Escolar Documentos
Profissional Documentos
Cultura Documentos
Rev.11/30/12MKB
ACL395ResistivityMeter
OPERATIONMANUAL
Meteriswarrantedforoneyearfromthedateofpurchaseonpartsandlabor.
Calibrationisrecommendedevery12months.
ACLIncorporated
840W.49thPlace
Chicago,IL60609USA
Page1of13
Tel:8479819212
info@aclstaticide.com
www.aclstaticide.com
ACL395OperationManual
Rev.11/30/12MKB
ACL395ResistivityMeter
TheACL395isalightweight,pocketsized,autorangingsurfaceresistivitymeter.Itisdesigned
totestconductive,dissipative,andinsulativesurfacesforelectricalresistivityaccordingtothe
ESDAsparallelresistivityprobemethodDINEN100015/1&ASTMD257.
Ifthemeterisusedwiththe5lbprobesavailableintheaccessorykit,itwillcomplyto
IEC6134041,ANSI/ESDAS4.1andANSI/ESDAS7.1.
ACL395ResistivityMeterincludes:
*Tester
*Twoaccordioncables(stereotobanana)
*9voltbattery
*Certificateofcalibration
ACL396AccessoryKitincludes:
*Two5lbprobes
*Foamlinedcarryingcase
Limits
*Resistivity:1031012ohms/square
*Resistance:1031012ohms
*Measuringvoltage:10vand100v
ACLIncorporated
840W.49thPlace
Chicago,IL60609USA
Page2of13
Tel:8479819212
info@aclstaticide.com
www.aclstaticide.com
ACL395OperationManual
Rev.11/30/12MKB
INTRODUCTION
TheACL395ResistivityMeterisaneasytousetesterformeasuringsurfaceresistivity.When
usedwiththeACL396AccessoryKit,theACL395ResistivityMeterisadependableauditkitfor
conductiveanddissipativesurfaces.Thismeterisdesignedforuseinallfacetsofmaterial
productionincludingengineering,maintenance,qualitycontrol,incominginspection,
manufacturingandresearch,orinsalesdepartmentsforthetestingofantistaticmats,floor
finishes,paints,wriststraps,smocks,footwear,bagsandcontainers.
Whenusingthebuiltinprobes,themeterstestvaluesforsurfaceresistivityareinohmsper
square(althoughtheyaredisplayedinohms).Whenusingtheexternal5lbprobes,themeters
testvaluesforresistanceareinohms.
DECADESCALE
103
=
1kilohm
4
10
=
10kilohms
105
=
100kilohms
300kilohms
3x105 =
6
10
=
1megohm
3x106 =
3megohm
7
10
=
10megohms
3x107 =
30megohms
=
100megohms
108
8
3x10 =
300megohms
109
=
1000megohms
3x109 =
3000megohms
10
10
=
10,000megohms
100,000megohms
1011 =
1012 =
1,000,000megohms
ThetestvalueisindicatedontheLEDdisplay.Halfdecadesprovidegreateraccuracybygivinga
closerapproximationtothemeasurementvalue.AnLEDwillbrightentotheaccordingtest
result.Colorssignifythetestvaluesfunction.
COLOR
Green
Yellow
Orange
Red
INDICATINGFUNCTION
Conductive
Dissipative:idealtestmeasurement
Dissipative,butclosetogoingoutofspec
Nearinsulativetoinsulative
103105
3x105109
3x1091010
10111012
ACLIncorporated
840W.49thPlace
Chicago,IL60609USA
Page3of13
Tel:8479819212
info@aclstaticide.com
www.aclstaticide.com
ACL395OperationManual
Rev.11/30/12MKB
TESTVOLTAGE
Thetestvoltagesare10vand100v.AccordingtoESDAssociation(ESDA)standardsS4.1and
S7.1,10vshouldbeappliedtosurfaceswithresistivityoflessthan106and100vshouldbe
appliedtosurfaceswithresistivityof106orgreater.TheACL395willautomaticallysimulate
thepropervoltageaccordingtothetestmeasurement.
AsdefinedbytheESDA,valuesindicatethefollowing:
Voltage
Range
Definition
10volt
<106ohmspersquare
Conductive
6
11
100volt
10 10 ohmspersquare
Dissipative
11
100volt
>10 ohmspersquare
Insulative
ANOTEABOUTVOLTAGE
Inpreviousyears,thosedesiringtomeasureresistivityorresistancefollowedtheASTMD264,
ASTM991,NFPA56AorNFPA99teststandards.Theseproceduresrequiredonetotestateither
500or1000volts.Thiscausedconcernregardingsafetyofthepersonperformingthetests.The
ESDAstandardizedthetestproceduressothatlowervoltscouldbeusedwithinspecificranges.
TheACL395meterusesa9voltbattery.Somemeterswith9voltbatteriesdonotgivethe
accuracyneededtoperformthetests,especiallyatvalueshigherthan107.TheACL395isbuilt
withatransformerthatconvertsthe9voltchargefromthebatteryto10voltsor100volts.The
meterappliesaconstantchargeoverthecompletevoltagerange.Accuracydependsonapplied
voltage,temperature,andhumidity.
TEMPERATUREANDHUMIDITY
Ambienthumidityandtemperatureaffecttheelectricalpropertiesofthematerialbeingtested.
Thecombinationoflowhumidityandlowtemperaturewillgivethehighestelectricalresistance
resultsandslowestdissipationtimes.Athighhumidity,athinlayerofwateriscondensedonor
absorbedintothematerialbeingtested.Thisistrueofhydroscopicadditivesthatare
incorporatedintoamaterialinordertoincreasetheelectricalconductivity.Theseadditiveswill
allowmoisturetobeabsorbedintothematerialstowhichtheyareadded.
Atelevatedtemperatures,themobilityoffreeelectronsincreases,therebyalsoincreasinga
materialsconductivity.Thisisespeciallytrueforcarbonblack,metallicoxides,metals,and
othersubstances.Whenthematerialisatalowertemperature,builtinstressesoccurwhich
mayincreaseresistanceduetoincreaseddistancebetweentheconductiveadditives.Thus,
humidityandtemperaturemustbeknown.
ACLIncorporated
840W.49thPlace
Chicago,IL60609USA
Page4of13
Tel:8479819212
info@aclstaticide.com
www.aclstaticide.com
ACL395OperationManual
Rev.11/30/12MKB
RECORDINGDATA
ANSI/ESDAandEuropeanCECCrecognizetheeffectenvironmentalconditionshaveontest
measurementsandspecifyintheirstandardsthattheymeasuredandrecorded.Itispossibleto
testormanufactureamaterialathighhumidityandpassallthetestspecifications,butwhen
thecustomerreceivesthematerialandusesitatalowerhumidityortemperaturethematerial
failstopassthespecifications.Thiscancauserejectionsandlossofproduct.
BothESDS4.1ESDProtectiveWorksurfacesections6.2.4andESDS7.11994Resistive
CharacterizationofMaterialsFloorMaterialssections5.2.4and5.3.3requirereportingof
temperatureandhumidityatthetimeoftesting.
CALIBRATION
Calibrationisrecommendedannually.TheACL395metercomeswithacertificateofcalibration
whichverifiesthecalibrationoftheinstrumentusingequipmentthatistraceabletoNational
StandardsandCADgeneratedtechniques.MeterscanbesentbacktoACLInc.forcalibrationfor
alabfeeortheycanbesenttoacertifiedcalibrationlab.(Seepage12forcalibration
instructions.)ThemetersalsocomewithaCEmarkapproval.
MEASURINGWITHINTERNALPROBES
TheparallelresistivityprobemethodcomplieswithASTMD257.Itisusedtogivefastelectrical
resistivitymeasurementsonflathomogeneousmaterials.Itmaybeusedonmultilayered
materials,butthisshouldbenotedalongwiththetemperatureandhumidityvaluesonthedata
report.
TherailsontheACL395aremadeofconductiverubber.Softrubberrailsprovidethehighestlevelof
surfacecontactforgreateraccuracy.
ACLIncorporated
840W.49thPlace
Chicago,IL60609USA
Page5of13
Tel:8479819212
info@aclstaticide.com
www.aclstaticide.com
ACL395OperationManual
Rev.11/30/12MKB
Whenthemeasurementistakenbetweenthemeterstworubberrails,themeterwillindicate
thesurfaceresistivityofthematerialbeingtested.
A. Priortotesting,makecertainthatsurfacestobetestedarecleanandfreeof
contaminants.
B. Allowthemetertoacclimatetotheatmosphereinwhichitwillbeused.Adjustmentto
newenvironmentalconditionsmaytakeaslongasahalfhour.
C. Placethemeteronthedesiredsurfacetobetested.
D. Pressandholdtheredintegratedtest
buttonwithapproximatelyfivepounds
ofappliedforce.Themeterwilldisplay
thesurfaceresistivity.
E. Whenusingthebuiltinprobes,the
meterstestvaluesforsurfaceresistivity
areinohmspersquare(althoughthey
aredisplayedinohms).
F. ThetestvalueisindicatedontheLED
display.Adecadewillbrightentothe
accordingtestvalue.Colorssignifythe
testvaluesfunction(seepage3).
MEASURINGWITHEXTERNALPROBES
Whenthemeasurementistakenusingthe5lbexternalprobesfromtheACL396accessorykit,
thetesterwillindicatetheresistanceofthematerialinohms.
Theexternalprobesareusedtogivefastelectricalresistivitymeasurementsonflat
homogeneousmaterials.Theymaybeusedonmultilayeredmaterials,butthisshouldbenoted
alongwiththetemperatureandhumidityvalueonthedatareport.
Whenthecableshavebeenpluggedinto
theappropriatesockets,theparallel
probes underthemeterdisengage.
ACLIncorporated
840W.49thPlace
Chicago,IL60609USA
Page6of13
Tel:8479819212
info@aclstaticide.com
www.aclstaticide.com
ACL395OperationManual
Rev.11/30/12MKB
Byconnectingthe5lbprobestotheACL395ssockets,itispossibletomeasureResistance
PointtoPoint(RTT),ResistancetoGround(RTG),andVolumeResistance.Usingtheseexternal
probeswillallowcompliancewithvariousstandardsincludingANSI/ESDAS4.1forWorksurface
ResistanceMeasurements,ANSI/ESDAS7.1ResistiveCharacterizationofMaterialsFloor
Materials.
Whenauditingisfinished,unplugthecablesandstoreprobesintheprotectivecase.
ResistancePointtoPoint(RTT)
RTTmeasurementscanbeusedfortheevaluationoffloors,chairs,carts,worksurfacesand
otherESDcontrolledmaterialsandproducts.Proceduresvaryregardingsamplepreparation,
probepreparationandspacingofthe5lbprobes.Selectandreadthecorrecttestprocedureor
standardforthedesiredmeasurement.Testprocedurescanbepurchasedfromwww.esda.org.
A. Connectoneendofeachbananatestleadintothesocketsofthemeter.Connectthe
otherendofthetestcoilcordsintothe5lbprobes.
B. Placebothprobesonthematerialaccordingtotestproceduresorstandardbeingused.
C. PresstheTESTbuttonandthevaluewillbedisplayedontheLEDs.Themeterwill
applythecorrectvoltage(10vor100v)accordingtothevalueofwhatismeasured.
D. Whenperformingtest,donottouchleadwiresorprobes.Avoidoverlappingoflead
wires.Thiswillensureaccuratereadings.
ACLIncorporated
840W.49thPlace
Chicago,IL60609USA
Page7of13
Tel:8479819212
info@aclstaticide.com
www.aclstaticide.com
ACL395OperationManual
Rev.11/30/12MKB
ANEXAMPLEOFMEASURINGRTTONDISSIPATIVEFLOORING:
Takingroutinemeasurementsoftilescoatedwithdissipativefloorfinishisakeycomponentof
properfloormaintenance.Anyproblemsthatarisewiththefloorwillbeeasilydetected.Keeping
arecordoftemperature,humidityandelectricalpropertieswillprovideareference,andwillpoint
towardablueprintoftrafficpatternsonthefloor.Goodrecordkeepingwillensuresuccesswhen
developingandmaintainingamaintenanceprogram.
Toobtainanaveragemeasurementofafloor,mapouta4x4sectionandconductfivetests
(oneatatime)withinthesquare.Conductatestforeachsideofthesquareandafinaltest
diagonallythroughthecenterasshowninthedrawingbelow.
IneachRTTtest,the5lbprobesareplacedthreefeetapart(36inches).Connectthetestleads
tothemeter.Attacha5lbprobetotheendofeachleadandplacethreefeetapartasindicated
below.PressandholdtheTESTbuttonontheACL395meteruntilavalueisdisplayed.
RTT#1
RTT#4
RTT#2
RTT#5
RTT#3
4x4Section
(nottoscale)
ACLIncorporated
840W.49thPlace
Chicago,IL60609USA
Page8of13
Tel:8479819212
info@aclstaticide.com
www.aclstaticide.com
ACL395OperationManual
Rev.11/30/12MKB
MEASURINGRESISTANCETOGROUND(RTG)
RTGmeasurementscanbeusedfortheevaluationoffloors,chairs,carts,worksurfacesand
otherESDcontrolledmaterialsandproducts.Keepingarecordoftestresultsfortemperature,
humidityandelectricalpropertieswillprovideareference.
ANEXAMPLEOFMEASURINGRTGONDISSIPATIVEFLOORING:
ForTestingResistanceonFloors,S7.1requiresaminimumoffiveRTGtestsper5,000sq.ft.
Connecttheleadsfortheexternalprobestothemeter.
Whenthecableshavebeenpluggedinto
theappropriatesocket,theparallel
probes underthemeterdisengage.
Attachoneleadtoa5lbprobeandplaceprobeontothefloortobetested.Attachtheother
leadtoanalligatorclipandconnecttoagroundablepoint(RTGP).Ifusingagroundadapter
plug,plugthebananaleadintotheadaptoraftertheadapterispluggedintothereceptacle.
5lbweightsitsonthefloor
Pushreddot forvalues
ACLIncorporated
840W.49thPlace
Chicago,IL60609USA
Page9of13
Tel:8479819212
info@aclstaticide.com
www.aclstaticide.com
ACL395OperationManual
Rev.11/30/12MKB
ANEXAMPLEOFMEASURINGRTGONDISSIPATIVETABLEMATS:
A. TotestRTGonaworkstation,connectthefirstleadtothemeterandtoa5lbprobe.Place
theprobeontheworksurface.Connectthesecondleadtothemeterandtoagroundable
point(RTGP).
B. ToattachleadtoRTGP,slipthealligatorcliptotheleadandconnectittothegroundsnapor
connectthebananaplugtoacommonpointgroundplug:
b. PresstheTESTbuttonandthevaluewillbedisplayedontheLED.Whileperformingtest,
donottouchleadwiresorprobe.Avoidoverlappingofleadwires.Thiswillensureaccurate
readings.
c. Resistancevaluesareinohms.Whenrecordingtestvaluesalsomakenoteofthe
temperatureandhumidity,astheenvironmentcanaffecttestresults.
ToGround
Clipalligatorcliptosnap
ACLIncorporated
840W.49thPlace
Chicago,IL60609USA
Page10of13
Tel:8479819212
info@aclstaticide.com
www.aclstaticide.com
ACL395OperationManual
Rev.11/30/12MKB
VolumeResistanceMeasurement
VolumeResistancemeasurestheelectricalpaththroughamaterial.
A.
Connectoneendofeachofthebananatestleadsintothesocketsofthemeter.
Connecttheotherendofthetestcoilcordsintothe5lbprobes.
B.
Placesamplematerialonaconductivemetalplate(suchasstainlesssteel).Placeoneof
the5lbprobesontothematerialsothatthematerialissandwichedbetweentheprobe
andmetalplate.(Seebelow)
C.
Placethesecond5lbprobeontheconductivemetalplate.
D.
PresstheTESTbuttonandthevaluewillbedisplayedontheLED.VolumeResistance
ismeasuredinohmscm.
ConductivePlate
TestMaterial
ACLIncorporated
840W.49thPlace
Chicago,IL60609USA
Page11of13
Tel:8479819212
info@aclstaticide.com
www.aclstaticide.com
ACL395OperationManual
Rev.11/30/12MKB
ACL395CALIBRATIONINSTRUCTIONS
Thismetershouldbecalibratedeverytwelvemonths.Toverifyifthetesteriswithin
specification,atestresistancecanbeappliedacrosstheparallelbarsusingaresistancedecade
box.CalibrationcanbeobtainedbycontactingACLIncorporated.
SPECIFICATIONS
Dimensions:
Weight:
Power:
Connections:
TestRange:
DissipativeRange:
UnitofMeasurement:
Accuracy:
Decaderange
10E3to4
10E4to5
10E5to3x10E5
3x10E5to10E6
10E6to3x10E6
3x10E6to10E7
10E7to3x10E7
3x10E7to10E8
10E8to3x10E8
3x10E8to10E9
10E9to3x10E9
3x10E9to10E10
10E10to11
ACLIncorporated
840W.49thPlace
Chicago,IL60609USA
70mmx130mmx35mm(approximate)
103grams(approximate)
BatteryoperatedPP39volt
2x3.5mmjackplugforearthconnection
103to1012+
3x105to3x109(decadebetweeneachdecadeonalogarithmicscale)
Surfaceresistivity
Ohmspersquare
Resistancepointtopoint
Ohms
+/0.5decadeinconductiverange
+/0.25decadeindissipativerange
Changeoverpoint
3x10E3
3x10E4
1.7x10E5
6x10E5
1.7x10E6
6x10E6
1.7x10E7
6x10E7
1.7x10E8
6x10E8
1.7x10E9
6x10E9
3x10E10
value
Page12of13
value
3,000
30,000
170,000
600,000
1,700,000
6,000,000
17,000,000
60,000,000
170,000,000
600,000,000
1,700,000,000
6,000,000,000
30,000,000,000
3Kohm
30Kohm
170Kohm
600Kohm
1.7Mohm
6Mohm
17Mohm
60Mohm
170Mohm
600Mohm
1.7Gohm
6Gohm
30Gohm
Tel:8479819212
info@aclstaticide.com
www.aclstaticide.com
ACL395OperationManual
Rev.11/30/12MKB
TESTEQUIPMENTUSED
*ResistanceDecadeBox
*TestLeads
Theresistancedecadeboxrequiredwillneedarangeoffrom1kilohm(103)to999megohms
(109).Measurementsgreaterthan109arecalculatedusingCADgeneratedtechniques,ashigh
resistancesgreaterthan109aredifficulttoverifywithatestvoltageof9volts.
Connectthetestleadsfromtheresistancedecadeboxtothetestprobesofthemeter,setthe
decadeboxtothedesiredresistancei.e.103=1K,thenpressandholdthemeterstestbutton.
The103LEDshouldlight,10K104LEDshouldlightandsoon.Tomeasurethechangeoverpoint
betweendecades,increasetheresistanceofthedecadeboxwhilepressingthemeterstest
button.RecordtheresistancewhenthenextLEDlightspermanently(thisisthechangeover
resistance).
Example:ThefirstgreenLEDisilluminated103=1kilohm.At3or4kilohms,104LEDis
illuminated.Thechangeoverpointis3or4kilohms.104=10kilohmssobetween3or4kilohms
and30or40kilohmswillbethechangeoverpointsfrom104to105.
Pleasenotethatthemeterhasnointernalpartstoadjust,soverificationofcalibration
canbeachievedbyusingtheaboveprocess.Ifverificationcannotbeachievedtheunit
shouldbereturnedtothesupplier.
ACLIncorporated
840W.49thPlace
Chicago,IL60609USA
Page13of13
Tel:8479819212
info@aclstaticide.com
www.aclstaticide.com