Escolar Documentos
Profissional Documentos
Cultura Documentos
99MBB203A1
SERIES No. 211
RA-120
Roundness/Cylindricity
Measuring System
User’s Manual
Read this User’s Manual thoroughly
before operating the instrument. After reading,
retain it close at hand for future reference.
CONVENTIONS USED IN THIS MANUAL
Safety Precautions
To ensure that instruments are operated correctly and safely, Mitutoyo manuals use
various safety symbols (Signal Words and Safety Alert Symbols) to identify and warn
against hazards and potential accidents.
Indicates an imminently hazardous situation which, if not avoided, will result in serious
injury or death.
DANGER
Indicates a potentially hazardous situation which, if not avoided, could result in serious
injury or death.
WARNING
Indicates a potentially hazardous situation which, if not avoided, may result in minor or
moderate injury or property damage.
CAUTION
Alerts the user to a specific hazardous situation. The given example means “Caution,
risk of electric shock”.
Prohibits a specific action. The given example means “Do not disassemble”.
No. 99MBB203A i
CONVENTIONS USED IN THIS MANUAL
Types of Notes
The following types of notes are used in this manual to help the operator obtain reliable
measurement data through correct instrument operation.
IMPORTANT • An important note provides information essential to the completion of a task. You cannot
disregard this note to complete the task.
• An important note is a type of precaution, which if neglected could result in a loss of
data, decreased accuracy or instrument malfunction/failure.
NOTE A note emphasizes or supplements important points of the main text. It also supplies
information about specific situations (e.g., memory limitations, equipment
configurations, or details that apply to specific versions of a program).
TIP A tip is a type of note that helps the user apply the techniques and procedures described
in the text to his or her specific needs.
It also provides reference information associated with the topic being discussed.
• Mitutoyo assumes no liability to any party for any loss or damage, direct or
indirect, caused by use of this instrument not conforming to this manual. However,
except as expressly specified in the Mitutoyo Software License Agreement, in no
event shall Mitutoyo be liable to any party for any loss of damage caused by use
of this instrument conforming or not conforming to this manual.
ii No. 99MBB203A
Safety Precautions (Essential Reading)
Observe the precautions/instructions given on the following pages in order to obtain the
highest performance from the machine and to ensure safety.
• Do not modify the machine, as this may cause fire or electric shock
• Do not rework, bend by force, twist and yank the power code, as this may cause fire or
WARNING electric shock.
• Do not remove the front panel, the rear panel, the side panel and the covers of the
machine, as this may cause electric shock.
• Use under the specified power supply voltage. Do not use under the unspecified power
supply voltage, as this may cause fire or electric shock.
• Do not install the machine on an unstable table, as this may cause injury due to
dropping or falling.
• Do not place a case with water and a vase, etc., around the machine. When the water
get inside the machine, it may cause fire or electric shock.
In case the water, etc., get inside the machine, first unplug the power cord from the
outlet then contact your dealer or the nearest Mitutoyo sales office.
• When performing the maintenance, be sure to unplug the power cord for safety.
If the maintenance is performed while connecting the power cord, it may cause electric
CAUTION shock.
• Do not pull the power cord when unplugging it, as this may damage the cord and cause
fire or electric shock.
• Do not insert or pull out the plug with wet hands, as this may cause fire or electric shock.
• Do not bring the power cord close to the heated equipment, as this may melt the
sheathing of the cord and cause fire or electric shock.
• The cable of the detector unit should be arranged so that it will not be caught by the
workpiece fixture or operation knobs when the turntable rotates.
Warranty
In the event that the Mitutoyo product, except software product, should prove defective in
workmanship or material, within one year from the date of original purchase for use, it will
be repaired or replaced, at out option, free of charge upon its prepaid return to us.
If the product fails or is damaged for any of the following reasons, it will be subject to a
repair charge, even if it is still under warranty.
This warranty is effective only where the instrument is properly installed and operated in
conformance with the instructions in this manual.
iv No. 99MBB203A
Export Control Compliance
This Product falls into the Catch-All-Controlled Goods or Program under the Category 16
of the Separate Table 1 of the Export Trade Control Order or the Category 16 of the
Separate Table of the Foreign Exchange Control Order, based on the Foreign Exchange
and Foreign Trade Law of Japan.
Further, this User’s Manual also falls into the Catch-All-Controlled Technology for use of
the Catch-All-Controlled Goods or Program, under the Category 16 of the Separate Table
of the Foreign Exchange Control Order.
If you intend re-exporting or re-providing the product or technology to any party other than
yourself, please consult with Mitutoyo prior to such re-exporting or re-providing.
No. 99MBB203A v
Contents
vi No. 99MBB203A
4.1 Calibrating the Detector................................................................................................................ 4-1
4.1.1 Performing static calibration ................................................................................................ 4-1
4.1.2 Performing dynamic calibration........................................................................................... 4-4
4.2 Modifying the Calibration Screen Conditions ............................................................................ 4-6
4.2.1 Switching between Static and Dynamic Calibrations ........................................................ 4-6
4.2.2 Modifying the static calibration conditions ........................................................................ 4-7
4.2.3 Modifying the dynamic calibration conditions ................................................................... 4-9
5 CENTERING/LEVELING MEASUREMENT.......................................................................................... 5-1
5.1 Centering/Leveling ........................................................................................................................ 5-1
5.1.1 Setting a workpiece ............................................................................................................... 5-2
5.1.2 Measuring the centering/leveling amounts......................................................................... 5-5
5.1.3 Adjusting the centering/leveling amounts ........................................................................ 5-16
5.2 Modifying the Centering Screen Conditions ............................................................................ 5-19
5.2.1 Modifying the measuring conditions for centering/leveling ........................................... 5-19
5.2.2 Modifying the centering/leveling adjustment conditions ................................................ 5-24
6 MEASUREMENT/ ANALYSIS ............................................................................................................... 6-1
6.1 Measuring Procedure of Each Analysis Element....................................................................... 6-2
6.1.1 Roundness (single cross section) ....................................................................................... 6-2
6.1.2 Coaxiality [3 cross sections] ................................................................................................ 6-5
6.1.3 Concentricity (dual cross section)....................................................................................... 6-8
6.1.4 Circular runout (radial direction) ........................................................................................6-11
6.1.5 Circular runout (axial direction) ......................................................................................... 6-14
6.1.6 Squareness (axis reference)............................................................................................... 6-18
6.1.7 Squareness (plane reference) ............................................................................................ 6-22
6.1.8 Thickness deviation ............................................................................................................ 6-25
6.1.9 Flatness (single cross section) .......................................................................................... 6-28
6.1.10 Parallelism............................................................................................................................ 6-31
6.2 Element Measurement Procedure ............................................................................................. 6-34
6.3 Listing or Deleting Measured Elements (Cross Sections) ...................................................... 6-38
6.3.1 Listing the measured elements (cross sections) ............................................................. 6-38
6.3.2 Deleting measured elements (cross sections) ................................................................. 6-38
6.4 Switching to Other Analysis Result Displays........................................................................... 6-39
6.5 Modifying the Measurement Screen Conditions...................................................................... 6-40
6.5.1 Procedures for changing items.......................................................................................... 6-41
6.6 Modifying the Result Screen Conditions .................................................................................. 6-47
6.6.1 Procedures for changing items.......................................................................................... 6-48
7 CONDITIONS REGISTRATION............................................................................................................. 7-1
7.1 Conditions Registration Function ............................................................................................... 7-1
7.2 Condition Registration Procedures............................................................................................. 7-2
7.2.1 Booting the registration screen ........................................................................................... 7-2
7.2.2 Calibration ratio setup........................................................................................................... 7-3
7.2.3 Centering condition setup .................................................................................................... 7-4
Service Network
No. 99MBB203A ix
MEMO
x No. 99MBB203A
1
1 INTRODUCTION
• The RA-120 employs a high precision air-bearing type turntable, offering enhanced
measurement accuracy, reliability, and durability. There is no accuracy deterioration due
to abrasion.
• Workpiece centering and leveling can be performed simply and speedily with the HELP
function.
• The built-in thermal line printer outputs measured data, setup parameters, and graphs
(circular cross section and development).
• The integrated construction of the measuring unit and display/control unit minimizes the
installation space required.
Column unit
Detector holder
SD scale
Detector
X-axis arm
Alignment table
Control panel
• To use the measuring unit safely, read "Safety Precautions" in this manual prior to use
and observe all the precautions.
CAUTION
The RA-120 employs Digimatic heads as centering knobs and leveling knobs for the D.A.T.
(Digital Adjustable Table) function.
Turntable diameterφ158
3-M5 Depth 8 (P.C.Dφ143) (circumference C3)
Center M5 Depth 10
CY centering Knob
Center point of inclination
LX leveling Knob
CX centering Knob
LY leveling Knob
+displacement
adjusting screw
• Two screws are mounted on the detector side to limit the detecting range and are used
to adjust the stylus displacement. The screws are locked in place with paint before
CAUTION shipping. They are adjusted for a detecting range of ±1000µm. Do not adjust these
screws unless specifically required.
Detector holder
Detector positioning mark
Holder clamp knob
Detector (Loosen the knob and switch the
detector holder orientation from
vertical to horizontal and vice
versa.)
X axis arm
• There are two detector base lines. Align the base line that corresponds to the measuring
direction with the detector positioning mark.
Measurements are possible in the setups illustrated below by changing the direction of the
detector holder and detector.
OD measurement ID measurement
Setting to OUT
OD measurement
• Do not rotate the X and Z-drive wheels at both ends of the travel limited by stoppers for
fear of damage to the mechanism.
CAUTION
Analysis items Keys • These keys are for setting analysis items.
NOTE • If there is no measured data in memory, the Result screen will not be selected.
• Do not manually pull the paper out of the printer and do not feed the paper in the
reverse direction. These actions may damage the printer.
CAUTION • The last meter of printer paper has red markings on both sides. Replace the print paper
with a new roll according to this marking guide. Do not print without paper. This may
cause the printer to fail.
Jog dial
Standard feed
(inside dial)
Measure key
ROTATE/STOP Key
Press this key to start/stop rotation of the
alignment table (Rotating direction/Speed:
Clockwise/6rpm).
-1000μm +1000μm
• The range of resetting is within the range indicated by parallel thick lines on the level
meter.
CAUTION • Adjust the detector position by manipulating the X/Z drive wheels to bring the meter
indicator into the range of resetting.
• Air inlet: Connects the air tube of the air filter unit.
• After the power switch has been turned on, information will not be displayed on the LCD
for approximately 10 seconds. Do not turn off the power switch within this time because
CAUTION the system executes initialization and a power-on check in this period.
Filter regurator
Connect main unit
Air filter
For information about how to replace elements in the air filter unit, refer to section 10.2.1,
"Replacing the elements in the air filter unit".
• Compressed air must always pass through the air filter unit.
• Ensure that the pressure of air to be supplied to the main unit is 0.39MPa (4.0kgf/cm2).
CAUTION • If compressed air is supplied directly to the main unit, the air bearing will be
contaminated with dust and oil mist, and unit failure will result.
• Select a stable installation site with minimum vibration, dust, splashing oil, and
temperature fluctuation.
• Maintain the environmental temperature at 20°C ±10% during operation and avoid a
sudden change in temperature.
• Avoid using the system at a site with high humidity. (Humidity should be less than
85%RH, with no condensation.)
• Use the power supply in a range between 90% and 110% of the rated voltage.
• Measurement made at a high sensitivity may be subject to even slight vibrations and
gentle air flows. Great care should be exercised when performing such high-sensitivity
measurements.
• Store this instrument in a place with an ambient temperature between -10°C and 60°C.
Detector holder
Stylus
(Insert into the detector unit)
X axis arm
Detector positioning mark
• There are two detector base lines. Align the base line that corresponds to the measuring
direction with the detector positioning mark.
φ6 tube
• If abnormalities such as heat radiation, smoke, and a burning smell are detected after
turning on the power switch, immediately turn off the power switch, then unplug the
CAUTION power cord from the outlet. If this instrument continues to be used under these
conditions, a fire or an electric shock may result. Contact your dealer or the nearest
Mitutoyo service network. To locate the nearest Mitutoyo sales office, refer to the service
network list on the back cover of this manual.
IMPORTANT • Be sure to perform the following operations while supplying the rated air pressure
0.39MPa (4.0kgf/cm2) to the main unit.
If operation is not performed at the rated air pressure, the air bearing may be damaged.
Unscrew the three screws to remove the fixture mounted across the upper face of the
alignment table.
NOTE • Do not discard the fixture. It is required for transporting the main unit.
Level
support leg A
Turn the two auxiliary legs (marked with B) on the bottom of the main unit to pull them in.
Place a level on the table in the direction shown below.
Adjust the height of three support legs (marked with A) on the bottom so that the main unit
is leveled with the level on the table oriented in two directions (90° to each other).
After leveling has been completed, gently bring the two auxiliary legs (marked with B) into
contact with the installation surface.
TIP • The battery lasts approximately 3.5 years. If a new battery is required, “B” will be
displayed on the LCD. If this occurs, immediately replace the battery.
At the factory, the alignment table top has been designated as the origin of the SD scale.
If the [ORIGIN] or [ZERO/ABS] button has been accidentally pressed, restore the original
condition according to the following procedure.
TIP • If the SD scale surface is stained, "ErrC" will be displayed on the LCD or the displayed
value will fluctuate. It is advisable to clean the scale surface and apply a small amount
of low-viscosity oil to repel water.
• If the Up/Down Feed knob is turned too quickly, the least significant digit on the LCD
may display ”E”, but measurement will not be affected. Continue the measurement
operation. If the SD scale surface is stained, “E” may be displayed on the LCD in the
position of the least significant digit. Clean the scale surface.
▪ The SD scale is turned on or off independently of the power switch on the RA-120 main
unit. Be sure to check the SD scale power before and after measurement.
Adjustment of the detector position is divided into two parts: adjustment in vertical detector
orientation and adjustment in horizontal detector orientation.
Detector holder
2.3.1 Adjustment in vertical detector orientation ルダ
(1) Perform centering of a work piece with
the detector unit set in vertical orientation
as shown in the figure at the right.
Detector stylus tip
(Refer to section 5 "CENTERING/LEVELING
MEASUREMENT".)
(3) While monitoring the stylus displacement display on the LCD screen, adjust the set
screw so that stylus displacement is maximized. If the set screw is turned clockwise from
the right side in the figure (or turned counterclockwise from the left side in the figure), the
stylus moves toward the front.
Fulcrum
Setscrew Setscrew
Counterclockwise Clockwise
(4) Increase the meter sensitivity, then repeat the procedure in step (1) to step (3).
NOTE • Particularly if measuring a small radius work piece at high accuracy, perform
measurement with the detector holder set in the vertical orientation.
(2) Align the stylus tip with the central position of the X axis on a work piece.
Adaptor
b) If the stylus is used with a tilt, lower the detector by the difference in stylus tip position,
as shown in the following figure.
(3) Without changing the vertical positional relationship of the detector, tilt the detector unit
in the horizontal orientation.
Confirm that the stylus tip does not deviate from the X-axis central position on the work
piece.
Power Switch
• Press the power switch on the [○] side to shut down the RA-120.
• After the power switch has been turned on, information is not displayed on the LCD for
approximately 10 seconds. Do not turn off the power switch within this time because the
CAUTION system executes initialization and a power-on check during this period.
• The SD scale is turned on or off independently of the power switch on the RA-120 main
unit. Be sure to check the SD scale power before and after measurement.
(2) Turn the power switch on the rear panel of the main unit to the [I] side to start the
RA-120.
IMPORTANT • Be sure to perform the following work with the main unit being supplied with the air of
specified pressure 0.39MPa (4.0kgf/cm2).
• The air bearing might be damaged when working with the air of the specified pressure
not supplied.
• After the power switch has been turned on, information is not displayed on the LCD for
approximately 10 seconds. Do not turn off the power switch within this time because the
CAUTION system executes initialization and a power-on check during this period.
• The SD scale (option) is turned on or off independently of the power switch on the
RA-120 main unit. Be sure to check the SD scale power before and after measurement.
NOTE • The level meter reading may electrically drift for about 10 minutes after the power-up of
the instrument. To ensure accurate measurements, it is recommended that the
instrument be turned on 10 minutes before starting measurement.
This section describes the procedure for setting up and roughly centering a workpiece on
the alignment table.
For the procedure for exactly centering a workpiece, refer to “5. Centering/Leveling
Measurement”.
Approximate center of
the table
alignment table
workpiece
(2) Set the indicating range of the level meter to ±1000μm by pressing the range key.
(3) Turn the X-axis drive wheel and move the detector stylus away from the workpiece.
(5) If the workpiece has been set up off-centered on the alignment table, the distance
between the stylus tip and workpiece varies during table rotation.
Stop the rotation of the table at the position where the distance between the stylus tip
and the workpiece is at the maximum.
a.Maximum distance
b.Minimum distance
c. Difference between the maximum and
minimum distances
(7) Turn the X-axis drive wheel and bring the stylus tip into contact with the workpiece until
the level meter reads an approximate zero (the center of the meter).
level meter
Approximately at the center
-1,000μm +1,000μm
-1,000μm +1,000μm
(9) Stop the table rotation when the level meter has deflected to the maximum on the plus
side.
-1,000μm +1,000μm
-200μm +600μm
-1,000μm +1,000μm
TIP • Tapping the workpiece slightly by such as the grip of a screwdriver will make it easy to
shift the workpiece.
Screwdriver
Workpiece
-1,000μm +1,000μm
(12) Repeat steps (8) through (10) if centering with a higher degree of accuracy is required.
After the level meter deflection has been reduced sufficiently at the displayed
magnification (±1000μm in the above case), increase the magnification for further
adjustment. This step by step adjustment facilitates centering.
The procedure shown above is a rough-and-ready method for the manual centering of a
workpiece.
To make centering with higher accuracy, use the centering/leveling function. (Refer to “5.
Centering/Leveling Measurement” for details.)
Registered conditions can be called all together at once by pressing the key to call
conditions. This saves the procedure of setting conditions otherwise needed for each time.
Four keys 1 through 4 are available allowing four sets of conditions to be registered
according to the workpieces to be measured.
Conditions that can be registered on the keys to call conditions are the following three
kinds.
Any combination of the following condition items can be registered:
*Refer to “7. Condition Registration” for the procedures of registration of conditions on the
keys to call conditions.
NOTE To perform measurement with higher accuracy in the range of ±100μm or lower, take the
following into consideration:
• Slight fluctuation takes place in detector displacement immediately after the detector
has been changed in orientation from the vertical to horizontal or vise versa or shifted
along the X or Z-axis. So, allow dozens of seconds for the detector to stabilize before
commencing measuring. When shifting the detector in the Z-axis direction, reciprocate
the detector a few times in the vertical direction for running-in around a target position
and finally stop the detector in position while shifting it upward. This reduces fluctuations
that may take place immediately after the detector shifting
• The effect of table speed fluctuation upon the start of table rotation can be reduced by
commencing measurement well after the table has started rotation.
(1) After setting up the workpiece to be measured, press the key with which the conditions
to be called have been registered.
(2) The calibration ratio registered with the key will be called. (If no calibration ratio has
been registered, the current one will apply.)
(3) The registered centering/leveling conditions will be called with automatic switching to
the display of the centering screen.
(4) Press the [START] key and perform the centering measurement .
For details of centering procedures, refer to “5. Centering/Leveling Measurement”.
(5) After the centering/leveling is over, press the [MEASURE] key to switch to the
measurement screen.
(6) Following the registered measurement conditions called on the screen, press the
[START] key to perform measurement.
For details of measurement/analysis procedures, refer to “6 Measurement/Analysis”.
(7) Press the [RESULT] key to switch to the result screen for the display of the
measurement results.
Refer to “6. Measurement/Analysis” for details of the result screen.
TIP • Automatic switching to the result screen after measurement is enabled if [2: Automatic
calculation [○] has been selected in the Environment setup/ Automatic execution setup.
Refer to “9.1 Automatic Execution Setup”.
(1) After setting up a workpiece, press the key with which the conditions to be called have
been registered.
(2) The calibration ratio registered will be called. (If no calibration ratio has been registered,
the current one will apply.)
(3) The registered measurement conditions will be called with automatic switching to the
measurement screen.
(5) Press the [RESULT] key to switch to the result screen for the display of the
measurement results.
Refer to “6. Measurement/Analysis” for details of result screen.
TIP • Automatic switching to the result screen after measurement is enabled if [2: Automatic
calculation [○]] has been selected in the Environment setup/Automatic execution setup.
Refer to “9.1 Automatic execution setup”.
(6) For measurement of the next workpiece, set up the next workpiece and return to step
(4).
Return to step (4) for re-measurement of the same workpiece.
Pressing the key to call conditions will call a desired calibration ratio.
If frequently used calibration ratio is registered with a key to call conditions, the key can
be used to call the calibration ratio with “one-key” operation.
Pressing the key to call conditions will call a desired centering/leveling condition with
automatic switching to the centering screen.
Switching to the centering screen, otherwise made from the menu screen, can be made
with “one-key" operation. Thus, the key registered with the centering/leveling conditions
can be used as a key for switching to the centering screen.
3.4.1 Printing
Analysis results will be printed by pressing the [PRINT] key under the result screen.
Result printing is possible with the built-in printer or with an external printer (option) using
the parallel output from the RA-120. Refer to “9.3 Instrument (RA-120 Hardware Setup)”
for procedures of switching the destination printers.
TIP • If the [4: AUTO PRINT] has been enabled in the Environment setup/Automatic
execution setup, the analysis results will be printed automatically upon switching to the
result screen after measurement is over. (Refer to “9.1 Automatic Execution Setup.”)
• Print items can be set up with [4: Print items] in the Environment setup/Parameter setup.
(Refer to “9.2 Parameter Setup”.)
TIP • Refer to “10.2.2. Replacing the printer paper” for how to replace the printer paper.
Refer to “8.1.2 Registering the result files” or “8.4 Outputting Data” for the procedures of
measurement results output.
TIP • Automatic execution setup has been enabled, automatic output is available for the
analysis results upon switching to the result screen after measurement is over. (Refer to
“9.1 Automatic Execution Setup.)
• ON/OFF of the SD scale (option) does not interrelate with the power switch of the
RA-120 main unit. To save the battery of the SD scale, ensure that SD scale has been
CAUTION switched OFF before and after measurement.
• Do not turn off the power switch of the RA-120 during the operation of the turntable. This
may cause instrument failure.
MEMO
TIP If [1: Startup calibration [O]] has been selected in Preferences Set/Auto-Execute setup,
the calibration screen automatically appears after the power is turned on. (Refer to
section9.1 "Setting auto-execution".)
(5) Insert the calibration film between the stylus and the reference plane, then press the
[ENT] key.
(7) Press the [ENT] key to enter the calibration film thickness (C value).
(9) This completes the calibration procedure and returns the LCD screen to step (1).
(1) Position the Dynamic calibration master at the center of the alignment table.
(2) Bring the stylus into contact with the measuring face on the Dynamic calibration master.
Adjust the alignment table so that the displayed stylus displacement is within the meter
sensitivity range while turning the table. For detailed information about the centering
procedure, refer to section 5 "CENTERING/LEVELING MEASUREMENT".
(4) The most recent calibration conditions are displayed on the screen. Confirm the setup
conditions of the following two items.
(6) Check the Dynamic calibration master roundness (C value) displayed on the screen.
If the displayed value is same as the
roundness of the Dynamic calibration
master, press the [ENT] key.
If the displayed value is different from the
roundness of the Dynamic calibration
master, adjust the displayed value to the
correct value by turning the jog dial, then
press the [ENT] key.
(8) This completes the calibration procedure and returns the LCD screen to step (1).
TIP This modification can only be made if the Calibration screen is displayed. The conditions
that can be set up differ depending on the calibration method.
Static calibration
Dynamic calibration
▪ Measuring orientation
▪ Calibration ratio
Measuring orientation
Calibration ratio
TIP The cursor moves in the following directions when the jog dial is turned.
Measuring orientation ← → Calibration ratio ← → Calibration method
→ : Clockwise rotation of the jog dial
← : Counterclockwise rotation of the jog dial
(2) Check the measuring orientation of the detector stylus. (Refer to section 2.2.2
"Mounting the detector".)
NOTE Be sure that the measuring orientation corresponds to the actual stylus orientation. If the
set measuring orientation differs from the stylus orientation, a correct stylus displacement
cannot be detected.
TIP A calibration ratio is a gain adjustment ratio by which a stylus displacement is multiplied.
Four kinds of calibration ratios have been registered. Those ratios can be used if, for
example, the current stylus is replaced with a different stylus.
Cutoff value
Calibration ratio
Calculation method
TIP The cursor moves in the following directions when the jog dial is turned.
Measuring orientation ← → Filter type ← → Cutoff value ← → Calculation method ← →
Calibration ratio ← → Calibration method
→ : Clockwise rotation of the jog dial
← : Counterclockwise rotation of the jog dial
TIP If the filter type has been set to [NON], no cutoff value can be selected.
5.1 Centering/Leveling
To perform measurement with high accuracy, it is necessary to accurately carry out
“centering” and “leveling” and correctly set up a workpiece to be measured in place. This is
an important work especially when performing measurement with a higher magnification.
Centering Leveling
▪ Centering
▪ Leveling
"Leveling" is an alignment procedure for adjusting the tilt of the alignment table so that a
measured cross-section contour may approach a circle.
If the axis of a cylindrical workpiece is tilted with respect to the rotational axis of the
alignment table, the measured profile will form an ellipse. This is critically detrimental to
obtaining a correct measurement.
Also, two methods of measurement data input are available: full-circumference input and
multi-point input.
This manual explains measurement based on the full-circumference input.
NOTE Before setting a workpiece on the alignment table be sure to confirm that the rated air
pressure has been supplied to the system.
(1) The adjustable range of the centering (CX and CY) knobs is ±3mm (±.12"). Confirm
that the rotary position of each knob is not off-center (it should be centered at the
adjustable range).
(2) Use the concentric circles scribed on the table to position a workpiece on the alignment
table so that the cross section to be measured comes as close as possible to the table
center.
Concentric circle
(3) Bring the stylus close to a measurement cross-section contour and then turn the
alignment table to check the deflection of the cross-section contour.
If the deflection exceeds 1mm(.04"), adjust the alignment table position according to the
following procedure.
1mm or less
Centering CX knob
(Gap check)
(Adjustment)
(Turn 180
degrees)
(5) Place the centering CY knob in the same position the CX knob was placed, as in the
above left figure, and then adjust the alignment table position using the same procedure
in step (4).
(6) This completes the workpiece setting procedure. Advance to section 5.1.2 "Measuring
the centering/leveling amounts".
(4) While turning the alignment table, adjust the detector position with the arm feed knob
so that the stylus displacement display is within the meter sensitivity range. Turn the jog
dial for fine adjustment and perform zero shifting.
(4) While turning the alignment table, adjust the detector position with the up/down feed
knob so that the stylus displacement display is within the meter sensitivity range. Turn
the jog dial for fine adjustment and perform zero shifting.
(3) Bring the stylus into contact with the first measurement position (cross-section contour)
on a workpiece.
(4) While turning the alignment table, adjust the detector position with the arm feed knob
so that the stylus displacement display is within the meter sensitivity range. Turn the jog
dial for fine adjustment and perform zero shifting.
(7) Bring the stylus into contact with the second measurement position (cross-section
contour).
(8) To measure the second cross-section contour, repeat the procedures in step (4) to step
(6).
(3) Bring the stylus into contact with the first measurement position (cross-section contour)
on a workpiece.
(4) While turning the alignment table, adjust the detector position with the arm feed knob
so that the stylus displacement display is within the meter sensitivity range. Turn the jog
dial for fine adjustment and perform zero shifting.
NOTE ▪ Be sure to enter a height from the top face on the alignment table.
▪ To measure the height of a cross-section contour, use a caliper, etc.
(7) Bring the stylus into contact with the second measurement position (cross-section
contour).
(8) To measure the second cross-section contour, repeat the procedures in step (4) to step
(6).
(4) While turning the alignment table, adjust the detector position with the arm feed knob
so that the stylus displacement display is within the meter sensitivity range. Turn the jog
dial for fine adjustment and perform zero shifting.
(7) Repeat the procedure in step (6) for the required number of points (at least 3 points).
(8) To determine the value for leveling, enter the value of height (Z value) or radius (X
value) of the first cross section.
(9) To perform dual cross-section contour measurement for centering/leveling, move the
stylus to the second cross-section contour, and then repeat the procedure in step (4) to
step (8).
▪ Detector method
Adjusting method by turning the centering knobs (CX, CY) and the leveling knobs (LX,
LY) while monitoring the adjustment value and the displacement of the detector
displayed on the LCD screen.
▪ D.A.T. method
Adjusting method by turning the knob only to adjust the value displayed on the LCD
screen while monitoring the digital display of the centering knobs (CX, CY) and the
leveling knobs (LX, LY)
(2) Turn the alignment table to adjust the table angle (θ value) displayed on the LCD
screen to 0[DEG]. Then the display of knob selection on the LCD screen changes to
[CX].
(3) Turn the CX knob to drive the displacement displayed on the LCD screen into 0.
(4) Turn the alignment table to adjust the table angle (θ value) displayed on the LCD
screen to 90 [DEG]. Then the display of knob selection on the LCD screen changes from
[CX] to [CY].
(5) Turn the CY knob in the same procedure as (3) to drive the displacement on the LCD
screen into 0.
(6) If turning the alignment table after centering has been completed, the LCD screen
returns to the initial screen in section 5.1.2 “Measuring the centering/leveling amounts”.
(2) Turn the arm feed knob to release the stylus from a workpiece.
(3) Press the [ZERO/ABS] button on the CX knob to zero-set the LCD display.
CY centering knob
(4) Turn the CX knob until its LCD display indicates the identical adjustment amount
displayed on the LCD screen of the main unit.
(5) Adjust the CY, LX and LY knobs in the same way the CX knob was adjusted.
(6) Press the [ENT] key to complete the centering and/or leveling adjustment. The LCD
screen returns to the initial screen in section 5.1.2 "Measuring the centering/leveling
amounts".
IMPORTANT ▪ Repeat the procedures in section 5.1.2 “Measuring the centering/leveling amounts" and
section 5.1.3 "Adjusting the centering/leveling amounts" while increasing the meter
sensitivity according to the accuracy required for a workpiece.
TIP ▪ This modification can only be used if the Centering screen is displayed.
▪ Measuring orientation
▪ Measuring method
▪ Notch processing (parameter)
▪ Centering/leveling method
▪ Adjusting method
Centering/
leveling method
Measuring orientation
Adjusting method
Measuring method
Notch processing
TIP The cursor moves in the following directions when the jog dial is turned.
Measuring orientation ←→ Adjusting method ←→ Measuring method ←→ Notch
processing (in case of measuring the entire circumference) ←→ Centering/leveling
method
→ : Clockwise rotation of the jog dial
← : Counterclockwise rotation of the jog dial
TIP ▪ The top face and bottom face cannot be selected for centering (single cross-section),
leveling and centering (dual cross-section), and leveling (dual cross-section).
▪ The outside face, inside face, and opposing face cannot be selected for leveling (single
plane).
NOTE ▪ The notch processing cannot be selected if the measuring method has been set to
[M-Point].
(6) Repeat steps (4) and (5) to modify the setting of each setup item.
(7) Turn the jog dial to select [Set & Exit], then press the [ENT] key.
(5) Turn the jog dial to display a required start point angle, then press the [ENT] key.
(6) Turn the jog dial to display a required end point angle, then press the [ENT] key.
(7) Repeat the procedure in step (4) to step (6) to specify the notched sections.
(8) Turn the jog dial to select [Set & Exit], then press the [ENT] key.
TIP ▪ It is possible to set the position (angle) for a maximum of 20 notched sections. Turn the
jog dial further clockwise if the cursor is on the [5:] line, then the setup screen switches
to that from line [6:] to line [10:].
TIP An alignment table with digimatic heads is required for using D.A.T. function.
TIP ▪ Adjustment position can only be modified, if the Detector mode has been selected as
the condition of “5.2.1.5 Modifying the adjustment method”.
▪ The only item which can be modified is the adjustment position.
(1) When the measurement has been completed, the following adjustment screen is
displayed.
Adjustment position
▪ Roundness
▪ Coaxiality (axial element)
▪ Concentricity
▪ Circular runout (radial direction)
▪ Circular runout (axial direction)
▪ Squareness (axis reference)
▪ Squareness (plane reference)
▪ Thickness deviation
▪ Flatness (1-element)
▪ Parallelism (2-element)
Also, the RA-120 has a versatile element measurement function in addition to being able
to perform the ten basic measurements described above.
For detailed information, refer to section 6.2 "Element Measurement Procedure".
▪ Slight fluctuation takes place in detector displacement immediately after the detector
has been changed in orientation from the vertical to horizontal or vise versa or shifted
along the X or Z-axis. So, allow dozens of seconds for the detector to stabilize before
commencing measuring. When shifting the detector in the Z-axis direction, reciprocate
the detector a few times in the vertical direction for running-in around a target position
and finally stop the detector in position while shifting it upward. This reduces fluctuations
that may take place immediately after the detector shifting.
▪ The effect of table speed fluctuation upon the start of table rotation can be reduced by
commencing measurement well after the table has started rotation.
(2) Press the [RNDNES] key to display the Roundness Measurement screen.
The previously backed-up measurement conditions are displayed in this
screen as the current setup conditions.
If the measurement conditions are not appropriate, make modifications
according to section 6.5 "Modifying the Measurement Screen
Conditions".
Measuring
orientation
Measurement
height:
Jog input
Number of
Tolerance: measurement cross
Jog input sections:
Max. 5 cross
Display graphics: sections.
Circular Jog input
Linear
Notch
Display Calculation conditions:
Magnification: Cutoff: None
Filter: method:
X5 X2K 15 CUT-LEVEL
NON LSC
X10 X5K 50 CUT-ANGLE
2CR-50 MZC
X20 X10K 150
2CR-75 MIC
X50 X20K 500
2CRPC-50 MCC
X100 X50K 15-150
2CRPC-75 15-500
X200 X100K
Gaussian 50-500
X500 X200K
X1K AUTO
(4) Adjust the detector position with the X/Z-axis drive wheel (coarse-feed) and jog dial
(fine-feed) so that the stylus displacement is within the meter sensitivity range on the
LCD.
When the measurement has been completed, the height (Z value) of the measurement
position is automatically entered.
NOTE ▪ Set the value of height (Z value) to be measured from the alignment table top.
▪ To measure the height of a cross-section contour, use a caliper, etc.
The measurement element (cross section) display "A1 [■]" on the LCD is lit.
However, if the Auto-Calculation has been set to "[○]" as the condition of Preferences
Set/Auto-Execute setup, the Result screen automatically appears after the
measurement required for each analysis item is completed.
If recalculating the measured cross section by changing evaluation conditions, press the
[CHANGE] key, then change the setup conditions, referring to section 6.6 "Modifying the
Result Screen Conditions".
Related analysis:
P= [um]
V= [um]
MLA= [um]
PC= [pc]
DX= [um]
DY= [um]
Comment input:
Display graphics: Alphabet,
Circular Numeric value
Linear etc.
Notch processing:
Arbitrary level
Arbitrary angle
Display Calculation Constant pitch
Filter Cutoff
Magnification method Specification release
Batch release
The conditions that have been set in the Setup end
Measurement screen are displayed.
To display the related analysis results, turn the jog dial clockwise.
(2) Press the [COAXAL] key to display the Coaxiality Measurement screen.
The previously backed-up measurement conditions are displayed in this screen as the
current setup conditions.
If the measurement conditions are not appropriate, make modifications according to
section 6.5 "Modifying the Measurement Screen Conditions".
Measuring
orientation
Measurement
height:
Jog input
Number of measurement
Tolerance: cross sections:
Jog input Jog modification.
Refer to section
6.5.1.3.
Notch conditions:
None
Display CUT-LEVEL
Magnification: Cutoff: Calculation CUT-ANGLE
X5 X2K 15 method:
Filter:
X10 X5K 50 LSC
NON
X20 X10K 150 MZC
2CR-50
X50 X20K 500 MIC
2CR-75
X100 X50K 2CRPC-50 15-150 MCC
X200 X100K 2CRPC-75 15-500
X500 X200K Gaussian 50-500
X1K AUTO
(4) Adjust the detector position with the X/Z-axis drive wheel (coarse-feed) and jog dial
(fine-feed) so that the stylus displacement is within the meter sensitivity range on the
LCD.
NOTE ▪ Set the value of height (Z value) to be measured from the alignment table top.
▪ To measure the height of a cross-section contour, use a caliper, etc.
(6) Set the stylus in the measurement position on the second cross section, and then
repeat the procedure in steps (4) and (5).
The measurement element (cross section) display "A2 [●]" (center) on the LCD is lit.
(7) Set the stylus in the measurement position on the third cross section, and then repeat
the procedure in steps (4) and (5).
The measurement element (cross section) display "A3 [■]" (right side) on the LCD is lit.
However, if the Auto-Calculation has been set to "[○]" as the condition of Preferences
Set/Auto-Execute setup, the Result screen automatically appears after the
measurement required for each analysis item is completed.
If recalculating the measured cross section by changing evaluation conditions, press the
[CHANGE] key, then change the setup conditions, referring to section 6.6 "Modifying the
Result Screen Conditions".
Related analysis:
DX= [um]
DY= [um]
Comment input:
Alphabet,
Numeric value
etc.
Notch processing:
Arbitrary level
Arbitrary angle
Display Calculation Constant pitch
Filter Magnification method Specification release
Cutoff
Batch release
The conditions that have been set in the Setup end
Measurement screen are displayed.
To display the related analysis results, turn the jog dial clockwise.
Evaluation Evaluation
cross section2 Reference cross section2 Reference
cross section1 cross section1
(2) Press the [CONCEN] key to display the Concentricity Measurement screen.
The previously backed-up measurement conditions are displayed in this screen as the
current setup conditions.
If the measurement conditions are not appropriate, make modifications according to
section 6.5 "Modifying the Measurement Screen Conditions".
Measuring
orientation
Measurement
height:
Jog input
Number of measurement
Tolerance: cross sections:
Jog input Jog modification.
Refer to section
6.5.1.3.
Notch conditions:
None
Display CUT-LEVEL
Magnification: Cutoff: Calculation CUT-ANGLE
X5 X2K 15 method:
Filter:
X10 X5K 50 LSC
NON
X20 X10K 150 MZC
2CR-50
X50 X20K 500 MIC
2CR-75
X100 X50K 2CRPC-50 15-150 MCC
X200 X100K 2CRPC-75 15-500
X500 X200K Gaussian 50-500
X1K AUTO
(3) Set the stylus in the measurement position on the first cross section.
(4) Adjust the detector position with the X/Z-axis drive wheel (coarse-feed) and jog dial
(fine-feed) so that the stylus displacement is within the meter sensitivity range on the
LCD.
NOTE ▪ Set the value of height (Z value) to be measured from the alignment table top.
▪ To measure the height of a cross-section contour, use a caliper, etc.
(6) Set the stylus in the measurement position on the second cross section, and then
repeat the procedure in steps (4) and (5).
The measurement element (cross section) display "B1 [■]" (right side) on the LCD is lit.
However, if the Auto-Calculation has been set to "[○]" as the condition of Preferences
Set/Auto-Execute setup, the Result screen automatically appears after the
measurement required for each analysis item is completed.
If recalculating the measured cross section by changing evaluation conditions, press the
[CHANGE] key, then change the setup conditions, referring to section 6.6 "Modifying the
Result Screen Conditions".
Related analysis:
DX= [um]
DY= [um]
Comment input:
Alphabet,
Numeric value
etc.
Notch processing:
Arbitrary level
Arbitrary angle
Display Calculation Constant pitch
Filter Magnification method Specification release
Cutoff
Batch release
The conditions that have been set in the Setup end
Measurement screen are displayed.
To display the related analysis results, turn the jog dial clockwise.
Reference
cross section2
Evaluation
cross section3
Reference
cross section1
(2) Press the [RUNOUT (RDL)] key to display the Circular Runout (radial direction)
Measurement screen.
The previously backed-up measurement conditions are displayed in this screen as the
current setup conditions.
If the measurement conditions are not appropriate, make modifications according to
section 6.5 "Modifying the Measurement Screen Conditions".
Measuring
orientation
Measurement
height:
Jog input
Number of measurement
cross sections:
Tolerance: Jog modification.
Jog input Refer to section
6.5.1.3.
Display graphics:
Circular
Linear Notch conditions:
None
Display Calculation CUT-LEVEL
Magnification: Filter: Cutoff: CUT-ANGLE
method:
X5 X2K NON 15
LSC
X10 X5K 2CR-50 50
MZC
X20 X10K 2CR-75 150
MIC
X50 X20K 2CRPC-50 500
MCC
X100 X50K 2CRPC-75 15-150
X200 X100K Gaussian 15-500
X500 X200K 50-500
X1K AUTO
(4) Adjust the detector position with the X/Z-axis drive wheel (coarse-feed) and jog dial
(fine-feed) so that the stylus displacement is within the meter sensitivity range on the
LCD.
NOTE ▪ Set the value of height (Z value) to be measured from the alignment table top.
▪ To measure the height of a cross-section contour, use a caliper, etc.
(6) Set the stylus in the measurement position on the second cross section, and then
repeat the procedure in steps (4) and (5).
The measurement element (cross section) display "A2 [●]" (center) on the LCD is lit.
(7) Set the stylus in the measurement position on the third cross section, and then repeat
the procedure in steps (4) and (5).
The measurement element (cross section) display "A3 [■]" (right side) on the LCD is lit.
However, if the Auto-Calculation has been set to "[○]" as the condition of Preferences
Set/Auto-Execute setup, the Result screen automatically appears after the
measurement required for each analysis item is completed.
If recalculating the measured cross section by changing evaluation conditions, press the
[CHANGE] key, then change the setup conditions, referring to section 6.6 "Modifying the
Result Screen Conditions".
Related analysis:
MAX= [um]
MIN= [um]
A1= [DEG]
A2= [DEG]
Comment input:
Alphabet,
Display graphics: Numeric value
Circular etc.
Linear
Notch processing:
Arbitrary level
Arbitrary angle
Display Calculation Constant pitch
Filter Magnification method Specification release
Cutoff
Batch release
The conditions that have been set in the Setup end
Measurement screen are displayed.
To display the related analysis results, turn the jog dial clockwise.
Evaluation
cross section3
(2) Press the [RUNOUT (AXL)] key to display the Circular Runout (axial direction)
Measurement screen.
The previously backed-up measurement conditions are displayed in this screen as the
current setup conditions.
If the measurement conditions are not appropriate, make modifications according to
section 6.5 "Modifying the Measurement Screen Conditions".
Measuring orientation
Measurement
height: Number of measurement
Jog input cross sections:
Jog modification.
Refer to section
Tolerance: 6.5.1.3.
Jog input
(3) Set the stylus in the measurement position on the first cross section.
(4) Adjust the detector position with the X/Z-axis drive wheel (coarse-feed) and jog dial
(fine-feed) so that the stylus displacement is within the meter sensitivity range on the
LCD.
NOTE ▪ Set the value of height (Z value) to be measured from the alignment table top.
▪ To measure the height of a cross-section contour, use a caliper, etc.
(7) Set the stylus in the measurement position on the third cross section, and then repeat
the procedure in steps (4) and (5).
However, if the Auto-Calculation has been set to "[○]" as the condition of Preferences
Set/Auto-Execute setup, the Result screen automatically appears after the
measurement required for each analysis item is completed.
If recalculating the measured cross section by changing evaluation conditions, press the
[CHANGE] key, then change the setup conditions, referring to section 6.6 "Modifying the
Result Screen Conditions".
Related analysis:
MAX= [um]
MIN= [um]
A1= [DEG]
A2= [DEG]
Comment input:
Alphabet,
Display graphics: Numeric value
Circular etc.
Linear
Notch processing:
Arbitrary level
Arbitrary angle
Display Calculation Constant pitch
Filter Magnification Cutoff method Specification release
Batch release
The conditions that have been set in the Setup end
Measurement screen are displayed.
To display the related analysis results, turn the jog dial clockwise.
Evaluation
cross section1
(2) Press the [SQRNES (AXS)] key to display the Squareness (axis reference)
Measurement screen.
The previously backed-up measurement conditions are displayed in this screen as the
current setup conditions.
If the measurement conditions are not appropriate, make modifications according to
section 6.5 "Modifying the Measurement Screen Conditions".
Measuring orientation
Measurement
height: Number of measurement
Jog input cross sections:
Jog modification.
Refer to section
Tolerance: 6.5.1.3.
Jog input
Notch conditions:
None
CUT-LEVEL
CUT-ANGLE
Display
Magnification: Filter: Cutoff:
X5 X2K 15 Calculation
NON
X10 X5K 50 method:
2CR-50
X20 X10K 150 LSC
2CR-75
X50 X20K 500 MZC
2CRPC-50
X100 X50K 15-150 MIC
2CRPC-75
X200 X100K 15-500 MCC
Gaussian
X500 X200K 50-500
X1K AUTO
(3) Set the stylus in the measurement position on the first cross section.
(4) Adjust the detector position with the X/Z-axis drive wheel (coarse-feed) and jog dial
(fine-feed) so that the stylus displacement is within the meter sensitivity range on the
LCD.
NOTE ▪ Set the value of height (Z value) to be measured from the alignment table top.
▪ To measure the height of a cross-section contour, use a caliper, etc.
(7) Set the stylus in the measurement position on the third cross section, and then repeat
the procedure in steps (4) and (5).
However, if the Auto-Calculation has been set to "[○]" as the condition of Preferences
Set/Auto-Execute setup, the Result screen automatically appears after the
measurement required for each analysis item is completed.
If recalculating the measured cross section by changing evaluation conditions, press the
[CHANGE] key, then change the setup conditions, referring to section 6.6 "Modifying the
Result Screen Conditions".
Related analysis:
ANG= [DEG]
Comment input:
Alphabet,
Numeric value
etc.
Notch processing:
Arbitrary level
Arbitrary angle
Constant pitch
Specification release
Batch release
Filter Display Cutoff Calculation Setup end
Magnification method
The conditions that have been set in the
Measurement screen are displayed.
To display the related analysis results, turn the jog dial clockwise.
Reference
cross section1
(2) Press the [SQRNES (PLN)] key to display the Squareness (plane reference)
Measurement screen.
The previously backed-up measurement conditions are displayed in this screen as the
current setup conditions.
If the measurement conditions are not appropriate, make modifications according to
section 6.5 "Modifying the Measurement Screen Conditions".
Measuring orientation
Measurement
height: Number of measurement
Jog input cross sections:
Jog modification.
Refer to section
Tolerance: 6.5.1.3.
Jog input
Notch conditions:
None
CUT-LEVEL
CUT-ANGLE
Display
Magnification: Filter: Cutoff:
X5 X2K NON 15 Calculation
X10 X5K 2CR-50 50 method:
X20 X10K 2CR-75 150 LSC
X50 X20K 2CRPC-50 500 MZC
X100 X50K 2CRPC-75 15-150 MIC
X200 X100K Gaussian 15-500 MCC
X500 X200K 50-500
X1K AUTO
(3) Set the stylus in the measurement position on the first cross section.
(4) Adjust the detector position with the X/Z-axis drive wheel (coarse-feed) and jog dial
(fine-feed) so that the stylus displacement is within the meter sensitivity range on the
LCD.
(7) Set the stylus in the measurement position on the second cross section, and then
repeat the procedure in steps (4) and (5).
(8) When the measurement has been completed, the height (Z value) of the measurement
position is automatically entered.
The measurement element (cross section) display "A1 [■]" (center) on the LCD is lit.
NOTE ▪ Set the value of height (Z value) to be measured from the alignment table top.
▪ To measure the height of a cross-section contour, use a caliper, etc.
(9) Set the stylus in the measurement position on the third cross section, and then repeat
the procedure in steps (4), (5), and (8). The measurement element (cross section)
display "A2 [■]" (right side) on the LCD is lit.
However, if the Auto-Calculation has been set to "[○]" as the condition of Preferences
Set/Auto-Execute setup, the Result screen automatically appears after the
measurement required for each analysis item is completed.
If recalculating the measured cross section by changing evaluation conditions, press the
[CHANGE] key, then change the setup conditions, referring to section 6.6 "Modifying the
Result Screen Conditions".
Related analysis:
ANG= [DEG]
Comment input:
Alphabet,
Numeric value
etc.
Notch processing:
Arbitrary level
Arbitrary angle
Constant pitch
Specification release
Batch release
Filter Display Cutoff Calculation Setup end
Magnification method
The conditions that have been set in the
Measurement screen are displayed.
To display the related analysis results, turn the jog dial clockwise.
Reference
Evaluation cross section1 cross section1
(2) Press the [THK-DV] key to display the Thickness Deviation Measurement screen.
The previously backed-up measurement conditions are displayed in this screen as the
current setup conditions.
If the measurement conditions are not appropriate, make modifications according to
section 6.5 "Modifying the Measurement Screen Conditions".
Measuring orientation
Measurement
height:
Jog input
Tolerance:
Jog input Notch conditions:
Display graphics: None
Circular CUT-LEVEL
Linear CUT-ANGLE
(4) Adjust the detector position with the X/Z-axis drive wheel (coarse-feed) and jog dial
(fine-feed) so that the stylus displacement is within the meter sensitivity range on the
LCD.
NOTE ▪ Set the value of height (Z value) to be measured from the alignment table top.
▪ To measure the height of a cross-section contour, use a caliper, etc.
(6) Set the stylus in the measurement position on the second cross section, and then
repeat the procedure in steps (4) and (5).
The measurement element (cross section) display "B1 [■]" (right side) on the LCD is lit.
However, if the Auto-Calculation has been set to "[○]" as the condition of Preferences
Set/Auto-Execute setup, the Result screen automatically appears after the
measurement required for each analysis item is completed.
If recalculating the measured cross section by changing evaluation conditions, press the
[CHANGE] key, then change the setup conditions, referring to section 6.6 "Modifying the
Result Screen Conditions".
Related analysis:
MAX= [um]
MIN= [um]
A1= [DEG]
A2= [DEG]
Comment input:
Alphabet,
Display graphics: Numeric value
Circular etc.
Linear
Notch processing:
Arbitrary level
Arbitrary angle
Display Constant pitch
Filter Magnification Cutoff Specification release
Batch release
The conditions that have been set in the Setup end
Measurement screen are displayed.
To display the related analysis results, turn the jog dial clockwise.
(2) Press the [FLTNES] key to display the Flatness Measurement screen.
The previously backed-up measurement conditions are displayed in this screen as the
current setup conditions.
If the measurement conditions are not appropriate, make modifications according to
section 6.5 "Modifying the Measurement Screen Conditions".
Measuring
orientation
Measurement
radius:
Jog input
Number of measurement
Tolerance: cross sections:
Jog input Max. 5 cross sections.
Jog input
Display graphics:
Circular
Linear Notch conditions:
None
Display CUT-LEVEL
Magnification: Cutoff: Calculation CUT-ANGLE
Filter: 15 method:
X5 X2K
NON 50 LS
X10 X5K
2CR-50 150 MZ
X20 X10K
2CR-75 500
X50 X20K
2CRPC-50 15-150
X100 X50K
2CRPC-75 15-500
X200 X100K
Gaussian 50-500
X500 X200K
X1K AUTO
(4) Adjust the detector position with the X/Z-axis drive wheel (coarse-feed) and jog dial
(fine-feed) so that the stylus displacement is within the meter sensitivity range on the
LCD.
However, if the Auto-Calculation has been set to "[○]" as the condition of Preferences
Set/Auto-Execute setup, the Result screen automatically appears after the
measurement required for each analysis item is completed.
If recalculating the measured cross section by changing evaluation conditions, press the
[CHANGE] key, then change the setup conditions, referring to section 6.6 "Modifying the
Result Screen Conditions".
Related analysis:
P= [um]
V= [um]
MLA= [um]
PC= [pc]
DX= [um]
DY= [um]
Comment input:
Alphabet,
Display graphics:
Numeric value
Circular
etc.
Linear
Notch processing:
Arbitrary level
Arbitrary angle
Display Calculation Constant pitch
Filter Magnification method Specification release
Cutoff
Batch release
The conditions that have been set in the Setup end
Measurement screen are displayed.
6.1.10 Parallelism
Measure the following positions to evaluate parallelism.
Evaluation
cross section
Evaluation
cross section
Reference Reference
cross section cross section
(2) Press the [PARLSM] key to display the Parallelism Measurement screen.
The previously backed-up measurement conditions are displayed in this screen as the
current setup conditions.
If the measurement conditions are not appropriate, make modifications according to
section 6.5 "Modifying the Measurement Screen Conditions".
Measuring
orientation
(4) Adjust the detector position with the X/Z-axis drive wheel (coarse-feed) and jog dial
(fine-feed) so that the stylus displacement is within the meter sensitivity range on the
LCD.
(6) Set the stylus in the measurement position on the second cross section, and then
repeat the procedure in steps (4) and (5).
The measurement element (cross section) display "C2 [■]" (right side) on the LCD is lit.
However, if the Auto-Calculation has been set to "[○]" as the condition of Preferences
Set/Auto-Execute setup, the Result screen automatically appears after the
measurement required for each analysis item is completed.
If recalculating the measured cross section by changing evaluation conditions, press the
[CHANGE] key, then change the setup conditions, referring to section 6.6 "Modifying the
Result Screen Conditions".
Related analysis:
ANG= [DEG]
Comment input:
Alphabet,
Numeric value
etc.
Notch processing:
Arbitrary level
Arbitrary angle
Constant pitch
Filter Display Cutoff Calculation Specification release
Magnification method Batch release
The conditions that have been set in the Setup end
Measurement screen are displayed.
To display the related analysis results, turn the jog dial clockwise.
For example, once the five cross sections (a to e) of the following workpiece have been
measured, it is possible to analyze roundness, coaxiality, circular runout, squareness, and
flatness.
Circular runout (radial direction): Cross section b with respect to cross section c and cross
section d
Squareness (axis reference): Cross section e with respect to cross section c and cross
section d
c
a
(3) Press the [CHANGE] key to release the setup conditions from being locked.
(8) Press the [CHANGE] key again to lock the setup conditions. The Measurement screen
is restored.
(10) Set the stylus in the measurement position on the first cross section a.
(11) Adjust the detector position so that the stylus displacement is within the meter
sensitivity range on the LCD
(15) Press the [RESULT] key to display the list of analysis items.
TIP ▪ Set the value of height (Z value) to be measured from the alignment table top.
▪ Analysis items once set up will remain valid until cleared and automatically apply to the
calculation of measurements to be made thereafter. In this case, procedures in the step
(16) through (18) can be omitted.
▪ To clear the analysis items, press the [CHANGE] key twice in succession and, when the
message appears for the confirmation of clearing, press the [ENT] key.
When switching to other analysis results refer to section 6.4 "Switching to Other
Analysis Result Displays".
(2) Turn the jog dial to select the measured element (cross section) to be deleted, then
press the [ENT] key.
(2) Turn the jog dial to select the analysis result to be displayed, then press the [ENT] key.
(3) The Result screen is restored to display the selected analysis result.
▪ Measuring orientation
▪ Number of measurement cross-sections
▪ Measuring sequence of evaluation elements
▪ Notch processing (parameter)
▪ Filter type
▪ Cutoff value
▪ Calculation method
▪ Record magnification
▪ Display graphic format
▪ Tolerance
▪ Analysis item
Analysis item
Measuring
orientation
Number of
measurement
cross-sections
Tolerance
/ Measuring
sequence
Display graphic
format Notch
processing
(parameter)
Record Cutoff value
magnification Filter type
Calculation
method
TIP This modification can only be made if the Measurement screen is displayed.
TIP The cursor moves in the following directions when the jog dial is turned.
Measuring orientation ←→Number of measurement cross-sections/Measuring sequence
←→ Notch processing (parameter) ←→ Filter type ←→ Cutoff value ←→ Calculation
method ←→ Record magnification ←→ Display graphic format ←→ Tolerance ←→
Analysis item
→: Clockwise rotation of the jog dial
←: Counterclockwise rotation of the jog dial
TIP ▪ The top face and bottom face cannot be selected as the reference elements of
roundness, coaxiality, concentricity, circular runout (axial direction) or circular runout
(radial direction), or as the evaluation element of squareness (axis reference) or the
reference element of squareness (plane reference).
▪ The outside face, inside face and opposing face cannot be selected as the evaluation
elements of flatness, parallelism or circular runout (radial direction), or as the reference
element of squareness (axis reference) or the evaluation element of squareness (plane
reference).
▪ The same measuring face (side face or plane) as that of the first cross-section can only
be selected as the second cross-section of thickness deviation.
TIP ▪ This modification can only be made if the analysis item is set to roundness, flatness or
element measurement.
TIP This modification is enabled only if the analysis item is set to coaxiality, concentricity,
circular runout (axial direction), circular runout (radial direction), squareness (axis
reference), or squareness (plane reference).
(3) Turn the jog dial to change the sequence of the evaluation element mark (□).
Determine the sequence of evaluation elements, then press the [ENT] key.
TIP ▪ Any calculation method cannot be selected for thickness deviation because there is no
calculation method for it.
▪ The calculation method of MIC/MCC cannot be selected for flatness or parallelism.
TIP In the Result screen it is possible to modify the record magnification directly (without
unlocking the setup conditions) with the Range key on the control panel.
NOTE ▪ The analysis items that can display their own developed graph are roundness (including
a overwriting print), circular runout (radial direction), circular runout (axial direction),
thickness deviation, and flatness.
▪ A developed graph is displayed with its vertical axis designated as displacement and
with its horizontal axis as angle.
TIP The element measurement can only be selected through the following operation. The
other basic measurement operations can also be selected with the analysis item keys on
the control panel.
▪ Record magnification
▪ Display graphic format
▪ Filter type
▪ Cutoff value
▪ Calculation method
▪ Analysis result
▪ Workpiece name
▪ Operator name
▪ Notch processing (option)
Analysis result
Workpiece
name
Operator name
Notch processing
(option)
Cutoff value
Calculation
Display graphic
Filter type Record method
format
magnification
TIP These modifications can only be made if the Result screen is displayed.
TIP The cursor moves in the following directions when the jog dial is turned.
Record magnification ←→Display graphic format ←→ Analysis result ←→ Workpiece
name ←→ Operator name ←→ Notch processing (option) ←→ Filter type ←→ Cutoff
value ←→ Calculation method
→: Clockwise rotation of the jog dial
←: Counterclockwise rotation of the jog dial
If the result of the analysis item that has been set is selected in step (3):
If [New Analysis]* is selected in step (3) (*: effective only in element measurement).
(4) Turn the jog dial to select a desired analysis item, then press the [ENT] key.
However, it is not possible to select an item that cannot be analyzed.
(5) Turn the jog dial to select an evaluation element (measurement cross-section),
then press the [ENT] key.
(6) Turn the jog dial to select the reference element (reference cross-section), then
press the [ENT] key.
However, this procedure is not required if either roundness, thickness deviation, or
flatness was selected as an analysis item.
(7) The result of the analysis item that has been newly set is displayed.
Turn the jog dial to select characters one at a time, then press the [ENT] key for each
character entry.
(4) Turn the jog dial to select [END] from the list, then press the [ENT] key (the input
character list disappears).
(5) Press the [ENT] key to load the entered workpiece name.
Turn the jog dial to select the characters one at a time, then press the [ENT] key for
each character entry.
(4) Turn the jog dial to select the [END] key, then press the [ENT] key (the input character
list disappears).
(5) Press the [ENT] key to load the entered operator name.
TIP If the selected element has been given notch processing, the element name will change
from uppercase to lowercase (e.g. from A1 to a1).
AE AS
Inside the range
of calculation
The registered conditions can be called by pressing the keys to call conditions, permitting
the operator to omit the condition setup procedures.
Each key to call conditions is available for registering a necessary condition selected from
among (1) calibration ratio, (2) centering conditions, and (3) measuring conditions.
(Registration of all kinds of conditions is also possible with each key.)
NOTE ▪ The conditions that appear at the time of the Condition Registration screen booting are
those of current setup and not the ones registered with a key to call conditions.
TIP ▪ Calibration ratios must have been obtained by calibration measurement beforehand.
▪ Refer to “4. CALIBRATION MEASUREMENT” for the procedures of calibration
measurement.
TIP ▪ If a calibration ratio has been registered, the registered calibration ratio is automatically
called upon press of the corresponding key.
▪ If a calibration ratio has not been registered, the calibration ratio update will not take
place upon press of the corresponding key and the existing ratio will be applied as it is.
Centering [─]
TIP ▪ If the centering conditions have been registered, the registered conditions can be
automatically called by pressing the corresponding key with a simultaneous shift to the
centering screen.
If no centering condition has been registered, the centering condition update will not
take place upon press of the corresponding key without shifting to the centering screen.
▪ Although automatic shifting to the centering screen will take place upon press of the key
to call the centering conditions, quitting the centering screen must be made manually.
Registered measurement
conditions are displayed as shown
below.
Measurement[○][○Roundness][ 3 ]
Measurement [─]
TIP ▪ If the measurement conditions have been registered, the registered measurement
conditions are automatically called by pressing the corresponding key.
▪ If no measurement condition registration has been made, no measurement condition
update takes place upon press of the corresponding key to call conditions.
If a name has been designated, the condition name will be attached to the title of the
measurement, results, centering, and calibration screens respectively called by pressing
the corresponding key to call conditions.
Condition name
Select characters one at a time by turning the jog dial and then pressing the [ENT] key for each
character entry.
(3) Turn the jog dial to select [END] from the list, then press the [ENT] key (the input
character list disappears).
TIP ▪ If part of the conditions has been changed manually after the call of conditions, the
registration name will disappear from the display.
NOTE ▪ Registering the conditions will discard the existing conditions registered with the
corresponding key.
[Internal memory]: Calls conditions that have been output to the internal memory
according to the procedures in “7.3.2 Condition output”.
[USB]: Calls conditions that have been output to a USB memory according
to the procedures in “7.3.2 Condition output”.
(Available only when the USB memory has been recognized.)
[Measurement file]: Calls measurement conditions that have been registered according
to the procedures in “8.1.1 Measurement file registration”.
IMPORTANT ▪ The USB memory cannot be recognized unless inserted prior to powering up.
▪ Use only the USB memory recommended by Mitutoyo.
USB memories other than those recommended may not be recognized.
(3) Pressing the [ENT] key calls the corresponding conditions, updating the conditions
displayed on the Condition Registration screen.
When [Internal memory] or [USB] has been selected in step (2) above:
IMPORTANT ▪ The USB memory cannot be recognized unless inserted prior to powering up.
▪ Use only the USB memory recommended by Mitutoyo.
USB memories other than those recommended may not be recognized.
Select characters one at a time by turning the jog dial and then pressing the [ENT] key
for each character entry.
NOTE Files saved in the USB memory cannot be deleted with the following procedures. They
should be deleted using a personal computer.
(3) A message asking for the confirmation of deletion appears upon press of the [ENT] key.
Press the [ENT] key and the file will be erased.
To initialize the conditions registered with the key to call conditions to the factory default
settings, follow the procedures below.
(4) Pressing the [ENT] key initializes the conditions and the menu screen will be returned
to.
TIP A result file can only be registered if the Measurement screen or Result screen is
displayed.
TIP If a measurement file other than a blank file indicating [-------] is selected, the message to
confirm overwriting is displayed.
Select characters one at a time by turning the jog dial and then pressing the [ENT] key
for each character entry.
(6) Turn the jog dial to select [END] from the list, then press the [ENT] key (the input
character list disappears).
(7) Press the [ENT] key to load the entered measurement file name.
Register measured data and analysis results along with set measurement conditions as a
result file. The registered result file can be recalled at any time for use.
TIP A result file can only be registered if the Result screen is displayed.
(1) Make sure that a USB memory for storing the result files has been inserted in the USB
port.
Select characters one at a time by turning the jog dial and then pressing the [ENT] key
for each character entry.
(7) Turn the jog dial to select [END] from the list, then press the [ENT] key (the input
character list disappears).
(8) Press the [ENT] key to load the entered result file name.
TIP A measurement file can only be recalled if the Measurement screen is displayed.
Recall a result file (including measurement conditions, measured data, and analysis
results).
It is possible to perform re-analysis by modifying the measured data conditions in the
recalled result file.
TIP A result file can only be recalled if the Measurement screen or Result screen is displayed.
(1) Ensure that a USB memory loaded with result files has been inserted in the USB port.
TIP A measurement file can only be deleted if the Measurement screen or Result screen is
displayed.
TIP A result file can only be deleted if the Measurement screen or Result screen is displayed.
(1) Ensure that a USB memory loaded with result files has been inserted in the USB port.
▪ RS-232C output (Measured result data or measurement point group data is output to the
external device.)
▪ USB output (Measurement result data or measurement point group data is saved on a
USB Memory.)
TIP The data output method and content are switched depending on the conditions of
[Preferences Set/Device Setup/Data Output]. For detailed information, refer to section
9.3, "Setting the Devices (RA-120 Hardware)".
NOTE ▪ In SPC output, the RA-120 outputs only the result of the analysis item that has been set
in the measurement condition.
▪ The results of concurrent analysis/re-analysis and related analysis will not be output.
IMPORTANT ▪ The USB memory cannot be recognized unless inserted prior to powering up.
▪ Use only the USB memory recommended by Mitutoyo.
USB memories other than those recommended may not be recognized.
Turn the jog dial to select characters one at a time, then press the [ENT] key for each
character entry.
(4) Turn the jog dial to select [END] from the list, then press the [ENT] key (the input
character list disappears).
Set up the operating environments of the RA-120 according to the following procedure.
The RA-120 retains the setup conditions without change, even if the power is turned off,
and implements the same settings at the next startup.
1 Startup Calib: Automatically displays the Calibration screen when the power
switch is turned on.
(6) Repeat the procedure in steps (4) and (5) to enable/disable the auto-execution function
for each of the required items.
(7) Turn the jog dial to select [0: Setup End], then press the [ENT] key.
2 Related Analysis: Switches between [DX/DY] and [DL/DA] of the related analysis.
5 Superimpose Print: Prints (or displays) multiple cross-sections by overlaying them (only
for roundness and flatness) on the same screen.
9 Meas Posi Input: Inputs a height (Z value) with the jog dial regardless of the presence
or absence of the SD scale unit.
NOTE If [3: Aux. Line], [4: Print Item] or [6: Check message] is selected, the confirmation
message is displayed for all parameters of the item. Set [○] or [▬] for each parameter.
(6) Repeat the procedure in steps (4) and (5) to modify the parameters of the required
items.
(7) Turn the jog dial to select [Setup End], then press the [ENT] key.
4 Data Output: Sets the data output destination and output data type.
6 Print Options: Sets the output contents (e.g. result print and hard copy of screen)
by pressing the [PRINT] key.
(5) Turn the jog dial to adjust the LCD contrast (0 to 10) or buzzer volume (0 to 10),
then press the [ENT] key. (Turn the jog dial to move the cursor up or down.)
(5) The option screen is displayed. Turn the jog dial to select a desired item, then
press the [ENT] key.
(5) The option screen is displayed. Turn the jog dial to select a desired item, then
press the [ENT] key.
NOTE ▪ If [○] has been set for Auto-Print at Preferences Preferences/Auto setup, a measured
result is printed after measurement, even if [Screen Printout] is selected.
(6) Repeat the procedure in steps (4) and (5) to modify the settings of all required items.
(7) Turn the jog dial to select [0: Setup End], then press the [ENT] key.
8 System Init: Restores the system setup descriptions to the initial state.
(5) Turn the jog dial to set a correct date and time, then press the [ENT] key.
(5) The message to confirm initialization is displayed. Press the [ENT] key to select
"Yes".
To cancel zero-shift initialization, press the [CANCEL] key to select "No".
(6) Repeat the procedure in steps (4) and (5) to modify the settings of all required items.
(7) Turn the jog dial to select [0: Setup End], then press the [ENT] key.
NOTE ▪ To enable the disabled keys all together, turn on the power while depressing the
[PRINT] key.
1 CONDITION: Enables or disables the four keys to call conditions, [Condition 1],
[Condition 2], [Condition 3], and [Condition 4] separately to each other.
4 Set. Altern.: Enables or disables the four keys, [MENU], [CHANGE], [CANCEL], and
[ENT] separately to each other.
5 Detector: Enables or disables the two keys, [ZERO SET], and [RANGE]
separately to each other.
6 Print: Enables or disables the two keys, [PRINT], and [FEED] separately to
each other.
8 Measure: Enables or disables the two keys, [START], and [STOP] separately to
each other.
NOTE A confirmation message will appear for each key operation if the following has been
selected, [1: CONDITION], [2: AnalysisItem], [3: Screen], [4: Sett. Altern.], [5: Detector],
[6: Print], or [8: Measure]. Choose “enabled” or “disabled” for every key.
(7) Turn the jog dial to select [Setup End], then press the [ENT] key.
1 Self-test
(6) Turn the jog dial to select [0: Setup End], then press the [ENT] key.
To obtain the highest performance from the RA-120 and use it safely for an extended
period, perform maintenance and inspection according to the following items.
The air bearing employed for the turntable requires the specified air pressure 0.39MPa
(4.0kgf/cm2). Unless the required pressure is supplied, it may be damaged. Check if the
Air NG message is displayed on the LCD screen and if the pressure gage on the air
regulator unit points at less than 0.39MPa (4.0kgf/cm2).
If the air pressure is below the rated pressure, adjust it by means of the
pressure-reducing valve (see the figure below) or by increasing the pressure at the
air-source.
If water and oil mist have gathered in parts “a” and “b” as shown in the figure below,
drain the water and oil mist off the regulators. To drain them away, close the cock to halt
the air supply to the air filter unit and then push in drain cocks “c” and “d” in the figure.
Pressure-reducing valve
Pressure gage
Filter regulator
C b d
Check if the air tube and signal cable between the main unit and air filter unit,
connecting cable of the detector, and power cord are connected properly, and if any of
them is scratched.
The alignment table top is a section of great importance to precision measurement. The
table is protected with rust-preventive oil during storage. Prior to measurement wipe out
the oil using an industrial paper towel or lint-free cloth.
Wipe off dust and dirt on the alignment table and apply rust-preventive oil to it.
Stop air supply to the air filter unit and drain water from the unit. (For more information,
refer to section 10.1.1 ” Inspection before measurement”.)
If the stylus tip of the detector is stained, the measuring accuracy may be deteriorated.
Always maintain the stylus clean.
• Do not use organic solvents such as benzene and thinner. Using either to wipe the main
unit may discolor the surface or peel off the coating. To clean the main unit, use a
CAUTION neutral detergent or a dry cloth.
• When the air pressure indicated on the gage reduces more than 0.098MPa from the
initial pressure by fully opening the reducing valve.
• Perform any work here after the air supply has been halted.
• After the work is complete, supply the air and make sure that there is no air leakage in
CAUTION the system.
To replace the air filter element, first remove the bowl by turning it counterclockwise.
Next, remove the element by turning it counterclockwise.
To assemble the air filter, reverse the above removing procedure. At this time confirm
that the O-ring is mounted on the bowl.
O-ring
Element(Order No.358592)
Bowl
To replace the filter regulator element, first remove the bowl by turning it
counterclockwise. Next, remove the element by turning it counterclockwise.
To assemble the filter regulator, reverse the above removing procedure. At this time
confirm that the O-ring is mounted on the bowl.
O-ring
Element (Order No.358593)
Bowl
Eject button
Open
Printer cover
(2) Load a roll of printer paper as shown in the figure. With the paper aligned parallel to the
printer and the paper edge fed out through the window of the printer cover, close the
printer cover.
Shut
(3) Press the [FEED] key to feed out the paper from the printer outlet.
• Do not manually pull the paper out of the printer and do not feed the paper in the
reverse direction. These actions may damage the printer.
CAUTION • The last meter of printer paper has red markings on both sides. Replace the print paper
with a new roll according to this marking guide. Do not print without paper. This may
cause the printer to fail.
Please follow the procedure below when “ “ is indicated on the LCD of the
digimatic head of the Centripetal aligning table Unit.
Battery
IMPORTANT •When the battery is set, first press the “ON/ZERO switch and do not turn the knob until
the normal value is fixed. If the knob is moved during this time frame, it may be incorrect
counting. If the knob is moved during this time, re-set the battery again.
Push the“ON/ZERO”switch.
ON/ZERO switch
TIP ▪ Our measuring instrument and user’s manual are conforming to four evaluation
methods: LSC, MZC, MIC and MCC.
11.2 Filters
11.2.1 About cutoff values
A work piece cross-section contour detected by the detector on a roundness measuring
instrument includes long-cycle waviness components such as elliptic and parabolic
waveforms, and short-cycle surface roughness components. Depending on the use and
property of a work piece it may be preferable to analyze the data by extracting only
waviness components while removing short-cycle roughness components for roundness
judgment or evaluation. As the unit of extracting extent (cutoff value) the number of
undulations per revolution (upr) is used. (e.g. 15 upr)
11.2.2.1 2CR75
This has the same attenuation characteristic as a pair of C-R circuits with an identical time
constant are connected in series.
The attenuation characteristic is -12dB/oct and the amplitude transmittance at the cutoff
value is 75%, as shown in Figure below.
amplitude transmission(%)
amplitude transmission(dB)
λ/λc
11.2.2.2 2CR50
This is same as the 2CR75 filter except that the amplitude transmittance at the cutoff value
is 50%. Its attenuation characteristic is shown in Figure below.
λ/λc
Attenuation characteristic of 2CR50 filter
1
Attenuation characteristic: H (λ ) = 2
⎛λ⎞
1 + ⎜⎜ ⎟⎟
⎝ λc ⎠
11.2.2.3 2CRPC75
This is a phase corrected type with the same amplitude characteristic as”11.2.2.1 2CR75”.
11.2.2.4 2CRPC50
This is a phase corrected type with the same amplitude characteristic as”11.2.2.2 2CR50”.
λ/λc
1
⎛ α ⋅ λc ⎞
2
⎛ ln 2 ⎞ 2
Attenuation characteristic: H (λ ) = 1 − e −π ⎜ ⎟ where α = ⎜ ⎟ 0.4697
⎝ λ ⎠ ⎝ π ⎠
The use of this filter creates a simple addition:
2CR 75
2CR 50
Gaussian Gaussian
λ/λc λ/λc
2CR75-Gaussian 2CR50-Gaussian
Roundness
Related Assessment Results: All evaluation methods, except for (5), are LSC or MZC
Peak height
Valley depth
Mean roundness
“Number of peaks” is determined from the number of peaks higher than the reference
circle A peak between the point where the data rises above the reference circle and that
where the data drops below the reference circle is referred to as one peak.
(5) Off-centering :
Coaxiality
11.3.1.3 Concentricity
“Concentricity” is determined by doubling the deviation of the
measured cross-section center from the reference measured
cross-section center, using the measurement data of two given
cross-sections.
(The centers are determined according to each roundness
evaluation method.)
Concentricity
Reference axis
(1) Angle :
Reference plane
“Number of peaks” is the number of peaks higher than the reference plane. The data
points between the point where the data rises above the reference plane and that where
the data below the reference plane is referred to as one peak.
(5) Off-centering :
LS MZ MI MC LS MZ MI MC
Roundness ○ ○ ○ ○ - - - -
*Concentricity ○ ○ ○ ○ ○ ○ ○ ○
*Coaxiality (axis-elem) ○ ○ ○ ○ ○ ○ ○ ○
Flatness (1-element) ○ ○ ○ ○ - - - -
*Parallelism ○ ○ - - ○ ○ - -
(2-elements)
Squareness ○ - - - ○ ○ ○ ○
(axial-reference)
FIX
Squareness ○ ○ ○ ○ ○ - - -
(plane-reference) FIX
Runout (radial) - - - - ○ ○ ○ ○
Runout (axial) - - - - ○ ○ ○ ○
LS: the least squares method MZ: the minimum zone method
MI: the maximum inscribed method MC: the minimum circumscribed method
*: The calculation methods of an evaluation element and a datum element are identical.
11.4 Datum
11.4.1 Kind of datum
DA0: Denotes the axis (rotation axis of the turntable) used for the mechanical center.
▪ An analysis items in the table is evaluated using either of the datum’s marked in the
table.
▪ For information about the configuration elements and the definition of each datum, refer
to other tables.
Concentricity - ○/● - -
Coaxiality (axis-elem) ● ● ○/● -
Parallelism (2-elements) - - - ○/●
Squareness (axial-reference) ● ● ○/● -
Squareness (plane-reference) - - - ○/●
Runout (radial) ● ● ○/● -
Runout (axial) ● ● ○/● -
○: Datum setup is possible for discrete datum measurements.
●: Datum setup is possible for element measurements.
11.4.3 Table of Datum and Its Necessary Number of Elements
·The axis of DA0 uses for mechanical center.
A (Outside)/B C (Top)/D
(Inside) (Bottom)
2) A4 or A3 size.
11.6.1.3 Printer
RA-120
PRINTER
(Canon,
EPSON)
D-sub 25pin
Centronics 36pin
Record magnification Record range (μm) Record range (μm) Record range (μm)
×5K ±2 ±4 ±8
×10K ±1 ±2 ±4
×20K ±0.5 ±1 ±2
11.7.1.5 Connection to PC
RA-120
DOS/V
Frame Frame
RA-120
DOS/V
Frame
D-sub 9pin
Frame
D-sub 25pin
Mm/inch
0 (0000) : M
1 (1000) : I
Measured data (D6~D11) Decimal point
Sign
0 (0000) : nnnnnn
0 (0000) : +
1 (1000) : nnnnn.n
8 (0001) : -
2 (0100) : nnnn.nn
3 (1100) : nnn.nnn
4 (0010) : nn.nnnn
5 (1010) : n.nnnnn
[Ex.] +16.45μm
Characteristic of AC
Symbol condition Min. Max. Unit
t1 Com 1 0 2 s
t2 Com 1 15 μs
t3 Com 1 100 μs
t4 Com 1 100 μs
t5 Com 1 0 - μs
t7 Com 2 - 80 ms
▪ I/F: USB2.0/1.1
+Cut level
-Cut level
-FULL
Cut angle
Item RA-120
Bearing type Air-bearing
Rotating accuracy: (0.04+6H/10000)μm
radial direction H: Measuring height (mm) JIS B 7451-1997
Rotating accuracy: (0.04+6X/10000)μm
axial direction X: distance from the rotation center (mm)
Rotating speed 6 rpm
Turntable unit Table diameter Ф150mm
Maximum loading
25kg
weight
Maximum probing Ф280mm (380mm: with the detector holder
diameter attached reversely, in vertical orientation only)
Maximum loading
Ф440mm
diameter
Traverse method Manual
Vertical travel amount 280mm from the turntable top
280mm from the turntable top
Vertical drive unit Maximum probing
(480mm: with the detector attached reversely, in
height
vertical orientation only)
Maximum probing
100mm (ID more than Ф30mm)
depth
Traverse method Manual
Radial drive unit Horizontal travel
-25mm to 50mm
amount
Measuring force 70 to 100mN (±30%) 0mN
Detector Stylus tip form Carbide ball Ф1.6mm
Measuring range ±1000μm [±.04”]
Measuring direction Switching styles with dual direction
NOTE • Only standard stylus is available for this detector, not long stylus.
• Use an optional auxiliary stage for measuring a workpiece whose diameter is 20mm or
shorter and whose height is 20mm or lower.
Remedy: Adjust the system to the current date and time from the System
Setup menu, then charge the battery. If the system power is
retained ON, the battery is automatically charged.
Full-charge time: 16 hours (The system can be backed up for 2
or 3 months.)
Cause: Only this analysis can be selected with the current measured
data.
Cause: The calibration ratio exceeds the calibration value range (0.5 to
5.3).
Remedy: Check the stylus for dust and dirt before displacement, then
calibrate the stylus again.
Check the setting value (master value), then calibrate the stylus
again.
Cause: The table was rotated while an alignment knob was being
adjusted for centering/leveling.
Cause: The number of measured data points is reduced less than the
effective range through the notch processing.
Remedy: Change the analysis method or reduce the deleting data points.
Cause: There is a mistake in setting the start or end point for notch
processing.
LSC: More than 45° as the minimum analysis data range
Other than LSC: More than 45° as the minimum analysis data
range, unless data range is more than 180°.
• E5001 No FD is inserted.
Remedy: Install a USB memory and then turn on the system again.
Remedy: Contact the nearest Mitutoyo service network with the detailed
state of this phenomenon.
Remedy: Contact the nearest Mitutoyo service network with the detailed
state of this phenomenon.
Cause: Invalid key was set with the Key Setup menu.
Remedy: Set the key valid with the Key Setup menu.
To validate all keys, turn off the power once, then turn on the
power again while holding down the [PRINT] key.