BIOLYSIS BIOGAS GAS ANALYSIS SYSTEM

BIOGAS COMPOSITION

Compound % volume
Methane (CH4) 50 - 75
Carbon Dioxide (CO2) 25 50
Nitrogen (N2) 0 -10
Hydrogen (H2) 0 1
Hydrogen Sulfide (H2S) 0 3
Oxygen (O-2) 0 - 0

In BIOGAS application, hydrogen sulfide require much more attention due to it corrosiveness
to certain type of metal or plastic parts, thus special type of material in used for field
instrument or method of reducing the hydrogen sulfide concentration is required to prevent
damage or failure causes to the field instrument or gas engine generator.

Hydrogen Sulfide (H2S) gas by itself is a colorless, very poisonous, corrosive agent and
classified as flammable gas having an explosive limit between 4.5% - 46%vol

BIOGAS contains fractions of water vapor as well; combination of hydrogen sulfide with
water vapor forming corrosive sulfuric acid (H2SO4).

BIOLYSIS Biogas analysis system, designed and built to prolong the life span of the
equipment functioning by selection right component considering component market
availability and best possible material in used considering to both hydrogen sulfide and weak
sulfuric acid reaction.

The technologies in used to measure the Methane and Carbon Dioxide is NDIR, the best
possible low cost technologies available comparing using expensive technologies like FID
detector for methane analysis; NDIR technologies having a weakness to measure accurate if
present of water vapor due to wave length absorption greater than any other measured
gases. Care is required by removing the water vapor to minimal possible in order for the
analyzer measure accurately.

In cases where hydrogen sulfide or oxygen measurement is require, BIOLYSIS offering
inexpensive electrochemical cell or partial sensor technology (Oxygen only).

BIOLYSIS system complete with sample filter probe taking care of possible present of
particulate in the pipe line as well as isolation valve, purge port for purging or portable unit
insertion port and pressure relief valve in event overpressure.

BIOLYSIS System equips with ventilation system, fire extinguisher, safety awareness
labeling and gas detector as standard package for safety purposes in the event of internal
gas leakage or possible of fire may happen.

Note: ANSI TECHNOLOGIES SDN BHD the reserve the right making any improvement to the
system without prior notice






ANSI TECHNOLOGIES SDN BHD
No. 78, Jalan BP5/11, Bandar Bukit Puchong Industrial Park, 43701 Puchong Selangor
Tel: +6012 476 6861 / Fax: +603 8061 2417
Page 2 of 3



ANSI TECHNOLOGIES SDN BHD (870848H)









TECHNICAL SPECIFICATION
Gas Analyzer Measurement Parameter
Methane (CH4)
Method: NDIR
Range: 0 100%vol
Resolution/Detection level: 0.10%vol
Accuracy: +/- 3 % rel., 0.5 % abs.
Response time (T90) < 45s
Carbon Dioxide (CO2)
Method: NDIR
Range: 0 100%vol
Resolution/Detection level: 0.10%vol
Accuracy: +/- 3 % rel., 0.3 % abs.
Response time (T90) < 45s
Hydrogen Sulfide (H2S)
Method: Electrochemical cell
Range: 0 1000/6000 ppm
Resolution/Detection level: 1 ppm
Accuracy: +/- 5 % rel., 5 ppm abs.
Response time (T90) < 45s
Oxygen (O2)
Method: Partial pressure
Range: 0 25%vol
Resolution/Detection level: 0.01 %vol
Accuracy: +/- 5 % rel., 0.2 % abs.
Response time (T90) < 45s
Gas Analyzer Built-up Specification
Supply 24VAC 120W
Analyzer gas pump Membrane pump; capacity 1.5LPM;
Liquid and solid Filtration Hydrophobic PTFE membrane filter
Data logger Independent data logger MMC 256MB storage card (As optional)
Interface RS232 (For Analyzer configuration only)
Analogue Output 4-20mA/ 0 10V active (require isolation for passive type interface)
Operating Temperature 0 - 50C
Ambient Humidity 5 90%RH non condensing
Calibration
(NDIR Sensor)
Zero Calibration:
Type, Auto programmable
Source of Zero Calibration gas:
Ambient air with special filtration element filter
Span Calibration:
Type- Manual Calibration
Source of Span Calibration gas: Cylinder gas traceable NIST standard
Analyzer Calibration Option Built Up:
10 point calibration, stored in EEPROM. Correction of the calibration curve possible
with zeroing and one point calibration.
Frequency of Calibration:
Every 6 monthly
Calibration Gas CH4/CO2 mixtures calibration gas blended traceable to NIST standard



ANSI TECHNOLOGIES SDN BHD
No. 78, Jalan BP5/11, Bandar Bukit Puchong Industrial Park, 43701 Puchong Selangor
Tel: +6012 476 6861 / Fax: +603 8061 2417
Page 3 of 3


ANSI TECHNOLOGIES SDN BHD (870848H)



Calibration
(Electrochemical & Partial Pressure Sensor)
Zero Calibration:
Type, Auto programmable
Source of Zero Calibration gas:
Ambient air with special filtration element filter
Span Calibration:
Type- Manual Calibration
Source of Span Calibration gas: Cylinder gas traceable to NIST standard
Frequency of Calibration:
Every 6 monthly
Calibration Gas CH4/CO2/H2S mixtures calibration gas blended traceable to NIST standard
Panel Built Up
Material Built-Up Stainless steel 304; (Stainless Steel 316 as optional)
Type Custom made single layer wall mouthing with transparent acrylic glass
(Floor standing as Optional)
Size Approximately 1000(H) x 800 (D) x 400 (D); subject to design changes
Sampling Conditioning/Sample Line/Sample Probe Built-up Specification
Internal Fitting and Tubing Stainless Steel 316/PTFE/Tygon
Sample Pump Diaphragm pump (Corrosive resistant to H2S gas and weak H2SO4)
Complete with pump cut off function; Low flow & Internal CH4 Gas leakage alarm
Gas Cooler Type: Peltier gas cooler
Water trap remover Water trap filter system complete with auto drain
Flow and Pressure Control 0 10LPM Variable Area flow meter with pressure and flow regulation control;
complete with low flow alarm.
Sample line Standard 25 meters 3/8 PTFE sample tube
Sample probe 2 Stainless steel 304 probe complete with sinter filter, isolation valve, purging
/portable measurement port and over pressure pressure relief valve function.
(Stainless steel 316 as optional)
Panel View Status Indication & Contact Output
Warm up Warm up LED pilot light
Voltage free contact output (As optional)
System in operation Normal LED pilot light
Voltage free contact output (As optional)
Low flow Low flow LED pilot light
Voltage free contact output (As optional)
Pump stop Pump stop LED pilot light
Voltage free contact output (As optional)
Internal flammable gas leakage CH4 Gas leakage pilot light
Voltage free contact output (As optional)
Reset Pump reset push button
Safety Features
Electrical surge protection Industrial auto voltage regulator complete with surge arrestor (As optional)
Labeling Standard safety labeling reminder sticker
Panel Ventilation 2 units axial fan
(Continuous and internal alarm activation 2
nd
fan activation function)
Gas Detection Unit Safe area flammable gas detection unit
Panel Purge System Require if field site temperature >50C (As Optional)
Panel purge for pressurize zone 1 (Exp) criteria non certification (As optional)
Fire Extinguisher 1KG ABC dry powder fire extinguisher