Escolar Documentos
Profissional Documentos
Cultura Documentos
by the instrument. It is advisable to validate the use when there is any change in software or reagent versions.
ALBUMIN
Instrument:
Reagent preparation
Reagent is ready to use.
RX-DAYTONA (=BQA)
Instrument settings
CHEMISTRY PARAMETERS Name Unit Assay Type Measuring Points 1 2 Wavelength Prim. Sampling Volume Rerun CALIBRATION Calculation 660 3 L 2 L Linear ALB g/L Endpoint 5 R1 R2 Reagent Name Reagent Name ALB Volume 300 L
Sec.
Decimal points 0 Normal Range (Male) 35 (Female) 35 (Child) 28 Technical Range (Conc.) 1.4 (mAbs/10) -35000 RPT Wash (R1) Sys Water (R2) Sys Water 0 Instrument Factor Stirring Speed
S1 S2
a 1 R1 Mid
DIRECT BILIRUBIN
The application parameters comprised here constitute a guide to facilitate the validation of our reagents by the instrument. It is advisable to validate the use when there is any change in software or reagent versions.
Instrument:
Reagent preparation
RX-DAYTONA (=BQA)
Working Reagent: Transfer the contents of one Reagent B vial into a Reagent A-D bottle. Mix thoroughly. Stable for 20 days at 2-8 C.
Instrument settings
CHEMISTRY PARAMETERS Name Unit Assay Type Measuring Points 1 2 Wavelength Prim. Sampling Volume Rerun CALIBRATION Calculation 546 20 L 10 L Linear BIL-D mg/dL Endpoint R1 R2 Reagent Name Reagent Name BIL-D BIL-D Volume 200 L Volume 50 L
23 Sec.
Decimal points 2 Normal Range (Male) 0.00 (Female) 0.00 (Child) Technical Range (Conc.) 0.00 (mAbs/10) -35000 RPT Wash (R1) Sys Water (R2) Sys Water 0 Instrument Factor Stirring Speed
S1 S2
a 1 R1 Mid
TOTAL BILIRUBIN
The application parameters comprised here constitute a guide to facilitate the validation of our reagents by the instrument. It is advisable to validate the use when there is any change in software or reagent versions.
Instrument:
Reagent preparation
RX-DAYTONA (=BQA)
Working Reagent: Transfer the contents of one Reagent B vial into a Reagent A-T bottle. Mix thoroughly. Stable for 20 days at 2-8 C.
Instrument settings
CHEMISTRY PARAMETERS Name Unit Assay Type Measuring Points 1 2 Wavelength Prim. Sampling Volume Rerun CALIBRATION Calculation 546 20 L 2 L Linear BIL-T mg/dL Endpoint R1 R2 Reagent Name Reagent Name BIL-T BIL-T Volume 200 L Volume 50 L
23 Sec.
Decimal points 2 Normal Range (Male) 0.00 (Female) 0.00 (Child) Technical Range (Conc.) 0.00 (mAbs/10) -35000 RPT Wash (R1) Sys Water (R2) Sys Water 0 Instrument Factor Stirring Speed
S1 S2
a 1 R1 Mid
CREATINE KINASE-MB
Inmunoinhibition The application parameters comprised here constitute a guide to facilitate the validation of our reagents by the instrument. It is advisable to validate the use when there is any change in software or reagent versions.
Instrument:
Reagent preparation
RX-DAYTONA (=BQA)
Working Reagent: Reconstitute the contents of a Reagent B vial with 2.5 mL (if 20x2.5 mL size) or 10 mL (if 10x10 mL size) of Reagent A. Mix gently. Stable for 15 days at 2-8C.
Instrument settings
CHEMISTRY PARAMETERS Name Unit Assay Type Measuring Points 1 2 Wavelength Prim. Sampling Volume Rerun CALIBRATION Calculation 340 10 L 5 L Linear CK-MB U/L Kinetic 21 30 R1 R2 Reagent Name Reagent Name CK-MB Volume 250 L
Sec.
Decimal points 0 Normal Range (Male) 0 (Female) 0 (Child) 0 Technical Range (Conc.) 0 (mAbs/10) -35000 RPT Wash (R1) Sys Water (R2) Sys Water 0 Instrument Factor Stirring Speed a 1 R1 Mid
S1 S2
LACTATE DEHYDROGENASE
Continuous-spectrophotometric PYRUVATE The application parameters comprised here constitute a guide to facilitate the validation of our reagents by the instrument. It is advisable to validate the use when there is any change in software or reagent versions.
Instrument:
Reagent preparation
RX-DAYTONA (=BQA)
Working Reagent: Pour the contents of the Reagent B into the Reagent A bottle. Mix gently. Stable for 2 months at 2-8C.
Instrument settings
CHEMISTRY PARAMETERS Name Unit Assay Type Measuring Points 1 2 Wavelength Prim. Sampling Volume Rerun CALIBRATION Calculation 340 5 L 2 L Linear LDH U/L Kinetic 2 11 R1 R2 Reagent Name Reagent Name LDH Volume 300 L
Sec.
Decimal points 0 Normal Range (Male) 207 (Female) 207 (Child) 207 Technical Range (Conc.) 0 (mAbs/10) -35000 RPT Wash (R1) Sys Water (R2) Sys Water 0 Instrument Factor Stirring Speed a 1 R1 Mid
414 414 414 1500 35000 Wash Sol.-3 Wash Sol.-3 b 0 R2 Mid
S1 S2
Spectrophotometric BIURET
PROTEIN
The application parameters comprised here constitute a guide to facilitate the validation of our reagents by the instrument. It is advisable to validate the use when there is any change in software or reagent versions.
Instrument:
Reagent preparation
Reagent is ready to be used.
RX-DAYTONA (=BQA)
Instrument settings
CHEMISTRY PARAMETERS Name Unit Assay Type Measuring Points 1 2 Wavelength Prim. Sampling Volume Rerun CALIBRATION Calculation 546 6 L 3 L Linear PROT g/L Endpoint 30 R1 R2 Reagent Name Reagent Name PROT Volume 300 L
Sec.
Decimal points 0 Normal Range (Male) 65 (Female) 65 (Child) 0 Technical Range (Conc.) 0 (mAbs/10) -35000 RPT Wash (R1) Sys Water (R2) Sys Water 0 Instrument Factor Stirring Speed a 1 R1 Mid
S1 S2