Você está na página 1de 24

Mach-I

Operating Manual

Einstein-II DGS

Einstein-II DGS
Operating Manual

Einstein-II DGS

CONTENTS

1> Introducing Einstein-II DGS ... 1


2> How does an Ultrasonic Flaw Detector work.2
3> Parts and Controls of Einstein-II DGS .. 5
4

4> Operating detail .. 9


5> Time/Date setting, Factory Default Setting 32
6> Interface of Printer ..33
7> Power Supply 34
8> Precaution ..35
9> Specification of Einstein-II DGS .36
10> Menu Tree ....40

Einstein-II DGS

MODSONIC

1 Introducing Einstein-II DGS


Einstein-II DGS is a portable, compact, light in weight and user friendly Digital
Ultrasonic Flaw Detector. Einstein-II DGS has COLOR LCD display, it provides wide
viewing angle and allows fast scanning speed. Fully charged battery gives continuous
working of eight hours and charging time is just 5 to 6 hours. Trace pattern can be
directly printed out on conventional PC printer having serial port for interface or can be
transferred to PC via RS232 Serial port for storage of data.

Einstein-II DGS
--1-MODSONIC

2 How does an Ultrasonic Flaw detector Work


Einstein-II DGS is a single channel Ultrasonic-testing instrument used for the inspection
of homogeneous material for the presence of inclusions, porosity and other
discontinuities that could affect the performance of material and components. It can also
be used for thickness gauging of homogeneous material, requiring access from only one
side of the test piece. High frequency sound (Ultrasonic sound) waves are introduced into
the test material/part from a transducer/probe that is usually coupled to the test piece by
water or other suitable coupling liquid. The transducer converts electric signals to
Ultrasound and vice versa. A short burst of Ultrasound is introduced into test material so
some or all of the energy is reflected by
discontinuities. The reflection of the ultrasound
energy is a function of the ratio between the
acoustic impedance of the discontinuity and the
base material. The greater the impedance ratio
the more sound energy will be reflected. The
principle of Ultrasonic testing is shown in fig. 1.
It shows the ultrasonic energy generated in the
test piece and resultant instrument display.
Thickness gauging with Einstein-II DGS
Operates on the principle of the time-of-flight
--2-Einstein-II DGS

MODSONIC
measurement. This principle utilizes the precise timing of the transit time of a short burst of
ultrasound energy, through a material under test. The ultrasound waves travel to the far side of the
test piece and reflect back to the transducer/probe and a measurement is obtained.
Basically three typed of transducers/probes are available for different types of applications:
1> Straight beam Probe (Normal probe)
2> TR probe (Dual crystal probe)
3> Angle beam Probe
1> Straight beam Probe: This probe introduces ultrasound normal to the test piece surface
utilizing longitudinal or compression waves. Normal beam probe is used mostly for flaw
detection and thickness gauging
(refer figure 1.)
2> TR probe: This probe contains separate transmitting
and receiving elements as showing in figure 2
usually mounted on delay lines. This design
improves near surface resolution by separating
the initial pulse from the received echoes. TR
probe is suitable for the thickness gauging of pitting
and corrosion and also for better surface resolution.

Einstein-II DGS

--3--

MODSONIC
3> Angle Beam Probe: This probe introduces ultrasound at an angle to the surface of the test
piece. In most angle beam probes,
the wave energy is mode converted from a
longitudinal wave to a shear wave by the
refraction principle. This probe is suitable for
inspection of the welds. The reason for this is
its ability to position the transducer/probe
away from the weld bead yet propagate energy
into the weld zone. Another reason to use angle
beam testing on the welds is to position the
sound beam more normal to the expected discontinuities since the flaws in welds are usually
perpendicular to the test piece (except porosity). Figure 3 shows the principle of angled beam
weld testing.

Einstein-II DGS

--4--

MODSONIC

Parts and Controls of Einstein-II DGS


16

17

19

18
15

14
13

12
3
11

10

9
1

Einstein-II DGS

21

20

--5--

MODSONIC
1. POWER key : Einstein-II DGS can be switched ON and OFF using this key.
2. ZOOM key : By pressing this key A-Scan spreads over to full screen. Menu
parameters disappear from LCD and only A-Scan is displayed. By
pressing any other key, A-Scan is restored to normal size and
menu parameters appear again .
3. FREEZE key : By pressing this key once A-Scan can freeze. By pressing second
time selects Peak Freeze mode. In this mode it hold echo dynamic
pattern which is helpful for a angle beam inspection to locate peak
signal.
4. LOCK key : By pressing this key all function (keys) get locked which prevents
accidental pressing of any key. To unlock, simultaneously press
Step key (9) and LOCK key (4).
5. LCD Screen : It displays A-Scan Trace pattern and Menu Parameters.
6. Battery Status : When all the three cells are displayed empty, then the battery
Indicator
needs to be recharged.
7. Gain Step

Einstein-II DGS

: This location displays Gain Step value. This value can be changed
by using dB STEP(10) key. It follows 0.5, 1, 2, 6, 12 and 20
dB Step sequence.
--6--

MODSONIC
8. Menu keys

: By using these keys the Menu can be changed.

9. Hot key

: When this key is pressed simultaneously with any other parameter


key then the change in value takes place in larger steps.

10. dB STEP key : Any Gain step out of 0.5, 1, 2, 6, 12 or 20 dB can be selected
using this key.
11. GAIN UP/DN : Gain can be incremented/decremented using these keys in the step
selected by dB STEP key.
12. UP/DN key

: Respective parameters can be incremented or decremented


using these keys.

13. REJECT
LED

: Rejection function activation is indicated by this LED.


When Rejection value is zero then LED is OFF. When rejection
value is not zero then LED illuminates.

14. MONITOR LED : It is visual indication of MONITOR getting triggered.(LED 1


and LED 2 represent Monitor Gage 1 and Gate 2 respective.)

Einstein-II DGS

--7--

MODSONIC
15. Parameter
Values

: All function (parameter) values are displayed in this area of


LCD.

16. RX Connector

: A probe is connected to the Receiver of Einstein-II DGS


through this connector.(choice of BNC and LEMO )

17. TX Connector

: A probe is connected to the Transmitter of Einstein-II DGS


through this connector.(choice of BNC and LEMO)

18. Charge Connector: Connect the charger here.


19. RS-232 Connector: Serial interface connector. For print-out on Serial PC
printer/Computer connect supplied cable on this connector.
20. Battery
Compartment

: Battery is fitted in this compartment.

21. Mode Status


Indicator

: Mode selection SINGLE or DUAL is displayed here.

Einstein-II DGS

--8--

MODSONIC

4 Operation detail
Einstein-II DGS is user friendly, menu driven Ultrasonic Flaw Detector. To access any
parameter, various menus are available. Details of all menus and parameters are as per below.
First menu is Basic menu. In this menu, Basic functions of an
Ultrasonic Flaw Detector (UFD) like Range, Fine Range,
Delay and Gain functions are covered.
ZERO
ZERO
: It is to calibrate Artificial zero of the probe.
2200
RANGE : The required test range can be set directly in terms of mm/
RANGE
inch. The range available for steel is from 10 to 5000 mm
0100 MM
(0.50 to 200 inch) it steps through 10,25,50, 100, 125,200,
VEL M/S
300,500,1000,2000,3000,4000 and 5000 mm with pressing
5920
hot key or in step of 1mm (0.5,1,2,5, 10, 15,20,25,30,50,100
150
and 200 inch) or in step of 0.2 inch).
DELAY
000.0MM VELOCITY : The required velocity can be set directly in terms of
met./Sec (inch/sec). By pressing hot key it steps through
GAIN 1
preset value. In fine mode 1 met./sec. (0.0001 in/us).
25.0 dB
DELAY
: Partial Range can be set using these keys. Exact shift/delay
in mm (Inch) is displayed at the bottom.

Einstein-II DGS

--9--

MODSONIC
GAIN STEP: Gain step value can be set by using GAIN STEP key, it follows dB in
steps of 0.5, 1, 2, 6, 12 or 20 dB.
GAIN
: Gain can be increased/decreased as per value of step by using GAIN
UP/DN keys.
Example: Einstein-II DGS is to be calibrated for 100 mm Test Range using Single
crystal (Normal) probe.
> Set required range to 100 mm using RANGE UP/DN keys.
> Couple the probe on 25mm thickness of V1 reference block and set first echo at 25mm
using ZERO UP/DN keys.
> Set 4th echo at 100 mm using VELOCITY UP/DN keys, if required.
> Repeat above two steps until both the echoes are at their respective positions.
> Amplitudes of the echoes can be varied using GAIN keys.

Einstein-II DGS

--10--

MODSONIC
Second Menu covers MONITOR GATE 1 functions.
GATE 1 : Gate can be changed to PLOGIC, NLOGIC, DAC, DAC 6,
GATE 1
DAC-14 trigger, EXPAND or OFF.
OFF
PLOGIC : When PLOGIC is selected, the MONITOR is
START
triggered if an echo is present within the gate and its
0030.0
height is equal to or crosses the set Threshold Level.
END
NLOGIC : When NLOGIC is selected, the MONITOR is
triggered if there is no echo within the gate or the
0100.0
height is lower than the Threshold Level.
LEVEL
DAC
: If DAC is turned ON and DAC is selected and if
040 %
any echo signal equals or crosses DAC curve then
monitor gates trigger.
GAIN 1
DAC-6 : If DAC is made ON 6/14 and DAC-6 is selected then if
25.0 dB
any echo signal cross DAC-6 dB curve then monitor
gates trigger.
DAC-14 : If DAC is made ON 6/14 and DAC-14 is selected then if echo
signal equals or crosses DAC-14 dB curve then monitor
gates trigger.
EXPAND: When EXPAND is selected then the Scale expansion is activated and
the GATE width is expanded to Full-screen width for a magnified view.
Einstein-II DGS
--11--

MODSONIC
START : Gate starting position can be set using START UP/DN keys. Using these
keys Gate start can be directly set in terms of mm/inch.
END
: Gate End position can be set using END UP/DN keys. Using these keys Gate
end can be directly set in terms of mm/inch.
LEVEL : Gate Threshold Level can be directly set in terms of the percentage of the
Full Scale Height (FSH).
Example: The MONITOR to be triggered when any echo crosses the 50% FSH
between 35mm and 80mm.
> In the Gate setting, select PLOGIC using GATE UP/DN keys.
> Set Gate START to 35mm using START UP/DN keys.
> Set Gate END to 80mm using END UP/DN keys.
> Set Gate LEVEL to 50% using LEVEL UP/DN keys.

Einstein-II DGS

--12--

MODSONIC
Third Menu covers MONITOR GATE 2 functions.
GATE 2
OFF
START
0040.0
END
0100.0

GATE 2 : Gate can be changed to PLOGIC, NLOGIC, or OFF.


PLOGIC : When PLOGIC is selected, the MONITOR is
triggered if an echo is present within the gate and its
height crosses the set Threshold Level.
NLOGIC: When NLOGIC is selected, the MONITOR is
triggered if there is no echo within the gate or the
height is lower than the Threshold Level.

LEVEL
050 %

START : Gate starting position can be set using START


UP/DN keys. Using these keys, Gate start can be
directly set in terms of mm.

GAIN 1
25.0 dB

END

LEVEL

: Gate End position can be set using END UP/DN keys.


Using these keys, Gate end can be directly set in
terms of mm.

: Gate Threshold Level can be directly set in terms of the percentage of


the Full Scale Height (FSH).

Einstein-II DGS

--13--

MODSONIC
Forth Menu covers Memory storage/recall and Trace printout functions.
MEMORY
A SCAN
MEM NO*
001
ACTION
RECALL
NOTE
ENTER
GAIN 1
25.0 dB

MEMORY : Memory function can be selected A SCAN or SET-UP


using MEMORY UP/DN keys.
MEM NO : Desired memory location can be selected using MEM NO
UP/DN keys. When changing Memory no. If Data is stored
and valid then asterisk (*) will be displayed next
to MEM NO.
ACTION : Required functions like SAVE, RECALL, SEND, PRINT or
DELETE, DETAIL can be selected using ACTION UP/DN
keys. When DETAIL is selected it shows Note data of
Stored A Scan it can be scroll by pressing MEM NO UP/DN
keys.
ENTER :By pressing Right key of the ENTER, the ACTION selected
Is performed. On pressing Left key Note menu will be
displayed and Side text menu will change for Text editing
as shown. If required then edit/change Note detail. By
pressing appropriate keys Note can be edited. For exit press
Exit UP/DN keys.

Einstein-II DGS

--14--

MODSONIC
OPERATOR:
ABC
JOB DETAIL:
NIL
DEFECT DETAIL:
NIL
REMARK:
NIL

CURSOR
MOVE
CURSOR
UP/DN
CHAR
SCROLL
NOTE
EXIT
GAIN 1
25.0 dB

Example: To SAVE current Calibration and A scan in Memory no 3.


> Get required pattern which is to be stored. Press FREEZE key.
> Select A SCAN using Memory UP/DN keys.
> Set MEM NO to 3 using MEM NO UP/DN keys.
Einstein-II DGS

--15--

MODSONIC
> Set ACTION to SAVE using ACTION UP/DN keys.
> Then press NOTE/ENTER Left key so Note menu will be displayed. If required
then edit it and exit from note menu.
> Then press NOTE/ENTER Right key to store A Scan. By pressing Right key
Current Calibration, A Scan pattern and Note DATA will be stored UP/DN key to
store the current calibration set-up at memory no. 3.
Note: If data is already stored at this location, then the message will be displayed on LCD
screen as OVER WRITE DATA YES - PRESS HOT KEY,
NO PRESS ANY KEY. So press HOT KEY to overwrite data.

Einstein-II DGS

--16--

MODSONIC
Fifth Menu covers Rejection of unwanted signal, Set Reference, Video setting and
Testing mode selection function.
REJECT
000 %
POWER
LOW
DAMPING
LOW

REJECT

: For the rejection of unwanted signals. Using REJECT


UP/DN keys Rejection level can be set as per requirement.
It can be set directly in terms of percentage of FSH.
Activation of this function is indicated by REJ LED(13).

POWER : Hi or Low Transmitter power can be selected using


POWER UP/DN keys.

MODE
SINGLE

DAMPING: Probe Damping Hi/Low can be selected using DAMPING


UP/DN keys.

GAIN 1
25.0 dB

MODE

Einstein-II DGS

: Test Mode for the use of Single or Dual Crystal probes can
be selected using MODE UP/DN keys.

--17--

MODSONIC
Sixth menu covers the required functions for DAC curve AWS or DGS curve as per selected in
menu 10 CONFIG parameter.
DAC
DAC
: DAC setting can be switched to DRAW, ON ,ON 6/14, TCG
ON or OFF using DAC UP/DN keys. DRAW mode, peak points
OFF
of echo signals are selected.
STRT G1
ON mode :The selected peak points are digitally connected to form a
030%
DAC curve.
POINT
ON 6/14 mode: -6dB DAC and 14dB DAC will be also displayed.
001
TCG ON: Time Gain Correction activated as per DAC so it brings echo
PRESS
signal to same height irrespective of depth.
ENTER
OFF mode :DAC curve will be not displayed on LCD.
GAIN 1
STRT G1 :Start of the GATE 1 can be set using STRT G1 UP/DN
keys. To select any peak of the echo signal, move the
25.0 dB
gate on the echo signal and then press ENTER UP/DN
key. POINT value is automatically incremented to indicate
next point.(If GATE 1 is OFF then turn to PLOGIC).
POINT : It indicates how many points are selected to draw DAC
curve.
PRESS
: Press ENTER UP/DN keys to accept the peak points
selected by the GATE 1.
Einstein-II DGS
--18--

MODSONIC
PRESS/ENTER : Parameter will be changed to T-CORR when DAC is On or TCG is
ON. Gain of machine can be change without curve height. It can be
used for Transfer/Scanning gain correction.
Example: To digitally draw DAC curve to join four peaks of the echo signals using 25mm
thick reference block and range of 100mm.
> Set Gain so that all four peaks are visible on the LCD screen above 5% FSH.
> Set DAC to DRAW using DAC UP/DN keys.
> Move cursor using STRT G1 UP/DN keys to cover the first echo. If GATE 1 is
OFF then switch to PLOGIC.
> Press ENTER UP/DN key to accept it and Point No. is incremented automatically.
> Repeat above two steps for second, third and forth echo.
> Set DAC to ON using DAC UP/DN key. Now DAC Curve will be displayed on
LCD.
Note: Cursor need not touch the echo for selection and acceptance of the peak point.

Einstein-II DGS

--19--

MODSONIC
AWS Function functions will be displayed as per selected in menu 10 CONFIG parameter.
: AWS setting can be switched to SET, MEASURE or OFF
using AWS UP/DN keys. SET mode, Make GATE 1 to
PLOGIC then Bring reference echo of reflector in between 40 to
60% , Press ENTER parameter key so gain of reference echo is
recorded and displayed as REF GAIN. (RG)
MEASURE mode : Display IL (Indication level) dB value to bring echo
to reference level.
AF
: Display attenuation factor to correct for the depth of the echo
signal.
IR
: Display indication rating as per AWS code. IR=IL-RG-AF.
Example:
GAIN 1
> Select MEASURE to ON.
25.0 dB
> Set GATE 1 to PLOGIC
> Couple probe on reference defect and set gain value so echo height within
40 to 60% of Screen.
> Set GATE 1 in such a way so echo signal within GATE.
> Then press SET REF /ENTER key so required gain to bring echo to 50% is
recorded and displayed as a REF Gain in GAIN box.
> Set AWS to MEASURE so it starts to display Indicating rating as per AWS
code.
Einstein-II DGS
--20-AWS
OFF
STRT G1
030%

AWS

MODSONIC
Indication rating is based below formula.
IR=IL-RG-AF
IL : Indicating Level its gain value to bring echo height to reference level (50%).
RG : Reference Gain.. Its Gain value to bring reference echo height to reference
level(50%)
AF : Attenuation Factor its attenuation of signal in material as per AWS standard.
Based on below formula.
AF= (Sound path (in Inch) -1) * 2
IR : Indicating Rating Its different in dB between the indication signal and reference signal with
attenuation correction.

Einstein-II DGS

--21--

MODSONIC
DGS functions will be displayed as per selected in menu 10 CONFIG parameter.
DGS
OFF
STRT G1
030%
DGS-REF
OFF
TL-CORR
00.0dB
GAIN 1
25.0 dB

DGS

: DGS setting can be switched to SELECT, ON or OFF using


DGS UP/DN keys. In SELECT mode probe type , reference
dia., reference type, required DGS curve dia. etc. can be
selected. Refer DGS selection menu. In ON mode DGS curve
will be displayed on Screen as per what is selected in DGS
selection menu.
STRT G1 : To select echo peak as reference move GATE 1 using
STRT G1 UP/DN keys. If GATE 1 is OFF then select to
PLOGIC in Menu 2.

SET-REF :When switched to ON, reference echo within gate is


registered and height of reference echo is brought to ref.
curve height. When DGS is turned to ON then SET-Ref will
automatically switched OFF.
TL-CORR : This correction is required if test object and reference block
have different surface qualities. By varying TL-CORR gain will
change but DGS curve height remains same.(It works when
DGS is selected to ON).

Einstein-II DGS

--22--

MODSONIC
DGS Selection Menu: When DGS is selected to SELECT DGS selection menu will
be displayed on Screen. By pressing
SCAL NO
D EFFT
DGS
Menu UP/DN keys arrow( ) shifts.
002
23.1 MM EXIT
SELECT
It indicates currently selected sub
PRBNAME
DGS-CRV
ATT-REF
STRT G1
menu. For eg. To change SCAL NO
M2SN
11.0 MM 00.0 dB
030%
move arrow to first bottom empty box
then press corresponding UP/DN keys.
PRBFREQ
REFECHO
ATT-OBJ
DGS-REF
2.00MHZ
FBH
00.0 dB
OFF
DEL-VEL

REFSIZE

AMPCOR

TL-CORR

2500M/S

11.0 MM

00.0 dB

00.0 dB

SCAL NO

GAIN

25.0 dB

: It scrolls from 0 to 18. Select 0 for custom probe which one is not listed
between 1 to 18. For custom probe select appropriate probe frequency and
probe effective dia. For predefined probe, 1 to 18, it is not required.

PRBNAME : It displays Model name of predefined probe.( for scale no 1 to 18 but for
scale 0 it will display CUSTOM). With each Memory set one Custom
probe parameters will be stored.
Einstein-II DGS

--23--

MODSONIC
PRBFREQ

: It displays probe frequency of predefined probe. For custom probe select


appropriate probe frequency by pressing appropriate UP/DN keys.

DEL-VEL

:It displays Ultrasound velocity of probe delay line. It is fixed for


predefined probes.

D-EFFT

: It displays Effective diameter of probe crystals. It is fixed for predefined


probe. Select appropriate value for Custom probe.

DGS-CRV

:Select appropriate value for which diameter DGS curve is required.

REF-ECHO :Select reference type weather it is FBH or BW (Back wall).


REF SIZE

:Select appropriate value as per reference reflector size.

ATT-REF

:Select Ultrasound Attenuation factor of Reference object. (dB/meter).

ATT-OBJ

:Select Ultrasound Attenuation factor of test object. (dB/meter).

AMPLCOR :This is gain correction. When Angle probe is used and reference echo is
taken from 100 mm or 25 mm arc of V1/V2 reference block.
Example: To plot DGS for 11 mm size equivalent reflector for M2SN probe.
> Switch GATE 1 to PLOGIC. (NLOGIC will also work).
> Select DGS in menu 10 under CONFIG parameter. It will not turn to DGS if DAC is
selected to OFF.
> Go to DGS menu 6.
Einstein-II DGS

--24--

MODSONIC
> Couple M2SN probe on 25 thick V1 block and calibrate for required range.
> Bring first echo within screen with height between 40 to 80 % FSH preferable.
> Set first back-wall echo signal within GATE 1 using STRT G1.
> Change DGS to SELECT by pressing DSG UP/DN keys.
> Move arrow below first sub menu by pressing Menu UP/DN Keys.
> Select SCAL NO to 2 by pressing corresponding UP/DN keys.
> Move arrow below second sub menu by pressing Menu UP/DN keys.
> Select DGS-CRV to 11.0 MM by pressing corresponding UP/DN keys.
> Select REFECHO to BW by pressing corresponding UP/DN keys.
>ATT-REF and ATT-OBJ to be set as per appropriate value. Here it is set to 0.
> For exit press EXIT UP/DN keys.
>Select SET-REF to ON so reference echo height is registered and brought to reference
curve. (make sure reference echo is within GATE 1 with 40 to 80% FSH preferable).
> Now press DGS UP/DN key so it turns to ON and appropriate DGS curve will be
displayed.
--25--

Einstein-II DGS

MODSONIC
Seventh menu covers the required functions for Thickness/Distance measurements.
MEASURE : MEASURE setting can be switched to ON or OFF using
MEASURE
MEASURE UP/DN keys. In ON mode it will display
thickness/distance
of echo signal which is in-side the gate.
OFF
In OFF mode it will not display thickness/distance.
X-OFF
X-OFF
: Set X-OFF value to same as Angle probe index point to
000.0
Front face distance so Surface Distance will be displayed
ANGLE
on LCD will be distance from angle probe front face to
000 DEG
defect position.
THICK
100.0MM

ANGLE

: Value to be set as per Angle of the probe. If angle value


not zero and Measure is set to ON then appropriate
Surface and Depth of defect is calculated and display on
LCD.

THICK

: Thickness value should be set as per the thickness of the


test piece, so as to enable in calculation of
depth/distance accounting for multiple skips.

GAIN 1
25.0 dB

Einstein-II DGS

--26--

MODSONIC
Example: To measure Beam path, Surface and Depth distance when 45 deg. probe is
used.
> Go to GATE 1 menu and make Gate to PLOGIC/NLOGIC.
> Set Gate START and Gate END value in such a way so defect signal within the GATE.
> Set Gain or Threshold so height of the echo/defect beyond the GATE.
> Go to measure menu and set MEASURE to ON.
> Set angle value to 45 DEG. using angle UP/DN keys.
> Set appropriate THICK value so it can account multiple skips for DEPTH distance
calculation.
> Go to menu 11 and set MEA PO1 to PD, MEA PO2 to DP and MEA PO3 to SD so
Sound path, Depth and Surface distance will be displayed on screen.

--27--

Einstein-II DGS

MODSONIC
Eighth menu covers the required functions for Thickness/Distance measurements.

TRIGGER
FLANK
HUD
OFF
SET REF
OFF
UNIT
INCHES

TRIGGER: Select Measurement point Weather trigger by FLANK or


on PEAK.
HUD

: In Head up Display selected measured value will be


displayed if MEASURE is selected to ON in Menu 7. If
ALL is selected then value of MEA PO1, MEA PO2 and
MEA PO3 will be displayed in Head up Display. (For
selection of measured value refer menu 11 detail)

SET REF

: To set any A-scan pattern as a reference, achieve


reference pattern, then set SET REF ON, so reference
pattern sets in background for easy comparison.
A-scan Pattern From Memory location can be recalled
and set as a reference.

GAIN 1
25.0 dB
UNIT

: To set Measurement/Calibration Unit. It can be select Inches


or Metric by pressing UNIT UP/DN keys.
Note: When changing UNIT Calibration will be change.

Einstein-II DGS

--28--

MODSONIC
Ninth Menu covers miscellaneous functions, which are described as below.

CLOCK
OFF
COLOR
005
VIDEO
ENVELOP

CLOCK

: Clock can be switched ON or OFF using CLOCK


UP/DN keys. If CLOCK is set to ON then date and
time will be displayed at bottom of A-Scan.

COLOR

: LCD Color combination can be selected using COLOR


UP/DN keys.

VIDEO

: Echo signals can be selected to appear as line


(ENVELOP) or solid (FILLED) using VIDEO UP/DN
keys. In ENVELOP mode, echo signals appear same as in
Analog Ultrasonic Flaw Detector. In FILLED mode, echo
signals looks solid for better visibility.

BRIGHT
050 %
GAIN 1
25.0 dB

BRIGHT : Brightness of LCD can be adjusted for better visibility


using BRIGHT UP/DN keys. Set Low brightness when
used for indoor application for the longer battery life.

--29--

Einstein-II DGS

MODSONIC
Tenth menu covers the required functions for HORN and Key Beep Setting.
CONFIG : Evaluation mode can be selected to DAC,AWSor DGS.As per
selection, menu 6 will turns to DAC, AWS or DGS setting.
This selection can change only if current active
DAC/AWS/DGS is selected to OFF.

CONFIG
DAC
HORN
OFF
BEEP
ON
PRF
HI
GAIN 1
25.0 dB

HORN

: Horn can be switched ON or OFF using HORN UP/DN


keys. When HORN is switched ON, then it gives out an
audible alarm each time MONITOR is triggered.

BEEP

: Key press will be acknowledge by sound if BEEP is


selected to ON. Otherwise it will not give beep sound when
set to OFF.

PRF
: Using PRF UP/DB keys HI or LO PRF can be selected.
In HI PRF is depends on current range selection. In Low
mode is PRF is low which is useful when ghost echo
pattern appears when testing low attenuate material.

Einstein-II DGS

--30--

MODSONIC
Eleventh menu covers selection of measured parameter to be displayed on screen.
MEA PO1 :Measured value will be displayed on screen at Left most
position below base line. It scrolls as per detail given below.
MEA PO1
OFF
Display OFF
SND PTH
Sound path of Echo signal
OFF
DEPTH
Depth of Echo signal
MEA PO2
SURF D
Surface dist. of Echo signal
OFF
ECH HI%
Percentage height of Echo signal
MEA PO3
DAC %
Percentage height with respect to DAC curve.
(When DAC is selected to ON)
OFF
ERS VAL
Equivalent Reflector Size of echo signal
(When DGS is selected to ON)
DGS CRV
Display currently selected DGS curve size.
DGS HDB
dB diff between DGS curve and echo signal
GAIN 1
(When DGS is selected to ON)
25.0 dB
Signal to be within GATE 1 and MEASURE is selected to ON in menu
7 then appropriate measured value will be displayed onscreen.
MEA PO2 : Measured value will be displayed on screen at Middle position below
base line. It scrolls as per detail given in MEA PO1.
MEA PO3 : Measured value will be displayed on screen at Right side below base line.
It scrolls as per detail given in MEA PO1.
Einstein-II DGS
--31--

MODSONIC

5 Setting of time Keep key 1 press and switch ON Einstein-II DGS. After display
of menu Parameter, release key 1 and again press key 1 so Time setting menu will be
displayed. Now set required Time/Data by pressing
1
UP/DN parameter key. Then press key 2 for to
store/set current time to value set as per display.
Press key 3 to exit without store/set current time.
Factory Default Calibration set-up
In Einstein-II DGS last calibration set-up remain
store So when further Einstein-II DGS Starts it
will start with last setting, But due to some reason
if stored calibration is corrupted then it will starts
with factory default calibration set-up. If any user
want to recall factory default calibration set-up
then press key 4, keep it press and Switch on
Einstein-II DGS. After displaying menu parameter
release key 4.
3
2

Einstein-II DGS

--32--

MODSONIC

6 Interface of Printer

Einstein-II DGS has direct printout facility of A-Scan pattern with full calibration set-up values.
It can be printed out on conventional PC printer having a Serial Communication connector. So
computer or special printer is not required. Printer should be IBM-PC compatible with serial
port. Serial port setting is as below. Serial printer EPSON FX/LQ type recommended. (To set
below setting in serial printer refer operating manual supplied with the printer).
Baud Rate 9600
Parity bit None
Data Bits 8
Stop Bits 1
To print out trace pattern, follow the below given steps.
> Connect Serial cable (supplied with Einstein-II DGS) between connector (19) of
Einstein-II DGS and serial printer.
> Switch ON Serial PC printer.
> Switch ON Einstein-II DGS.
> Get the required Trace pattern and freeze using Freeze key (3).
> Using Menu key go to memory menu (4).
> Using ACTION UP/DN key select PRINT.
> Press ENTER UP/DN key to initiate printing.

For Interface with PC Using Serial port Refer Help.txt file supplied in Software
CD.
Einstein-II DGS

--33--

7 Power Supply

MODSONIC

Einstein-II DGS is powered by a Lithium-Ion battery pack ATB1, which is fitted in Einstein-II
DGS using two screws. It electrically connects to Einstein-II DGS via two internal terminal
contact points. With fully charged battery pack ATB 1, Einstein-II DGS can operate
continuously for eight hours, under typical operating conditions. The battery pack ATB 1 can be
charged when inside Einstein-II DGS by connecting Battery charger ATC 1 to the connector
(18). Alternatively, the battery pack ATB 1 can be charged outside Einstein-II DGS which
allows uninterrupted use of Einstein-II DGS using multiple battery packs.
Battery Charging
The battery Charger ATC 1 supplied with Einstein-II DGS is a special quick- charge charger.
The charging time is just 5 to 6 hours. ATC 1 battery charger is suitable for use with mains
supply from 100 to 240 volts AC/50 Hz. The LED on charger illuminates to indicate power on.
When the battery voltage is low (discharge condition) the LED glows RED. When the battery is
almost charged the LED turns to yellow. When the battery is full charged the LED glows green.
The output voltage from the charger is 16.4 Volts DC. The charging current is constant at 2.0
Amps.

Einstein-II DGS

--34--

MODSONIC

8 Precaution
> Avoid shock to the Einstein-II DGS and probes.
> Wipe the couplant off the probe and Einstein-II DGS after use.

> When battery pack is out-of Einstein-II DGS, keep battery contacts away from
metal object
> Never heat the battery pack ATB 1 nor throw into a fire.

Einstein-II DGS

--35--

9 Specification of Einstein-II DGS


Test Range

Velocity

Delay
Gain
Rejection
Rectification
Frequency
Test Modes
Transmitter
Damping
Connectors
Einstein-II DGS

MODSONIC

: 10 mm to 5 meter (0.500 in to 200 in) (in steel). In Hot Key mode


it has 13 preset values. Fine mode adjustable in step of
1mm (0.2 in).
: 1000 met./sec to 9999 met./sec (0.0400 in/s to 0.4000 in/s). In
Hot Key mode it has 7 preset values. Fine Mode adjustable in
step of 1 met/sec (0.0001 in/s).
: Variable from 0 to 3000 mm (0 to 100 in). In Hot Key mode 5mm
(0.2 in) . Fine mode adjustable in step of 0.5mm (0.02 in).
: 100dB calibrated gain adjustable in 0.5, 1, 2, 6, 12 or 20 dB step.
: 0 to 100%FSH with LED indicator.
: Full-wave rectified with filtering.
: Broad Band amplifier 0.5 MHz. to 10MHz.
: Pulse echo and Transmit/Receive.
:Transmission pulse Negative spike with selectable high or low
power.
: Damping high/low is selectable.
: BNC and LEMO (Size 1) both are provided.
--36--

MODSONIC
Freeze/Peak Freeze

Connectors
Monitor
Gate Expand
A-Scan memory
Calibration Set-up
Software
Printer Attachment

Display
Full Screen

Einstein-II DGS

: A-Scan freeze and Peak Freeze available. In Peak freeze it holds


echo dynamic pattern which is useful for angle beam probing to
locate peak signal.
: BNC and LEMO (Size 1) both are provided.
: Dual gate adjustable in 1% of Screen width with
Positive/Negative Logic, Gate Expand modes.
: Expands Range to width of the gate.
: 200 Trace Patterns can be stored (with Note/Detail),recalled, print
or transferred to PC via RS-232 serial port.
: 50 different calibration set-ups can be Stored and Recalled.
: EinSoft DGS Interface software for transferring A-Scan from
Einstein-II DGS to PC supplied.
: IBM compatible Printer having serial port can be directly
attached to main unit for printing of stored A-Scan waveform
with calibration data.
: High brightness Color TFT LCD Display. Display area 320 x 240
pixel (122 x 92 mm). Eleven different color scheme option.
: By pressing Zoom key A-Scan can be displayed in Full Screen
area.
--37--

MODSONIC
Reference A-Scan
DAC

TCG
AWS
DGS

Digital Readout

Measurement Unit
Einstein-II DGS

: Reference A-Scan pattern of standard test object can saved and


recalled in the background for easy comparison during testing.
: Dynamic DAC curve can be Digitally plotted (Smooth parabolic
curve) on screen with 6dB and 14dB DAC curves.. DAC curve
can be set as flaw monitor gate. Using minimum 2 to
maximum 10 point DAC curve can be plotted.
: As per plotted DAC Time Corrected Gain activated so it brings
echo signal to same height irrespective of depth.
: Built-in Software for evaluation of defect in accordance with
AWS standards
: Defect size evaluation 18 predefined probe settings and one
custom probe set-up per memory location. Defect size is directly
displayed in ERS value. (Equivalent Reflector Size).
: Thickness/Depth can be displayed in digital readout when using
a normal probe and Sound path, Surface Distance and Depth are
directly displayed when angle probe is in use. Measurement point
can be selected to be Peak or Flank. Echo height, ERS value,
dB diff of DGS curve to signal height, Echo height WRP to DAC
can be measured.
: Metric or British unit of measurement is selectable.
--38--

MODSONIC
Video Output
PRF
Update Rate
Power
Charger
Temperature
Size
Weight

: Composite video signal (PAL) output for connection to a TV.


: High or low PRF is selectable.
: 50 Hz.
: Lithium-Ion Battery pack 14.4VDC, 4AH, gives 8 hours
continuous operation from fully charged battery.
: Input voltage 100 to 240 VAC. Charge status indicators provided.
: 0 to 55 C.
: 152 x 255 x 70 mm (HxWxD).
: 2.3 kg. with Battery.

--39--

Einstein-II DGS

MODSONIC

10 Menu Tree
First Menu
Second Menu

Third Menu

Forth Menu

ZERO
GATE 1
OFF
PLOGIC
NLOGIC
DAC
DAC 6
DAC 14
EXPAND
GATE 2
OFF
PLOGIC
NLOGIC
MEMORY
SET-UP

RANGE
START

VEL
END

DELAY
LEVEL

START

END

LEVEL

MEM NO

ACTION
RECALL
SAVE
DELETE
RECALL
DETAIL

NOTE
ENTER

A SCAN
Einstein-II DGS

--40--

MODSONIC

Fifth Menu

REJECT

Sixth Menu (1) DAC


OFF
DRAW
ON
ON 6/14
TCG ON
AWS
OFF
SET
MEASURE
Sixth Menu (2) DGS
OFF
SELECT
ON
Seventh Menu MEASURE
Einstein-II DGS

POWER
LOW
HI
STRT G1

DELETE
PRINT
SEND
SAVE
DAMPING
LOW
HI
POINT

START G1
START G1
START G1
STRT G1

SET REF
IL ---- AF---DGS-REF

RATING
TL-CORR

X-OFF

ANGLE

THICK

MODE
SINGLE
DUAL
PRESS
ENTER

--41--

MODSONIC

Eighth Menu

Ninth Menu

Tenth Menu

Eleventh Menu

Einstein-II DGS

OFF
ON
TRIGGER
FLANK
PEAK

CLOCK
OFF
ON
CONFIG
DAC
DGS
MEA PO1
OFF
SND PTH

HUD
SND PTH
DEPTH
SURF D
ECH HI%
DAC %
ERS VAL
DGS CRV
DGS HDB
ALL
COLOR

HORN
OFF
ON
MEA PO2
OFF
SND PTH
--42--

SET REF
OFF
ON

UNIT
INCHES
METRIC

VIDEO
ENVELOP
FILLED
BEEP
OFF
ON
MEA PO3
OFF
SND PTH

BRIGHT

PRF
HI
LOW

MODSONIC
DEPTH
SURF D
ECH HT%
DAC %
ERS VAL
DGS CRV
DGS HDB

Einstein-II DGS

DEPTH
SURF D
ECH HT%
DAC %
ERS VAL
DGS CRV
DGS HDB

DEPTH
SURF D
ECH HT%
DAC %
ERS VAL
DGS CRV
DGS HDB

--43--

MODSONIC
All possible care has been taken in the preparation of this operation manual. If you find
any errors or would like to make suggestions for improvement to this operation manual,
then please write to:

MODSONIC INSTRUMENTS MFG. CO. (P) LTD.


Plot No.33, Phase III, GIDC Industrial Estate, Naroda, Ahmedabad 382 330 INDIA.
Tel.:- +91 (079) 2281 1217, 2281 3131, 2284 1294 Fax.:- +91 (079) 2282 0012.
E-mail: modsonic@modsonic.com Web site: www.modsonic.com

MODSONIC reserves the rights to make changes without notice.

OP22 (1)