Escolar Documentos
Profissional Documentos
Cultura Documentos
OF
GAS CHROMATOGRAPHY
The
The
The
use of gases
Gases
Sample
Introduction
Column
Temperature
Control
Qualitative
Analysis
Detector
Electronics
Output
Device
Quantitative
Analysis
GASES
Definitions
Gases
Carrier Gas
introduction
Oxygen Trap
Trace amounts of Oxygen can damage columns, particularly
capillary columns. Oxygen also degrades performance of
ECD's. An Oxygen trap (p/n 3150-0414), may be connected
between the molecular sieve drier and inlet supply fittings to
the instrument.
Molecular Sieve Trap "Ambersorb" Trap
See Appendices for more information on :
General
information
Conditioning
Installation
Regeneration
Repacking
Tubing
GC
Plastic
Pl ti
ttubing
bi iis permeable
bl tto O2 and
d other
th
contaminants. It may also outgas detectable
impurities.
Precondition
Maintenance of Gases
PURITY
Filters need to be changed at the manufacturer's
manufacturer s
recommended interval to prevent contamination
breakthrough (every 3 cylinders).
LEAKS
All external fittings should be checked on a routine
basis for leaks (every 4 - 6 months).
INLET SYSTEMS
Inlet Systems
In this section you will be able to :
Know
K
the
h diff
different types off Inlet
I l Systems
S
Identify
Understand
Calculate
Inlet Systems
Purpose : To allow the insertion of a sample into the gas
chromatograph in a repeatable, reproducible
manner. The sample should be representative of
manner
the bulk, and unless specifically desired should
be inserted without chemical change.
column inlet
inlet
pneumatic controlled (EPC) on column
10
11
Split/Splitless Inlet
Split Mode for -------------------> Major Component Analysis
Splitless Mode for -------------> Trace Component Analysis
12
13
14
15
Column
ID (um)
A
Total
Flow
B
Column
Flow
C
Septum
Purge
D
Split
Flow
Split
Ratio
200
320
530
100
100
100
0.5
1.0
3.0
2.0
2.0
2.0
97.5
97.0
95.0
196:1
98:1
33:1
B
Total Flow
Column
Head Pressure
Septum Purge
A
Split Vent
Septum Vent
16
= 3.14
r = column radius (cm)
L = column length (cm)
tn = retention time of a
non-retained species
5. Again, measure the "Split vent" flow and adjust the "Total Flow" knob
to set the split vent flow for the desired split ratio.
Split Vent Flow = (Split Ratio) (Column Flow) - Column Flow
6. Set the auxiliary or make-up flow rate
7. Set Other appropriate detector gas flow rates
17
25
50
100
20 psi
10 psi
5 psi
1 psi
0 5 psi
0.5
40 psi
20 psi
10 psi
3 psi
1 psi
80 psi
40 psi
20 psi
5 psi
3 psi
>100 psi
80 psi
40 psi
10 psi
5 psi
Liners
18
Injection Port
Typical problems
Non-representative sample
Leaks
Leaks
Blockage
Incorrect Temperature
Contamination
Detecting Leaks
Snoop
Leak Gun- Matheson
Pressure Test
Chromatographic Changes
Compare shifts in retention time and area counts
Septum
Types :
Standard Low Bleed
T fl Faced
Teflon
F
d
Multilayer
Gray
White
Blue
etc.
Avoid Problems By :
Changing regularly
Installing "Hand-tight"
Hand-tight
Using septum purge when available
Using 11mm septum with autosampler
19
Septa Considerations
Should be used within their recommended temperature limits to avoid :
Leaks
Decomposition
Sample Loss
Reduced Column or Split Vent Flow
Ghosting
Ghosting can be reduced or eliminated using several combined techniques including :
An inlet with septum purge which continuously sweeps the exposed surface of the
septum to remove potential contaminants
Septa which are designed to minimize ghosting (preconditioned, low bleed)
Sample volumes and solvents consistent with inlet volume, temperature, and injection
technique
Smaller sample size
Lower inlet temperature
Larger liner
"High-Temp" septa for applications requiring high injection port temperature (>350)
llowestt practical
ti l ttemperature
t
Purge
Use
septum
with flow
liners
Solvent
Clean
cleaning
syringes
20
Sample Introduction
Perform
Sample Introduction
Purpose : To introduce the sample on to the column
in the vapor state
Syringe Injection :
Manual Injection
Autosampler Injection
Valve Injection :
Gas Sampling Valves
Liquid Sampling Valves
Auxiliary Sampling Devices :
Purge and Trap
Headspace
21
Syringe Injection
Sample should be injected as a plug
For reproducible, sharp peaks injection should be
R id and
Rapid
d Consistent
C
i t t
Prefered technique :
1. Load the desired volume of liquid, then draw the plunger back to pull the liquid out of the needle.
2. Insert the needle quickly through the septum as far as it will go.
3. Press START, depress the plunger, and immediately remove the needles from the injection port.
22
COLUMN
23
Mobile Phase
Stationary Phase
Gas Solid
Chromatography
(GSC)
Gas
Solid
Gas Liquid
q
Chromatography
(GLC)
Gas
Liquid
Separation is
a Partitioning Process
24
Column Types
LENGTH (meters)
I.D. (mm)
PACKED
SERIES 530
SCOT
WSCOT(wide)
WCOT (narrow)
0.5 10
5 - 100
25 -100
5 - 100
5 - 100
24
0.530
0.5 - 0.75
0.3 - 0.75
0.1 - 0.25
10 60
4 - 30
0.5 - 30
1 - 30
0.3 - 1.0
10 - 90
1 - 20
1 - 40
1 - 40
SAMPLE CAPACITY
100 ng/peak
100 ng/peak
5 - 90
50 ng/peak
25
GENERAL PERFORMANCE
CHARACTERISTICS
Packed, Megabore, and Capillary Columns
Packed
Megabore
y (plates/meter
(p
Efficiency
Parameter
Capillary
++
Practical length
--
++
General Inertness
++
++
Phase selectivity
++
++
High-temp analysis
++
Speed of analysis
--
++
Trace Analysis
--
++
++
--
Quantitation
++
++
++
++
++
Ease Of Operation
++
++
Cost
++
Megabore
5m x 0.53mm
Conventional
15m x 0.32mm
26
Typical Chromatogram
Retention time
Parameter used to identify a sample component
Peak Area
Parameter used to measure the quantity of the sample component
5% OV101
on 80/100 Chromosorb
Capillary
30m x 0.53mm x 0.88um
Conventional
30m x 0.32mm x 0.25um
27
Column Technology
Column Selection/Column Care
In this section you will be able to:
Identify the primary differences between packed and
capillary columns
Make preliminary choices in column selection for various
types of applications
Identify the steps in column installation for packed and
capillary columns
Identify
y the p
proper
p steps
p for column care
Perform isothermal and temperature programming oven
operation
Column Types
LENGTH (meters)
I.D. (mm)
PACKED
SERIES 530
SCOT
WSCOT(wide)
WCOT (narrow)
0.5 10
5 - 100
25 -100
5 - 100
5 - 100
24
0.530
0.5 - 0.75
0.3 - 0.75
0.1 - 0.25
10 60
4 - 30
0.5 - 30
1 - 30
0.3 - 1.0
10 - 90
1 - 20
1 - 40
1 - 40
SAMPLE CAPACITY
100 ng/peak
100 ng/peak
5 - 90
50 ng/peak
28
GENERAL PERFORMANCE
CHARACTERISTICS
Packed, Megabore, and Capillary Columns
Packed
Megabore
Capillary
General Inertness
Parameter
++
++
Phase selectivity
++
++
High-temp analysis
++
++
--
++
++
++
Ease Of Operation
++
++
Cost
++
29
Example :
30
Resolution
Selectivity
We would like to know the actual time the component spends in the stationary phase.
tR' = tR - tm
tR
n = 5.545 (------)2
Wh
n
L
31
Injecting
5 cc of butane vapor from a disposable lighter and using the retention time of
the butane peak.
Flowrate in ml/min can be calculated from the linear velocity using the formula:
Flowrate (ml/min) = r260
where
32
Optimum Flowrate
Generally the following guidelines may be used for finding the optimum
values for a particular column:
Packed Column
Capillary Column
Column Diameter
1/4"
1/4
1/8"
Optimum carrier
Flowrate
50-60
50
60 ml/min
20-30 ml/min
Optimum Carrier
Linear Velocity
2 6-3
2.6
3.2
2 cm/sec
4.2-6.3 cm/sec
3-5 ml/min
1-3 ml/min
0.5-1 ml/min
0.2-0.5 ml/min
22-38 cm/sec
20-62 cm/sec
26-53 cm/sec
42-106 cm/sec
33
Judging Efficiency
Packed 1/8"
Column Efficiency
HETP, cm
HETP, cm
Excellent
Good
Fair
> 500
300-500
200-300
< 0.06
0.06-0.10
0.10-0.15
>5000
3000-5000
2000-3000
< 2x10-4
2-3x10-4
3-5x10-4
Dubious
< 100
> 0.3
< 1000
> 1x10-3
Where :
N= n
L
HETP = L
n
1.
2.
3.
4.
5.
NOTE :
34
Resolution
1.18 (RT2-RT1)
(Wh + Wh)
R=
1.5
35
1.
2.
3
3.
4.
5.
6.
7.
8.
9.
C11
4-chlorophenol
1-decylamine
C13
methyl caprate
C14
acenaphthylene
1-dodecanol
C15
50% phenyl
h
l methyl
th l silicone
ili
Conclusions
Capillary Columns:
9Are 10X as efficient as packed columns
9Are 10X as long as packed columns
9Exhibit 100X as many plates as packed columns
9Have lower capacity than packed columns
9Exhibit the same selectivity as packed columns
9Provide enhanced resolution by significantly
9increasing the number of theoretical plates available.
A "brute force" approach to gas phase separation.
36
37
Gasoline Premium
Column 15 m/ 0.3 mm SE-52
38
9Leaks.
9Flow
controller operation.
9Damaged
9Detector
or faulty syringe.
or integrator problems.
too dilute.
9Adsorption
9Incorrect
9Incorrect
flow rates.
9Leaks.
9Sample
9Defective
9Poor
injection technique
9Leaks.
9Incorrect
9Incorrect
detector sensitivity.
Tailing Peaks
9Adsorption
9Injection
9Poor
39
9Choose
9Disconnect
the column from the detector end and verify carrier flow through the column.
9Conditioning
may be done by
exposure to an oven temperature well below the maximum temperature
of the liquid phase.
Repeated oven temperature programming to an elevated temperature until column
performance improves.
Overnight
40
Column Restoration
Removing
Turning
Rinsing
the column. This is most extreme measure. In general, only 50% of the
columns rinsed can be regenerated. Therefore, complete loss of the column is
possible with this method.
41
Rinsing
Column Storage
The column should also be safeguarded when not in use.
There are two storage safeguards of greatest importance:
1. Store the column so it will not be scratched. If scratched,
the stress to the column upon heating may be great
enough to allow the column to break at the weak point.
2. seal the column ends to protect the liquid phase against
diffusing oxygen and contaminants.
When using fused-silica columns, remember that fused silica is a
glass material and eye protection should be used.
42
Temperature Programmed
Isothermal Operation
9Set the injector zone temp
9Set the detector zone temp
9Set the "initial time" to run time for one analysis
9Set the rate to 0
9Set the desired oven temperature
43
44
Temperature Programmed
Characteristics Of Temperature
Programmed Runs
9Used
9Reduces
analysis time
9Produces
sharper peaks
9Produces
9Increased
baseline
9Decrease
45
Blank
Run
Column
C
l
Compensated
Run
46
DETECTORS
Detectors
In this section you will be able to :
Identify
Identify
Identify
Identify
Compare
47
GC Detector - A Definition
A GC Detector is a device which senses the presence of a
component different from the carrier gas, and converts that
information to an electrical signal.
9Thermal
Detectors
Types of Detectors
Brief Description
TCD
FID
ECD
NPD
FPD
48
Detectors
Types of Detectors
Brief Description
ELCD
PID
MSD
IRD
Molecules
o ecu es absorb
abso b infrared
a ed e
energy,
e gy, tthe
e frequencies
eque c es o
of
which are characteristic of the bonds within that molecule.
AED
Detectors Response
Characteristics
Sensitivity
Selectivity
49
Comparison Of GC Detectors
TCD
FID
ECD
AED
PID
IRD
N-P(N)
N-P(P)
FPD(S)
MSD
(SIM)
(SCAN)
ELCD
(X)
ELCD
-15
10
fg
pptrillion
(SCAN)
-12
10
pg
ppb
-9
10
ng
ppm
-6
10
ug
ppthousand
-3
10
mg
percent
1 ng
1 mg
1 ul
1 ppm = ------- = ------- = ------1 ul
liter
liter
Packed Columns
Area
Height
Noise
= 2500
=5
=1
Capillary
p
y Columns
S/N = 5
Area
Height
Noise
= 2500
= 10
=1
S/N =10
50
Dynamic Range
Dynamic range is a measure of response vs. Quantity
Response is the signal produced by the sample.
Dynamic Range
Response
Response
Quantity
51
Flow
When the carrier gas is
contaminated by sample, the
amount of cooling changes.
The difference in cooling is used
to generate the detector signal.
52
TCD Maintenance
Operate at Highest Practical Temperature
Turn Off Filament Current
To measure flows
To change septum
To change column
When gases are off
g used
When not being
53
FID Schematic
54
NH3
H2
CO
CO2
HCOH
CS2
COS
O2
N2
HCOOH
Good
Solution :
CO
or
CO2
Nickel 350
+ H2 ------------------------> CH4
Catalyst tube
CO2 H2O
CHO+
CO2
CHO+
CHO+
H2O
H2O
H2
H2
H2
H2
H2
H2
C
CH4
CH4
CH4
CH4
CH4
CH4
H2
H2
H2
H2
H2
H2
The FID is a
destructive, mass
sensing
i d
detector.
t t
Cations generated in
the flame are counted
and produce the
detector signal.
Analytes that have the
greatest number of
low oxidation state
carbons produce the
largest signal.
55
FID Response
Relative
Response
(by wt.)
FID Variables
56
FID Selection
Part No.
18789-80070
0.030
Use
18710-20119
0.018
Packed Column
(Standard, FID and NPD)
19244-80560
0.011
Capillary Column
(FID and NPD)
(FID: high sensitivity, packed
column)
57
Out
9Spiking
9Low
Sensitivity
9Noise
9Drift
FID Maintenance
9Highest
9Clean
Practical Temperature
Jet and Collector on Regular Basis
FID Troubleshooting
9Check
Flows
Jet
Connections
O-Rings
OK ?
Clean ?
9Column
9Tank
Pressure ?
58
Noisy Baseline
Turn Off
Air/H2
Still Noisy?
No
Contamination
Problem
Yes
Electronic
Problem
Call Berca
For Service
Check:
Filter, Traps
p Gases
Clean Collector
Clean Jet
Verify Col. Max Temp.
Septum
59
60
DATA ACQUISITION
AND
DATA ANALYSIS
61
Data Acquisition
Identify
retention times
9Identify
peak area
9The
9The
9The
62
Producing + 1 V Output
------------------------------|
[ Range 2 (^) ]
(mV, High
Gain)
ECD (kHz)
0
1
2
3
4
5
6
7
8
9
10
11
12
13
1.0 x 10^3
2.0 x 10^3
4.0 x 10^3
8.0 x 10^3
1.6 x 10^4
3.2 x 10^4
6.4 x 10^4
1.3 x 10^5
2.6 x 10^5
10 5
5.1 x 10^5
1.0 x 10^6
2.0 x 10^6
4.1 x 10^6
8.2 x 10^6
25
50
|
|
|
|
|
|
|
|
|
|
|
|
800
|
|
|
|
|
|
|
|
|
|
|
|
|
10
20
40
80
160
320
|
|
|
|
|
|
|
|
63
3396 Integrator
Integration Parameters
Steps in finding and measuring peaks :
Accept data
I
Prepare the data for integration
I
Search the prepared data for peaks
I
Measure the peaks
I
Construct the chromatographic
g p
baseline
I
Correct the peak measurement for baseline
I
Save data
Peak Recognition
Attenuation on the Integrator
Affects only the chromatographic picture and not the data being integrated
Peak Width
Is set to approximate the width of the peak at half height
It filters or matches component peak widths to this setting in order to
eliminate noise spikes and broad ghost peaks
Peak width must be more than 1/4 but less than 2 times the actual width
Area Reject
Sets the minimum area count value necessary for inclusion of the peak in the report.
64
Threshold
THRSH (Threshold) controls sensitivity. By reducing sensitivity,
we make the HP 3396 less responsive to noise in the signal.
Peak Width
Peak Width : Specifies peak width at half height (minutes)
65
66
Area Reject
When a retention time appears on the plot but not in the
report, the peak has been recognized and integrated, but
f il d tto meett th
failed
the minimum
i i
area criteria.
it i
67
Codes
Four Codes
Only one warning code can be printed per peak
> Peak exceeds 1000mV, reduce sample size
I Peaks ends prematurely
< Signal less than -10mV, adjust output for electrical zero
N Negative peak inverted INTG (11) operative
Solvent
Codes
Peaks treated in special manner:
S Designated a solvent peak
T Tangent skimmed from a downslope of a declared solvent peak
Baseline
B
li
C
Codes
d
Indicates how peaks start and end and how baseline was constructed
B Peak begins or ends on baseline
V A valley point occurs when a peak begins before the previous peak ends
P Baseline penetrated, reset to lowest point at the beginning or end of the peak
H Horizontal baseline which extends from the last declared baseline point
Calibration or Standardization
Distinguish
68
Calibration or Standardization
It is the process of relating a particular peak
g or area q
quantity
y with a component
p
height
concentration or quantity.
Methods of Calibration:
Area
Normalization
Area % Calculation
Fame
Area Counts
Area%
C12
280
26.2
C14
250
23 4
23.4
C16
220
20.6
C18
320
29.9
Total:
1070
100.1
Analysis of a sample containing unknown amounts of C-12, C-14, C-16 and C-18.
Fatty Acid Methyl Esters (FAME) by Area % method.
Area % of C12
Area %
280
280 + 250 + 220 + 320
Area (pk)
Area (pk)
x 100% = 26.2 %
x 100%
69
Conc.(pk-std)
Area (pk-std)
RF (pk)
x Area (pk)
Normalization
Normalization ((Norm)) % :
NORM %
x 100 %
70
9Readily
available
9Chemically
9Same
9Does
similar to sample
9Elutes
near component
p
of interest
9Isolated,
clean peaks
9Chromatographically
stable
71
ISTD Calculation
STANDARD
C12
C14
C15
C16
C18
Area Count
250
290
310
330
370
RF
0.100
0.086
0 081
0.081
0.076
0.068
Single point calibration of a standard sample containing 25 ng each of C-12, C-14, C-15, C-16
and C-18 FAME where the C-15 FAME acts as the internal standard.
SAMPLE
C12
C14
C15
C16
C18
Area Count
280
250
280
220
320
RF x Area (ESTD)
28.0 ng
21.5 ng
22 7 ng
22.7
16.7 ng
21.8 ng
ISTD Result
x 1.10 =
x 1.10 =
x 1.10
1 10 =
x 1.10 =
x 1.10 =
30.8 ng
23.7 ng
25 0 ng
25.0
18.4 ng
24.0 ng
Summary of Methods
ADVANTAGES
AREA %
ESTD
NORM %
ISTD
DISADVANTAGES
No calibration required
Injection size not critical
72
73