Operating Manual
Einstein-II DGS
Einstein-II DGS
Operating Manual
Einstein-II DGS
CONTENTS
Einstein-II DGS
MODSONIC
Einstein-II DGS
--1-MODSONIC
MODSONIC
measurement. This principle utilizes the precise timing of the transit time of a short burst of
ultrasound energy, through a material under test. The ultrasound waves travel to the far side of the
test piece and reflect back to the transducer/probe and a measurement is obtained.
Basically three typed of transducers/probes are available for different types of applications:
1> Straight beam Probe (Normal probe)
2> TR probe (Dual crystal probe)
3> Angle beam Probe
1> Straight beam Probe: This probe introduces ultrasound normal to the test piece surface
utilizing longitudinal or compression waves. Normal beam probe is used mostly for flaw
detection and thickness gauging
(refer figure 1.)
2> TR probe: This probe contains separate transmitting
and receiving elements as showing in figure 2
usually mounted on delay lines. This design
improves near surface resolution by separating
the initial pulse from the received echoes. TR
probe is suitable for the thickness gauging of pitting
and corrosion and also for better surface resolution.
Einstein-II DGS
--3--
MODSONIC
3> Angle Beam Probe: This probe introduces ultrasound at an angle to the surface of the test
piece. In most angle beam probes,
the wave energy is mode converted from a
longitudinal wave to a shear wave by the
refraction principle. This probe is suitable for
inspection of the welds. The reason for this is
its ability to position the transducer/probe
away from the weld bead yet propagate energy
into the weld zone. Another reason to use angle
beam testing on the welds is to position the
sound beam more normal to the expected discontinuities since the flaws in welds are usually
perpendicular to the test piece (except porosity). Figure 3 shows the principle of angled beam
weld testing.
Einstein-II DGS
--4--
MODSONIC
17
19
18
15
14
13
12
3
11
10
9
1
Einstein-II DGS
21
20
--5--
MODSONIC
1. POWER key : Einstein-II DGS can be switched ON and OFF using this key.
2. ZOOM key : By pressing this key A-Scan spreads over to full screen. Menu
parameters disappear from LCD and only A-Scan is displayed. By
pressing any other key, A-Scan is restored to normal size and
menu parameters appear again .
3. FREEZE key : By pressing this key once A-Scan can freeze. By pressing second
time selects Peak Freeze mode. In this mode it hold echo dynamic
pattern which is helpful for a angle beam inspection to locate peak
signal.
4. LOCK key : By pressing this key all function (keys) get locked which prevents
accidental pressing of any key. To unlock, simultaneously press
Step key (9) and LOCK key (4).
5. LCD Screen : It displays A-Scan Trace pattern and Menu Parameters.
6. Battery Status : When all the three cells are displayed empty, then the battery
Indicator
needs to be recharged.
7. Gain Step
Einstein-II DGS
: This location displays Gain Step value. This value can be changed
by using dB STEP(10) key. It follows 0.5, 1, 2, 6, 12 and 20
dB Step sequence.
--6--
MODSONIC
8. Menu keys
9. Hot key
10. dB STEP key : Any Gain step out of 0.5, 1, 2, 6, 12 or 20 dB can be selected
using this key.
11. GAIN UP/DN : Gain can be incremented/decremented using these keys in the step
selected by dB STEP key.
12. UP/DN key
13. REJECT
LED
Einstein-II DGS
--7--
MODSONIC
15. Parameter
Values
16. RX Connector
17. TX Connector
Einstein-II DGS
--8--
MODSONIC
4 Operation detail
Einstein-II DGS is user friendly, menu driven Ultrasonic Flaw Detector. To access any
parameter, various menus are available. Details of all menus and parameters are as per below.
First menu is Basic menu. In this menu, Basic functions of an
Ultrasonic Flaw Detector (UFD) like Range, Fine Range,
Delay and Gain functions are covered.
ZERO
ZERO
: It is to calibrate Artificial zero of the probe.
2200
RANGE : The required test range can be set directly in terms of mm/
RANGE
inch. The range available for steel is from 10 to 5000 mm
0100 MM
(0.50 to 200 inch) it steps through 10,25,50, 100, 125,200,
VEL M/S
300,500,1000,2000,3000,4000 and 5000 mm with pressing
5920
hot key or in step of 1mm (0.5,1,2,5, 10, 15,20,25,30,50,100
150
and 200 inch) or in step of 0.2 inch).
DELAY
000.0MM VELOCITY : The required velocity can be set directly in terms of
met./Sec (inch/sec). By pressing hot key it steps through
GAIN 1
preset value. In fine mode 1 met./sec. (0.0001 in/us).
25.0 dB
DELAY
: Partial Range can be set using these keys. Exact shift/delay
in mm (Inch) is displayed at the bottom.
Einstein-II DGS
--9--
MODSONIC
GAIN STEP: Gain step value can be set by using GAIN STEP key, it follows dB in
steps of 0.5, 1, 2, 6, 12 or 20 dB.
GAIN
: Gain can be increased/decreased as per value of step by using GAIN
UP/DN keys.
Example: Einstein-II DGS is to be calibrated for 100 mm Test Range using Single
crystal (Normal) probe.
> Set required range to 100 mm using RANGE UP/DN keys.
> Couple the probe on 25mm thickness of V1 reference block and set first echo at 25mm
using ZERO UP/DN keys.
> Set 4th echo at 100 mm using VELOCITY UP/DN keys, if required.
> Repeat above two steps until both the echoes are at their respective positions.
> Amplitudes of the echoes can be varied using GAIN keys.
Einstein-II DGS
--10--
MODSONIC
Second Menu covers MONITOR GATE 1 functions.
GATE 1 : Gate can be changed to PLOGIC, NLOGIC, DAC, DAC 6,
GATE 1
DAC-14 trigger, EXPAND or OFF.
OFF
PLOGIC : When PLOGIC is selected, the MONITOR is
START
triggered if an echo is present within the gate and its
0030.0
height is equal to or crosses the set Threshold Level.
END
NLOGIC : When NLOGIC is selected, the MONITOR is
triggered if there is no echo within the gate or the
0100.0
height is lower than the Threshold Level.
LEVEL
DAC
: If DAC is turned ON and DAC is selected and if
040 %
any echo signal equals or crosses DAC curve then
monitor gates trigger.
GAIN 1
DAC-6 : If DAC is made ON 6/14 and DAC-6 is selected then if
25.0 dB
any echo signal cross DAC-6 dB curve then monitor
gates trigger.
DAC-14 : If DAC is made ON 6/14 and DAC-14 is selected then if echo
signal equals or crosses DAC-14 dB curve then monitor
gates trigger.
EXPAND: When EXPAND is selected then the Scale expansion is activated and
the GATE width is expanded to Full-screen width for a magnified view.
Einstein-II DGS
--11--
MODSONIC
START : Gate starting position can be set using START UP/DN keys. Using these
keys Gate start can be directly set in terms of mm/inch.
END
: Gate End position can be set using END UP/DN keys. Using these keys Gate
end can be directly set in terms of mm/inch.
LEVEL : Gate Threshold Level can be directly set in terms of the percentage of the
Full Scale Height (FSH).
Example: The MONITOR to be triggered when any echo crosses the 50% FSH
between 35mm and 80mm.
> In the Gate setting, select PLOGIC using GATE UP/DN keys.
> Set Gate START to 35mm using START UP/DN keys.
> Set Gate END to 80mm using END UP/DN keys.
> Set Gate LEVEL to 50% using LEVEL UP/DN keys.
Einstein-II DGS
--12--
MODSONIC
Third Menu covers MONITOR GATE 2 functions.
GATE 2
OFF
START
0040.0
END
0100.0
LEVEL
050 %
GAIN 1
25.0 dB
END
LEVEL
Einstein-II DGS
--13--
MODSONIC
Forth Menu covers Memory storage/recall and Trace printout functions.
MEMORY
A SCAN
MEM NO*
001
ACTION
RECALL
NOTE
ENTER
GAIN 1
25.0 dB
Einstein-II DGS
--14--
MODSONIC
OPERATOR:
ABC
JOB DETAIL:
NIL
DEFECT DETAIL:
NIL
REMARK:
NIL
CURSOR
MOVE
CURSOR
UP/DN
CHAR
SCROLL
NOTE
EXIT
GAIN 1
25.0 dB
--15--
MODSONIC
> Set ACTION to SAVE using ACTION UP/DN keys.
> Then press NOTE/ENTER Left key so Note menu will be displayed. If required
then edit it and exit from note menu.
> Then press NOTE/ENTER Right key to store A Scan. By pressing Right key
Current Calibration, A Scan pattern and Note DATA will be stored UP/DN key to
store the current calibration set-up at memory no. 3.
Note: If data is already stored at this location, then the message will be displayed on LCD
screen as OVER WRITE DATA YES - PRESS HOT KEY,
NO PRESS ANY KEY. So press HOT KEY to overwrite data.
Einstein-II DGS
--16--
MODSONIC
Fifth Menu covers Rejection of unwanted signal, Set Reference, Video setting and
Testing mode selection function.
REJECT
000 %
POWER
LOW
DAMPING
LOW
REJECT
MODE
SINGLE
GAIN 1
25.0 dB
MODE
Einstein-II DGS
: Test Mode for the use of Single or Dual Crystal probes can
be selected using MODE UP/DN keys.
--17--
MODSONIC
Sixth menu covers the required functions for DAC curve AWS or DGS curve as per selected in
menu 10 CONFIG parameter.
DAC
DAC
: DAC setting can be switched to DRAW, ON ,ON 6/14, TCG
ON or OFF using DAC UP/DN keys. DRAW mode, peak points
OFF
of echo signals are selected.
STRT G1
ON mode :The selected peak points are digitally connected to form a
030%
DAC curve.
POINT
ON 6/14 mode: -6dB DAC and 14dB DAC will be also displayed.
001
TCG ON: Time Gain Correction activated as per DAC so it brings echo
PRESS
signal to same height irrespective of depth.
ENTER
OFF mode :DAC curve will be not displayed on LCD.
GAIN 1
STRT G1 :Start of the GATE 1 can be set using STRT G1 UP/DN
keys. To select any peak of the echo signal, move the
25.0 dB
gate on the echo signal and then press ENTER UP/DN
key. POINT value is automatically incremented to indicate
next point.(If GATE 1 is OFF then turn to PLOGIC).
POINT : It indicates how many points are selected to draw DAC
curve.
PRESS
: Press ENTER UP/DN keys to accept the peak points
selected by the GATE 1.
Einstein-II DGS
--18--
MODSONIC
PRESS/ENTER : Parameter will be changed to T-CORR when DAC is On or TCG is
ON. Gain of machine can be change without curve height. It can be
used for Transfer/Scanning gain correction.
Example: To digitally draw DAC curve to join four peaks of the echo signals using 25mm
thick reference block and range of 100mm.
> Set Gain so that all four peaks are visible on the LCD screen above 5% FSH.
> Set DAC to DRAW using DAC UP/DN keys.
> Move cursor using STRT G1 UP/DN keys to cover the first echo. If GATE 1 is
OFF then switch to PLOGIC.
> Press ENTER UP/DN key to accept it and Point No. is incremented automatically.
> Repeat above two steps for second, third and forth echo.
> Set DAC to ON using DAC UP/DN key. Now DAC Curve will be displayed on
LCD.
Note: Cursor need not touch the echo for selection and acceptance of the peak point.
Einstein-II DGS
--19--
MODSONIC
AWS Function functions will be displayed as per selected in menu 10 CONFIG parameter.
: AWS setting can be switched to SET, MEASURE or OFF
using AWS UP/DN keys. SET mode, Make GATE 1 to
PLOGIC then Bring reference echo of reflector in between 40 to
60% , Press ENTER parameter key so gain of reference echo is
recorded and displayed as REF GAIN. (RG)
MEASURE mode : Display IL (Indication level) dB value to bring echo
to reference level.
AF
: Display attenuation factor to correct for the depth of the echo
signal.
IR
: Display indication rating as per AWS code. IR=IL-RG-AF.
Example:
GAIN 1
> Select MEASURE to ON.
25.0 dB
> Set GATE 1 to PLOGIC
> Couple probe on reference defect and set gain value so echo height within
40 to 60% of Screen.
> Set GATE 1 in such a way so echo signal within GATE.
> Then press SET REF /ENTER key so required gain to bring echo to 50% is
recorded and displayed as a REF Gain in GAIN box.
> Set AWS to MEASURE so it starts to display Indicating rating as per AWS
code.
Einstein-II DGS
--20-AWS
OFF
STRT G1
030%
AWS
MODSONIC
Indication rating is based below formula.
IR=IL-RG-AF
IL : Indicating Level its gain value to bring echo height to reference level (50%).
RG : Reference Gain.. Its Gain value to bring reference echo height to reference
level(50%)
AF : Attenuation Factor its attenuation of signal in material as per AWS standard.
Based on below formula.
AF= (Sound path (in Inch) -1) * 2
IR : Indicating Rating Its different in dB between the indication signal and reference signal with
attenuation correction.
Einstein-II DGS
--21--
MODSONIC
DGS functions will be displayed as per selected in menu 10 CONFIG parameter.
DGS
OFF
STRT G1
030%
DGS-REF
OFF
TL-CORR
00.0dB
GAIN 1
25.0 dB
DGS
Einstein-II DGS
--22--
MODSONIC
DGS Selection Menu: When DGS is selected to SELECT DGS selection menu will
be displayed on Screen. By pressing
SCAL NO
D EFFT
DGS
Menu UP/DN keys arrow( ) shifts.
002
23.1 MM EXIT
SELECT
It indicates currently selected sub
PRBNAME
DGS-CRV
ATT-REF
STRT G1
menu. For eg. To change SCAL NO
M2SN
11.0 MM 00.0 dB
030%
move arrow to first bottom empty box
then press corresponding UP/DN keys.
PRBFREQ
REFECHO
ATT-OBJ
DGS-REF
2.00MHZ
FBH
00.0 dB
OFF
DEL-VEL
REFSIZE
AMPCOR
TL-CORR
2500M/S
11.0 MM
00.0 dB
00.0 dB
SCAL NO
GAIN
25.0 dB
: It scrolls from 0 to 18. Select 0 for custom probe which one is not listed
between 1 to 18. For custom probe select appropriate probe frequency and
probe effective dia. For predefined probe, 1 to 18, it is not required.
PRBNAME : It displays Model name of predefined probe.( for scale no 1 to 18 but for
scale 0 it will display CUSTOM). With each Memory set one Custom
probe parameters will be stored.
Einstein-II DGS
--23--
MODSONIC
PRBFREQ
DEL-VEL
D-EFFT
DGS-CRV
ATT-REF
ATT-OBJ
AMPLCOR :This is gain correction. When Angle probe is used and reference echo is
taken from 100 mm or 25 mm arc of V1/V2 reference block.
Example: To plot DGS for 11 mm size equivalent reflector for M2SN probe.
> Switch GATE 1 to PLOGIC. (NLOGIC will also work).
> Select DGS in menu 10 under CONFIG parameter. It will not turn to DGS if DAC is
selected to OFF.
> Go to DGS menu 6.
Einstein-II DGS
--24--
MODSONIC
> Couple M2SN probe on 25 thick V1 block and calibrate for required range.
> Bring first echo within screen with height between 40 to 80 % FSH preferable.
> Set first back-wall echo signal within GATE 1 using STRT G1.
> Change DGS to SELECT by pressing DSG UP/DN keys.
> Move arrow below first sub menu by pressing Menu UP/DN Keys.
> Select SCAL NO to 2 by pressing corresponding UP/DN keys.
> Move arrow below second sub menu by pressing Menu UP/DN keys.
> Select DGS-CRV to 11.0 MM by pressing corresponding UP/DN keys.
> Select REFECHO to BW by pressing corresponding UP/DN keys.
>ATT-REF and ATT-OBJ to be set as per appropriate value. Here it is set to 0.
> For exit press EXIT UP/DN keys.
>Select SET-REF to ON so reference echo height is registered and brought to reference
curve. (make sure reference echo is within GATE 1 with 40 to 80% FSH preferable).
> Now press DGS UP/DN key so it turns to ON and appropriate DGS curve will be
displayed.
--25--
Einstein-II DGS
MODSONIC
Seventh menu covers the required functions for Thickness/Distance measurements.
MEASURE : MEASURE setting can be switched to ON or OFF using
MEASURE
MEASURE UP/DN keys. In ON mode it will display
thickness/distance
of echo signal which is in-side the gate.
OFF
In OFF mode it will not display thickness/distance.
X-OFF
X-OFF
: Set X-OFF value to same as Angle probe index point to
000.0
Front face distance so Surface Distance will be displayed
ANGLE
on LCD will be distance from angle probe front face to
000 DEG
defect position.
THICK
100.0MM
ANGLE
THICK
GAIN 1
25.0 dB
Einstein-II DGS
--26--
MODSONIC
Example: To measure Beam path, Surface and Depth distance when 45 deg. probe is
used.
> Go to GATE 1 menu and make Gate to PLOGIC/NLOGIC.
> Set Gate START and Gate END value in such a way so defect signal within the GATE.
> Set Gain or Threshold so height of the echo/defect beyond the GATE.
> Go to measure menu and set MEASURE to ON.
> Set angle value to 45 DEG. using angle UP/DN keys.
> Set appropriate THICK value so it can account multiple skips for DEPTH distance
calculation.
> Go to menu 11 and set MEA PO1 to PD, MEA PO2 to DP and MEA PO3 to SD so
Sound path, Depth and Surface distance will be displayed on screen.
--27--
Einstein-II DGS
MODSONIC
Eighth menu covers the required functions for Thickness/Distance measurements.
TRIGGER
FLANK
HUD
OFF
SET REF
OFF
UNIT
INCHES
SET REF
GAIN 1
25.0 dB
UNIT
Einstein-II DGS
--28--
MODSONIC
Ninth Menu covers miscellaneous functions, which are described as below.
CLOCK
OFF
COLOR
005
VIDEO
ENVELOP
CLOCK
COLOR
VIDEO
BRIGHT
050 %
GAIN 1
25.0 dB
--29--
Einstein-II DGS
MODSONIC
Tenth menu covers the required functions for HORN and Key Beep Setting.
CONFIG : Evaluation mode can be selected to DAC,AWSor DGS.As per
selection, menu 6 will turns to DAC, AWS or DGS setting.
This selection can change only if current active
DAC/AWS/DGS is selected to OFF.
CONFIG
DAC
HORN
OFF
BEEP
ON
PRF
HI
GAIN 1
25.0 dB
HORN
BEEP
PRF
: Using PRF UP/DB keys HI or LO PRF can be selected.
In HI PRF is depends on current range selection. In Low
mode is PRF is low which is useful when ghost echo
pattern appears when testing low attenuate material.
Einstein-II DGS
--30--
MODSONIC
Eleventh menu covers selection of measured parameter to be displayed on screen.
MEA PO1 :Measured value will be displayed on screen at Left most
position below base line. It scrolls as per detail given below.
MEA PO1
OFF
Display OFF
SND PTH
Sound path of Echo signal
OFF
DEPTH
Depth of Echo signal
MEA PO2
SURF D
Surface dist. of Echo signal
OFF
ECH HI%
Percentage height of Echo signal
MEA PO3
DAC %
Percentage height with respect to DAC curve.
(When DAC is selected to ON)
OFF
ERS VAL
Equivalent Reflector Size of echo signal
(When DGS is selected to ON)
DGS CRV
Display currently selected DGS curve size.
DGS HDB
dB diff between DGS curve and echo signal
GAIN 1
(When DGS is selected to ON)
25.0 dB
Signal to be within GATE 1 and MEASURE is selected to ON in menu
7 then appropriate measured value will be displayed onscreen.
MEA PO2 : Measured value will be displayed on screen at Middle position below
base line. It scrolls as per detail given in MEA PO1.
MEA PO3 : Measured value will be displayed on screen at Right side below base line.
It scrolls as per detail given in MEA PO1.
Einstein-II DGS
--31--
MODSONIC
5 Setting of time Keep key 1 press and switch ON Einstein-II DGS. After display
of menu Parameter, release key 1 and again press key 1 so Time setting menu will be
displayed. Now set required Time/Data by pressing
1
UP/DN parameter key. Then press key 2 for to
store/set current time to value set as per display.
Press key 3 to exit without store/set current time.
Factory Default Calibration set-up
In Einstein-II DGS last calibration set-up remain
store So when further Einstein-II DGS Starts it
will start with last setting, But due to some reason
if stored calibration is corrupted then it will starts
with factory default calibration set-up. If any user
want to recall factory default calibration set-up
then press key 4, keep it press and Switch on
Einstein-II DGS. After displaying menu parameter
release key 4.
3
2
Einstein-II DGS
--32--
MODSONIC
6 Interface of Printer
Einstein-II DGS has direct printout facility of A-Scan pattern with full calibration set-up values.
It can be printed out on conventional PC printer having a Serial Communication connector. So
computer or special printer is not required. Printer should be IBM-PC compatible with serial
port. Serial port setting is as below. Serial printer EPSON FX/LQ type recommended. (To set
below setting in serial printer refer operating manual supplied with the printer).
Baud Rate 9600
Parity bit None
Data Bits 8
Stop Bits 1
To print out trace pattern, follow the below given steps.
> Connect Serial cable (supplied with Einstein-II DGS) between connector (19) of
Einstein-II DGS and serial printer.
> Switch ON Serial PC printer.
> Switch ON Einstein-II DGS.
> Get the required Trace pattern and freeze using Freeze key (3).
> Using Menu key go to memory menu (4).
> Using ACTION UP/DN key select PRINT.
> Press ENTER UP/DN key to initiate printing.
For Interface with PC Using Serial port Refer Help.txt file supplied in Software
CD.
Einstein-II DGS
--33--
7 Power Supply
MODSONIC
Einstein-II DGS is powered by a Lithium-Ion battery pack ATB1, which is fitted in Einstein-II
DGS using two screws. It electrically connects to Einstein-II DGS via two internal terminal
contact points. With fully charged battery pack ATB 1, Einstein-II DGS can operate
continuously for eight hours, under typical operating conditions. The battery pack ATB 1 can be
charged when inside Einstein-II DGS by connecting Battery charger ATC 1 to the connector
(18). Alternatively, the battery pack ATB 1 can be charged outside Einstein-II DGS which
allows uninterrupted use of Einstein-II DGS using multiple battery packs.
Battery Charging
The battery Charger ATC 1 supplied with Einstein-II DGS is a special quick- charge charger.
The charging time is just 5 to 6 hours. ATC 1 battery charger is suitable for use with mains
supply from 100 to 240 volts AC/50 Hz. The LED on charger illuminates to indicate power on.
When the battery voltage is low (discharge condition) the LED glows RED. When the battery is
almost charged the LED turns to yellow. When the battery is full charged the LED glows green.
The output voltage from the charger is 16.4 Volts DC. The charging current is constant at 2.0
Amps.
Einstein-II DGS
--34--
MODSONIC
8 Precaution
> Avoid shock to the Einstein-II DGS and probes.
> Wipe the couplant off the probe and Einstein-II DGS after use.
> When battery pack is out-of Einstein-II DGS, keep battery contacts away from
metal object
> Never heat the battery pack ATB 1 nor throw into a fire.
Einstein-II DGS
--35--
Velocity
Delay
Gain
Rejection
Rectification
Frequency
Test Modes
Transmitter
Damping
Connectors
Einstein-II DGS
MODSONIC
MODSONIC
Freeze/Peak Freeze
Connectors
Monitor
Gate Expand
A-Scan memory
Calibration Set-up
Software
Printer Attachment
Display
Full Screen
Einstein-II DGS
MODSONIC
Reference A-Scan
DAC
TCG
AWS
DGS
Digital Readout
Measurement Unit
Einstein-II DGS
MODSONIC
Video Output
PRF
Update Rate
Power
Charger
Temperature
Size
Weight
--39--
Einstein-II DGS
MODSONIC
10 Menu Tree
First Menu
Second Menu
Third Menu
Forth Menu
ZERO
GATE 1
OFF
PLOGIC
NLOGIC
DAC
DAC 6
DAC 14
EXPAND
GATE 2
OFF
PLOGIC
NLOGIC
MEMORY
SET-UP
RANGE
START
VEL
END
DELAY
LEVEL
START
END
LEVEL
MEM NO
ACTION
RECALL
SAVE
DELETE
RECALL
DETAIL
NOTE
ENTER
A SCAN
Einstein-II DGS
--40--
MODSONIC
Fifth Menu
REJECT
POWER
LOW
HI
STRT G1
DELETE
PRINT
SEND
SAVE
DAMPING
LOW
HI
POINT
START G1
START G1
START G1
STRT G1
SET REF
IL ---- AF---DGS-REF
RATING
TL-CORR
X-OFF
ANGLE
THICK
MODE
SINGLE
DUAL
PRESS
ENTER
--41--
MODSONIC
Eighth Menu
Ninth Menu
Tenth Menu
Eleventh Menu
Einstein-II DGS
OFF
ON
TRIGGER
FLANK
PEAK
CLOCK
OFF
ON
CONFIG
DAC
DGS
MEA PO1
OFF
SND PTH
HUD
SND PTH
DEPTH
SURF D
ECH HI%
DAC %
ERS VAL
DGS CRV
DGS HDB
ALL
COLOR
HORN
OFF
ON
MEA PO2
OFF
SND PTH
--42--
SET REF
OFF
ON
UNIT
INCHES
METRIC
VIDEO
ENVELOP
FILLED
BEEP
OFF
ON
MEA PO3
OFF
SND PTH
BRIGHT
PRF
HI
LOW
MODSONIC
DEPTH
SURF D
ECH HT%
DAC %
ERS VAL
DGS CRV
DGS HDB
Einstein-II DGS
DEPTH
SURF D
ECH HT%
DAC %
ERS VAL
DGS CRV
DGS HDB
DEPTH
SURF D
ECH HT%
DAC %
ERS VAL
DGS CRV
DGS HDB
--43--
MODSONIC
All possible care has been taken in the preparation of this operation manual. If you find
any errors or would like to make suggestions for improvement to this operation manual,
then please write to:
OP22 (1)
Einstein-II DGS