Use the ALT + w key combination to open newest watch file.
Enter BIOS
To enter BIOS, press F2 several times during switch on the instrument.
Password for BIOS
The Password is ?????
Installing, uninstalling CD-writer
Install / uninstall new drivers for CD-writer (Plexwriter)
Uninstall: 1. IN CD program
Install:
2. Driver for the CD-writer
Note:

1. Driver for the CD-writer


2. IN CD program

If it is not possible to read any data from a CD-RW, you have to install the
UDF- Reader program on your computer to read files from CD-RW.

Drivers are local at c:\ installs


Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

29

of 39

Opening the Test Tools program

VISUPAC lite is on and in the patient or image


capture mode.
Click the ? in the program bar.
The window with Help Topics and About

opens .
Click About

The About windows opens.


Press the ??? + ? key combination.
Note: This key combination is only available to
service staff and is provided during service
training.

The Test Tools selector window opens,


providing access to different modules of the
Test Tool.
Note: Service module is designed for service
use. Other modules are not necessary to
use.

Click Service

The Test Tool windows opens.

Overview Test Tools program


There are 3 windows to check VISUCAM
in / out window
checks the Input Output
functions

lite

functions or setup data:

stepper motors window


stepper motor of exciter filter
stepper motor of barrier filter
stepper motor of camera
selctor

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Data window
shows or setup data
factory settings

Page

30

of 39

Service work
Software
Database

The Database folder contains the Visucam.mdb database,


the Image folder, the Sequences folder and the TNI`s folder.
Database Images

The Image folder contains all Images (b/w and color) in BMP format.
Database Sequences

The Sequences folder contains all Video clips in AVI format.


Database TNIs

The TNI`s folder contains all small Thumbnails (pictures in overview mode) in JPG format.

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

31

of 39

VISUCAM Save and Update folder

In the Save folder the files Main.ini and visucam.mdb are saved as the last correct files during
start up.
In the Update folder there are temporarily data during software update. Normally empty!
VISUCAM Installs folder
A lot of different drivers are placed in this installs folder.

VISUCAM HP Driver folder

In the Hawlett-Packard/HPZ/Glue folder, the install driver of the HP Deskjet 940c are placed.
See also printer installation.
Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

32

of 39

In and Outputs of the


electronic
Lamp control
2 -255
Flash control
Temperature sensor
20C38, 50C85, 60C-105
Hardware Version
(reverence voltage) 128 1
Voltage +5V, +12V
+12V 200 5, 5%
+ 5V 200 5, 5%
Finder camera illumination
Fan on/ off
Port A Inputs
0 - fire button
1 barrier filter sensor
2 exciter filter sensor
3 camera sensor

Data setup
Serial number of camera
Temperature limits
Sensor positions
Default settings

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

33

of 39

Stepper Motor control exciter filter

Stepper Motor control barrier filter

Stepper Motor control


camera shift
Overflow position error for
B/W Camera is <5 and >50
between the light barrier and
B/W sensor.
Switch the color sensor on:
Click color button
Click +1 button
Click 1 button
The color sensor is on now.
Positions of sensors:
Light barrier
B/W
color
sensor
sensor
+1 step, Image moves
- 1 step, Image moves
Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

34

of 39

Testing focus drive

Focus the illumination ring to the first area by moving


lite
the VISUCAM camera by hand. You
should see the structure off the light guide
inside the illuminated ring.
Focus with focus drive to the area with the
vertical lines and expect a sharp picture at
the screen.

Note: The distance between the front area


and the back area is 40mm and it
should be possible to focus in this area.

Testing grabber function and quality of the sensors


Use the VISUCAM lite in a dimmed area and
switch off the room light.
Put the gray filter in between the light guide and
the halogen lamp.
Use capture mode.
Focus to the gray field of the color table.
Setup Angle 45
Fluo mode, barrier and exciter filter
b/w sensor
Select maximum brightness of the halogen lamp
Inspect the image on screen. No lines only
snow or noise should be visible.
Snap the image and save it to a test patient to
compare it later time.

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Grey filter

Page

35

of 39

Measuring the operating voltages


+12V
+12V
0V
0V
+5V

Input voltage of the power supply unit (x4)


Voltage supplied to the instrument interface

+5V
+5V
+5V
0V
0V
0V
+12V
+12V
0V

0V
0V
+5V
+5V

Voltage supplied to the motherboard


Output voltage of the power supply unit
+5V (red)
+12V (yellow)
0V (black)
Voltages present at Interface GIF PCB

C13
X21

X4 Sensor left right


movement

U1

X22

1
+5V

2
0V

3
4
P22 +12V

1
+5V

2
0V

3
4
P23 +12V

C2

D2

R2

X1

C3

R3
C4

R18

L5

R1

X4

44
R12

R11

X3

D3

R13

Display
left
OS
right OD

C6
C8
C7
C5

R5*

R8
R10
R14

Sensor P22 pin 3


+5.2V without metal
+0.8V with metal and LED on
at the sensor

R4

C9

X2

C11

X7 C14

X6

C12
R7

C10
X8*

R17

X5
1

X7 Fixation lamp

R6

2
0V

L6

1
+5V

+5V logic
0V
0V
+12V lamp
+12V cameras

X6 Fan
X5 Voltage supply
camera head

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

2
+7V

Page

1
0V

36

of 39

Voltages present at the VISUCAM


X4
X5
X6
X7
X8
X9

1
+5V+R

2
GND

3
IN

4
GND

lite

control PCB

5
+5V

1
2
3
4
OUT 2A OUT 2B OUT 1A OUT 1B

1
2
3
+5V1 Haloout GND

X11

1
GND

X14

2
OUT

3
GND

+ in -

GIF

1
REL

X2

1
ADR0

Finder cam.

2
3
4
GND Videoin VidGND

2
3
4
HPBeo HPBel HPSen

2
ADR1

7
Luftout

8
9
GND Tempin

10
GND

5
frei

Lamp control

TMS
TDI
TDO
TCK
GND
+5V

GIF

X1

6
GND

Base: Switch

Excitercoup.Camerascoup.

1
+5V

5
GND

Base: Poti
+ in -

GIF: ADU

X10

4
+5V1

4
GND

3
ADR2

4
ADR3

Voltages present at the VISUCAM

++

Control
G I F 12 V
SW-Camera
12V
Cameras

7
+12V

8
OUT

5
frei

6
Dout

9
+5V

5
BIT0

6
BIT1

7
BIT2

8
BIT3

9
GND

lite

X16

10

11

12V ColorCamera

1
2
3
4
Gamma DGND FieldFr DGND

12

5
Gain

6
AGND

13
GND

10
11
12
VIDIN VIDOUT +12V

lamp Control PCB

0V 0V 0V +12V +12V

X41

X43

Checking the electrical functions


Check electrical function by using the Test Tools program.
Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

37

of 39

System check
Mechanical check
Check that the camera head can be swiveled all the way from the left to the right. ........... 
lite
Check that VISUCAM fundus camera be easily moved and securely locked on the base
within the entire working range .......................................................................................... 
Check that the ophthalmoscope lens is clean .................................................................... 
Check that the connectors are firmly seated. ..................................................................... 
Check that the built-in filters and optics are clean. ............................................................. 
Check all cables for any damage or traces of wear............................................................ 
Check that all labels and lettering are legible and complete............................................... 
Check the glide plate for any traces of wear.. .................................................................... 
Electrical function check
Check that all electrical connections have been made and securely mounted. .................. 
lite
After switching on VISUCAM , check the booting process of the computer. ................... 
Listen for any unusual noise made by the fan and the hard disk........................................ 
Check whether the operating software is starting properly................................................. 
Check whether the exciter filter drive is initialized. ............................................................. 
Check whether the barrier filter drive is initialized. ............................................................. 
Check the brightness control of the halogen lamp and its remaining service life................ 
Check the illumination for the finder camera. ..................................................................... 
Check the proper functioning of the finder camera and the image produced by it on the
monitor............................................................................................................................... 
Check that the cameras for black/white and color images can be selected ....................... 
Check that the focusing system is functioning properly and moving easily. ....................... 
Check whether images can be captured using the black/white camera and whether the
images can be saved ......................................................................................................... 
Check whether images can be captured using the color camera and whether the
images can be saved. ........................................................................................................ 
Check whether archiving on an external unit is possible (optional). ................................... 
Check that you can print out the images with the printer connected (optional)................... 
Check the screen settings (contrast, brightness, resolution) ............................................. 

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

38

of 39

Adjustment Circle
This Circle can be used for finder camera adjustment.

Service insructions

Issue: 08.07.2003

SM-30-4043-A1-en

Replaces issue: 20.09.2002

Page

39

of 39