Escolar Documentos
Profissional Documentos
Cultura Documentos
Method Development
Introduction
This work describes the use of Fusion AE™ for Galaxie™ - an integrated method development
software application that enables a true Quality by Design approach to develop and optimize a
biopharmaceutical HPLC separation. This method required the separation a single large
product peak from several co- or closely eluting impurity peaks in the shortest time possible.
The legacy method used to separate these components involved a very long (over 60 minute)
two-step gradient method that did not meet all method performance requirements, including
peak resolution and robustness.
USP Resolution: 1.5 — 2.5, USP Tailing: 0.8 — 1.5, Robustness Cp: ≥ 1.0.
Figure 1. Experimental Design Template -
experimental parameters (study variables)
including Flow Rate, Gradient Slope,
Gradient Time and Column Temperature
were entered into a standardized template.
40.0
52.5
65.0
Figure 6. Single Response Surface plots for Peak 3 data showing the effect of
increasing flow rate and column temperature on resolution
Size of unshaded
region around
optimizer answer
indicates very
stable method in
terms of meeting
goals.
1. Responses were evaluated and modeled for all four critical peaks. All models fitted the
data.
(all coefficients are statistically significant, model prediction error ≈ experimental error). (3)
A Resolution Cp value was computed for each run using a modified Monte Carlo simulation
approach (8) that employed (a) the Resolution models obtained from analysis of the
experimental results, (b) user-defined tolerance limits on the instrument parameters, and
(c) the user-defined acceptance limits for the response.
Robustness Cp – ≥1.0.
(means ±3σ of Resolution variation is within acceptance limits)
4. Software Automated Optimizer answer was a simple (one-step) 30 minute gradient, flow
rate of 1.4ml.min, initial organic modifier concentration of 20% and column temperature
o
setting of 46.7 C.
a)
Figure 8. Chromatograms
showing increasing degree of
resolution between product
(Peak 2) and impurity peaks.
Method parameters for each
run were:
Chromatogram 1, 2
a) 1 ml.min, 10 minute
gradient, 32% initial organic,
o
52.5 C
b) 0.5 ml.min, 50 minute
gradient, 20% initial organic,
52.5 oC
c) 1.5ml.min, 30 minute b)
gradient, 32% initial organic,
52.5 oC
c)
Figure 9. Chromatogram resulting from a partially optimized method showing baseline
resolution of impurity peaks from product peak.
Conclusions
Fusion AE for Galaxie was able to successfully develop a HPLC method, optimized for
flow rate, gradient time, percent organic modifier and column temperature, that met all
the initial biopharmaceutical product purification separation goal settings.
Single response surface plots indicated that the major peak resolution response
effectors were a combination of high flow rate and elevated column temperature.
The Overlay Graphics plot showed that, at the predicted optimum method settings, the
method was extremely robust with respect to gradient time and column temperature
settings.
Acknowledgements
The Authors would like to thank Ms Shan Lin for her assistance with this project and
George Cooney, S-Matrix Corp., Eureka, CA, for his help with data analysis.