A leader in the field of sensor instrumentation,
Eutech Instruments is a dynamic company
rapidly positioning itself globally as a
leading manufacturer of electrochemical
instrumentation for water analysis. A pioneer
in developing ASIC-based (Application
Specific Integrated Circuit) instruments,
Eutech Instruments has gained international
awards and recognition for its achievements
in sensor technology, software programming
and product design. Underlying its strong
commitment to Research & Development, is
the constant drive by Eutech engineers to apply
new, emerging technologies to the design and
manufacture of advanced sensor instruments.
Eutech offers a comprehensive range of
laboratory and field instrumentation for water
analysis and continuous online process
instruments for the monitor and control of pH,
Conductivity, Total Dissolved Solids (TDS),
Turbidity, Temperature, Ion Concentration,
Redox Potential (ORP) and Dissolved Oxygen
(DO). The Digi-Sense range of temperature
products, now manufactured by Eutech,
complements Eutechs product line-up in
electrochemistr y. Eutechs products are
certified to comply with global standards for
electromagnetic emission and interference.
Eutech Instruments is dedicated to producing
convenience products for Water analysis and
Thermometry.
Introduction
Digi-Sense
Thermocouple
Thermometers
Digi-Sense RTD
Thermometers
Digi-Sense
Thermistor
Thermometers
Digi-Sense
Handheld
Humidity
Dataloggers
Digi-Sense
12-Channel
Scanning
Thermometers
80
70
60
50
40
30
10 Digi-Sense
Temperature
Controllers
20
11 Temperature
Probes
10
13 Application Guide
1
Introduction
Digi-Sense thermometers deliver quick, reliable and accurate
temperature readings. Eutech Instruments manufactures a wide range
of portable hand-held meters with distinctive features for a wide
range of applications. These intrinsically safe UL-listed thermometers
meet the needs of chemical, laboratory, petrochemical and HVAC
specialty applications.
Temperature Introduction
Temperature Instrument Ranges
-273C
-460F
-156C
-18C
121C
260C
398C
537C
676C
815C
954C
1093C
1648C
250F
0F
250F
500F
750F
1000F
1250F
1500F
1750F
2000F
3000F
2204C 3037C
4000F
5500F
-183
Typical Accuracies
Platinum RTD
Probes : 0.2 to 0.35% of reading
Meter : 0.1% of reading and
1 C (1 F)
-40
Thermistor
-40
400 series
probes
500 series
probes
700 series
probes
Meters
225
70
: 0.15 C (0.27 F)
500 Series
: 0.1 C to 0.2 C
(0.2 to 0.4 F)
Type J, K and E
probes
Type T probes
Type R and S
probes
Temperature
F = (1.8 x C) +32
C = (F32) x 0.555
Kelvin = C + 273.2
Rankin = F + 459.67
-250
-250
752
Type J
184
1260
Type K
700
1000
752
1375
Thermocouple
-190.0
Thermocouple
Meters
Thermistor
Type T
871
Type E
899
Type R
1482
1766
Type S
1482
1766
Digi-Sense Thermocouple
Thermometers
Meter Specifications
Range
Resolution
Accuracy
Display
Power
Dimensions
Weight
: 0.5 kg
Ordering Information
60010-00 Type J Thermocouple Thermometer
60010-10 Type K Thermocouple Thermometer
60010-20 Type T Thermocouple Thermometer
Probe to be ordered separately. Refer to page 11 and 12 for selection.
Digi-Sense Thermocouple
Thermometers
Meter Specifications
Range
K
T
E
Resolution
Accuracy
Display
Power
Dimensions
Weight
: 0.5 kg
Ordering Information
60010-40
Digi-Sense Thermocouple
Thermometers
Meter Specifications
Range
Resolution
Accuracy
: Type J, K, T, E, and N
Below -150C (-238F): 0.25% of reading plus 1C (2F)
Above -150C (-238F): 0.1% of reading plus 0.4C (0.7F)
Type R, S, and B
0.1% of reading plus 1C (2F)
Display
Output
Power
Dimensions
Weight
: 0.5 kg
Ordering Information
60010-50
91100-52
91100-55
91100-85
RTD Thermometer
Automatic user
calibration for
improved meter
accuracy
Meter Specifications
Range
Resolution
Accuracy
: 2 C (4 F)
: From -200 to -100 C (-328 to -148 F)
0.2 C (0.4 F) : From -99.9 to 199.9 C (-148 to 392 F)
2 C (4 F)
: From 200 to 850 C (392 to 1562 F)
Alpha Coefficient
: 0.003850 //C
Display
Power
Dimensions
Weight
: 0.5 kg
Ordering Information
60010-80
RTD Thermometer
EC-TEM6TEM01P
Meter Specifications
Range
Resolution
: 0.1C/F
0.01C/F
0.1C/F
1 C/F
Accuracy
: 0.1C/F
Display
Output
Power
Dimensions
Weight
: 0.5 kg
Ordering Information
60010-85
EC-TEM6TEM01P
91100-52
91100-55
91100-85
Digi-Sense Thermistor
Thermometers
Meter Specifications
Thermistor Thermometer
Automatic calibration recognises
freezing point for improved meter
accuracy, minimises probe errors
This easy-to-use thermistor thermometer
features auto-power-off, selectable C/F
mode, low-battery indicator and HOLD
function; ideal for both laboratory and
field applications.
Accepts a large variety of YSI series 400
or series 500 thermistor probes with 6.35
mm phono plug.
Range
Resolution
: 0.1C/F
Accuracy
: 0.2C (0.4F)
Probes
Display
Power
Dimensions
Weight
: 0.5 kg
Ordering Information
60010-70
Thermistor Thermometer
Meter Specifications
Range
Resolution
Accuracy
: 0.03C (0.06F)
0.1C (0.1F)
0.2C(0.4F)
Display
Output
60010-75
Power
91100-52
91100-55
91100-85
: 0.5 kg
Ordering Information
Digi-Sense Handheld
Humidity Datalogger
Temperature/Humidity Datalogger
Retain up to 1000 data sets complete with date and time
Measure and display relative humidity, temperature and dewpoint
simultaneously
Transfer data to an infrared printer wirelessly or to a computer
via optional RS-232 adapter for permanent documentation
0 to 100% RH range - Use in a variety of environments
External probe allows accessibity to hard-to-reach areas
This rugged portable datalogger is suitable for applications where relative
humidity measurement is critical, such as laboratories, storage areas,
and research facilities. It measures and displays all 3 parameters - relative
humidity (RH), temperature, and dew point simultaneously on its easy-toread LCD.
Datalogger features minimum/maximum readings, easy data transfer
via infra-red or RS232, HOLD function, and low battery indicator.
Calibration salts are available for field calibration (order separately).
Meter Specifications
Humidity Range
Resolution
: 0 to 100% RH
: 0.1% RH
Accuracy
Temperature Range
Resolution
Accuracy
Response Time
: 15 seconds
Display
Output
Power
Dimensions
Weight
: 0.5 kg
Ordering Information
60020-52
60020-62
Replacement probe
37950-60
37950-61
91100-52
91100-55
91100-85
Meter Specifications
Range
J
K
T
E
N
R
S
B
Resolution
:
:
:
:
:
:
:
:
:
Accuracy
Display
Display Update Rate
Scan Rate
:
:
:
:
:
Dimensions
Shipping Weight
Power
MENU key lets you select high or low alarm set points for each
channel. Easily enable or disable built-in audible alarm and sets
alarm hysteresis; program for manual or automatic alarm reset.
Thermocouple inputs are electrically isolated from each other
and from data outputs provides safe and accurate measurements
even when probes are exposed to various electrical conditions.
Ordering Information
Cat no.
Description
69200-00
69202-30
69210-00
69212-30
12-Channel
Benchtop
12-Channel
Wall-mount
115
230
115
230
VAC
VAC
VAC
VAC
50/60
50/60
50/60
50/60
Hz
Hz
Hz
Hz
10
Temperature scale
Selectable to read in C, F, K (Kelvin), and R (Rankine)
Separate fuses
Sensor-calibrate function
Easy to connect
115 VAC models include 2m power cord and U.S. standard plug; 230 VAC
models feature an IEC socket power connector for interchangeable cord sets
(specify country of destination when ordering).
Loop-break alarm
Shuts off output if a broken sensor or heater is detected
Meter Specifications
Input type
Range
Resolution
J:
-190 to 1000C
(-310 to 1832F)
K:
-200 to 1372C
(-328 to 2502F)
T:
-200 to 400C
(-328 to 752F)
E:
-200 to 1000 C
(-328 to 1832F)
R:
0 to 1768C
(32 to 3214F)
S:
0 to 1768C
(32 to 3214F)
N:
-200 to 1300C
(-328 to 2372F)
B:
200 to 800C
(392 to 3272F)
100 Pt RTD *
Accuracy
Types J, K, T, E and N:
0.1;
1 above 999.9;
1 below -99.9
Ordering Information
Controller Type
Control output
68900-01
68900-03
Code No.
Standard
1150 Watts
2300 Watts
68900-11
68900-13
Deluxe
1752 Watts
3450 Watts
Ramp/Soak
Recorder Output
Power
Shipping Weight
No
None
3.2 kg
Yes; 9 programs
16 segments
4-20/20-4 mA,
selectable
4.1 kg
11
Temperature Probes
Thermocouple Type K Probes
Ordering Code
Dimension
Sensor Tip
2 mm
EC 3K220
Time
Constant
Class of
Precision
-60 to +600
-60 to +800
-60 to +800
-60 to +1100
6 mm -60 to +1100
5 mm -60 to +600
15
200 mm
Cable 1.5m
Temperature
C
liquid/air
3 mm
EC 3K320
200 mm
Cable 1.5m
3 mm
EC 3K350
500 mm
Cable 1.5m
6 mm
EC 3K6100
1000 mm
Cable 1.5m
EC 3K650
500 mm
Cable 1.5m
EC 3K520C
200 mm
Cable 1.5m
EC 3K1200
contact
-60 to +400
2000 mm
200 mm
Cable 1.5m
EC 3K415
3 mm -50 to +350
4 mm -50 to +350
-50 to +350
-50 to +350
-50 to +250
contact
150 mm
Cable 1.5m
contact
2 mm
EC 3T215
150 mm
Cable 1.5m
contact
5 mm
EC 3T520C
200 mm
Cable 1.5m
contact
4 mm
EC 3T420A
200 mm
contact
Cable 1.5m
EC 3T210
100 mm
Cable 1.5m
12
RTD Pt 100 Probes
Model
Dimension
Sensor Tip
Temperature
C
Time
constant
Class of
Precision
-50 to +400
30
5 mm
-50 to +400
120
4 mm
-50 to +400
60
4 mm
-50 to +250
60
3 mm
-50 to +400
30
1/5 DIN
-100 to +350
30
-50 to +400
60
-50 to +400
60
3 mm
EC PT56L
200 mm
Cable 1.5m
Liquid/Air
EC PT56C
200 mm
Cable 1.5m
Contact
EC PT56P
500 mm
Cable 1.5m
Penetration
EC PT56A
1000 mm
Cable 1.5m
Air
EC PT56L 1/5DIN
500 mm
Cable 1.5m
4 mm
EC PT56 TFE
100 mm
Cable 1.5m
contact
6 mm
EC PT56/500
500 mm
Cable 1.5m
6 mm
EC PT56/1000
1000 mm
Cable 1.5m
Sensor
Temperature Range
Class
Tolerance
Themocouple
Type K
-60...+1200C
Type T
-40...+350C
Type J
-40...+750C
Pt 100
-100C
+0C
+100C
+200C
Class A
0.35C
0.15C
0.35C
0.55C
Pt 100
-100C
+0C
+100C
+200C
Class 1/5
0.16C
0.06C
0.16C
0.26C
Time Secs
Specification
/T
couple K
Thermo
% of final value
13
Application Guide
Temperature Measurement
Temperature is the most commonly measured parameter. To obtain an
accuracy better than 0.2C (0.4F), extreme care is needed. Quite
often errors can occur due to presence of temperature gradients or
drafts, sensor non-linearities, poor thermal contact, calibration drifts,
radiant energy and sensor self-heating.
Sensor Type
Output
Range C
Accuracy C
Robustness
Cost
Thermocouple
Platinum RTD
Nickel RTD
Thermistor
Semiconductor
Non-contact
Fiber Optic
Cryogenic
Bimetallic
40 V/C
0.4%/C
0.4%/C
5%/C
10 mV/C or 1 A/C
Millivolts
Various
Various
Hi/Low
-270 to 2300
-200 to 600
-200 to 600
-50 to 200
-40 to 125
0 to 6000+
-100 to 200
-273.15 to -200
-100 to 200
1.5
0.2
0.3
0.2
1.5
2
1
Various, to 0.001
2
High
Medium
Medium
High
Medium
Low
Medium
Various
High
Low
Medium
Low
Medium
Low
High
Very High
Various
Low
The above table provides a general summary and should not be taken as definitive statement on temperature sensors, especially on the accuracy.
Generally the accuracy of all sensor types can be greatly improved by calibration.
Temperature Sensors
Thermocouple sensors consist of a measuring junction formed by
two dissimiliar wires that complete an electrical circuit to the meter. A
millivolt signal is generated proportional to the temperature gradient
between the measuring and reference junctions. Thermocouples can
be used in a wide variety of applications and offer a broad measuring
range.
RTD sensors (resistance-temperature detector) use an element with
RTD
Thermistor
100
Platinum thin-film
400
Two-wire, single
ceramic element
500
Miniature two-wire,
single ceramic
element
Three-wire,
dual ceramic
element
+ / white / Iron
/ red / Constantan
+ / yellow / Chromel
/ red / Alumel
700
Thermocouple
(T/C)
J
K
+ / blue / Copper
/red / Constantan
+ / purple / Chromel
/ red /Constantan
+ / orange /
87% Platinum,
13% Rhodium
/ black / Platinum
+ / red /
87% Platinum,
13% Rhodium
/ red / Platinum
+ / red /
90% Platinum,
10% Rhodium
/ black / Platinum
Features
Applications
Max. Range
0.1%
of reading
-200 to 849C
-330 to 1560F
-40 to 150C
-40 to 302F
-40 to 150C
-40 to 302F
0 to 1000C
-328 to 1832F
-250 to 1375C
-418 to 2507F
-250 to 400C
-418 to 752F
-250 to 1000C
-118 to 1832F
-250 to 1300C
-400 to 2372F
0 to 1450C
32 to 1450F
0 to 1450C
32 to 2642F
Traditional Sensor
Agriculture Research
Automotive
Thermistor, Pt100
Pt100
Cryogenics
Environmental Research
General Industry
Pt100
Hobby, Education
HVAC
Thermistor, Pt100
In Manufactured Goods
Metallurgy
Error vs Temperature
for a Typical Thermocouple System
8
System = Probe & Meter
6
Error C
To determine your system accuracy when you have a separate meter and probe, add the
tolerance of the meter to the tolerance of the probe. The diagram (on the right) shows the
accuracy versus temperature for a typical thermocouple meter and probe. For example,
using the diagram, the meter error at 250C is 0.5C and the probe error at 250C is
1.0C, thus the system accuracy would be 1.5C or better. The result is often
conservative the actual system accuracy is usually better. Glass thermometers, labels,
paints, bimetal thermometers, handheld infra-red thermometers, and digital indicators
include a general system accuracy statement.
Probe
Meter
1250
1000
750
500
250
Warranty
-250
Temperature C
Eutech Instruments warrants its handheld meters free from manufacturing defects for 3
years, scanning thermometers and temperature controllers for 1 year and probes for 6
months.
Note: We reserve the right to make changes, improvements and modifications to products shown.
www.digi-sense.com
marketing@eutechinst.com
Singapore USA Netherlands Malaysia China
55, Ayer Rajah Crescent, #04-16/24, Singapore 139949.
Tel: (65) 6778 6876 Fax: (65) 6773 0836
YSI is a registered trademark of YSI Inc. & Windows is a registered trademark of Microsoft Corporation