Guidelines for Philips Healthcare

Scanner Detector Rows Retrospective Gating Recommended mA


Brilliance 64

64 Retrospective-gated helical BMI < 25: 120kVp, 600-800 mAs (eff) 25 > BMI < 30: 120kVp, 800-1000 mAs (eff) 30 > BMI < 35: 120 kVp, 1000 1500 mAs (or alternatively 140 kVp, 1000 mAs (eff))

Step & Shoot Cardiac BMI < 25: 120 kVp, 150-160 mAs 25 > BMI < 30: 120 kVp, 160-190 mAs 30 > BMI < 32 : 120 kVp, 200210 mAs Step and Shoot not recommended for BMI > 32 140 msec buffer provided 80, 120, 140 64 x 0.625 mm

ECG-pulsing Adaptive pitch Available kV Collimation Pitch Rotation time Default kernel Recommended kV 80, 120, 140 64 x 0.625 mm 0.2 0.4 CB/XCB BMI < 25: 120 25 > BMI < 30: 120 30 > BMI < 35: 120/140

Prospective gating Padding Other notes Automatic exposure control Other techniques Manufacturer's expected dose Other info Vendor-specific mAs Definition

0.4 CB/XCB BMI < 25: 120 25 > BMI < 30: 120 30 > BMI < 35: 120 HR < 65 140 msec buffer provided

iDose (iterative may be available for purchase in late 2010) Step and Shoot Cardiac: 2.4-2.6 mSv; Retrospective-gated helical 10-12 mSv (120 mm coverage; no pulsing used) http://netforum.medical.philips.com/Defa ult.aspx Helical scans: mAs (eff) = mAs normalized to pitch, i.e., mAs(eff) = (mA x rotation time) /pitch factor Step & Shoot Cardiac: mAs = mA x x-ray ON time.

iDose (iterative may be available for purchase in late 2010)

RESCUE Diagnostic Testing Protocol Recommendations Appendix A: Vendor Specific Dose Reduction Guidelines for CT Philips Healthcare Version 2.0 Page 1 of 3
This document cannot be replicated, presented, or published without the specific agreement and consent of the MGH CVIC.

Scanner Detector Rows Retrospective Gating Recommended mA

Brilliance iCT

128 rows, 256 slice scanner Retrospective-gated helical BMI < 25: 100kVp, 600-800 mAs (eff); 120kVp, 600-800 mAs (eff) 25 > BMI < 30: 120kVp, 8001000 mAs (eff) 30 > BMI < 35: 120kVp, 10001500 mAs (eff); 140kVp, 1000 mAs (eff); alternatively 0.33s rotation time Step & Shoot Cardiac BMI < 25: 100 kVp, 180-200 mAs; 120kVp, 150-160 mAs 25 > BMI < 30: 120 kVp, 160190 mAs 30 > BMI < 35 : 120 kVp, 200210 mAs; alternatively 0.33s rotation time

ECG-pulsing Adaptive pitch Available kV Collimation Pitch Rotation time Default kernel Recommended kV

80, 100, 120, 140 128 x 0.625 mm 0.16 0.27 CB/XCB BMI < 25: 100/120 25 > BMI < 30: 120 30 > BMI < 35: 120/140

80, 100, 120, 140 128 x 0.625 mm 0.27 CB/XCB BMI < 25: 100/120 25 > BMI < 30: 120 30 > BMI < 35: 120 HR < 75 3% &5% phase tolerance provided

Prospective gating Padding Other notes Automatic exposure control Other techniques

Manufacturer's expected dose

Dynamic helical collimation (Eclipse DoseRight Collimator) available and enabled on all helical scans Step & Shoot Cardiac: 1.4 1.6 mSv (100 kV); 2.4 2.6 mSv (120 kV); Retrospective-gated helical: 6.5 8.5 mSv (100 kV); 10 12 mSv (120 kV) (120 mm coverage no pulsing) Helical scans: mAs (eff) = mAs normalized to pitch, i.e., mAs(eff) = (mA x rotation time) /pitch factor Step & Shoot Cardiac: mAs = mA x xray ON time.

Other info Vendor-specific mAs Definition

RESCUE Diagnostic Testing Protocol Recommendations Appendix A: Vendor Specific Dose Reduction Guidelines for CT Philips Healthcare Version 2.0 Page 2 of 3
This document cannot be replicated, presented, or published without the specific agreement and consent of the MGH CVIC.

Scanner Detector Rows Retrospective Gating Recommended mA

Brilliance iCT-SP

64 rows, 128-slice scanner Retrospective-gated helical BMI < 25: 100kVp, 600-800 mAs (eff); 120kVp, 600-800 mAs (eff) 25 > BMI < 30: 120kVp, 8001000 mAs (eff) 30 > BMI < 35: 120kVp, 10001500 mAs (eff); 140kVp, 1000 mAs (eff); alternatively 0.33s rotation time Step & Shoot Cardiac BMI < 25: 100 kVp, 180-200 mAs ; 120 kVp, 150-160 mAs) 25 > BMI < 30: 120 kVp, 160190 mAs 30 > BMI < 35: 120 kVp, 200 210 mAs.; alternatively 0.33 s rotation time

ECG-pulsing Adaptive pitch Available kV Collimation Pitch Rotation time Default kernel Recommended kV

64 x 0.625 mm 0.16 0.3 (0.27 optional) CB/XCB BMI < 25: 100/120 25 > BMI < 30: 120 30 > BMI < 35: 120/140

64 x 0.625 mm 0.3 (0.27 optional) CB/XCB BMI < 25: 100/120 25 > BMI < 30: 120 30 > BMI < 35: 120 HR < 75 3% &5% phase tolerance provided

Prospective gating Padding Other notes Automatic exposure control Other techniques Manufacturer's expected dose Approximately the same as in Brilliance iCT

Other info Vendor-specific mAs Definition

Helical scans: mAs (eff) = mAs normalized to pitch, i.e., mAs(eff) = (mA x rotation time) /pitch factor Step & Shoot Cardiac: mAs = mA x xray ON time.

RESCUE Diagnostic Testing Protocol Recommendations Appendix A: Vendor Specific Dose Reduction Guidelines for CT Philips Healthcare Version 2.0 Page 3 of 3
This document cannot be replicated, presented, or published without the specific agreement and consent of the MGH CVIC.