Escolar Documentos
Profissional Documentos
Cultura Documentos
Engineering Div.
Highett, Victoria Australia
PP-T001
1. SCOPE
c)
d)
2.
e)
3. REFERENCED DOCUMENTS
5. INTRODUCTION
This procedure makes reference to the
following documents;
a)
b)
b)
c)
d)
4. DEFINITIONS
a)
b)
PCCP T001 v4
Date of issue 23/02/2012
e)
f)
g)
Page 1 of 9
PP-T001
PAINTING CONTRACTOR CERTIFICATION PROGRAM
TECHNICAL NOTE - CALIBRATION OF INSPECTION & TEST EQUIPMENT
6. REQUIREMENTS
6.1 Dry film thickness gauges
c)
a) AS 3894.3 Determination of Dry Film
Thickness details the use and calibration of
various instruments used for dry film
thickness measurement. The contractor will
be required to have a set of Primary
Standards and a set of Secondary
Standards for use in the field. Both are
shims, usually of plastic or coated metal.
b) Primary (or Master) standards need to be
purchased separately in addition to those
supplied with the instrument. They need to
be kept secure and away from general
access. They are only to be used by the
calibrating officer when instrument checking
is required. Refer Appendix A below.
c) Secondary standards are usually supplied
with the instrument on purchase. Because
they are used daily in the field they are
subject to wear and tear and need to be
checked regularly to ensure continued
accuracy Refer Table 1 section A below.
d)
e)
f)
g)
h)
6.2 Thermometers
a)
b)
c)
d)
b)
j)
k)
l)
PCCP T001 v4
Date of issue 23/02/2012
i)
Page 2 of 9
PP-T001
PAINTING CONTRACTOR CERTIFICATION PROGRAM
TECHNICAL NOTE - CALIBRATION OF INSPECTION & TEST EQUIPMENT
7. CALIBRATION FREQUENCY
a)
b)
8. ACKNOWLEDGMENT
This technical note is based on AS 3894 and
NATA calibration requirements for chemical
testing.
PCCP T001 v4
Date of issue 23/02/2012
Page 3 of 9
PP-T001
PAINTING CONTRACTOR CERTIFICATION PROGRAM
TECHNICAL NOTE - CALIBRATION OF INSPECTION & TEST EQUIPMENT
2. Secondary Standards
(working shims)
Either;
As per Appendix A (9.3B) below
or
Externally using an accredited facility
Either:
External calibration with supplied certificate
3. Primary Standards
(reference shims)
or
Against externally calibrated micrometer
4. Vernier calipers
Either;
As per Appendix A (9.3B) below
or
Externally using an accredited facility
10 yearly
Annually
Annually
B. Humidity measurement
1. Electronic gauge
2. Sling psychrometer
Weekly
Annually
C. Temperature measurement
1. Working (liquid in
glass) thermometers
2. Working electronic
(IR) or other surface
(eg magnetic)
thermometer
3. Reference
thermometer
Annually
Monthly
10 yearly
D. Linear measurement
1. Tape measures
On use
6 monthly
E. Time measurement
1. Digital or analogue
timer
PCCP T001 v4
Date of issue 23/02/2012
Page 4 of 9
PP-T001
PAINTING CONTRACTOR CERTIFICATION PROGRAM
TECHNICAL NOTE - CALIBRATION OF INSPECTION & TEST EQUIPMENT
10. APPENDIX A SUGGESTED CHECKING METHOD FOR DRY FILM THICKNESS
10.1 Scope
This document suggests a method to be used for the verification or checking of the accuracy of an
electronic instrument for the measurement of dry film builds (DFB). Although this method is satisfactory for
PCCP use, there may be other methods that would also provide the same level of surety of readings.
10.2 Apparatus
The following apparatus, plant or equipment is needed to carry out this checking of DFB instrument;
a) An electronic DFB instrument needing checking
b) A set of checking shims (Secondary Standards, typically supplied with the instrument)
c) A set of Primary Standards shims with a valid external calibration certificate.
10.3 Process
A. Daily checking
a) Prior to beginning taking the days first reading on a surface with dry paint film on it, the instrument
needs to be checked according to the manufacturers directions as supplied with the instrument.
b) Frequency of checking shall be as nominated in Table 1 above and shall include a zero check.
c) Repeat at the frequency nominated in Table 1 above.
B. Annual checking
a) Place the instrument, its Secondary Standards and the Primary Standards in the same room with a
constant temperature and allow their temperatures to stabilise for 30 mins minimum.
b) Zero the instrument using its designated probe.
c) Take readings on the first Primary Standard shim. Record the readings and the average reading on a
suitable calibration form/document.
d) Take readings on the closest corresponding thickness Secondary Standard shim. Record the readings
and the average reading on the same form/document.
e) Calculate the difference in readings in microns & percent (using the formula;
100(T1 T2) / T1
where T1 = thickness of the Primary Standard
and T2 = thickness reading obtained on the Secondary Standard).
f)
Repeat steps c) to e) for each of the shims applicable to the range of coating thicknesses likely to be
encountered in jobs the organisation normally undertakes.
g) Repeat at the frequency nominated in Table 1 above.
Page 5 of 9
PP-T001
PAINTING CONTRACTOR CERTIFICATION PROGRAM
TECHNICAL NOTE - CALIBRATION OF INSPECTION & TEST EQUIPMENT
11 APPENDIX B SUGGESTED CHECKING METHOD FOR HUMIDITY METERS
11.1 Scope
This document suggests a method to be used for the verification or checking of the accuracy of electronic
dew meters or hygrometers. External calibration by a suitably accredited organisation is equally acceptable in
technical terms but may not provide the company with sufficient financial savings or turn around times to
satisfy their internal objectives.
11.2 Apparatus
The following apparatus, plant or equipment is needed to carry out this checking of the electronic dew
meter instrument;
a)
b)
c)
d)
e)
f)
g)
11.3 Process
A. Calibration of the sling psychrometer
a) Place some distilled water, ice, the Reference Thermometer and the two thermometers from the sling
psychrometer into a container and allow to stabilise for 10 minutes in a room with constant (ambient)
temperature (the calibration room).
b) Record the readings on each of the three (3) thermometers on a suitable calibration form/document.
c) Remove the thermometers and dry. Place them on a surface in the room within 5cm of each other but
not touching, cover with a clean rag or paper towel to exclude any drafts and allow to stabilise for 10
minutes.
d) Record the readings on each thermometer and any resulting calibration decision on a suitable
calibration form/document.
e) Repeat at the frequency nominated in Table 1 above.
B. Verification of electronic hygrometers
a) Place the hygrometer to be checked in the calibration room and switch on. Allow to stabilise for
15mins.
Note 1: The calibration room to be used can be any suitable room inside a building and complying with 10.2d)
above, if a calibrated sling psychrometer is to be used as the standard instrument. If data from a BoM web site
is to be used a the standard, the calibration room shall be outside in the open but as free of drafts as
practicable.
b) Record the temperature and relative humidity on both the standard instrument and the test
hygrometer.
c) Repeat step 10.3Bb) at least twice during the first day with at least 3h between readings.
d) Repeat steps 10.3Bb) and c) for additional days until a clear picture of the accuracy and consistency is
achieved.
Note 2: A greater number of readings will be required to establish a case for high consistency, typically 50+
readings. If it is clear that the readings are inaccurate and inconsistent, a lower number of readings is
required.
e) Repeat steps 10.3Bb) to d) monthly and record results including verification or otherwise of accuracy
and consistency.
PCCP T001 v4
Date of issue 23/02/2012
Page 6 of 9
PP-T001
PAINTING CONTRACTOR CERTIFICATION PROGRAM
TECHNICAL NOTE - CALIBRATION OF INSPECTION & TEST EQUIPMENT
Note 3: The frequency of checking of accuracy and consistency may be extended where results obtained
show good consistency eg if the inaccuracy has been consistently +4%RH for several months, The checking
frequency can be increased under certain circumstances see clause 7 above.
C. Checking of the instrument
a) Place the electronic dew meter to be checked in the calibration room and switch on.
b) Fill the (in-calibration) sling psychrometer reservoir with distilled water, insert the wick and allow 30
mins for the wick to wet completely.
c) Whirl the sling psychrometer for a period of 3 minutes and record the readings on both thermometers.
d) Repeat step g) for a further 3 minutes and record the temperatures registered. If these are different
from the first record, repeat until a constant reading is obtained.
e) From the Dew Point chart, using the stabilised temperatures, determine the relative humidity (%) and
record.
f) Record the reading of relative humidity from the electronic dew meter.
g) Calculate the difference in readings in percent (using the formula;
100(H1 H2) / H1
where H1 = relative humidity obtained from the sling psychrometer
and H2 = relative humidity obtained from the electronic dew meter)
h) Repeat at the frequency nominated in Table 1 above.
PCCP T001 v4
Date of issue 23/02/2012
Page 7 of 9
PP-T001
PAINTING CONTRACTOR CERTIFICATION PROGRAM
TECHNICAL NOTE - CALIBRATION OF INSPECTION & TEST EQUIPMENT
12.2 Apparatus
The following apparatus, plant or equipment is needed to carry out this checking of thermometers;
a)
b)
c)
d)
12.3 Process
A. Working liquid-in-glass thermometers
i) Place some distilled water, ice, the Reference Thermometer and the working thermometer in a
container so that the thermometers are submerged to at least 75mm of their length.
j) Allow to stabilise for 10 minutes in a room with constant (ambient) temperature (the calibration room).
k) Record the readings on each thermometer on a suitable calibration form/document.
l) Remove the thermometers and dry. Place them on a surface in the room within 5cm of each other but
not touching, cover with a clean rag or paper towel to exclude any drafts and allow to stabilise for 10
minutes.
m) Record the readings on each thermometer on a suitable calibration form/document.
n) Repeat at the frequency nominated in Table 1 above.
B. Infra-red thermometers
a) Place the piece of steel, the reference thermometer and the infra-red thermometer in the calibrating
room. Switch on the instrument and allow everything to stabilise for 30 minutes.
b) Record the reading on the Reference thermometer on a suitable calibration form/document.
c) Use the instrument to take several readings on the steel block and record the readings. Calculate and
record the average reading.
d) Calculate the difference in readings in percent (using the formula;
100(t1 t2) / t1
where t1 = temperature obtained from the Reference Thermometer
and t2 = temperature obtained from the instrument
e) Repeat at the frequency nominated in Table 1 above.
C. Magnetic surface thermometer
a) Use method 11.3B above but place the magnetic surface thermometer into contact with the piece of
steel.
PCCP T001 v4
Date of issue 23/02/2012
Page 8 of 9
PP-T001
PAINTING CONTRACTOR CERTIFICATION PROGRAM
TECHNICAL NOTE - CALIBRATION OF INSPECTION & TEST EQUIPMENT
PCCP T001 v4
Date of issue 23/02/2012
Page 9 of 9