Escolar Documentos
Profissional Documentos
Cultura Documentos
Full-Scale Coiled Tubing Fatigue Tests With Tubing Pressures to 15,000 psi
D.C. Headrick and R.S. Rosine, SPE, Halliburton Energy Services, Inc.
Abstract
In recent years, the performance envelope of coiled tubing (CT)
has expanded because of improvements in tubing manufacture
and the prediction of low-cycle fatigue life. More accurate life
predictions have been achieved with data from laboratory-scale
and full-scale CT fatigue tests. To date, most testing has evaluated only tubing pressures below 10,000 psi. Fatigue models at
higher pressures remain unvalidated.
This paper presents results from new, full-scale fatigue tests
performed at 7,500 to 15,000 psi. CT with 90,000-psi minimum
yield strength was cycled to failure at internal pressures of 7,500,
10,000, 12,500, and 15,000 psi. The measured fatigue-life
values helped validate the fatigue model used for CT job design,
and allowed the models further refinement.
Additional data are presented for full-scale fatigue tests
performed with extended tubing-stroke lengths. This testing
verifies the correlation between the tubing life observed in
controlled, full-scale, short-stroke fatigue tests and the fatigue
life observed in field applications with significantly longer
stroke lengths. Measurements and analyses of tubing-diameter
growth during high-pressure cycling are included.
Introduction
For many years, the coiled tubing service industry considered
5,000-psi internal pressure and 3,000-psi wellhead pressure to
be the limits for routine CT operations. Recent technological
advances have allowed applications with internal and wellhead
pressures as high as 10,000 psi. This paper focuses on the
technical issues of fatigue damage and diameter growth that can
occur when CT internal pressures exceed 10,000 psi.
References at the end of the paper.
FULL-SCALE COILED TUBING FATIGUE TESTS WITH TUBING PRESSURES TO 15,000 PSI
However, laboratory-scale machines usually provide a simplified loading, which may not accurately duplicate the fatigue
damage observed with full-scale equipment.
To understand CT limitations and develop an accurate
mathematical model of fatigue and diameter growth, operators
should consider data from all three sources. Although this paper
presents results from full-scale tests only, efforts are underway
to obtain similar data from field operations and laboratory-scale
fatigue machines.
Fatigue Test Procedures
The following fatigue tests were performed on 1.5-in. CT with a
90,000-psi nominal-yield strength. Most of the tests used a
conventional CT service unit, rated for 60,000 lb. Fig. 1 shows
the setup used for full-scale fatigue tests. Fig. 3 (Page 4) shows
the actual test arrangement.
In the first set of tests, 1.50-in. OD 0.188-in. wall tubing
was cycled to failure at internal pressures of 7,500, 10,000,
12,500, and 15,000 psi. Ten individual tests were run: one at
7,500 psi, three at 10,000 psi, three at 12,500 psi, and three at
15,000 psi. The tubing was run into a water-filled test well,
containing 4.0-in. ID casing open to the atmosphere (wellhead
pressure = 0 psi). The CT was sealed off at the tip, and pressurized with water through the reel to the required test pressure. A
tubing stroke length was selected on the basis of providing a
25-ft CT section that would encounter all bending events and run
into the well with each stroke.
In the second set of tests, 1.50-in. OD 0.156-in. wall tubing
was cycled to failure at various stroke lengths. The longer stroke
lengths more closely simulated actual field conditions. The
internal pressure was maintained at 10,000 psi for all tests in this
series. The CT was run into 2 7/8-in. tubing, rather than the 4.0in. ID casing used on previous tests.
Results of Tubing-Pressure Variations
Table 1 (Page 3) contains results of the full-scale fatigue testing
on 1.50-in. OD 0.188-in. wall tubing at various internal
pressures. A single fatigue test was performed with 7,500-psi
tubing pressure for evaluating equipment, procedures, and safety
issues. Then, testing proceeded with tubing pressures of 10,000
psi and higher.
Nearly all failures occurred while the tubing was coming off
or going onto the reel. A pinhole developed during the 7,500-psi
tests (Fig. 4, Page 5). Tests at higher pressures resulted in a
blowout of the tubings underside (Fig. 5, Page 5). In almost
every case, the tubing remained in one piece.
Results of Stroke-Length Variations
Table 2 (Page 3) presents results of the full-scale fatigue testing
on 0.156-in. wall tubing at 10,000-psi internal pressure. The
varied stroke length produced the test specimen lengths shown.
Because of the amount of tubing required for the tests and time
limitations, only one test was performed at each of the stroke
lengths greater than 100 ft.
SPE 54482
Avakov, V.A. and Martin, J.: Large Coiled Tubing Fatigue Life,
paper SPE 38407 presented at the 1997 SPE/ICoTA North American Coiled Tubing Round Table, Montgomery, TX, Apr. 1-3.
Avakov, V.A., Foster, J.C., and Smith, E.J.: Coiled Tubing Life
Prediction, paper OTC 7325 presented at the 1993 SPE Offshore
Technology Conference, Richardson, TX, May 3-6.
SPE 54482
165
117
69
39
112
66
40
105
71
33
165
111
69
37
60
49
46
34
55
54
56
49
46
34
Tubing Pressure
(psi)
Diameter Growth
Per Stroke (in.)
7,500
10,000
12,500
15,000
10,000
0.0006
0.0018
0.0037
0.0045
0.0034
1.50 0.156
om000303
FULL-SCALE COILED TUBING FATIGUE TESTS WITH TUBING PRESSURES TO 15,000 PSI
om000253
Fig. 360K rigup used for testing CT fatigue at tubing pressures as high as 15,000 psi.
SPE 54482
SPE 54482
Fig. 4A crack in tubing during a fatigue test with 7,500-psi tubing pressure.
FULL-SCALE COILED TUBING FATIGUE TESTS WITH TUBING PRESSURES TO 15,000 PSI
SPE 54482
om000304
Fig. 6Measured fatigue life vs. fatigue life predicted by a fatigue model for QT-900 coiled tubing.
om000305
SPE 54482
Fig. 8Typical display showing percentage of fatigue life used vs. position along the CT string.
om000388
Fig. 9OD growth of 1.50- 0.188-in. CT during cycling while internal pressure is varied.