Você está na página 1de 6

ARTICLE IN PRESS

Journal of Biomechanics 38 (2005) 1932–1937


www.elsevier.com/locate/jbiomech
www.JBiomech.com
Technical note

Is the force-length relationship a useful indicator of contractile


element damage following eccentric exercise?
Timothy A. Butterfield, Walter Herzog
Human Performance Laboratory, Faculty of Kinesiology, University of Calgary, 2500 University Drive NW, Calgary, Alta. Canada T2N 1N4
Accepted 11 August 2004

Abstract

Eccentric exercise has been shown to have a measurable effect on the force–length relationship (FLR), as peak force is shifted to
longer muscle lengths following exercise. Recently, this shift in the FLR has been proposed as a ‘‘simple, reliable indicator’’ for
assessing contractile element damage following eccentric exercise. However, eccentric exercise causes fatigue and damage, and there
is evidence that fatigue alone may also cause a shift in the FLR. The purpose of this paper was to assess the role of fatigue on the
FLR (as measured by a torque–joint angle relationship) following isometric and eccentric exercise in the New Zealand white (NZW)
rabbit. Six NZW rabbits were divided into two groups for eccentric or isometric contractions of the hindlimb dorsiflexor muscles.
Pre- and post-exercise torque–joint angle relationships were measured, and the shift from the pre- to the post-exercise relationship
was measured as the change in joint angle at which peak torque was produced. Eccentric exercise resulted in a rightward shift of
seven degrees; isometric exercise, which is thought to not cause damage, resulted in a shift of four degrees. Furthermore, torque
production was reduced to a greater extent at short compared to long muscle lengths for the eccentric and isometric exercise,
resulting in a post-exercise torque–joint angle relationship that was altered in shape. We conclude from these results, that the shift in
peak torque may not be a simple and reliable indicator of muscle damage, but is caused by a combination of damage and post-
exercise fatigue.
r 2004 Elsevier Ltd. All rights reserved.

Keywords: Muscle; Force–length relationship; Eccentric; Isometric; Fatique

1. Introduction (Brockett et al., 2001; Morgan and Allen, 1999; Morgan


et al., 1996; Brockett et al., 2002). Recently, the rightward
Eccentric exercise has been shown to have a measurable shift in the FLR has been proposed as a ‘‘simple, reliable
effect on the force–length relationship (FLR), as peak indicator’’ for assessing contractile element damage
force is produced at a longer muscle length following following eccentric exercise (Whitehead et al., 2003).
exercise (Talbot and Morgan, 1996; Jones et al., 1997; The rightward shift in the FLR was first identified
Wood et al., 1993). Several mechanisms producing this with contractile element injury by Katz (1939), who
rightward shift in the FLR following exercise have been proposed that contractile elements were transformed
proposed. These include stretch-induced damage to into passive elastic elements by muscle stretch. Morgan
calcium handling structures, (Balnave and Allen, 1995; (1990) proposed that this transformation of contractile
Warren et al., 1993; Endo, 1972), visco-elastic creep elements during eccentric contractions occurs through a
(Fowles et al., 2000) , and contractile element damage ‘popping’ of a small number of sarcomeres as they are
stretched beyond myofilament overlap, resulting in an
Corresponding author. Tel: +1-403-220-8525; fax: +1-403-284- increase in muscle compliance. As a consequence, the
3553. muscle would produce peak force at a longer length,
E-mail address: walter@kin.ucalgary.ca (W. Herzog). thereby exhibiting a rightward shift of the FLR.

0021-9290/$ - see front matter r 2004 Elsevier Ltd. All rights reserved.
doi:10.1016/j.jbiomech.2004.08.013
ARTICLE IN PRESS
T.A. Butterfield, W. Herzog / Journal of Biomechanics 38 (2005) 1932–1937 1933

This rightward shift of the FLR following eccentric nerves, superior to the gastrocnemius and distal to the
contractions has been observed in a variety of studies branching of the sciatic nerve. In this manner, all
and muscle preparations, including human plantarflex- dorsiflexors of the tibiotarsal joints were stimulated
ors (Whitehead et al., 1998; Jones et al., 1997), and effectively. Following implantation, the skin was stapled
hamstrings (Brockett et al., 2001), single toad fibers closed.
(Jones et al., 1997), single frog fibers (Morgan et al.,
1996), and individual motor units from the cat medial 2.3. Exercise and measurement protocol
gastrocnemius (Brockett et al., 2002). In each of these
studies, the shift was associated with contractile element 2.3.1. Eccentric exercise, group 1
damage. In addition, histological findings post-exercise Rabbits were placed supine in a stereotaxic frame
supported the concept of contractile element injury with the knee at 901 flexion (full extension=01). The
within the muscle following eccentric exercise (Jones et foot was strapped to a servo-motor footplate (Parker
al., 1997), although the possible contribution of fatigue Hannifin Corporation, Irwin, PA, USA) and ankle
on the FLR was not considered. movement was controlled via computer (Motion Plan-
Unlike eccentric contractions, isometric exercise does ner, Rohnert Park, CA, USA). The tibiotarsal
not produce extensive contractile element damage (Koh joint angle was set at 901 (full extension=1801) which
and Brooks, 2001), and presents a unique model to served as the reference angle for the remainder of
study the effects of fatigue on the FLR in the absence of the experiment. The peroneal nerve cuff leads were
muscle injury. Although previous studies have assessed attached to a stimulator (Grass S8800, Astro-Med
the effects of fatigue on the force–length relationship, Inc., Longueil, QC, Canada), and the a-motoneuron
the use of single fibers (Yeung et al., 2002a, b) or lack of threshold was determined (pulse duration=0.1 ms,
direct comparison to the post eccentric exercise FLR frequency=150 Hz).
(Cutlip et al., 2004) leaves the contribution of fatigue to First, a pre-exercise, isometric torque–joint angle
the rightward shift of the FLR following eccentric relationship was determined by supramaximally stimu-
exercise unclear. Therefore, the purpose of this study lating (3  a-motoneuron threshold voltage, pulse dur-
was to directly test the hypothesis that fatigue con- ation=0.1 ms, frequency=150 Hz, train duration=
tributes to the rightward shift in the FLR of the ankle 2000 ms) the dorsiflexor muscles, beginning at a
dorsiflexors in New Zealand white rabbits following tibiotarsal angle of 551 and progressing in 51 increments
eccentric exercise. to 1551. To limit the potential effects of creep of passive
elements at long muscle lengths, the foot was parked at a
551 joint angle (short muscle length) between isometric
2. Materials and methods contractions. Therefore, following each isometric con-
traction, the foot was returned passively to a dorsiflexed
2.1. Experimental design position (551 tibiotarsal angle) for 2 min of rest between
contractions.
Eight skeletally mature NZW rabbits (5.670.2 kg, The exercise protocol consisted of eccentric contrac-
Riemens, St. Agatha, Ontario, Canada) were divided tions performed from a tibiotarsal angle of 70–1051 of
into two study groups for measurement of a pre-exercise plantar flexion at 701 s 1(3  a-motoneuron threshold
and post-exercise torque–joint angle relationship. voltage, pulse duration=0.1 ms, frequency=150 Hz,
Group one consisted of five rabbits (nine hindlimbs) train duration=500 ms). Each stretch-shortening cycle
subjected to eccentric exercise. Group two consisted of lasted 1000 ms. Stimulation started at the onset of
three rabbits (six hindlimbs) subjected to an equivalent plantar flexion and ceased at the end of the eccentric
isometric exercise. All procedures were approved by the exercise (1051), resulting in an eccentric exercise of
Animal Care Committee of the University of Calgary. 500 ms duration. Immediately following the eccentric
contraction, the foot was passively returned to the 701
2.2. Surgical procedure starting position at 701 s 1(500 ms passive shortening).
To induce muscle injury associated with eccentric
Rabbits were tranquilized with 0.18 ml (10 mg/ml) contractions, it has been shown that greater than one
acepro-25 (MTC Pharmaceuticals, Cambridge, ON, contraction is required when working within the
Canada) and held under anesthesia with 1.5% isoflur- physiologic range of motion (Cutlip et al., 2004;
ane, 0.6 L/min N2O and 0.8 L/min O2. An incision was Gosselin and Burton, 2002). Therefore, the exercise
made on the posterior aspect of the right and left protocol consisted of five sets of ten eccentric contrac-
hindlimbs, anterior to the sciatic vein, and the biceps tions, with 2 min rest between sets. Immediately after the
femoris and semimembranosis were separated exposing eccentric protocol, a post-exercise torque-joint angle
the peroneal nerve. Nerve cuff stimulating electrodes relationship was measured in a manner identical to the
were secured over the right and left common peroneal pre-exercise relationship. However, the rest between
ARTICLE IN PRESS
1934 T.A. Butterfield, W. Herzog / Journal of Biomechanics 38 (2005) 1932–1937

isometric contractions was reduced to 30 s to limit the expressed as a percent decrease from the pre-exercise
recovery from fatigue during measurement of the FLR values. All statistical analyses were performed using
post-exercise. Therefore, the foot was returned to a SPSS version 11.5 (SPSS Incorporated, Chicago, IL,
dorsiflexed position (551 tibiotarsal joint angle) between USA). For all parameters measured, means7SE are
contractions, but for only 30 s of rest before moving to reported. Statistical significance was set at po0.05, and
the next joint angle. analyzed using Student’s paired, or independent, t-tests.

2.3.2. Isometric exercise, group 2


Using identical stimulation parameters as described 3. Results
above, the pre-exercise torque–joint angle relationship
was measured. Then, isometric contractions (3x a- Following eccentric exercise, peak torque values at all
motoneuron threshold voltage, pulse duration=0.1 ms, angles were significantly lower compared to those
frequency=150 Hz, train duration=500 ms, 500 ms rest) obtained pre-exercise for all joint angles measured
were performed at a tibiotarsal angle of 951. This joint (po0.001, Fig. 1). Analysis of the torque–angle relation-
angle was chosen for the isometric contractions for ships revealed a rightward shift for all subjects in the
several reasons: Firstly, pilot data indicated that it was, eccentric group, and all but one subject in the isometric
on average, the angle at which peak isometric torque group (Table 1). Overall, mean peak torque production,
was produced by the dorsiflexors; secondly, it lies within as calculated from the polynomial fitted to the mean
the range of motion for our eccentric protocol; and torque values following eccentric exercise, occurred at a
lastly, we found that little or no passive tension was joint angle of 1041 compared to the pre-exercise peak of
associated with this muscle length, thus limiting the 981, resulting in a 61 rightward shift. The rightward shift
effects of creep during testing. The 500 ms train calculated as the average of individual fitted polyno-
duration, and 500 ms rest between contractions in each mials was 71 (Table 1). In addition, the shape of the
protocol, assured that stimulation duration and timing torque-angle relationship was rotated in a counter-
were identical between the two exercise groups. The clockwise manner following eccentric exercise (Fig. 1,
protocol consisted of five sets of 10 isometric contrac- inset).
tions with 2 min rest between sets. A post-exercise Following isometric exercise, peak torque values at all
torque–joint angle relationship was measured immedi- angles were significantly lower compared to those
ately following exercise as described above. obtained pre-exercise for all joint angles measured
(po0.001, Fig. 2). Analysis of joint torque decrement
2.4. Torque-joint angle evaluation between protocols revealed the greatest decrement in
joint torque production following isometric exercise was
Two precision strain gauges (Vishay Micro-Measure- 40.573.0% at a tibiotarsal joint angle of 601. The
ments, Raleigh, NC, USA. Item No. CEA-06-125WT- greatest torque decrement following eccentric exercise
350) were arranged in a full bridge configuration, was 55.073.0%, and was measured at the smallest
creating an instrumented cam connecting the footplate tibiotarsal joint angle, 551 (Fig. 3). Torque decrement
to the motor. Strain gauge calibration confirmed the following eccentric exercise was not significantly differ-
linearity of output voltage as a function of strain ent compared to the decrement following isometric
(R240.99) through a range greater than required for exercise, except at short muscle lengths (po0.002),
our protocol. The output signal was routed through corresponding to the smallest joint angles (55–701,
a strain gauge amplifier/conditioner (Vishay 2100, Fig. 3). The mean peak torque value, calculated from
Vishay Micro-Measurements, Raleigh, NC, USA. Gain the polynomial fitted to the mean torque values
8.44  100; Excitation 10 Vdc; low pass filter 10 Hz) and following isometric exercise, was shifted to the right
sampled at 250 Hz with WinDaq data acquisition by four degrees, from 1011 to 1051, and the curve
software (Dataq Instruments, Akron, Ohio, USA). was rotated in a counterclockwise direction (Fig. 2,
Peak torque values were plotted for all 21 tibiotarsal inset). Calculating the rightward shift as the average
joint angles. Mean torques across all joints were plotted of individual shifts resulted in a mean rightward shift of
pre- and post-exercise. In order to calculate the angle of 31, (Table 1).
peak torque occurrence, torque values475% of the
absolute peak torque were normalized, fit with a second-
order polynomial, and peak torque was calculated as the 4. Discussion
peak value of the polynomial approximation. In
addition, the mean rightward shift, calculated as the Although a rightward shift in the FLR has been
average of the shifts for each subject, was calculated for associated with histological evidence of contractile
comparison (Jones et al., 1997; Whitehead et al., 2001). element damage following eccentric exercise, the results
For force decrement data, post-exercise torques were of this study suggest that a portion of the rightward shift
ARTICLE IN PRESS
T.A. Butterfield, W. Herzog / Journal of Biomechanics 38 (2005) 1932–1937 1935

1.1
1.0 R2 = 0.98
R2 = 0.99
0.9
1.2
0.8
0.7
1.0 0.6
0.5
Joint Torque [Nm]

60 70 80 90 100 110 120 130 140


0.8

0.6

0.4

0.2

0.0
50 60 70 80 90 100 110 120 130 140 150 160
Tibiotarsal Joint Angle [degrees]

Fig. 1. Torque–angle relationship for NZW dorsiflexors before (J) and after () a bout of eccentric exercise. Note the apparent shift to the right in
peak force production (from 100 to 1051) and the change in shape of the torque-joint angle relationship following exercise. Torque production
following eccentric exercise was significantly reduced compared to pre-exercise values at all joint angles tested (po0.001). Values are mean7SE
(n=9). Inset. Normalized torque–angle relationship. Here the rightward shift and counterclockwise rotation are evident. Data were fitted with a
second-order polynomial to calculate the angle of peak torque production pre-exercise (981) and post-exercise (1041).

Table 1 to 43% by calculating the shift as the mean shift


Shift in peak torques from pre- to the post-exercise torque–angle obtained from the individual subjects (Table 1).
relationship for individual subjects in the eccentric and isometric test
groups.
Jones et al. (1997) also observed a rightward shift in
the FLR using passive stretch combined with isometric
Eccentric Shift in Peak Isometric Shift in Peak exercise in human plantar flexor muscles that amounted
exercise Torque Exercise Torque to 50% of the shift following an eccentric exercise
ECC1L +71 ISO1L +31 protocol. However, the authors deemed the isometric
ECC2L +41 ISO1R +31 shift insignificant. In the same paper, a significant
ECC2R +71 ISO2L 01 rightward shift in optimum force production of 3.5%
ECC3L +121 ISO2R +51 muscle length was also reported for single toad fibers
ECC3R +41 ISO3L +31
following a protocol of passive stretch and isometric
ECC4L +91 ISO3R +31
ECC4R +121 contractions (Jones et al., 1997). Although these results
ECC5L +41 are similar to ours, a direct comparison is difficult as the
ECC5R +51 use of passive stretch may have produced damage (Koh
MEAN +71 MEAN +31 and Brooks, 2001), that could have contributed to the
A positive value indicates a rightward shift of the peak torque towards shift.
increased muscle lengths. The mean rightward shift was calculated as Other studies have reported conflicting results com-
the average of the individual shift values. ECC=eccentric, pared to those presented here. Whitehead et al. (2003)
1,2,3,4,5=subject number; L, R=left and right. ISO=isometric, showed a small shift in peak force production following
1,2,3=subject number; L,R,=left and right.
isometric exercise in the cat medial gastrocnemius,
which would only account for a small proportion of
the shift reported following eccentric exercise (o10%).
and change in shape of the torque–angle relationship However, their fatigue protocols resulted in much
after exercise is contributed by the effects of fatigue. In smaller force decrements at optimal length following
our study, the isometric exercise resulted in a rightward exercise (7% and 13%) compared to ours (29%).
shift of 41 fitting mean torque values with one Conversely, single fiber experiments have shown no
polynomial (Fig. 2). This amounts to 67% of the 61 shift in peak force production following isometric
rightward shift obtained following the eccentric exercise exercise (Yeung et al., 2002a). However, a direct
protocol using the same procedure (Fig. 1). The overall comparison between their in vitro protocol and our
contribution of fatigue to the rightward shift is reduced study is difficult.
ARTICLE IN PRESS
1936 T.A. Butterfield, W. Herzog / Journal of Biomechanics 38 (2005) 1932–1937

1.1

1 2 2
R = 0.98 R = 0.99
0.9

0.8
1.6
0.7

1.4 0.6
0.5
1.2 60 70 80 90 100 110 120 130 140
Joint Torque [Nm]

0.8

0.6

0.4

0.2

0
50 60 70 80 90 100 110 120 130 140 150 160
Tibiotarsal Joint Angle [degrees]

Fig. 2. Torque–angle relationship for NZW dorsiflexors before (J) and after () a bout of isometric exercise. Note the apparent lack of rightward
shift in peak force production (1051). Torque production following isometric exercise was significantly reduced compared to pre-exercise values at all
joint angles tested (po0.001). Values shown are mean7SE (n=6). Inset. Normalized torque-angle relationship. Here the rightward shift and
counter-clockwise rotation are evident. Data were fitted with a second order polynomial to calculate the angle of peak torque production pre-exercise
(1011) and post-exercise (1051).

70 following both eccentric exercise (Fig. 1, inset) and


isometric exercise (Fig. 2, inset). Decreased isometric
Post Exercise Torque Decrement [%]

60 * force production has been shown to be greatest at short


*
*
* muscle lengths following isometric (Gauthier et al.,
50
1993; Cutlip et al., 2004) and eccentric exercise (Cutlip et
al., 2004), implicating metabolic changes as the common
40
ECC factor in the altered shape of the ascending limb of the
ISO
30 FLR following exercise. Here, we showed that isometric
exercise produced the greatest reduction in joint torque
20 at short muscle lengths. Similarly, the greatest reduction
in joint torque following eccentric exercise
10 (55.073.0%), also occurred at the smallest joint angle
tested (551, Fig. 3), corresponding to the shortest muscle
0 length.
50 60 70 80 90 100 110 120 130 140 150 160
Tibiotarsal Joint Angle [degrees] In this study, the eccentric exercise was performed on
the ascending limb and plateau region of the FLR. It has
Fig. 3. Percent decrease in joint torque production immediately been shown that markers of muscle injury, including the
following 50 eccentric contractions (B) and 50 isometric contractions
(’). Visual inspection of the graph shows similar shaped traces for
shift in FLR, are greater on the descending, compared
both the eccentric (ECC) and the isometric (ISO) groups, as the to ascending, limb of the FLR (Whitehead et al., 2003;
greatest loss of torque production following both protocols occurred at Cutlip et al., 2004; Talbot and Morgan, 1998). There-
short muscle lengths. Analysis of the data indicate a significantly fore, it is possible that the contribution of fatigue to the
greater reduction in joint torque in the ECC group compared to the FLR shift following eccentric exercise at long muscle
ISO group at joint angles of 55–701, as indicated by (*). Significance of
po0.05 was reported. Values are means7SE (ECC n=4; ISO n=6).
lengths would be diminished, due to a greater rightward
shift following eccentric exercise.
Since fatigue has been shown to have a greater effect
The contribution of fatigue to the rightward shift of at short compared to long muscle lengths (Gauthier et
the FLR may be associated with the change in shape of al., 1993; Cutlip et al., 2004), torque production is
the FLR following both types of exercise, as the shapes reduced to a greater extent at short compared to long
of the torque–angle curves were similar between the lengths, producing a post-exercise torque–angle rela-
exercise groups (Figs. 1 and 2). Simply, the torque–angle tionship that is altered in shape. Torque decrement at
relationship appears to be rotated counterclockwise short muscle lengths following eccentric exercise is
ARTICLE IN PRESS
T.A. Butterfield, W. Herzog / Journal of Biomechanics 38 (2005) 1932–1937 1937

similar to that seen following isometric exercise (Fig. 3), Gauthier, A.P., Faltus, R.E., Macklem, P.T., Bellemare, F., 1993.
but increased in magnitude for the eccentric compared Effects of fatigue on the length-tetanic force relationship of the rat
diaphragm. Journal of Applied Physiology 74, 326–332.
to the isometric protocol. Thus, we concluded that
Gosselin, L.E., Burton, H., 2002. Impact of initial muscle length on
metabolic changes are implicated in the altered shape of force deficit following lengthening contractions in mammalian
the ascending limb of the FLR following eccentric skeletal muscle. Muscle Nerve 25, 822–827.
exercise. This, in conjunction with the use of curve Jones, C., Allen, T., Talbot, J., Morgan, D.L., Proske, U., 1997.
fitting to predict the peak of the FLR, although well Changes in the mechanical properties of human and amphibian
accepted (Wood et al., 1993; Jones et al., 1997; Talbot muscle after eccentric exercise. European Journal of Applied
Physiology and Occupational Physiology 76, 21–31.
and Morgan, 1998; Whitehead et al., 1998), may Katz, B., 1939. The relation between force and speed in muscular
produce the rightward shift due to fatigue. contraction. Journal of Physiology 96, 45–64.
Koh, T.J., Brooks, S.V., 2001. Lengthening contractions are not
required to induce protection from contraction-induced muscle
injury. American Journal of Physiology-Regulatory Integrative and
5. Conclusion Comparative Physiology 281, R155–R161.
Morgan, D.L., 1990. New insights into the behavior of muscle during
We demonstrated that a rightward shift in the FLR active lengthening. Biophysical Journal 57, 209–221.
following exercise may be partially attributed to the Morgan, D.L., Allen, D.G., 1999. Early events in stretch-induced
muscle damage. J.Appl.Physiol 87, 2007–2015.
effects of fatigue. Although it is well accepted that Morgan, D.L., Claflin, D.R., Julian, F.J., 1996. The effects of repeated
contractile element damage plays an important role in active stretches on tension generation and myoplasmic calcium in
the altered mechanical properties of muscle following frog single muscle fibres. Journal of Physiology 497 (Part 3),
eccentric contractions, the role of fatigue with respect to 665–674.
the shift in peak force production seems to have been Talbot, J.A., Morgan, D.L., 1996. Quantitative analysis of sarcomere
non-uniformities in active muscle following a stretch. Journal of
overlooked. Based on the current results, fatigue may be
Muscle Research and Cell Motility 17, 261–268.
responsible for up to 50% of the measured rightward Talbot, J.A., Morgan, D.L., 1998. The effects of stretch parameters on
shift in peak torque production following eccentric eccentric exercise-induced damage to toad skeletal muscle. Journal
exercise. Therefore, the shift in peak torque is not of Muscle Research and Cell Motility 19, 237–245.
necessarily a simple and reliable indicator of muscle Warren, G.L., Lowe, D.A., Hayes, D.A., Karwoski, C.J., Prior, B.M.,
Armstrong, R.B., 1993. Excitation failure in eccentric contraction-
damage.
induced injury of mouse soleus muscle. Journal of Physiology 468,
487–499.
Whitehead, N.P., Allen, T.J., Morgan, D.L., Proske, U., 1998.
References Damage to human muscle from eccentric exercise after training
with concentric exercise. Journal of Physiology 512 (Part 2),
Balnave, C.D., Allen, D.G., 1995. Intracellular calcium and force in 615–620.
single mouse muscle fibres following repeated contractions with Whitehead, N.P., Morgan, D.L., Gregory, J.E., Proske, U., 2003.
stretch. Journal of Physiology 488 (Part 1), 25–36. Rises in whole muscle passive tension of mammalian muscle after
Brockett, C.L., Morgan, D.L., Gregory, J.E., Proske, U., 2002. eccentric contractions at different lengths. Journal of Applied
Damage to different motor units from active lengthening of the Physiology 95, 1224–1234.
medial gastrocnemius muscle of the cat. Journal of Applied Whitehead, N.P., Weerakkody, N.S., Gregory, J.E., Morgan, D.L.,
Physiology 92, 1104–1110. Proske, U., 2001. Changes in passive tension of muscle in humans
Brockett, C.L., Morgan, D.L., Proske, U., 2001. Human hamstring and animals after eccentric exercise. Journal of Physiology 533,
muscles adapt to eccentric exercise by changing optimum length. 593–604.
Medicine and Science in Sports and Exercise 33, 783–790. Wood, S.A., Morgan, D.L., Proske, U., 1993. Effects of repeated
Cutlip, R.G., Geronilla, K.B., Baker, B.A., Kashon, M.L., Miller, eccentric contractions on structure and mechanical properties of
G.R., Schopper, A.W., 2004. Impact of muscle length during toad sartorius muscle. American Journal of Physiology 265,
stretch-shortening contractions on real-time and temporal muscle C792–C800.
performance measures in rats in vivo. Journal of Applied Yeung, E.W., Balnave, C.D., Ballard, H.J., Bourreau, J.P., Allen,
Physiology 96, 507–516. D.G., 2002a. Development of T-tubular vacuoles in eccentrically
Endo, M., 1972. Stretch-induced increase in activation of skinned damaged mouse muscle fibres. Journal of Physiology 540, 581–592.
muscle fibres by calcium. Nature-New Biology 237, 211–213. Yeung, E.W., Bourreau, J.P., Allen, D.G., Ballard, H.J., 2002b. Effect
Fowles, J.R., Sale, D.G., MacDougall, J.D., 2000. Reduced strength of eccentric contraction-induced injury on force and intracellular
after passive stretch of the human plantarflexors. Journal of pH in rat skeletal muscles. Journal of Applied Physiology 92,
Applied Physiology 89, 1179–1188. 93–99.

Você também pode gostar