Você está na página 1de 72

ADDENDUM SPESIFIKASI TEKNIS PENGADAAN PERALATAN LABORATORIUM (LELANG ULANG) SUMBER DANA : APBN TAHUN ANGGARAN 2012 Perubahan

dalam spesifikasi teknis pengadaan peralatan laboratorium : 1. Item No. 27 (Automatic Serial Soxhlet System), Chiller, Footprint : tertulis 210 x 450 x570 mm, seharusnya 210 x 400 x500 mm or greater 2. Item No. 29 (Rotary Vacuum Evaporator + Chiller), Recirculating Chiller, Footprint of chiller : 280 x 400 x 500 mm or greater, seharusnya 210 x 400 x 500 mm or greater 3. Item No. 39 (Distillation Unit) dan 40 (Distillation Unit For Analisys SO2 With Titrator), tertulis Protective door with sensor, seharusnya Protective door (tulisan with sensor dihilangkan) 4. Item No. 54 (Cabinet Sterilizing) Perubahan Dimensions tertulis at least 860 x 480 x 1300 mm, seharusnya at least 860 x 490 x 1750 mm NO 1. INSTRUMEN NAME and TECHNICAL SPECIFICATION ATOMIC ABSORPTION SPECTROPHOTOMETER *) Specifications : a. Analysis methods b. Optic system : c. Measured wavelength range d. Detector e. Background correction f. lamps g. Gas Controllers h. Burner head i. furnace (1 unit ) - autosampler - Programming temperature ENTITY Unit AMOUNT 1

: : : : : : : :

Flame & Furnace double-beam method 185 - 900 nm or greater Photomultiplier or equivalent D2 and Zeeman method 8-lamps Automatic Titanium or the corotion resistant material

: 40 to 3000 C

Includes of: a. Training b. Original Operasional manual (plus bahasa indonesia) c. Guarantee min. A year d. Installation e. Factory Certificate f. Blower + Vent + Dusting (2 set) g. Vapor Generator accessory (pump units; gas liquid separator; quartz cell; reagent container; tubing, Flow Through Mercury Cell and spares kit) h. Hollow Cathode Lamps + Standard Solution for : Pb, Hg, Cd, Zn, As, Mg, Ca, Na, K,Cu, Fe, Cr, Se, Sn, Ag (15 buah) i. Argon gas with tank & regulator (1 set) j. C2H2 gas with tank & regulator (1 set) k. N20 gas with tank & regulator (1 set)

l. m. n. o. p. q. r. 2 LSMSMS *)

Compressor + air dryer (1 set) UPS 10 KVA (2 unit) Dust cover + tool kit (1 unit) Furnace Camera (1 unit) Recirculating Water Chiller (1 unit) PC + Software + printer (1 set) Online Calibration & Dilution System Unit 1

Specifications : MASS SPECTROMETRY a. Ion Sources b. Ion Sources Transfer Optics c. Detector d. Vacuum System e. Mass range f. Scan Speed g. Mass resolution h. Mass sensitivity POMPA a. Type b. Flow rate range c. Flow rate precision d. Flow rate accuracy e. Maximum operating pressure AUTOSAMPLER a. Injection volume b. Injection precision c. Sampel capacity d. Sample carryover OVEN KOLOM Column temperature range KOMPUTER DAN SOFTWARE

: ESI dan APCI : High Efficiency stacked ring ion guide : Photomultiplier or equivalent : single : 2 to 2000 or wider : 5000 - 10000 Da/s : 1,0 Da FWHM : ESI(-) , S/N 180:1, 10 pg Reserpine or better

: Quartenary : minimum 0,010 to 2,000 ml/min : 0,075 % RSD : 1,0 % : 8500 psi or greater

: : : :

0,1 to 10,0 L 1 % RSD min 90 < 0,005 %

: 20 to 90 C or greater

Hardware

Software LIBRARY compatible with system N2 generator min 20 L/sec, with internal compressor Ar gas + cylinder + regulator, tubibg, connector, reducer UPS min 6 KVA Magnetic Switch Sample Cone Solvent for 1 year : Methanol 4 x 2,5 L Acetonitril 2 x 2,5 L NH4OH Formic Acid 1 L Includes of: 1. Installation 2. Training 3. Performance Test 4. After Sales Support 5. Factory Certificate 6. Guarantee 1 year 7. Spare Part Availability min. 5 year 8. Original manual book 9. Column C18 10. Column C8 11. Guard column for C 18 12. Guard column for C 8 13. Pre Column Filter Holder (fitting included) 14. 0,2 m Filter Cartridge, 5 / pk 15. Tube 50 ml /AOAC-acetat 16. Tube 2 ml- AOAC/C18 17. Solvent Clarification Kit with pump 220 V/50 Hz 18. Membran filter diameter 25 mm, 0,2 m ; 50 /pk 19. Sampel vial min 1 ml (100 pc/pk)

: Min Pentium 4, 3,2 GHz/800MHz, 2 MB L2 Cache, Keyboard & Scroll Mouse & USB, 160 GB 7200 RPM, Windows XP Pro, DVD Multi Burner, Com 2 cable, System Restore DVD, Monitor, Printer. : Compatible

: : : : : : : : : : :

min 3 pcs min 3 pcs min 3 pkgs min 3 pkgs min 1 pk min 1 pk min 1 pk min 1 pk min 1 kit min 3 pkgs min 3 pkgs

GC Auto injector With Detector FID, TCD, FPD *) Each set consist of : 1 Coloumn oven 2 Injection Port 3 Flow Control Unit 4 Split / Splitless, Programmable Temperature Vaporization injection unit 5 Detectors 6 Workstation 7 Autoinjektor 1 Column Oven Specifications : a. Temperature range b. Cool down range 2 Injection Port Specifications : Split/Splitless, Programable Temperature Vaporization Injection Unit - Split ratio setting range - Total Flow setting - Temperature range 3 Flow Control Unit Specification : a. Pressure setting range b. Pressure control precision 4 Detectors Flame Ionization Detector (FID) Specification : a. Temperature range Thermal Conductivity Detector (TCD) Specification : a. Temperature range

Unit

: : : : : : :

1 unit minimal 1 unit 1 unit 1 unit 3 unit 1 set minimal 1 unit

: to 450 C : up to 50 C or more cooler : minimum 1

: 1:7500 or better : 0 - 200 mL/min or better : up to 400 C or greater

: 0 to 100 psi or greater : 0,001 psi or greater

: up to 400C or greater

: Up to 400oC or greater

Flame Photometric Detector (FPD) Specification : a. Temperature range

: Up to 350oC or greater

5 a. Hardware Minimum Intel Processor core I3, Windows XP, Hard Disk min. 120 GB, RAM min 2 GB, CDROM Drive, DVD/CD Writer, LCD Monitor 17", Color printer. b. Software : - Compatible with Windows XP - Data Acquisition - Hardware Functions - Data Processing and Analysis - Data Storage 7 Autoinjector a. number of vials : min 12 or greater 8 Accessories : 1. Hydrogen gas 2. Helium Gas 3. Compress Air 4. Regulator for gases 5. 10 mL screw vial 6. Capillary Column 7. UPS min 5 KVA 8. GC Data Systems 9. Driver for SW 10. Gas Clean Filter 9. The system includes of : 1. Instalation 2. Training 3. Dust Cover 4. Guarantee min. 1 year 5. Spare Part Availability min. 5 year 6. Original operation manual 7. Factory Certificate 8. Tool Kit

: : : : : : : : : :

1 pcs 1 pcs 1 pcs 3 pcs 1 pcs 2 pcs 1 pcs 1 pcs 1 pcs 1 pcs

9. BAPETEN license for ECD 10. Training at Representative GC Training Centre 4 GC Auto injector With Detector FID, TCD, ECD *) Each set consist of : 1. Coloumn oven 2. Injection Port 3. Flow Control Unit 4. Split / Splitless, Programmable Temperature Vaporization injection unit 5. Detectors 6. Workstation 7. Autoinjektor 1 Column Oven Specifications : a. Temperature range b. Cool down rate 2 Injection Port Specifications : Split/Splitless, Programable Temperature Vaporization Injection Unit a. Split ratio setting range b. Flow rate setting range c. Temperature range 3 Flow Control Unit Specification : a. Pressure setting range b. Pressure control precision 4 Detectors Flame Ionization Detector (FID) Specification : a. Temperature range Thermal Conductivity Detector (TCD) Specification : Unit : : : : : : : 1 unit min 1 unit 1 unit 1 unit 3 unit 1 set min 1 unit 1

: to 450 C : up to 50 C or more cooler : minimum 1

: 1:7500 or better : 0 -200 mL/min or better : up to 400 C or greater

: 0 to 100 psi or greater : 0,001 psi or greater

: up to 400C or greater

a. Temperature range Electron Capture Detector (ECD) Specification : a. Temperature range

: Up to 400oC or greater

: Up to 350oC or greater

5 a. Hardware Minimum Intel Processor core I3, Windows XP, Hard Disk min. 120 GB, RAM min 2 GB, CDROM Drive, DVD/CD Writer, LCD Monitor 17", Color printer. b. Software : a. Compatible with Windows XP b. Data Acquisition c. Hardware Functions d. Data Processing and Analysis e. Data Storage 7 Autoinjector a. number of vials 8 Accessories : 1. Hydrogen gas 2. Helium Gas 3. Compress Air 4. Nitrogen Gas 5. Regulator for gases 6. 10 mL screw vial 7. Capillary Column 8. UPS min 5 KVA 9. GC Data Systems 10. Driver for SW 11. Gas Clean Filter 9. The system includes of : 1. Instalation 2. Training 3. Dust Cover 4. Guarantee min. 1 year

: min 12 or greater

: : : : : : : : : : :

1 pcs 1 pcs 1 pcs 1 pcs 4 pcs 1 pcs 2 pcs 1 pcs 1 pcs 1 pcs 1 pcs

5. 6. 7. 8. 9. 10. 5

Spare Part Availability min. 5 year Original operation manual Factory Certificate Tool Kit BAPETEN license for ECD Training at Representative GC Training Centre Unit : : : : : : : : : : : : : : : : maximum 2 unit Quaternary 0,0001-5,0 ml/min or greater 1% 0.1% RSD 5000 - 6000 psi 1% min 3 1 unit D2 lamp 200 - 600 nm or greater 0,1 nm 1 nm 2,5 AU Dual Wavelength 5 x 10-6 AU 4

HPLC UV-Vis with Auto injector *) Specifications Pompa a. Type b. Flow rate range c. Flow rate accuracy d. Flow rate precision e. Maximum pressure f. Gradient accuracy g. Number of solvent Detektor UV-Vis a. Lamp b. Wavelength range c. Wavelength Precision d. Wavelenght Accuracy e. Linearity f. Function g. Noise Column Oven With Cooler Temperature setting range Autosampler With Sample Cooler a. Injection Accuracy b. Temperatur control c. Injection volume d. Sample Capacity

: 1 unit : 4 C to 65 C or higher : : : : : 1 unit 2% or better 4 C - 40 C 0,1-100ul min 100 vial @ 1 - 2 ml

Degasser

: 1 unit

Data Processor : 1 set - Minimum Core i3 2100, 3.1 GHz, Intel GMA, Win 7 Pro 32-bit, HDD 500 GB 7200RPMSATA, USB Optical Mouse & Keyboard, Media Card Reader, Monitor LCD 18,50", Printer. - Compatible Software for Data analysis and reporting Include 1 Instalation 2 Training 3 Tool kits (1 set) 4 Maintenance & autorinsing Kit for pump (1 set) 5 Dust Cover 6 C18 coloumn, 5m, 25 cm: minimum 2 pcs 7 Sampel vial min 1 ml (100 pc/pk) 8 Micro Syringe 100 L, minimum 1 pcs 9 Compatible mixer, min 1 set 10 Guarantee min. 1 year 11 Spare Part Availability min. 5 year 12 Reservoir Tray (1 pcs) 13 Original manual book 14 Factory certificate 15 Stabilizer min 3 KVA 16 Needle wash (1 set) 17 HUB W/ Cable 6 HPLC PDA with Auto injector *) Spesifications 1. Pompa (maksimal 2 unit) a. Tipe b. c. d. e. f. g. Rentang laju alir Akurasi laju alir Presisi laju alir Gradient accuracy Jumlah wadah fase gerak Max. Pressure Unit 1

: Quarternary : : : : : : 0,0001-5,0 ml/min or greater 1% 0.1% RSD 1% minimal 3 5000 - 6000 psi

2. Detektor (1 unit) a. Lampu b. Rentang panjang gelombang c. Presisi Panjang gelombang d. Akurasi Panjang gelombang e. Lineritas f. Photodiodes 3. Oven Kolom (1 unit) a. Rentang suhu b. Presisi suhu 4. Autosampler (1 unit) a. Injection Accuracy b. Injection Volume range c. Sampel Capacity 5. Manual Injektor & Stand Kit (1 unit) 6. Degasser (1 unit) 7. Data Processor Included of 1. Instalasi 2. Pelatihan 3. Kolom C18, 5m,25 cm (min 2 pcs ) 4. Garansi: minimum 1 tahun 5. Tool kits (1 set) 6. Buku petunjuk pemakaian asli (plus bahasa indonesia) 7. Factory certificate 8. Needle wash (1 set) 9. dust cover (1 set) 10. Sampel vial min 1 ml (100 pc/pk) 11. HUB W/ Cable 12. Stabilizer min. 3 KVA

: : : : : :

D2 lamp 190-800 nm atau lebih besar 0.1 nm 1 nm 2.5 AU 512

: 4 C - 65 C or higher : 1% atau lebih kecil

: 2% or better : 0.1 - 100 l or better : min 100 Vial @1 - 2 ml

: Intel Core i3-2100 Processor,min 2GB, Windows 7 Professional,500GB SATA,DVD+RW, Monitor LCD 18,50",Software PDA,Printer

13. 14. 15. 16. 17. 7

Maintenance & autorinsing Kit for pump Micro syringe 100l, min 1 pcs compatible mixer min.1 set Spare part availability min. 5 year Reservoir Tray 1 set Unit : : : : 1 nm 190 - 1100 nm or better 0.2 nm at 656.1 nm or better 0.05 % or less (220 nm & 340 nm) 1

SPECTROPHOTOMETER UV-VIS *) Specifications: a. Spectral bandwitch b. Wavelength range c. Wavelength accuracy d. Stray light e. Photometric system f. Photometric range g. Photometric accuracy h. Photometer repeatability

: Double beam optic : Absorbance : (-) 2 to 2.0 Abs or higher Transmitance: 0.0 to 300 % or higher : 0.002 Abs ( at 0.5 Abs ) ; or better : 0.001 Abs ( at 0.5 Abs)

i. Baseline Stability j. Baseline flatness k. Light source l. Noice level Included of : 1. Instalation 2. Training 3. Guarantee min.a year 4. Original operation manual 5. Dust Cover 6. Tools kit 7. Glass Filter Holmium Oxide 8. Shutter Block

: 0.0003 Abs : 0.0006 abs (200 to 900) or better : W1 and D2 lamp. : 0.00015Abs (at 500 nm)

: : : : :

1 set (plus bahasa indonesia) 1 set 1 set 1 set 1 set

9. 10. 11. 12. 13. 14.

Cuvette/cell 1-cm Stabilizer 1 KVA Multi-cell sample compartment unit with 6-cell holder Factory Certificate Rectangular Quartz Cell, 10 mm Branded personal computer and printer: - Minimum suplied : Intel Pentium Core II Duo or similar

: 1 set : 1unit

: 2 Pcs/set

DISSOLUTION TESTER *) 1. Specifications : a. Flask temperatur - Range - Accuracy b. Bath capacity c. Spindle speed - Range - Accuracy d. Vibration e. Voltage f. Supporting systems including with

Unit : 30 to 40 C or better : 0.2 C or better : 8 flasks : : : : : 25 to 250 rpm or better 1 rpm or better < 0.1 mils 220 V-240 V, 50/60 Hz - 8pcs USP/NF glass dissolution flask - 8 pcs USP paddles - 8 basket assembly 40 mesh basket - 8 pcs sampling probe 8 Pcs 8 Pcs 8 Pcs 8 Pcs 30 tablet 1 rol Pack of 100 8 pcs

2 Accessories : a. USP / NF Glass dissolution flask b. USP all teflon paddle c. basket assembly, 40 mesh basket d. sampling probe e. USP Calibrator Prednisone Tablet f. Stainless wire to sink capsule g. Filter h. Sinker Japanese Pharmacopoeia Cage 3 Included of : 1. Instalation 2. Training 3. Dust cover 4. Guarantee min. A year 5. original operation manual(plus bahasa indonesia) 9 DISSOLUTION TESTER WITH AUTOSAMPLER *) Each unit consist of 1. Dissolution Tester

: : : : : : : :

Unit

2. Autosampler 1. Dissolution Tester Specifications : a. Flask temperatur Range Accuracy b. Bath capacity c. Spindle speed Range Accuracy d. Voltage e. Supporting systems including with - 8pcs USP/NF glass dissolution flask - 8 pcs USP paddles - 8 basket assembly 40 mesh basket - 8 pcs sampling probe -USP Calibrator Prednisone (30 tablets) -Stainless wire to sink capsule 1 roll -Filter (pack of 100)

: 30 to 40 C or better : 0.5 C or better : 8 flasks : 25 to 250 rpm or better : 1 rpm or better : 220 V-240 V, 50/60 Hz

2. Autosampler Specifications : Include precision syringe pump sampling Change and save minimum up to 20 protocols, easy-to-use routines, password protect Built-in validation, calibration and diagnostics, plus unique advanced features Teflon valves, media replace and system rinse & wash. Stainless probes for media replace and waste; 8 sample system 24 hours sampling time (sustained release) Media replace and system rinse 3 Accessories : a. Validation tool kit b. Replacement filters c. Base-kit, including Device box d. Sinker Japanese Pharmacopoeia Cage e. Sample and return probe kit f. E-prob actuator kit g. Vessel cover h. Return Harness,6 ft i. Evaporation caps for 16 mm test tube j. Certified calibrated tachometer k. Certified calibrated thermokopel

: 1 set : pack of 100 : : : : : : : : 8 pcs 8 pcs 8 pcs 8 pcs 8 pcs pack of 50 1 set 1 set

4 Included of : 1. Instalation 2. Training 3. Dust cover 4. Guarantee min. 1 year 5. Spare Part Availability min. 5 year 6. original operation manual (plus bahasa indonesia) 7. Alignment kit 10 MIKROSKOP KAMERA *) Specifications: A. Microscope body a. Lamp b. Optical system c. Eyepieces d. Objective e. Able to pre -focus & tension adjusment B. Accessories : a. Compatible Digital Camera b. PC + Sofware for capturing image and recording movie files,can be attached to any desktop or laptop c. Memory C. Included of 1. Object glass and cover 2. instalation 3. training 4. guarantee min. a year 5. original operation manual (plus bahasa indonesia) 6. dust cover 11 FUME HOOD PROVIDED WITH MICROPROCESSOR CONTROL *) Specifications : a. Exterior dimension (W x D X H in mm) b. Interior work zone dimension (W x D X H in mm) c. Max. sash opening d. e. f. g. Exhaust air volume for minimum face velocity Other exhaust outlet diameter Light intensity for work surface Construction : - Work surface Unit 1

: : : :

halogen illumination UIS up to 10x up to 10x, 40x, 100x

: min 12 mpx or better.

: CompactFlash,microdrive,xD-Picture Card : min 100 pcs

Unit : (1100-1250) x (750-900) x (1100-1600) : max. 996 x 672 x 1240 : 740 mm (vertical rising sash) : : : : : 500 cmh or greater 305 mm > 1000 lux (>100 foot candles) or greater Tri wall construction Phenolic Resin Laminate

- Main Body - Coating h. Controller

Include of 1. Installation 2. Guarantee min a year 3. Training 4. Original operation manual 5. Sash stop kit and standar accessories for fume hood (blower, ducting, piping, etc) 12 WATER PURIFICATION SYSTEM *) Specifications : Complete System Type I : for AAS,GC, HPLC Type II : for Micro media, feed water Type I Type I Daily Volume Delivery Flow rate-maximum Inorganics ( resitivity @ 25C) Organics/TOC Bacteria (with Biofilter) Endotoxin (with Biofilter) DNAse (with Biofilter) RNA (with Biofilter)

: Fabricated of electro-galvanized steel : Epoxy polyester hybrid antimicrobial powder coating : Microprocessor Completed with airflow alarm kit :

Unit

: : : : : : : :

min. 10 litre Up to 2 L/min 18.2 M.cm < 5 ppb < 1 cfu/10 ml or better < 0.001 EU/ml < 20 pg/ml or better <0.002 ng/ml

Type II Daily Volume Daily output Inorganics ( resitivity @ 25C) Organics/TOC Bacteria Recirculating Supply voltage Included of : 1. Tank 25 L with 5 level switches, 2. Deionization Cartrige ( 2 ) 3. Purification Pack ( 2 ) 4. Biofilter ( 2 ) 5. POU filter ( 2 ) 6. Warranty 1 year

: : : : : : :

min. 7 litre/hour 170 L /day or better 10 to > 10 M - cm < 20 ppb < 1 cfu/10 ml or better Fully Deionization 100 -240 V ac / 50 -60 Hz

7. Install 8. Certificate of Conformity 13 DIGITAL BURET Spesifikasi: Dispensing range Display volume buret Included 1. safety valve 2. microbatteries 2 piece 3. telescopic aspiration tube 4. bottle adapters minimal 3 piece 5. adjustable discharge tube 6. recirculation tube 7. original operation manual (plus bahasa indonesia) 8. instalation 9. guarantee min. a year 10. training 14 AUTOMATIC TLC SAMPLER *) Specifications: a. Automatic sample application for qualitative and quantitative b. The system is Controlled by software c. Aplication of spots, bands or wider d. External gas supply Include of 1. Dosing syringe starter kit including 2. 25 ul Dosing syringe 3. Needle for spray-on application 4. Needle for contact application 5. Heated spray nozzle 6. Sample vial rack for 66 vials 2 ml (12 x 32 mm) 7. Waste bottle 250 ml complete 8. Syringe rinsing tube 9. TLC-foil down-holder 10. Compensation ledge for 10 x 10 cm plates or wider 11. Screw cap bottle 250 ml, suitable as waste or rinsing bottle 12. Gas N2 + Regulator 13. Monitor 17" LCD 14. Printer 15. Instal Unit : 0.01ml - 999.9 ml : LCD : minimal 20 ml : 3

Unit

16. Training 17. original operation Manual (plus bahasa indonesia) 18. garansi minimal 1 tahun 15 SEMI AUTOMATIC TLC SAMPLER Specifications: a. System b. Aplication of spots c. Temperatur range d. Display f. Kapasitas chanel g. Plate dimension h. Power Include of 1. Syringe 50 l & 100 l minimal 10 pcs 2. Syringe 25 l minimal 20 pcs 3. Instal 4. Training 5. original operation Manual (plus bahasa indonesia) 6. garansi minimal 1 tahun 16 AUTOMATIC TITRATOR *) Spesifications : a. Application b. Burette c. operation d. Interface e. max. number of sampel Accessories : a. Magnetic stirer b. pp beaker set of 500 pcs c. Glass Beaker 80 ml set of 20 pcs d. Burettes for 2 titration e. IQ/OQ f. Printer Included of : 1. Installation 2. Training 3. Guarantee min. a year 4. Original operational manual (plus bahasa indonesia) 17 ULTRASONIC BATH Unit 2 Unit : : : : : redox & iodometric titration & Complexometric titration. 10 ml Color LCD touch panel or color TFT touch control RS 232 min 100 1 Unit : Semi Automatic , qualitative and quantitative : : : : : 85C - 90C or wider digital or equivalent up to 18 or more 10x10cm ; 20x20 cm or more 220 V/ 50 hz 1

Specifications: a. Capacity b. Temperatur range c. Timer d. Frequency e. Display f. Material Include of 1. Guarante min. A year 2. Original operation manual (plus bahasa indonesia) 3. Dust cover ( 1 unit) 4. Tray/basket (stainless steel) (1 unit) 5. Hinged Lids made of Stainless Steel to be mounted on the unit 18 SEPARATORY FUNNEL SHAKER Specifications : a. Shaking method b. Shaking speed c. Speed control d. Timer e. Motor f. Voltage Accesories a. Separatory funnel holder b. Fuse c. Carbon Brush Included of 1. Training 2. Guarantee min.1 year 3. Manual book 4. Certificate of Origin 19 SHAKER Specifications : a. Motion b. Speed - Speed frequency (min-1) c. Load capacity d. Amplitude (mm) e. Power f. Timer

: : : : : :

10 liters up to 70 o C or greater minimum 1-30 minutes or greater minimum 35 KHz LED Stainless steel or equivalent

Unit : : : : : : Vertical and rotary shaking Approx 20 to 210 rpm or greater Stepless speed control by thyristor 0-60 min or more DC motor, 90 W 100/220V, single phase, 2.2/1.0 A /50/60 Hz

: 1 set (6 funnel) for 100 - 1000 ml : 1 pcs : 1 set

Unit : Orbital : : : : : 0 to 200 or greater minimal 8 kg min 10 mm 115-230 V, 50/60 HZ Up to 60 min or greater

g. Dimention h. Shaking platform Acessories 1. fastening bar 545 mm lengthways minimal 2 unit 2. clamping rolls 375 mm, widhtways minimal 3 unit 3. clamping rolls 375 mm, widhtways minimal 3 unit Included of 1. Instalation 2. Guarantee min. 1 year 3. Spare part availability min. 5 year 4. Original operation manual 20 SEMIMICRO BALANCE *) Specifications : a. Capacity max b. Readability c. Reapeatability d. Liniearity e. Power f. Display Acessories 1. Printer 2. Antistatic brush 3. Stainless steel spatulla, 6 x 180 mm @ 5 pcs 4. Stabilizer 500 mA 5. External Calibrator 100 and 200 g E2 Included of 1. Installation 2. Training 3. Original operational manual book 4. Guarantee min 1 year 5. Spare part availability min 1 year 6. Dust cover 21 CENTRIFUGE Specifications : a. Max Speed (rpm) b. Max capacity c. Control system d. Display digital fuse number speed time

: min (380 W x510 D x 120 H) mm : minimal 300 x 300 mm :

Unit : : : : : : : minimal 200 g or more 0.01 mg 0,07 mg 0,2 mg or better 100 - 240 V, 50-60 Hz LCD or better

Unit : : : : minimal 4000 rpm swing / minimal 4000 rpm angle up to 4 x 200 ml (swing out) Microprocessor yes

e. Timer f. Power requirements Accessories : 1. - Angle- Rotor Include of : 1. Guarantee min. 1 year 2. Original operation manual (plus Bahasa indonesia) 3. Spare part availability min. 5 year 22 pH Meter Berdiri *) Specifications : a. pH Range b. pH Relative accuracy c. pH Resolution d. mV Range e. mV Relative Accuracy f. mV Resolution g. Temperature range h. Acessories

: 1-99 minutes : 220-240; 50-60 Hz

: 6 x 50 ml

Unit : : : : : : : : 0.00 to 14.00 or wider 0.005 or less variable 0.001 / 0.01/0.1 -1800 to 1800 or wider 0.2 or less 0.1 or better 0 to 100 0C or wider Technical buffer pH = 3 point : acid, neutral, base @ min 250 ml Power adapter Stabilizer

Included of 1 Installation 2 Training 3 Guarantee min 1 year 4 Original Operation manual (plus bahasa indonesia) 5 Storage solution 6 spare part availability 5 year 7 electrolyte solution 23 ALAT HIDROLISIS LEMAK Specifications : a. Max. Temperature b. Capacity perbatch c. Power Consist of: 1 The crude fibre digestion unit (1) 2 Plus a filter with 6 funnels 3 Soft polyethylene bottle with a narrow mouth (1000 ml) : 1 4 100 folded filter as well as 250 gr boiling stones 5 Glass condensers

Unit : min 350 C : 6 sample : 230 V, 50/60Hz

6 7 8 9 10 24

Ring shaped rest Silicone rubber tubings Beakers (1000 ml) Power cord Standard Accessories Unit : : : : 500 ml up to 4500C digital Nichrome Wire or equivalent 4

HEATING MANTLE Specifications : a. Flask capacity b. Temperature range c. Temperature controller d. Material Include of : 1 Guarantee min. 1 year 2 Spare part availability min. 5 year 3 Original operation manual (plus bahasa indonesia)

25

ANALYTICAL BALANCE with PRINTER *) Specifications : a. Capacity max b. Readability c. Reapeatability d. Liniearity Included of: 1 Printer 2 Instalation 3 Dust cover 4 Guarantee min 1 year 5 Spare part availability min 5 year 6 Original operational manual book 7 Weight Standart Class E2 Capacity 100 & 200 g

Unit : min 220 g : 0,0001 g : 0,1 mg : 0,2 mg or less

26

APPARATUS FOR DETERMINATIONS OF SAPONIFICATIONS NUMBER AND OH NUMBER Specifications : a. Digestion Unit : 1 b. Sample positions : 6 samples c. Dimension (W x H x T) : approx 31 x 90 x 38 cm d. Weight : approx 28 kg e. Nominal Voltage of the heating block : 220 - 240 VAC 50/60 Hz f. Power Consumptions : approx 800 Watts

Unit

Included of 1 Instalation 2 Training 3 Heating block 4 Controller 5 Glassware ( 6 pcs) 6 Straight Air Condenser 27 AUTOMATIC SERIAL SOXHLET SYSTEM *) Completed with Waterbath Each set consist of : 1 Extraction system 2. Chiller 1. Extraction system Consists of Extraction Unit and Control Unit Specifications : a. Solvent volume b. Capacity per-batch c. Power supply of control unit Accessories 1. Extraction beaker glass 2. Holder for extraction Thimbles 3. Extraction thimbles 33 x 80 mm 4. Pair of glass tongs 5. Box of boiling stones 250 g 6. Holder for 6 extraction beaker 7. Tubing set complete 8. Line filter 16 A (just for 230V) 2. Chiller Specifications : a. temperature range b. temperature ontrol c. Footprint 3. Includes of 1. Instalation 2. Training 3. Dust cover 4. Guarantee min 1 year 5. Spare part availability min 5 year 6. Original operation manual

Unit : 1 unit : 1 unit

: minimum 100 ml : 6 samples : 230V/50/60 Hz : 12 pcs : 6 pcs : 25 pcs

: 0 C to 100 C : 0.2 C : 210 x 400 x 500 mm or greater

28

GLASS WARE DRYER Specifications : a. Dimensions ( W x D x H ) b. Nominal Voltage c. Number of drying tubes d. Temperature in drying vessels e. Electric timer ROTARY VACUUM EVAPORATOR + CHILLER *) Consist of : A. Rotary Evaporator (1 unit) Specifications : a. Operating Voltage b. Rotation Speed B. Heating Bath (1 unit) a. Temperature Range b. Bath Capacity C. Vacuum Pump with controller (1 unit) a. Capacity b. Final vacuum (absolute) D Recirculating Chiller (1 unit) a Temperature range b Power Supply c Footprint of chiller E Acessories : 1. vapor temperature sensor 2. condenser 3. Manual vacuum controller 4. Woulff botle minimal 1 pcs F Included of : 1 Installation 2 Training 3 Guarantee min 1 year 4 Original Operation Manual (plus bahasa indonesia)

Unit : : : : : Approx 380 x 380 x 1150 mm 230 V ac / 50 Hz / 1000 Watt max. min. 56 tubes, consist of small & large vessels up to 50C Approx 120 minutes Unit

29

: 100-240V, 50-60 Hz : 20 to 280 rpm or greater

: 20 to 150 0C or wider : 4 L or greater

: 1.5 m3/h or greater : 10 mbar

: Fix or adjustable : 220/240 V, AC, 50/60 Hz : 210 x 400 x 500 mm or greater :

30

MICROWAVE DIGESTION Spesification : a. Temperature control

Unit : every vessel

b. c. d. e. f.

Jumlah vessel Volume vessel Vessel material Temperature Presisi temperatur

: : : : : :

min 10 vessel min 10 ml PTFE/Teflon 50-2600 C 1C at 200C

Accessories 1. Turntable with 10 pcs of Vessels 60 mL - Forming Systems for Vessels 60 mL - Consumables for Vessels 60 mL For approx 2500 digestions - Spare of Vessels 60 mL , pk/10 2. Turntable with 24 pcs Quartz Vessels 10 mL - Forming Tools for Quartz Vessels - Rupture Disc for Quarts Vessels, pk/5 - Lids for Quartz Vessels, pk/4 - Fixations Plate of Rupture Disc - Spare of Quartz Vessels 10 mL, pk/24 Included of 1. instalation 2. training 3. guarantee min. a year 4. original operation manual (plus bahasa indonesia) 31 WATERBATH Spesifications a. Capacity b. Casing, container and cover made of c. Temperature range d. Resolution h. Display i. Timer Accessories a. Flat stainless steel cover with openings& concentric ring sets b. Gable stainless steel cover mounted for condensate drainage c. water level control Included of : 1. Instalation 2. Guarantee min. a year 3. Original operation manual (plus bahasa indonesia)

Unit : : : : : : 14 liter Stainless steel 10 - 95C or better 0.1C LED 1 minutes to 99 hours or better

: 6 holes

32

SPIRAL SERIAL FLASK HEATER Spesifikasi: a. Temperature b. Voltage c. Frequency d. Wattage e. For Flask f. Heating place g. Dimensi h. Weight Included of 1. original operation Manual (plus bahasa indonesia) 2. instalasi 3. garansi min 1 tahun UPS Spesification : a. Capacity b. Input voltage c. Input Frequency d. Output Voltage e. Output Frequency f. Overload g. Runtime Included of 1. Installation 2. garansi spare parts,bateray & service minimal a year 3. Original operation manual (plus bahasa indonesia)

Unit : : : : : : : : up to 425C 230 V 50 - 60 Hz 1800 W 50 - 250 ml Minimal 6 Min 590 x 220 x 125 or wider Min 7 kg

33

Unit : : : : : : : : min 10 Kva 220 volt (min 1 fase) 45 - 65 Hz 220 volt (min 1 fase) 50 / 60 Hz up to 120% 0r better approx 10 minutes

34

UV CABINET Specifications : a. UV lamps in combination with UV viewing box b. Consist of : UV lamps for 2 wavelength 254/365nm (2 x 6W) or better Included of : 1. Guarantee min. 1 year 2. Spare part availability min. 5 year 3. Original operation manual (plus bahasa indonesia)

Unit

35

HOT PLATE STIRRER Specifications : a. Heating plate material b. Stirring capacity with HO

Unit : glass Ceramic or equivalent : minimal 4 L

c. Speed range d. Temperature range e. Power Supply Included of : 1. Magnetic bar minimal 5 pcs 2. Guarantee min.a year 3. Original operation manual (plus bahasa indonesia) 36 HOT PLATE SEDANG Specifications : a. Min Temperature b. Max Temperature c. Heating Surface (width x Depth) d. Power e. Type f. Approvals g. Overall dimensions (W x H x D) HOT PLATE BESAR Specifications : a. Digital LCD display b. Anodized c. Temperature range up to 300C d. Plate size(mm) approx 300 x 600 e. Accuracy max 0.3C f. Operating manual prosedur (English and Indonesian) g. Min guarantee 1 year h. Power supply 230 V/ 50-60 Hz i. Dimensions at least 650 x 350 x 140 MUFFLE FURNACE Specifications : a. Capacity b. Temperature range c. Interior Dim ( W x D x H) d. Exterior Dim ( W x D x H) e. Temp Uniformity f. Temp Stability g. Power Output h. Voltage i. Temperature control Includes: 1. Warranty 2. Supplied with cord and plug set

: up to 1100 rpm or higher : up to 550 C : 220-240 V / 50-60 Hz

Unit : : : : : : : 5C 400C 10 % approx 26.0 x 26.0 cm 220 - 240 V / 6.40 A / 1530 W Ceramic Top Hot Plate or equivalent CSA, CE, NRTLc (CCSAUS) 28,7 x 41.1 x 10.1 cm 10 % Unit

37

38

Unit : : : : : : : : : 14 L 10 % up to 1100C 325 x 254 x 172 mm 10 % 485 x 510 x 533 mm 10 % 2.2C at 1000C 0.3C at 1000C max. 3100 Watts 12.9A / 208-240V Digital single set point control

: min 24 month

39

40

DISTILLATION UNIT*) Specifications : a. Kjeldahl nitrogen and protein distillation unit; Alcohol& phenol distillation; volatil acids & Cyanide determinations b. Acid-resistant pump c. Safety features: - Protective door - Sample tube detector - Service door sensor - Drip tray - Cooling water flow sensor d. Methods can be store : e. Tube volume : f. Recovery rate : g. Reproducibility (RSD) : h. Detection limit : i. Voltage : j. Steam Performance : k. Power consumption : l. Set of storage tank : DISTILLATION UNIT FOR ANALISYS SO2 WITH TITRATOR *) Specifications : A. Destillation Unit a. Aplikasi determination of sulfur dioxida. Can be used for Kjeldahl nitrogen and protein distillation unit; Alcohol& phenol distillation; volatil acids & Cyanide determinations , b. Acid-resistant pump c Safety features: - Protective door - Sample tube detector - Service door sensor - Drip tray - Cooling water flow sensor d Methods can be store e Tube volume f. Recovery rate g. Reproducibility (RSD) h. Detection limit i. Voltage j.. Power consumption k. Set of storage tank B. Titrator

Unit

9 or more min 250 ml (1 set) min. 99.5% minimum 1% minimum 0,1 mg 220 V - 240 V, 50/60 Hz 160 ml / 5 min or better 2200 W or less 2 x 20 liter Unit 1

: : : : : : :

50 or more min 250 ml (1 set) min. 99.5% 1% minimum 0,1 mg 220 V - 240 V, 50/60 Hz 2200 W or less 4 x 20 liter

a. End point titration automatic b. Display of consumtion of titration solution c. Display of pH value

41

VACCUM PUMP with GLASS FILTER HOLDER Each set consist of 1. Pump 2. Glass Filter Holder Pump Specifications : a. Presssure b. Power Supply Glass Filter Holder Specifications : a. Diameter b. Joint connection c. Funnel and flask d. Flask capacity e. stopper f. clamp g. fritted base Accessories 1. Silicon rubber laboratory tubing, compatible,L= 2 m 2. Support Screen

Unit : 1 unit : 1 unit

: 4 Bar : 220-240 V/ 50-60 Hz

: : : :

47 mm Compatible Glass 1000 ml

42

MIKRO BALANCES PROVIDE WITH ANTIVIBRATION TABLE *) Specifications : a. b. c. d. e. Weighing capacity Readability Repeatability Linearity Display : : : : : 6-7 g 0,001 mg 0,001 mg or better 0,004 mg or better touchscreen

Unit

Accessories 1. Antivibration Table (1 set) 2. External calibrator 2 and 5 gram E2 (1 set) 3. Printer (1 unit) 4. Filter weighing kit (1 set) 5. Antistatic brush (1 set) 6. Dust Cover (1 set)

7. Stabilizer 500 VA (1 unit)

Include of : 1. Instalation 2. Original operational manual book (plus bahasa indonesia) 3. Guarantee min 1 year 4. Training 43 THERMOHYGROMETER DIGITAL Specifications : a. Temperature Range b. Humidity Range c. Temperature Accuracy d. Humidity Accuracy Accessories 1. Battery compatible @ 2 pcs 2. With clock 3. accesory for wall mounting (screws or adhesive tape). Include of 1. Original Operation Manual (plus bahasa indonesia) 2. Guarantee min 6 months 44 STOP WATCH DIGITAL Specifications : a. 7 digits b. Measurable 10 hours c. Net time, split time, 1st/2nd place times Included of : 1. Guarantee min. 1 year 2. original operation manual (plus bahasa indonesia) 45 VACUUM FILTER HOLDER Spesifications a. autoclavable b. polysulfone or polycarbonate Unit 2 Unit : : : : : (-5)o to 55 oC or wider 20 -95 % or wider 1C 5 % rF (rh) 7

Unit

c. upper capacity at least 250 ml d. lower capasity at least 250 ml e. connection for vacuum pump f. With steril membrane 0.45 g. Membrane diameter 47mm 46 ULTRASONIC PIPPET WASHERS Spesifications a. dimension at least 300 x 400 x 800mm b. capacity at least 12 L for: 1 ml = 250 pcs 5 ml = 200 pcs 10 ml = 130 pcs c. temperatur water for cleaning solution up to 40degC with detergent d. vertical tank e. stainless steel cabinet Include of 1. Basket (mm) at least 2 Min guarantee 1 year 3. Trainning 4. Standar operating procedur (English and Indonesia) 5. Power supply 100V/ 50-60 hz 6 Step up & down transformer 7 Rack 8 Manual Pump 9 Water supply hose 10 Drain hose 11 Detergent Unit 1

146 (dia) x 550 (H)

47

AUTOMATIC COLONY COUNTING AND ANTIBIOTIC ZONE READER Spesifications a. Integral PC (all in one PC), CCD camera, and lens b. USB integral camera, Min camera 1.4m pixel c. Light shield 2 sliding doors to prevent excessive ambient light d. Complies with 21 CFR part 11, GLP e. With software and Database store all data f. Detection: fully automatic with manual over ride g. Count Modes : Separation of touching colonies, colur mode, shape mode, SRD zone sizing and statistical analysis h. Used standard up to 150 mm petri dish i. Smallest detectable colony at least 45 micron j. Inc mouse and keyboard, IQ/OQ documenting.

Unit

k. Trainning l. Min guarantee 1 year m. Standar operating prosedur (English and indonesian) 48 THERMOMETER AIR RAKSA SUHU -38 S/D 50C Spesifications a. made from glass with mercury b. temperature range -38 s.d 50C c. min guarantee 1 year d. standar operating procedure VORTEX MIXER Spesifications a. speed range up to 3000 rpm b. used for at least 1.5 ml test tube c. with analog display and control scale d. orbit diameter 4 mm e. speed control with knob for different modes of mixing f. power supply 230 V/ 50-60 hz g. min guarantee 1 year h. standar operating procedure (English and Indonesia) MINI PERSONAL CENTRIFUGE Spesifications a. Max speed b. Max rcf c. Rotor capacity d. Noise level e. Timer f. Power supply g. Power requerement h. Min guarantee 1 year i. Operating manual procedur (English and Indonesia) j. Inc with rotor SEROLOGI PIPET Spesifications a. baterry rechargeable b. autoclavable c. used for pipette types 1 ml to 100 ml d. at least 7 hours operation e. aspirating speed at least 25 ml in 5 second f. membrane filter replaceable g. touch control adjustable speed Unit 1

49

Unit

50

Unit : : : : : : : at least 12.000 rpm at least 14.000 rpm at least 12 x 1.5/2 ml test tube < 56 dB(A) up to 99 min 230 V/50-60 Hz up to 100 W

51

Unit

Include 1. 2. 3. 4. 52

holder Min guarantee 1 year Standar operating procedur (English and indonesia) Comfortable design for operation Unit : 1 unit 2

DIGITAL THERMOMETER WITH SENSOR Temperatur range -250degC to 400degC Spesifications inc: impact absorbing holster and two 80PK-1 beaded probe thermocouples a. temperatur range for -250degC to 400degC: b. Batteries 3 AA c. Accuracy above -100degC: 0.05% + 0.3degC d. Min guarantee 1 year e. Standar operating procedure (English and Indonesia)

53

THERMOMETER AIR RAKSA SUHU -10 S/D 50C Spesifications a. made from glass with mercury b. temperature range -10 s.d 50 C c. min guarantee 1 year d. standar operating procedure CABINET STERILIZING Spesifications a. with UV lamp b. dimensions c. accesories d. interior decoration of a stainless steel e. include fall prevention metal fittings f. Power supply AC 100 V 50/60 Hz g. Step up & down transformer h. with timer and analog display for clock i. number of sheet up to 10 j. 2 doors k. min guarantee 1 year l. Operating manual prosedur (english and Indonesian) m Trainning TURBIDIMETRI Spesifications

Unit

54

Unit

: at least 860 x 490 x 1750 mm

55

Unit

a. for measurement 0.0 up to 4.0 McFarland b. automated digital display when used c. data store software d. used flexible pipette from glass or plastic 12mm x 75 mm e. verification meansurement with standart kit Included of 1. Instrument kit - 4 baterai AAA alkaline - instrument user manual CD - product certified - RoHS certified 2. Standart kit - scale 0 McF, 0.5 McF, 2.0 McF, 3.0 McF 3. Min guarantee 1 year 4. Trainning 5. Standart operation procedur (English and indonesia) 56 SAFETY CABINET CLASS II *) Spesifications a. external dimensions max 1500 x 900 x 1700 mm b. HEPA filter with 99,999% efficiency at 0.1 to 0.3 microns or better c. microprossesor control system d. with UV lamp, support stand e. dual fan system f. stainless steel main body with stainless steel armrest g. epoxy coated h. sound emision < 60 dBA i. downflow velocity 0.3 0.1 m/s j. Inflow avarege 0.4 0.1 m/s Included of 1. Min guarantee 1 year 2. Trainning 3. Standar operation procedure (English and Indonesia) 57 COOLED INCUBATOR Spesifications a. Volume b. Temperatur range c. Interior chamber d. Over-and undertemperatur protecting with alarm e. Digital display f. Works calibration certificate g. interior dan exterior door Unit : 25010 liter : -12C to 60C5C : Stainless steel 2 Unit 1

h. i. j. k.

Fan Speed adjustable from 10 - 100 % in 100 % steps electtronic microprossesor temperatur controllor temperatur setup: 3 point calibration on controllor electrical supply 230 V/ 50-60 Hz

Included of 1. Min guarantee 1 year 2. Trainning 3. Operating manual prosedur (english and indonesian) 58 PCR HOOD Spesifications a. with UV lamp and timer b. HEPA filter H14 or >99.99% at 0.1 to 0.3 microns c. Internal dimension at least 900 (max 1200) x 500 x 500 mm d. external dimensions max 1200 x 700 x 1000 mm e. epoxy powder coated with antimicrobial coating f. pre filter g. air flow velocity average 0.3 m/s (60 fpm) h. power supply 220-230 v/ 50-60 hz Included of 1. Min guarantee 1 year 2. Trainning 3. Operating manual prosedur (english and indonesian) 59 STERITEST PUMP *) Spesifications a. pump with bottle holder and canister holder b. autoclavable c. power supply 230 V/ 50-60 Hz d. Materials of Construction Frame, pump head, Pump head cover, and bottle holder keypad and screen window: Drain Tray Drain Tray tubing e. Dimensions f. compatible with canister for sterility test Unit 1 Unit 1

: pump: approx. 35 x 30 x 25 cm : pump with bottle handle, canister handle approx. 59 x 34 x 53 cm

Included of 1. consumable and accesories Ampoule breaker for sterility test

(1 pk)

Canister with membran for sterility test for liquids in ampoules or collapsible bags (2 pk) Canister with membran for sterility test for antibiotic, 10 / pack (2 pk) Canister with membran for sterility test for solvent, creams, ointment (2pk) veterinary injectables, 10 / pack 2. Min guarantee 1 year 3. Trainning 4. Standar operation operation (English and Indonesia) 60 INKUBATOR KERING Spesifications a. compatible with thermo block for at least 20 x 1.5ml, 20 x 2 ml, 8 x 15ml, 4 x 50ml test tubes b. temperatur setting at least 4degC to 99degC c. temperatur range at least 30degC to 99degC d. digital timer at least 1 min to 90 hours e. incubating accuracy at least 1.0degC f. heating speed at least 5degC/min g. cooling speed at least 5degC/min between 99degC and 25degC h. power supply 220-230V/50-60 hz Included of 1. min guarantee 1 year 2. trainning 3. standar operation precedur (English and Indonesia) Assesories 1. thermoblock for test tube 1,5ml 2. thermoblock for test tube 2ml 3. thermoblock for test tube 15ml 4. thermoblock for test tube 50ml 5. box for all tube types, interior dimensions at least 13 x 8 x 5 cm 61 MICROCENTRIFUGE Spesifications a. speed b. max capacity c. timer d. dimension e. refrigerated f. temperatur control g. power supply 230 v/ 50-60 Hz h. LCD display Included of 1. rotor for 1.5/2ml, 15ml, 50ml, microplate rotor Unit : min 16,000 rpm : at least 24 x 1.5/2ml, 6 x 50ml ml : up to min 60s : max 40 x 70 x 35 cm : at least -10degC to 40degC 1 Unit 1

2. Trainning 3. Min guarantee 1 year 4. Operating manual prosedur (English and Indonesia) 62 PIPET MIKRO DAN ACCESSORIES ukuran 0,5 - 10 l 10 - 100 l 100 - 1000 l 1000 l 0,5 - 5 ml 1 - 10 ml Stands pipette Spesifications a. autoclavable b. daya tahan yang tinggi terhadap asam, basa, dan tumbuhnya jamur c. stans pipette at least for 6 pipet d. adjustment display to a spesific volume e. min guarantee 1 year

Unit Unit Unit Unit Unit Unit Unit

1 2 1 1 1 1 1

MEREK / TYPE

Você também pode gostar