Você está na página 1de 125

X-Ray Energy Dispersive Spectroscopy In the Electron Microscope

Nestor J. Zaluzec
zaluzec@microscopy.com zaluzec@aaem.amc.anl.gov

Micro-Analytical Capabilities

High Spatial Resolution Quantitative Microanalysis

Brief Review of X-ray Generation Instrumentation: Detector Systems Instrumentation: EM Systems Data Analysis and Quantification: Additional Topics

Electron Excitation of In ner Shell Processes

Incident Electron

The Emission Process:


Inelastically Scattered Primary Electron Ejected Inner Shell Electron

1-Excitation 2-Relaxation 3-Emission

Electron Distribution Relaxation

Internal Conversion and Auger Electron Emission

X-ray Photon Emission

L shell
1 2 3

M shell
12 345

d-band s-band

DOS

XEDS/XRF

}
EELS/XAS Incident Electrons

AES XPS/UPS

Incident Photons

Schematic Diagram Illustrating Sources of Signals Resulting from Inelastic Scattering

Experimental XEDS, XPS, and EELS data from the Copper L shell. Note the differences in energy resolution, and spectral features. features

Nomenclature for Principle X-ray Emission Lines

L# K! K# K" L"

L!

M"

M# N Shell

M Shell L Shell K Shell

Characteristic X-ray Line Energy

= E final - E initial

Recall that for each atom every shell has a unique energy level determined by the atomic conguration for that element.

! X-ray line energies are unique.

Relative Intensities of Major X-ray Lines K"1 K"2 K#1 K#2 K#3 = = = = = 100 50 15-30 1-10 6-15 L" 1 L" 2 L #1 L #2 L #3 L #4 L $1 L $3 L% L& = = = = = = = = = = 100 50 50 20 1-6 3-5 1-10 0.5-2 1 1-3 M"1,2 = 100 M# = 60

Mosleys Law

Characteristic X-Ray Spectrum Illustrating KLM lines

Barium L

Copper K

Yttrium K

Oxygen K Barium M Copper L Yttrium L

10

12

14

16

18

20

Energy (keV)
TEM Specimen: Y1Ba2 Cu3O6.9 Superconductor - 120 kV - UTW Detector Note: As Z increases the Kth shell line energy increases. If K-shell is excited then all shells are excited (Y, Cu, Ba) but may not be detected. Severe spectral overlap may occur for low energy lines.

Electron Excitation also generates Continuum (background) signal

Intensity
0

Photon Energy

!o

Energy Range - Continuous Distribution Maximum = Incident Electron Energy (Least Frequent) Minimum = E plasmon~ 15-30 eV (Most Frequent) Spectral Distribution will reflect this range, modified by detector response function

Electron Excitation of Continuum (Background) Intensity


Oxygen K Barium M Copper L Yttrium L

Barium L

Copper K

Yttrium K

Intensity
0

10

12

14

16

18

20

Energy (keV)
Spectral background will be influenced by: 1.) Specimen composition 2.) Detector efficiency 3.) TEM generated artifacts

Instrumentation: Detector Systems


Wavelength Dispersive Spectrometers (WDS) Energy Dispersive Spectrometers (EDS) Si(Li) Si(Li) Detectors HPGe Detectors Spectral Artifacts of the EDS System Detector Efficiency Functions Light Element Detectors Superconducting Calorimeters/Bolometers Silicon Drift Detectors Multichannel Analyzers

Energy Dispersive Spectrometers: (Solid State Detector) Operates on Energy Deposition Principle
Simple, Nearly Operator Independent Large Solid Angles (0.05-0.5 sr) sr) Virtually Specimen Position Independent No Moving Parts Parallel Detection Quantification by Standardless or Standards Methods Poor Energy Resolution (~ 130 eV) eV) ** SuperConducting Systems ( ~ 20 eV) eV) Poor Peak/Background Ratios ( 100:1) Detection Efficiency Depends upon X-ray Energy

Wavelength Dispersive Spectrometers : (Diffractometer ) (Diffractometer) Operates using Diffraction Principles (Bragg's Law) Excellent Energy Resolution (~ 5 eV) eV) High Peak/Background Ratios (10000:1) Good Detection Efficiency for All X-rays High Counting Rates Good Light Element Capabilities Complex Mechanical Devices, Operator Intensive Specimen Height dependant focus Moving Components in the AEM Limited Solid Angles (<0.01 sr) sr) Serial Detection Quantification Requires Standards

WDS system in a TEM EMMA-4 System

Circa 1972

Modern Applications WDS Valence Band/Electronic Structure Measurements


TEM

CCD
Grating

Diamond C K-emission spectrum

Enery (eV)

M. Terauchi , M. Kawana Tohoku University

Y. Ito NIU Physics

Comparison of EDS and WDS Spectrometers


Parameter Construction Energy Resolution Efficiency Input Count Rate Peak/Background* Atomic Number Range Number of Elements Solid Angle Collection Time Beam Current Detector Stability Spectral Artifacts Operation Wavelength Dispersive Mechanical Device moving components 5 eV < 30 % 30-50 K cps 10000 Z > 4 (Be) 1 per Detector 0.001-0.01 sr Tens of Minutes High Stability Required Good Short Term Neglegible Skilled (?) Energy Dispersive Solid State no moving parts 130 eV 100 % (3-15keV) 10 K cps 100 Z > 11 (Na) Z > 5 (B) All in Energy Range 0.02-0.3 sr Minutes Low Stability Required Excellent Important Novice

* Values depend on definition, specimen, and operating conditions

How is the X-ray Signal Measured?

Properties of Intrinsic Silicon . Attaching HV electrodes to the two surfaces the Si(Li) Si(Li) crystal will act similiar to a capacitor with free charges developing on the electrical contacts. Charge developed in the crystal is N = E/'. (E= x-ray Energy, ' = 3.8 eV/e-h pair) ==> 10 kV X-ray produces ~2630 electrons = 4.2 x 10 -16 Coulombs.

Solid State Detector Construction


-HV

X-Rays

Intrinsic (Active) Zone

LN2 Cold Finger

FET

Dead Layer p-type SS Cryostat Housing Au Electrical Contact Dead Layer n-type Environmental Isolation Window (Be, Hydro-Carbon,Windowless)

Relative Detection Efficiency Solid State Detectors: Si(Li) or Instrinsic (High Purity) Ge Using a simple absorption model define the relative detector efficiency '(E) by the following procedure:
x

T = I o exp (- x ) = I o exp (-

[ ]! !

x)

Calculated Si(Li) Detector Efficiency by Active Layer Thickness & Window Type

Windowless vs. Conventional Detectors Comparision of XEDS measurement on NiO using a Windowless versus Beryllium Window detector
1.00

Ni L
Normalized Intensity (Arb. Units)
0.80

Windowless XEDS Detector

Ni K

0.60

OK
0.40

Beryllium Window XEDS Detector


0.20

0.00 0 2 4 6 8 10

Energy (keV)
Note the enhanced detection efficiency below 1 keV for the WL detector. Both spectra are normalized to unity at the Ni K" Line (7.48 keV)

Windowless vs. vs. Conventional Detectors

K Shell Spectra using Windowless Detector Boron -> Silicon

L Shell Spectra Using Windowless Detector Titanium ->Zinc Note Potential Overlaps with K shell Lines

Comparision Light Element Spectroscopy Resolution XEDS

'= 3.8 eV (in Si) Si) / 2.9 eV (in Ge) Ge) -> But the electrons produced are in a Poissson Distribution this gives rise to a spread in the number of electron F = Fano Factor ~ 0.1 E= X-ray Energy Noise = Electronic Noise (mainly in the FET) Nominal FWHM Values in Modern Si(Li) Si(Li) Detectors: O K" (0.52 keV) = 80 to 100 eV keV) Mn K" (5.9 keV) = 140 to 160 eV keV) Mo K" (17.5 keV) keV) = 210 to 230 eV

Comparison of EDS and WDS Spectra

Resolution will also vary with Microphonic & Electronic Noise, and Counting Rate!

1000.00 High Noise 800. 00

Intensity

600. 00 400. 00 200. 00 0.00 0 0.4 0.8 1.2

Intermediate

None

1.6

Energy (keV)

WL & UTW detectors are particuliarly sensitive to low energy noise and microphonics. microphonics. Observe the changes in the spectra ( width of the peaks)

Silicon Drift Detector

~ 0.3 mm

X-rays

Both sides are reversed biased Electrons travel along the central potential well Radial drift gradiant sweeps electrons to the Anode

Silicon Drift Detector Construction

Detector Area = 50 mm2 Peltier Cooled -> No LN2 Low Capacitance (250 fF)

Silicon Drift Detector Construction

(ANL ~ 0.41 sr

Resolution vs Count Rate

350

Tc = 1 sec
FWHM Resolution (eV)
300

250

200

150 Si Ka 100 0 50 100 150 200 250 Cu Ka Mo Ka

Input Count Rate (Kcps)

Count Rate = ~ 200 Kcps Time = 100 sec Integral = 20, 000, 000 counts !

Mo-Si/Cu Slot Grid

Dead Time vs Count Rate

Si(Li)

SDD

Si(Li) vs SDD

0.8

Relative Efficiency

0.6

0.4 SDD - 0.35 mm 0.2 SDD - 1.0 mm Si(Li) - 3.0 mm

0 0 5 10 15 20 25 30

Energy (eV)

UTW Comparison Si(Li) vs SDD

Instrumentation: AEM Systems


The AEM as a system Spectral Artifacts in the AEM Uncollimated Radiation Systems Peaks Artifacts at High Electron Energy Specimen Contamination & Preparation Optimizing Experimental Conditions

Spectral Artifacts in the AEM Uncollimated Radiation: The Hole Count

Electrons X -r ay s

Fixed First Condensor Aperture Variable Second Condensor Aperture

Upper Objective Pole Piece

Specimen and Goniometer Stage

Objective Aperture Lower Objective Pole Piece

Spectral Artifacts in the AEM Uncollimated Radiation Solutions

Electrons X -r ay s

Electrons X -r ay s Thick Fixed

Fixed First Condensor Aperture Variable Second Condensor Aperture

First Cond ensor Aperture

Thick

Varia b l e

Second Con densor Aperture

Non-Beam

De f i n i n g

Collimator
Upper Objective Pole Piece

Upper Objective Pole Piece

Specimen and Goniometer Stage

Specimen and Goniometer Stage

Objective Aperture Lower Objective Pole Piece

Objective Aperture Lower Objective Pole Piece

Spectral Artifacts in the AEM


Hole Count Effects: Modified C1 and C2 Apertures
e-

Hole Count Effects: Modified C1 and C2 & NonBeam Defining Apertures


e-

Specimen

Specimen

15000

Ni K

Intensity

Intensity

10000

Ni4Mo Specimen Thick C1/C2 Apertures

15000

Ni K Ni K

10000

5000

Mo L Ni L Fe K Mo K

5000

Mo L Mo L Ni L Ni L

Ni4Mo Ni4Mo Specimen Specimen Thick C1/C2 Thick C1/C2 Apertures Apertures Non-Beam Non-Beam Defining Defining Aperture Aperture Mo K Mo K

Fe K Fe K 5 10 15

0 1000 0 5 10 15 Ni4Mo Hole Count Thick C1/C2 Apertures 20

0 0 300 20 Ni4Mo Hole Count Thick C1/C2 Apertures Non-Beam Defining Aperture

Intensity

500

Intensity

200

100

0 0 5 10 15 20

0 0 5 10 15 20

Energy (keV)

Energy (keV)

Optimizing Experimental Conditions


Choice of X-ray Line K- series L- series M- series Detector/Specimen Geometry Elevation Angle Solid Angle Detector Collimation Choice of Accelerating Voltage Relative Intensity Peak/ Background Systems Peaks/Uncollimated Radiation Choice of Electron Source Spatial Resolution Tungsten Hairpin LaB6 Field Emission

Detector/Specimen Geometry
eCharacteristic X-ray Emission Characteristic Angular Dependance ~IsotropicDistribution) (Isotropic

XEDS

Collimator

High Energy

Low Energy

Bremsstrahlung Continuum Angular Dependance (Aniostropic Distribution ) Highly Anisotropic

Detector/Specimen Geometry TEM


e -

!E
Specimen
Eucentric Height

Designation

Elevation Angle )E 0o 0o 15-30o 68-72o

Azimuthal Angle )A 45o 90o 90o 0o

Manufacturer

Low Intermediate High

JEOL JEOL, FEI, VG FEI, JEOL, Hitachi,VG Hitachi, JEOL

Geometrical / Collection Solid Angle

2# a 2 + d 2 $ d a 2 + d 2 S "= 2 = R a2 + d2

Geometrical / Collection Solid Angle

10

2*

Solid Angle (sR)

50 mm^2 30 mm^2 10 mm^2 1 mm^2 0.3mm^2 0.2 mm^2 0.1 mm^2 0.1

0.01

0.01

0.1

1 Distance (mm)

10

2# a 2 + d 2 $ d a 2 + d 2 S "= 2 = R a2 + d2

Comparison of AEM Systems with XEDS Detectors ANL/UIUC/NU

X-ray Collection Efficiency

0.30 0.25 0.20 0.15 0.10 0.05 0.00 0 2000 4000 6000 8000 10000

X-Ray Photon Energy (eV) ANL AAEM ------- WL ANL CM30 ------- WL UIUC & ANL ---BE

UIUC ---------- UTW NU HF2000 ----- UTW

Cu? In Steel ? What is the source

Subtending Solid Angle

Where do Systems Peaks Come from?

Detection of System Peaks Effects of the Collimator & Stage

Detection & Removal of System Peaks

Removal of Stage System Peaks by use of Beryllium Gimbals Ge specimen 10,000 in Ge K" peak in both spectra Left Standard Single Tilt Cu Stage, Right Be Gimbal DT Stage

k,k* = Constants Px = Characteristic Signal from element X (P/B)x = Peak to Background ratio for element X Io = Incident electron flux Jo = Incident electron current density do = Probe diameter T = Analysis time

Experimental Peak/Background Variation with Voltage

Visualizing Minimum Detectable Mass

1000

What is your spatial resolution?

Spatial Resolution /Beam Spreading Monte Carlo Calculations

Al

DC Joy's MC Program
100 kV

Aluminium

Gold

Al

Al

Au

500

Au

2000

100 kV

400 kV

100 kV

400kV

Monte Carlo Calculations of B (Newbury & Myklebust -1979) Element Carbon Aluminium Copper Gold Z 6 13 29 79 10nm 0.22 0.41 0.78 1.71 Thickness 50nm 100nm 1.9 3.0 5.8 15.0 4.1 7.6 17.5 52.2 500nm 33.0 66.4 244.0 1725.0

Element Carbon Aluminium Copper Gold

Z 6 13 29 79

10nm 0.16 0.26 0.68 15.5

50nm 1.8 1.9 7.6 17.3

100nm 5.13 8.12 21.4 *

500nm 57.4 90.9 * *

*model invalid at higher kV and/or high scattering angles

What are the Limits - Today?

Specimen Contamination
Example:
The figure at the right shows the results of contamination formed when a 300 kV probe is focussed on the surface of a freshly electropolished 304 SS TEM specimen. The dark deposits mainly consist of hydrocarbons which diffuse across the surface of the specimen to the immediate vicinity of the electron probe. The amount of the contamination is a function of the time spent at each location. Here the time was varied from 15 - 300 seconds.
15 sec 30 sec 60 sec 120 sec 300 sec

Plasma Cleaning of Electropolished 304 SS


Untreated Specimen

After 5 minutes Argon Processing

After 5 minutes of additional Oxygen Processing

Specimen Preparation Artifacts

Electropolishing: HCl residue

Ion Milling:Redeposition of Fe,Ni, Cu from SS holder

Data Analysis and Quantification: Spectral Processing Thin Film Quantification Methods Specimen Thickness Effects: Absorption Fluorescence

Spectral Processing : XEDS Spectrum = Characteristic Peaks + Background


Data Reduction Simple: Linear Background Fit & Integration Curve Fitting: Non-Linear Background & Profile Matching Frequency (Digital) Filtering: Background Suppression & Reference Spectra Fitting Deconvolution: Deconvolution: Fourier Method for Resolution Enhancement Background Modeling Simple - Linear and/or Polynominal Interpolation Modeling - Parametric Fits of Analytic Expressions Phenomenological Expressions Modified Bethe Heitler Model Digital Filtering - Mathematical Supression

Spectral Processing : XEDS Simple Data Reduction

Note: Must use peak integrals (I) and not peak amplitudes (A) Recall that for a Gaussian Peak

Hence for the ratio of Intensities

Spectral Processing : XEDS Background Modeling : Power Law/Parametric Fits

Polynominal expansion of Kramers Law

Spectral Processing : XEDS Curve Fitting : Linear Modeling


! A A
1 2

Using simple matrix algebra solve for A Fast and simple procedure Presumes operator knows all elements present * System must be calibrated (i.e. Ec and +E must be accurately known)

Spectral Processing : XEDS Curve Fitting : Linear Modeling

Spectral Processing : XEDS Digital Filtering Background Suppression by Mathematical modeling - Replace Data by new spectra formed by the following linear operation.

Operator independent Introduces severe spectral distortion

Spectral Processing : XEDS Digital Filtering

Data Analysis and Quantification: Spectral Processing Thin Film Quantification Methods Specimen Thickness Effects:

Okay we can detect elements How do we Quantify things?

X-Ray Production

Quantitative Analysis Equations


For a thin specimen

IA + , W No . C %o t ' (

= = = = = = = = = = = =

Measured x-ray intensity per unit area Kth-shell ionization cross-section Kth-shell fluorescence yield Kth-shell radiative partition function Atomic Weight Avagodro's number Density Composition (At %) Incident electron flux Specimen thickness Detector efficiency Detector solid angle

Ionization Cross-Section

X-ray Fluorescence Yield has Systematic Variation With Atomic Number

, K shell

, K vs , L shell

Quantitative Analysis using XEDS Standardless Method Invoke the Intensity Ratio Method, that is consider the ratio of x-ray lines from two elements

This simple equation states that the relative intensity ratio of any two characteristic x-ray lines is directly proportional to the relative composition ratio of their elemental components multiplied by some "constants" and is independent of thickness. NOTE: The kAB factor is not a universal constant!! Only the ratio of /A//B is a true physical constant and is independant of the AEM system. The ratio of 'A/'B is not a constant since no two detectors are identical over their entire operational range. This can cause problems in some cases as we shall see.

The analysis to this point has only yielded the relative compositions of the specimen. We need one additional assumption to convert the relative intensity ratio's (Ii/Ij) into compositions namely:

One now has a set of N equations and N unknowns which be solved algebraically solved for the individual composition values. Thus for a simple two element system we have: and or

Solving for CB and CA

Efficiency

Variation in Measured Composition on 308 SS for Different Labs Example in which K-factor is stable Cr, Fe, Ni Note: Detector efciency ~ 100% in this energy range

1.0

Si(Li)

0.8

Detector

0.6

HP Ge

Detector

Parameters 0 nm 20 nm 100 nm 3 mm 200 nm 3 mm

0.4

Calculated

0.2

Be Window: Gold Contact: Si Dead Layer: Si Active Layer: Ge Dead Layer: Ge Active Layer:

0.0 0 1 2 3 X-ray 4 5 6 Energy 7 (keV) 8 9 10

Photon

Variation in K-factor with AEM/Detector System Specimen: Uniform NiO lm on Be Grid

Instrument

Experimental K -Factor 5.17 1.56 1.25 1.25 1.24 1.13 1.10 1.03 0.74
1.0

Apparent Variation in Composition 80.9/19.1 56.1/43.9 50.6/49.4 50.6/49.4 50.4/49.5 48.1/51.9 47.4/52.6 45.8/54.2 37.8/62.2
Si(Li)

1 2 3 4 5 6 7 8 9

Efficiency Detector

From: Comparison of UTW/WL X-ray Detectors on TEM/STEMs and STEMs Thomas, Charlot, Charlot, Franti, Franti, Garratt-Reed, Garratt-Reed, Goodhew, Goodhew, Joy, Lee, Ng, Plicta, Plicta, Zaluzec. Zaluzec. Analytical Electron Microscopy-1984

0.8

0.6

HP Ge

Detector

Parameters 0 nm 20 nm 100 nm 3 mm 200 nm 3 mm

0.4

Calculated

0.2

Be Window: Gold Contact: Si Dead Layer: Si Active Layer: Ge Dead Layer: Ge Active Layer:

0.0 0 1 2 3 X-ray 4 5 6 Energy 7 (keV) 8 9 10

Photon

Determining the kAB-1 Factor Experimental Measurements Prepare thin-lm standards of known composition then measure relative intensities and solve explicitly for the kAB factor needed. Prepare a working data base. This is the "best" method, but - specimen composition must be veried independently - must have a standard for every element to be studied Theoretical Calculations Attempt rst principles calculation knowing some fundamental parameters of the AEM system Start with a limited number of kAB factor measurements, then t the AEM parameters to best match the data. Extrapolate to systems where measurements and/or standards do not exist.
Method 1. (Goldstein etal) Assume values for -,,,' and determine the best s to t kAB. This procedure essentially iterates the t of s to the data. Method 2. (Zaluzec) Assume values for -,,,+ determine the best e to t kAB. This procedure essentially iterates the t of e (detector window parameters) to the data.

Sources of values for kAB Calculations W -(K) - International Tables of Atomic Weights - Schreiber and Wims , X-ray Spectroscopy (1982) Vol 11, p. 42 - Scoeld, Atomic and Nuclear Data Tables (1974) Vol 14, #2, p. 121

-(L)

, (K) - Bambynek etal, Rev. Mod. Physics, Vol 44, p. 716 Freund, X-ray Spectrometry, (1975) Vol 4, p.90 ,(L) - Krause, J. Phys. Chem. Ref. Data (1974) Vol 8, p.307 - Inokuti, Rev. Mod. Physics, 43, No. 3, 297 (1971) - Goldstein etal, SEM 1, 315, (1977) - Chapman etal, X-ray Spectrometry, 12,153,(1983) 12,153,(1983) - Rez, X-ray Spectrometry, 13, 55, (1984) - Egerton, Ultramicroscopy, 4, 169, (1969) - Zaluzec, AEM-1984, San Fran. Press. 279, (1984) - Use mass absorption coefcients from: -Thinh and Leroux; X-ray Spect. (1979), 8, p. 963 -Henke and Ebsiu, Adv. in X-ray Analysis,17, Analysis,17, (1974) -Holton and Zaluzec, AEM-1984, San Fran Press,353,(1984)

+(Eo)

' (E)

Quantitative Analysis using XEDS


Thin Film Standards Method
Invoke the Intensity Ratio Method, but now consider the ratio of the same x-ray line from two different specimens, where one is from a standard of known composition while the other is unknown:

This simple equation states that the relative intensity ratio of same characteristic x-ray line is directly proportional to the relative composition ratio of the two specimens multiplied by a some new parameters. % = incident beam current . = local specimen density t = local specimen thickness

X-ray Absorption

Quantitative Analysis using XEDS : Absorption Correction

Quantitative Analysis using XEDS Specimen Thickness Effects For finite thickness specimens, what is a thin film? Previous Assumptions: Energy loss, X-ray absorption, No X-ray fluorescence

X-Ray Fluorescence Correction

Specimen Homogenity
In this and all other derivations we have assumed that over the excited volume, as well as along th exiting pathlength, pathlength, the specimen is homogeneous in composition. If this assumption is invalid, one must reformulate the absorption correction and take into account changes in : /., ., and t along the exiting pathlength. pathlength. Effects of Beam Broadening Parallel Slab Model: No Change in absorption pathlength Wedge Model: There is a correction the magnitude of which varies with the wedge angle. Effects of Irregular Surface This cannot be analytically modeled but must be understood!

Additional Topics Heterogeneous Specimens Composition Proles Electron Channeling Radiation Damage Spectral Imaging

Grain Boundary Segregation

Cr

Cr

Ni

Ni

Fe

Fe

-200

Distance (nm)

+200

-200

Distance (nm)

+200

C *(x,y) = C(x,y,z)* d(x,y,z)


C * (x,y) = Apparent profile measured C(x,y,z) = Actual composition profile d(x,y,z) = Incident beam profile * = Convolution operator F,F- 1 = Fourier and Inverse Fourier Transforms In the 2 dimensional limit one can deconvolute the measured profile using: $ "F{ C * (x,y)}' " # & C(x,y) = F-1" ! F{ d(x,y) } " % Realistically, it is better to decrease the probe diameter and specimen thickness

Electron Channeling Induced X-ray Emission

Characteristic X-ray Emission is not truly isotropic in crystalline materials! Original Observations of Effect
Duncumb 62, Hall 66, Cherns etal 73

Predicted Applications
Cowley 64, 70

ACHEMI Technique Tafto 79, Spence & Tafto 83

Multi-Variate Statistical Analysis Rossouw etal , Anderson and others


late80 late80-90 -90s

High Angular Resolution Electron Channeling X-ray Spectroscopy

Orientation Dependance in Homogeneous Alloys

Applications in Ordered Systems

HARECXS - Mo K/L Shell Expt Calculated K Shell

Expt Calculated L Shell


with M. Oxley, L. Allen - Melbourne

ALCHEMI Atom Location by CHanneling EMIssion Tafto & Spence - Science 1982

Spectral Imaging

20nm

FEI Tecnai F20

How much data is generated in PRD Expt's


Compare with Spectrum Imaging

20nm

XY E 256 512 1024

1282

2562

5122

10242

4M 8M 16M

16M 32M 64M

67M 134M 268M

262M 524M 1G

Energy

Spectrum Imaging allows for Off-line Projections/Summations Operations

Po

sit

/ ion

An

gle

Position/Angle

E1 E2 E3 E4 E5 En

But it requires massive data strorage

XY E 256

2 128

256

2 512

2 1024

4M

16 M

67 M

262 M

512 Spectrum Imaging 1024

8M

32 M

134 M

524 M

16 M

64 M

268 M

1G

Energy

it os

/ ion

An

gle

Position/Angle XY t 1s 2 128 2 2 512 2 1024

256

E1 E2 E3 E4 E5 En Spectrum Imaging

16 ms

65 ms

0.26 s

1.05 s

1ms

16 s

65 s

4.4 min ~3 days ~6 mnths

17.5 min ~12 days ~2 yrs

1s

~4.5 hrs ~11 days

~18 hrs ~1.5 mnths

1 min

Você também pode gostar