Calibration Guidelines pH Meter Introduction: You can calibrate the pH meter for three pH levels, but pH7 is most

important. If you have limited time and/or resources, its OK if you just test for pH7 only. If you are testing all three solutions, test pH7 first, and then pH4 and pH10 in any order following. It is best to calibrate the pH meter(s) up to 24-48 hours before the water monitoring event, or less. Record your calibration on the provided sheet for pH7 only, not pH4 or pH10. Check for a BAT symbol flashing on the pH meter, which indicates low batteries. Be sure to record low batteries on the calibration sheet, and notify a San Diego Coastkeeper water monitoring staff member. Be aware that the meters will automatically turn off after 3-5 min. as a battery saving precaution, so dont leave the meter sitting for more than 5 min. or it will turn off. Materials: pH meters you wish to calibrate Calibration solutions pH4, pH7 & pH10 Three 80mL beakers fill with enough calibration solution to submerge the instruments electrodes De-ionized DI H2O for rinsing electrodes Step-by-Step Instructions: 1) Pour the pH7 calibration solution into an 80mL beaker. You want to be able to submerge the entire electrode, so be sure to pour enough to make this possible, but not so much that there is a lot of excess so we dont waste calibration solution. 2) Rinse the electrode with DI H2O and gently shake off excess H2O. 3) Insert the pH meter into the pH7 calibration solution, deep enough so the entire electrode is submerged. 4) Swirl the electrode for a few seconds, and then leave still in the solution for up to 2 minutes. 5) Wait until the pH value stops fluxuating and settles on one value; record that value on the calibration sheet under current reading. 6) Now, you are ready to calibrate. Press and release the CAL button. On the pHTestr3 instruments, the value will start to blink, however on the pHTestr30 instruments the value does not blink. Once settled, press the HOLD button. 7) The value will settle at the calibrated value, which should take no longer than 30 seconds. [The pH meter will automatically settle at the pH value closest to the pre-calibration value. Ex: If the pre-calibration value was 7.25, the pH meter should calibrate to 7.00, or something very close to that. is this correct? Or is it ] The value should then read 7.00. 8) Record the value after calibration on the provided sheet under reading after calibration, even if its not exactly 7.00. You only have to record the pre and post calibration values for the pH7 calibration, not the subsequent pH4 and pH10

San Diego Coastkeeper Calibration Guidelines, courtesy of SDSU

calibrations. 9) Its a good idea to check the calibrated reading again. It should read the exact same calibrated value, even if the calibrated value wasnt exactly 7.00. 10) If the up/down fluxuation is 1.0 or greater (ex: reading 1 = 7.5; reading 2 = 6.4), recalibrate. If the problem persists, record in the notes, dont use the pH meter and notify a San Diego Coastkeeper water monitoring staff member. 11) Rinse with DI H2O if continuing the calibration process with the pH4 and pH10 calibration solutions and follow the same procedure you did for calibrating pH7. 12) Now that the pH meter is calibrated, you need to complete a control chart. To do this, you will record the pH level of the same solution (use the pH7 calibration solution to complete these tests because its a known value) seven times. Be sure to rinse the electrode with DI H2O in between each reading. Once this procedure is complete, plot your results on the provided charts (pH values on the Y axis, reading number on the X axis). This will help us see if the pH meter is taking accurate readings, or if it needs to be replaced. If the readings are going up and down slightly each time (a horizontal zigzag), that is optimal. If the readings are going consistently up or down from readings 1 to 7, the pH meter needs to be recalibrated. If the problem persists you should not use the pH meter, as it probably needs to be replaced. You do not need to complete a control chart for pH4 or pH10, only pH7. 13) Once calibration is complete, store the electrode damp. To do this, rinse the electrode with DI H2O, gently shake the excess H2O off and cap to store. If necessary, you can wipe the pH meter with a cloth, but do not touch the electrode with the cloth. Storing the electrode wet helps to keep the circuit closed and prolong the pH meters life and accuracy. Clean-up: You can safely pour all pH calibration solutions down the sink drain; they are not hazardous. Return the pH calibration solutions and DI H2O squeeze bottle(s) to the appropriate cabinet for storage. Wash all beakers and dry on the drying rack; return to the appropriate cabinet once dry for storage. Make sure all calibrated pH meters are turned off and put in the correct monitoring box(es), or stored all together and clearly dated and labeled calibrated. Make sure youve filled out all the fields on the calibration sheet and return to the clipboard on the wall to the right of the sink.

Conductivity Meter

San Diego Coastkeeper Calibration Guidelines, courtesy of SDSU

Introduction: You can calibrate the conductivity meters for up to 3 standards, but we only calibrate 1 standard (1,413). You should change the calibration solution after every 5 instruments calibrated. It is best to calibrate the conductivity meters on the same day as the water monitoring event, or at least within 24 hours. Materials: Conductivity meters you wish to calibrate Traceable Conductivity Calibration Standard Solution (1,413) One 80mL beaker fill with enough calibration solution to submerge the instruments electrodes De-ionized DI H2O for rinsing electrodes Step-by-Step Instructions: 1) Rinse electrode with DI H2O and gently shake off excess H2O, you can gently towel off the excess water if needed just dont towel off the electrodes. 2) Turn on the instrument and insert the Conductivity meter into the calibration solution, deep enough so the entire electrode is submerged. 3) Swirl the electrode for a few seconds, and then leave still in the solution for up to 2 minutes. 4) Wait until the conductivity value stops fluxuating and settles on one value; record that value on the calibration sheet under reading before calibration. 5) Hold down the CAL button for a few seconds and release. It will immediately start blinking 1,413 a few times, and then will settle on the actual calibrated value. 6) Record the value after calibration on the provided sheet under reading after calibration, it should be exactly 1,413. 7) Now that the conductivity meter is calibrated, you need to complete a control chart. To do this, you will record the conductivity level of the same solution (use the same calibration solution you calibrated the conductivity meter with) seven times. Be sure to rinse the electrode with DI H2O in between each reading. You can test in the same solution because you are testing the same instrument. Once this procedure is complete, plot your results on the provided charts (conductivity values on the Y axis, reading number on the X axis). This will help us see if the conductivity meter is taking accurate readings, or if it needs to be replaced. If the readings are going up and down slightly each time (a horizontal zigzag), that is optimal. If the readings are going consistently up or down from readings 1 to 7, the conductivity meter needs to be re-calibrated. If the problem persists, you should not use the conductivity meter, as it probably needs to be replaced. 8) A 5% swing on either side of 1,413 is generally bad and should be recorded in the notes section of the calibration sheet. However, you will likely see a swing greater than that on the control chart (around 10-15%), as the conductivity meters are highly variable. Still make a note of it on the calibration sheet. 9) Once calibration is complete, store the electrode damp. To do this, you do not need to rinse the electrode with DI H2O, simply take out of the calibration

San Diego Coastkeeper Calibration Guidelines, courtesy of SDSU

solutions and gently shake the excess solution off and cap to store. If necessary, you can wipe the conductivity meter with a cloth, but do not touch the electrode with the cloth. Storing the electrode wet helps to keep the circuit closed and prolong the conductivity meters life and accuracy. Clean-up: You can safely pour the Traceable Conductivity Calibration Standard Solution down the sink drain; it is not hazardous. Return the Traceable Conductivity Calibration Standard Solution and DI H2O squeeze bottle(s) to the appropriate cabinet for storage. Wash all beakers and dry on the drying rack; return to the appropriate cabinet once dry for storage. Make sure all calibrated conductivity meters are put in the correct monitoring box(es), or stored all together and clearly dated and labeled calibrated. Make sure youve filled out all the fields on the calibration sheet and return to the clipboard on the wall to the right of the sink.

Dissolved Oxygen Meter

San Diego Coastkeeper Calibration Guidelines, courtesy of SDSU

Introduction: This meter reads dissolved oxygen (DO) in solution, or the oxygen in the air. The DO in natural/healthy spring water is usually around 8 or 9 (the value is variable depending on several conditions such as aeration, bacteria levels, etc.), and the oxygen in air (20.9 moles per liter). Therefore, we calibrate using the air because its a known value. Its important to look for indicators that the membrane may need to be replaced, or the probe tip is cracked/broken. Those indicators are 1) if the value doesnt settle on one particular value when trying to calibrate and 2) if the value is reading 0 when trying to calibrate or test. You can also check the probe tip each time you calibrate or use it to look for cracks. Be sure to mark this on the calibration sheet and notify a San Diego Coastkeeper water monitoring staff member. It is best to calibrate in an open, well aerated/circulated room 24-48 hours before the water monitoring event, or less. Materials: DO meters you wish to calibrate New membranes (only if you need to change the membranes should only be performed by a San Diego Coastkeeper water monitoring staff member) Step-by-Step Instructions: 1) Make sure the DO meter is reading O2, not mg/L. This will allow you to calibrate using air. 2) Without attaching the probe, turn the meter on. It should read 0. 3) Plug in the probe and take off the cap. Be very careful to remove the cap gently, easing off from the bottom of the cap. Never pull from the top or twist, which are the motions that tend to break the membrane and/or crack the probe tip. Leave probe open to the air in the room for a few minutes (can prop up on something to keep the DO meter stable) until the value settles on one value. 4) Once the value is stable, press the Ox Cal button and release to calibrate. 5) You do not need to complete a control chart, or fill out the calibration sheet. You do need to check off and date the calibration event on the checklist provided to record that the calibration was completed. 6) Once the calibration is complete, turn off the DO meter. Be sure to change the reading units to mg/L to set the DO meter to test water during the upcoming water monitoring event. 7) Unplug probe before storing. Clean-up: Make sure all calibrated DO meters are put in the correct monitoring box(es), or stored all together and clearly dated and labeled calibrated. Make sure youve completed the calibration checklist and return to the clipboard on the wall to the right of the sink. Return the membranes to the appropriate cabinet if necessary.

San Diego Coastkeeper Calibration Guidelines, courtesy of SDSU