Proline Promag 55S


Electromagnetic Flow Measuring System
Flow rate measurement of liquids with solids content or
inhomogeneous liquids

Application Approvals for hazardous area:


Electromagnetic flowmeter for bidirectional meas- • ATEX, FM, CSA
urement of liquids with a minimum conductivity
Connection to process control system:
of ≥ 5 μS/cm – in particular fluids with solids, and fluids
• HART, PROFIBUS PA, FOUNDATION Fieldbus
which are abrasive, inhomogeneous or tend to build-up,
for example: Your benefits
Promag measuring devices offer you cost-effective flow
• Chemical/mechanical pulps, paper pulp or wood pulp
measurement with a high degree of accuracy for a wide
with solids contents up to 15 Vol.-%
range of process conditions.
• Fruit mashes, fruit concentrates and final products
(salad dressings, soups with vegetable pieces) The Proline transmitter concept comprises of:
• Slurries containing high amounts of sand or stone with • High degree of efficiency due to the modular device
an abrasive effect, e.g. ore slurry or mortar and operating concept
• Chemically inhomogeneous fluids (e.g. additives) • Software options for: electrode cleaning, advanced
• Thick wastewater sludges diagnostics, calculation of mass flow and solids content
• Flow measurement up to 9600 m3/h (42267 gal/min) The robust Promag S sensors offer:
• Can be used up to +180 °C (+356 °F) • Universal devices, even for difficult fluids
and max. 40 bar (580 psi) • Excellent accuracy and repeatability
• Fitting lengths as per DVGW/ISO • High resistance to abrasion thanks to industry-
optimized linings and measuring electrodes
Application-specific linings and electrodes:
• Optimum operational safety due to advanced,
• Natural rubber, hard rubber, polyurethane, PTFE or
permanent self-diagnosis
PFA linings
• Simple installation and commissioning
• Flat, bullet nose, neck, bow or brush electrodes
• Insensitive to vibration
• No pressure loss

TI071D/06/ae/12.07
71067186
Proline Promag 55S

Table of contents

Function and system design. . . . . . . . . . . . . . . . . . . . . 3 Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42


Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Material load diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Fitted electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Process connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Surface roughness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Human interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Operable flow range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Display elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Input signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Operating elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Language groups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Language groups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Certificates and approvals . . . . . . . . . . . . . . . . . . . . . 45
Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Low flow cutoff . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 C-tick mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Galvanic isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Ex certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Switching output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Sanitary compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Pressure measuring device approval . . . . . . . . . . . . . . . . . . . . . . 45
FOUNDATION Fieldbus certification . . . . . . . . . . . . . . . . . . . . . 45
Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 PROFIBUS PA certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Electrical connection measuring unit . . . . . . . . . . . . . . . . . . . . . . . 6 Other standards, guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Electrical connection terminal assignment . . . . . . . . . . . . . . . . . . . 7
Electrical connection remote version . . . . . . . . . . . . . . . . . . . . . . . 7
Electrical connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Cable entries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Cable specifications (remote version) . . . . . . . . . . . . . . . . . . . . . . 8 Supplementary documentation . . . . . . . . . . . . . . . . . 46
Power consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Power supply failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Potential equalization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Registered trademarks . . . . . . . . . . . . . . . . . . . . . . . . 46

Performance characteristics. . . . . . . . . . . . . . . . . . . . 13 Ordering information. . . . . . . . . . . . . . . . . . . . . . . . . 47


Reference conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Promag 55S . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Maximum measured error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Repeatability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

Operating conditions: Installation . . . . . . . . . . . . . . . 14


Installation instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

Operating conditions: Environment. . . . . . . . . . . . . . 20


Ambient temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Shock and vibration resistance . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Electromagnetic compatibility (EMC) . . . . . . . . . . . . . . . . . . . . . 20

Operating conditions: Process . . . . . . . . . . . . . . . . . . 21


Medium temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Conductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Limiting medium pressure range (nominal pressure) . . . . . . . . . . 22
Pressure tightness (lining) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Nominal diameter and flow rate . . . . . . . . . . . . . . . . . . . . . . . . . 24
Pressure loss . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Measuring tube specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

Mechanical construction . . . . . . . . . . . . . . . . . . . . . . 30
Design, dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

2 Endress+Hauser
Proline Promag 55S

Function and system design


Measuring principle Faraday’s law of induction states that a voltage is induced in a conductor moving in a magnetic field.
In electromagnetic measuring, the flowing medium corresponds to the moving conductor. The induced voltage
is proportional to the flow velocity and is detected by two measuring electrodes and transmitted to the ampli-
fier. Flow volume is computed on the basis of the pipe’s diameter. The constant magnetic field is generated by
a switched direct current of alternating polarity.

Ue

V
B
I
L

A0003191

Ue = B · L · v
Q=A·v
Ue induced voltage
B magnetic induction (magnetic field)
L electrode gap
v flow velocity
Q volume flow
A pipe cross-section
I current strength

Measuring system The flow measuring system consists of the following components:
• Promag 55 transmitter
• Promag S sensor (DN 15 to 600 / ½ to 24")
Two versions are available:
• Compact version: Transmitter and sensor form a single mechanical unit.
• Remote version: Transmitter and sensor are installed separately.

Measurement of solids flow rates


In combination with a density meter, e.g. with "Gammapilot M" from Endress+Hauser, Promag 55S also
determines the throughput of solids in mass, volume or percentage rates.
The following order specifications are required for this: order option for software function "Solids content flow"
(F-CHIP) and order option for a current input.

rM

2
rC m
rS

1 V

A0006118

Solids content flow measurement (m) with the aid of a density and flow measuring device. If the solid density (ρS ) and the
density of the carrier liquid (ρC ) are also known, they can be used to calculate the solids flow.
1 Flow measuring device (Promag 55S) → volume flow (V). The solid density (ρS ) and the density of the transport
liquid (ρC ) must be entered in the transmitter.
2 Density measuring device (e.g. "Gammapilot M") → total fluid density (ρM ) (transport liquid and solids)

Endress+Hauser 3
Proline Promag 55S

Input
Measured variable • Flow rate (proportional to induced voltage)
• Conductivity (without temperature compensation)

Measuring range • Flow rate: Typical v = 0.01 to 10 m/s (0.03 to 33 ft/s) with the specified measuring accuracy
• Conductivity σ = 5 to 2000 μ/cm
not for sensors without reference electrode (Promag S with brush electrodes)

Operable flow range Over 1000 : 1

Input signal Status input (auxiliary input):


U = 3…30 V DC, Ri = 5 kΩ, galvanically isolated
Configurable for: totalizer(s) reset, positive zero return, error-message reset

Current input:
active/passive selectable, galvanically isolated, full scale value adjustable, resolution: 3 μA,
temperature coefficient: typically 0.005% o.f.s./°C; (0.003% o.f.s./°F)
• active: 4 to 20 mA, Ri ≤ 150 Ω, Uout = 24 V DC, short-circuit proof
• passive: 0/4 to 20 mA, Ri ≤ 150 Ω, Umax = 30 V DC

Output
Output signal Current output:
active/passive selectable, galvanically isolated, time constant selectable (0.01 to 100 s),
full scale value adjustable, temperature coefficient: typically 0.005% o.f.s./°C (0.003 % o.f.s./°F),
resolution: 0.5 μA
• active: 0/4 to 20 mA, RL < 700 Ω (for HART: RL ≥ 250 Ω)
• passive: 4 to 20 mA; supply voltage VS : 18 to 30 V DC; Ri ≥ 150 Ω

Pulse/frequency output:
active/passive selectable (Ex i version passive only), galvanically isolated
• active: 24 V DC, 25 mA (max. 250 mA over 20 ms), RL > 100 Ω
• passive: open collector, 30 V DC, 250 mA
• Frequency output: end frequency 2 to 10000 Hz (fmax = 12500 Hz), on/off ratio 1:1, pulse width max. 10 s
• Pulse output: pulse value and pulse polarity selectable, pulse width configurable (0.05 to 2000 ms)

PROFIBUS PA interface:
• Transmission technology (Physical Layer): IEC 61158-2 (MBP), galvanically isolated
• Profile version 3.0
• Current consumption: 11 mA
• Permissible supply voltage: 9 to 32 V
• Bus connection with integrated reverse polarity protection
• Error current FDE (Fault Disconnection Electronic): 0 mA
• Function blocks: 2 x analog input, 3 x totalizer
• Output data: volume flow, calculated mass flow, totalizer 1 to 3
• Input data: positive zero return (ON/OFF), totalizer control, value for local display
• Cyclic data transmission compatible with previous model Promag 35S
• Bus address adjustable via miniature switches or local display (optional) at the measuring device

4 Endress+Hauser
Proline Promag 55S

FOUNDATION Fieldbus interface:


– FOUNDATION Fieldbus H1
– Transmission technology (Physical Layer): IEC 61158-2 (MBP), galvanically isolated
– ITK version 5.0
– Current consumption: 12 mA
– Inrush current: <12 mA
– Error current FDE (Fault Disconnection Electronic): 0 mA
– Permissible supply voltage: 9 to 32 V
– Bus connection with integrated reverse polarity protection
– Function blocks:
– 5 x Analog Input (execution time: 20 ms each)
– 1 x PID (50 ms)
– 1 x Descrete Output (20 ms)
– 1 x Arithmetic (20 ms)
– 1 x Signal Characterizer (20 ms)
– 1 x Input Selector (20 ms)
– 1 x Integrator (25 ms)
– Total VCRs: 48
– Total link objects in VFD: 40
– Output data: volume flow, calculated mass flow, temperature, totalizer 1 to 3
– Input data: positive zero return (ON/OFF), reset totalizer
– Link Master (LM) functionality is supported

Signal on alarm Current output:


Failsafe mode selectable (e.g. according to NAMUR recommendation NE 43)
Pulse/frequency output:
Failsafe mode selectable
Relay output:
"de-energized" in the event of a fault or power supply failure

Load See "output signal"

Low flow cutoff Switch points for low flow cut off freely selectable.

Galvanic isolation All circuits for inputs, outputs, and power supply are galvanically isolated from each other.

Switching output Relay output:


Normally closed (NC or break) or normally open (NO or make) contacts available
(default: relay 1 = NO, relay 2 = NC),
max. 30 V / 0.5 A AC; 60 V / 0.1 A DC, galvanically isolated.
Configurable for: error messages, empty pipe detection (EPD), direction of flow, limit values.

Endress+Hauser 5
Proline Promag 55S

Power supply
Electrical connection
measuring unit
A B

a
a
b d (d)

HART PROFIBUS PA*


FOUNDATION Fieldbus*

– 27 PA – / FF – 27
+ 26 PA + / FF + 26
– 25 d – 25 d
+ 24 + 24
f – 23 e f – 23 e
+ 22 + 22
– 21 – 21
+ 20 + 20

N (L-) 2 c N (L-) 2 c
L1 (L+)1 L1 (L+) 1
b b

A0006187

Connecting the transmitter, cable cross-section max. 2.5 mm 2


A View A (field housing)
B View B (wall-mount housing)
*) Fixed communication boards
a Connection compartment cover
b Cable for power supply: 20 to 260 V AC / 20 to 64 V DC
Terminal No. 1: L1 for AC, L+ for DC
Terminal No. 2: N for AC, L– for DC
c Ground terminal for protective conductor
d Signal cable: see "Electrical connection terminal assignment"
Fieldbus cable:
Terminal No. 26: PA + / FF + (with polarity protection)
Terminal No. 27: PA – / FF – (with polarity protection)
e Ground terminal for signal cable shield / Fieldbus cable
f Service adapter for connecting service interface FXA193 (Fieldcheck, FieldCare)

6 Endress+Hauser
Proline Promag 55S

Electrical connection Terminal No. (inputs / outputs


terminal assignment Order version
20 (+) / 21 (–) 22 (+) / 23 (–) 24 (+) / 25 (–) 26 (+) / 27 (–)

Fixed communication boards (fixed assignment)

Current output
55***-***********A - - Frequency output
HART

Current output
55***-***********B Relay output 2 Relay output 1 Frequency output
HART

PROFIBUS PA
55***-***********H

FOUNDATION
55***-***********K
Fieldbus

Flexible communication boards

Current output
55***-***********C Relay output 2 Relay output 1 Frequency output
HART

Current output
55***-***********D Status input Relay output Frequency output
HART

Current output
55***-***********L Status input Relay output 2 Relay output 1
HART

Current output
55***-***********M Status input Frequency output 2 Frequency output 1
HART

Current output 1
55***-***********2 Relay output Current output 2 Frequency output
HART

Current output 1
55***-***********3 Current input Current output 2 Frequency output
HART
Current output
55***-***********4 Current input Relay output Frequency output
HART

Current output
55***-***********5 Status input Current input Frequency output
HART

Electrical connection
remote version Electrode circuit Coil circuit
Meas.signal Pipe EPD

S1 E1 E2 S2GND E S

a 6 5 7 8 4 37 36 42 41

c d
2
1

n.c. n.c. n.c.

5 7 4 37 42 41

E1 E2 GND E

A0004323-en

Connecting the remote version


a Connection compartment, wall-mount housing
b Cover of connection housing, sensor
c Signal cable
d Coil current cable
n.c. unconnected, insulated cable shields
Terminal no. and cable colors: 6/5 = brown; 7/8 = white; 4 = green; 36/37 = yellow

Endress+Hauser 7
Proline Promag 55S

Electrical connections 20 to 260 V AC, 45 to 65 Hz


20 to 64 V DC

Cable entries Power-supply and signal cables (inputs/outputs):


• Cable gland M20 x 1.5 (8 to 12 mm / 0.31 to 0.47 inch)
• Cable entries for thread ½" NPT, G ½"
Fieldbus cable:
• Fieldbus connector for PROFIBUS PA, M12 x 1 / PG 13.5 plus adapter PG 13.5 / M20.5
• Fieldbus connector for FOUNDATION Fieldbus, 7/8-16 UNC x M20
Connecting cable for remote version:
• Cable gland M20 x 1.5 (8 to 12 mm / 0.31 to 0.47 inch)
• Cable entries for thread ½" NPT, G ½"

Cable specifications Coil cable


(remote version) • 2 x 0.75 mm2 (18 AWG) PVC cable with common, braided copper shield (∅ ∼ 7 mm / 0.28")
• Conductor resistance: ≤37 Ω/km (≤0.011 Ω/ft)
• Capacitance: core/core, shield grounded: ≤120 pF/m (≤37 pF/ft)
• Operating temperature:
– Cable not permanently routed: –20 to +80 °C (–4 to +176 °F)
– Cable permanently routed: –40 to +80 °C (–40 to +176 °F)
• Cable cross-section: max. 2.5 mm2 (14 AWG)
Signal cable
• 3 x 0.38 mm2 (20 AWG) PVC cable with common, braided copper shield (∅ ∼ 7 mm / 0.28") and indivi-
dually shielded cores
• With Empty Pipe Detection (EPD): 4 x 0.38 mm2 (20 AWG) PVC cable with common, braided copper shield
(∅ ∼ 7 mm / 0.28") and individually shielded cores
• Conductor resistance: ≤50 Ω/km (≤0.015 Ω/ft)
• Capacitance: core/shield: ≤420 pF/m (≤128 pF/ft)
• Operating temperature:
– Cable not permanently routed: –20 to +80 °C (–4 to +176 °F)
– Cable permanently routed: –40 to +80 °C (–40 to +176 °F)
• Cable cross-section: max. 2.5 mm2 (14 AWG)

1
2
3
4
5
6
7
a b
A0003194

a Signal cable
b Coil current cable
1 Core
2 Core insulation
3 Core shield
4 Core jacket
5 Core reinforcement
6 Cable shield
7 Outer jacket

As an option, Endress+Hauser can also deliver reinforced connecting cables with an additional, reinforcing
metal braid. We recommend such cables for the following cases:
• Directly buried cable
• Cables endangered by rodents
• Device operation which should comply with the IP 68 (NEMA 6P) standard of protection

8 Endress+Hauser
Proline Promag 55S

Operation in zones of severe electrical interference


The measuring device complies with the general safety requirements in accordance with EN 61010-1,
the EMC requirements of IEC/EN 61326 and NAMUR recommendation NE 21.

" Caution!
Grounding of the shield is by means of the ground terminals provided for the purpose inside the connection
housing. Keep the stripped and twisted lengths of cable shield to the terminals as short as possible.

Power consumption AC: <45 VA at 260 V AC; <32 VA at 110 V AC (incl. sensor)
DC: <19 W (including sensor)
Switch-on current:
• max. 2.00 A (<700 ms) at 20 V AC
• max. 2.28 A (<5 ms) at 110 V AC
• max. 5.5 A (<5 ms) at 260 V AC

Power supply failure Lasting min. 1 power cycle:


• EEPROM or HistoROM/T-DAT saves measuring system data if power supply fails
• HistoROM/S-DAT: exchangeable data storage device which stores sensor characteristic data (nominal
diameter, serial number, calibration factor, zero point etc.)

Potential equalization Standard case


Perfect measurement is only ensured when the medium and the sensor have the same electrical potential. Most
Promag sensors have a reference electrode installed as standard, which guarantees the required potential equa-
lization. This usually means that additional potential equalization measures are unnecessary.
Promag S:
• Reference electrode is standard for electrode materials 1.4435/316L, Alloy C-22, tantalum, Duplex,
tungsten carbide coating (for electrodes made of 1.4435)
• Reference electrode is optional for electrode material platinum/rhodium 80/20
• Reference electrode not present in measuring tubes with natural rubber lining and brush electrodes

! Note!
For installation in metal pipes, it is advisable to connect the ground terminal of the transmitter housing to the
piping. Also, observe company-internal grounding guidelines.

A0004375

Potential equalization by means of the transmitter's ground terminal

" Caution!
• For sensors without reference electrodes or without metal process connections, carry out potential
equalization as per the instructions for special cases described below. These special measures are particularly
important when standard grounding practice cannot be ensured or extremely strong matching currents are
expected.
• As sensors with a natural rubber lining do not have a reference electrode, ground disks must be installed if
necessary to ensure sufficient potential equalization to the fluid. This applies in particular to ungrounded
metal pipes → Page 10.

Endress+Hauser 9
Proline Promag 55S

Special cases

Metal, ungrounded piping


In order to prevent outside influences on measurement, it is necessary to use ground cables to connect each
sensor flange to its corresponding pipe flange and ground the flanges. Connect the transmitter or sensor con-
nection housing, as applicable, to ground potential by means of the ground terminal provided for the purpose
(see diagram).
The ground cable for flange-to-flange connections can be ordered separately as an accessory from
Endress+Hauser → Page 46.
• DN ≤ 300 (12"): The ground cable is in direct connection with the conductive flange coating and is secured
by the flange screws (A).
• DN ≥ 350 (14"): The ground cable connects directly to the metal transport bracket (B).

" Caution!
Also, observe company-internal grounding guidelines.

A B

A0004376

Potential equalization with equalizing currents in ungrounded, metal pipes


(ground cable: copper wire, at least 6 mm 2/10 AWG)
A Installing ground cable at DN ≤ 300 (12")
B Installing ground cable at DN ≥ 350 (14")

10 Endress+Hauser
Proline Promag 55S

Pre-installed ground cable for DN ≤ 300/12" (order option)


Ground cables which are preinstalled on the sensor flange, are also available. These ground cables can be
mounted and connected electrically to the piping in different ways:
• Using a screw on the side of the pipe flange (a)
• Using the flange screws (b)
• Using a pipe clip installed around the pipe (c)

c a

A006117

Possibilities for connecting and mounting pre-installed ground cables


(ground cable: copper wire at least 6 mm 2/10 AWG)

Plastic pipes and isolating lined pipes


Normally, potential is matched using the reference electrodes in the measuring tube. However, in exceptional
cases it is possible that, due to the grounding plan of a system, matching currents flow over the reference elec-
trodes. This can lead to destruction of the sensor, e.g. through electro-chemical decomposition of the electro-
des. In such cases, e.g. for fiberglass or PVC pipings, it is therefore essential that you use additional ground
disks for potential equalization. This applies also to two-phase or two-component flow, where the fluid is not
well mixed or its constituents are not mixable.

" Caution!
• Risk of damage by electrochemical corrosion. Note the electrochemical insulation rating, if the ground disks
and measuring electrodes are made of different materials.
• Also, observe company-internal grounding guidelines.

A0004377

Potential equalization/ground disks in the case of plastic pipes or isolating lined pipes
(ground cable: copper wire at least 6 mm 2/10 AWG)

Endress+Hauser 11
Proline Promag 55S

Plastic pipes and isolating lined pipes


In such cases, install the measuring instrument without potential in the piping:
• When installing the measuring device, make sure that there is an electrical connection between the two
piping runs (copper wire, at least 6 mm2/10 AWG).
• When using ground disks in plastic or isolating lined pipes, ensure that they are electrically connected with
each other (copper wire at least 6 mm2/10 AWG).
• Make sure that the mounting material used does not establish a conductive connection between the pipe
and the measuring device and that the mounting material withstands the torques applied when the threaded
fasteners are tightened during installation.
• Check the galvanic isolation using an insulation tester (protection against contact).
• Also comply with the regulations applicable to potential-free installation.

! Note!
For the remote version, both the sensor and the transmitter must be installed so that they are potential-free.

2 2
1

A0004378

Potential equalization and cathodic protection (connecting cable: copper wire at least 6 mm2/10 AWG)
1 Isolation transformer power supply
2 Electrically insulated

12 Endress+Hauser
Proline Promag 55S

Performance characteristics
Reference conditions To DIN EN 29104 and VDI/VDE 2641:
• Fluid temperature: +28 °C ± 2 K
• Ambient temperature: +22 °C ± 2 K
• Warm-up time: 30 minutes
Installation:
• Inlet run >10 x DN
• Outlet run > 5 x DN
• Sensor and transmitter grounded.
• Sensor centered relative to the pipe.

Maximum measured error Pulse output:


• Standard: ±0.2% o.r. ± 2 mm/s (o.r. = of reading)
• With brush electrodes (Option): ±0.5% o.r. ± 2 mm/s (o.r. = of reading)
Current output:
in addition typically ± 5 μA

! Note!
Supply-voltage fluctuations have no effect within the specified range.

[%]
2.5

2.0

1.5

1.0

0.5

0
0 1 2 4 6 8 10 [m/s]
v
0 5 10 15 20 25 30 32 [ft]

A0004456-ae

Max. measured error in % of reading

Conductivity
• Max. measuring error not specified
• Without temperature compensation (cell constant is a factory setting)

Repeatability Volume flow


• Standard: max. ±0.1% o.r. ± 0.5 mm/s (o.r. = of reading)
• With brush electrodes (Option): max. ±0.2% o.r. ± 0.5 mm/s (o.r. = of reading)

Conductivity
• Max. ±5% o.r. (o.r. = of reading)

Endress+Hauser 13
Proline Promag 55S

Operating conditions: Installation

Installation instructions Location


The accumulation of air or gas bubbles in the measuring tube could result in an increase in measuring errors.
Avoid the following locations:
• At the highest point of a pipeline. Risk of air accumulating.
• Directly upstream from a free pipe outlet in a vertical pipeline.

h ³ 2 x DN

A0003202

Installing pumps
Do not install the sensor on the intake side of a pump. This precaution is to avoid low pressure and the conse-
quent risk of damage to the lining of the measuring tube. Information on the lining's resistance to partial
vacuum → Page 22.
It might be necessary to install pulse dampers in systems incorporating reciprocating, diaphragm or peristaltic
pumps. Information on the measuring system's resistance to vibration and shock → Page 20.

A0003203

14 Endress+Hauser
Proline Promag 55S

Partially filled pipes


Partially filled pipes with gradients necessitate a drain-type configuration. The Empty Pipe Detection function
offers additional protection by detecting empty or partially filled pipes.

" Caution!
Risk of solids accumulating. Do not install the sensor at the lowest point in the drain. It is advisable to install a
cleaning valve.

³ 2 x DN

³ 5 x DN

A0003204

Down pipes
Install a siphon or a vent valve downstream of the sensor in down pipes longer than 5 meters (16 ft). This pre-
caution is to avoid low pressure and the consequent risk of damage to the lining of the measuring tube. This
measure also prevents the system losing prime, which could cause air inclusions.
Information on the lining's resistance to partial vacuum → Page 22.

2
h

A0003205

Measures for installation in a down pipe (h > 5 m/16 ft)


1 Vent valve
2 Siphon

Endress+Hauser 15
Proline Promag 55S

Orientation
An optimum orientation position helps avoid gas and air accumulations and deposits in the measuring tube.
Promag, nevertheless, supplies a range of functions and accessories for correct measuring of problematic fluids:
• Electrode Cleaning Circuit (ECC) for applications with fluids producing build-up, e.g. electrically conductive
deposits → "Description of Device Functions" manual.
• Empty Pipe Detection (EPD) ensures the detection of partially filled measuring tubes, e.g. in the case of
degassing fluids or varying process pressures

Vertical orientation
A vertical orientation is ideal in the following cases:
• For self-emptying piping systems and when using empty pipe detection.
• For sludge containing sand or stones and where the solids cause sedimentation.

A0003206

Horizontal orientation
The measuring electrode plane should be horizontal. This prevents brief insulation of the two electrodes by
entrained air bubbles.

" Caution!
Empty Pipe Detection functions correctly with the measuring device installed horizontally only when the
transmitter housing is facing upward (see diagram). Otherwise there is no guarantee that Empty Pipe Detection
will respond if the measuring tube is only partially filled or empty.

A
1

2 2

A
3

A0003207

1 EPD electrode for empty pipe detection (not for lining made of natural rubber)
2 Measuring electrodes for signal detection
3 Reference electrode for potential equalization (not for lining made of natural rubber)

16 Endress+Hauser
Proline Promag 55S

Vibrations
Secure and fix both the piping and the sensor if the vibrations are severe.

" Caution!
It is advisable to install sensor and transmitter separately if vibration is excessively severe. Information on the
permitted resistance to vibration and shock → Page 20.

A0003208

Measures to prevent vibration of the measuring device (L > 10 m/33 ft)

Foundations, supports
If the nominal diameter is DN ≥ 350 (14"), mount the sensor on a foundation of adequate load-bearing
strength.

" Caution!
Risk of damage.
Do not support the weight of the sensor on the metal casing: the casing would buckle and damage the internal
magnetic coils.

A0003209

Endress+Hauser 17
Proline Promag 55S

Adapters
Suitable adapters to DIN EN 545 (double-flange reducers) can be used to install the sensor in larger-diameter
pipes. The resultant increase in the rate of flow improves measuring accuracy with very slow-moving fluids.
The nomogram shown here can be used to calculate the pressure loss caused by cross-section reduction.

! Note!
The nomogram applies to fluids of viscosity similar to water.
1. Calculate the ratio of the diameters d/D.
2. From the nomogram, read off the pressure loss as a function of fluid velocity (downstream from the
reduction) and the d/D ratio.

[mbar]
100

8 m/s

7 m/s

6 m/s

10
5 m/s

4 m/s
max. 8°
3 m/s
d D
2 m/s
1

1 m/s

0.5 0.6 0.7 0.8 0.9


d/D

A0003213

Pressure loss due to adapters (for pressure losses in US units, please contact Endress+Hauser)

Inlet and outlet runs


If possible, install the sensor well clear of fittings such as valves, T-pieces, elbows, etc.
Compliance with the following requirements for the inlet and outlet runs is necessary in order to ensure meas-
uring accuracy.
• Inlet run ≥ 5 x DN
• Outlet run ≥ 2 x DN

 5 x DN  2 x DN

A0003210

18 Endress+Hauser
Proline Promag 55S

Length of connecting cable


In order to ensure measuring accuracy, comply with the following instructions when installing the remote
version:
• Secure the cable run or route the cable in an armored conduit. Movement of the cable can falsify the meas-
uring signal, particularly if the fluid conductivity is low.
• Route the cable well clear of electrical machines and switching elements.
• Ensure potential equalization between sensor and transmitter, if necessary.
• The permissible cable length Lmax depends on the fluid conductivity (see Figure).

[µS/cm]

50

40

30
L max
20

10
5
0 [m]
0 10 20 30 40 50 L max
[ft]
0 50 100 150
A0006116-ae

Permitted lengths for connecting cable in remote version, as a function of the conductivity of the fluid
Gray shaded area = permissible area
Lmax = Length of connecting cable

Endress+Hauser 19
Proline Promag 55S

Operating conditions: Environment

Ambient temperature Transmitter:


• Standard:
– Compact version: –20 to +50 °C (–4 to +122 °F)
– Remote version: –20 to +60 °C (–4 to +140 °F)
• Optional:
– Compact version: –40 to +50 °C (–40 to +122 °F)
– Remote version: –40 to +60 °C (–40 to +140 °F)

! Note!
At ambient temperatures below –20 °C (–4 °F), the readability of the display may be impaired.
Sensor:
• Flange material carbon steel: –10 to +60 °C (+14 to +140 °F)
• Flange material stainless steel: –40 to +60 °C (–40 to +140 °F)

" Caution!
Do not exceed the min. and max. temperatures for the lining of the measuring tube (→ "Medium temperature
range").
Note the following points:
• Install the device at a shady location. Avoid direct sunlight, particularly in warm climatic regions.
• If both fluid and ambient temperatures are high, install the transmitter at a remote location from the sensor
(→ "Medium temperature range").

Storage temperature The storage temperature corresponds to the operating temperature range of the transmitter and sensor.

Degree of protection • Standard: IP 67 (NEMA 4X) for transmitter and sensor


• Optional: IP 68 (NEMA 6P) for remote version of Promag S sensor

Shock and vibration resistance Acceleration up to 2 g by analogy with IEC 600 68-2-6
(High temperature version: no data available)

Electromagnetic compatibility According to IEC/EN 61326 and NAMUR recommendation NE 21


(EMC)

20 Endress+Hauser
Proline Promag 55S

Operating conditions: Process

Medium temperature range The permitted temperature depends on the lining of the measuring tube:
• 0 to +60 °C (+32 to +140 °F) for natural rubber (DN 65 to 600 / 2½ to 24")
• 0 to +80 °C (+32 to +176 °F) for hard rubber (DN 65 to 600 / 2½ to 24")
• –20 to +50 °C (–4 to +122 °F) for polyurethane (DN 25 to 1000 / 1 to 40")
• –20 to +180 °C (–4 to +356 °F) for PFA (DN 25 to 200 / 1 to 8"), restrictions → see diagrams
• –40 to +130 °C (–40 to +266 °F) for PTFE (DN 15 to 600 / ½ to 24"), restrictions → see diagrams

HT
TA
[°F] [°C]

140 60

40
100

20

PFA
0 -20
m
PTFE
-40 -40
-40 -20 0 20 40 60 80 100 120 140 160 180 [°C]
TF
-40 0 100 200 300 360 [°F]

A0006119-ae

Promag S compact versions (with PFA or PTFE lining)


TA Ambient temperature
TF Fluid temperature
HT High temperature version with insulation
➀ Gray shaded area → temperature range from –10 to –40 °C (–14 to –40 °F) applies only to stainless steel flanges

HT
TA
[°F] [°C]

140 60

40
100

20

PFA
0 -20
m
PTFE
-40 -40
-40 -20 0 20 40 60 80 100 120 140 160 180 [°C]
TF
-40 0 100 200 300 360 [°F]

A0002671-ae

Promag S remote versions (with PFA or PTFE lining)


TA Ambient temperature
TF Fluid temperature
HT High temperature version with insulation
➀ Gray shaded area → temperature range from –10 to –40 °C (–14 to –40 °F) applies only to stainless steel flanges

Endress+Hauser 21
Proline Promag 55S

Conductivity Minimum conductivity:


• ≥5 μS/cm for all liquids (incl. demineralized water)

! Note!
In the remote version, the required minimum conductivity is also influenced by the length of the cable
→ Page 19

Limiting medium pressure • EN 1092-1 (DIN 2501): PN 10 (DN 200 to 600 / 8 to 24"), PN 16 (DN 65 to 600 / 2½ to 24"),
range (nominal pressure) PN 25 (DN 200 to 600 / 8 to 24"), PN 40 (DN 15 to 150 / ½ to 6")
• ANSI B16.5: Class 150 (DN ½ to 24"), Class 300 (DN ½ to 6")
• JIS B2220: 10K (DN 50 to 300 / 2 to 12"), 20K (DN 15 to 300 / ½ to 12")
• AS 2129: Table E (DN 25/1", DN 50/2")
• AS 4087: Cl. 14 (DN 50/2")

Pressure tightness (lining) Resistance to partial vacuum in SI units [mbar]

Promag S Measuring tube Resistance of measuring tube lining to partial vacuum (SI units)
Nominal lining Limit values for abs. pressure [mbar] at various fluid temperatures
diameter
[mm] 25 °C 50 °C 80 °C 100 °C 130 °C 150 °C 180 °C
25 to 600 Polyurethane 0 0 - - - - -
65 to 600 Natural rubber 0 0 - - - - -
65 to 600 Hard rubber 0 0 0 - - - -

Promag S Measuring tube Resistance of measuring tube lining to partial vacuum (SI units)
Nominal lining Limit values for abs. pressure [mbar] at various fluid temperatures
diameter
[mm] 25 °C 80 °C 100 °C 130 °C 150 °C 180 °C
15 PTFE 0 0 0 100 – –
25 PTFE / PFA 0/0 0/0 0/0 100/0 –/0 –/0
32 PTFE / PFA 0/0 0/0 0/0 100/0 –/0 –/0
40 PTFE / PFA 0/0 0/0 0/0 100/0 –/0 –/0
50 PTFE / PFA 0/0 0/0 0/0 100/0 –/0 –/0
65 PTFE / PFA 0/0 * 40/0 130/0 –/0 –/0
80 PTFE / PFA 0/0 * 40/0 130/0 –/0 –/0
100 PTFE / PFA 0/0 * 135/0 170/0 –/0 –/0
125 PTFE / PFA 135/0 * 240/0 385/0 –/0 –/0
150 PTFE / PFA 135/0 * 240/0 385/0 –/0 –/0
200 PTFE / PFA 200/0 * 290/0 410/0 –/0 –/0
250 PTFE 330 * 400 530 – –
300 PTFE 400 * 500 630 – –
350 PTFE 470 * 600 730 – –
400 PTFE 540 * 670 800 – –
450 PTFE
500 PTFE Partial vacuum is impermissible
600 PTFE
* No value can be quoted.

22 Endress+Hauser
Proline Promag 55S

Pressure tightness in US units [psia = pounds/inch2]

Promag S Measuring tube Resistance of measuring tube lining to partial vacuum (US units)
Nominal lining Limit values for abs. pressure [mbar] at various fluid temperatures
diameter
[inch] 77 °F 122 °F 176 °F 212 °F 266 °F 302 °F 356 °F
1 to 24" Polyurethane 0 0 - - - - -
3 to 24" Natural rubber 0 0 - - - - -
3 to 24" Hard rubber 0 0 0 - - - -

Promag S Measuring tube Resistance of measuring tube lining to partial vacuum (US units)
Nominal lining Limit values for abs. pressure [mbar] at various fluid temperatures
diameter
[inch] 77 °F 176 °F 212 °F 266 °F 302 °F 356 °F
½" PTFE 0 0 0 1.5 – –
1" PTFE / PFA 0/0 0/0 0/0 1.5/0 –/0 –/0
– PTFE / PFA 0/0 0/0 0/0 1.5/0 –/0 –/0
1½" PTFE / PFA 0/0 0/0 0/0 1.5/0 –/0 –/0
2" PTFE / PFA 0/0 0/0 0/0 1.5/0 –/0 –/0
– PTFE / PFA 0/0 * 0.6/0 1.9/0 –/0 –/0
3" PTFE / PFA 0/0 * 0.6/0 1.9/0 –/0 –/0
4" PTFE / PFA 0/0 * 2.0/0 2.5/0 –/0 –/0
– PTFE / PFA 2.0/0 * 3.5/0 5.6/0 –/0 –/0
6" PTFE / PFA 2.0/0 * 3.5/0 5.6/0 –/0 –/0
8" PTFE / PFA 2.9/0 * 4.2/0 5.9/0 –/0 –/0
10" PTFE 4.8 * 5.8 7.7 – –
12" PTFE 5.8 * 7.3 9.1 – –
14" PTFE 6.8 * 8.7 10.6 – –
16" PTFE 7.8 * 9.7 11.6 – –
18" PTFE
20" PTFE Partial vacuum is impermissible
24" PTFE
* No value can be quoted.

Endress+Hauser 23
Proline Promag 55S

Nominal diameter and The diameter of the pipe and the flow rate determine the nominal diameter of the sensor. The optimum flow
flow rate velocity is between 2 and 3 m/s (6 to 10 ft/s). The flow velocity (v), moreover, has to be matched to the phy-
sical properties of the fluid:
• v < 2 m/s (<6 ft/s): for abrasive fluids where solids do not cause sedimentation (e.g. lime milk)
• v > 2 m/s (>6 ft/s): for fluids producing build-up (e.g. wastewater sludge)
• v > 2 m/s (>6 ft/s): for abrasive sludge with a high sand or stone content and where the solids easily cause
sedimentation (e.g. ore slurry)

! Note!
Flow velocity can be increased, if necessary, by reducing the nominal diameter of the sensor through the use
of adapters → Page 18.

Flow rate characteristic values - Promag S (SI units)

Nominal Recommended flow rate Factory settings


diameter

min./max. full scale value Full scale value Pulse value Low flow cut off
[mm] (v ≈ 0.3 or 10 m/s) (v ≈ 2.5 m/s) (≈ 2 pulse/s) (v ≈ 0.04 m/s)

15 4 to 100 dm3/min 25 dm3/min 0.20 dm3 0.5 dm3/min

25 9 to 300 dm3/min 75 dm3/min 0.50 dm3 1 dm3/min

32 15 to 500 dm3/min 125 dm3/min 1.00 dm3 2 dm3/min

40 25 to 700 dm3/min 200 dm3/min 1.50 dm3 3 dm3/min

50 35 to 1100 dm3/min 300 dm3/min 2.50 dm3 5 dm3/min

65 60 to 2000 dm3/min 500 dm3/min 5.00 dm3 8 dm3/min

80 90 to 3000 dm3/min 750 dm3/min 5.00 dm3 12 dm3/min

100 145 to 4700 dm3/min 1200 dm3/min 10.00 dm3 20 dm3/min

125 220 to 7500 dm3/min 1850 dm3/min 15.00 dm3 30 dm3/min

150 20 to 600 m3/h 150 m3/h 0.025 m3 2.5 m3/h

200 35 to 1100 m3/h 300 m3/h 0.05 m3 5.0 m3/h

250 55 to 1700 m3/h 500 m3/h 0.05 m3 7.5 m3/h

300 80 to 2400 m3/h 750 m3/h 0.10 m3 10 m3/h

350 110 to 3300 m3/h 1000 m3/h 0.10 m3 15 m3/h

400 140 to 4200 m3/h 1200 m3/h 0.15 m3 20 m3/h

450 180 to 5400 m3/h 1500 m3/h 0.25 m3 25 m3/h

500 220 to 6600 m3/h 2000 m3/h 0.25 m3 30 m3/h

600 310 to 9600 m3/h 2500 m3/h 0.30 m3 40 m3/h

24 Endress+Hauser
Proline Promag 55S

Flow rate characteristic values - Promag S (US units)

Nominal Recommended flow rate Factory settings


diameter

min./max. full scale value Full scale value Pulse value Low flow cut off
[inch] (v ≈ 1.0 or 33 ft/s) (v ≈ 8.2 ft/s) (≈ 2 pulse/s) (v ≈ 0.1 ft/s)

½" 1.0 to 27 gal/min 6 gal/min 0.05 gal 0.10 gal/min

1" 2.5 to 80 gal/min 18 gal/min 0.20 gal 0.25 gal/min

1¼" 4 to 130 gal/min 30 gal/min 0.20 gal 0.50 gal/min

1½" 7 to 190 gal/min 50 gal/min 0.50 gal 0.75 gal/min

2" 10 to 300 gal/min 75 gal/min 0.50 gal 1.25 gal/min

2½" 16 to 500 gal/min 130 gal/min 1 gal 2.0 gal/min

3" 24 to 800 gal/min 200 gal/min 2 gal 2.5 gal/min

4" 40 to 1250 gal/min 300 gal/min 2 gal 4.0 gal/min

5" 60 to 1950 gal/min 450 gal/min 5 gal 7.0 gal/min

6" 90 to 2650 gal/min 600 gal/min 5 gal 12 gal/min

8" 155 to 4850 gal/min 1200 gal/min 10 gal 15 gal/min

10" 250 to 7500 gal/min 1500 gal/min 15 gal 30 gal/min

12" 350 to 10600 gal/min 2400 gal/min 25 gal 45 gal/min

14" 500 to 15000 gal/min 3600 gal/min 30 gal 60 gal/min

16" 600 to 19000 gal/min 4800 gal/min 50 gal 60 gal/min

18" 800 to 24000 gal/min 6000 gal/min 50 gal 90 gal/min

20" 1000 to 30000 gal/min 7500 gal/min 75 gal 120 gal/min

24" 1400 to 44000 gal/min 10500 gal/min 100 gal 180 gal/min

Pressure loss • No pressure loss if the sensor is installed in a pipe of the same nominal diameter.
• Pressure losses for configurations incorporating adapters according to DIN EN 545 → Page 18

Endress+Hauser 25
Proline Promag 55S

Measuring tube specifications


Measuring tube specifications - Promag S (SI units)

Nominal Pressure rating Inside diameter of


diameter measuring tube

EN (DIN) AS 2129 AS 4087 ANSI JIS with PFA with PTFE PU1) HR1)
[mm] [bar] [lbs] [mm] [mm] [mm] [mm]

15 PN 40 – – Cl 150 20K – 15 – –

25 PN 40 Table E – Cl 150 20K 23 26 24 –

32 PN 40 – – – 20K 32 35 32 –

40 PN 40 – – Cl 150 20K 36 41 38 –

50 PN 40 Table E Cl.14 Cl 150 10K 48 52 50 –

65 PN 16 – – – 10K 63 67 66 66

80 PN 16 – – Cl 150 10K 75 80 79 79

100 PN 16 – – Cl 150 10K 101 104 102 102

125 PN 16 – – – 10K 126 129 127 127

150 PN 16 – – Cl 150 10K 154 156 156 156

200 PN 10 – – Cl 150 10K 201 202 204 204

250 PN 10 – – Cl 150 10K – 256 258 258

300 PN 10 – – Cl 150 10K – 306 309 309

350 PN 10 – – Cl 150 – – 337 342 342

400 PN 10 – – Cl 150 – – 387 392 392

450 PN 10 – – Cl 150 – – 432 437 437

500 PN 10 – – Cl 150 – – 487 492 492

600 PN 10 – – Cl 150 – – 593 594 594

1)
Abbreviations (lining): PU = Polyurethane, HR = Hard rubber

26 Endress+Hauser
Proline Promag 55S

Measuring tube specifications - Promag S (US units)

Nominal Pressure rating Inside diameter of


diameter measuring tube

EN (DIN) AS 2129 AS 4087 ANSI JIS with PFA with PTFE PU1) HR1)
[inch] [bar] [lbs] [inch] [inch] [inch] [inch]

½" PN 40 – – Cl 150 20K – 0.59 – –

1" PN 40 Table E – Cl 150 20K 0.91 1.02 0.94 –

– PN 40 – – – 20K 1.26 1.34 1.26 –

1½" PN 40 – – Cl 150 20K 1.42 1.61 1.50 –

2" PN 40 Table E Cl.14 Cl 150 10K 1.89 2.05 1.97 –

– PN 16 – – – 10K 2.48 2.64 2.60 2.60

3" PN 16 – – Cl 150 10K 2.95 3.15 3.11 3.11

4" PN 16 – – Cl 150 10K 3.98 4.09 4.02 4.02

– PN 16 – – – 10K 4.96 5.08 5.00 5.00

6" PN 16 – – Cl 150 10K 6.06 6.14 6.14 6.14

8" PN 10 – – Cl 150 10K 7.91 7.92 8.03 8.03

10" PN 10 – – Cl 150 10K – 10.08 10.16 10.16

12" PN 10 – – Cl 150 10K – 12.05 12.17 12.17

14" PN 10 – – Cl 150 – – 13.27 13.46 13.46

16" PN 10 – – Cl 150 – – 15.24 15.43 15.43

18" PN 10 – – Cl 150 – – 17.01 17.20 17.20

20" PN 10 – – Cl 150 – – 19.17 19.37 19.37

24" PN 10 – – Cl 150 – – 23.35 23.39 23.39

1)
Abbreviations (lining): PU = Polyurethane, HR = Hard rubber

Endress+Hauser 27
Proline Promag 55S

Nominal diameter Inside diameter of Thickness of lining material


measuring tube with natural rubber natural rubber

[mm] [mm] [mm]

65 52 10

80 65 10

PN 16/150 lbs
100 91 10

125 116 10

150 142 12

200 190 12
250 244 12

300 292 13

350 322 14
PN 10

400 369 16

450 417 14

500 466 17
600 562 20

250 243 12

300 291 12

350 320 13
150 lbs

400 368 14

450 417 14

500 465 16
600 563 16

28 Endress+Hauser
Proline Promag 55S

Nominal diameter Inside diameter of Thickness of lining material


measuring tube with natural rubber natural rubber

[inch] [inch] [inch]

2½" 2.05 0.39

3" 2.56 0.39

PN 16/150 lbs
4" 3.58 0.39

5" 4.57 0.39

6" 5.59 0.47

8" 6.30 0.47


10" 9.61 0.47

12" 11.50 0.51

14" 12.68 0.55

PN 10
16" 14.53 0.63

18" 16.42 0.55

20" 18.35 0.67


24" 22.13 0.79

10" 9.57 0.47

12" 11.46 0.47

14" 12.60 0.51


150 lbs

16" 14.49 0.55

18" 16.42 0.55

20" 18.31 0.63


24" 22.17 0.63

Endress+Hauser 29
Proline Promag 55S

Mechanical construction
Design, dimensions Dimensions: Wall-mount housing (non hazardous area and II3G / zone 2)

D
Esc
- + E

B
C
F G
A E

H J K J

N
S S

M
R

L
P Q P O

A0001150

Metric units [mm]

A B C D E F G H J K L M N O P Q R

215 250 90.5 159.5 135 90 45 >50 81 53 95 53 102 81.5 11.5 192 8xM5

US units [inch]

A B C D E F G H J K L M N O P Q R

8.46 9.84 3.56 6.28 5.32 3.54 1.77 >1.97 3.19 2.09 3.74 2.09 4.02 3.21 0.45 7.56 8xM5

30 Endress+Hauser
Proline Promag 55S

There is a separate mounting kit for the wall-mounted housing. It can be ordered from Endress+Hauser as an
accessory. The following installation variants are possible:
• Panel-mounted installation
• Pipe mounting

Installation in control panel

+0.5 (+0.019)
–0.5 (–0.019)
210 (8.27)

+0.5 (+0.019)
–0.5 (–0.019)
245 (9.65)

mm (inch) ~110 (~4.33)

A0001131-ae

Pipe mounting

Ø 20…70
(Ø 0.79…2.75)

~155 (~ 6.1)

mm (inch)

A0001132-ae

Endress+Hauser 31
Proline Promag 55S

Compact version DN ≤ 300, Flange connections to EN (DIN) / JIS / AS

227 187
207 168

160
Esc
- + E

B
A
C
E
K L (DVGW)

A0005423-en

DN L A B C K E
1)
EN (DIN) / JIS /AS
[mm] [mm] [mm] [mm] [mm] [mm] [mm]

15 200 341 257 84 120 94

25 200 341 257 84 120 94

32 200 341 257 84 120 94


40 200 341 257 84 120 94

50 200 341 257 84 120 94

65 200 391 282 109 180 94


80 200 391 282 109 180 94

100 250 391 282 109 180 94

125 250 472 322 150 260 140


150 300 472 322 150 260 140

200 350 527 347 180 324 156

250 450 577 372 205 400 156

300 500 627 397 230 460 166

The fitting length (L) is always the same, regardless of the pressure rating.
1)
Only DN 25 and DN 50 are available for flanges according to AS.

High temperature version DN ≤ 300 (12")


B1
110 (4.33)

A1

mm (inch)

A0005529-ae

Measurement A1, B1 = Measurement A, B of the standard compact version plus 110 mm (4.33 inch)

32 Endress+Hauser
Proline Promag 55S

Compact version DN ≤ 300 (12"), Flange connections to ANSI

227 (8.94) 187 (7.36)


207 (8.15) 168 (6.61)

160 (6.30)
Esc
- + E

B
A
C
E
K L (DVGW) mm (inch)

A0005423-ae

DN L A B C K E

ANSI
[inch] [inch] [inch] [inch] [inch] [inch] [inch]

½" 7.87 13.43 10.12 3.31 4.72 3.70

1 7.87 13.43 10.12 3.31 4.72 3.70


1½" 7.87 13.43 10.12 3.31 4.72 3.70

2" 7.87 13.43 10.12 3.31 4.72 3.70

3" 7.87 15.39 11.10 4.29 7.09 3.70


4" 9.84 15.39 11.10 4.29 7.09 3.70

6" 11.81 18.58 12.68 5.91 10.24 5.51

8" 13.78 20.75 13.66 7.09 12.76 6.14

10" 17.72 22.72 14.65 8.07 15.75 6.14

12" 19.69 24.69 15.63 9.06 18.11 6.54

The fitting length (L) is always the same, regardless of the pressure rating.

Endress+Hauser 33
Proline Promag 55S

Remote version DN ≤ 300, Flange connections to EN (DIN) / JIS / AS

129 163
143

102
B
A
C
E
K L (DVGW)

A0003219-en

DN L A B C K E

EN (DIN) / JIS / AS 1)
[mm] [mm] [mm] [mm] [mm] [mm] [mm]
15 200 286 202 84 120 94

25 200 286 202 84 120 94

32 200 286 202 84 120 94


40 200 286 202 84 120 94

50 200 286 202 84 120 94

65 200 336 227 109 180 94

80 200 336 227 109 180 94

100 250 336 227 109 180 94

125 250 417 267 150 260 140


150 300 417 267 150 260 140

200 350 472 292 180 324 156

250 450 522 317 205 400 156

300 500 572 342 230 460 166

The fitting length (L) is always the same, regardless of the pressure rating.
1)
Only DN 25 and DN 50 are available for flanges according to AS.

High temperature version DN ≤ 300 (12")


B1
110 (4.33)

A1

mm (inch)

A0005570-ae

Measurement A1, B1 = Measurement A, B of the standard compact version plus 110 mm (4.33 inch)

34 Endress+Hauser
Proline Promag 55S

Remote version DN ≤ 300 (12"), Flange connections to ANSI

129 (5.08) 163 (6.42)


143 (5.63)

102 (4.02)

B
A
C
E
K L (DVGW) mm (inch)

A0003219-ae

DN L A B C K E

ANSI
[inch] [inch] [inch] [inch] [inch] [inch] [inch]
½" 7.87 11.26 7.95 3.31 4.72 3.70

1" 7.87 11.26 7.95 3.31 4.72 3.70

1½" 7.87 11.26 7.95 3.31 4.72 3.70


2" 7.87 11.26 7.95 3.31 4.72 3.70

3" 7.87 13.23 8.94 4.29 7.09 3.70

4" 9.84 13.23 8.94 4.29 7.09 3.70

6" 11.81 16.42 10.51 5.91 10.24 5.51

8" 13.78 18.58 11.50 7.08 12.76 6.14

10" 17.72 20.55 12.48 8.07 15.75 6.14


12" 19.69 22.52 13.46 9.06 18.11 6.54

The fitting length (L) is always the same, regardless of the pressure rating.

Endress+Hauser 35
Proline Promag 55S

Compact version DN ≥ 300 (12"), Flange connections to EN (DIN) und ANSI

227 (8.94) 187 (7.36) mm (inch)


207 (8.15) 168 (6.61)

160 (6.30)
Esc
- + E

B
A
C
E

K L

A0005424-ae

DN L A B C K E

EN (DIN)
[mm] [mm] [mm] [mm] [mm] [mm] [mm]

350 550 738.5 456.5 282.0 564 276

400 600 790.5 482.5 308.0 616 276


450 650 840.5 507.5 333.0 666 292

500 650 891.5 533.0 358.5 717 292

600 780 995.5 585.0 410.5 821 402


The fitting length (L) is always the same, regardless of the pressure rating.

DN L A B C K E

ANSI
[inch] [inch] [inch] [inch] [inch] [inch] [inch]
14" 21.65 29.07 17.97 11.10 22.20 10.87

16" 23.62 31.12 19.00 12.12 24.25 10.87

18" 25.59 33.09 19.98 13.11 26.22 11.50

20" 25.59 35.10 20.98 14.12 28.23 11.50

24" 30.71 39.19 23.03 16.16 32.32 15.83

The fitting length (L) is always the same, regardless of the pressure rating.

36 Endress+Hauser
Proline Promag 55S

Remote version DN ≥ 300 (12"), Flange connections to EN (DIN) und ANSI

129 (5.08) 163 (6.42) mm (inch)


143 (5.63)

102 (4.02)

B
A
C
E
K L

A0003220-ae

DN L A B C K E

EN (DIN)
[mm] [mm] [mm] [mm] [mm] [mm] [mm]

350 550 683.5 401.5 282.0 564 276

400 600 735.5 427.5 308.0 616 276


450 650 785.5 452.5 333.0 666 292

500 650 836.5 478.0 358.5 717 292

600 780 940.5 530.0 410.5 821 402


The fitting length (L) is always the same, regardless of the pressure rating.

DN L A B C K E

ANSI
[inch] [inch] [inch] [inch] [inch] [inch] [inch]

14" 21.65 26.91 15.81 11.10 22.20 10.87

16" 23.62 28.96 16.83 12.13 24.25 10.87

18" 25.59 30.93 17.81 13.12 26.22 11.50

20" 25.59 32.93 18.82 14.11 28.23 11.50

24" 30.71 37.03 20.87 16.16 32.32 15.83

The fitting length (L) is always the same, regardless of the pressure rating.

Endress+Hauser 37
Proline Promag 55S

Ground disk for flange connections to EN (DIN) / JIS / AS

DN £ 300 DN ³ 350

Ø9
Ø 6.5

H
H

ØB
A
Ø ØD
ØB
A ØD
Ø

t=2 t=2

A0003221-en

DN 1) A B D H

EN (DIN) / JIS / PTFE, PFA, PU, HR 5) NR 5)


AS 4) [mm] [mm] [mm] [mm] [mm] [mm]
15 16 – 43 61.5 73

25 26 – 62 77.5 87.5

32 35 – 80 87.5 94.5

40 41 – 82 101 103

50 52 – 101 115.5 108

65 68 53 121 131.5 118

80 80 66 131 154.5 135

100 104 91.5 156 186.5 153

125 130 117 187 206.5 160

150 158 143.5 217 256 184

200 206 192 267 288 205

250 260 245 328 359 240


2) 312 294.5 375 413 273
300
3)
300 310 – 375 404 268
2)
350 343 323.5 433 479 365

400 2) 393 371 480 542 395


2)
450 439 420 538 583 417
2)
500 493 469 592 650 460

600 2) 593 566 693 766 522


1) Ground disks at DN 15 to 250 can be used for all flange standards/nominal pressures available as standard.
2)
PN 10/16, Cl 150
3)
PN 25, JIS 10K/20K
4) Only DN 25 and DN 50 are available for flanges according to AS
5)
Abbreviations (lining): PU = Polyurethane, NR = Natural rubber, HR = Hard rubber

38 Endress+Hauser
Proline Promag 55S

Ground disk for flange connections to ANSI

DN £ 300 (12") DN ³ 350 (14")


mm [inch]
Ø 9 (0.35)
Ø 6.5 (0.26)

H
H

ØB
A
Ø ØD
ØB
A ØD
Ø

t = 2 (0.08) t = 2 (0.08)

a0003221-ae

DN 1) A B D H

ANSI PTFE, PFA, PU, HR 2) NR 2)


[inch] [inch] [inch] [inch] [inch] [inch]
½" 0.63 – 1.69 2.42 2.87

1" 1.02 – 2.44 3.05 3.44

1½" 1.61 – 3.23 3.98 4.06

2" 2.05 – 3.98 4.55 4.25

3" 3.15 2.60 5.16 6.08 5.31

4" 4.09 3.60 6.14 7.34 6.02

6" 6.22 5.65 8.54 10.08 7.24

8" 8.11 7.56 10.51 11.34 8.07

10" 10.24 9.65 12.91 14.13 9.45

12" 12.28 11.59 14.76 16.26 10.75

14" 13.50 12.74 17.05 18.86 14.37

16" 15.47 14.61 18.90 21.34 15.55

18" 17.28 16.54 21.18 22.95 16.42

20" 19.41 18.46 23.31 25.59 18.11

24" 23.35 22.28 27.28 30.16 20.55


1) Ground disks can be used for all pressure ratings.
2)
Abbreviations (lining): PU = Polyurethane, NR = Natural rubber, HR = Hard rubber

Endress+Hauser 39
Proline Promag 55S

Weight Weight data in SI units [kg]

Weight in kilograms [kg]

Nominal Compact version Remote version (without cable)


diameter
Sensor Transmitter

[mm] [inch] EN (DIN) JIS ANSI EN (DIN) JIS ANSI (Wall-mount housing)
/ AS* / AS*

15 ½" 6.5 6.5 6.5 4.5 4.5 4.5 6.0

25 1" 7.3 7.3 7.3 5.3 5.3 5.3 6.0


PN 40

PN 40
32 1¼" 8.0 7.3 – 6.0 5.3 – 6.0

40 1½" 9.4 8.3 9.4 7.4 6.3 7.4 6.0

50 2" 10.6 9.3 10.6 8.6 7.3 8.6 6.0

65 2½" 12.0 11.1 – 10.0 9.1 – 6.0


10K

10K
80 3" 14.0 12.5 14.0 12.0 10.5 12.0 6.0
PN 16

PN 16
100 4" 16.0 14.7 16.0 14.0 12.7 14.0 6.0
Class 150

Class 150
125 5" 21.5 21.0 – 19.5 19.0 – 6.0

150 6" 25.5 24.5 25.5 23.5 22.5 23.5 6.0

200 8" 45 41.9 45 43 39.9 43 6.0

250 10" 65 69.4 75 63 67.4 73 6.0

300 12" 70 72.3 110 68 70.3 108 6.0


350 14" 115 175 113 173 6.0
PN 10

PN 10

400 16" 135 205 133 203 6.0

450 18" 175 255 173 253 6.0


500 20" 175 285 173 283 6.0

600 24" 235 405 233 403 6.0

Transmitter (compact version): 3.4 kg


High temperature version: +1.5 kg
(Weight data valid for standard pressure ratings and without packaging material)
* Only DN 25 and DN 50 are available for flanges according to AS

40 Endress+Hauser
Proline Promag 55S

Weight data in US units [lbs = pounds]

Nominal Weight in pounds [lbs]


diameter

Compact version Remote version (without cable)

Sensor Transmitter

[mm] [inch] EN (DIN) JIS ANSI EN (DIN) JIS ANSI (Wall-mount housing)
/ AS* / AS*

15 ½" 14 14 14 10 10 10 13

25 1" 16 16 16 12 12 12 13

PN 40

PN 40
32 1¼" 18 16 – 13 12 – 13

40 1½" 21 18 21 16 14 16 13
50 2" 23 21 23 19 16 19 13

65 2½" 26 24 – 22 20 – 13

10K

10K
80 3" 31 28 31 26 23 26 13
PN 16

PN 16
100 4" 35 32 35 31 28 31 13

Class 150

Class 150
125 5" 47 46 – 43 42 – 13

150 6" 56 54 56 52 50 52 13

200 8" 99 92 99 95 88 95 13
250 10" 143 153 165 139 149 161 13

300 12" 154 159 243 150 155 238 13

350 14" 254 386 249 381 13


PN 10

PN 10

400 16" 298 452 293 448 13

450 18" 386 562 381 558 13

500 20" 386 628 381 624 13

600 24" 518 893 514 889 13

Transmitter (compact version): +7.5 lbs


High temperature version: +3.3 lbs
(Weight data valid for standard pressure ratings and without packaging material)
* Only DN 25 (1") and DN 50 (2") are available for flanges according to AS

Endress+Hauser 41
Proline Promag 55S

Materials Transmitter housing:


• Compact and remote version: Powder-coated die-cast aluminum
Sensor housing:
• DN 15 to 300 (½ to 12"): Powder-coated die-cast aluminum
• DN 350 to 600 (14 to 24"): Painted steel (Amerlock 400)
Measuring tube:
• DN < 350 (14"): Stainless steel 1.4301 (SS 304) or 1.4306/304L. For flanges of carbon steel with Al/Zn
protective coating.
• DN > 300 (12"): Stainless steel 1.4301/304. For flanges of carbon steel with Amerlock 400 paint finish.
Flange:
• EN 1092-1 (DIN 2501): 316L / 1.4571 (SS 316Ti); RSt37-2 (S235JRG2) / C22 / FE 410W B
(DN < 350/14": with Al/Zn protective coating; DN > 300/12" with Amerlock 400 paint finish)
• ANSI: A105; F316L
(DN < 350/14" with Al/Zn protective coating; DN > 300/12" with Amerlock 400 paint finish)
• JIS: RSt37-2 (S235JRG2) / HII / 1.0425 / 316L
(DN < 350 with Al/Zn protective coating; DN > 300/12" with Amerlock 400 paint finish)
• AS 2129:
– DN 25 (1"): A105 or RSt37-2 (S235JRG2), with Al/Zn protective coating
– DN 50 (2"): A105 or St44-2 (S275JR), with Al/Zn protective coating
• AS 4087:
– DN 50 (2"): A105 or St44-2 (S275JR), with Al/Zn protective coating
Ground disks: 1.4435/316L or Alloy C-22
Electrodes: 1.4435/316L, platinum/rhodium 80/20, Alloy C-22, tantalum, tungsten carbide coating
(for electrodes made from 1.4435), 1.4310/302 (for brush electrodes)
Seals: according to DIN EN 1514-1

Material load diagrams Caution!


The following diagrams contain material load curves (reference curves) for various process connections relating
to the fluid temperature. But the maximal permissible fluid temperature always depends on the lining material
of the sensor and/or the sealing material.

Flange connection to EN 1092-1 (DIN 2501)


Flange material: RSt37-2 (S235JRG2) / C22 / Fe 410W B

[psi] [bar]
600
40
PN 40
500 35

30
400
25
PN 25
300 20

200 15
PN 16
10
PN 10
100
5 PN 6

0 0
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 [°C]

-40 0 100 200 300 360 [°F]

A0005594-ae

42 Endress+Hauser
Proline Promag 55S

Flange connection to EN 1092-1 (DIN 2501)


Flange material: 316L / 1.4571

[psi] [bar]
600
40
PN 40

500 35

30
400
25
PN 25
300 20

200 15 PN 16

10
PN 10
100
5 PN 6

0 0
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 [°C]

-40 0 100 200 300 360 [°F]

A0005304-ae

Flange connection to ANSI B16.5


Flange material: A105

[psi] [bar]
900 60
800
700 50 Class 300
600 40
500
400 30
300 20 Class 150
200
10
100
0 0
-40 -20 0 20 40 60 80 100 120 140 160 180 [°C]

-40 0 100 200 300 360 [°F]

A0003226-ae

Flange connection to ANSI B16.5


Flange material: F316L

[psi] [bar]
900 60
800
700 50
600 40 Class 300
500
400 30
300 20
200 Class 150
10
100
0 0
-40 -20 0 20 40 60 80 100 120 140 160 180 [°C]

-40 0 100 200 300 360 [°F]

A0005307-ae

Endress+Hauser 43
Proline Promag 55S

Flange connection to JIS B2220


Flange material: RSt37-2 (S235JRG2) / H II / 1.0425

[psi] [bar]

30
400

20K
300 20

200
10K
10
100

0 0
-40 -20 0 20 40 60 80 100 120 140 160 180 [°C]

-40 0 100 200 300 360 [°F]

A0003228-ae

Flange connection to AS 2129 Table E or AS 4087 Cl. 14


Flange material: A105 / RSt37-2 (S235JRG2) / St44-2 (S275JR)

[psi] [bar]

300 20

200 15

10
100
5
0 0
-40 -20 0 20 40 60 80 100 120 140 160 [°C]

-40 0 100 200 300 [°F]

A0005595-ae

Fitted electrodes Reference and empty pipe detection electrodes:


• Comes as standard with: 1.4435/SS 316L, Alloy C-22, tantalum, Duplex,
tungsten carbide coating (for electrodes made of 1.4435)
• Optionally available: only for measuring electrodes made of platinum/rhodium 80/20
• Not available: for measuring tubes with natural rubber lining and brush electrodes

Process connections Flange connection:


• EN 1092-1 (DIN 2501; DN < 300: Form A; DN > 300: Form B; DN 65 PN 16 and DN 600 PN 16
exclusively according to EN 1092-1)
• ANSI B16.5
• JIS B2220
• AS 2129 Table E
• AS 4087 Cl. 14

Surface roughness • Lining with PFA: ≤0.4 μm (16 μin)


• Electrodes: 0.3 to 0.5 μm (12 to 20 μin)
All data relate to wetted parts.

44 Endress+Hauser
Proline Promag 55S

Human interface
Display elements • Liquid-crystal display: illuminated, four lines with 16 characters per line
• Custom configurations for presenting different measured values and status variables
• 3 totalizers
• At ambient temperatures below –20 °C (–4 °F), the readability of the display may be impaired.

Operating elements • Onsite operation with three optical sensor keys (S/O/F)
• Application-specific Quick Setup menus for straightforward commissioning

Language groups Language groups available for operation in different countries:


• Western Europe and America (WEA):
English, German, Spanish, Italian, French, Dutch, Portuguese
• Eastern Europe/Scandinavia (EES):
English, Russian, Polish, Norwegian, Finnish, Swedish, Czech
• South and East Asia (SEA):
English, Japanese, Indonesian
• China (CN):
English, Chinese

! Note!
You can change the language group via the operating program "FieldCare."

Language groups via HART protocol

Certificates and approvals


CE mark The measuring system described in these Operating Instructions therefore complies with the legal requirements
of the EU Directives. Endress+Hauser confirms this by affixing the CE mark to it and by issuing the CE
declaration of conformity.

C-tick mark The measuring system meets the EMC requirements of the "Australian Communications and Media Authority
(ACMA)".

Ex certification Information on the currently available Ex-rated versions (ATEX, FM, CSA, etc.) is available on request from
your Endress+Hauser sales outlet. All information relevant to explosion protection is available in separate docu-
ments that you can order as necessary.

Sanitary compatibility No applicable approvals or certification

Pressure measuring device All measuring devices, including those with a nominal diameter smaller than or equal to DN 25, correspond
approval to Article 3(3) of the EC Directive 97/23/EC (Pressure Equipment Directive) and have been designed and
manufactured according to good engineering practice. For nominal diameters greater than DN 25 (depending
on the fluid and process pressure), there are additional optional approvals according to category II/III.

FOUNDATION Fieldbus The flow device has successfully passed all the test procedures carried out and is certified and registered by the
certification Fieldbus Foundation. The device thus meets all the requirements of the following specifications:
• Certified to FOUNDATION Fieldbus Specification
• The device meets all the specifications of the FOUNDATION Fieldbus H1
• Interoperability Test Kit (ITK), revision status 5.0 (device certification number: on request)
• The device can also be operated with certified devices of other manufacturers
• Physical Layer Conformance Test of the Fieldbus Foundation

PROFIBUS PA The flowmeter has successfully passed all the test procedures carried out and is certified and registered by the
certification PNO (PROFIBUS User Organization). The device thus meets all the requirements of the following
specifications:
• Certified in accordance with PROFIBUS Profile Version 3.0 (device certification number: available on
request)
• The measuring device can also be operated with certified devices of other manufacturers (interoperability)

Endress+Hauser 45
Proline Promag 55S

Other standards, guidelines • EN 60529


Degrees of protection by housing (IP code)
• EN 61010-1
Protection Measures for Electrical Equipment for Measurement, Control,
Regulation and Laboratory Procedures
• IEC/EN 61326
"Emission in accordance with requirements for class A".
Electromagnetic compatibility (EMC requirements).
• NAMUR NE 21
Electromagnetic compatibility (EMC) of industrial process and laboratory control equipment.
• NAMUR NE 43
Standardization of the signal level for the breakdown information of digital transmitters with analog output
signal.
• NAMUR NE 53
Software of field devices and signal-processing devices with digital electronics.

Accessories
Various accessories are available for the transmitter and the sensor. These can be ordered separately from
Endress+Hauser.

! Note!
For detailed information on specific order codes, please contact the Endress+Hauser service organization.

Supplementary documentation
• Flow Measurement (FA005D/06/en)
• Operating Instructions Promag 55 (BA119D/06/en, BA120D/06/en)
• Operating Instructions Promag 55 PROFIBUS PA (BA124D/06/en, BA125D/06/en)
• Operating Instructions Promag 53 FOUNDATION Fieldbus (BA126D/06/en, BA127D/06/en)
• Supplementary documentation on Ex-ratings: ATEX, FM, CSA

Registered trademarks
HART®
Registered trademark of HART Communication Foundation, Austin, USA
PROFIBUS ®
Registered trademark of the PROFIBUS User Organisation, Karlsruhe, Germany
FOUNDATIONTM Fieldbus
Registered trademark of the Fieldbus Foundation, Austin, USA
HistoROM™, S-DAT®, T-DAT®, F-CHIP ®, FieldCare®, Fieldcheck®, Applicator®
Registered or registration-pending trademarks of Endress+Hauser Flowtec AG, Reinach, CH

46 Endress+Hauser
Proline Promag 55S

Ordering information
! Note!
Endress+Hauser reserves the right to change or modify product, specifications, and ordering information at any
time without notice. Please consult Endress+Hauser or your local representative for the most recent
information.

Promag 55S
010 020 030 040 050 060 070 080 090 100 110 120
Promag 55S -
Nominal Diameter 070 Housing
15 1/2” A NEMA 4X (IP 67) compact aluminum housing
25 1” C NEMA 4X (IP 67) remote wall-mounted (only for approvals A or R)
40 1-1/2” K NEMA 6P sensor, wall-mounted housing (only for approvals A or R, not for liner B and N)
50 2” P NEMA 4X (IP 67) compact aluminum housing, HE (harsh environment),
80 3” compact and remote sensor sizes up to 12” only (not for liner N) *
1H 4” S NEMA 4X (IP 67) remote wall-mounted housing, HE (harsh environment),
1F 6” remote sensor sizes 14” to 78” (only for approvals A and R, not for liner N) *
2H 8” 1 -40°F (-40°C) ambient temperature, NEMA 4X (IP 67), compact, aluminum field
2F 10” housing (only for approvals A and R)
3H 12” 3 -40°F (-40°C) ambient temperature, NEMA 4X (IP 67), compact (only for approvals
3F 14” A and R)
4H 16” 5 -40°F (-40°C) ambient temperature, NEMA 4X (IP 67) wall mount housing, for
4F 18” NEMA 6P sensor (only for approvals A and R)
5H 20” 080 Cable for remote
6H 24” 0 Without cable
010 Liner 1 15 foot coil and signal cable
A PFA 2 30 foot coil and signal cable
B PFA, high temperature version (1” to 8” only) 4 Coil and signal cable, specify length in meters (maximum of 50 meters
D Hard rubber, NSF 61 drinking plus warm water approval depending on conductivity of process material)
E PTFE 5 Coil and signal cable, specify length in feet (maximum 160 ft depending
H Hard rubber on conductivity of process material)
N Natural rubber (3” to 24”), only with brush electrodes 090 Cable entries
P Polyurethane, NSF 61 drinking water approval B 1/2” NPT
U Polyurethane (1” to 24”) L 1/2” NPT fieldbus connector
020 Process connection X Not used, sensor only
L Class 150 ANSI B16.5 CS steel A105 flanges 100 Power supply / display
M Class 300 ANSI B16.5 CS steel A105 flanges (not for 8” and larger) A 85 to 260 VAC, WEA, 4-line + touch control (language: EN, ES, FR, IT, NL, PT, DE)
R Class 150 ANSI B16.5 316L SS flanges B 20 to 55 VAC / 20 to 64 VDC, WEA, 4-line + touch control (language: EN, ES, FR, IT,
S Class 300 ANSI B16.5 316L SS flanges (not for 8” and larger) NL, PT, DE)
030 Electrodes C 85 to 260 VAC, SEA, 4-line + touch control (language: EN, JA, ID
0 Measuring, reference and EPD electrodes / 316L SS D 20 to 55 VAC / 20 to 64 VDC, SEA, 4-line + touch control (language: EN, JA, ID)
1 Measuring, reference and EPD electrodes / Alloy C22 E 85 to 260 VAC, EES, 4-line + touch control (language: EN, SV, FI, NO, RU, PL, CS)
2 Measuring, reference and EPD electrodes / Tantalum F 20 to 55 VAC / 20 to 64 VDC, EES, 4-line + touch control (language: EN, SV, FI, NO,
3 Platinum/Rhodium 80/20 measuring electrodes only RU, PL, CS)
5 Platinum/Rhodium 80/20, reference and EPD electrodes G 85 to 260 VAC, CN, 4-line + touch control (language: EN, ZH)
6 Duplex 1.4462 alloy (PFA liner only, 2” to 8”) H 20 to 55 VAC / 20 to 64 VDC, CN, 4-line + touch control (language: EN, ZH)
A Duplex 1.4462 alloy, grounding on flange (PFA liner only, 2” to 8”) 7 85 to 260 VAC, without display, remote configuration only (not for wall mount housing)
G Measuring, reference and EPD electrodes, bullet nose / 316L SS 8 20 to 55 VAC / 16 to 62 VDC, without display, remote configuration only
H Measuring, reference and EPD elecctrodes, bullet nose / Alloy C-22 (not for wall mount housing)
J Measuring, reference and EPD electrodes, Tungsten carbide coated 316L SS X Sensor only (without transmitter, only available up to 8”)
(only for PU liner, not for 1” sensor) 110 Software
K Brush type electrodes / 302 SS, measuring electrodes only, natural rubber A Factory setup, Standard software
liner only B Customized, basic version, volume flow
M Measuring, reference and EPD electrodes / 316L SS, grounding on flange C ECC electrode cleaning circuit
N Measuring, reference and EPD electrodes / Alloy C22, grounding on flange D Customized, ECC electrode cleaning circuit
P Measuring, reference and EPD electrodes / Tantalum, grounding on flange E Factory setup, advanced diagnostics
R Platinum/Rhodium 80/20 measuring electrodes only, grounding on flange F Customized, advanced diagnostics
S Platinum/Rhodium 80/20, reference and EPD electrodes, grounding on flange G Factory setup, advanced diagnostics + ECC cleaning circuit (not for natural rubber liner)
T Bullet nose electrodes / 316L SS, grounding on flange H Customized, advanced diagnostics + ECC cleaning circuit (not for natural rubber liner)
U Bullet nose electrodes / Alloy C22, grounding on flange J Factroy setup, solid content flow (Input/Output modules 4 and 5 only)
040 Calibration K Customized, solid content flow (Input/Ouput modules 4 and 5 only)
A 0.5% L Factory setup, solid content flow + ECC (I/O modules 4 and 5 only)
B 0.2% M Customized, solid content flow + ECC (I/O modules 4 and 5 only)
D 0.5%, 3-point, traceable ISO/IEC17025 N Factory setup, solid content flow + advanced diagnostics (I/O modules 4 and 5 only)
E 0.2%, 3-point, traceable ISO/IEC17025 P Customized, solid content flow + advanced diagnostics (I/O modules 4 and 5 only)
050 Certificates R Factory setup, solid content flow + EEC + advanced diagnostics
1 Standard, no certificate (I/O modules 4 and 5 only)
2 3.1B material certificate for pipe and flanges S Customized, solid content flow + ECC + advanced diagnostics (I/O modules
3 2.3 pressure test certificate (1.5 x PN, 3 minutes) for sensors up to 12” only 4 and 5 only)
4 3.1B material and 2.3 pressure test certificate for sensors up to 12” only 120 Outputs / Inputs
5 CRN approval (ANSI flanges only, not for grounding on flange) Fixed communication boards
8 CRN approval + material certificate + pressure test A Current HART, frequency
(ANSI flanges only, not for grounding on flange) B Current HART, frequency, 2 relays
9 Special version, to be specified H Profibus-PA
060 Approvals K Foundation Fieldbus
A For use in non-hazardous areas Flexible communication boards
R FM non-incendive Class I, Div. 2 / CSA Class I, Div. 2 C Current HART, frequency, 2 relays, flexible module
D Current HART, frequency, relay, and status input, flexible module
L Current HART, 2 relays and status input/output
M Current HART, 2 frequency output, status input
2 Current HART, relay, current, frequency ouputs
3 Current HART, current input, frequency output, current output
4 Current HART, relay, frequency outputs, current input
5 Current HART, status input, current input, frequency output
X Sensor only

* Harsh environment (HE) option is available for process conditions where cool process temperatures in tropical (high
humidity) environments or process fluids which undergo large cyclical temperature variations which can cause high
amounts of moisture that could condense onto the measurement tube.
A0008959

Endress+Hauser 47
USA Canada México Instruments International

Endress+Hauser, Inc. Endress+Hauser Canada Endress+Hauser, México, S.A. de C.V.F Endress+Hauser


2350 Endress Place 1075 Sutton Drive Fernando Montes de Oca 21 Edificio A Piso 3 Instruments International AG
Greenwood, IN 46143 Burlington, ON L7L 5Z8 Fracc. Industrial San Nicolas Kaegenstrasse 2
USA Canada 54030. Tlalnepantla de Baz 4153 Reinach
Estado de México Switzerland
Tel. 317-535-7138 Tel. 905-681-9292
Fax 317-535-8498 800-668-3199 México Tel. +41 61 715 81 00
Sales 888-ENDRESS Fax 905-681-9444 Tel. +52 55-5321-2080 Fax +41 61 715 25 00