Escolar Documentos
Profissional Documentos
Cultura Documentos
Lesson 14
BIOFEEDBACK
Relaxation and Arousal
Manual Revision
08232001.PL3.6.6-ML3.0.7
William McMullen
Vice President
BIOPAC Systems, Inc.
Faster
Heart
Rate
Slower
Higher
GSR
Arousal
Level
Lower
Page 2
I. INTRODUCTION
Recently, there has been a renewed interest in mind-body interactions. There are many processes in the
body in which you exert voluntary control. For example, if you want a glass of water, you move your
body and go through the motions to take out a glass, fill it with water, and drink. You are consciously
aware of the motions that are under your voluntary control. However, once you start to swallow the
drink of water, you are not usually aware of the regulatory processes which follow automatically (e.g.,
you secrete saliva, rhythmical constrictions in the esophagus move the water down to the stomach, the
stomach starts churning, etc).
These processes are regulated by the autonomic nervous system and do not require conscious control by
the cerebral cortex. Muscle movement to obtain a drink of water involves some voluntary controls
(wherein your brain and body interact in a loop between sensory reinforcement of movement and the
brain) but there is usually no loop of feedback between consciousness and the involuntary actions
regulated by the autonomic nervous system, e.g., actions of the gastrointestinal tract.
The autonomic nervous system has two regulatory divisions, which can affect the same organs or tissues
but exert contrasting effects:
sympathetic division short-term response to acute stress, fight-or-flight response
parasympathetic division daily routine maintenance of homeostasis
For example, both systems are constantly regulating heart rate, but when the parasympathetic system
dominates, the heart rate is lower than when the sympathetic nervous system dominates.
Essentially, biofeedback completes the loop between autonomic functions and conscious awareness.
Biofeedback training is a learning process whereby people exert conscious control over physiological
processes controlled by the autonomic nervous system. Biofeedback instruments unobtrusively monitor
physiological functions (e.g., heart rate) and provide feedback in real time. The equipment provides
feedback using a signal that changes with the monitored variable. The person can then use the signal to
enhance the desired response.
With biofeedback training, people have been able to regulate many processes: lower
heart rate; lower blood pressure, control headaches, and manage responses to
stressful situations. For example, biofeedback training has been shown to be effective
for controlling hypertension or high blood pressure.
One training method consists of a hand warming response that is, the
biofeedback signal is associated with the temperature of the hands. What does hand
warming have to do with high blood pressure? Well, blood pressure is a result of
cardiac output (volume output of the heart per minute) and peripheral resistance.
Peripheral resistance is inversely proportional to the amount of blood flow to the
periphery. Because blood is warmed in the body, when blood flow to the skin
increases, the skin is warmer. Therefore, warmer hands signify increased blood flow
that occurred when peripheral resistance and blood pressure decreased.
Biofeedback training has also been used to teach stress management techniques. In physiological terms,
relaxation using biofeedback training teaches people to activate specific controls of the parasympathetic
part of the autonomic nervous system, e.g., to lower heart rate. At the same time, biofeedback can be used
to decrease activity of the sympathetic nervous system.
The galvanic skin response (GSR) or electrodermal activity is one variable traditionally associated with
sympathetic nervous system activity. The galvanic skin response is affected by sweat gland activity and
skin responses on the palmar surface of the hand. Unlike the heart, the sweat glands are only activated by
Page 3
sympathetic activity. If the sympathetic branch of the autonomic nervous system is highly aroused, then
sweat gland activity increases and consequently, so does GSR. Because of this association, the GSR is
traditionally used as an index of sympathetic activity. When a person is relaxed, then the GSR should be
low.
GSR has been used as a measure of emotional and sympathetic response. For
biofeedback purposes, however, the relationship of GSR response as an index of
sympathetic activity is complex for many reasons. GSR responses are delayed
1-3 seconds. Responses are not consistent when given the same stimulus level.
External factors, e.g., temperature and humidity, affect GSR measurements,
adding to inconsistent results. Internal factors, e.g., medications, can change
GSR measurements. A final complication is that the neurotransmitter for palmar
sweat glands is acetylcholine, not norepinephrine as is common for most of the
sympathetic nervous system.
In this lesson, the biofeedback signal will be plotted on the screen as a thermometer style bar chart that
will rise and fall with changes in heart rate and GSR, allowing the Subject to become conscious of his/her
heart rate and arousal (GSR). The Subject will try to change the reading(s) without physical movements
and should be able to see that heart rate and level of arousal are independent.
III. MATERIALS
BIOPAC disposable vinyl electrodes (EL503), 3 electrodes per Subject
Electrode lead (SS2L)
Page 4
marker
marker tools
channel boxes
(Data analysis mode only)
vertical scales
marker label
vertical (amplitude)
scroll bar
channel labels
selection tool
zoom tool
I-Beam cursor
The symbols explained below are used throughout Experimental Methods and Analysis.
Key to Symbols
If you encounter a problem or need further explanation of a concept, refer to the
Orientation Chapter for more details.
The data collected in the step needs to be recorded in the Data Report (in the section
indicated by the alpha character). You can record the data individually by hand or choose
Edit > Journal > Paste measurements to paste the data to your journal for future
reference.
Most markers and labels are automatic. Markers appear at the top of the window as
inverted triangles. This symbol is used to indicate that you need to insert a marker and key
in a marker label similar to the text in quotes. You can insert and label the marker during or
after acquisition. On a Mac, press ESC and on a PC, press F9.
Each section is presented in a two-column format, as described below.
FAST TRACK STEPS
This side of the lesson (left, shaded
column) is the FAST TRACK through
the lesson, which contains a basic
explanation of each step.
A.
Page 5
SET UP
FAST TRACK Set Up
SS2L
Electrode Lead Set
SS3L
GSR
Transducer
Fig. 14.1
3. Turn the MP30/35 Data Acquisition
Unit ON.
4. Select a Subject (who has not
exercised within the last hour).
5. Place three electrodes on the Subject
(Fig. 14.2).
Subject must not have done any rigorous exercise within the
last hour.
one on right forearm
(just above wrist)
Set Up continues
Fig. 14.2
Place one electrode on the medial surface of the right
leg, just above the ankle bone
Page 6
right forearm
WHITE lead
right leg
left leg
BLACK lead
RED lead
(ground)
Fig. 14.3
Each of the pinch connectors on the end of the electrode
cable needs to be attached to a specific electrode. The
electrode cables are each a different color, and you should
follow Fig. 14.3 to ensure that you connect each cable to the
proper electrode.
The pinch connectors work like a small clothespin, but will
only latch onto the nipple of the electrode from one side of
the connector.
7. Fill the electrode cavities with
(GEL100) gel and then attach the
GSR transducer (SS3L) to the
Subjects fingertip (Fig. 14.4).
Note Electrodes should be in
place for at least five
minutes prior to calibration
or recording.
Set Up continues
You need to fill both the electrode cavities with electrode gel
before attaching to the fingertips.
HINT
For a good signal to be picked up, it helps if
the subjects have a little sweat on their
hands (not a lot, but enough so that their
hands are not completely smooth or cold). If
subjects wash their hands just prior to the
recording or if they have been sitting in a cold room, then
they must do something to activate the sweat glands before
beginning calibration or recording. If subjects begin with
colder hands, the scale will be diminished and the signal will
be easily saturated once they warm up during the lesson.
Page 7
Sensors attach to
bottom of fingertips
Velcro straps
wrap around
fingers
Fig. 14.4
The SS3L is typically placed on the index and middle fingers of
the left hand, with the sensor on the bottom of the fingertips
(the part without the fingernail). Wrap the Velcro tape around
the finger so the transducer fits snugly (Fig. 14.4).
8. Start the Biopac Student Lab Program
9. Choose Lesson 14 (L14-Biofbk-1).
END OF SET UP
Page 8
B.
CALIBRATION
The Calibration procedure establishes the hardwares internal parameters (such as gain, offset, and
scaling) and is critical for optimum performance. Pay close attention to the Calibration procedure.
FAST TRACK Calibration
Five seconds into the recording, the Subject should take one
deep inhale and exhale, and then return to normal breathing.
The program needs to see some variation in the GSR and BPM
data.
Note: The Subject should inhale and exhale without tensing
the chest muscles too much, as this may cause excessive
EMG artifact.
Figure 14.5
Both recording channels should show some fluctuation. If there
is not any fluctuation, then it is possible that a transducer is not
connected properly and you must redo calibration.
If the lower recording of ECG rate shows large fluctuations (20
BPM or more) during the interval from 4-10 seconds, then it is
possible that excessive EMG artifact during inhale and exhale
caused loss of BPM tracking and you should redo the
calibration. And have the Subject breathe less deeply.
END OF CALIBRATION
If you did not hear a beep during the calibration, then you
should make sure your computer volume is turned up and redo
calibration. Redo by clicking on the Redo Calibration button
and repeating the entire calibration sequence.
C.
Page 9
Faster
Heart
Rate
Slower
Higher
GSR
Arousal
Level
Lower
The recording will begin and will run continuously until you
click on the STOP button or the time indicated runs out.
Recording continues
Page 10
4. After 90 seconds, Subject should try to To move the bars up, you need to become more aroused.
move the bars up by increasing arousal
Recorder should insert a marker 90 seconds into the recording to
and Recorder should insert marker.
indicate that the Subject is now attempting arousal versus
relaxation. To insert Markers:
insert marker at 90 seconds
(No marker label needed.)
PC = F9 key.
Markers can also be entered after the data is recorded.
Hints for achieving arousal:
a) Think of a stressful situation.
b) Have the Director clap behind you.
c) Make a scowl with your face.
d) Hold your breath.
5. Click on Stop twice.
The first time you click on Stop, the data window containing the
Stop button will be activated.
When you click on Stop the second time, the bar chart data
display will be hidden and the data window will change to a
standard three-channel recording with GSR, Raw ECG and ECG
Rate. This data represents the entire recording.
If all went well, your data should look similar to Fig. 14.7 and
you can proceed to Step 7.
Note: The recording will vary greatly from person to person,
and it is hard to generalize about the results. Hopefully, you
were able to manipulate your physiological responses to some
degree. This may take some practice, as many people are not
used to thinking that they have some control over things like
their heart rate and nervous system.
Recording continues
Page 11
A pop-up window with four options will appear. Make your
choice, and continue as directed.
If choosing the Record from another subject option:
a) Attach the electrodes and transducer to the new Subject
per Set Up Steps 5 and 6, and continue the entire lesson
from Set Up Step 9.
b) Each the new Subject will need to use a unique file
name.
END OF RECORDING
Remove the electrode cable pinch connectors, and peel off the
electrodes. Throw out the electrodes (BIOPAC electrodes are not
reusable). Wash the electrode gel residue from the skin, using
soap and water. The electrodes may leave a slight ring on the
skin for a few hours. This is normal, and does not indicate that
anything is wrong.
Remove the GSR transducers and clean out the electrode gel
cavity using alcohol.
Note: GSR is a measure of electrical conductance and the units of the GSR recording are given
in mho. In 1885, Oliver Heaviside coined the term mho to represent the units of electrical
conductance. Electrical conductance is the reciprocal of electrical resistance, which is measured
in ohm (). Therefore, the mho unit was adopted by spelling ohm backwards and is represented
by an upside down Greek letter omega. The SI unit of electrical conductance is the siemens (S).
1 S = 1 mho = 1/
Page 12
V.
DATA ANALYSIS
FAST TRACK Data Analysis
Displays
CH 41
heart rate
CH 42
GSR
Fig 14.8
2. Turn OFF Channel 2, Raw ECG.
Autoscale horizontal
Autoscale waveforms
Zoom Tool
4. Set up the measurement boxes as
follows:
Channel
Measurement
CH 41
value
CH 42
value
A
Page 13
It takes a few heart rate cycles for the BPM data to be correct,
so measuring data 10 seconds into the record will be an
accurate baseline. Simply clicking the cursor at Time =10
seconds will update the value measurements.
Fig 14.10
6. Scroll to the point of maximal
relaxation.
Use the horizontal scroll bar to move through the data. Select a
point at the end of the first interval (before the marker)
representing the point of maximal relaxation.
Fig. 14.11
7. Take the measurements on this point as
needed.
A
8. Scroll to the point of maximal arousal.
Select a point at the end of the second interval (after the marker)
representing the point of maximal arousal.
A
10. Save or print the data file.
You may save the data to a floppy drive, save notes that are in
the journal, or print the data file.
END OF LESSON 14
Complete the Lesson 14 Data Report that follows.
Page 14
BIOFEEDBACK
Relaxation and Arousal
DATA REPORT
Computer Number:
Lab Section:
Date:
Height
Age
Weight
Gender:
Male / Female
A.
Table 14.1
II.
Measurement
Heart Rate
(BPM)
CH. #
CH 40
value
GSR
(MHO)
CH 41
value
Baseline
(10 sec.)
Condition
Maximal
Relaxation
Maximal
Arousal
Questions
B. Based on the data from Table 14.1, did the effects of the parasympathetic nervous system change with
biofeedback? Explain the physiological mechanisms causing the results.
Page 15
C. Name the branches of the autonomic nervous system and explain their function.
E. What change, if any, did your GSR recording show when you were aroused? Relaxed?
F.