Escolar Documentos
Profissional Documentos
Cultura Documentos
m
Proboscis
Proboscis
Labium
(sheath)
Head
4.5 1 mm
100 m
50
Labrum
Maxilla
Maxillary palp
Phase 1
Phase 2
Phase 3
Phase 4
Movement
of head
Movement
of body
Movement of
maxillary palp
Fig. 4. Observation results.
Soft slime sheet
Up Down
Fig. 5. Schematic motion of mosquito proboscis based
on observation. The head and three needles are
gradually lowered from Phase 1 to Phase 4. These
phases are repeated at several Hz (2Hz, ranging 1-7Hz).
Moving direction
Head
Labrum
Maxilla
Phase 1 Phase 2 Phase 3 Phase 4
2297
distribution related to proboscis motion has not been cleared
yet. We tackled this problem. The dead mosquitoes were
soaked in formaldehyde. After taken out, they were
embedded in paraffin. The bulk paraffin embedding
mosquitoes was sliced to thin films using a microtome (Leica
Microsystems, RM2125RT), each thickness of which is 5 m.
The films were stretched and stuck on preparation glasses,
followed by a staining process using hematoxylin and eosin
solutions. An example of results is shown in Fig. 7. The
stained area corresponds to a cell nucleus. Taking account
that cell nucleuses are lined up in muscle fibers, muscle
distribution can be obtained. Observing many samples, we
have judged that the muscles are only located on the two
places in the jaw, which move two maxillae, respectively. We
have also found that there are no muscles for moving the
labrum, and no muscles on the needles themselves. It means
that the mosquito moves two maxillae by two muscles of the
jaw, and it moves the labrum by its head and body.
III. PROPOSAL OF EQUIVALENTLY NEGATIVE STIFFNESS
MECHANISM
We noticed that the maxilla has a jagged harpoon-like
shape on its tip area. This shape is said to be functional for
easy insertion [1]. There are many hypotheses for the function
of jagged shape [1, 12]: however, none of them has been
confirmed either theoretically or experimentally at the present
state. One of hypotheses is that the jagged parts may have the
function to anchor the maxillae to the skin organization and
may assist the labrums easy insertion by some mechanism,
which is focused on in this section.
Based on the observation results mentioned in the previous
section, an analytical model of mosquitos head and proboscis
including muscles is constructed, as shown in Fig. 8. In the
following of this section, this figure is referred to.
As for the maxilla, the following equation holds true:
(1)
where 1 i means that the right maxilla is considered, and
2 i means that the left one is considered. i
m is the mass of
maxilla, i
x is the displacement of maxilla (downward is
positive direction),
i skin
f is the frictional force given by skin
organization,
i muscle
f is the driving force given by the muscle
located at the base of the jaw.
As for the labrum with head, the following equation holds:
(2)
where c
m is the total mass of head and labrum,
c skin
f is the
frictional force given by skin organization,
c muscle
f is the
driving force given by the muscle located between the body
and the head (see Fig. 8), the term of
1 2 )
(
muscle muscle
f f is
the reactive force given by muscles for driving maxillae.
As the insertion displacement becomes large, the frictional
force becomes large due to the increase of interface area
between the needle and the skin organization. Therefore, the
following equations hold true for the maxillae:
(3)
where
i forward
k is the equivalent stiffness when the maxilla
moves forward, i.e., downward, and
i back
k is that when it
moves backward, i.e., upward. Due to the harpoon-like
jagged shape on maxillas tip area,
i forward i back
k k holds true.
The validities of these equations have been experimentally
confirmed using microneedles fabricated by
micro-electro-mechanical systems (MEMS) technology
(details are explained in the next section). Similarly, the
following equation holds true for the labrum with head:
(4)
The mass of maxilla is negligible compared with those of
head and labrum, i.e., holds true. Although the three
needles move synchronously with certain phase differences,
they gradually go forward, which means their displacements
are assumed to be approximately the same level, i.e., i c
x x
holds true.
Under these assumptions, the following equation is
obtained using Eqs. (1)-(4) in the case that the labrum goes
forward and the two maxillae go backward, i.e., in Phases 2
and 4 shown in Figs. 4 and 5:
(5)
In this equation, the term of ( 2 )
c forward i back
k k is the
equivalent stiffness in pushing out the labrum. If this term can
be set below zero, the mechanism would have equivalently
negative stiffness. In other words, the central needle is easily
inserted to the skin by no forward driving force on it ideally,
while backward driving forces are applied on the side needles.
We named this mechanism as negative stiffness
mechanism. Considering that the stiffness of pulling up
maxilla is considerably large due to anchoring effect arising
from its jagged harpoon-like shape, there is possibility that
the mosquito is actually utilizing this mechanism. This
Fig. 7. Stained sectional cell organization of mosquitos
jaw area. Two muscle fibers are seen, which drive two
maxillas. There is not a muscle fiber for driving labrum.
Muscle fiber
m
i
x
i
=f
i skin
+ f
i muscle
, i=1,2,
m
c
x
c
=f
c skin
+f
c muscle
- f
1 muscle
- f
2 muscle
,
f
i skin
= -k
i forward
x
i
(x
i
>0), k
i back
x
i
(x
i
<0) ,
f
c skin
= -k
c forward
x
c
(x
c
>0), k
c back
x
c
(x
c
<0) .
m
c
x
c
+( k
c forward
2k
i back
)x
c
=f
c muscle
.
m
i
m
c
Fig. 8 Analytical model of mosquitos head and proboscis.
(a) Schematic front view. (b) Model including muscles. m
c
Head and labrum, m
1
Right maxilla, m
2
Left maxilla.
Magnified
proboscis
(a)
x
c
x
1
, x
2
m
2
Leg
m
1
Head
Body
m
c
(b)
Muscle
2298
mechanism could be applied to artificial medical needle
systems for reducing the pain in future.
IV. EXPERIMENTAL INVESTIGATION ON EQUIVALENT
STIFFNESS USING ARTIFICIAL SILICON MICRONEEDLE
IMITATING MOSQUITO
A. Purpose of Experiment
It is difficult to estimate the actual parameters of mosquito
used in the model described in the previous section.
Especially, we want to know the parameters of
c forward
k ,
c back
k ,
i forward
k , and
i back
k of mosquito, and we want to
investigate whether the negative stiffness mechanism holds
true or not. However, the estimation of these parameters is
impossible at present state, limited by the available
measurement instrumentation. Therefore, we fabricated
microneedles artificially, which really imitate mosquito
proboscis in terms of shape and size. Mosquito proboscis is
made of organic material of chitin. It is difficult to fabricate a
real imitation using this biomaterial at present. Thus, we
employed silicon material, which has a good compatibility to
MEMS fabrication. Although the mechanical properties such
as Youngs modulus are different from mosquitos, the
experimental data of
c forward
k ,
c back
k ,
i forward
k , and
i back
k
using the silicon imitation would be useful for discussing the
proposed equivalent stiffness issue.
B. Fabrication of Microneedle Using MEMS Technology
Silicon microneedles were fabricated using MEMS
technology. The fabrication process is shown in Fig. 9. A thin
silicon wafer having 50 m in thickness is employed. The
surface of it is oxidized, followed by photolithography to
define the etching part of silicon. Using the defined SiO
2
area
as an etching mask, silicon is etched by deep reactive ion
etching (DRIE) process using SF
6
/C
4
F
8
gas. Photoresist and
SiO
2
are removed by wet etching, leaving a silicon
microneedle, which has two-and-half dimensionally sharp tip.
For sharpening a tip to make a real three dimensionally
conical shape, an electrochemical etching method, which was
invented by the authors [13], were employed, as shown in Fig.
10. Optical microscope images of the fabricated microneedles
are shown in Fig. 11. One is straight shaped; another is
harpoon-like jagged shaped. Almost the same size and the
shape as mosquitos proboscis were successfully achieved.
There sizes are far smaller compared to currently used
commercial stainless needles.
C. Experimental Estimation of Equivalent Stiffness
The resistance force during inserting/extracting the
fabricated needles to/from an artificial skin of silicone rubber
was detected, as shown in Fig. 12 [12]. A silicone rubber
(thickness is 1 mm, Youngs modulus is 2.2 MPa) is attached
to a load cell (the rated load is 2 N, the linearity is 0.5% to the
full range). A personal computer can detect displacement and
load in real time with resolution of 0.5 mN. As for further
details of system, please see ref. [12], due to want of space.
Experimental result of the resistance force relative to the
forward moving distance is shown in Fig. 13. That in the
backward case is shown in Fig. 14.
The equivalent stiffness, which is the ratio of the resistance
force relative to the moving distance, can be obtained using
the data in Figs. 13 and 14. They are summarized in Table I.
D. Discussion
It is confirmed that the equivalent stiffness in the backward
motion of jagged needle takes large value due to anchoring
effect. Assume the fabricated needles are used corresponding
to the mosquitos central labrum and side two maxillae, then,
the term of ( 2 )
c forward i back
k k , which is equivalent stiffness
in pushing out central needle, takes the minus value of -45
gf/mm according to Table I. It indicates that the proposed
Chemical reactions:
Si + 2H
2
O SiO
2
+ 4H
+
SiO
2
+ 6HF H
2
SiF
6
+ 2H
2
O
Fig. 10. Schematic of electrochemical etching for
sharpening a needle.
Hydrofluoric acid (HF) 10%
Anode Cathode
Si Si
Condition:
200 V, 100 mA, 10~15 s
Fig. 9. Micromachining process of silicon needle.
(d) Etching Si using DRIE
(b) Photolithography
(e) Electrochemical etching of
tip of needle after removal
of photoresist and SiO
2
(c) Etching SiO
2
by CHF
3
Si
(a) Oxidation
SiO
2
Photoresist Sharp tip
Length: 2 mm
Width: 30 m
Thickness: 50
m
Length: 2 mm
Width: 15 m
Thickness: 50 m
Enlargement
Fig. 11. Optical images of the fabricated needles with
sharp tip. (a) Straight needle. (b) Jagged needle.
(a) (b)
2299
negative stiffness mechanism holds true in this experimental
case, which would make the insertion process easy.
On the other hand, the equivalent stiffness in pulling out
central needle, which is expressed as ( 2 )
c back i forward
k k , also
takes the minus value of -2 gf/mm. The results indicate that
the mosquitos body supporting its proboscis is always pulled
to the human skin independent of labrum moving direction.
We anticipate that the legs support the body against the
pulling force. In the contrary case that body is pushed against
the skin, the body is lifted up; the legs can no longer support it,
since legs seem not to have anchoring function on their tip.
V. CONCLUSIONS
Based on the observation using two high speed camera
systems, the easy insertion mechanism using three needles
was proposed. It is as follows: the equivalent stiffness, which
is the ratio of the resistance force relative to the moving
distance, was considered. The total stiffness in pushing out
the central needle can be made smaller than zero due to the
anchoring effect of the side needles. We named it as negative
stiffness mechanism. Microneedles were fabricated using
MEMS technology. The obtained experimental equivalent
stiffness showed the validity of proposed mechanism.
In the present paper, the proposed mechanism was
preliminarily discussed using experimental data on a single
needle. The further experimental discussion using three
needles like real mosquito proboscis is the ongoing work.
ACKNOWLEDGMENT
This work was supported by JSPS.KAKENHI(23656188).
REFERENCES
[1] T. Ikeshouji, The Interface between Mosquitoes and Humans,
University of Tokyo Press, 1999 (in Japanese).
[2] K. Oka, S. Aoyagi, Y. Arai, Y. Isono, G. Hashiguchi, and H. Fujita,
Fabrication of a Micro Needle for a Trace Blood Test, Sens.
Actuators, vol. 97-98C, pp. 478-485, 2002.
[3] T. Maeno, Structure and Function of Finger Pad and Tactile
Receptors, J. The Robotics Society of Japan, vol. 18, no. 6, pp.
761-771, 2000.
[4] S. Henry, D. V. MecAllister, M. G. Allen, and M. R. Prausnitz,
Micromachined Needles for the Transdermal Delivery of Drugs, in
Proc. IEEE Int. Conf. Micro-electro-mechanical Systesms (MEMS
98), pp. 494-498, 1998.
[5] P. K. Campbell, K. E. Jones, R. J. Huber, K. W. Horch, and R. A.
Normann, A Silicon-Based, Three-Dimensional Neural Interface:
Manufacturing Processes for an Intracortical Electrode Array, IEEE
Trans. Biomed. Eng., vol. 38, no. 8, pp.758-768, 1991.
[6] S. J. Moon and S. S. Lee, Fabrication of Microneedle Array Using
Inclined LIGA Process, in Tech. Digest Transducers 03, pp.
1546-1549, 2003.
[7] K. Najafi, J. Ji, and K. D. Wise, Scaling Limitations of Silicon
Multichannel Recording Probes, J. Biomed. Eng., vol. 37, no.1, pp.
1-11, 1990.
[8] S. Chandrasekaran, J. D. Brazzle, and A. B. Frazier, Surface
Micromachined Metallic Microneedles, J. MEMS, vol. 12, no. 3, pp.
281-288, 2003.
[9] P. A. Stupar and A. P. Pisano, Silicon, Parylene, and Silicon/Parylene
Micro-Needles for Strength and Toughness, in Tech. Digest
Transducers 01, pp. 1386-1389, 2001.
[10] http://www.terumo.co.jp/English
[11] S. Aoyagi, H. Izumi, T. Aoki, and M. Fukuda, Development of a
Micro Lancet Needle Made of Biodegradable Polymer for Low
Invasive Medical Treatment, in Tech. Digest Transducers05, pp.
1195-1198, 2005.
[12] S. Aoyagi, H. Izumi, and M. Fukuda, Biodegradable Polymer Needle
with Various Tip Angles and Effect of Vibration and Surface Tension
on Easy insertion, Sens. Actuators, vol. A143, pp. 20-28, 2008.
[13] H. Izumi, M. Suzuki, S. Aoyagi, and T. Kanzaki, Realistic Imitation of
Mosquitos Proboscis: Electrochemically Etched Sharp and Jagged
Needles and Their Cooperative Inserting Motion, Sens. Actuators, vol.
A165, pp. 115-123, 2011.
[14] H. Saito, S. Yanai, Y. Ohta, and T. Ogawa, Comparison of Vibration
Waveform and Frequency Effecting to Reduce the Needle Puncturing
Force, in Proc. Jpn. Soc. Medical and Biological Eng., 2003, p. 199,
2003 (in Japanese).
[15] H. Saito, K. Mitsubayashi, and T. Togawa, Detection of Needle
Puncture to Blood Vessel by Using Electric Conductivity of Blood for
Automatic Blood Sampling, Sens. Actuators, vol. A125, no.2, pp.
446-450, 2006.
[16] M. K. Ramasubramanian, O. M. Barham, and V. Swaminathan,
Mechanics of a mosquito bite with applications to microneedle design,
IOP journal on Biomimetics and Bioinspiration, vol. 3, 2008, pp.
0460001.
[17] A. N. Clement, The Biology of Mosquitoes, CABI Publishing, 2000.
Fig. 12. Experimental equipment for detecting force during
insertion and extraction of needle. Details are in ref. [12].
Strain gauge amp.
Encoder
pulse
Silicone rubber
Load cell
Encoder (8000 pulse/R)
Controller
Program
PC
PC
Command
Motor
Voltage
Microneedle
One-axis moving stage
Holder
TABLE I
EXPERIMENTAL RESULTS OF EQUIVALENT STIFFNESS
Straight needle (gf/mm) Jagged needle (gf/mm)
Forward
c forward
k =11
i forward
k =7
Backward
c back
k =12
i back
k =28
0
1
2
3
4
5
6
7
8
0 0.1 0.2 0.3 0.4 0.5
Jagged needle
Straight needle
Fig. 13. Experimental result of resistance force in forward motion.
Straight needle
Jagged needle
c forward
k
i forward
k
R
e
s
i
s
t
a
n
c
e
f
o
r
c
e
[
g
f
]
Displacement [mm]
-6
-5
-4
-3
-2
-1
0
1
0 0.1 0.2 0.3 0.4 0.5 0.6
Jagged needle
Straight needle
Fig. 14. Experimental result of resistance force in backward motion.
Straight needle
Jagged needle
R
e
s
i
s
t
a
n
c
e
f
o
r
c
e
[
g
f
]
c back
k
i back
k
Displacement [mm]
2300