HEMATOCRIT/HCT AND CALCULATED HEMOGLOBIN/HB

Hematocrit is determined conductometrically. The measured conductivity, after correction for electrolyte concentration, is inversely related to the hematocrit. See below for information on factors affecting results. Certain substances, such as drugs, may affect analyte levels in vivo.1 If results appear inconsistent with the clinical assessment, the patient sample should be retested using another cartridge. Intended Use The test for hematocrit, as part of the i-STAT System, is intended for use in the in vitro quantification of packed red blood cell volume fraction in arterial, venous, or capillary whole blood. Hematocrit measurements can aid in the determination and monitoring of normal or abnormal total red cell volume status including, but not limited to, conditions such as anemia, erythrocytosis, and blood loss related to trauma and surgery. Contents Each i-STAT cartridge contains one reference electrode (when potentiometric sensors are included in the cartridge configuration), sensors for the measurement of specific analytes, and a buffered aqueous calibrant solution of known conductance that contains known concentrations of analytes and preservatives. Metrological Traceability The i-STAT System test for hematocrit measures packed red blood cell volume fraction in arterial, venous, or capillary whole blood (expressed as the % packed cell volume) for in vitro diagnostic use. Hematocrit values assigned to i-STATs working calibrators are traceable to the U.S. National Committee for Clinical Laboratory Standards (CLSI) H7-A3 procedure for determining packed cell volume by the microhematocrit method2. Further information regarding metrological traceability is available from Abbott Point of Care Inc.. Expected Values Test/Abbreviation Hematocrit/Hct Hemoglobin/Hb Units* %PCV Fraction g/dL g/L mmol/L Reportable Range 10 75 0.10 0.75 3.4 25.5 34 255 2.1 15.8 Reference Range3 38 51** 0.38 0.51 12 17 120 170 7 11

* The i-STAT System can be configured with the preferred units. **The reference ranges for hematocrit and hemoglobin span both female and male populations.

Rev. Date: 01-Aug-11

Art: 714178-00L

To convert a result from %PCV to fraction packed cell volume, divide the %PCV result by 100. For the measurement of hematocrit, the i-STAT System can be customized to agree with methods calibrated by the microhematocrit reference method using either K3EDTA or K2EDTA anticoagulant. Mean cell volumes of K3EDTA anticoagulated blood are approximately 2-4% less than K2EDTA anticoagulated blood.2 While the choice of anticoagulant affects the microhematocrit method to which all hematocrit methods are calibrated, results from routine samples on hematology analyzers are independent of the anticoagulant used. Since most clinical hematology analyzers are calibrated by the microhematocrit method using K3EDTA anticoagulant, the i-STAT System default customization is K3EDTA. The reference range programmed into the analyzer and shown above is intended to be used as a guide for the interpretation of results. Since reference ranges may vary with demographic factors such as age, gender and heritage, it is recommended that reference ranges be determined for the population being tested. Clinical Significance Hematocrit is a measurement of the fractional volume of red blood cells. This is a key indicator of the bodys state of hydration, anemia or severe blood loss, as well as the bloods ability to transport oxygen. A decreased hematocrit can be due to either overhydration, which increases the plasma volume, or a decrease in the number of red blood cells caused by anemias or blood loss. An increased hematocrit can be due to loss of fluids, such as in dehydration, diuretic therapy, and burns, or an increase in red blood cells, such as in cardiovascular and renal disorders, polycythemia vera, and impaired ventilation. Performance Characteristics The typical performance data summarized below was collected in health care facilities by health care professionals trained in the use of the i-STAT System and comparative methods. Precision data were collected in multiple sites as follows: Duplicates of each control fluid were tested in the morning and in the afternoon on five days for a total of 20 replicates. The averaged statistics are presented below. Method comparison data were collected using CLSI guideline EP9-A4. Venous blood samples, collected in lithium heparin Vacutainer tubes, were analyzed in duplicate on the i-STAT System and on the comparative methods for hematocrit within 20 minutes of collection. Deming regression analysis5 was performed on the first replicate of each sample. In the method comparison table, n is the number of specimens in the data set, Sxx and Syy refer to estimates of imprecision based on the duplicates of the comparative and the i-STAT methods respectively, Sy.x is the standard error of the estimate, and r is the correlation coefficient.* Method comparisons will vary from site to site due to differences in sample handling, comparative method calibration and other site specific variables. Interference studies were based on CLSI guideline EP7-P.6
*The usual warning relating to the use of regression analysis is summarized here as a reminder: For any analyte, if the data is collected over a narrow range, the estimate of the regression parameters are relatively imprecise and may be biased. Therefore, predictions made from these estimates may be invalid.5 The correlation coefficient, r, can be used as a guide to assess the adequacy of the comparative method range in overcoming this problem. As a guide, the range of data can be considered adequate if r>0.975.

Precision Data (%PCV) Whole Blood Control Mean Low High 30.0 49.0

SD 0.44 0.50

%CV 1.5 1.0

HB - 2

Art: 714178-00L

Rev. Date: 01-Aug-11

Method Comparison (%PCV) Coulter S Plus n Sxx Syy Slope Intt Sy.x Xmin Xmax r 142 0.50 1.09 0.98 1.78 2.03 18 51 0.952 Nova STAT Profile 5 192 0.46 1.31 1.06 -3.98 2.063 21 50 0.932 Abbott Cell-Dyn 4000 29 0.41 0.77 1.06 -1.42 1.13 19 46 0.993 Sysmex SE9500 29 0.53 0.76 1.11 -4.19 0.98 24 47 0.980

Factors Affecting Results* The measurement of certain blood samples with high erythrocyte sedimentation rates (ESR) may be affected by analyzer angle. While testing blood samples, beginning ninety (90) seconds after the cartridge is inserted, the analyzer should remain level until a result is obtained. A level surface includes running the handheld in the downloader/recharger. Interferent WBC Total Protein Effect Grossly elevated white blood cell counts may increase results. Hematocrit results are affected by the level of total protein as follows: Displayed Result
HCT < 40 %PCV

TP < 6.5 g/dL


Hct decreased by ~1% PCV for each decrease of 1 g/dL TP Hct decreased by ~0.75 % PCV for each decrease of 1 g/dL TP

TP > 8.0 g/dL


Hct increased by ~1% PCV for each increase 1 g/dL TP Hct increased by ~0.75 %PCV for each increase 1 g/dL TP

HCT > 40 % PCV

Total protein levels may be low in neonatal and burn patient populations, as well as in additional clinical populations listed in Statland3. Total protein levels may also be decreased in patients undergoing cardiopulmonary bypass (CPB) or ECMO, and with patients receiving large volumes of saline-based IV fluids. Care should be taken when using hematocrit results from patients with total protein levels below the adult reference range (6.5 to 8 g/dL). The CPB sample type can be used to correct the hematocrit result for the dilutional affect of the pump prime in cardiovascular surgery. The CPB algorithm assumes that cells and plasma are diluted equally and that the pump priming solution has no added albumin or other colloid or packed red blood cells. Since perfusion practices vary, it is recommended that each practice verify the use of the CPB sample type and the length of time in which the CPB sample type should be used during the recovery period. Note that for hematocrit values above 30 %PCV, the CPB correction is 1.5 %PCV; the size of the correction at this level should not impact transfusion decisions. Lipids Sodium Abnormally high lipids may increase results. Interference from lipids will be about two thirds the size of the interference from protein. The sample electrolyte concentration is used to correct the measured conductivity prior to reporting hematocrit results. Factors that affect sodium will therefore also affect hematocrit.

*It is possible that other interfering substances may be encountered. These results are representative and your results may differ somewhat due to test-to-test variation. The degree of interference at concentrations other than those listed might not be predictable.

Rev. Date: 01-Aug-11

Art: 714178-00L

HB - 3

Sample Collection and Handling Erroneous hematocrit results can be obtained by improper sample handling. n Hematocrit results can be affected by the settling of red blood cells in the collection device. The best way to avoid the affect of settling is to test the sample immediately. If there is a delay in testing of a minute or more, the sample must be remixed thoroughly: o o If the sample is in a collection tube, invert the tube gently 10 times. If the sample is in a syringe, roll the syringe between the palms for five seconds in one direction, then roll in a second direction for five seconds, then gently invert repeatedly for five seconds. Note that it may not be possible to adequately mix a blood sample in a 1 mL syringe. Samples from 1 mL syringes should not be used to determine hematocrit if testing is delayed. Discard one or two drops of blood from a syringe before filling a cartridge.

Low hematocrit results can be caused by contamination of flush solutions in an arterial or venous line. o Back flush a line with a sufficient amount of blood to remove intravenous solutions, heparin or medications that may contaminate the sample. Five to six times the volume of the catheter, connectors and needle is recommended.

Cartridge Comparison The performance characteristics of the sensors are equivalent in all cartridge configurations. System difference analysis was performed on 40 patient samples using the i-STAT 6+ and i-STAT E3+ cartridges. In the 1530 %PCV range the average difference was 0.462. In the 3050 %PCV range the average difference was 0.097. Calculated Result for Hemoglobin The i-STAT System provides a calculated hemoglobin result which is determined as follows7: hemoglobin (g/dL) = hematocrit (% PCV) x 0.34 hemoglobin (g/dL) = hematocrit (decimal fraction) x 34 To convert a hemoglobin result from g/dL to mmol/L, multiply the displayed result by 0.621. The calculation of hemoglobin from hematocrit assumes a normal MCHC.

HB - 4

Art: 714178-00L

Rev. Date: 01-Aug-11

References 1. D.S. Young, Effects of Drugs on Clinical Laboratory Tests, 3rd ed. (Washington, DC: American Association of Clinical Chemistry, 1990). 2. CLSI. Procedure for Determining Packed Cell Volume by the Microhematocrit Method; Approved Standard - Third Edition. CLSI document H7-A3 [ISBN 1-56238-413-9]. CLSI, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania 19087-1898 USA, 2000. 3. B.E. Statland, Clinical Decision Levels for Lab Tests (Oradell, NJ: Medical Economic Books, 1987). 4. CLSI. Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline. CLSI document EP9-A [ISBN 1-56238-283-7]. CLSI, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania 19087-1898, USA 1995. 5. P.J. Cornbleet and N. Gochman, Incorrect Least-Squares Regression Coefficients in MethodComparison Analysis, Clinical Chemistry 25:3, 432 (1979). 6. CLSI. Interference Testing in Clinical Chemistry; Proposed Guideline. CLSI document EP7-P [ISBN 156238-020-6]. CLSI, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania 19087-1898, USA 1986. 7. J.D. Bower, P.G. Ackerman and G. Toto, eds., Evaluation of Formed Elements of Blood, in Clinical Laboratory Methods (St. Louis: The C.V. Mosby Company, 1974).

i-STAT is a registered trademark of the Abbott Group of Companies in various jurisdictions. Vacutainer is a registered trademark of Becton Dickinson and Company, Franklin Lakes, NJ USA. Coulter S Plus is a registered trademark of Beckman Coulter Incorporated, Fullerton, CA USA. Cell-Dyn is a registered trademark of Abbott Laboratories, Abbott Park, IL USA. SE9500 is a trademark of Sysmex America Inc., Mundelein, IL USA. STAT Profile is a registered trademark of Nova Biomedical, Waltham, MA USA.

Rev. Date: 01-Aug-11

Art: 714178-00L

HB - 5

Abbott Point of Care Inc. Abbott Park, IL 60064 USA Emergo Europe Molenstraat 15 2513 BH, The Hague The Netherlands Tel: (31)70 345 8570 Fax: (31)70 346 7299

2011 Abbott Point of Care Inc.. All rights reserved. Printed in USA.

Rev. Date: 01-Aug-11

Art: 714178-00L

HB - 6