Envirom Bedhead Trunking System

The Horizontal H-Sys bedhead trunking system and Vertical V-Sys/


NV-Sys headwall shall be constructed from custom designed extruded
aluminium sections with customer specied powder coated 60%
gloss nish fascia panels. Fascia panels shall be cut prior to painting to
ensure all surfaces are coated, providing a tight seal between panels
to prevent dust traps. Cover strips on the front fascia panels shall not
be allowed. All visible extruded aluminium sections shall be powder
coated RAL9010 60% gloss by a DuPont/Akzo Nobel approved powder
coating specialist, ofering a minimum guaranteed service life of 25
years. End caps shall be manufactured from 2.5mm thick UV stabilised
and re retardant high-impact Fabex 578. A removable UV stabilised
polymer extrusion shall cover the fascia xing screws, providing a tight
seal to prevent dust traps. A UV stabilised elastomeric wall seal shall
run the full length of the bedhead unit, providing a dust tight seal
between the bedhead unit and the wall and shall cater for a 10mm
tolerance in the atness of the mounting surface. Three segregated
service compartments (H-Sys) and a segregated service compartment
(V-Sys) shall run the length of the unit to carry medical gas pipes, low-
voltage electrical cables and ELV/data, with segregation of services
being maintained throughout. Each bedhead unit shall be supplied
pre-piped, wired and certied.
The design and conguration of the bedhead units shall fully comply
with all relevant applicable standards, including HTM 2007, HTM
2011, HTM 2015, HTM 2020, HTM 2022, HTM02-01, HTM08-03, BS
EN ISO 11197, BS EN 60601-1, BS EN 60598-1 and BS EN 60598-2-25,
BS 6496, BS 7671, BS EN 60439, IEC 60364-7-710, CIE, CIBSE LG2,
CIBSE LG3.
Medical Gases
The H-Sys compartment for housing medical gas services shall be
capable of running 3 pipes of 15mm diameter with axes on a common
vertical plane to facilitate simple on-site brazing to the piped
distribution system.
The V-Sys compartment for housing medical gas services shall be
capable of running pipes of 15mm diameter generously spaced to
facilitate simple on-site brazing to the piped distribution system.
The headwall shall be capable of housing at least three terminal units
in a horizontal array.
Copper pipes shall be manufactured from phosphorous de-oxidised
non-arsenical copper to BS EN 1412:1996 grade CW024A and be
manufactured to metric outside diameters in accordance with BS EN
13348:2001R250 (half hard). Degreasing of pipe shall be such that
there is less than 20mg/m2 (0.002mg/cm2) of hydrocarbons on the
degreased surface when tested by the method specied in ASTM B280
clause 12. The type of terminal unit installed shall be specied by the
client. Hoses shall not be used to connect the medical gas terminal
units to the distribution system.
Lighting
Difusers shall be manufactured from extruded re-retardant Lexan
ML3290 polycarbonate resin, incorporating prismatic inner surfaces to
maximise efciency of light distribution from the chosen source.
Efciency shall be further enhanced by the use of mirror nish
reectors manufactured from Alanod Miro4 or Miro27 aluminium,
achieving a minimum clarity and total reection to TR-2 or
DIN 5036-3 of 95%.
Envirom Bedhead Trunking and Headwall Systems
SPECIFICATION
2002639
28/07/10
Page 1 of 2
Issue 4
BeaconMeds Telford Crescent, Staveley, Derbyshire S43 3PF Tel: +44 (0) 1246 474 242 Fax: +44 (0) 1246 472 982
Company Reg No. 2957933 (English) Company VAT Reg No. 598612590 www.beaconmedaes.com gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 CHAS accredited contractor SAFEcontractor Reg No. N00184450
E
N
V
I
R
O
M

B
E
D
H
E
A
D
T
R
U
N
K
I
N
G

S
Y
S
T
E
M
Luminaires shall be provided with electronic ballasts suitable for use
with TL5 high efciency uorescent tubes, with a power factor rating of
at least cos=0.93.Lighting controls shall be as agreed with the
client and shall include options for local and/or remote control,
control via the nursecall handset or control via a Digital Addressable
Lighting Interface (DALI) or equivalent system.
Electrical Sockets
Electrical sockets shall normally be tted in the side panels of the V-Sys
headwall, with additional sockets being tted to the front fascia panel
as required. Electrical sockets shall be wired in ring or radial mains to
circuits as specied by the customer.
Communications
Provision for or tting of the nursecall system shall be co-ordinated by
the bedhead unit supplier. Data sockets, including, but not limited to
RJ45 and telephone sockets shall be installed in the bedhead unit at
the time of manufacture.
Medical Equipment Rail
Medical equipment rails shall be designed in accordance with BS EN
12218:1999, manufactured from a hollow rectangular stainless steel
prole of 30mm high by 10mm deep. It shall be possible to retrot
further medical rails to the bedhead unit after installation without the
need for power tools and without the need to disrupt the continuity of
services provided by the bedhead unit.
LEXAN is a registered trademark of the General Electric Company
Horizontal H-Sys Cross Sectional View
2
6
5
101
88
130
121
1
6
6
Direct/Reading Light
Indirect/Room Light
Low Voltage
Gas Pipes
Retro-fit rail
ELV/Data
In an efort to continuously improve our products,
the right is reserved to change the specication of the
items described herein at any time. Please contact us
for further information and up to date specications.
2002639
28/07/10
Page 2 of 2
Issue 4
Typical Conguration - Horizontal Unit (H-Sys)
8
EM
0 0
O2
E DA S
1E 0
6
M
G m e E DA S
1 0
4 A M
G m e 1 EM0
Ge m
c a V
S A DE
8 0 0
S A DE E
M
2 O
6 6 0 0 8 6 0 8 0
e
01m G EE DA
M
S
4 A
m 1 0
e G
M
E E DA S
Vac
m 1 0
e
M
G
Relay
downdrop
Optional
G
e
m

1
0

O
x
y
g
e
n
G
e
m

1
0

M
e
d
i
c
a
l

A
i
r
E
q
u
i
p
m
e
n
t

R
a
i
l
G
e
m

1
0

V
a
c
u
u
m
N
u
r
s
e
c
a
ll
P
r
o
v
is
io
n
R
J
4
5

c
a
t
.

5
e
D
o
w
n
l
i
g
h
t

s
w
i
t
c
h
/
d
i
m
m
e
r
U
p
l
i
g
h
t

s
w
i
t
c
h
/
d
i
m
m
e
r
T
e
l
e
p
h
o
n
e

s
o
c
k
e
t
(
s
e
c
o
n
d
a
r
y
)
T
w
i
n

u
n
s
w
i
t
c
h
e
d

1
3
A
(
n
o
r
m
a
l

c
i
r
c
u
i
t
)
T
w
i
n

u
n
s
w
i
t
c
h
e
d

1
3
A
(
e
s
s
e
n
t
i
a
l

c
i
r
c
u
i
t
)
T
w
i
n

u
n
s
w
i
t
c
h
e
d

1
3
A
(
n
o
r
m
a
l

c
i
r
c
u
i
t
)
Feed
1
8
0
0
m
m
t
o

o
o
r
6 8 0 0 6 8 0 0
Cross-Section
1900
1240
670
160 130 160 131 2 x 24W T5 Reading Light 150
Equipment rail 545mm
135 135 85
200
90
1 x 54W T5 Indirect Room Light
1095
170
8
6
1
0
8
1
3
0
In an efort to continuously improve our products,
the right is reserved to change the specication of the
items described herein at any time. Please contact us
for further information and up to date specications.
BeaconMeds Telford Crescent, Staveley, Derbyshire S43 3PF Tel: +44 (0) 1246 474 242 Fax: +44 (0) 1246 472 982
Company Reg No. 2957933 (English) Company VAT Reg No. 598612590 www.beaconmedaes.com gbn.info@beaconmedaes.com
Constructionline Reg No. 75463 CHAS accredited contractor SAFEcontractor Reg No. N00184450
m
6 0 8 0
E E DA SM1 0 e G
6 0 0 8
EE DA M S m 1 0 e G
8 0 0
S A DE E M
2 O
6
e01m G
G m 1 0 e
6
M EE DA S
0 0 8
Ge01m S M A DE E
8 0 0 6
Ge01 M
c a V
S A DEE
0 8 0 6
m
M 4 A
4 A M
O2
V c a
Vac
8 0 0
S A DE E M
6
e01m G
8 0 0
S A DE E M
2 O
6
e01m G
G m 1 0 e
6
M EE DA S
0 0 8
4 A M
(essential circuit)
Unswitched 13A sockets
Unswitched 13A sockets
(normal circuit)
Telephone socket
(secondary)
Nursecall Provision
RJ45 cat. 5e
Twin unswitched 13A sockets
(normal circuit)
Medical Equipment
Rail 30mmx 10mm
Rail 30mmx 10mm
Medical Equipment
Wall Seals
Gem10 Oxygen
Gem10 Medical Air
Gem10 Vacuum
Examination Lamp
Envirom NV Sys Envirom V Sys
452
438
715
1
0
0
1
0
0
2
0
0
Side
View
125
Typical Conguration - Vertical Headwall (V-Sys) and Narrow Vertical Headwall (NV-Sys)
Envirom Colour Selection Chart
Pure White
RAL9010
Light Pink
RAL3015
Pastel Blue
RAL5024
Pastel Green
RAL6019
Telegrey 4
RAL7047
Cream
RAL9001

Você também pode gostar