Application The Coriolis measuring principle operates independently of physical fluid properties, such as viscosity and density. Extremely accurate measurement of liquids and gases such as oils, lubricants, fuels, liquefied gases, cleaning agents and solvents, sterile media (blood plasma), foodstuffs and paints Fluid temperatures up to +150 C Process pressures up to 100 bar Mass flow measurement up to 180 t/h Approvals for hazardous area: ATEX, FM, CSA, TIIS Approvals in the food industry/hygiene sector: 3A, EHEDG Connection to all common process control systems: HART, PROFIBUS PA/DP, FOUNDATION Fieldbus, MODBUS Relevant safety aspects: Secondary containment (up to 40 bar), Pressure Equipment Directive, SIL-2
Your benefits The Promass measuring devices make it possible to simultaneously record several process variables (mass/density/temperature) for various process conditions during measuring operation. The uniform Proline transmitter concept includes: Modular device and operating concept resulting in a higher degree of efficiency Software options for batching and concentration measurement for extended range of application Diagnostic ability and data back-up for increased process quality The Promass sensors, tried and tested in over 100000, offer: Multivariable flow measurement in compact design Insensitivity to vibrations thanks to balanced singletube measuring system Efficient protection against forces from piping thanks to robust construction Easy installation without taking inlet and outlet runs into account
TI075D/06/en/12.06 71036053
Table of contents
Function and system design. . . . . . . . . . . . . . . . . . . . . 3
Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Mechanical construction . . . . . . . . . . . . . . . . . . . . . . 22
Design / dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Material load curves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Process connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 37 38 38 40
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Operable flow range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Input signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Human interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Display elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Unified control concept for both types of transmitter . . . . . . . . . . Language groups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Remote operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 41 41 41
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Low flow cut off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Galvanic isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Switching output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Electrical connection Measuring unit . . . . . . . . . . . . . . . . . . . . . 10 Electrical connection, terminal assignment . . . . . . . . . . . . . . . . . 11 Electrical connection Remote version . . . . . . . . . . . . . . . . . . . . . 12 Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Cable entries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Cable specification Remote version . . . . . . . . . . . . . . . . . . . . . . . 13 Power consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Power supply failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Potential equalization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Performance characteristics. . . . . . . . . . . . . . . . . . . . 13
Reference operating conditions . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Maximum measured error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Repeatability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Influence of fluid temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Influence of fluid pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Endress+Hauser
A B 1 2
A B 3
A B
a0003383
The phase difference (A-B) increases with increasing mass flow. Electrodynamic sensors register the tube oscillations at the inlet and outlet. The system balance required for proper measurement is created by exciting an eccentrically arranged swinging mass to antiphase oscillation. This patented TMB system (Torsion Mode Balanced System) guarantees perfect measurements, even in changing process and environmental conditions. Therefore, the device is just as easy to install as the familiar two-tube systems! Consequently, no special measures for attachment are required in front of or behind the sensor. The measuring principle operates independently of temperature, pressure, viscosity, conductivity and flow profile. Density measurement The measuring tube is continuously excited at its resonance frequency. A change in the mass and thus the density of the oscillating system (comprising the measuring tube and fluid) results in a corresponding, automatic adjustment in the oscillation frequency. Resonance frequency is thus a function of fluid density. The microprocessor utilizes this relationship to obtain a density signal. Temperature measurement The temperature of the measuring tube is determined in order to calculate the compensation factor due to temperature effects. This signal corresponds to the process temperature and is also available as an output.
Endress+Hauser
Measuring system
The measuring system consists of a transmitter and a sensor. Two versions are available: Compact version: transmitter and sensor form a mechanical unit. Remote version: transmitter and sensor are mounted physically separate from one another. Transmitter
Promass 80 Two-line liquid-crystal display Operation with push buttons
Esc
a0003671
Promass 83
Esc
Four-line liquid-crystal display Operation with Touch control Application-specific Quick Setup Mass flow, volume flow, density and temperature measurement as well as calculated variables (e.g. fluid concentrations)
a0003672
Sensor
I Straight single-tube instrument. Minimal shear stress on fluid, hygienic design, low pressure loss Nominal diameters DN 8 to 80 Tube material: titanium Documentation No. TI 075D/06/en
a0003678
a0003679
General purpose sensor, ideal replacement for volumetric flowmeters. Nominal diameters DN 8 to 50 Tube material: stainless steel
a0002271
Endress+Hauser
Universal sensor for fluid temperatures up to 200 C. Nominal diameters DN 8 to 250 Tube material: stainless steel or Alloy C-22
a0003673
F (High-temperature)
Universal high-temperature sensor for fluid temperatures up to 350 C. Nominal diameters DN 25, 50, 80 Tube material: Alloy C-22
a0003675
Robust sensor for extreme process pressures, high requirements for the secondary containment and fluid temperatures up to 150 C Nominal diameters DN 8 to 80 Tube material: titanium
a0003676
Single bent tube. Low pressure loss and chemically resistant material Nominal diameters DN 8 to 50 Tube material: zirconium
a0003677
Single bent tube. Hygienic design, low pressure loss, for fluid temperatures up to 150 C Nominal diameters DN 8 to 50 Tube material: stainless steel
a0006828
Endress+Hauser
Input
Measured variable Mass flow (proportional to the phase difference between two sensors mounted on the measuring tube to register a phase shift in the oscillation) Fluid density (proportional to resonance frequency of the measuring tube) Fluid temperature (measured with temperature sensors) Measuring ranges for liquids
DN 8 15 15 FB 25 25 FB 40 40 FB 50 50 FB 80 FB = Full bore versions of Promass I Range for full scale values (liquids) gmin(F) to gmax(F) 0 to 2000 kg/h 0 to 6500 kg/h 0 to 18000 kg/h 0 to 18000 kg/h 0 to 45000 kg/h 0 to 45000 kg/h 0 to 70000 kg/h 0 to 70000 kg/h 0 to 180000 kg/h 0 to 180000 kg/h
Measuring range
Measuring ranges for gases The full scale values depend on the density of the gas. Use the formula below to calculate the full scale values: mmax(G) = mmax(F) (G) / 160 [kg/m3] mmax(G) = max. full scale value for gas [kg/h] mmax(F) = max. full scale value for liquid [kg/h] (G) = Gas density in [kg/m3] at operating conditions Here, mmax(G) can never be greater than mmax(F) Calculation example for gas: Sensor type: Promass I, DN 50 Gas: air with a density of 60.3 kg/m3 (at 20 C and 50 bar) Measuring range (liquid): 70000 kg/h Max. possible full scale value: mmax(G) = mmax(F) (G) : 160 [kg/m3] = 70000 kg/h 60.3 kg/h : 160 kg/m3 = 26400 kg/h Recommended full scale values See information in the "Limiting flow" section Page 20 ff. Operable flow range Greater than 1000: 1. Flow rates above the preset full scale value do not overload the amplifier, i.e. the totalizer values are registered correctly. Status input (auxiliary input): U = 3 to 30 V DC, Ri = 5 k, galvanically isolated. Configurable for: totalizer reset, positive zero return, error message reset, zero point adjustment start, batching start/stop (optional). Status input (auxiliary input) with PROFIBUS DP and MODBUS RS485: U = 3 to 30 V DC, Ri = 3 k, galvanically isolated. Switching level: 3 to 30 V DC, polarity-independent. Configurable for: totalizer reset, positive zero return, error message reset, batching start/stop (optional), batch totalizer reset (optional).
Input signal
Endress+Hauser
Current input (only Promass 83) Active/passive selectable, galvanically isolated, resolution: 2 A Active: 4 to 20 mA, RL < 700 , Uout = 24 V DC, short-circuit proof Passive: 0/4 to 20 mA, Ri = 150 , Umax = 30 V DC
Output
Output signal Promass 80 Current output: Active/passive selectable, galvanically isolated, time constant selectable (0.05 to 100 s), full scale value selectable, temperature coefficient: typically 0.005% o.f.s./C, resolution: 0.5 A Active: 0/4 to 20 mA, RL < 700 (for HART: RL 250 ) Passive: 4 to 20 mA; supply voltage US 18 to 30 V DC; Ri 150 Pulse/frequency output: Passive, open collector, 30 V DC, 250 mA, galvanically isolated. Frequency output: full scale frequency 2 to 1000 Hz (fmax = 1250 Hz), on/off ratio 1:1, pulse width max. 2 s Pulse output: pulse value and pulse polarity selectable, pulse width configurable (0.5 to 2000 ms) PROFIBUS PA interface: PROFIBUS PA in accordance with EN 50170 Volume 2, IEC 61158-2 (MBP), galvanically isolated Profile Version 3.0 Current consumption: 11 mA Permitted supply voltage: 9 to 32 V Bus connection with integrated reverse polarity protection Error current FDE (Fault Disconnection Electronic) = 0 mA Data transmission rate: 31.25 kBit/s Signal encoding: Manchester II Function blocks: 4 x Analog Input, 1 x Totalizer Output data: Mass flow, Volume flow, Density, Temperature, Totalizer Input data: Positive zero return (ON/OFF), Zero point adjustment, Measuring mode, Totalizer control Bus address can be configured via miniature switches or via the local display (optional)
Promass 83 Current output: Active/passive selectable, galvanically isolated, time constant selectable (0.05 to 100 s), full scale value selectable, temperature coefficient: typically 0.005% o.f.s./C, resolution: 0.5 A Active: 0/4 to 20 mA, RL < 700 (for HART: RL 250 ) Passive: 4 to 20 mA; supply voltage US 18 to 30 V DC; Ri 150 Pulse/frequency output: Active/passive selectable, galvanically isolated Active: 24 V DC, 25 mA (max. 250 mA during 20 ms), RL > 100 Passive: open collector, 30 V DC, 250 mA Frequency output: full scale frequency 2 to 10000 Hz (fmax = 12500 Hz), on/off ratio 1:1, pulse width max. 2 s Pulse output: pulse value and pulse polarity selectable, pulse width configurable (0.05 to 2000 ms)
Endress+Hauser
PROFIBUS DP interface: PROFIBUS DP in accordance with EN 50170 Volume 2 Profile Version 3.0 Data transmission rate: 9.6 kBaud to 12 MBaud Automatic data transmission rate recognition Signal encoding: NRZ Code Function blocks: 6 x Analog Input, 3 x Totalizer Output data: Mass flow, Volume flow, Corrected volume flow, Density, Reference density, Temperature, Totalizers 1 to 3 Input data: Positive zero return (ON/OFF), Zero point adjustment, Measuring mode, Totalizer control Bus address can be configured via miniature switches or via the local display (optional) Available output combination Page 11 PROFIBUS PA interface: PROFIBUS PA in accordance with EN 50170 Volume 2, IEC 61158-2 (MBP), galvanically isolated Data transmission rate: 31.25 kBit/s Current consumption: 11 mA Permitted supply voltage: 9 to 32 V Bus connection with integrated reverse polarity protection Error current FDE (Fault Disconnection Electronic): 0 mA Signal encoding: Manchester II Function blocks: 6 x Analog Input, 3 x Totalizer Input data: Positive zero return (ON/OFF), Zero point adjustment, Measuring mode, Totalizer control Output data: Mass flow, Volume flow, Corrected volume flow, Density, Reference density, Temperature, Totalizers 1 to 3 Bus address can be configured via miniature switches or via the local display (optional) Available output combination Page 11 MODBUS interface: MODBUS device type: slave Address range: 1 to 247 Supported function codes: 03, 04, 06, 08, 16, 23 Broadcast: supported with the function codes 06, 16, 23 Physical interface: RS485 in accordance with EIA/TIA-485 standard Supported baud rate: 1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200 Baud Transmission mode: RTU or ASCII Response times: Direct data access = typically 25 to 50 ms Auto-scan buffer (data range) = typically 3 to 5 ms Possible output combinations Page 11 FOUNDATION Fieldbus interface: FOUNDATION Fieldbus H1, IEC 61158-2, galvanically isolated Data transmission rate: 31.25 kBit/s Current consumption: 12 mA Permitted supply voltage: 9 to 32 V Error current FDE (Fault Disconnection Electronic): 0 mA Bus connection with integrated reverse polarity protection Signal encoding: Manchester II ITK Version 4.01 Function blocks: 7 x Analog Input, 1 x Digital Output, 1 x PID Output data: Mass flow, Volume flow, Corrected volume flow, Density, Reference density, Temperature, Totalizers 1 to 3 Input data: Positive zero return (ON/OFF), Zero point adjustment, Measuring mode, Reset totalizer Link Master (LM) function is supported
Endress+Hauser
Signal on alarm
Current output: Failsafe mode selectable (e.g. in accordance with NAMUR Recommendation NE 43) Pulse/frequency output: Failsafe mode selectable Status output (Promass 80): "Nonconductive" in the event of a fault or if the power supply fails Relay output (Promass 83): "Dead" in the event of a fault or if the power supply fails
see "Output signal" Switch points for low flow cut off are selectable. All circuits for inputs, outputs, and power supply are galvanically isolated from each other. Status output (Promass 80): Open collector, max. 30 V DC / 250 mA, galvanically isolated. Configurable for: error messages, Empty Pipe Detection (EPD), flow direction, limit values. Relay output (Promass 83): Normally closed (NC or break) or normally open (NO or make) contacts available (factory setting: relay 1 = NO, relay 2 = NC), max. 30 V / 0.5 A AC; 60 V / 0.1 A DC, galvanically isolated.
Endress+Hauser
Power supply
Electrical connection Measuring unit
d g b a a b d/(g) (d) HART* PROFIBUS PA* FOUNDATION Fieldbus*
PA()/FF() PA(+)/FF(+) + f + + 27 26 25 24 23 22 21 20
A d g b
27 + 26 25 + 24 23 + 22 21 + 20
d e
d e
N (L-) 2 L1 (L+)1
c b
N (L-) 2 L1 (L+) 1
c b
PROFIBUS DP*
A (RxD/TxD-N) 27 B (RxD/TxD-P) 26 25 + 24 f 23 + 22 21 + 20
d e g c b
d e g c b
a0002441
N (L-) 2 L1 (L+) 1
N (L-) 2 L1 (L+) 1
Connecting the transmitter, cable cross-section: max. 2.5 mm2 A B C *) **) a b View A (field housing) View B (stainless steel field housing) View C (wall-mount housing) fixed communication board flexible communication board Connection compartment cover Cable for power supply: 85 to 260 V AC, 20 to 55 V AC, 16 to 62 V DC Terminal No. 1: L1 for AC, L+ for DC Terminal No. 2: N for AC, L for DC Ground terminal for protective ground Signal cable: see Terminal assignment Page 11 Fieldbus cable: Terminal No. 26: DP (A) / PA (+) / FF (+) / MODBUS RS485 (A) / (PA, FF: with reverse polarity protection) Terminal No. 27: DP (B) / PA () / FF () / MODBUS RS485 (B) / (PA, FF: with reverse polarity protection) Ground terminal for signal cable shield / fieldbus cable / RS485 line Service adapter for connecting service interface FXA193 or FXA291 (Fieldcheck, ToF Tool - Fieldtool Package) Signal cable: see Terminal assignment Page 11 Cable for external termination (only for PROFIBUS DP with permanent assignment communication board): Terminal No. 24: +5 V Terminal No. 25: DGND
c d
e f g g
10
Endress+Hauser
Promass 80
Terminal No. (inputs/outputs) Order version 80***-***********A 80***-***********D 80***-***********H 80***-***********S 80***-***********T 80***-***********8 20 (+) / 21 () Status input Status input 22 (+) / 23 () Status output Frequency output 24 (+) / 25 () Frequency output Frequency output Frequency output Ex i, passive Frequency output Ex i, passive Current output 2 26 (+) / 27 () Current output, HART Current output, HART PROFIBUS PA Current output Ex i Active, HART Current output Ex i Passive, HART Current output 1, HART
Promass 83 The inputs and outputs on the communication board can be either permanently assigned (fixed) or variable (flexible), depending on the version ordered (see table). Replacements for modules which are defective or which have to be replaced can be ordered as accessories.
Terminal No. (inputs/outputs) Order version 20 (+) / 21 () 22 (+) / 23 () 24 (+) / 25 () 26 (+) / 27 ()
Fixed communication boards (permanent assignment) 83***-***********A 83***-***********B 83***-***********F 83***-***********G 83***-***********H 83***-***********J 83***-***********K 83**************Q 83***-***********R Relay output Relay output Frequency output Frequency output +5V (ext. termination) Status input Current output 2 Ex i, active Frequency output Ex i, passive Frequency output Ex i, passive Current output 2 Ex i, passive Current output HART Current output HART PROFIBUS PA, Ex i FOUNDATION Fieldbus Ex i PROFIBUS PA PROFIBUS DP FOUNDATION Fieldbus MODBUS RS485 Current output 1 Ex i active, HART Current output Ex i Active, HART Current output Ex i Passive, HART Current output 1 Ex i passive, HART
Flexible communication boards 83***-***********C 83***-***********D 83***-***********E 83***-***********L Relay output 2 Status input Status input Status input Relay output 1 Relay output Relay output Relay output 2 Frequency output Frequency output Current output 2 Relay output 1 Current output HART Current output HART Current output 1 HART Current output HART
Endress+Hauser
11
Terminal No. (inputs/outputs) Order version 83***-***********M 83**************N 83**************P 83**************V 83***-***********W 83***-***********0 83***-***********2 83***-***********3 83***-***********4 83***-***********5 83***-***********6 83**************7 20 (+) / 21 () Status input Current output Current output Relay output 2 Relay output Status input Relay output Current input Current input Status input Status input Relay output 2 22 (+) / 23 () 24 (+) / 25 () 26 (+) / 27 () Current output HART MODBUS RS485 PROFIBUS DP PROFIBUS DP Current output 1 HART Current output 1 HART Current output 1 HART Current output 1 HART Current output HART Current output HART Current output HART MODBUS RS485
Frequency output 2 Frequency output 1 Frequency output Frequency output Relay output 1 Current output 3 Current output 3 Current output 2 Relay output Relay output Current input Current input Relay output 1 Status input Status input Status input Current output 2 Current output 2 Frequency output Current output 2 Frequency output Frequency output Current output 2 Status input
S1 S1 S2 S2 GND TM TM TT TT + + + + 4 5 6 7 8 9 10 11 12 41 42
d d d e
4 5 6 7 8 9 10 11 12 + + + + S1 S1 S2 S2 GND TM TM TT TT
41 42
a0003681
Connecting the remote version a b c d e Wall-mount housing: non-hazardous area and ATEX II3G / zone 2 see separate "Ex documentation" Wall-mount housing: ATEX II2G / Zone 1 /FM/CSA see separate "Ex documentation" Remote version, flanged version Cover for connection compartment or connection housing Connecting cable
Terminal No.: 4/5 = gray; 6/7 = green; 8 = yellow; 9/10 = pink; 11/12 = white; 41/42 = brown
Supply voltage
12
Endress+Hauser
Cable entries
Power-supply and signal cables (inputs/outputs): Cable entry M20 x 1.5 (8 to 12 mm) Thread for cable entries, 1/2" NPT, G 1/2" Connecting cable for remote version: Cable entry M20 x 1.5 (8 to 12 mm) Thread for cable entries, 1/2" NPT, G 1/2"
6 x 0.38 mm2 (PVC cable with common shield and individually shielded cores Conductor resistance: 50 /km Capacitance: core/shield: 420 pF/m Cable length: max. 20 m Permanent operating temperature: max. +105 C
Operation in zones of severe electrical interference: The measuring device complies with the general safety requirements in accordance with EN 61010, the EMC requirements of EN 61326/A1, and NAMUR recommendation NE 21/43. Power consumption AC: <15 VA (including sensor) DC: <15 W (including sensor) Switch-on current: Max. 13.5 A (< 50 ms) at 24 V DC Max. 3 A (< 5 ms) at 260 V AC Power supply failure Promass 80 Lasting min. 1 power cycle: EEPROM saves measuring system data if the power supply fails HistoROM/S-DAT: exchangeable data storage chip with sensor specific data (nominal diameter, serial number, calibration factor, zero point, etc.) Promass 83 Lasting min. 1 power cycle: EEPROM and T-DAT save the measuring system data if the power supply fails. HistoROM/S-DAT: exchangeable data storage chip with sensor specific data (nominal diameter, serial number, calibration factor, zero point, etc.) Potential equalization No measures necessary.
Performance characteristics
Reference operating conditions Error limits following ISO/DIS 11631: 20 to 30 C 2 to 4 bar Calibration systems as per national norms Zero point calibrated under operating conditions Field density calibrated (or special density calibration) The following values refer to the pulse/frequency output. Measured error at the current output is typically 5 A. o.r. = of reading Mass flow (liquid): Promass 80 0.175% [(zero point stability / measured value) 100]% o.r. Promass 83 0.125% [(zero point stability / measured value) 100]% o.r.
Endress+Hauser
13
Mass flow (gas) 0.50% [(zero point stability / measured value) 100]% o.r. Volume flow (liquid) 0.50% [(zero point stability / measured value) 100]% o.r. Zero point stability:
Maximum full scale value DN 8 15 15 FB 25 25 FB 40 40 FB 50 50 FB 80 FB = Full bore versions of Promass I [kg/h] 2 000 6500 18000 18000 45000 45000 70000 70000 180000 180000 Zero point stability [kg/h] 0.20 0.65 1.8 1.8 4.5 4.5 7.0 7.0 18.0 18.0
Sample calculation
[%] 1.0
0.5
Promass 83 Promass 80
0.2
0 0 1 2 4 6 8 10 12 14 16 18 t/h
a0004611
Calculation example (mass flow, liquid): Given: Promass 83 I / DN 25, flow measured value = 8000 kg/h Max. measured error: 0.125% [(zero point stability / measured value) x 100]% o.r. Max. measured error: 0.125% 1.8 kg/h : 8000 kg/h 100% = 0.15% Density (liquid) 1 g/cc = 1 kg/l Standard calibration: 0.02 g/cc
14
Endress+Hauser
Special density calibration (optional), calibration range: 0.8 to 1.8 g/cc, 5 to 80 C: 0.004 g/cc After field density calibration or under reference conditions: 0.0020 g/cc Temperature 0.5 C 0.005 T (T = fluid temperature in C) Repeatability Mass flow (liquid): 0.05% [1/2 (zero point stability / measured value) 100]% o.r. Mass flow (gas): 0.25% [1/2 (zero point stability / measured value) 100]% o.r. Volume flow (liquid): 0.20% [1/2 (zero point stability / measured value) 100]% o.r. o.r. = of reading Zero point stability: see "Max. measured error" Page 13 ff. Calculation example (mass flow, liquid): Given: Promass 83 I / DN 25, flow measured value = 8000 kg/h Repeatability: 0.05% [1/2 (zero point stability / measured value) x 100]% o.r. Repeatability: 0.05% 1/2 1.8 kg/h : 8000 kg/h 100% = 0.061% Density measurement (liquid) 1 g/cc = 1 kg/l 0.001 g/cc Temperature measurement 0.25 C 0.0025 T (T = fluid temperature in C) Influence of fluid temperature When there is a difference between the temperature for zero point adjustment and the process temperature, the typical measured error of the Promass sensor is 0.0002% of the full scale value / C. The table below shows the effect on accuracy of mass flow due to a difference between calibration pressure and process pressure.
Promass I DN 8 15 15 FB 25 25 FB 40 40 FB 50 50 FB 80 FB = Full bore versions of Promass I 3/8" 1/2" 1/2" FB 1" 1" FB 1 " 1 " FB 2" 2" FB 3" 0.006 0.006 0.003 0.003 % o.r./bar 0.006 0.004 0.006 0.006 No influence No influence 0.0004 0.0004 0.0002 0.0002 % o.r./psi 0.0004 0.0003 0.0004 0.0004
Endress+Hauser
15
a0003605
Mounting location
Notwithstanding the above, the installation proposal below permits installation in an open vertical pipeline. Pipe restrictions or the use of an orifice with a smaller cross-section than the nominal diameter prevent the sensor running empty while measurement is in progress.
3 4
5
a0003597
Installation in a down pipe (e.g. for batching applications) 1 = Supply tank, 2 = Sensor, 3 = Orifice plate, pipe restriction (see Table), 4 = Valve, 5 = Batching tank
DN Orifice plate, pipe restriction [mm] "FB" = Full bore versions of Promass I
8 6
15 10
15FB 15
25 14
25FB 24
40 22
40FB 35
50 28
50FB 54
80 50
16
Endress+Hauser
Orientation Make sure that the direction of the arrow on the nameplate of the sensor matches the direction of flow (direction of fluid flow through the pipe. Vertical (view V) Recommended orientation with upward direction of flow. When fluid is not flowing, entrained solids will sink down and gases will rise away from the measuring tube. The measuring tubes can be completely drained and protected against solids buildup. Horizontal The transmitter can be installed in any orientation in a horizontal pipe run.
Standard, compact
a0004572
m = To ensure that the maximum permitted ambient temperature for the transmitter (20 to +60 C, optionally 40 to +60 C) is not exceeded, for low-temperature fluids, we recommend the horizontal orientation with the transmitter head up (Fig. H1) or the vertical orientation (Fig. V).
Endress+Hauser
Standard, remote
17
Heating Some fluids require suitable measures to avoid heat transfer at the sensor. Heating can be electric, e.g. with heated elements, or by means of hot water or steam pipes made of copper.
"
Caution! Risk of electronics overheating! Consequently, make sure that the adapter between the sensor and transmitter and the connection housing of the remote version always remain free of insulating material. Note that a certain orientation might be required, depending on the fluid temperature. Page 17 If using an electric trace heating system whose heating is regulated via phase angle control or pulse packages, influence on the measured values cannot be ruled out due to magnetic fields (i.e. for values that are greater than the values approved by the EN standard (sine 30 A/m)). In such cases, the sensor must be magnetically shielded. The secondary containment can be shielded with tin plates or electric sheets without preferential direction (e.g. V330-35A) with the following properties: Relative magnetic permeability r 300 Plate thickness d 0.35 mm Information on permitted temperature ranges Page 20 Special heating jackets, which can be ordered separately from Endress+Hauser as an accessory, are available for the sensors. Zero point adjustment All measuring devices are calibrated to state-of-the-art technology. The zero point determined in this way is imprinted on the nameplate. Calibration takes place under reference conditions. Page 13 ff. Therefore, a zero point adjustment is generally not required! Experience shows that the zero point adjustment is advisable only in special cases: When the highest measuring accuracy is required and the flow rates are very low. Under extreme process or operating conditions (e.g. very high process temperatures or very high viscosity fluids). Please note the following before carrying out the adjustment: The adjustment can only be performed with fluids that have no gas or solid contents. Zero point adjustment is performed with the measuring tubes completely filled and at zero flow (v = 0 m/s). This can be achieved, for example, with shutoff valves upstream and/or downstream of the sensor or by using existing valves and gates. Normal operation valves 1 and 2 open Zero point adjustment with pump pressure valve 1 open / valve 2 closed Zero point adjustment without pump pressure valve 1 closed / valve 2 open
a0003601
18
Endress+Hauser
There are no installation requirements regarding inlet and outlet runs. Max. 20 meters (66 feet) (remote version) It is important to ensure that cavitation does not occur, because it would influence the oscillation of the measuring tube. No special measures need to be taken for fluids which have properties similar to water under normal conditions. In the case of liquids with a low boiling point (hydrocarbons, solvents, liquefied gases) or in suction lines, it is important to ensure that pressure does not drop below the vapor pressure and that the liquid does not start to boil. It is also important to ensure that the gases that occur naturally in many liquids do not outgas. Such effects can be prevented when system pressure is sufficiently high. Therefore, the following locations should be preferred for installation: Downstream from pumps (no danger of vacuum) At the lowest point in a vertical pipe
!
Storage temperature Degree of protection Shock resistance Vibration resistance Electromagnetic compatibility (EMC)
Endress+Hauser
19
Warning! In case a danger of measuring tube failure exists due to process characteristics, e.g. with corrosive process fluids, we recommend the use of sensors whose secondary containment is equipped with special pressure monitoring connections (ordering option). With the help of these connections, fluid collected in the secondary containment in the event of tube failure can be bled off. This is especially important in high pressure gas applications. These connections can also be used for gas purging (gas detection). Dimensions Page 22 ff. See information in the "Measuring range" section Page 6 Select nominal diameter by optimizing between required flow range and permissible pressure loss. See the "Measuring range" section for a list of maximum possible full scale values. The minimum recommended full scale value is approx. 1/20 of the max. full scale value. In most applications, 20 to 50% of the maximum full scale value can be considered ideal Select a lower full scale value for abrasive substances such as fluids with entrained solids (flow velocity <1 m/s). For gas measurement the following rules apply: Flow velocity in the measuring tubes should not be more than half the sonic velocity (0.5 Mach). The maximum mass flow depends on the density of the gas: formula Page 6
Limiting flow
Pressure loss
Pressure loss depends on the fluid properties and on the flow rate. The following formulae can be used to approximately calculate the pressure loss:
Reynolds number
Re =
4m pdnr
a0003381
Re 2300 *
Dp = K n
0.25
1.75
0.75
m K3 r
a0004631
Re < 2300 p = pressure loss [mbar] = kinematic viscosity [m2/s] m = mass flow [kg/s]
m2 Dp = K1 n m K3 r
a0004633
= fluid density [kg/m3] d = inside diameter of measuring tubes [m] K to K3 = constants (depending on nominal diameter)
* To compute the pressure loss for gases, always use the formula for Re 2300.
20
Endress+Hauser
11.38 103 17.07 103 17.07 103 25.60 103 25.60 103 35.62 103 35.62 103 54.8 103 54.8 103
2.3 106 4.1 105 4.1 105 7.8 104 7.8 104 1.3 104 1.3 104 2.3 103 2.3 103
1.3 107 3.3 106 3.3 106 8.5 105 8.5 105 2.0 105 2.0 105 5.5 104 5.5 104
Pressure loss data includes interface between measuring tube and piping FB = Full bore versions of Promass I
[mbar]
1000
DN 8
DN 15
DN 25 DN 15 *
DN 40 DN 25 *
DN 50 DN 40 * DN 80 DN 50 *
100
10
0.1 0.01
1 2
a0004608
0.1
10
100 [t/h]
Pressure loss diagram for water 1 2 Standard versions Full bore versions (*)
Endress+Hauser
21
Mechanical construction
Design / dimensions
Dimensions: Dimensions of field housing compact version, powder-coated die-cast aluminum Dimensions: stainless steel field housing Dimensions: Remote field housing (II2G / zone 1) Dimensions: Remote version Dimensions: Wall-mount housing (non hazardous area and II3G / zone 2) Flange connections EN (DIN), ASME B16.5, JIS Tri-Clamp connections DIN 11851 connections (sanitary connection) Connections DIN 11864-1 Form A (couplings) Flange connection DIN 11864-2 Form A (flat flange) ISO 2853 connections (couplings) SMS 1145 connections (sanitary connection) Promass purge connections / secondary containment monitoring Purge connections / secondary containment monitoring Page 37 Page 25 Page 24 Page 24 Page 22 Page 23 Page 26 Page 30 Page 32 Page 33 Page 34 Page 35 Page 36
a0002516
DN 8
1)
A 118.5 118.5 118.5 118.5 118.5 118.5 118.5 118.5 118.5 118.5
B 137.5 137.5 137.5 137.5 137.5 137.5 137.5 137.5 137.5 137.5
C 138 138 138 138 152 152 167 167 196 196
15 15
2)
25 25
2)
40 40
2)
50 50
2)
80
1) 2)
DN 8 with DN 15 flanges as standard "FB" = Full bore versions of Promass I All dimensions in [mm]
22
Endress+Hauser
Esc
D C B
F A H J E K
L
a0001150
A 215 K 53
B 250 L 95
C 90.5 M 53
D 159.5 N 102
E 135 O 81.5
F 90 P 11.5
G 45 Q 192
H >50 R 8xM5
M
J 81 All dimensions in [mm]
Endress+Hauser
23
Nicht-eigensichere Stromkreise durch IP40-Abdeckung geschtzt Non-intrinsically safe circuits Ip40 protected Boucles de courant sans scurit intrinsque protges par Ip40
Esc
E J K L M
G H
a0002128
A* 242 G 100
B 240 H 123
B* 217 J 100
Esc
B 153
C 171
24
Ne
C 206 K 133
D 186 L 188
E 167 M 355
a0002245
Endress+Hauser
A A*
B C
Esc
E F G
G L
+1.5 2.0
di
A0006964
A 227
A* 207
B 187
C 168
D 160
DN 8 1) 15 15 2) 25 25 2) 40 40 2) 50 50 2) 80
1) 2)
E 291 291 291 291 305 305 320 320 349 349
F 59 59 59 59 72 72 86 86 110 110
G 350 350 350 350 377 377 406 406 458.1 458.1
L
3) 3) 3) 3) 3) 3) 3) 3) 3) 3)
di
3) 3) 3) 3) 3) 3) 3) 3) 3) 3)
DN 8 with DN 15 flanges as standard "FB" = Full bore versions of Promass I 3) dependent on respective process connection For dimensions, see the following pages All dimensions in [mm]
Endress+Hauser
25
N
Esc
G
+1.5 2.0
di
U LK G
a0003313
Flange EN 1092-1 (DIN 2501) / PN 40: 1.4301/304, fluid wetted parts: titanium Surface roughness (flange): EN 1092-1 Form B1 (DIN 2526 Form C), Ra 6.3 to 12.5 m DN 8
1)
L 402 438 572 578 700 708 819 827 1210 1210
N 4 x 14 4 x 14 4 x 14 4 x 14 4 x 14 4 x 18 4 x 18 4 x 18 4 x 18 8 x 18
S 20 20 19 23 22 26 24 28 40 37
U 17.30 17.30 17.07 28.50 25.60 43.10 35.62 54.50 54.8 82.5
di 8.55 11.38 17.07 17.07 25.60 25.60 35.62 35.62 54.8 54.8
15 15
2)
25 25
2)
40 40
2)
50 50
2)
80
1)
All dimensions in [inch]; Further dimensions Page 22 ff. DN 8 with DN 15 flanges as standard 2) "FB" = Full bore versions of Promass I
Flange EN 1092-1 (DIN 2501) / PN 63: 1.4301/304, fluid wetted parts: titanium Surface roughness (flange): EN 1092-1 Form B2 (DIN 2526 Form E), Ra 1.6 to 3.2 m DN 50 50 1) 80
1)
N 4 x 22 4 x 22 8 x 22
S 34 45 41
All dimensions in [inch]; Further dimensions Page 22 ff. "FB" = Full bore versions of Promass I
26
Endress+Hauser
Flange EN 1092-1(DIN 2501) / PN 100: 1.4301/304, fluid wetted parts: titanium Surface roughness (flange): EN 1092-1 Form B2 (DIN 2526 Form E), Ra 1.6 to 3.2 m DN 8 1) 15 15 2) 25 25 2) 40 40 2) 50 50 2) 80
1)
G 105 105 105 140 140 170 170 195 195 230
L 402 438 578 578 706 708 825 832 1210 1236
N 4 x 14 4 x 14 4 x 14 4 x 18 4 x 18 4 x 22 4 x 22 4 x 26 4 x 26 8 x 26
S 25 25 26 29 31 32 33 36 48 58
U 17.30 17.30 17.07 28.50 25.60 42.50 35.62 53.90 54.8 80.9
di 8.55 11.38 17.07 17.07 25.60 25.60 35.62 35.62 54.8 54.8
All dimensions in [inch]; Further dimensions Page 22 ff. DN 8 with DN 15 flanges as standard 2) "FB" = Full bore versions of Promass I
Flange according to ASME B16.5 / Cl 150: 1.4301/304, fluid wetted parts: titanium Surface roughness (flange): Ra 3.2 to 6.3 m DN 8 1) 15 15 2) 25 25 2) 40 40 2) 50 50 2) 80
1)
G 88.9 88.9 88.9 108.0 108.0 127.0 127.0 152.4 152.4 190.5
L 402 438 572 578 700 708 819 827 1210 1210
N 4 x 15.7 4 x 15.7 4 x 15.7 4 x 15.7 4 x 15.7 4 x 15.7 4 x 15.7 4 x 19.1 4 x 19.1 4 x 19.1
S 20 20 19 23 22 26 24 28 40 37
LK 60.5 60.5 60.5 79.2 79.2 98.6 98.6 120.7 120.7 152.4
di 8.55 11.38 17.07 17.07 25.60 25.60 35.62 35.62 54.8 54.8
All dimensions in [inch]; Further dimensions Page 22 ff. DN 8 with DN 15 flanges as standard 2) "FB" = Full bore versions of Promass I
Endress+Hauser
27
Flange according to ASME B16.5 / Cl 300: 1.4301/304, fluid wetted parts: titanium Surface roughness (flange): Ra 3.2 to 6.3 m DN 8 1) 15 15 2) 25 25 2) 40 40 2) 50 50 2) 80
1)
G 95.3 95.3 95.3 124.0 124.0 155.4 155.4 165.1 165.1 209.5
L 402 438 572 578 700 708 819 827 1210 1210
N 4 x 15.7 4 x 15.7 4 x 15.7 4 x 19.1 4 x 19.1 4 x 22.4 4 x 22.4 8 x 19.1 8 x 19.1 8 x 22.3
S 20 20 19 23 22 26 24 28 43 42
LK 66.5 66.5 66.5 88.9 88.9 114.3 114.3 127.0 127 168.1
di 8.55 11.38 17.07 17.07 25.60 25.60 35.62 35.62 54.8 54.8
All dimensions in [inch]; Further dimensions Page 22 ff. DN 8 with DN 15 flanges as standard 2) "FB" = Full bore versions of Promass I
Flange according to ASME B16.5 / Cl 600: 1.4301/304, fluid wetted parts: titanium Surface roughness (flange): Ra 3.2 to 6.3 m DN 8 1) 15 15 2) 25 25 2) 40 40 2) 50 50 2) 80
1)
G 95.3 95.3 95.3 124.0 124.0 155.4 155.4 165.1 165.1 209.5
L 402 438 578 578 706 708 825 832 1210 1222
N 4 x 15.7 4 x 15.7 4 x 15.7 4 x 19.1 4 x 19.1 4 x 22.4 4 x 22.4 8 x 19.1 8 x 19.1 8 x 22.3
S 20 20 22 23 25 28 29 33 46 53
LK 66.5 66.5 66.5 88.9 88.9 114.3 114.3 127.0 127 168.1
U 13.80 13.80 17.07 24.40 25.60 38.10 35.62 49.30 54.8 73.7
di 8.55 11.38 17.07 17.07 25.60 25.60 35.62 35.62 54.8 54.8
All dimensions in [inch]; Further dimensions Page 22 ff. DN 8 with DN 15 flanges as standard 2) "FB" = Full bore versions of Promass I
Flange JIS B2238 / 10K: 1.4301/304, fluid wetted parts: titanium Surface roughness (flange): Ra 3.2 to 6.3 m DN 50 50 1) 80
1)
N 4 x 19 4 x 26 8 x 18
S 28 48 37
U 50 54.8 82.5
All dimensions in [inch]; Further dimensions Page 22 ff. "FB" = Full bore versions of Promass I
28
Endress+Hauser
Flange JIS B2238 / 20K: 1.4301/304, fluid wetted parts: titanium Surface roughness (flange): Ra 3.2 to 6.3 m DN 8 1) 15 15 2) 25 25 2) 40 40 2) 50 50 2) 80
1)
L 402 438 572 578 700 708 819 827 1210 1210
N 4 x 15 4 x 15 4 x 15 4 x 19 4 x 19 4 x 19 4 x 19 8 x 19 8 x 19 8 x 23
S 20 20 19 23 22 26 24 28 42 36
di 8.55 11.38 17.07 17.07 25.60 25.60 35.62 35.62 54.8 54.8
All dimensions in [inch]; Further dimensions Page 22 ff. DN 8 with DN 15 flanges as standard 2) "FB" = Full bore versions of Promass I
Flange JIS B2238 / 40K: 1.4301/304, fluid wetted parts: titanium Surface roughness (flange): Ra 3.2 to 6.3 m DN 8 1) 15 15 2) 25 25 2) 40 40 2) 50 50 2) 80
1)
G 115 115 115 130 130 160 160 165 165 210
L 402 438 578 578 706 708 825 827 1210 1210
N 4 x 19 4 x 19 4 x 19 4 x 19 4 x 19 4 x 23 4 x 23 8 x 19 8 x 19 8 x 23
S 25 25 26 27 29 30 31 32 43 46
di 8.55 11.38 17.07 17.07 25.60 25.60 35.62 35.62 54.8 54.8
All dimensions in [inch]; Further dimensions Page 22 ff. DN 8 with DN 15 flanges as standard 2) "FB" = Full bore versions of Promass I
Endress+Hauser
29
Flange JIS B2238 / 63K: 1.4301/304, fluid wetted parts: titanium Surface roughness (flange): Ra 3.2 to 6.3 m DN 8 1) 15 15 2) 25 25 2) 40 40 2) 50 50 2) 80
1)
G 120 120 120 140 140 175 175 185 185 230
L 402 438 578 578 706 708 825 832 1210 1226
N 4 x 19 4 x 19 4 x 19 4 x 23 4 x 23 4 x 25 4 x 25 8 x 23 8 x 23 8 x 25
S 28 28 29 30 32 36 37 40 47 55
di 8.55 11.38 17.07 17.07 25.60 25.60 35.62 35.62 54.8 54.8
All dimensions in [inch]; Further dimensions Page 22 ff. DN 8 with DN 15 flanges as standard 2) "FB" = Full bore versions of Promass I
Tri-Clamp connections
Esc
di
+1.5 2.0
U G
a0003314
All dimensions in [inch]; Further dimensions Page 22 ff. 3A version (Ra 0.8 m/150 grit) Optional: Ra 0.4 m/240 grit) 2) DN 15 "FB" = Full bore versions of Promass I
30
Endress+Hauser
G 25.0 25.0
L 426 462
U 9.5 9.5
di 8.55 11.38
All dimensions in [inch]; Further dimensions Page 22 ff. 3A version (Ra 0.8 m/150 grit) Optional: Ra 0.4 m/240 grit)
G 50.4 50.4
L 427 463
U 22.1 22.1
di 8.55 11.38
see 3/4" Tri-Clamp connection 1" 1" 1 " 1 " 2" 2 " 3" 50.4 50.4 50.4 50.4 63.9 77.4 90.9 603 730 731 849 850 1268 1268 22.1 22.1 34.8 34.8 47.5 60.3 72.9 17.07 25.60 25.60 35.62 35.62 54.8 54.8
All dimensions in [inch]; Further dimensions Page 22 ff. 3A version (Ra 0.8 m/150 grit) Optional: Ra 0.4 m/240 grit) 2) "FB" = Full bore versions of Promass I
Endress+Hauser
31
Esc
di
+1.5 2.0
U G
a0003322
G Rd 34 x 1/8" Rd 34 x 1/8" Rd 34 x 1/8" Rd 52 x 1/6" Rd 52 x 1/6" Rd 65 x 1/6" Rd 65 x 1/6" Rd 78 x 1/6" Rd 78 x 1/6" Rd 110 x 1/4"
L 427 463 602 603 736 731 855 856 1268 1268
U 16 16 16 26 26 38 38 50 50 81
di 8.55 11.38 17.07 17.07 25.60 25.60 35.62 35.62 54.8 54.8
25 25 2) 40 40 2) 50 50
2)
80
1)
All dimensions in [inch]; Further dimensions Page 22 ff. 3A version (Ra 0.8 m/150 grit) 2) "FB" = Full bore versions of Promass I
Sanitary connection DIN 11851 Rd 28 x 1/8" / 3A version 1): titanium DN 8 15 G Rd 28x 1/8" Rd 28x 1/8" L 426 462 U 10 10 di 8.55 11.38
All dimensions in [inch]; Further dimensions Page 22 ff. 1) 3A version (Ra 0.8 m/150 grit)
32
Endress+Hauser
Esc
di
+1.5 2.0
U G
a0003317
G Rd 28x 1/8" Rd 34 x 1/8" Rd 34 x 1/8" Rd 52 x 1/6" Rd 52 x 1/6" Rd 65 x 1/6" Rd 65 x 1/6" Rd 78 x 1/6" Rd 78 x 1/6" Rd 110 x 1/4"
L 428 463 602 603 734 731 855 856 1268 1268
U 10 16 16 26 26 38 38 50 50 81
di 8.55 11.38 17.07 17.07 25.60 25.60 35.62 35.62 54.8 54.8
15 15
3)
25 25
3)
40 40
3)
50 50
3)
80
1)
All dimensions in [inch]; Further dimensions Page 22 ff. 3A version (Ra 0.8 m/150 grit) Optional: Ra 0.4 m/240 grit) 2) DN 8 with DN 10 threaded adapter 3) "FB" = Full bore versions of Promass I
Endress+Hauser
33
N
Esc
di
+1.5 2.0
U LK G
a0003318
G 54 59 70 82 94 94 133
N 4 x 9 4 x 9 4 x 9 4 x 9 4 x 9 4 x 9 4 x 11
S 10 10 10 10 10 10 12
LK 37 42 53 65 77 77 97.7
U 10 16 26 38 50 50 81
15 25 40 50 50
3)
80
All dimensions in [inch]; Further dimensions Page 22 ff. 1) 3A version (Ra 0.8 mm/150 grit) Optional: Ra 0.4 mm/240 grit) 2) DN 8 with DN 10 threaded adapter 3) DN 50 "FB" = Full bore versions of Promass I
34
Endress+Hauser
Esc
di
+1.5 2.0
U G
a0003319
Coupling ISO 2853 / 3A version 1): titanium DN 8 2) 15 15 3) 25 3) 40 40 3) 50 50 3) 80 G 37.13 37.13 37.13 37.13 50.65 50.65 64.16 64.1 91.19 L 435 471 610 744 737 859 856 1268 1268 U 22.6 22.6 22.6 22.6 35.6 35.6 48.6 48.6 72.9 di 8.55 11.38 17.07 25.60 25.60 35.62 35.62 54.8 54.8
All dimensions in [inch]; Further dimensions Page 22 ff. 1) 3A version (Ra 0.8 m/150 grit) Optional: Ra 0.4 m/240 grit) 2) DN 8 with DN 15 threaded adapters as standard 3) "FB" = Full bore versions of Promass I
Endress+Hauser
35
Esc
di
+1.5 2.0
U G
a0003320
40 40
2)
50 50
2)
80
1)
All dimensions in [inch]; Further dimensions Page 22 ff. 3A version (Ra 0.8 m/150 grit) 2) "FB" = Full bore versions of Promass I
36
Endress+Hauser
"
Caution! The secondary containment is filled with dry nitrogen (N2). Do not open the purge connections unless the containment can be filled immediately with a dry inert gas. Use only low gauge pressure to purge. Maximum pressure: 5 bar.
G L L
a0003321
DN 8
1)
H 90.65 90.65 90.65 90.65 90.65 103.35 103.35 117.75 145.5 145.5
G 1/2" NPT 1/2" NPT 1/2" NPT 1/2" NPT 1/2" NPT 1/2" NPT 1/2" NPT 1/2" NPT 1/2" NPT 1/2" NPT
15 15
2)
25 25
2)
40 40
2)
50 50
2)
80
1)
All dimensions in [inch]; Further dimensions Page 22 ff. DN 8 with DN 15 flanges as standard 2) "FB" = Full bore versions of Promass I
Weight
Compact version: see table below Remote version Sensor: see table below Wall-mount housing: 5 kg
Promass I / DN Compact version Remote version "FB" = Full bore versions of Promass I 8 12 10 15 15 13 15FB 19 17 25 20 18 25FB 40 38 40 41 39 40FB 65 63 50 67 65 50FB 120 118 80 124 122
All values (weight) refer to devices with EN/DIN PN 40 flanges. Weight information in [kg].
Endress+Hauser
37
Materials
Transmitter housing: Compact housing: stainless steel 1.4301/304 Compact housing: powder coated die-cast aluminum Wall-mount housing: powder coated die-cast aluminum Remote field housing: powder-coated die-cast aluminum
Sensor housing / containment: Acid and alkali-resistant outer surface Stainless steel 1.4301/304 Connection housing, sensor (remote version): Stainless steel 1.4301/304 Process connections Flanges EN 1092-1 (DIN 2501) / ASME B16.5 / JIS B2238 stainless steel 1.4301/304 Flange DIN 11864-2 Form A (flat flange) Grade 2 titanium Sanitary connection DIN 11851 / SMS 1145 Grade 2 titanium Coupling ISO 2853 / DIN 11864-1 Grade 2 titanium Tri-Clamp (OD-Tubes) Grade 2 titanium Measuring tubes: Grade 9 titanium Grade 2 titanium (flange disk) Seals: Welded process connections without internal seals Material load curves Flange connection to EN 1092-1 (DIN 2501) Flange material: 1.4301/304 Fluid wetted parts: titanium
[bar]
100
PN 100
90 80 70 60 50 40
PN 40 PN 63
PN 16
38
Endress+Hauser
Flange connection according to ASME B16.5 Flange material: 1.4301/304 Fluid wetted parts: titanium
[bar] 90
80 70 60 50 40 30 20 10 0
Class 600
Class 300
Class 150
20
40
60
Flange connection to JIS B2238 Flange material: 1.4301/304 Fluid wetted parts: titanium
[bar]
70 60 50 40
40K 63K
30 20
20K
10 0
10K
20
40
60
[bar] 30
20 10 0 -60 -40 -20 0 20 40 60 80 100 120 140 160 180 200 [C]
a0003305-en
PN 16
Endress+Hauser
39
[bar]
50 40
DN 8...40
DN 50 / 40 FB
Flange connections to DIN 11864-2 Form A (flat flange) Connection material: titanium
[bar] 40
30 20 10 0
DN 8...40
DN 50 / 40 FB
20
40
60
[C]
a0003307-en
20 10 0 -60 -40 -20 0 20 40 60 80 100 120 140 160 180 200 [C]
a0003308-en
Tri-Clamp process connection The load limit is defined exclusively by the material properties of the outer clamp used. This clamp is not included in the scope of delivery. Process connections Promass I (welded process connections): Flanges EN 1092-1 (DIN 2501), ASME B16.5, JIS B2238 Sanitary connections: Tri-Clamp, couplings (DIN 11851, SMS 1145, ISO 2853, DIN 11864-1), flange to DIN 11864-2 Form A (flat flange)
40
Endress+Hauser
Human interface
Display elements Liquid-crystal display: backlit, two lines (Promass 80) or four lines (Promass 83) with 16 characters per line Selectable display of different measured values and status variables At ambient temperatures below 20 C the readability of the display may be impaired. Promass 80: Local operation with three keys (, +, E) Quick Setup menus for straightforward commissioning Promass 83: Local operation with three optical keys (S/O/F) Application-specific Quick Setup menus for straightforward commissioning Language groups
Note! The language group is changed using the "ToF Tool Fieldtool Package" operating program. Language groups available for operation in different countries: Western Europe and America (WEA): English, German, Spanish, Italian, French, Dutch and Portuguese Eastern Europe/Scandinavia (EES): English, Russian, Polish, Norwegian, Finnish, Swedish and Czech South and Eastern Asia (SEA): English, Japanese, Indonesian Only Promass 83: China (CN): English, Chinese
Remote operation
Promass 80: Remote operation via HART, PROFIBUS PA Promass 83: Remote operation via HART, PROFIBUS PA/DP, FOUNDATION fieldbus
C-Tick symbol
Ex approval
Sanitary compatibility
Endress+Hauser
41
The flow device has successfully passed all the test procedures carried out and is certified and registered by the Fieldbus FOUNDATION. The device thus meets all the requirements of the following specifications: Certified to FOUNDATION Fieldbus Specification The device meets all the specifications of the FOUNDATION Fieldbus H1. Interoperability Test Kit (ITK), revision status 4.0 (device certification number: on request) The device can also be operated with certified devices of other manufacturers Physical Layer Conformance Test of the Fieldbus FOUNDATION
The flow device has successfully passed all the test procedures carried out and is certified and registered by the PNO (PROFIBUS User Organization). The device thus meets all the requirements of the following specifications: Certified in accordance with PROFIBUS Profile Version 3.0 (device certification number: available on request) The device can also be operated with certified devices of other manufacturers (interoperability)
MODBUS certification
The measuring device meets all the requirements of the MODBUS/TCP conformity test and has the "MODBUS/TCP Conformance Test Policy, Version 2.0". The measuring device has successfully passed all the test procedures carried out and is certified by the "MODBUS/TCP Conformance Test Laboratory" of the University of Michigan. EN 60529 Degrees of protection by housing (IP code) EN 61010-1 Protection Measures for Electrical Equipment for Measurement, Control, Regulation and Laboratory Procedures. EN 61326/A1 (IEC 1326) "Emission in accordance with Class A requirements". Electromagnetic compatibility (EMC requirements) NAMUR NE 21 Electromagnetic compatibility (EMC) of industrial process and laboratory control equipment. NAMUR NE 43 Standardization of the signal level for the breakdown information of digital transmitters with analog output signal. NAMUR NE 53 Software of field devices and signal-processing devices with digital electronics
Measuring devices with a nominal diameter smaller than or equal to DN 25 correspond to Article 3(3) of the EC Directive 97/23/EC (Pressure Equipment Directive) and have been designed and manufactured according to good engineering practice. For larger nominal diameters, optional approvals according to Cat. II/III are available when required (depends on fluid and process pressure). SIL -2: accordance IEC 61508/IEC 61511-1 (FDIS) "420 mA HART" output according to the following order code:
Functional safety
42
Endress+Hauser
Promass 80
Promass80***-************A Promass80***-************D Promass80***-************S Promass80***-************T Promass80***-************8
Promass 83
Promass83***-************A Promass83***-************B Promass83***-************C Promass83***-************D Promass83***-************E Promass83***-************L Promass83***-************M Promass83***-************R Promass83***-************S Promass83***-************T Promass83***-************U Promass83***-************W Promass83***-************ Promass83***-************2 Promass83***-************3 Promass83***-************4 Promass83***-************5 Promass83***-************6
Ordering information
The Endress +Hauser service organization can provide detailed ordering information and information on the order codes on request.
Accessories
Various accessories, which can be ordered separately from Endress+Hauser, are available for the transmitter and the sensor.
Documentation
Flow measurement (FA005D/06/en) Technical Information Promass 80F, 80M, 83F, 83M (TI053D/06/en) Technical Information Promass 80E, 83E (TI061D/06/en) Technical Information Promass 80A, 83A (TI054D/06/en) Technical Information Promass 80H, 83H (TI074D/06/en) Technical Information Promass 80S, 83S (TI076D/06/en) Technical Information Promass 80P, 83P (TI078D/06/en) Operating Instructions Promass 80 (BA057D/06/en) Operating Instructions Promass 80 PROFIBUS PA (BA072D/06/en) Operating Instructions Promass 83 (BA059D/06/en) Operating Instructions Promass 83 FOUNDATION Fieldbus (BA065D/06/en) Operating Instructions Promass 83 PROFIBUS DP/PA(BA063D/06/en) Operating Instructions Promass 83 MODBUS (BA107D/06/en) Description of Device Functions Promass 80 (BA058D/06/en) Description of Device Functions Promass 80 PROFIBUS PA (BA073D/06/en) Description of Device Functions Promass 83 (BA060D/06/en) Description of Device Functions Promass 83 FOUNDATION Fieldbus (BA066D/06/en) Description of Device Functions Promass 83, PROFIBUS DP/PA (BA064D/06/en) Description of Device Functions Promass 83 MODBUS (BA108D/06/en) Supplementary documentation on Ex-ratings: ATEX, FM, CSA Functional safety manual Promass 80, 83 (SD077D/06/en)
Endress+Hauser
43
Registered trademarks
KALREZ and VITON Registered trademarks of E.I. Du Pont de Nemours & Co., Wilmington, USA TRI-CLAMP Registered trademark of Ladish & Co., Inc., Kenosha, USA SWAGELOK Registered trademark of Swagelok & Co., Solon, USA HART Registered trademark of HART Communication Foundation, Austin, USA PROFIBUS Registered trademark of the PROFIBUS User Organization, Karlsruhe, Germany FOUNDATION Fieldbus Registered trademark of the Fieldbus FOUNDATION, Austin, USA MODBUS Registered trademark of the MODBUS Organization HistoROM, S-DAT, T-DAT, F-CHIP, ToF Tool - Fieldtool Package, FieldCare, Fieldcheck, Applicator Registered or registration-pending trademarks of Endress+Hauser Flowtec AG, Reinach, CH