Elastic Modulus Measurement
Application:
Measurement on Young's Modulus and Shear Modulus of Elasticity, and Poisson's ratio, in nondispersive isotropic engineering materials.

Background:
Young's Modulus of Elasticity is defined as the ratio of stress (force per unit area) to corresponding strain (deformation) in a material under tension or compression. Shear Modulus of Elasticity is similar to the ratio of stress to strain in a material subjected to shear stress. Poisson's Ratio is the ratio of transverse strain to corresponding axial strain on a material stressed along one axis. These basic material properties, which are of interest in many manufacturing and research applications, can be determined through computations based on measured sound velocities and material density. Sound velocity can be easily measured using ultrasonic pulseecho techniques with appropriate equipment. The general procedure outlined below is valid for any nondispersive material and sample geometry (i.e., velocity does not change with frequency). This includes most metals, ceramics, and glasses as long as cross sectional dimensions are not close to the test frequency wavelength. Rubber usually cannot be characterized ultrasonically because of its high dispersion and nonlinear elastic properties. In the case of anisotropic materials, elastic properties vary with direction, and so does longitudinal and/or shear wave sound velocity. Generation of a full matrix of elastic moduli in anisotropic specimens typically requires six different sets of ultrasonic measurements.

Equipment:
The technique requires ultrasonic pulser-receiver such as a 5072PR or 5077PR, an ultrasonic thickness gage such as a Model 35DL or 38DL PLUS, or a flaw detector with velocity measurement capability such as the EPOCH series instruments. It also requires two transducers appropriate to the material being tested, for pulse-echo sound velocity measurement in longitudinal and shear modes, Commonly used transducers include an M112 or V112 broadband longitudinal wave transducer (10 MHz) and a V156 normal incidence shear wave transducer (5 MHz). These work well for many common metal and fired ceramic samples. Different transducers will be required for very thick, thin, or highly attenuating samples. The test sample may be of any geometry that permits clean pulse/echo measurement of sound transit time through a section on thickness. Ideally this would be a sample at least 0.5 in. (12.5 mm) thick, with smooth parallel surfaces and a width or diameter greater than the diameter of the transducer being used. Caution must be used when testing narrow specimens due to possible edge effects that can affect measured pulse transit time.

Procedure:
Measure the longitudinal and shear wave sound velocity of the test piece using the appropriate transducers. A Model 35DL or 38DL PLUS thickness gage can provide a direct readout of material velocity based on an entered sample thickness, and an EPOCh series flaw detector can measure velocity through a velocity calibration procedure. In either case, follow the recommended procedure for velocity measurement as described in the instrument's operating manual. If using a pulser/receiver, simply record the round-trip transit time through an area of known thickness with both longitudinal and shear wave transducers, and compute:

Convert units as necessary to obtain velocities expressed as inches per second or centimeters per second. (Time will usually have been measured in microseconds, so multiply in/uS or cm/uS by 106 to obtain in/S or cm/s.) The velocities thus obtained may be inserted into the following equations.

English (Change)

Home About Us

Site Map Global Site

Inspection & Measurement Systems

SEARCH

Products

Applications

Knowledge

News

Events

Support

Contact Us

Home > Applications > Application Notes

Note on units:If sound velocity is expressed in cm/S and density in g/cm3, then Young's modulus will be expressed in units of dynes/cm2. If English units of in/S and lbs/in3 are used to compute modulus in pounds per square inch (PSI), remember the distinction between "pound" as a unit of force versus a unit of mass. Since modulus is expressed as a force per unit area, when calculating in English units it is necessary to multiply the solution of the above equation by a mass/force conversion constant of (1 / Acceleration of Gravity) to obtain modulus in PSI. Alternately, if the initial calculation is done in metric units, use the conversion factor 1 psi = 6.89 x 104 dynes/cm 2. Another alternative is to enter velocity in in/S, density in g/cm 3, and divide by a conversion constant of 1.07 x 104 to obtain modulus in PSI.

For shear modulus simply multiply the square of the shear wave velocity by the density. Again, use units of cm/S and g/cm 3to obtain modulus in dynes/cm 2or English units of in/S and lbs/in 3and multiply the result by the mass/force conversion constant.

Bibliography For further information on ultrasonic measurement of elastic modulus, consult the following: 1. Moore, P. (ed.), Nondestructive Testing Handbook, Volume 7, American Society for Nondestructive Testing, 2007, pp. 319-321. 2. Krautkramer, J., H. Krautkramer, Ultrasonic Testing of Materials, Berlin, Heidelberg, New York 1990 (Fourth Edition), pp. 13-14, 533534.

Products used for this application


35
The pocket-sized 35 is a handheld ultrasonic thickness gage for precision measurements on metal, plastic, rubber, glass, and composites using single element transducers. Wide thickness range from 0.08 mm to 635.0 mm (0.0030 in. to 25.0 in.) Optional live Waveform and Adjust Mode.

38DL PLUS
The 38DL PLUS is an advanced ultrasonic thickness gage. Uses dual element transducer for internally corroded applications, features include THRU-COAT and Echo-to-Echo. Uses single element transducers for very precise thickness measurements of thin, very thick, or multilayer materials.

5072PR
The 5072PR manually controlled ultrasonic pulser-receiver is for both conventional and high frequency applications. 35 MHz (-3 dB) bandwidth and spike pulser is ideal for general testing. Optimal broadband excitation resulting in superior time domain recovery especially between 15-30 MHz.

5077PR
The 5077PR manually controlled ultrasonic pulser-receiver with 35 MHz (-3 dB) bandwidth and square wave pulser-receiver is ideal for maximizing response in scattering materials. The square wave pulser is particularly advantageous when testing with transducers of 10 MHz or lower.

Advanced NDT Solutions


Weld Inspection Solutions Corrosion Inspection Solutions Composite Inspection Solutions Tube Inspection Solutions Industrial Scanners Aerospace Inspection Solutions

Thickness Gages
Precision Thickness Gages Corrosion Thickness Gages

Videoscopes, Borescopes
Industrial Videoscopes Industrial Fiberscope Industrial Rigid Borescopes Light Sources Turning Tools

Applications
Application Notes Applications Support

Microscope Solutions
Laser Confocal Microscopes Opto-digital Microscopes Semiconductor & Flat Panel Display Inspection Microscopes Upright Metallurgical Microscopes Inverted Metallurgical Microscopes Modular Microscopes Polarizing Microscopes

Support
Service Centers PDF Library Video Gallery Software Downloads Training Academy Obsolete Products ISO Certifications MSDS Datasheets

High Speed Video Cameras


i-SPEED LT i-SPEED 2 i-SPEED TR i-SPEED 3 i-SPEED FS

Flaw Detectors
Ultrasonic Flaw Detectors Phased Array Eddy Current Products Eddy Current Array Products

BondMaster Probes and Transducers Pulser-Receivers

Measuring Microscopes Stereo Microscopes Objective Lenses Digital Cameras Image Analysis Software

i-SPEED PL i-SPEED Comparison i-SPEED Video Examples

Integrated Inspection Systems


Bar Inspection Systems Tube Inspection Systems NDT Systems Instrumentation

Innov-X XRF and XRD Analyzers


Alloy QA/QC Petrochemical & Power Precious Metals Scrap Recycling, Metal, Glass, & Plastics Geochemistry & Mining Oil & Concentrate Analyses Environmental & IH Consumer Products Safety Archaeometry, Academic & R&D Pharmaceutical & Nutraceutical Government & Security Applications Solutions Key

Optical Metrology
Laser Confocal Microscopes Opto-digital Microscopes Measuring Microscopes

Copyright 2011 OLYMPUS CORPORATION, All rights reserved. Terms Of Use | Privacy Statement

Você também pode gostar