Escolar Documentos
Profissional Documentos
Cultura Documentos
Talanta
journal homepage: www.elsevier.com/locate/talanta
a r t i c l e
i n f o
Article history:
Received 2 August 2008
Received in revised form
27 November 2008
Accepted 29 November 2008
Available online 6 December 2008
Keywords:
Polyaminoanthraquinone
LeadII) ion-selective electrode
Conducting polymer
Lead(II) ionophore
Performance optimization
Membrane electrode
a b s t r a c t
A new polyvinylchloride membrane electrode was facilely prepared by using polyaminoanthraquinone
(PAAQ) microparticles with an intrinsically electrical conductivity as a lead(II) ionophore. It is found
that the electrode performance will signicantly be improved with adding 1 wt% PAAQ microparticles and decreasing the membrane thickness. A 90 m-thick membrane electrode consisting of
PAAQ(salt):polyvinyl chloride:dioctylphthalate:sodium tetraphenylborate of 1:33:66:1 (wt) but without
any traditional lead(II) ionophore achieved the optimal performance and exhibited a good Nernstian
response for Pb(II) ions over a wide concentration range from 2.5 106 to 0.1 M with a slope of
28.9 mV/decade and a detection limit down to 776 nM. A reasonably short response time of 12 s was
revealed together with a long lifetime over a period of around 4 months in a wide pH range between 2.8
and 5.2. A xed interference method indicated that the electrode has an excellent selectivity for lead(II)
ion over alkali, alkaline earth and other heavy metal ions. The proposed electrode has been also found to
be a powerful indicator electrode for potentiometric titration of Pb(II) ions with EDTA. The electrode can
be used to accurately monitor the Pb(II) pollution in environmental waters.
2009 Published by Elsevier B.V.
1. Introduction
Lead(II)ISE is of great signicance because the lead is ubiquitous in the environment and extremely hazardous to human health.
A large number of lead(II)ISEs reported are mainly based on polymeric membrane containing ionophores [16]. The key part of these
electrodes is a highly sensing ionophore having strong afnity for
a particular metal ion but giving low or even no response to others.
Enormous efforts have been made to design and synthesize suitable materials that are highly selective to lead(II) ions. Macrocyclic
crown ethers [24], calixarenes [57], and Schiff base [8,9], have
been widely investigated for this purpose. Other ligands such as
porphyrin [10,11], pyridinecarboximide [12], piroxicam [13], tetrabenzyl pyrophosphate [14], capric acid [15], phenyl disulde [16],
dithiodibenzoic acid [17], and quinaldic acid derivatives [18] are
also served for the fabrication of the lead(II)ISEs. In spite of availability of various lead(II)ISEs, the narrow working concentration
range, high detection limit, slow response rate and especially poor
selectivity over interfering ions have restricted their widespread
application. In particular, the ionophores studied are almost all the
lipophilic compounds that easily leak out from matrix. Only one
report on the polymer as ionophores with good anti-leaking ability and then long lifetime was found [19]. However, the response
time is long because of its intrinsically high membrane resistance.
Therefore, it is of great challenge and signicance to further search a
new ionophore that possesses an intrinsically low membrane resistance and therefore high sensitivity and selectivity to lead(II) ions
so as to develop lead(II)ISEs with excellent detection performance
including quick responsibility and long duration.
In recent years, it is found that the conducting polymers especially some polymers from aromatic diamines have a unique ability
to form stable complexes with some heavy metal ions such as Pb(II),
Hg(II) and Ag(I) ions [2026]. Polyphenylenediamine synthesized
in our laboratory was found to possess a strong capability to adsorb
lead ions through complexation between Pb2+ ions and NH/ = N
groups in the macromolecular chains [23]. Concerning these properties, attempts have been made to employ these polymers to
extract and sense heavy metal ions. Carbon paste electrode modied with poly(1,8-diaminonaphthalene) was successfully used
to determine lead(II) ions in a concentration range from 0.2 to
10 M [21]. The detection process was complicated due to an indispensable combination of ion pre-enrichment and anodic-stripping
voltammogram, accompanying with a quite narrow work concentration range. Nevertheless, these results have indicated the
possibility of some aromatic conducting polymers with strong
afnity to lead(II) ions to act as ionophores in lead(II)ISEs. Unfortunately, no relevant reports have been found up to now.
A novel multifunctional polyaminoanthraquinone (PAAQ) was
successfully synthesized through a chemically oxidative polymerization [26]. PAAQ exhibits a strong adsorbability to lead ions owing
to its strong complexibility with lead ion through N and O coordination centers in the polymer chains. It seems that the PAAQ could be a
potential ionophore for construction of a new membrane sensor for
lead ions. The ingredient and thickness of the membrane were optimized to provide unique Pb2+ ISE that could result in reproducible,
noiseless and stable potentials. The excellent performance of the
electrode containing PAAQ as the sensing ionophore to lead(II) ions
in the determination of lead(II) ions has been elaborated for the rst
time.
499
2. Experimental
2.1. Reagents
1-Aminoanthraquinone
(AAQ),
ammonium
persulfate
((NH4 )2 S2 O8 ), HClO4 (70%), acetonitrile, high molecular weight
polyvinyl chloride (PVC), dioctylphthalate (DOP), sodium
tetraphenyl borate (NaTPB), tetrahydrofuran (THF) and lead
nitrate of analytical reagent grade were commercially obtained
and used as received. 0.1 M Pb(NO3 )2 stock solution was prepared
by dissolving lead nitrate in distilled deionized water and standardized whenever necessary. The working solutions of different
concentrations were confected by gradually diluting the stock
solutions.
2.2. Synthesis of ne PAAQ microparticles as ionophore
PAAQ particles used as ionophore were simply prepared by a
chemically oxidative polymerization of the AAQ monomer [26]. A
typical procedure for the preparation of PAAQ particles is as follows: AAQ monomer (446 mg, 2 mmol) and HClO4 (11.8 M, 0.17 mL)
were added into acetonitrile (40 mL) in a 100 mL glass ask in a
water bath at 20 C and the mixture was then stirred vigorously for
30 min. (NH4 )2 S2 O8 (456 mg, 2 mmol) was dissolved separately in
deionized water of 0.75 mL to prepare an oxidant solution. The oxidant solution was then added dropwise into the monomer solution
at a rate of one drop (around 60 L) every 3 s. The reaction mixture
was stirred continuously for 24 h at 20 C. After reaction, the resulting polymer particles as precipitates were isolated from the reaction
mixture by ltration and washed with ethanol to remove the residual oxidant, remaining monomers and soluble oligomers. The PAAQ
particles were left to dry in ambient air at 50 C for 3 days. The
bluish black PAAQ particles obtained have nominal macromolecular structural formula in Scheme 1 and bulk electrical conductivity
of 5.0 105 S/cm at 15 C.
500
Fig. 2. FTIR spectra of PAAQ ionophores (base and salt) before and after the exposure
of PAAQ in the 0.001 M Pb(NO3 )2 solution for 1 h.
Fig. 1. Potentiometric responses of Pb2+ selective electrodes based on PAAQ
microparticles at two doping states of salt and base at PAAQ:PVC:DOP:NaTPB of
1:33:66:1 and membrane thickness of 160 m.
501
Table 1
Composition and performance characteristics of Pb(II) ISEs in Fig. 3.
PAAQ:PVC: DOP:NaTPB (weight ratio)
1:25:50:1
1:33:66:1
1:50:100:1
1.0 105
to 0.1
4.0 106 to 0.1
5
1.0 10 to 0.024
Slope (mV/decade)
26.7
28.1
27.9
3.29 106
16
14
17
1.86 106
2.92 106
4.0 106 to 0.1 M. The ISEs based on the other three plasticizers
only give a diminished response slope (<27 mV/decade) over narrow working concentration range. Therefore, DOP was proved to be
the most effective plasticizer in preparing the proposed Pb2+ ISEs
based on PAAQ.
3.4. Lipophilic anion additive content
The presence of lipophilic anion additive in cation-selective
membrane electrodes can not only reduce the ohmic resistance and
anion interference but also improve the sensitivity and selectivity
[11]. The potential responses of the Pb2+ ISEs containing lipophilic
anion additive (NaTPB) were investigated to examine the effect
of additive content. It can be seen from Fig. 5 that the potential
response of the Pb2+ ISE without NaTPB gives the weakest sensitivity with a diminished slope of 25.3 mV/decade over a narrow linear
range from 105 to 102 M. The sensitivity can be greatly improved
by the addition of NaTPB. The Pb2+ ISE with PAAQ:NaTPB of 1:1
(wt) displays the best response characteristics with a higher Nernstian response slope of 28.1 mV/decade over a wider linear range
of 4.0 106 to 0.1 M and a lower detection limit of 1.86 106 M.
Higher content of additive (PAAQ:NaTPB = 1:1.5) is also unfavorable
for the improvement of sensitivity. More additive may deteriorate
the mechanical property of membrane and the lifetime of electrode. Therefore, the optimal lipophilic additive content is 100 wt%
relative to the ionophore.
3.5. Thickness of the membrane containing PAAQ microparticles
Three electrodes with different membrane thicknesses but
the same membrane composition of PAAQ:PVC:DOP:NaTPB = 1:33:
66:1 were fabricated to examine the effect of membrane thickness.
As shown in Fig. 6 and Table 2, a signicant improvement is achieved
in the response characteristics of electrodes including linear range,
502
Table 2
Performance characteristics of Pb(II) ISEs in Fig. 6.
Membrane thickness (m)
220
160
90
1.0 105
to 0.1
4.0 106 to 0.1
6
2.5 10 to 0.1
detection limit, slope and response time with decreasing the membrane thickness. At the smallest membrane thickness of 0.09 mm,
the electrode exhibits an almost Nernstian slope of 28.9 mV/decade
and a widest working concentration range from 2.5 106 to 0.1 M,
which has been broadened nearly by four times compared with that
at the thickness of 0.22 mm. The electrode also possesses the lowest
detection limit of 7.76 107 M and the shortest response time of
12 s. It seems that the optimal PAAQ membrane exhibits comparable or even better performance than 10 representative lead(II)ISEs
with PVC matrix in Table 3.
It is reported that the membrane thickness of the PVC-based
Pb2+ ISEs is mostly between 0.2 and 0.5 mm [1,10,11,27,28]. The
sensing unit supported by a highly porous polymeric layer in these
electrodes is practically liquid state comprising a lipophilic organic
ionophore dissolved in oil phase such as DOP. The detection limits
of these electrodes commonly keep at the order of 106 or 107 M,
Slope (mV/decade)
28.7
28.1
28.9
4.47 106
15
14
12
1.86 106
7.76 107
beyond which the potential response would deviate from the Nernstian equation, because a zero-current ux of the primary ions from
the membrane into sample solutions would result in ion activity
of the primary ions maintaining as high as 106 or 107 M in the
local domain near the membrane surface at the sample side. The
ux of primary ions is generated by a transmembrane concentration gradient which occurs when the concentrations of the sample
and inner solutions are not the same. The detection limit could be
improved through reducing the transmembrane concentration gradient by adjusting the concentration of the inner solution to the
value of sample solution, decreasing the total ion concentration,
or increasing the thickness of the membrane [29]. However, the
result here is reverse to the above-mentioned circumstance. The
potential response characteristics of the proposed electrodes can be
enhanced by decreasing the thickness of the membrane. This may
be relevant to the existing state of the ionophore. The ne PAAQ
particle ionophore that is insoluble in DOP oil phase, disperses uniformly in the PVC membrane as the solid state. The transporting
rate of lead(II) ions in solid phase is much slower and more arduous than that in liquid phase. Hence the ux of lead(II) ions is not
encouraged in solid membrane. Nevertheless, the response is still
very fast due to the relatively thin membrane, as discussed below.
Considering the remarkable decrease of membrane strength, it is
not suggested to further reduce the membrane thickness. All further
detailed studies were carried out on the electrode with the composition of PAAQ:PVC:DOP:NaTPB (1:33:66:1) and the membrane
thickness of 0.09 mm.
Another result which disagrees with the theory that increasing membrane thickness might improve the detection limit was
also reported in the investigation on the electrode based on
diphenylmethyl-N-phenylhydroxamic acid ionophore [30]. Among
three styrene/acrylonitrile copolymer membranes with different
thicknesses of 0.14, 0.21, and 0.45 mm, the 0.21 mm-thick membrane demonstrates the lowest detection limit.
3.6. Effect of pH
The dependence of the potentiometric response of the proposed
ISE on the pH value of the Pb2+ solution was tested at three Pb2+
Table 3
Comparison of performance characteristics of the proposed electrode with the PVC membrane lead(II) ISEs reported in literature.
Ionophore
Working concentration
range (M)
Slope (mV/decade)
Lifetime (month)
Refs.
N,N -Dibenzyl-1,4,10,13-tetraoxa7,16-diazacyclooctadecane
Hexathia-18-crown-6-tetraone
2 -Methoxyethoxyl
sym-dibenzo-16-crown-5-ether
Tetrakis(p-carboxyphenyl)azo]-8tetrahydroxy
calix[4]arene
Meso-tetrakis(2-hydroxy-1naphthyl)
porphyrin
Chiral 2,6-bis-pyridinecarboximide
Piroxicam
Tetrabenzyl pyrophosphate
Capric acid
Phenyl disulde
Solid PAAQ particles
30.0
8.2 106
10
[1]
29.0
28.9
8.0 107
1.0 106
40
30
2
3
[2]
[3]
29.4
8.0 107
20
[6]
29.2
3.5 106
10
[11]
27.9
30.0
28.7
29.0
29.3
28.9
2.2 106
4.0 106
3.0 106
6.0 106
1.2 106
7.8 107
25
45
10
15
45
12
0.5
3
0.7
3
1.7
4
[12]
[13]
[14]
[15]
[16]
This study
concentrations (1.0 102 , 1.0 103 and 1.0 104 M) over the pH
range between 1 and 6. It is seen from Fig. 7 that the potential
response remains almost constant over the pH range from 2.8 to
5.2, beyond which a gradual change in potential can be observed.
As a result, this range can be taken as the working pH range of the
proposed electrode. The declined potential at higher pH values may
be ascribed to the formation of some hydroxy complexes of Pb2+
such as Pb(OH)+ and Pb(OH)2 , leading to a decreased Pb2+ concentration, while at lower pH, the abundant H+ ions can protonate the
N atoms of PAAQ and even cause the decomplexation of Pb2+ PAAQ
complex. The H+ ions itself can make interference to the electrode
response simultaneously. Both of them could result in the rise of
the potential.
The wide working pH range of the proposed electrode is comparable to the reported Pb2+ ISEs based on the following ionophores:
[1],
N,N -dibenzyl-1,4,10,13-tetraoxa-7,16-diazacyclooctadecane
hexathia-18-crown-6-tetraone [2], phosphorylated calyx[4]arene
[31], and 5,11-dibromo-25,27-dipropoxycalix[4]arene [32]. The
lower limit of the working pH range is expected to be further
improved in case that the binding ability of Pb2+ onto PAAQ
chains is strengthened. The electrode based on 2,12-dimethyl7,17-diphenyltetrapyrazole with strong complex capability shows
wider working pH range of 1.66.0, whereas the electrode based
on 5,11-dibromo-25,27-dipropoxycalix[4]arene with a relatively
weaker complex capability has narrower pH range of 2.36.0 [32].
3.7. Response and lifetime of the electrode
The response time of the PAAQ ISE was determined by measuring the time required to achieve a steady potential in Pb2+
solution with three concentrations of 1.0 102 , 1.0 103 and
1.0 104 M. It is found from Fig. 8 that the response time
is 12 s, approaching to the fastest response of Pb2+ ISEs based
on N,N -dibenzyl-1,4,10,13-tetraoxa-7,16-diazacyclooctadecane[1]
and N,N -bis(salicylidene)-2,6-pyridinediamine [33] as ionophores.
This fast response is relevant to the fast kinetic process of complexation between Pb2+ and PAAQ ionophore in the doped state. On
the other hand, the ne particles of the PAAQ salts, together with
their higher intrinsic conductivity (5.0 105 S/cm), may promote
the formation of a uniform, thin, and highly conducting composite membrane. All of these features are benecial to the transition
of charges in membrane, resulting in the fast response of the electrode. There was no difference for the potential responses for each
503
concentration recorded from high concentration to low concentration and from low concentration to high concentration, indicating
that the ISE has an excellent reversibility. The standard deviations
of ten replicate potential measurements of 1 mM and 0.1 mM sample solutions were 0.58 and 0.65 mV, respectively, indicating the
good reproducibility of the prepared ISE. It is also observed that
the electrode could be satisfactorily used over a period of 4 months
without any signicant loss of the performance characteristics such
as working concentration range, slope and response time. Dislike
the ordinary ionophore, PAAQ salts existing as ne solid particles do
not leak from the matrix while contacting with aqueous solution,
ensuring a reasonably long lifetime of the proposed electrode.
3.8. Electrode selectivity
The selectivity is one of the most important characteristics for
selective electrodes, because it determines the feasibility and utility of the electrodes in real sample analysis. The potentiometric
selectivity coefcient K POT
has been served to reect the rela2+
Pb
,B
,B
where Pb2+ (DL) is the detection limit of Pb2+ ion, B the activity
of the interfering ion, and z the charge of the interfering ion [34]. It
is noteworthy that the selectivity coefcients for all diverse cations
listed in Table 4 are in the order of 102 or 103 , indicating that
the proposed electrode is highly selective over all the interfering
ions studied. The interferences of alkali metal ions (Na+ and K+ ) and
alkaline-earth metal ions (Ba2+ and Ca2+ ) are almost negligible with
the selectivity coefcients in the order of 103 . Some heavy metal
ions and noble metal ions (Cu2+ , Hg2+ , Au3+ and Ag+ ) might cause a
weak interference, which may arise from the complex capability of
PAAQ to some transition metal ions. It is discovered that aromatic
amine polymers can complex Hg2+ and Ag+ ions as well as Pb2+
ions through the NH2 and NH groups in the chains [2225].
504
Table 4
Potentiometric selectivity coefcients of the PAAQ salt particles-based Pb2+ ISE
obtained by the xed interference method at an interfering ion concentration of
1.0 103 M.
Interfering ions
Hg2+
Ag+
Cu2+
Au3+
NH4 +
Na+
K+
Ba2+
Ca2+
8.51 102
6.97 102
5.82 102
4.68 102
7.59 103
7.18 103
4.27 103
3.55 103
3.09 103
Pb
,B
Table 5
Recovery studies of the PAAQ-based Pb2+ ISE on the detection of lead(II) in tap and
rain waters.
Water sample
Recovery (%)
Tap water 1
Tap water 2
Rain water 1
Rain water 2
50
200
50
200
48.9
205
51.5
195
97.8
102.5
103.0
97.5
[1]
[2]
[3]
[4]
V.K. Gupta, A.K. Jain, P. Kumar, Sens. Actuators B 120 (2006) 259.
M. Shamsipur, M.R. Ganjali, A. Roihollahi, Anal. Sci. 17 (2001) 935.
C.C. Su, M.C. Chang, L.K. Liu, Anal. Chim. Acta 432 (2001) 261.
M.R. Ganjali, M. Hosseini, F. Basiripour, M. Javanbakht, O.R. Hashemi, M.F. Rastegar, M. Shamsipur, G.W. Buchanen, Anal. Chim. Acta 464 (2002) 181.
[5] F. Cadogan, P. Kane, M.A. McKervey, D. Diamond, Anal. Chem. 71 (1999) 5544.
[6] J.Q. Lu, R. Chen, X.W. He, J. Electroanal. Chem. 528 (2002) 33.
505
[21] S. Majid, M.E. Rhazi, A. Amine, A. Curulli, G. Palleschi, Microchim. Acta 143
(2003) 195.
[22] X.G. Li, R. Liu, M.R. Huang, Chem. Mater. 17 (2005) 5411.
[23] M.R. Huang, Q.Y. Peng, X.G. Li, Chem. Eur. J. 12 (2006) 4341.
[24] Q.F. L, M.R. Huang, X.G. Li, Chem. Eur. J. 13 (2007) 6009.
[25] X.G. Li, H. Li, M.R. Huang, Chem. Eur. J. 13 (2007) 8884.
[26] X.G. Li, H. Li, M.R. Huang, China Patent CN1810852, 2008.
[27] M.F. Mousavi, S. Sahari, N. Alizadeh, M. Shamsipur, Anal. Chim. Acta 414 (2000)
189.
[28] A. Rouhollahi, M.R. Ganjali, M. Shamsipur, Talanta 46 (1998) 1341.
[29] Z. Szigeti, T. Vigassy, E. Bakker, E. Pretsch, Electroanalysis 18 (2006) 1254.
[30] K.C. Gupta, M.J. DArc, IEEE Sens. J. 1 (2001) 275.
[31] M.R. Yaftian, S. Rayati, D. Emadi, D. Matt, Anal. Sci. 22 (2006) 1075.
[32] A.K. Jain, V.K. Gupta, L.P. Singh, J.R. Raisoni, Electrochim. Acta 51 (2006) 2547.
[33] T. Jeong, H.K. Lee, D.C. Jeong, S. Jeon, Talanta 65 (2005) 543.
[34] E. Bakker, E. Pretsch, P. Bu1hlmann, Anal. Chem. 72 (2000) 1127.
[35] V.K. Gupta, R. Mangla, S. Agarwal, Electroanalysis 14 (2002) 1127.
[36] V.S. Bhat, V.S. Ijeri, A.K. Srivastava, Sens. Actuators B 99 (2004) 98.