Users Manual

RENEW FACE & BODY

Rev. 0 dated 08/03/2011

2 / 46

RENEW FACE & BODY

ESTETIC ITALIA congratulates you for purchasing a RENEW FACE & BODY
radio frequency unit for aesthetic use.
RENEW FACE & BODY represents the most advanced application of the
latest technologies in radio frequency aesthetic use.

This publication is intended for use as support documentation


Not for sale
No part of this manual may be reproduced in any way without the authorization from ESTETIC
ITALIA.
Prepared by the Engineering Dept.

3 / 46

RENEW FACE & BODY

INDEX
Index
Information on use manual
Typographic conventions
The radio frequency: RENEW FACE&BODY
What is it
Where it acts and effects
General applications modalities in aesthetic
The use in aesthetic
Treatment protocol
Preparing the client
Preliminary treatment protocol
Declaration of responsibility
RENEW FACE&BODY recommended programmes
Duration and costs of treatment
Ideal use
Settings to start a treatment with RENEW FACE&BODY
Editing the settings
To start a FACE treatment
To start a BODY treatment
General information
Description of unit
Control panel
Power supply and output panel
Accessories
Unit set-up
Unpacking
Installing
Prescriptions regarding the installing environment
Maintenance
Hygienisation modality
Work place and storage
Maintenance and controls
Electric safety
Personnel training
Malfunctions
Electromagnetic interferences
Troubleshooting
Technical features
Appendixes
A - Environmental protection
B - Labels
Indications/Contraindications
Conclusions
Warranty
After-sales assistance request form
Declaration of Conformity

4 / 46

Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page

4
5
5
6
6
6
6
6
7
7
7
8
9
9
10
11
11
12
17
24
24
25
26
27
28
28
29
30
31
34
34
35
35
35
35
36
37
38
39
39
40
41
42
42
44
45

RENEW FACE & BODY

INFORMATION ON USE MANUAL


This manual provides information on setting up and using the RENEW FACE & BODY radio frequency unit for
aesthetic use.
It is an indispensable reference guide for the user: it is fundamental to carefully read the content of this manual
before installing and using the units and to always keep it at hand for rapid consultation.
Failure to comply with the instructions in this manual, in whole or in part, may lead to malfunctions and damages
to the unit, resulting in the warranty becoming void.
Only by carefully following the manufacturer instructions and recommendation can users be sure to achieve the
best results and, when necessary, fast and efficient after-sales assistance.

TYPOGRAPHIC CONVENTIONS
Certain parts of the manual are given particular emphasis by underlining.
NOTE
Notes are used to highlight particularly important information.
CAUTION
Caution messages are placed immediately before operations which, if not complied with, may lead to damages
to the unit and/or to its accessories.

! WARNING !
WARNING messages indicate operations or situations which, if ignored or not complied with, may cause
problems to the user.

5 / 46

RENEW FACE & BODY


THE RADIO FREQUENCY: RENEW FACE&BODY
What is it
Radio Frequency treatments are one of the new frontiers in aesthetic medicine. They are successfully
applied as a non-invasive method for skin rejuvenation and achieve results that are very close to those
obtained by surgical facelifts.
This powerful technique is also widely applied in the treatment of cellulite, loose tissues and localized
adiposity.
Radio Frequency treatments are based on emissions of high-frequency radio waves and exploit the highpower thermal energy produced by the tissues when they are exposed to radio waves. The intensity of the
thermal effects depends on the conductivity of the tissues subjected to treatment. Tissues having greater
impedance, such as adipose (fatty) tissues, generate more heat and the thermal effects are consequently
greater.
Radio Frequency treatments are based on the principle that treated areas exhibit an increase in
temperature, conveying energy to the deep dermis and the subcutaneous layers.
This improves cell functions and blood circulation, stimulates the lipolysis (breakdown) of subcutaneous fat
and increases the synthesis of collagen.
Where it acts and effects
Radio Frequency treatments act on the tissues and produce the following results:
- three-dimensional contraction of collagen fibres, with a 360 volume effect
- Drawing of liquids up to the surface
- Improvements in the micro-circulation system
- Removal of toxins
- Synthesis of new collagen
- Tightening of tissues
- Smoothing of tissues (lifting effect)
- Restructuring of the skin
- Increased skin density in the short term
- Oxygenation of tissues
General applications modalities in aesthetic
Radio Frequency treatments can be carried out at any time of the year without upsetting the subjects daily
lifestyle.
The initial effects can be seen immediately and gradually progress during the following weeks.
The use in aesthetic
The Radio Frequency treatments are applied on the following imperfections:
- Body tissue relaxation
- Cellulite
- Fat
- Stretch marks
- Atony of breast muscles
- Aging of face and neck tissues
- Wrinkles
- Scars
- Tissue gaps
- Asphyxial skin
- Acne

6 / 46

RENEW FACE & BODY

TREATMENT PROTOCOL
Depending on the type of problem to be treated, the operator will choose the more suitable program,
following the table on page 9.
Preparing the client
The treatment room must have a comfortable temperature in order to avoid shivers or muscular tremors.
The client must be in as comfortable position as possible; it is recommended, like in other cases of physical
treatment, to explain to the client, prior to treatment, which sensation is to be expected.
Also ensure that you are aware and able to follow all indications relating to safety.

PRELIMINARY TREATMENT PROTOCOL

Estetic Italia proposes the method to rejuvenate, smooth and regenerate tissues, quickly and without
contraindications, thanks to the use of Renew Face & Body.
Before every face and body treatment, the use of the Softlight laser unit is recommended, with
programmes (1-13-16-27) for 15 minutes. This treatment allows to create a vasodilatation which favours
the microcirculation, the collagen synthesis, and notably improves the oxygenation of tissues.
It is recommended to subsequently apply the Bodyson 8 ultrasound unit with gel for 10 minutes, using the
face applicator or body plates which, thanks to the mechanical, thermal and chemical effects, prepares the
tissue causing the permeability and temperature increase of the cellular membrane, as well as the
acceleration of the tissues particles.
This pre-treatment system allows the subsequently applied radio frequency, to act deeply, completely and
in a long lasting manner

7 / 46

RENEW FACE & BODY

IT IS RECOMMENDED TO HAVE THE CLIENT SIGN THE FOLLOWING DECLARATION BEFORE


STARTING THE RENEW FACE&BODY TREATMENT CYCLE:

DECLARATION OF RESPONSIBILITY
I, the undersigned declare, under my responsibility that:

I do not have autoimmune diseases affecting collagen tissues e.g. lupus

I do not have serious decompensated diabetes

I am not pregnant

I am not wearing a pace maker

I do not have Sclerodermia

I do not have moles in the area to be treated

I do not have mycete

I also declare to have been informed of the indications and contraindications of the RENEW FACE&BODY
treatment

Date

SIGNATURE

DURATION AND COSTS OF TREATMENT


We recommend two sessions a week for two months.
-the treatment cost varies from 60 to 150 euro for each single treatment.
The above-said treatment duration is purely indicative.
Only the operator will be able to evaluate and recommend the number of treatments and duration of each
single treatment.

8 / 46

RENEW FACE & BODY

RENEW FACE & BODY RECOMMENDED PROGRAMMES


AREA TO BE TREATED

TOTAL
DURATION
OF THE
SITTING
MINUTES

INITIAL POWER %

Deep toning
Toning
Toning
Toning
Rejuvenating
Deep rejuvenating
Invigorating
Lifting

Face
Peri-ocul. Peri-lab.
Neck
Breast cleavage
Face
Face
Face
Face

30
10
15
20
25
25
25
25

50
50
50
50
50
50
50
50

9
10
11
12
13

Toning
Toning
Toning
Toning
Toning

45
40
45
45
35

14

Toning

50
50
50
50
50
50

15
16

Toning
Rejuvenating

17

Rejuvenating

18

Lifting

Body
Stretch marks
Stomach
Buttocks
Inside thigh
Back and
surroundings
Arms
Stretch marks
Back and
surroundings
Body

NUMBER
OF THE
PROGRAM

TYPE OF TREATMENT

1
2
3
4
5
6
7
8

45
45
40
40
40

50
50
50
50

IT IS RECOMMENDED TO START THE TREATMENT WITH POWER THAT DOES NOT EXCEED 50%,
WHICH WILL THEN BE PROGRESSIVELY INCREASED ON THE BASIS OF THE TREATMENT TO BE
PERFORMED
THE POWER PERCENTAGE WILL VARY DEPENDING ON THE TYPE OF INSERT, SKIN AND AREA TO BE
TREATED.
IT IS THEREFORE RECOMMENDED TO KEEP THE APPLICATOR OR THE EARTHING PLATE AS NEAR
AS POSSIBLE TO THE AREA TO BE TREATED, SO THAT THE IDEAL TEMPERATURE IS REACHED
QUICKLY.

9 / 46

RENEW FACE & BODY

IDEAL USE

! WARNING !
Operators must make sure that the electric connections have been properly made before turning on the main
power switch
.
! WARNING !
Operators must make sure that the application plates
of the inserts/electrodes are in order before turning on the main power switch and starting a session.

After having applied the cable for connecting the adaptors to the appropriate connector, insert the power plug
into the wall socket (230 Vac) and activate the unit by switching the main ON/OFF switch on the back panel to
"ON".
This will turn on the RENEW FACE & BODY unit and the back-lit LCD will light up, meaning that the device is
ready for use.
The LCD display will light up showing the logo:

Fig.1

10 / 46

RENEW FACE & BODY

Press the encoder knob to go to the first page (Fig. 2). Here you can choose between three operating modes by
pressing the relative function key.

Fig.2

SETTINGS TO START A TREATMENT WITH RENEW FACE&BODY


TO EDIT THE SETTINGS
If wanting to edit and save the basic settings which will be automatically recalled every time the unit is switched
on, in the internal memory, press the function key relating to the SETTINGS button (see fig. 2), the following
screen will be loaded (see fig. 3):

Fig. 3
Press the function key relating to the MODIFY button to toggle between the screens shown in fig. 3 and fig. 4:

11 / 46

RENEW FACE & BODY

Fig.4
Here you can choose which parameters to edit among those available (LCD brightness and contrast) for
adjusting the displays visibility to the local conditions.
Select the parameter to edit and turn the Encoder knob clockwise to increase its intensity or anticlockwise to
reduce it.
Press the function key relating to the EXIT button to return to the page shown in fig. 2.
The page also shows the version of the software currently installed on the unit.
STARTING A FACE TREATMENT
Press the function key relating ot the FACE TREATMENT button (see fig. 2), the following screen is loaded with
the indications on preparing the subject undergoing the treatment (see fig. 5):

Fig.5
.

12 / 46

RENEW FACE & BODY


Turn the encoder knob clockwise to go to more pages containing information on how to proceed (see fig. 6, 7
and 8):

Fig. 6

Fig. 7

13 / 46

RENEW FACE & BODY

Fig. 8
Once you reach the last page (see fig. 8), turning the encoder knob anticlockwise will show the information
pages in reverse order.
These pages can be used as an on-line Help and provide operators with the basic information required before
starting a treatment session, for example:
-

The procedure to follow to carry out the selected treatment


Information on any contraindications for the selected treatment
What persons should not be subjected to the selected treatment

After following the instructions in the previous screens, press the encoder knob to set the treatment parameters.
The following screen will appear (see fig. 9):

Fig. 9

14 / 46

RENEW FACE & BODY

Press the function key relating to the MODIFY button to toggle between the screens shown in fig. 9 and fig. 10:

Fig. 10
On these screens you can set the parameters for a standard treatment, such as:
- The issued power (see fig. 9), which is set by default at 50%; operators can change the value before or
during the treatment
-

The treatment time (see fig. 10), which is set by default at 20 minutes; operators can change the value
before but not during the treatment

Once these parameters are set, press START to start the treatment.
On the screens shown in fig. 9 and fig. 10 there are two boxes that show a series of dots before you press
START: this is read-only value that cannot be edited by the operator.
Press START to start the treatment: the treatment starts as shown by the countdown (see fig. 11).

Fig. 11
The unit will periodically check the treatment temperature and will continue to work until reaching the
temperature set by the operator through the power.

15 / 46

RENEW FACE & BODY

CAUTION
When carrying out facial treatments the face applicator can only be used
with the electrodes having diameter 30 mm or 40 mm.

When the applicator comes into contact with the part being treated, a power will be detected (expressed as a
percentage) whose numeric value will replace the previously met dots.
The above-mentioned power represents the power actually absorbed by the RENEW FACE & BODY unit as
referred to the maximum power the unit is capable of supplying (equal to 100% of the power parameter).
When a treatment is under way, pressing the STOP button once will pause the unit and temporarily interrupt the
treatment.
Following which (see fig. 12):
- The operation phase field shows the message "In pause..."
- the measured field shows a series of dots instead of the temperatures numeric value, as the treatment
is currently suspended

Fig. 12
The operator can change the settings if necessary and decide whether to resume the treatment or definitively
end it.
Press START to resume the treatment: the unit will resume the treatment from the point at which it was
interrupted.
To definitively end the treatment press STOP again: the treatment will be ended and the remaining time
parameter will revert to the value shown in the Preset time parameter (see Fig. 10).

16 / 46

RENEW FACE & BODY


The body part to be treated must be covered with an abundant coating of contact "GEL" or cream (which
enhances the sound conduction of the radiant energy). The operator must distribute the radiant energy by gently
massaging the body part, i.e. by moving the radiating head over the part in order to attain an equal distribution
of the energy dose.
CAUTION
For the correct outcome of the application, it is fundamental that the subject
undergoing treatment tightly holds an earthing applicator (also called
mass or reference applicator) during treatment and that the same is as close as possible to the part to be
treated.

STARTING A BODY TREATMENT


Press the function key relating to the BODY TREATMENT button (see fig. 5) to load the following screen (see
fig. 13):

Fig. 13
this page contains information on how to prepare the subject for treatment.
Turn the Encoder knob clockwise to go to more windows containing information on how to proceed (see fig. 14,
15, 16, 17, 18 and 19):

17 / 46

RENEW FACE & BODY

Fig. 14

Fig. 15

18 / 46

RENEW FACE & BODY

Fig. 16

Fig. 17

19 / 46

RENEW FACE & BODY

Fig. 18

Fig. 19
Once you reach the last screen (see fig. 19), turning the encoder knob anticlockwise will show the information
screens in reverse order.
These pages can be used as an on-line Help and provide operators with the basic information required before
starting a treatment session, for example:
-

The procedure to follow to carry out the selected treatment


Information on any contraindications for the selected treatment
What persons should not be subjected to the selected treatment

20 / 46

RENEW FACE & BODY


After following the instructions in the previous screens, press the encoder knob to set the treatment parameters.
The following screen will appear (see fig. 20):

Fig. 20
Press the function key relating to the MODIFY button to toggle between the screens shown in fig. 20 and fig. 21:

Fig. 21
On these screens you can set the parameters for a standard treatment, such as:
-

The issued power (see fig. 20), which is set by default at 50%; operators can change the value before or
during the treatment

The treatment time (see fig. 21), which is set by default at 10 minutes; operators can change the value
before but not during the treatment
The value set as skin temperature is the temperature that will be reached on the part of the face subjected to
treatment, while the treatment time is the time for which the selected treatment is intended to last.
-

On the screens shown in fig. 20 and fig. 21 there is one boxes that show a series of dots before you press
START: this is read-only value that cannot be edited by the operator.

21 / 46

RENEW FACE & BODY


All the parameters can be changed before starting a treatment session.
Once these parameters are set, press START to start the treatment. The following screen will appear showing
the countdown (see fig. 22) .

Fig. 22

The unit will periodically check the treatment temperature and will continue to work until reaching the
temperature set by the operator.
When the applicator comes into contact with the part being treated, a power will be detected (expressed as a
percentage) whose numeric value will replace the previously met dots.
The above-mentioned power represents the power actually absorbed by the RENEW FACE & BODY unit as
referred to the maximum power the unit is capable of supplying (equal to 100% of the power parameter).
When a treatment is under way, pressing the STOP button once will pause the unit and temporarily interrupt the
treatment. Following which (see fig. 23):
-

The operation phase field shows the message "In pause..."


the measured field shows a series of dots instead of the temperatures numeric value, as the treatment
is currently suspended

22 / 46

RENEW FACE & BODY

Fig. 23
The operator can change the settings if necessary and decide whether to resume the treatment or definitively
end it.
Press START to resume the treatment: the unit will resume the treatment from the point at which it was
interrupted.
To definitively end the treatment press STOP again: the treatment will be ended and the remaining time
parameter will revert to the value shown in the Preset time parameter (see Fig. 21).
The body part to be treated must be covered with an abundant coating of contact "GEL" (which enhances the
sound conduction of the radiant energy). The operator must distribute the radiant energy by gently massaging
the body part, i.e. by moving the radiating head over the part in order to attain an equal distribution of the energy
dose".
CAUTION
For the correct outcome of the application, it is fundamental that the subject
undergoing treatment has the earthing plate applied to an area of the body (also called
mass or reference plate) during the same treatment.
GENERAL INFORMATION
EME srl has recently developed a complete series of devices, accessories and equipment designed and built
according to the highest standards, applying the state-of-the-art technology, in full compliance with the directives
and regulations in force.
Special attention has been devoted to design, user-friendliness, functionality and safety.
The result is a compact unit, equipped with a modern line, able to offer an extremely logical operating sequence,
supported by an easy-to-read display.
The many possible aesthetic applications and the absolute safety for subjects and operators (the unit complies
with all relevant international standards) put the RENEW FACE & BODY radio frequency unit for aesthetic use at
the top of its category.

23 / 46

RENEW FACE & BODY

DESCRIPTION OF UNIT
RENEW FACE & BODY is a radio frequency unit for aesthetic use which uses, inside the body, a 470 KHz
frequency emission without causing sensory stimulation of the motor nerves or causing muscle contractions and
discomfort.
The biological effect of the heat produced by Radio Frequency waves on the skin consists in the denaturation of
collagen fibres (5 to 20% of total fibres), which cause their immediate contraction and a progressive effect lasting
for the following 4 to 6 months.
As required by Van't Hoffs principle, any reaction whose speed depends on heat is accelerated by an increase
in temperature. The increase in the metabolism depends directly on the increase in heat.
The increased metabolism causes an increase in the demand for oxygen and nutrients and the production of
catabolites and metabolites.
The heating of the body parts subjected to treatment produces various results:
-

Improvement and stimulation of local blood circulation


Reabsorption and drainage of liquids in excess
Reabsorption of local edemas (swellings)
Removal of fat deposits
Reduction of cellulite
Improvement of tissue elasticity

The unit for RENEW FACE & BODY radio frequency for aesthetic use
The RENEW FACE & BODY radio frequency unit for aesthetic use is equipped with a control panel optimized for
its specific field of application and for the functions for which the unit is intended.
All operation parameters are controlled in real time by a sophisticated electronic microprocessor circuit and the
functions are clearly shown on a backlit LCD display and indicated by specific sound signals.

24 / 46

RENEW FACE & BODY

CONTROL PANEL
Figure 1 shows the front panel with its controls and indications and their relative meanings.
Fig. 1 : Front control panel

Meanings:
1.
2.
3.
4.
5.
6.
7.

Name of unit
LCD Graphic Display
Left side function buttons
Right side function buttons
Output START button
STOP / PAUSE button
Encoder knob for: selecting / confirming button

Buttons 3) and 4) have the function that is shown on the display at any given time.

25 / 46

RENEW FACE & BODY


POWER SUPPLY AND OUTPUT PANEL
Figures 2 and 3, together with the respective meanings, show the panel containing the integrated power supply
module (main switch, fuse box and connector for the network power cable) and the output connectors.
Fig. 2 : Power supply panel

Meanings:
1. Double-pole main ON/OFF switch
2. Fuse box
3. Three-pole (male) socket for network power cable
4. Ground terminal
Fig. 3 : Output panel

CAUTION
The connections between the applicators
and their respective connectors are
compulsory: one applicator will not fit into
the others connector and vice versa.

Meanings:
5. Output connector hole
face applicator/body applicator
6. Output connector hole
earthing applicator/metal plate

26 / 46

RENEW FACE & BODY


ACCESSORIES
The unit is compatible with the following supplied accessories:

1
1
1
1
1
1
1
1
1
1

face applicator (ACC1308)


body applicator (ACC1307/1)
30mm diam. insert/electrode for differential treatments (ACC1307/3)
40mm diam. insert/electrode for differential treatments (ACC1307/4)
50mm diam. insert/electrode for differential treatments (ACC1307/5)
60mm diam. insert/electrode for differential treatments (ACC1307/6)
70mm diam. insert/electrode for differential treatments (ACC1307/7)
260 ml bottle of gel (ACC1311)
metal plate (21x8 cm) (ACC1310)
earthing applicator (ACC1309)

All the above accessories make up a kit for radio frequency units.

The body applicator can be fitted with any of the inserts/electrodes supplied for body treatments.
The face applicator can be fitted with the following supplied inserts/electrodes:

27 / 46

RENEW FACE & BODY

30mm diam. insert/electrode


40mm diam. insert/electrode

for facial treatments.


Both the body applicator and the face applicator are easy to connect: simply plug the applicator cable into the
connector on the units back panel.
Apply the metal plate to the body part opposite the part being treated. The plate is usually placed beneath the
buttocks, the back or the thigh.
Make sure the plate is in close contact with the subjects body: if the plate is too loose the subject might feel
uncomfortable or warming on the corresponding body part.
The earthing applicator is a handle that must be held tightly by the subject being treated; its use is recommended
when carrying out facial treatments.
The earthing applicator and the metal plates are easy to connect: simply plug the applicator cable into the
connector on the units back panel.

UNIT SET-UP
UNPACKING
The RENEW FACE & BODY radio frequency unit for aesthetic use comes packed in a box filled with packing
filler and designed for easy storage and safe transportation.
To unpack the unit, put the box on a flat, solid surface and remove the polystyrene top.
Carefully take out the unit.
The unit and its accessories are wrapped in protective sheets of transparent polyethylene. The packing also
contains:

Use Manual;
n. 1 cable for network power supply;
n. 2 spare fuses (see technical features).

Check the contents of the box.


If any item is missing, immediately contact your authorized EME dealer.
CAUTION
Do not discard the original packing : it will be necessary in case of return shipments.

INSTALLING

28 / 46

RENEW FACE & BODY


The RENEW FACE & BODY radio frequency unit for aesthetic use is simple to install and requires no special
procedures.
The recommended environmental features for installing RENEW FACE&BODY are the following:
- ambient temperature:
from +10C to +40C;
- relative humidity:
from 10% to 80%, with no condensate;
- avoid exposure to direct sunlight, chemical products and vibrations.
CAUTION
Do not use RENEW FACE & BODY in places where it could get wet.

! WARNING !
RENEW FACE & BODY UNIT IS GUARANTEED TO WORK CORRECTLY,
RESPECTING THE INSTALLATION AND USE INSTRUCTIONS INDICATED,
ONLY WITH THE USE OF ORIGINAL ACCESSORIES AND SPARE PARTS.
In case of problems or difficulties during installation
contact EME srl Customer Service.

CAUTION
Do not use accessories other than the ones provided: they might damage
the unit, causing the warranty to become void.

The back of the unit is fitted with an integrated module for network power supply, consisting of a three-pole
connector for the power cable, a removable fuse box with two fuses (see technical features) and a double-pole
main switch.
Connect the power cables three-pole female plug into the integrated module and make sure it is perfectly inside
the connector.

! WARNING !
Before connecting the cable to the network plug, make sure the unit has not been
damaged during transport and ensure the electric energy supply features
on the available current socket, satisfy the data on the plate, reported at the back of the unit.
! WARNING !
The electricity powering the unit is VERY DANGEROUS.
Be sure the power cable is disconnected from the mains
before you connect or disconnect it from the connector on the unit.
! WARNING !

29 / 46

RENEW FACE & BODY


DO NOT INSTALL RENEW FACE & BODY unit close
to heat sources or in direct sunlight.

! WARNING !
For safety reasons, the power cable is supplied with a plug with protective earthing.
Use only adequately grounded power sockets.
The unit must be connected only to power systems complying with safety regulations.
If any extensions are necessary, make sure they are provided with an efficient earthing protection conductor.
Failure to observe this recommendation may lead to the accidental electrocution
of persons and affect the units operation.

CAUTION
If an extension is used between the RENEW FACE & BODY unit and other units,
check that the total current absorption of the connected units does not exceed the
maximum current allowed for that type of cable or however not higher than 15A.

Instructions for the installing environment


The room where the unit will be installed must be:

equipped with electric plant in compliance with the regulations in force


correctly screened and realised complying with safety regulations
correctly identified at the entrance, with visual indications which highlight the type of units installed inside
controlled and access to it limited, correctly indicating the authorised personnel who has access
Also, no life-monitoring devices must be present in the room

! WARNING !
This unit is not suitable for use with flammable liquids
with air and oxygen or with nitrogen protoxide, do not install in such rooms.
! WARNING !
Avoid installing and/or leaving the unit near sources of heat,
exposed to rain, humidity, or where it might get wet.

For the safety of the client and the operator, the following obligations must be respected:

30 / 46

RENEW FACE & BODY

Avoid using extensions


Avoid using double or triple plugs
Position the unit with its air inlets at least 20 cm away from any wall
Position the unit away from sources of heat
Position the power cables so as to avoid they are crushed or stepped on
Do not keep liquid containers near the unit
CAUTION

Before switching the unit on make sure that:


It is connected to a 220-240V 50 Hz current socket ( 110-120 V 60 Hz upon request )
Any air inlets are not totally or partially obstructed
The fuses are in their casing

MAINTENANCE
The RENEW FACE & BODY radio frequency unit for aesthetic use does not require special maintenance. Clean
the applicators regularly to ensure ideal operation and safety.
! WARNING !
For safety reasons
before servicing or cleaning the unit,
YOU MUST
turn off the rear main switch
and disconnect the power cable from the mains.

To clean the outside of the unit only use a soft cloth dampened in hot water, or a non-flammable, non-abrasive
liquid cleaning agent.
Clean the front control panel in the same way.
The applicators, and especially the treatment head, should be regularly cleaned using water and denaturised
alcohol.

31 / 46

RENEW FACE & BODY

Store the applicators with care after every treatment session.


For information on original accessories and spare parts please contact EME srl authorised dealers.
CAUTION
Handle the applicators with care:
any misuse may affect their performance and features.

! WARNING !
Operators should be informed about the need to carry out periodic maintenance on the applicators. In particular:
- inspect the treatment head for any cracks that might let in the conducting fluid;
- make sure the applicators cables and connectors are in good conditions.

CAUTION
The applicators may be opened and/or disassembled only by personnel authorized by EME srl
:
tampering will cause the immediate voiding of all warranty rights,
as well as damaging the applicator.
CAUTION
Do not use thinners, detergents, acid solutions, aggressive solutions or flammable liquids
to clean the external parts of the unit and accessories.
Using these substances, or misusing the accessories,
will cause the immediate voiding of all warranty rights
, as well as irreparably damaging the unit and the electrodes.

Do not spray or pour liquids on the external casing of the RENEW FACE & BODY radio frequency units for
aesthetic use, nor on the air vents.
Do not dip the unit in water.
Do not dip the applicators or cables in liquid solutions: their immersion in such solutions will cause the
deterioration of the electronic components and comprise the insulation of the electric cables.
After cleaning the outside of the box, perfectly dry all parts before starting the unit again.
Do not under any circumstances disassemble the unit in order to clean it or inspect it: there are no parts that
need cleaning inside the unit, and at any rate RENEW FACE & BODY radio frequency units for aesthetic use
may be opened only by specialized personnel authorized by EME srl.

32 / 46

RENEW FACE & BODY


CAUTION
For safety reasons, before servicing or cleaning
IT IS NECESSARY TO
turn off the main switch
and disconnect the power cable from the mains.
NOTE
For an ideal use of the unit and to guarantee the best performance,
we recommend to correctly carry out maintenance in the times and with the prescribed methods.
NOTE
The ordinary maintenance must be carried out on the device
before start up to verify its integrity and safety.
In particular, before start up it is necessary to:

check that the power supply cable is intact and connected to the electric network,
visually examine the whole unit for any damages.
Check there is no humidity near the device, particularly on the connectors.
At the end of the treatment, switch off the unit and disconnect the power supply cable if the unit is not to be
used immediately.
! WARNING !
Should there be damages on the cable, replace it before start up
of unit with one the same as that provided.

The RENEW FACE&BODY system correctly installed, in good maintenance state and used according to
regulations, does not present dangers, neither for operator nor for the client during his therapeutic work.

Before using such system check the network voltage is 230 V, 50 Hz.
Avoid placing objects on the unit.
Avoid installing or leaving the system near sources of heat, or exposed to rain or humidity
If any object should penetrate inside or liquid be poured on, immediately disconnect the unit from the network
and have the specialised personnel check it, before using it again.
Clean the external casing with a soft cloth, eventually slightly damp (not dripping) with water and neutral
detergent or alcohol or products for furniture; do not use abrasive products or solvents.
Do not, for any reason, open or disassemble the unit.
Periodically check the state of the cable and plug, in case of wear, replace them.

Some verifications can be made by the personnel operating on the device, whereas others, more technical, must
be carried out by trained personnel and authorised in writing by the manufacturer.
To guarantee the safety and the correct functioning of the unit, respect the following maintenance program.

33 / 46

RENEW FACE & BODY

ACTION

COMPANY

Cleaning and hygienisation of container


Cleaning / hygienisation of accessories in direct
contact with client
Visual check of power supply cable integrity (in case
of deterioration contact the technical assistance
service)
Visual check of applicator cable integrity (in case of
deterioration contact the technical assistance
service)

Operator
Operator

CADENCE
Daily
every session with different client

Operator

Weekly

Operator

Weekly

! WARNING !
RENEW FACE&BODY is an electronic device, therefore to guarantee its safety
the periodic maintenance will have to be carried out exclusively at the manufacturer
laboratory or by adequately trained personnel, and authorised in writing by the manufacturer.

Hygienisation mode
The device must be hygienised periodically. Disconnect the power supply plug from the socket.
Do not use compressed air or nebulising products; any dripping of any liquid can compromise the units
functioning.
In case of accidental liquid or foreign materials infiltration inside the casing, do not attempt to open the device to
remove the penetrating material, but contact the technical assistance service and do not use the unit.
Cleaning must be carried out using non abrasive materials dampened with non corrosive hygieniser products.

Work place and storage


The device has been designed to function in an ideal environment, like: rooms set up for aesthetic use.
Rooms with strong temperature changes, high humidity level, dripping of water or any other liquid, dusty
atmosphere or polluted by chemical agents, materials or flammable gasses, are to be considered unsuitable both
for the functioning and for the storing of the device.
The unit can be used in the presence of other devices normally used in the above listed rooms.
The use of the unit in rooms polluted by radio-frequencies, electro-magnetic fields (mobile telephones, etc.),
ionising radiation, mechanical and/or sound vibrations and, anyhow, in the presence of devices not normally
used in medical, physiotherapy and aesthetic environments and not provided with the electro-magnetic
compatibility requisites, may cause anomalies to the device in question, making it dangerous for the operator
and the client to use it.

34 / 46

RENEW FACE & BODY


Maintenance and controls
The trials and testing carried out at our offices, represent the peculiar premise to satisfy the safety requirements
of an electro-medical unit.
To guarantee client and operators safety, the latter have to request the technical assistance service to
periodically carry out safety checks and maintenance interventions on the units in their possession.
According to the current standard, to obtain good reliability and efficiency, it is necessary to respect the following:

The installation, the set up, periodical checks or repairs are carried out by the manufacturer or personnel
authorised by him;
The electric plant is in compliance with the safety regulations;
The unit is used according to the use manual (keep the manual in a safe place);
The unit is used with the accessories supplied by the manufacturer.

Electric Safety
According to the current C.E.I. 62.5 standards, the safety, the reliability and the performance of the system can
be guaranteed only if:

Assembly, set up or repairs are carried out by the manufacturer or persons by him authorised.
The electric plant of the room in which it is installed, is in compliance with the respective regulations.
The unit is used in compliance with the use instructions.

Measuring of the dispersion currents and resistance of the protection conductor and checking the integrity of the
part connected to the network, is to be carried out every two years, according to the C.E.I. n.12764 dated
November 1989 publication.

PERSONNEL TRAINING
The RENEW FACE & BODY unit should be used only by operators who have been professionally trained to
carry out perfectly safe treatments.

MALFUNCTIONS
The RENEW FACE & BODY radio frequency units for aesthetic use have been designed and built applying
advanced technology and high-quality components in order to ensure efficient and reliable continuous use.
However, if any malfunction should occur, we recommend consulting the following guide before contacting an
authorized service centre.
CAUTION
Only trained personnel can access the internal parts of the unit
authorised by the manufacturer.
For repairs or information please contact EME srl
or an authorized service centre.

35 / 46

RENEW FACE & BODY

! WARNING !
DO NOT OPEN THE RENEW FACE & BODY unit: the inside contains
HIGH VOLTAGE ELECTRICITY which can be DANGEROUS.

Should any of the following events occur, disconnect the unit from the mains and contact EME srl customer
service:

the cable or the integrated rear power supply module are damaged or show signs of wear;
a liquid has penetrated inside the unit;
the unit was left in the rain.

ELECTROMAGNETIC INTERFERENCE
The RENEW FACE & BODY radio frequency units for aesthetic use have been designed and built in compliance
with European Directive 89/336/EEC on ELECTROMAGNETIC COMPATIBILITY with the purpose of providing a
reasonable degree of protection against harmful interferences in homes or health centres. The EC marking on
the unit guarantees the units conformity to the Directive.
All the necessary tests and measures were carried out in EMEs Internal Lab for Tests, Measures and Trials
(LPMC) and at specialized external centres. Upon request, customers have been allowed to view the EMC test
reports on the companys premises.
On the grounds of their operating principle, RENEW FACE & BODY radio frequency units for aesthetic use
produce an acceptable amount of radio frequency energy and have an adequate degree of immunity from
radiating electromagnetic fields: In these conditions, no interferences can occur that are damaging for radioelectric communications, electromedical equipment used for monitoring, diagnostics, treatments and surgery,
office equipment such as computers, printers, photocopiers, fax machines, etc., and for any other electric or
electronic device used in those places, provided they comply with the EMC Directive on ELECTROMAGNETIC
COMPATIBILITY.
However, in order to prevent any possible interferences, radio frequency units should be used at a safe distance
from critical life-monitoring devices and treatments carried out on subjects carrying pacemakers should be
performed with caution.

36 / 46

RENEW FACE & BODY

TROUBLESHOOTING

PROBLEM

POSSIBLE CAUSE
Plug not completely inserted in power
socket.

Check plug and socket are working.

Power cable not completely inserted in


the unit connector.

Insert plug and cable completely into the unit


connector.

Neither the lights nor the


Power cable worn and cut.
LCD display on the front
panel turn on: the unit
does not work.
Power switch off.

Some controls on the


front control panel are
not working properly.

The unit does not emit


heat.

The unit seems to work


regularly, but treatment
is significantly less
efficient.

SOLUTION

Replace power cable (or contact an EME srl


service centre).
Turn power switch on.

Fuse or fuses damaged or blown.

Replace missing, faulty or blown fuse(s).

Failure in electronic control


circuit.

Contact an EME srl service centre.

Faulty keys or buttons.


Failure in electronic control circuit.

Contact an EME srl service centre.

Faulty connections in output circuit


applied by the operator.

Accurately check output connections.

Applicator cable cut or improperly


connected.
Output cable worn and/or with faulty
contacts.

Replace faulty applicator showing signs of


wear on head and cable.

Failure in electronic current


generator circuit.

Contact an EME srl service centre.

Faulty connection on applicator output


circuit.
Applicator piezoelectric transducer
damaged.

Carry out prescribed maintenance. Install


and position unit as described.
Make sure applicator cable and connector
are in good conditions.

Mechanical damages to applicator


(following fall or violent impact), in
particular to radiating head.

Make sure the radiating head adheres


closely to the part being treated.

37 / 46

RENEW FACE & BODY

TECHNICAL SPECIFICATIONS

RENEW FACE & BODY


230 Vac, 50-60 Hz, 10%

Power supply from mains

115 Vac, 50-60 Hz, 10% *


Max. absorbed power

250 VA

Double protection fuse on power supply


Backlit LCD for displaying and controlling
the operative parameters

230 Vac

1.6A Ret. - 5 x 20 mm

115 Vac

3.15A Ret. - 5 x 20 mm

graphic 320 x 240 pixels

Hazard class

Programmable treatment time

up to 30 minutes

Applicator emission frequency

470 KHz 10%

Temperature

(3445)C 1C

Output channels

Wheeled container, external dimensions

39x30x92h cm

Unit weight

32 Kg

* on request

38 / 46

RENEW FACE & BODY


APPENDIXES
Appendix A
ENVIRONMENTAL PROTECTION
The RENEW FACE & BODY radio frequency units for aesthetic use have been designed to have the smallest
possible impact on the environment.
Design criteria included minimizing waste, toxic materials, noise, undesired radiations and energy consumption.
Particular efforts were put into optimizing the units performance for energy saving purposes.

This symbol means that the product should not be disposed of as household waste.
When the time comes to discard the units, users should hand them over to collection centres for the recycling of
electric and electronic equipment.

39 / 46

RENEW FACE & BODY

Appendix B
LABELS

Nameplate
(on back of unit)

Label next to equipotential connection (earthing)

40 / 46

RENEW FACE & BODY

INDICATIONS
The RENEW FACE & BODY radio frequency unit for aesthetic use is applied on the following imperfections:

Body tissue relaxation


Cellulite
Fat
Stretch marks
Atony of breast muscles
Aging of face and neck tissues
Wrinkles
Scars
Tissue gaps
Asphyxial skin
Acne

CONTRAINDICATIONS

Autoimmune diseases affecting collagen tissues, ex. lupus

Serious decompensated diabetes

pregnancy

Holders of pacemakers

sclerodermia

moles

mycete

treatment on central area of breasts

In case of doubt, consult a physician.


NOTE: Do not use cellphones near radio frequency devices when they are in use, as they may cause
electromagnetic interferences.

CONCLUSIONS
As with every treatment, mechanical or manual, the responsibility of eventual consequences on the subject
being treated is of the operator, who must be properly trained on the use of the ESTETIC ITALIA units.
The Company is lifted from any responsibility, in case of consequences and/or erroneous applications of
the units, on the treated subject.
In case of need, we invite the user to contact an ESTETIC ITALIA authorised dealer.

41 / 46

RENEW FACE & BODY

WARRANTY
EME srl guarantees the quality of its units for a period of 24 months from the date of purchase, provided they are
used in accordance with the instructions contained in this manual.
During the warranty period, any faulty products will be repaired or replaced by the manufacturer at its own
discretion.
Units will not be replaced under any circumstances.
The warranty does not cover malfunctions or damages caused by:

inappropriate positioning, installation and set-up;


improper use or use non conforming to the instructions contained in this manual;
improper or inadequate maintenance performed by user;
operation non conforming to the environmental requirements indicated for the product;
unauthorized opening of external casings;
unauthorized tampering and/or modifications;
using non-original spare parts.

The warranty is effective free delivered at the establishment of EME srl.


In case of return shipments follow the packing instructions below.
Attach a copy of the purchase receipt.
Insuring the shipment is recommended.
We recommend consulting the chapters MAINTENANCE and MALFUNCTIONS before returning the unit
because of suspected malfunctions. difficulties are often caused by inadequate maintenance or technical hitches
that can be easily solved by the user.
A simple phone call to EME srl Customer Service can provide effective help in identifying and solving problems.
CAUTION
Customers shall be responsible for damages caused by inadequate packing.
When repacking, use the original packing material if possible.

Instructions for packing and returning the unit:

disconnect the power cables and the cables connecting the applicators, etc.;
carefully clean and disinfect all the parts and accessories that have been in contact with the subject.
For obvious sanitary reasons protecting the health of technical personnel (Directive concerning health in the
workplace, Law no. 626 dated 19 Sept.1994), any devices considered hygienically unsafe will not be
examined by the acceptance staff;

remove all accessories and any mechanical supports;


use the original box and packing materials;
attach the Assistance Request Form and state the causes for requesting an inspection, along with the
problems or malfunctions observed. this information will be very useful to our personnel and will
considerably shorten the time required for repairs.

42 / 46

RENEW FACE & BODY

AFTER-SALES ASSISTANCE REQUEST FORM (rev. 0 dated


10/01/2007)
(must be completely filled in!)

COMPANY

PRODUCT

CONTACT PERSON

TEL

E-MAIL

FAX

SERIAL
NUMBER

PURCHAS
E DATE

ENCOUNTERED FAULT

APPROV.

Rules and regulations for after-sales assistance:


The goods will have to be sent FREE PORT to our offices of VIA DEGLI ABETI, 88/1 61100 PESARO, complete with accessories necessary for
working. In case the material is damaged, it will be returned to the sender out of warranty, debiting the same with the transport expenses for
returning the goods. The packing must be ideal for transporting fragile material and all electronic devices must be packed in an antistatic product.

Necessary condition for the sending of the goods is the sending of the present FORM completely filled in, to be attached to the
shipment, possibly together with a detailed TECHNICAL REPORT.
If not, we will not be able to solve the problem.

Stamp and signature as acceptance of regulations

Requested date

_______________________________________

____________

Eme s.r.l. Via degli Abeti, 88/1 61100 Pesaro PU


Tel. +39-0721-400-791 Fax +39-0721-26-385
www.eme-srl.com info@eme.srl.com

43 / 46

RENEW FACE & BODY

DECLARATION OF CONFORMITY

The manufacturer
EME s.r.l. via degli Abeti 88 / 1
61100 Pesaro (PU) ITALY
DECLARES UNDER ITS OWN RESPONSIBILITY THAT THE PRODUCT:
Unit for radio frequency treatments models:

RENEW BODY RENEW FACE & BODY

IS IN COMPLIANCE WITH THE FOLLOWING DIRECTIVES:


SAFETY OF ELECTRIC UNITS FOR AESTHETIC USE
regulations:
IEC 62-39
connected regulations:
IEC 62-5
IEC 62-24
BT 73/23/EEC (mod. 2006/95/EC)
ELECTROMAGNETIC COMPATIBILITY
Directive EC 2004/108 - EEC 92/31 EC 2006/95
regulations:

IEC 62-50

Electromagnetic emission conducted on the power supply:

Radiating electromagnetic emissions:


Immunity to electromagnetic shock:
Immunity to radiating electromagnetic fields:
Immunity to fast transient: (burst 5/50 msec
on the power supply lines)

EN 55011 group 2
class A (IEC 110.6)
EN 61000 4 1
EN 61000 4 2
EN 61000 4 3
EN 61000 4 4

IT IS ALSO DECLARED THAT THIS UNIT IS NOT DESTINED FOR MEDICAL USE, IS CLASS I AND IS
DENOMINATED "ELECTRIC UNIT FOR AESTHETIC USE" (62/39- 10/1992)

EC Mark for electromagnetic compatibility


Pesaro , 15-01-2008

44 / 46

RENEW FACE & BODY

45 / 46

RENEW FACE & BODY

Via degli Abeti 88/1 61100 Pesaro Italy


Tel. +39.0721.400791 (6 lines) Fax +39.0721.26385
www.eme-srl.com

46 / 46