Features
Precise bi-directional and highly dynamic flow measurement with the non-intrusive clamp-on technology High precision at fast and slow flow rates, high temperature and zero point stability ATEX certified FLUXUS G800 is presented in a flameproof housing (IP66) and can be operated by a magnet pen without opening the housing All stainless steel and seawater resistant FLUXUS G801 is ATEX certified and thus suited for offshore applications Automatic loading of calibration data and transducer detection for a fast and easy set-up (less than 5 min), providing precise and long-term stable results User-friendly design Transducers available for a wide range of inner pipe diameters ( 7...1600 mm ) and fluid temperatures ( -40...+200 C) ATEX, IECEx approved transducers for hazardous areas available Measurement is unaffected by gas density, viscosity, composition, dust, humidity, temperature or pressure
FLUXUS G800
Applications
Designed for industrial use in harsh environments, in gas processing and natural gas extraction, chemical industry and in the petroleum industry. Practical applications: Measurement on natural gas pipelines and in natural gas storage installations Measurement of synthesized gas and injection gas Measurement for the gas supply industry
FLUXUS G801
TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
FLUXUS G80x
Technical Specification
Table of Contents
Function ........................................................................................................................................................... 3 Measurement Principle ..................................................................................................................................... 3 Calculation of Volumetric Flow Rate ................................................................................................................. 3 Number of Sound Paths.................................................................................................................................... 4 Typical Measurement Setup ............................................................................................................................. 5 Standard Volumetric Flow Rate ........................................................................................................................ 5 Flow Transmitter ............................................................................................................................................. 6 Technical Data .................................................................................................................................................. 6 Dimensions ....................................................................................................................................................... 9 Wall and 2 " Pipe Mounting Kit........................................................................................................................ 10 Terminal Assignment ...................................................................................................................................... 11 Transducers................................................................................................................................................... 13 Transducer Selection ...................................................................................................................................... 13 Transducer Order Code .................................................................................................................................. 16 Technical Data ................................................................................................................................................ 17 Transducer Mounting Fixture ...................................................................................................................... 23 Coupling Materials for Transducers............................................................................................................ 25 Damping Mats (optional) .............................................................................................................................. 26 Connection Systems ..................................................................................................................................... 28 Transducer Cable............................................................................................................................................ 28 Junction Box ................................................................................................................................................. 29 Technical Data ................................................................................................................................................ 29 Dimensions ..................................................................................................................................................... 29 2 " Pipe Mounting Kit (optional)....................................................................................................................... 30 Terminal Assignment ...................................................................................................................................... 30
TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
Technical Specification
FLUXUS G80x
Function
Measurement Principle
In order to measure the flow of a medium in a pipe, ultrasonic signals are used, employing the transit time difference principle. Ultrasonic signals are emitted by a transducer installed on the pipe and received by a second transducer. These signals are emitted alternately in the flow direction and against it. As the medium in which the signals propagate is flowing, the transit time of the ultrasonic signals in the flow direction is shorter than against the flow direction. The transit time difference, t, is measured and allows the flowmeter to determine the average flow velocity along the propagation path of the ultrasonic signals. A flow profile correction is then performed in order to obtain the area averaged flow velocity, which is proportional to the volumetric flow rate. Two integrated microprocessors control the entire measuring process. This allows the flowmeter to remove disturbance signals, and to check each received ultrasonic wave for its validity which reduces noise.
J
J
TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
FLUXUS G80x
Technical Specification
a - transducer distance
TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
Technical Specification
FLUXUS G80x
transmitter
RS485 (optional)
analog/binary outputs
Example of a gas flow measurement in reflection arrangement with standard volumetric flow rate output
The operational pressure p and the operational temperature T of the medium will be entered directly as fixed values into the transmitter. The gas compressibility coefficient K of the gas is entered in the transmitter: as fixed value or as approximation according to e.g. AGA8 or GERG
TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
FLUXUS G80x
Technical Specification
Flow Transmitter
Technical Data
FLUXUS G800C24 G800 G800LC24 G800L G800P G800LP explosion proof field device G801 G801P G801C24
design
measurement measurement principle flow velocity repeatability medium temperature compensation accuracy volumetric flow rate flow transmitter power supply
transit time difference correlation principle 0.01...35 m/s, depending on pipe diameter 0.15 % of reading 0.01 m/s all acoustically conductive gases, e.g. nitrogen, air, oxygen, hydrogen, argon, helium, ethylene, propane corresponding to the recommendations in ANSI/ASME MFC-5.1-2011 1...3 % of reading 0.01 m/s depending on application 0.5 % of reading 0.01 m/s with field calibration 100...240 V/50...60 Hz or 20...32 V DC or on request: 11...16 V DC < 10 W 1, optional: 2 24 V DC 10 % 100...240 V/50...60 Hz or 20...32 V DC or on request: 11...16 V DC < 10 W 1, optional: 2 24 V DC 10 %
power consumption number of flow measuring channels signal attenuation measuring cycle (1 channel) response time housing material
<4W 1, optional: 2
<4W 1, optional: 2
degree of protection according to IEC/EN 60529 dimensions weight fixation ambient temperature display menu language
0...100 s, adjustable 100...1000 Hz 1 s (1 channel), option: 70 ms cast aluminum stainless steel 316/316L (1.4401, 1.4404, 1.4432) G800, G800P, G800C24: powder coated G800L, G800LP, G800LC24: special offshore coating IP66 IP66 see dimensional drawing 6 kg -20...+60 C wall mounting, 2 " pipe mounting -20...+50 C -20...+50 C 2 x 16 characters, dot matrix, backlight English, German, French, Dutch, Spanish 8.5 kg -20...+50 C
TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
Technical Specification
FLUXUS G80x
FLUXUS
G800 G800L G800P G800LP 1 G800: 0637 II2G Ex d e IIC T6 Ta -20...+60 C G800L: 0637 II2G Ex d e IIB T6 Ta -20...+60 C G800P: 0637 II2G Ex d e IIC T4 Ta -20...+60 C
G800C24 G800LC24
G801 G801P
G801C24
1 G800C24: 0637 II2G Ex d e [ib] IIC T4 Ta -20...+50 C G800LC24: 0637 II2G Ex d e [ib] IIB T4 Ta -20...+50 C
1 G801: 0637 II2G Ex d e IIC T6 Ta -20...+50 C G801P: 0637 II2G Ex d e IIC T4 Ta -20...+50 C
G800LP: 0637 II2G Ex d e IIB T4 Ta -20...+60 C IBExU01ATEX1064 IBExU01ATEX1064 IBExU05ATEX1078 IBExU05ATEX1078 electronics compartment: electronics compartment: electronics compartment: electronics compartment: flameproof enclosure flameproof enclosure flameproof enclosure flameproof enclosure connection compartment: connection compartment: connection compartment: connection compartment: increased safety increased safety increased safety increased safety output circuits: intrinsic safety Um = 250 V AC intrinsically safe outputs: Ui = 28.2 V Pi = 0.76 W Li, Ci negligible output circuits: intrinsic safety Um = 250 V AC intrinsically safe outputs: Ui = 28.2 V Pi = 0.76 W Li, Ci negligible
measuring functions physical quantities totalizer calculation functions diagnostic functions data logger loggable values capacity communication interface
operating volumetric flow rate, standard volumetric flow rate, mass flow rate, flow velocity volume, mass average, difference, sum (2 measuring channels necessary) sound speed, signal amplitude, SNR, SCNR, standard deviation of amplitudes and transit times all physical quantities, totalized values and diagnostic values > 100 000 measured values - process integration (optional): RS485 (emitter) or Modbus/RS485 (emitter) or HART - diagnosis: RS2321 - diagnosis: RS2321 - process integration (optional): RS485 (emitter) or Modbus/RS485 (emitter) or HART - diagnosis: RS2321 - diagnosis: RS2321
serial data kit (optional) software (all Windows ver- - FluxData: download of measurement data, graphical presentation, conversion to other formats (e.g. for Excel) sions) - FluxKoef: creating medium data sets able adapter
1
connection of the interface RS232 outside of explosive atmosphere (housing cover open)
TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
FLUXUS G80x
Technical Specification
FLUXUS
G800C24 G800LC24
G801 G801P
G801C24
outputs (optional) number The outputs are galvanically isolated from the transmitter. current output: 1...2 current output: 1 current output: 1...2 and and and binary output (open binary output (open binary output (open collector (A)): 1...4 collector (A)): 1 collector (A)): 1...4 or current output: 1...2 and binary output (open collector (A)): 1...2 and binary output (Reed relays): 0...2 or frequency output: 1 and binary output (open collector (B)): 1 current output I1, I2 - range - accuracy - active output or current output: 1...2 and binary output (open collector (A)): 1...2 and binary output (Reed relays): 0...2 or frequency output: 1 and binary output (open collector (B)): 1 current output current output: 1 and binary output (open collector (A)): 1
- passive output
0/4...20 mA 4...20 mA 0/4...20 mA 4...20 mA 0.1 % of reading 15 A 0.1 % of reading 15 A 0.1 % of reading 15 A 0.1 % of reading 15 A G800, G800L: G801: Rext < 500 Rext < 500 G800P, Uext = 4...28.2 V, Uext = 4...28.2 V, G800LP: depending on Rext depending on Rext G801P: Rext < 1 k Rext < 1 k Uext = 4...26.4 V, Uext = 4...26.4 V, depending on Rext depending on Rext intrinsic safety intrinsic safety Rext < 1 k Rext < 1 k 4...20 mA Uext = 10...24 V 0...5 kHz 30 V/100 mA optional: 8.2 V DIN EN 60947-5-6 (NAMUR) 48 V/0.25 A 24 V/4 mA 30 V/100 mA limit, change of flow direction or error 0.01...1000 units 1...1000 ms 4...20 mA Uext = 10...24 V frequency output (optional) (G800P, G801P) 0...5 kHz 30 V/100 mA optional: 8.2 V DIN EN 60947-5-6 (NAMUR) binary output 48 V/0.25 A 24 V/4 mA 24 V/4 mA intrinsic safety 30 V/100 mA limit, change of flow direction or error 0.01...1000 units 1...1000 ms limit, change of flow direction or error 0.01...1000 units 1...1000 ms -
Reed relay open collector (A) open collector (B), optional binary output as alarm output - functions binary output as pulse output - pulse value - pulse width
24 V/4 mA intrinsic safety limit, change of flow direction or error 0.01...1000 units 1...1000 ms
TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
Technical Specification
FLUXUS G80x
Dimensions
FLUXUS G800
301.5 259 thread, cable gland: 2x 3/4 NPT 195 137
202
FLUXUS G801
277 189
132
292
in mm
TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
178
FLUXUS G80x
Technical Specification
FLUXUS G801
10
TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
Technical Specification
FLUXUS G80x
Terminal Assignment
FLUXUS G800, G800L, G800P (transmitter without frequency output), G800LP
upper housing, back view
A R S
A V S
B R S
B V S
A R
A V
B V
B R
outputs
1 0 1 1 1 2 1 3 1 4 1 5 1 2 3 4 5 6 7 8 9
power supply outputs equipotential bonding terminal lower housing, front view transducers power supply
L 1 /L + N /L P E
power supply
AC terminal PE N L1 earth neutral phase connection PE LL+ terminal earth + DC connection
transducers
terminal AV AVS ARS AR cable gland measuring channel A connection transducer , signal transducer , internal shield transducer , internal shield transducer , signal external shield terminal BV BVS BRS BR cable gland measuring channel B connection transducer , signal transducer , internal shield transducer , internal shield transducer , signal external shield
outputs
terminal 1(-), 2(+) 3(-), 4(+) 5(-), 6(+) 7(-), 8(+) 9(a), 10(b) 11(a), 12(b) 13(B-), 14(A+) connection current output I1 current output I2 (optional) binary output B1 (open collector) binary output B2 (open collector, optional) binary output B3 (open collector or Reed relay, optional) binary output B4 (open collector or Reed relay, optional) RS485 (optional)
TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
11
FLUXUS G80x
Technical Specification
transducers power supply outputs equipotential bonding terminal power supply lower housing, front view transducers
power supply
AC (G800P, G801P) connection earth neutral phase DC terminal PE LL+ earth + connection
terminal PE N L1
transducers
terminal AV AVS ARS AR S cable gland measuring channel A connection transducer , signal transducer , internal shield transducer , internal shield transducer , signal not connected external shield terminal BV BVS BRS BR S cable gland measuring channel B connection transducer , signal transducer , internal shield transducer , internal shield transducer , signal not connected external shield
outputs
color of terminals terminal 1(-), 2(+) 5(-), 6(+) G800C24, G800LC24, G801C24 blue (intrinsic safety) current output I1 binary output B1 (open collector) G800P, G801P green connection frequency output F1 binary output B1 (open collector)
12
TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
Technical Specification
FLUXUS G80x
Transducers
Transducer Selection
Step 1a Select a Lamb wave transducer:
transducer order code GLG GLH GLK GLM GLP GLQ 0.5...1 5 10 15 20 25 pipe wall thickness [mm] 2 1...3 4 5 7 9 11 15 23
Step 1b If the pipe wall thickness is not in the range of the Lamb wave transducers, select a shear wave transducer:
transducer order code GSG GSK GSM GSP 1.5 2.5 3 5 10 15 20 25 pipe wall thickness [mm] recommended possible 5 5 9 14 23
Step 2 inner pipe diameter d dependent on the flow velocity v of the medium in the pipe The transducers are selected from the characteristics (see next page). Lamb wave transducers are selected from the left column, shear wave transducers from the right column. Lamb wave transducers: If the values d and v are not in the range, the diagonal arrangement with 1 sound path may be used, i.e. the same characteristics can be used with doubling the inner pipe diameter. If the values are still not in the range, shear waves transducers regarding the pipe wall thickness have to be selected in step 1b.
TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
13
FLUXUS G80x
Technical Specification
GSG
10
15
20
25
30
35
vms
10
15
20
25
30
35
vms
GLH
10
15
20
25
30
35
GLK
GSK
10
15
20
25
30
35
vms
0 D mm
10
15
20
25
30
35
vms
GLM
GSM
10
15
20
25
30
35
vms
0
D mm
10
15 20 GDP
25
30
35
vms
GSP
GLP
10
15
20
25
30
35
vms
10
15
20
25
30
35
vms
GLQ
10
15
20
25
30
35
vms
inner pipe diameter and max. flow velocity for a typical application with natural gas, nitrogen, oxygen in reflection arrangement with 2 sound paths (Lamb wave transducers)/1 sound path (shear wave transducers)
14
TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
Technical Specification
FLUXUS G80x
shear wave transducer transducer medium pressure1 [bar] order code metal pipe plastic pipe min. min. extended min. GSG 30 20 1 GSK GSM GSP 30 30 30 20 20 20 1 1 1
15 15 (d > 120 mm) 10 (d < 120 mm) 10 (d > 60 mm) 5 (d < 60 mm) 10 (d > 35 mm) 5 (d < 35 mm) 10 (d > 15 mm) 5 (d < 15 mm)
10
depending on application, typical absolute value for natural gas, nitrogen, compressed air
Example
step 1 2 pipe wall thickness selected transducer inner pipe diameter max. flow velocity selected transducer mm mm m/s 12 GLG or GLH 800 15 GLG 12 GLG or GLH 600 15 GLG or GLH 12 GLG or GLH 800 30 values not in the range of the characteristics, but by using diagonal arrangement with 1 sound path, the inner pipe diameter in the characteristics is doubled: GLG 17 GLG 30 GS 300 15 GSK
bar
17 GLG
17 GLG or GLH influence of acoustic noise is reduced with increased transducer frequency, thus recommended: GLH
35 GSK
Step 4 for the characters 4...11 of the transducer order code (ambient temperature, explosion protection, connection system, extension cable) see page 16 Step 5 for the technical data of the selected transducer see page 17 et seqq.
TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
15
FLUXUS G80x
Technical Specification
explosion protection
GL GS G H K M P Q N E A1 TS XXX
option
set of ultrasonic flow transducers for gas measurement, Lamb wave set of ultrasonic flow transducers for gas measurement, shear wave 0.2 MHz 0.3 MHz (Lamb wave only) 0.5 MHz 1 MHz 2 MHz 4 MHz (Lamb wave only) normal temperature range extended temperature range (shear wave transducers with transducer frequency M, P) ATEX zone 1/IECEx zone 1 direct connection or connection via junction box cable length in m, for max. length of extension cable see page 28 connection system TS: 0 m: without junction box > 0 m: with junction box JB01 IP68 OS degree of protection IP68 housing with stainless steel 316 Lamb wave transducer 0.5 MHz, normal temperature range, ATEX zone 1/IECEx zone 1, connection system TS with with junction box JB01 and extension cable 30 m /
example GL K N A1 TS 030
16
TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
Technical Specification
FLUXUS G80x
Technical Data
Shear Wave Transducers (zone 1)
technical type order code transducer frequency medium pressure1 min. extended min. GDG1N81 GSG-NA1TS GSG-NA1TS/OS MHz 0.2 bar bar metal pipe: 20 metal pipe: 30 plastic pipe: 1 250 380 810 1100 14 PEEK with stainless steel cap 304 (1.4301), option OS: 316L (1.4404) PEEK IP65 GDK1N81 GSK-NA1TS GSK-NA1TS/OS 0.5 metal pipe: 20 metal pipe: 30 plastic pipe: 1 70 80 500 720 5 PEEK with stainless steel cap 304 (1.4301), option OS: 316L (1.4404) PEEK IP65 GDM2N81 GSM-NA1TS GSM-NA1TS/OS 1 metal pipe: 20 metal pipe: 30 plastic pipe: 1 30 40 80 120 2.5 PEEK with stainless steel cap 304 (1.4301), option OS: 316L (1.4404) PEEK IP65 GDP2N81 GSP-NA1TS GSP-NA1TS/OS 2 metal pipe: 20 metal pipe: 30 plastic pipe: 1 15 20 40 60 1.5 PEEK with stainless steel cap 304 (1.4301), option OS: 316L (1.4404) PEEK IP65
inner pipe diameter d2 min. extended mm min. recommended mm max. recommended mm max. extended mm pipe wall thickness min. mm max. mm material housing contact surface degree of protection according to IEC/ EN 60529 transducer cable type length dimensions length l width b height h dimensional drawing
m mm mm mm
1699 5 129.5 51 67
1699 4 64 32 40.5
1699 4 64 32 40.5
ambient temperature min. C -40 max. C +130 temperature x compensation explosion protection gas: 2G dust: 2D category zone 1 21 explosion protection temperature (pipe surface) min. C -55 C +180 A max. T marking II2G 0637 E II2D Ex e q IIC T6...T3 Gb X / Ex tb IIIC TX Db I certification ATEX IBExU07ATEX1168 X E certification IECEx IECEx IBE 08.0007X C type of protection gas: increased safety, E powder filling x dust: protection by enclosure transducer x mounting fixture necessary
1 2
-40 +130 x
-40 +130 x
-40 +130 x
dust: 2D 21
dust: 2D 21
dust: 2D 21
II2G II2D Ex e q IIC T6...T3 Gb Ex tb IIIC TX Db IBExU07ATEX1168 X IECEx IBE 08.0007X gas: increased safety, powder filling dust: protection by enclosure x 0637
II2G II2D Ex e q IIC T6...T3 Gb Ex tb IIIC TX Db IBExU07ATEX1168 X IECEx IBE 08.0007X gas: increased safety, powder filling dust: protection by enclosure x 0637
II2G II2D Ex e q IIC T6...T3 Gb Ex tb IIIC TX Db IBExU07ATEX1168 X IECEx IBE 08.0007X gas: increased safety, powder filling dust: protection by enclosure x 0637
depending on application, typical absolute value for natural gas, nitrogen, compressed air shear wave transducer: typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on request pipe diameter min. recommended/max. recommended/max. extended: in diagonal arrangement and for a flow velocity of 15 m/s
17
TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
FLUXUS G80x
Technical Specification
inner pipe diameter d2 min. extended mm min. recommended mm max. recommended mm max. extended mm pipe wall thickness min. mm max. mm material housing contact surface degree of protection according to IEC/ EN 60529 transducer cable type length dimensions length l width b height h dimensional drawing
m mm mm mm
2550 12 72 32 46
2550 12 72 32 46
ambient temperature min. C -40 max. C +100 temperature x compensation explosion protection category gas: 2G dust: 2D zone 1 21 explosion protection temperature (pipe surface) min. C -55 A max. C +180 T marking II2G 0637 E II2D X Ex q IIC T6...T3 Gb / Ex tb IIIC TX Db I IBExU07ATEX1168 X E certification ATEX IECEx IBE 08.0007X C certification IECEx gas: powder filling E type of protection dust: protection by x enclosure transducer x mounting fixture necessary
1 2
-40 +100 x
-40 +100 x
-40 +100 x
dust: 2D 21
dust: 2D 21
dust: 2D 21
II2G II2D Ex q IIC T6...T3 Gb Ex tb IIIC TX Db IBExU07ATEX1168 X IECEx IBE 08.0007X gas: powder filling dust: protection by enclosure x 0637
II2G II2D Ex q IIC T6...T3 Gb Ex tb IIIC TX Db IBExU07ATEX1168 X IECEx IBE 08.0007X gas: powder filling dust: protection by enclosure x 0637
II2G II2D Ex q IIC T6...T3 Gb Ex tb IIIC TX Db IBExU07ATEX1168 X IECEx IBE 08.0007X gas: powder filling dust: protection by enclosure x 0637
depending on application, typical absolute value for natural gas, nitrogen, compressed air shear wave transducer: typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on request pipe diameter min. recommended/max. recommended/max. extended: in diagonal arrangement and for a flow velocity of 15 m/s test conditions: 3 months/2 bar (20 m)/20 C
18
TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
Technical Specification
FLUXUS G80x
inner pipe diameter d2 min. extended mm min. recommended mm max. recommended mm max. extended mm pipe wall thickness min. mm max. mm material housing contact surface degree of protection according to IEC/ EN 60529 transducer cable type length dimensions length l width b height h dimensional drawing
m mm mm mm
6111 4 64 32 40.5
6111 4 64 32 40.5
ambient temperature min. C -30 max. C +200 temperature x compensation explosion protection category gas: 2G dust: 3D zone 1 22 explosion protection temperature (pipe surface) min. C -45 C +225 A max. T marking II2G 0637 E II2D Ex e q IIC T6...T2 Gb X / Ex tb IIIA TX Db I certification ATEX IBExU07ATEX1168 X E certification IECEx IECEx IBE 08.0007X C type of protection gas: increased safety, E powder filling x dust: protection by enclosure transducer x mounting fixture necessary
1 2
-30 +200 x
dust: 3D 22
II2G II2D Ex e q IIC T6...T2 Gb Ex tb IIIA TX Db IBExU07ATEX1168 X IECEx IBE 08.0007X gas: increased safety, powder filling dust: protection by enclosure x 0637
depending on application, typical absolute value for natural gas, nitrogen, compressed air shear wave transducer: typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on request pipe diameter min. recommended/max. recommended/max. extended: in diagonal arrangement and for a flow velocity of 15 m/s
TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
19
FLUXUS G80x
Technical Specification
inner pipe diameter d2 min. extended mm min. recommended mm max. recommended mm max. extended mm pipe wall thickness min. mm max. mm material housing contact surface degree of protection according to IEC/ EN 60529 transducer cable type length dimensions length l width b height h dimensional drawing
190 220 900 1600 11 23 PPSU with stainless steel cap 304 (1.4301), option OS: 316L (1.4404) PPSU IP65
120 140 600 1000 7 15 PPSU with stainless steel cap 304 (1.4301), option OS: 316L (1.4404) PPSU IP65
m mm mm mm
ambient temperature min. C -40 -40 -40 max. C +170 +170 +170 temperature x x x compensation explosion protection category gas: 2G dust: 2D gas: 2G dust: 2D gas: 2G dust: 2D zone 1 21 1 21 1 21 explosion protection temperature (pipe surface) min. C -55 -55 -55 A max. C +140 +140 +140 T marking II2G II2G II2G 0637 0637 0637 E II2D II2D II2D X Ex e q IIC T6...T3 Gb Ex e q IIC T6...T3 Gb Ex e q IIC T6...T3 Gb / Ex tb IIIC TX Db Ex tb IIIC TX Db Ex tb IIIC TX Db I IBExU07ATEX1168 X IBExU07ATEX1168 X IBExU07ATEX1168 X E certification ATEX IECEx IBE 08.0007X IECEx IBE 08.0007X IECEx IBE 08.0007X C certification IECEx gas: increased safety, powder gas: increased safety, powder gas: increased safety, powder E type of protection filling filling filling x dust: protection by enclosure dust: protection by enclosure dust: protection by enclosure transducer x x x mounting fixture necessary 1 depending on application, typical absolute value for natural gas, nitrogen, compressed air 2 Lamb wave transducer: typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on request pipe diameter min. recommended/max. recommended: in reflection arrangement and for a flow velocity of 15 m/s pipe diameter max. extended: in diagonal arrangement and for a flow velocity of 25 m/s
20 TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
Technical Specification
FLUXUS G80x
inner pipe diameter d2 min. extended mm min. recommended mm max. recommended mm max. extended mm pipe wall thickness min. mm max. mm material housing contact surface degree of protection according to IEC/ EN 60529 transducer cable type length dimensions length l width b height h dimensional drawing
m mm mm mm
1699 4 74 32 40.5
1699 4 74 32 40.5
1699 3 42 22 25.5
ambient temperature min. C -40 max. C +170 temperature x compensation explosion protection category gas: 2G dust: 2D zone 1 21 explosion protection temperature (pipe surface) min. C -55 A max. C +140 T marking II2G 0637 E II2D X Ex e q IIC T6...T3 Gb / Ex tb IIIC TX Db I IBExU07ATEX1168 X E certification ATEX IECEx IBE 08.0007X C certification IECEx gas: increased safety, powder E type of protection filling x dust: protection by enclosure transducer x mounting fixture necessary remark
1 2
-40 +170 x
-40 +170 x
dust: 2D 21
dust: 2D 21
II2G II2D Ex e q IIC T6...T3 Gb Ex tb IIIC TX Db IBExU07ATEX1168 X IECEx IBE 08.0007X gas: increased safety, powder filling dust: protection by enclosure x 0637
II2G II2D Ex e q IIC T6...T3 Gb Ex tb IIIC TX Db IBExU07ATEX1168 X IECEx IBE 08.0007X gas: increased safety, powder filling dust: protection by enclosure x 0637 on request
depending on application, typical absolute value for natural gas, nitrogen, compressed air Lamb wave transducer: typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on request pipe diameter min. recommended/max. recommended: in reflection arrangement and for a flow velocity of 15 m/s pipe diameter max. extended: in diagonal arrangement and for a flow velocity of 25 m/s
21
TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
FLUXUS G80x
Technical Specification
inner pipe diameter d2 min. extended mm min. recommended mm max. recommended mm max. extended mm pipe wall thickness min. mm max. mm material housing contact surface degree of protection according to IEC/ EN 60529 transducer cable type length dimensions length l width b height h dimensional drawing
190 220 900 1600 11 23 PPSU with stainless steel cap 316Ti (1.4571) PPSU IP68 3
120 140 600 1000 7 15 PPSU with stainless steel cap 316Ti (1.4571) PPSU IP68 3
m mm mm mm
ambient temperature min. C -40 -40 -40 max. C +100 +100 +100 temperature x x x compensation explosion protection category gas: 2G dust: 2D gas: 2G dust: 2D gas: 2G dust: 2D zone 1 21 1 21 1 21 explosion protection temperature (pipe surface) min. C -55 -55 -55 A max. C +140 +140 +140 T marking II2G II2G II2G 0637 0637 0637 E II2D II2D II2D X Ex q IIC T6...T3 Gb Ex q IIC T6...T3 Gb Ex q IIC T6...T3 Gb / Ex tb IIIC TX Db Ex tb IIIC TX Db Ex tb IIIC TX Db I IBExU07ATEX1168 X IBExU07ATEX1168 X IBExU07ATEX1168 X E certification ATEX IECEx IBE 08.0007X IECEx IBE 08.0007X IECEx IBE 08.0007X C certification IECEx gas: powder filling gas: powder filling gas: powder filling E type of protection dust: protection by dust: protection by dust: protection by x enclosure enclosure enclosure transducer x x x mounting fixture necessary 1 depending on application, typical absolute value for natural gas, nitrogen, compressed air 2 Lamb wave transducer: typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on request pipe diameter min. recommended/max. recommended: in reflection arrangement and for a flow velocity of 15 m/s pipe diameter max. extended: in diagonal arrangement and for a flow velocity of 25 m/s 3 test conditions: 3 months/2 bar (20 m)/20 C
22 TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
Technical Specification
FLUXUS G80x
transducer
fixation
VL V K M Q D R S M L S W N 002 004 T36 013 036 092 200 IP68 OS Z example VL K D S S 200 /
option
size
Variofix L Variofix C transducers with transducer frequency G, H, K transducers with transducer frequency M, P transducers with transducer frequency Q reflection arrangement or diagonal arrangement reflection arrangement small medium large tension straps welding without fixation 10...20 mm 20...40 mm 40...360 mm 10...130 mm 130...360 mm 360...920 mm 920...2000 mm degree of protection IP68 housing with stainless steel 316 special design Variofix L and tension straps for transducers with transducer frequency G, H, K
TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
23
FLUXUS G80x
Technical Specification
material: stainless steel 304 (1.4301), 301 (1.4310), 410 (1.4006), 410 (1.4006) option OS: 316 (1.4571), 316L (1.4404), 17-7PH (1.4568) inner length: VLK: 348 mm, option IP68: 368 mm VLM: 234 mm VLQ: 176 mm dimensions: VLK: 423 x 90 x 93 mm, option IP68: 443 x 94 x 105 mm VLM: 309 x 57 x 63 mm VLQ: 247 x 43 x 47 mm
Variofix C (VC)
material: stainless steel 304 (1.4301), 301 (1.4310) option OS: 316 (1.4571) inner length: VCK-*L: 500 mm VCK-*S: 350 mm VCM: 400 mm VCQ: 250 mm dimensions: VCK-*L: 560 x 122 x 102 mm, option IP68: 560 x 126 x 120 mm VCK-*S: 410 x 122 x 102 mm, option IP68: 410 x 126 x 120 mm VCM: 460 x 96 x 80 mm VCQ: 310 x 85 x 62 mm
24
TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
Technical Specification
FLUXUS G80x
Technical Data
type coupling compound type N coupling compound type E coupling compound type H coupling foil type VT order code 990739-1 990739-2 990739-3 990739-0 990739-6 990739-14 990739-5 ambient temperature material C -30...+130 mineral grease paste -30...+200 -30...+250 -10...+200 silicone paste fluoropolymer paste fluoroelastomer for transducers with transducer frequency G, H, K for shear wave transducers with transducer frequency M, P for shear wave transducers IP68 and Lambwave transducers with transducer frequency M, P for transducers with transducer frequency Q remark
TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
25
FLUXUS G80x
Technical Specification
reflection arrangement
installation length
Technical Data
type width thickness length (per roll) weight ambient temperature properties mm mm m kg/m C E30R4 225 0.7 10 1.015 -30...+80 self-adhesive E30R3 50
26
TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
Technical Specification
FLUXUS G80x
Dimensioning
transducer transducer order mounting fixture code type number of layers damping mat transducer damping mat transducer damping mat + 2x pipe damping mat max. installation length number of rolls1 [mm] standard extended 1830 9 9 4 1 2 10 10 5 2 2 1 1 1 1 1 11 11 5 2 3 7 7 4 1 2 2 2 1 1 1 13 11 7 2 3 14 11 7 2 3 2 2 1 1 1 15 12 8 2 3 10 8 5 2 2 3 2 1 1 1
max. installation length [mm] VarioFix L VLK GLG GSG GLH GLK GSK GLG GSG GLH GLK GSK GLM GSM GLP GSP GLQ E30R4 3 3 2 1 1 E30R4 3 3 2 1 1 E30R3 1 1 1 1 E30R3 1 E30R4 3 3 2 1 1 E30R4 3 3 2 1 1 E30R3 1 1 1 1 E30R3 1 890 4 4 2 1 1 5 5 2 1 1 1 1 1 1 1 6 6 3 1 2 4 4 2 1 1 1 1 1 1 1
VLK-**-****/IP68
930
1910
VLM
660
1360
VLQ Variofix C VCK-*L GLG VCK-*L-****/IP68 GSG GLH GLK GSK VCK-*S GLG VCK-*S-****/IP68 GSG GLH GLK GSK VCM GLM GSM GLP GSP VCQ GLQ
1
540 1160
1120 2360
860
1760
960
1960
660
1360
calculation on the base of: - max. installation length (installation of one transducer mounting fixture per transducer in reflection arrangement) and - max. recommended pipe diameter (standard) or max. extended pipe diameter (extended) (for inner pipe diameter max. recommended and max. extended see Technical Data of the Transducers from page 17) calculation for the number of rolls when both transducers are mounted in one transducer mounting fixture (reflection arrangement) or in diagonal arrangement: number of rolls/2 and round up to the nearest integer
TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
27
FLUXUS G80x
Technical Specification
Connection Systems
connection system TS
connection with extension cable JB01 direct connection transducers technical type *****8*
x
transmitter
transmitter
JB01
****LI*
x x
transducer frequency (3d character of transducer order code) T S cable length m cable length (option IP68) m
G, H, K
M, P
transmitter
transmitter
x 5 12
l 300 300
x 4 12
l 300 300
x 3 -
l 90 -
x 2 -
l 40 -
Transducer Cable
Technical Data
type standard length max. length ambient temperature properties 1699 see table above -55...+200 transducer cable 2550 (option IP68) 6111 extension cable 2615 see table above -40...+70 halogen free fire propagation test according to IEC 60332-1 combustion test according to IEC 60754-2 sheath material stainless steel 304 (1.4301) option OS: 316L (1.4404) 8 PTFE 2.9 0.3 brown x PUR 5.2 0.2 0.9 gray x stainless steel 304 (1.4301) option OS: 316L (1.4404) 8 PFA 2.7 0.5 white x PUR 12 2 black x
m m C
-100...+225
outer diameter cable jacket material outer diameter thickness color shield
mm
mm mm
28
TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
Technical Specification
FLUXUS G80x
Junction Box
Technical Data
technical type dimensions fixation material housing gasket degree of protection according to IEC/ EN 60529 ambient temperature min. C max. C explosion protection zone A marking T E X / I certification ATEX E certification IECEx C type of protection E x JB01S4E3M see dimensional drawing wall mounting, optional: 2 " pipe mounting stainless steel 316L (1.4404) silicone IP67
-40 +80 1 II2G II2D Ex e mb II (T6)...T4 Ta -40...+(70)80 C Ex tD A21 IP67 T 100 C IBExU06ATEX1161 IECEx IBE 08.0006 gas: increased safety decoupled network: encapsulation dust: protection by enclosure 0637
Dimensions
wall mounting holder
70 163.5
156
in mm
TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
174
29
FLUXUS G80x
Technical Specification
Terminal Assignment
JB01
transducers
terminal strip KL1 terminal V VS RS R cable gland connection transducer , signal transducer , internal shield transducer , internal shield transducer , signal external shield
T R S
T V S
T G
V S V
T V
T G
T R
R S R
K L 2
K L 1
shield terminal
extension cable
terminal strip KL2 terminal TV TVS TRS TR shield terminal connection signal internal shield internal shield signal external shield
30
TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18
FLUXUS G80x
Technical Specification
FLEXIM GmbH Wolfener Str. 36 12681 Berlin Germany Tel.: +49 (30) 93 66 76 60 32 Fax: +49 (30) 93 66 76 80
internet: www.flexim.com e-mail: info@flexim.com Subject to change without notification. Errors excepted. FLUXUS is a registered trademark of FLEXIM GmbH. TSFLUXUS_G8V1-5-3EN_Leu, 2012-12-18