Escolar Documentos
Profissional Documentos
Cultura Documentos
S8 TIGER
Operator's Manual
think forward
XRF
The reproduction, transmission or use of this document or its contents is not permitted without express written
authority. Offenders will be liable for damages. All rights reserved.
We have checked the contents of this manual for agreement with the hardware and software described. Since
deviations cannot be precluded entirely, we cannot guarantee full agreement. However, the data in this manual are
reviewed regularly and any necessary corrections are included in subsequent editions. Suggestions for improvement
are welcome.
All configurations and specifications are subject to change without notice.
Order no. DOC-M80-EXX032 V3. Updated: November 30, 2009.
2007 - 2009 Bruker AXS GmbH, Karlsruhe, Germany.
All trademarks and registered trademarks are the sole property of their respective owners.
Printed in the Federal Republic of Germany.
S8 TIGER
X-Ray Spectrometer
Operators Manual
Contents
Contents
ii
Contents
5.3.2
Sorting and Printing Results .................................................................................. 5-15
5.3.2
Invalidating Results................................................................................................ 5-15
5.4
Status ........................................................................................................................... 5-16
5.4.1
Schedule................................................................................................................ 5-18
5.5
Tools ............................................................................................................................ 5-19
5.5.1
Appearance ........................................................................................................... 5-19
5.5.2
Measure................................................................................................................. 5-20
5.5.3
Options .................................................................................................................. 5-21
5.5.4
User Editor............................................................................................................. 5-23
5.5.5
S8Tools.................................................................................................................. 5-26
5.5.6
Help & About.......................................................................................................... 5-28
5.6
Alarms.......................................................................................................................... 5-29
iii
Contents
iv
Contents
Contents
vi
Contents
Figures
Fig. 2-1:
Fig. 2-2:
Fig. 2-3:
Fig. 2-4:
Fig. 2-5:
Fig. 2-6:
Fig. 3-1:
Fig. 3-2:
Fig. 3-3:
Fig. 3-4:
Fig. 3-5:
Fig. 3-6:
Fig. 3-7:
Fig. 3-8:
Fig. 3-9:
Fig. 3-10:
Fig. 3-11:
Fig. 3-12:
Fig. 4-1:
Fig. 4-2:
Fig. 4-3:
Fig. 4-4:
Fig. 4-5:
Fig. 4-6:
Fig. 4-7:
vii
Contents
Fig. 4-8:
Fig. 5-1:
Fig. 5-2:
Fig. 5-3:
Fig. 5-4:
Fig. 5-5:
Fig. 5-6:
Fig. 5-7:
Fig. 5-8:
Fig. 5-9:
Fig. 5-10:
Fig. 5-11:
Fig. 5-12:
Fig. 5-13:
Fig. 5-14:
Fig. 5-15:
Fig. 5-16:
Fig. 5-17:
Fig. 5-18:
Fig. 5-19:
Fig. 5-20:
Fig. 5-21:
Fig. 5-22:
Fig. 6-1:
Fig. 6-2:
Fig. 6-3:
Fig. 6-4:
Fig. 6-5:
Fig. 6-6:
Fig. 6-7:
Fig. 6-8:
viii
Fig. 6-9:
Fig. 6-10:
Fig. 6-11:
Fig. 6-12:
Fig. 6-13:
Fig. 6-14:
Fig. 6-15:
Fig. 7-1:
Fig. 7-2:
Fig. 7-3:
Fig. 7-4:
Fig. 7-5:
Fig. 7-6:
Fig. 7-7:
Fig. 7-8:
Fig. 7-9:
Fig. 7-10:
Fig. 7-11:
Fig. 7-12:
Fig. 7-13:
Fig. 8-1:
Fig. 8-2:
Fig. 8-3:
Fig. 8-4:
Fig. 8-5:
Fig. 8-6:
Fig. 8-7:
Fig. 8-8:
Fig. 8-9:
Fig. 8-10:
Fig. 8-11:
Contents
ix
Contents
Tables
Tab. 3-1:
Tab. 3-2:
Tab. 6-1:
Tab. 6-2:
Tab. 6-3:
Tab. 6-4:
Tab. 7-1:
Tab. 7-2:
Tab. 8-1:
Tab. 11-1:
Tab. 11-2:
Tab. 11-3:
Tab. 11-4:
Tab. 11-5:
Tab. 11-6:
Tab. 11-7:
Tab. 11-8:
Tab. 11-9:
Tab. 11-10:
Tab. 11-11:
Tab. 11-12:
Tab. 11-13:
Tab.12-1:
Safety Guidelines
Caution!
To prevent accidents and system damage, read these instructions before installation or
use.
Only operate the S8 TIGER after you have understood and followed the safety warnings
and instructions!
Warning Symbols
All of the following symbols (except the Biohazard Warning) are displayed on the
S8 TIGER system. Strictly obey all instructions and the warning text printed on the labels attached to
the parts of the equipment.
Radiation Danger!
1-1
Safety Guidelines
Follow all valid national, state, and local regulations for disposal.
1-2
Safety Guidelines
Qualified Personnel
Danger: Potentially toxic or hazardous samples!
To prevent any risk for health and safety the analysis of samples with the S8 TIGER, including handling and/or preparation steps, must be entrusted only to persons who have
qualifications in accordance with their activities and functions and who, based on their
technical training, their knowledge and experience and their knowledge of the applicable
standards, are able to evaluate the work assigned to them and recognize possible hazards. This also includes the knowledge of applicable accident prevention regulations, generally recognized safety regulations, EC guidelines and country-specific standards and
regulations. Carefully read and follow all safety and handling advice in the appropriate material safety data sheets.
The S8 TIGER spectrometer system or its parts may only be set up and operated in conjunction with
this manual. Only qualified personnel may be allowed to work on this equipment.
The installation, maintenance and repair of the spectrometry system may be carried out only by personnel who are authorized by Bruker AXS.
All repairs, adjustments and alignments performed on any components of the S8 TIGER system (including the host computer) must be carried out in strict accordance with the established safety practices and standards of the country in which the equipment is installed.
Correct Usage
The instrument was designed to perform element analysis by means of X-ray fluorescence. This device and its components may only be used for the applications described in the technical description or
in the Bruker AXS literature (e.g. XRF Lab Reports or XRF Application Reports). You can request
those reports from your local Bruker AXS Service or Sales Representative. It may only be used in
connection with devices or components from other manufacturers that have been approved or recommended by Bruker AXS, for example for automation solutions.
1-3
Safety Guidelines
Warning!
This product can only function correctly and safely if it is transported, stored, set up, and
installed correctly, and if it is operated and maintained as recommended by Bruker AXS.
Disclaimer and Liability
Bruker AXS has checked the contents of this manual for agreement with the hardware, firmware, and
software described. Since deviations cannot be excluded entirely, Bruker AXS cannot guarantee full
agreement. However, the information in this manual is reviewed regularly and any necessary corrections are included in subsequent editions. Suggestions for improvements are welcome. Send them to
info@bruker-axs.de.
1-4
Safety Guidelines
Note
The X-ray generator is controlled by a three position Keyed switch (OFF/NEUTRAL/ON)
similar to the ignition switch of a car. The key can only be removed when the switch is in
the OFF position. The switch will remain in the OFF position until the key is inserted and
turned. Therefore the removal of the key will terminate production of X-ray.
1-5
Safety Guidelines
Warning: High-Voltage!
Inside the S8 TIGER, voltages up to 60 kVDC occur which are however not accessible
from the outside. High voltages exist in the high-voltage generator and X-ray tube, as well
as in the connecting high-voltage cable. In addition, the radiation detectors are operated
with high voltages up to 2.3 kVDC.
If one of these high-voltage components is damaged, switch off the system immediately
and contact your local Bruker AXS Service Representative.
1-6
Safety Guidelines
1-7
Safety Guidelines
Caution: Oil!
The vacuum pump uses oil as a lubricant. Disposal of oil must comply with all applicable
national, state, and local regulations.
Caution: Batteries!
Some circuit boards inside the S8 TIGER are equipped with batteries. Disposal of batteries must comply with all applicable national, state, and local regulations.
1-8
Safety Guidelines
Warning: Beryllium!
Do not touch the X-ray tubes radiation outlet window or scintillation detectors front windows, as these windows contain beryllium.
In case of an implosion of the X-ray tube, you could get hurt by beryllium fragments. This
is why you should wear safety goggles and gloves when cleaning the sample chamber or
handling the X-ray tube.
Fumes or dust from Beryllium and its compounds can be hazardous if inhaled! Corrosion
of the beryllium may occur during use. Beryllium must not be cut, machined, or handled in
any way.
Disposal of beryllium must comply with all applicable national, state, and local regulations.
1-9
Safety Guidelines
Warning: Thallium!
The analyzer crystal TLAP contains thallium, an extremely toxic heavy metal. This crystal
is water-soluble and can cause poisoning by skin contact.
Please handle this crystal with maximum care and avoid any skin contact! Strictly obey the
applicable national, state, and local safety regulations.
Disposal of thallium-containing parts must comply with all applicable national, state, and
local regulations.
1-10
Safety Guidelines
1-11
Safety Guidelines
Biological Effects
(from Princeton University web page)
Mechanisms of Damage
Injury to living tissue results from the transfer of energy to atoms and molecules in the cellular structure. Ionizing radiation causes atoms and molecules to become ionized or excited. These excitations
and ionizations can:
Produce free radicals.
Break chemical bonds.
Produce new chemical bonds and cross-linkage between macromolecules.
Damage molecules that regulate vital cell processes (e.g. DNA, RNA, proteins).
The cell can repair certain levels of cell damage. At low doses, such as that received every day from
background radiation, cellular damage is rapidly repaired.
At higher levels, cell death results. At extremely high doses, cells cannot be replaced quickly enough,
and tissues fail to function.
Tissue Sensitivity
In general, the radiation sensitivity of a tissue is:
proportional to the rate of proliferation of its cells
inversely proportional to the degree of cell differentiation
1-12
Safety Guidelines
For example, the following tissues and organs are listed from most radiosensitive to least radiosensitive:
Most Sensitive: Blood-forming organs
Reproductive organs
Skin
Bone and teeth
Muscle
Least sensitive: Nervous system
This also means that a developing embryo is most sensitive to radiation during the early stages of
differentiation, and an embryo/fetus is more sensitive to radiation exposure in the first trimester than in
later trimesters.
1-13
Safety Guidelines
1-14
Production of X-rays
X-rays are photons that are emitted when electrons strike a metal target and interact with the electrons
orbiting the nuclei of the metal atoms. The electromagnetic waves in the -range 0.1 - 2 are emitted
in all directions.
In X-ray tubes, electrons are emitted by a heated cathode, accelerated, and strike an anode material.
The cathode consists of a tungsten filament. For the generation of X-rays, high voltages of 20-60 kV
are required. Much of the energy supplied to the tube is transferred to heat, while less than 1% of it is
converted into X-rays. The anode consists of a metal (for example, Rh, Mo or Cr) that produces a
characteristic wavelength. A water cooling system prevents it from overheating.
X-ray tubes usually have a thin Beryllium windows by which the X-rays escape to the outside. The
tube is partly clad with a heavy absorbing material such as lead to prevent the X-rays emitted in other
directions from penetrating to the outside.
Electron beam
Cathode (ring, tungsten filament)
X-ray beam
Fig. 2-1:
2-1
Continuous Spectrum
When accelerated electrons collide with atoms of the metal target, they can be stopped on impact, in
which case all their energy is converted into radiant energy, or they can be decelerated in steps, in
which case a part of their total energy is released in the form of X-rays of greater wavelengths at each
step until it is expended. The incident electron enters the electric field between the electrons and the
nucleus of an atom and is deflected and decelerated by it at the same time. This process is repeated
when the deflected electron enters the electric fields of other atoms.
Fig. 2-2:
The waves of X-rays generated in this way are called continuous radiation because they form a continuous spectrum of wavelengths.
2-2
Characteristic Spectrum
High-energy electrons can also interact with the atoms of the metal target to produce another type of
radiation called characteristic radiation because the wavelengths of the photons emitted are characteristic of the anode metal.
In this case, the incident electron collides with an electron of the first, innermost electron shell of an
atom and knocks it out of its orbit. To restore stability to the energy state of the atom, the ejected electron is replaced by an electron from an outer shell. Because the energy levels of shells increases with
their distance from the nucleus of the atom, the transition of a replacement electron from an outer to
an inner shell is accompanied by a loss in energy, which is released in the form of X-rays.
Fig. 2-3:
2-3
Fig. 2-3 is a simplified representation of the process by which characteristic radiation is produced
through the transition of electrons at the atomic level. In this diagram the conventional notation for the
electron shells and the photons produced by the transition of their respective electrons has been
adopted. K radiation is produced by an electron transfer from shell L to shell K, K by a transfer from
shell M to shell K, and L by a transfer from M to L. Note that the photons have different characteristic
wavelengths.
In fact, on closer examination of the atom and its electron shells, it can be seen that the shells are
divided into subshells as shown in Fig. 2-4 below. For this reason, K radiation can be divided into K1
and K2 radiation, K radiation into K1 and K2 radiation and so on. The difference between the wavelengths of photons produced by the transfer of electrons of different subshells within the same shell is
very small.
Fig. 2-4:
2-4
In Fig. 2-5 below, the wavelength distribution of both continuous and characteristic radiation is shown
as a function of intensity or the number of pulse counts per second. The characteristic peaks K1, Ka2
and K1 are clearly visible above the continuous spectrum.
The point on the wavelength scale at which the continuous spectrum begins is known as the short
wavelength limit (SWL) and corresponds to the radiation produced by the electrons that give up all
their energy on the initial impact. Thus for a particular supply voltage, these electrons give rise to photons of maximum energy (minimum wavelength). With an increase in supply voltage, the SWL as well
as the point of maximum intensity of the continuous spectrum moves down the wavelength scale. In
other words, an increase in operating potential produces a shift of the continuous spectrum in the direction of lower wavelengths and higher energy. On the other hand, the characteristic spectrum, because it is characteristic of the anode metal, merely undergoes an increase in intensity (pulse count
rate).
Fig. 2-5:
2-5
Fig. 2-6:
2-6
Fig. 2-6 shows a monochromatic beam of X-rays incident on the surface of a crystal at an angle . P,
Q and R represent a family of planes separated by the distance d. Plane P reflects AX in XD. Similarly,
plane Q reflects BY in YE at the same angle . Although the beam penetrates many more planes we
need only consider the top two.
Since Q is lower than P, the beam path BYE is longer than AXD by the amount GY + YH. This is
called the path difference.
Since angle AXG = + angle PXG = 90
Then angle PXG = 90 -
Since angle PXY = PXG + GXY = 90
Then angle GXY =
Similarly, you can show that angle YXH =
sin
GY
d
Therefore GY = d sin
YH
d
Therefore YH = d sin
2-7
Scintillation counter
A scintillation counter measures ionizing radiation. The scintillator consists of a transparent crystal that
fluoresces when struck by ionizing radiation. A sensitive photomultiplier tube (PMT) measures the light
from the crystal. When a charged particle strikes the scintillator a flash of light is produced. The PMT is
attached to an electronic amplifier and other electronic equipment to count and possibly quantify the
amplitude of the signals produced by the photomultiplier. The association of a scintillator and photomultiplier with the counter circuits forms the basis of the scintillation counter. The intensity of the light
flash depends on the energy of the charged particles. Cesium iodide (CsI) in crystalline form is used
as the scintillator for the detection of protons and alpha particles; sodium iodide (NaI) containing a
small amount of thallium is used as a scintillator for the detection of gamma waves.
Proportional counter
A proportional counter is a type of gaseous ionization detector. It works on the same principle as the
Geiger-Mller counter, but uses a lower operating voltage. An inert gas is used to fill the tube, with a
quench gas added as a stabilizer. A common proportional gas mixture is 90% argon, 10% methane,
known as P-10. An incoming ionizing particle, if it has sufficient energy, liberates electrons from the
atomic orbitals of the gas atoms, leaving an electron and positively charged atom (ion pair). As the
charged particle travels through the chamber it leaves a trail of ion pairs along its trajectory. The electrons created in this process drift under the influence of the electric field toward the electrode (anode).
At the same time, the positive ions drift towards the cathode, at much lower speed. A proportional
counter differs from an ionization chamber in that the operating voltage is sufficiently high that the
drifting electrons gain enough energy over a mean free path to create further ion pairs when they collide with other neutral atoms of the gas. The electrons created in these new events also drift toward
the electrode and can create further ion pairs themselves. If the operating voltage is chosen carefully,
each avalanche process occurs independently of other avalanches which derive from the same initial
ionizing event. Therefore, even though the total number of electrons liberated can increase exponentially with distance, the total amount of charge created remains proportional to the amount of charge
liberated in the original event.
2-8
3-1
The base instrument allows maximum settings of 50 kV and 50 mA, while the power is limited to 1
kW. The 1 kW version can be operated without external water cooling.
The intermediate system runs at 3 kW with 60 kV and 150 mA as voltage and current limits. The
increase of the voltage to 60 kV allows a better detection of heavier elements.
The high power system allows operation at up to 4 kW. The maximum high voltage remains at 60
kV, but the current can be increased up to 170 kV, which is unique in the market. For light element detection this allows shorter measurement times.
3-2
EasyLoadTM: An 75 position loader with two trays for 51mm rings, automatic liquid sample detection, and 20 sample rings.
SampleCareTM: An automatic mask changer with one closed position for goniometer protection
during pump-down, a 28 mm and a 34 mm mask installed, and a Be foil tube window protection.
Please check your shipping and order information if which options were included in your instrument.
Please note: The S8 TIGER 1kW version can probe to energies up 50 keV.
3-3
Fig. 3-1:
3-4
The X-ray fingerprints of the elements are detected by the S8 TIGER detectors. The measured intensities are used to determine the elemental concentrations.
The beam path is shown in Fig. 3-2. The spectrum of the primary beam can be changed by inserting a
primary beam filter. This can be useful if the excitation spectrum should be biased, for example to
excited heavier elements more easily. To reduce the erroneous signals and background, the sample
beam is collimated and made parallel by a collimator mask and a collimator. The collimator mask can
be changed automatically like all other components in the beam path. The collimator is labelled by its
opening angle. This limits the beam divergence of the excited beam. Smaller values lead to a better
energy resolution albeit a smaller intensity. There are up to four collimators available.
Filter wheel
(primary beam filter)
Sample
Mask changer
Vacuum seal
Collimator changer
X-ray source
Crystal
changer
Scintillation counter
(heavy elements)
Fig. 3-2:
Proportional counter
(light elements)
3-5
After leaving the collimator the beam impinges on an analyzer crystal. Those crystals are organic or
artificially produced multi-layers. Depending on the lattice spacing and obeying Braggs Law the X-ray
beam is diffracted at certain angles. A total of eight crystal positions can be used. As a default, three
positions are used for PET, XS-55, and LiF 200 analyzer crystals. A quick overview of the elemental
range of those standard crystals is shown in Fig. 3-3. In addition, Bruker AXS offers a large variety of
optional crystals optimized for specific applications. The number of options is continuously growing.
Please contact your local application or sales representative for detailed and up-to-date information.
Fig. 3-3:
3-6
Overview of the typical elemental range for the three standard analyzer crystals. For a
more detailed periodic table, please, consult the appendix.
The radiation from the analyzer crystal is detected by one of the two detectors. One detector is optimized for soft, one for hard radiation1. To reduce the background even further both detectors are
equipped with soller slits (not shown in Fig. 3-2).
The detectors are mounted onto the 2-theta goniometer, the sample on the theta-goniometer. For the
measurement the respective detector is either positioned at a given angular position (peak measurement) or it is moving during the measurement (scan measurement). The angular position of constructive interference is determined by Braggs Law. This means the 2-theta position of a peak depends
both on the d-spacing of the analyzer crystal and the characteristic wavelength of the element probed.
Since all elements are measured subsequently, the instrument is referred to as sequential X-ray spectrometer.
The beam path is inside a housing to shield the user from X-ray radiation. A safety circuitry assures
that the radiation can only be produced if both the spectrometer chamber is closed and the protective
covers are installed. The S8 TIGER can be operated with a variety to sample magazines.
Warning
It is illegal to hamper or defeat the radiation protection system.
3-7
Instrument Components
The S8 TIGER is a fully automated X-ray spectrometer, both operation and analysis can run with very
little user interaction. In this manual we will focus on the general instrument operation.
All the connection lines and cables can be found at the back of the instrument. Through those lines
you supply electricity and water. You can also connect a local area network via an Ethernet connection. Depending on your configuration and application, there might be two gas hoses, one for supplying P-10 to the flow counter and one for supplying Helium or Nitrogen to the sample chamber.
From the front and either side of the instrument, see Fig. 3-4, you can see a field of LEDs that are
illuminated when X-rays are produced. These are the X-ray warning lights. When standing in front of
the spectrometer to the left you see the on and the off button, an emergency off switch and the generator turn-on key switch. To the right are the touchscreen and the panel with four LEDs that display
an abbreviated instrument status.
When you open the instruments top you can have a close look at the sample magazine and the sample robot.
Warning
Do not use the handle of the top cover to pull the instrument!
Towards the back left the sample chamber can be found. There are many magazine types available.
The whole magazine and robotics is often referred to as the loader. Prior to the measurement the test
specimen is placed into a sample container, like a ring or a cup, and subsequently this ring or cup is
placed into the loader.
3-8
Top cover
X-ray
warning light
X-ray
warning light
Touchscreen
Status
LEDs
On and Off
switches
Fig. 3-4:
3-9
Sample
grabber
Sample
position
Sample
chamber
Fig. 3-5:
During operation, all side panels must be mounted, since they are part of the X-ray safety system.
During installation and system commissioning and in case of a service visit, panels will be removed.
The internal arrangement of the instrument will be revealed.
3-10
Status LEDs
Electronics
control rack
Spectrometer
chamber
Safety board
Generator
Fig. 3-6:
View on the instrument after removing the front and the right side panel with several
components labelled.
3-11
X-ray warning
light
On/off
switch panel
X-ray tube
Safety board
Spectrometer
chamber
Back side of
the generator
Fig. 3-7:
3-12
View on the instrument after removing the front and the left side panel with several components labelled.
Mask changer
drive
Spectrometer
chamber
Vacuum pump
Fig. 3-8:
View on the 3 or 4kW version after removing the back and the left (viewed from front) side
panel with several components labelled.
3-13
Electric power
distribution
and mains
connection
Refill port for deionized water
Electronics
control rack
Generator
Water pump
Vacuum pump
Heat exchanger
Fig. 3-9:
3-14
View on the 3 or 4 kW version after removing the back and the right (viewed from front)
side panel with several components labelled.
Fig. 3-10:
Status display
LED Guide
Tab. 3-1:
LED guide
Ready
(orange)
permanently on
blinking
On
(green)
permanently on
blinking
Alarm
(red)
permanently on
blinking
system alarm
system warning is pending
Busy
(yellow)
permanently on
blinking
measurement is running
a ready flag is missing, data taking is interrupted
All LEDs
blinking at a frequency
of about 2 Hz
3-15
3-16
Tab. 3-2:
Overview of the S8 TIGER atmospheric measurement modus. Please note that some
modes are optional.
Mode
Samples
Remarks
Vacuum
Metals,
fused beads,
pressed pellets
glass
Separated vacuum
Pressed powders
(dusty samples)
The vacuum seal protects the goniometer and compartments against dust particles released from the
sample.
IMPORTANT:
It is strongly recommended to prepare pressed pellets
as stable as possible by using the appropriate binder
and applying best pressure
Helium
(reduced pressure)
Liquid samples
Oils, water
Loose powders in a cup
with polymer film
To avoid sample evaporation and the resulting contamination and/or damages, liquid samples must not
be measured under vacuum. The helium mode with
reduced pressure is recommended for liquids with a
high boiling point. The helium pressure is reduced to
gain better intensity for light elements and to keep
helium consumption low.
Atmospheric Helium
Liquid samples
automotive fuels or
solvents
3-17
Several pressure sensors monitor the vacuum continuously. Only after the pressure is sufficiently low,
is possible to start a measurement.
The S8 TIGER is a wavelength-dispersive spectrometer. Intensities for individual wavelengths are
determined in what is also known as Bragg-Brentano geometry with, in our case, the sample acting as
radiation source. The analyzer crystal and detector are moved in a fixed angular relationship of 1:2.
This gives great flexibility to the instrument. It can be adjusted to almost any new analytical task. On
the other hand, this also means that measurements of different samples are executed in sequence.
So, before the sample is in measurement position, the instrument has to know for which elements to
probe. Based on this input, the optical components (beam filter, mask, collimator, analyzer crystal, and
detector) are moved into position while the chamber is being evacuated. Most motions can be executed in parallel.
Once in position, the measurement starts immediately. All components move in parallel to get ready
for the measurement. After all measurement tasks are finished, the sample is removed from the sample chamber. During sample transfer, no X-rays are produced for your safety. There are two types of
measurements: Peak Measurement (faster; only the intensity at one position is measured) and Scan
Measurement (the goniometer performs a coupled scan, the peak shape and intensity over an angular
range is determined).
More background on XRF analysis and some instrumentation background can be found in the Bruker
AXS publication Introduction to X-ray Fluorescence Analysis (DOC-M84-EXX001). Additional references are given therein.
Note
Please consult the Introductory User Manual for an in-depth description of the safety system.
3-18
NO
Error message
Move up (z).
NO
Open shutter.
Wait.
Move down.
Check if height is okay.
YES
De-magnetize (leave sample).
NO
Error
3-19
X-ray On
Ramp up HV for X-ray tube.
Check if V, I for X-ray
tube ready and stable.
IF NOT
IF NOT
IF NOT
Error and stop
measurement.
IF OK
Execute measurement program (method).
End measurement.
Move grabber to sample input position (xy).
Move in vacuum seal.
Ramp down X-ray HV and current.
Flood sample chamber with air.
Remove sample.
Fig. 3-11:
3-20
Network Configuration
The SPECTRAplus PC and the S8 TIGER are connected together via a network. Depending on your
network configuration this may be a simple cable from the PC to the S8, or the devices may be connected via a hub or switch. Alternatively both devices may be connected to your local-area-network
(LAN). The S8 TIGER will request an IP address via DHCP, and if a DHCP server is not available, it
will assign itself an automatic private IP. It is recommended that the SPECTRAplus PC is configured for
DHCP to ease any network configuration problem. The network settings of the devices leaving the
factory are guaranteed to work. If you need to reconfigure the devices for your network please contact
your Bruker representative for advice and instructions.
C:\
Embedded OS
TouchControl
Measurement Server
D:\SPECplus
C:\Program Files\SPECPlus
SPECTRAplus programs
e.g.
Loader.exe
Eval2.exe
etc.
Fig. 3-12:
T:\SPECplus
3-21
The SPECTRAplus PC accesses data on the S8 TIGER via a Windows mapped drive. It is recommended to use the SPBrowser program map the S8 drive automatically.
Fig. 3-12 shows an overview of the software and data files for the PC-S8 system. It is important to
note that there is a separation between data files and programs. This is to allow the S8 TIGER to work
as a stand alone instrument, allowing routine measurements even when the PC is switched off.
Use of an External PC
You can operate the S8 TIGER with TouchControl on the touchscreen or with the measurement
software SPECTRAplus on an external PC via network.
3-22
2.
Check that the external cooling water system is operating correctly. All valves connecting to the
instrument are open and hoses correctly connected to the S8 TIGER.
3.
Check the P-10 and helium supply (if used) is turned on and set correctly. The P10 should be set
to 0.5 bar (~7 PSI). Helium should be set to 2 bar (~29 PSI).
4.
The system configuration should be set in the Default parameters pane to Airlock mode = Spectrometer mode (Standby parameters) and vacuum with seal (Mode). The default settings see in
Fig. 4-1.
5.
Press the green power-on button on the front of the spectrometer (Fig. 4-2). All four lights will
come on for about 30 sec. If the lights do not come on verify the STOP button is out by turning it
clockwise then press the green button again.
Note
For this procedure to work auto setup1 must be enabled. This is the default setting.
4-1
Fig. 4-1:
4-2
STOP button
power-off button
power-on button
Fig. 4-2:
Control panel
Fig. 4-3:
Status display
6.
When the S8 TIGER is finished initializing the green On and the red Alarm lights should be on
(Fig. 4-3).
7.
Immediately afterward, the z drive moves up and the sample seal (shutter) opens and closes for
an automated safety circuit check. This takes up to 30 seconds because the sample chamber
might need to be flooded with air.
4-3
Note
You must leave the cover closed; otherwise the start up routine is interrupted.
8.
The X-ray generator can now be turned on by inserting the key and turning the key clockwise.
9.
10. HV is turned on, the X-ray warning lamps turn on and the Ready LED is permanently illuminated.
11. Release the key. It is spring-loaded and should return to the middle position.
Note
If the S8 TIGER has been turned off for a while it will take a few hours to return to its operating temperature. Until the operating temperature has been reached the spectrometer
will show a temperature warning with a flashing alarm light.
12. For TouchControlTM users: log on to the touchscreen with username: admin and password: pass.
4-4
Fig. 4-4:
4-5
2.
3.
Then confirm you want to shutdown by pressing Yes (see Fig. 4-6).
4.
Wait until the message It is now safe to switch off your machine appears on the screen.
5.
Press the power-off button on the front of the machine (see Fig. 4-2).
6.
4-6
Shutdown button
Fig. 4-5:
7.
Shutdown.
To shutdown the S8 TIGER, first press the Shutdown button shown above and confirm the shutdown by pressing Yes (see Fig. 4-6).
Fig. 4-6:
4-7
Start S8 Tools.
2.
Fig. 4-7:
3.
4.
4-8
Only when the air mode has been achieved, turn the X-ray key to the left.
STOP button
power-off button
power-on button
Fig. 4-8:
6.
Control panel
4-9
4-10
TouchControl Operation
5.1 Introduction
This chapter describes how to use the TouchControl interface. The interface is designed to be easy to
use and intuitive to allow operation with minimal training, yet powerful enough for all routine operations. TouchControl is not used for creating or editing calibrations which requires the SPECTRAplus
software package to be installed on a standard PC.
Caution
In order to avoid damage to the TouchControl, the use of sharp objects should be
avoided. The interface has been designed to be used with your fingers only. Gloves may
be used to avoid contamination. To clean the touchscreen, avoid the use of abrasives or
aggressive solvents. Only use a soft damp cloth with a mild detergent if necessary.
Note
The touchscreen interface is available in different languages. Please contact your Bruker
AXS sales representative to check if your choice is available.
5-1
TouchControl Operation
Disabled
Button
Selection
buttons
Thumbwheel
The Thumbwheel allows you to scroll items in a vertical direction. Press and hold while sliding your
finger up or down to rotate the wheel.
The Thumbwheel is also used in the Selection List control (see below).
5-2
TouchControl Operation
Selection List
3
This control is used to select an item from a list. The highlighted text is the
selected item. To select another item, the following methods can be used:
2
1
1.
2.
3.
Screen Selector
Once you have logged into the S8 TIGER, this control is always located at the top of the screen. This is used
to navigate through the various screens in the software. Since only one screen can be displayed at any time,
only one button is highlighted. Simply press the button to display that screen.
The Stop button becomes active during measurements and can be used to interrupt the current
measurement and to pause the measurement queue. When this button is pressed the text will change to
Restart which will restart the measurement queue.
5-3
TouchControl Operation
Screen Keyboard
The Screen Keyboard is used to enter text and operates in the same way as a normal keyboard. The text is
inserted into a Textbox (see below).
The Caps Lock key is used to toggle between lower and upper case characters.
Textbox
This is the active Textbox. Text entered with the Screen Keyboard will be
inserted here.
The Textbox is used to enter details such as sample ID's, usernames, passwords etc. An active Textbox has a
thick border and receives text entered from the Screen Keyboard. An inactive Textbox has no border and will
not receive text from the keyboard. Touching an inactive Textbox will make it active.
5-4
TouchControl Operation
Date Selector
Status Bar
The Status Bar is always located at the bottom of the screen and provides the following machine information:
Current machine action and any warning message.
Status light.
Green = On line normal operation.
Red = Alarm active or communication problem with measurement server.
Current time.
5-5
TouchControl Operation
Login information
User Name
5-6
admin
Password
pass
TouchControl Operation
Screen-Keyboard
Fig. 5-1:
To log-in to the S8 TIGER you will need to enter your username and password. Touch the
textboxes and use the Screen-Keyboard to enter your details.
5-7
TouchControl Operation
Main screens
Loader
This screen is used to define and start your measurements. It displays all samples
defined on the magazine including analytical results. All users have access to this
screen.
Results
This screen allows access to the measurement database to recall all previous
measurements performed on the S8 TIGER. You may restrict access to this screen for
certain users (see section 4.3).
Status
This screen displays the current status of the machine, spectrometer values and a
schedule of queued measurements. You may restrict access to this screen for certain
users (see section 4.4).
Tools
This screen includes various pages which allow the users to change options, edit user
details, perform diagnostics and view the on-line help. You may restrict access to this
screen for certain users (see section 4.5).
After logging into TouchControl the Loader screen will be displayed and is described in the next section.
5-8
TouchControl Operation
Sample Magazine
Application
Chooser
Fig. 5-2:
The Loader screen will display the Application Chooser when an empty position is selected on the Sample Magazine.
5-9
TouchControl Operation
Search results
Search
Options
Fig. 5-3:
5-10
The Results screen allows access to all stored results on the S8 TIGER.
TouchControl Operation
Fig. 5-4:
5-11
TouchControl Operation
Fig. 5-5:
5-12
TouchControl Operation
Quick Search
Search results
Advanced
Search
Fig. 5-6:
5-13
TouchControl Operation
Fig. 5-7:
5-14
TouchControl Operation
Fig. 5-8:
Search Options.
5-15
TouchControl Operation
Fig. 5-9:
5.4 Status
The Status screen shows the current status of the machine as is shown in Fig. 5-10. The information is
grouped into five sections:
1.
They show whether the device is ready for measurement (Ready), is counting (Busy) and if any warnings or alarms are present. If a warning is present details will be displayed in the Status Bar at the
bottom of the screen. For alarms see section 4.6 for details.
2.
Loader
It shows the x, y and z coordinates of the sample grabber. Also shows whether the loader cover is
open or closed and shows the magazine position of the active sample.
3.
Drive 1 represents the theta drive and Drive 2 represents the 2theta drive. The lights show if the drives
are ready and running and will also show any errors present.
4.
Spectrometer
The section shows details of the spectrometer such as the theta, 2theta, tube conditions, and count
rates. It also shows the selected primary filter, crystal and mask.
5-16
5.
TouchControl Operation
Ready flags
Before a measurement can be performed all ready flags have to be set. This section shows what parts
of the machine are ready for measurement and can be useful for diagnostics.
Fig. 5-10:
5-17
TouchControl Operation
5.4.1 Schedule
The Status screen also contains a Schedule button which shows all samples waiting to be measured.
The Schedule page is shown in Fig. 5-11. This page shows the samples position in the queue (#), the
magazine position, the sample ID, the expected start time, the expected end time, and for active samples, a progress bar. To close the page, simply press the OK button.
Fig. 5-11:
5-18
TouchControl Operation
5.5 Tools
The Tools screen contains seven pages of options and information which are described in this section.
5.5.1 Appearance
The Appearance page contains options that affect the appearance of the Loader screen. Here you can
select a 3D or 2D graphical loader, the movement direction of the cursor and whether the magazine
(tray) position is displayed. These options do not affect the measurement process in any way.
Fig. 5-12:
The Appearance page contains options that affect the appearance of the Loader screen.
5-19
TouchControl Operation
5.5.2 Measure
The Measure options affect how measurements are started and the priority at which they run and are
shown in Fig. 5-13. By default after defining a sample (see section 4.3.1) the sample will not be added
to the measurement queue and the cursor will move onto the next empty magazine position (Automove). This is suitable when a large number of samples need to be defined but you do not want to
measure the samples immediately. The Auto-move option can be switch off by simply pressing the
Off button. If you wish to send the sample to the measurement queue immediately after it has been
defined, simply switch the Auto-measure option on.
Samples can also be defined and measured using the SPECTRAplus software. The Priority options set
the priority level of the TouchControl samples. High priority will interrupt SPECTRAplus samples, normal priory will measure TouchControl samples first but will not interrupt SPECTRAplus and low priority
will measure TouchControl last. Normal priority is recommended.
Fig. 5-13:
5-20
TouchControl Operation
5.5.3 Options
The Option page allows you to choose your Application Shortcuts (see section 4.3.1), choose your
language, set your preferred date-format & temperature scale, and enable the auto log-off option. Fig.
5-14 shows the Options page.
Fig. 5-14:
Fig. 5-15 shows how to add or remove an Application Shortcut. Shortcuts are applications which are
represented as a button in the Loader screen. This allows a user to choose their application with a
single button press. It is recommended that you include frequently used applications in your shortcuts.
5-21
TouchControl Operation
Other Applications
Fig. 5-15:
Shortcuts
To add an application to the Shortcuts, select the application using the selection list on
the left and press 'Add'. To remove a Shortcut, select the application on the right and
press 'Remove'.
The auto log-off option will automatically log the current user off if there has been a period of inactivity.
The period of inactivity can be set as either 5 or 10 minutes. This option is recommended to prevent
unauthorized operators using the machine if the current operator forgets to log off.
TouchControl is available in various languages; simply select your language by pressing the appropriate button.
The date format used in TouchControl can be switch between month/day/year & day/month year.
5-22
TouchControl Operation
Fig. 5-16:
The User editor page is used to add a new user or change the privileges of an existing
user.
5-23
TouchControl Operation
Changing a password
To change a password, select the user from the Selection List on the left-hand side and press the
'Change Password' button. You will now be asked to enter the new password, which is entered using
the Screen-Keyboard. Press the OK button when you have finished. Remember to make a note of
your new password and store it in a safe place. If you forget your password you will need to contact
your Bruker AXS representative.
Adding new users
It is recommended that you add a user for every operator that will use the S8 TIGER. This will enable
you to track who measured a particular sample and if you wish, grant limited access to the software
options. You may want to spend some time using the S8 TIGER before adding new users, so that you
can decide what options to set. Alternatively you can add users straightaway and change their privileges at a later date.
To add a new user, press the 'Add User' button. A dialog will be displayed which is shown in Fig. 5-17.
Enter the username and password using the Screen Keyboard.
Fig. 5-17:
5-24
TouchControl Operation
Fig. 5-18:
5-25
TouchControl Operation
5.5.5 S8Tools
The S8Tools page allows a manual mask to be inserted into the machine, initialize the device and start
the advanced S8Tools which can help diagnose any problem with the device. The S8Tools screen is
shown in Fig. 5-19.
Fig. 5-19:
Manual mask
Some applications require a specific mask size for measurement. The S8 TIGER is equipped with an
automatic mask changer for up to 3 masks (automatic masks). If an additional mask size is required
5-26
TouchControl Operation
for measurement, it can be inserted using the S8Tools page (manual mask). The S8Tools pages show
the current installed mask and a list of available manual masks.
Mask changes can only be performed when the machine is idle and not measuring a sample. To insert
a mask press the Change Mask button and the dialog shown in Fig. 5-20 will be displayed. Wait for
the chamber shutter to open before opening the magazine cover. Insert the mask into the correct location.
Fig. 5-20:
Initialize S8
If there is a problem with the device you may try to re-initialize it by pressing this button.
S8Tools (advanced)
If a problem persists you can start S8Tools (advanced) by pressing the Start S8Tools (Advanced)
button. Using the separate S8Tools program requires specific training.
The S8Tools program can also be started from the external PC.
5-27
TouchControl Operation
Fig. 5-21:
5-28
TouchControl Operation
5.6 Alarms
If the machine experiences a problem an alarm may be displayed. All alarms are displayed in the
Loader screen. While an alarm is active no measurement can be run. An example of an active alarm is
shown in Fig. 5-22. Follow the instructions on the Alarm message to clear the alarm. You may be advised to use S8Tools (advanced) to diagnose and clear the alarm. If you have difficultly clearing an
alarm or you feel you are experiencing problems you should contact your Bruker AXS representative.
Fig. 5-22:
5-29
TouchControl Operation
5-30
Note
All systems are equipped with a temporary installation of QUANT-EXPRESS.
The example in the following sections shows how to measure an unknown sample with the QUANTEXPRESS application.
Start material specific and other applications in the same way. Some sample information might not be
required and can be skipped.
You can measure samples with either vacuum or helium modes. The following table offers guidance
on which mode to choose.
Method
Shortcut
Sample
Atmospheric Helium
AtmHe
Volatile liquids
Helium
He
Separated vacuum
Svac
Pressed samples
Vacuum
Vac
Compact solids
6-1
Sample Preparation
The sample is prepared according to best practice experience and using the respective preparation
equipment in the laboratory. Tab. 6-1 gives an overview of the most common materials and the recommended sample preparation method:
Tab. 6-1:
Shape
Typically Material
Compact
solids
Metals
Polymers
Important
Polishing
Milling
Hot melting
Flat surface
Pressed pellet
(Milling and Pressing)
Homogenous sample
(Using wax as a binder to form a stable specimen)
Homogenous sample
Reduction of matrix effects and grain size effects
Powders
Chemicals
Geological
samples
Granulates
Chemicals
Polymers
Liquids
Fuels
Solvents
Water
Oils, Grease
Fused bead
6-2
Routine Check
To verify the correct functionality of the S8 TIGER and to recognize the need of maintenance for the
system we recommend that you run a system check regularly. Bruker AXS has already installed two
methods on the S8 TIGER, the S8 CHECK and the QE CHECK.
For material specific applications we recommend that you apply the same quality procedure based on
a stable sample (please refer to the "SPECTRAplus V2 User Manual").
Note
We recommend that you perform the routine check once a week. To comply with even
stricter quality rules we recommend that you run it on a daily basis.
6-3
Load the sample STG 2 in a sample cup with 34 mm mask size and place it on the loader.
Fig. 6-1:
2.
6-4
Now select the appropriate position on the screen by touching it with the finger.
3.
Press the button S8 CHECK. If the button S8 CHECK is not displayed, first select and press
others. Now select S8 CHECK.
4.
The loader of the S8 TIGER now picks up the sample and loads it automatically. The sample is analyzed under vacuum. After the measurement the result is displayed at the bottom of the page.
Fig. 6-2:
6-5
Note
Please note that your measurement results might differ from the results of the example
shown in the figure above. This is due to changes between the reference sample lots.
Compare your results and tolerances with the results of the acceptance protocol in your
supplement folder chapter 3. These results apply particularly to your
S8 TIGER system.
If the results are outside the limits please refer to the chapter "Troubleshooting" in this manual.
6-6
Load the sample STG 2 in a sample cup with 34 mm mask size and place it on the loader.
Fig. 6-3:
2.
3.
Press the button QE CHECK. If the button QE CHECK is not displayed, first select and press
others. Now select QE CHECK
4.
The loader of the S8 TIGER now picks up the sample and loads it automatically into the sample
chamber. The sample is now analyzed under vacuum. Upon completion, the result is displayed at the
bottom of the page.
6-7
Fig. 6-4:
Note
Please note that your measurement results might differ from the results of the example
shown in the figure above. This is due to changes between the reference sample lots.
Compare your results and tolerances with the results of the acceptance protocol in your
supplement folder chapter 3. These results apply particularly to your
S8 TIGER system.
If the results are outside the limits please refer to the chapter "Troubleshooting" in this manual.
6-8
6-9
Fig. 6-5:
Note
If the button QUANT-EXPRESS is not shown on your screen, press OTHER to achieve a
listing of all installed applications where you will find also QUANT-EXPRESS. The button
can be created on the tool page.
6-10
Now you are asked to enter the sample ID using the keyboard on the touchscreen. This is important
for the correct identification of the sample. All related information is stored using this sample ID in the
measurement database. Please confirm the sample ID by pressing the button NEXT>.
Fig. 6-6:
6-11
Step 2: Material
For sample evaluation, QUANT-EXPRESS can integrate additional sample information about the matrix composition and light elements. Even if those elements cannot be measured directly by XRF. This
is needed to achieve a good quantitative result. By default, QUANT-EXPRESS comes with three
evaluation models.
Tab. 6-2:
Method
Matrix
Material
Elements
none
Oxides
none
Organics
CH2
Choose the right evaluation model for your sample, here Elements for a metal sample and press
NEXT>.
Note
Additional matrix information about your sample or concentrations for light elements which
has been determined by another method can be included during the interactive evaluation
using the program EVAL2 and will further improve the results.
6-12
Fig. 6-7:
Sample materials.
The program Application Setup is used to create customized evaluation models (.evm file). The
evaluation models are based on user input like sample type and competition. The .evm files must be
stored in the directory SPECplus/Libraries/Calibrations on the hard drive on the S8 TIGER. At the next
start-up of QUANT-EXPRESS these models will be shown automatically on the screen.
6-13
Step 3: Preparation
QUANT-EXPRESS automatically calculates and displays the concentration values of the original sample prior to preparation. Effective sample dilution and chemicals used during the preparation of the
specimen, e.g. wax or flux, are taken into account during the evaluation. Also the information about
the sample shape (diameter, thickness, weight) will help QUANT-EXPRESS to properly correct for
sample effects.
All preparation methods stored in the central database fluo.mdb are displayed. Press the preparation
button according to the sample and press NEXT>. In our case the metal sample is defined as
SOLID, the diameter of the metal disk is 40 mm and infinitely thick for XRF.
6-14
Fig. 6-8:
6-15
Step 4: Mode
All atmospheric modes can be used to perform measurements with QUANT-EXPRESS. Compact solid
samples are typically measured in VACUUM. The highest sensitivity for light elements is achieved in
vacuum. If your S8 TIGER is equipped with the vacuum seal of SampleCareTM and helium flushing
unit, three additional modes are available: SEPARATE VACUUM, HELIUM and ATMOSPERIC HELIUM.
The metal disk is run under vacuum by pressing VACUUM and Next>.
Note
QUANT-EXPRESS automatically recognizes sample preparation methods which cannot
be run under vacuum. If the sample is filled into a liquid cup (with a polymer film as bottom) the vacuum modes are automatically disabled. This is true only if the preparation
method is properly defined.
6-16
Fig. 6-9:
Also if the instrument is not equipped with helium flushing unit and a vacuum seal only the vacuum
mode is displayed. In this case the instrument is configured to analyze only compact solid samples.
6-17
Step 5: Diameter
Within this step the sample diameter is defined. According to the instrument configuration all available
masks are displayed in the combo box. The S8 TIGER allows the use of two mask types: Automatic
masks in the mask changer, this is recommended. If you own manual masks from other Bruker instruments, you can also use most of them. Also, some of the low background small area masks are
only available as manual mask.
If an automatic mask is selected, the S8 TIGER automatically installs the corresponding mask before
the measurement. If a manual mask is chosen, you will be asked before the measurement to place the
corresponding mask in the sample chamber. Note that after this measurement you may need to
change back to another mask to run new samples.
Select the appropriate mask, in our example 34mm is selected.
6-18
Fig. 6-10:
6-19
Method
Typical duration
Target
Fast Screening
~ 3 minutes
Full Analysis
~ 8 minutes1
Best Detection
~ 15 minutes1
Note
Prolonging the measurement time especially for trace elements will improve data quality
and result in lower detection limits for these elements.
Note
When setting measurement times for measurement methods, take into account to which
degree your sample and the element distribution remains stable during the measurement.
In liquids one often observes element separation, e.g. heavier elements collect at the bottom close to the foil.
The total measurement time for each sample may change according to the sample composition. The measurement methods of QUANT EXPRESS run the samples using the automatic count rate reduction to avoid detector
saturation. The current of the X-ray tube is automatically reduced so that the intensity of each major element is
measured in the linear range of the detector.
6-20
Measurement methods with customized settings can be stored in the path \Measurement methods\QuantExpress. The file name must include settings for the mode and mask size, the name of the
file must end with the corresponding naming, e.g. *-Vac34mm.mm for the vacuum mode and 34 mm
mask.
Fig. 6-11:
Select now Full Analysis to finish the sample definition. The sample is now ready to be measured.
6-21
Fig. 6-12:
You may now start the sample by pressing the button MEASURE. The sample grabber picks up the
cup and places the sample in the chamber. The S8 TIGER automatically adjusts mode and sample
6-22
mask and performs the measurement. When the measurement is finished, the sample is automatically
placed to its original position. The next sample will be loaded.
Fig. 6-13:
6-23
Fig. 6-14:
6-24
The results could also be retrieved performing a database search on the screen "Results.
Fig. 6-15:
Result screen.
You may now check the results. It is recommended to check the quality of the result in EVAL2 interactively. By default QUANT-EXPRESS performs the measurement in scan mode. The major benefit is
that the user can change interactively later analytical settings (analytical lines), check for line overlays,
provide more detailed sample information (matrix compounds) or correct sample preparation parameters. The quality of the analytical result may improve and the new evaluation can be stored in the database as a new record.
6-25
Evaluation Method
<My application.evm>
links all files and information for the measurement and the
evaluation of the sample
Measurement Method
<My application.mm>
Calibration file
<My calibration.fcl>
Line Library
S8-LineLibrary.fll
contains all element lines with their measurement parameters (collimator, crystal, detector type and discriminators,
dead time, peak and background positions, drift correction
settings and information,)
Preparation database
Fluo.mdb
contains all information about standards and their preparation, also about preparation types
Results database
Measure.mdb
Specification database
Specification.mdb
Instrument configuration
S8-configuration.cnf
6-26
Note
Solvents should be disposed of in accordance with local regulations.
Larger particles of dirt on the sample magazine or inside the loader can be removed with a vacuum
cleaner.
7-1
Operators Manual
Touchscreen
To clean the touchscreen, avoid the use of abrasives or aggressive solvents. Only use a soft damp
cloth with a mild detergent if necessary.
Keep the samples in a safe, clean and dry place (e.g. desiccator).
Clean the surface of the regular base must be cleaned with pure alcohol and a lintfree tissue.
This procedure allows retaining the same high quality on results over time. In case of sample damage
or high deviation of the results contact the Bruker AXS representative for a replacement of the sample.
7-2
Operators Manual
7-3
Operators Manual
7-4
Operators Manual
The S8 TIGER shows warning if the water level in the reservoir is low.
The S8 TIGER shows alarm if the water level in the reservoir is below a critical limit.
Warning
Only use deionized water!
Warning
If the system shows ALARM the HV immediately switch off, and can not be switched on
again!
7-5
Operators Manual
In case of an Alarm the ALARM LED is flushing permanently, the HV can not be switched on.
Instument|Warnings
Instrument|Alarms
Fig. 7-1:
7-6
Operators Manual
Switch the S8 TIGER off like described in chapter 3 Switching the Instrument On and Off (on
page 3-6). Disconnect the S8 TIGER from the mains power (pull the mains plug).
Fig. 7-2:
7-7
Operators Manual
Remove the rear panel of the S8 TIGER. Use a slit screwdriver and open the 3 lock screws (1-3),
turn clockwise:
Fig. 7-3:
7-8
Backside view.
Fig. 7-4:
Operators Manual
It is not necessary to remove the gas pipes; theyre fixed internally, not on the rear panel. If you put the
rear panel carefully to the backside you will automatically loose the ground plugs.
Note
Be careful, not to disconnect the USB and the LAN-Plugs!
The water tank is placed on the left side, above the vacuum pump (see Fig. 7-5).
7-9
Operators Manual
Fig. 7-5:
Open both of the thumb screws and remove the cover of the cooling water reservoir. Use a watering can or the funnel and hose delivered with the system to refill the reservoir. Fill just up to 5mm
from the top of the reservoir.
Warning
Only use deionized water!
Put the cover to the tank and fix it with the thumb screws. Put the rear panel back, be careful with
the gas hoses and dont forget to plug the ground cable, USB and LAN if they were disconnected.
7-10
Reconnect the system to the mains power. Plug in the mains connector.
Fig. 7-6:
Operators Manual
Switch on the S8 TIGER like described in chapter 3 Switching the Instrument On and Off (on
page 3-1).
7-11
Operators Manual
Reason
Due to slightly different mixtures and tanks it is necessary to readjust the high voltage settings for the
system after a tank change. The realignment can be done by the help of a SPECTRAplus program
called SpectrometerAlignment.exe.
Select Meascontrol.
Type RC0.
Select S8 TOOLS.
Select ONLINE.
7-12
Fig. 7-7:
Operators Manual
Select Utilities.
7-13
Operators Manual
Fig. 7-8:
7-14
Fig. 7-9:
Operators Manual
7-15
Operators Manual
Warning
Close the MAIN VALVE on the tank and detach the SERVICE label. Screw on the valve
protector BEFORE removing the tank from the wall or cabinet. NEVER leave a tank standing WITHOUT the protection cap and WITHOUT any attachment to prevent the tank from
tipping over.
REMOVE the used tank before attaching a new tank.
Secure the new tank before removing the protection cap. Remove the protection cap and
attach the regulator. Tighten the connection! Be sure not to change the secondary pressure setting.
Check with some foamed up soap water for leaks in the connection to the regulator. The
regulator should be set to 0.5bars/7psi. You can only adjust the secondary side IF gas is
flowing.
7-16
Fig. 7-10:
Operators Manual
Wait until the Counter Gas set flag comes on (second red circle), see Fig. 7-11.
7-17
Operators Manual
Fig. 7-11:
If you encounter any errors adjust the pressure accordingly and/or check the supply.
7-18
Fig. 7-12:
Operators Manual
SpectrometerAlignment.
Verify that the sample position for the STG samples is correct, or enter the positions where they are
located at.
Hit start.
7-19
Operators Manual
Check for a message (a green status light means OK according to the specifications, a yellow
status light means WARNINGS and a red status light means PROBLEM).
If the message relates to warnings or errors check the SUMMARY1.tsv file and the SpectrometerAlignment.log file, which are located in the SPECPLUS\datafiles\info directory. If the Summary files
are not created please refer to the appendix on how to have them created.
In any case it is advisable to have a look at the SUMMARY1 file. Locate the summary section for your
current alignment and compare the settings for HV and Resolution with the previous settings obtained.
Please check the deviations carefully as they are typically in the range of only a few Volts. If the deviations are much higher something has gone wrong (e.g. the composition of the gas changed a lot).
If the resolution increased by more than 50% from the FIRST resolution the wire needs to be changed
(as well as the FC window).
S8 TIGER
Summary of Peak alignment function on 16-Sep-2008 23:07:05
Sample
Line
QuantExpress-STG
QuantExpress-STG
QuantExpress-STG
QuantExpress-STG
QuantExpress-STG
QuantExpress-STG
7-20
Initial Measured
KCps Status
Al KA1/S8-Check
Ca KA1/S8-Check
Fe KA1/S8-Check
Na KA1/S8-Check
Si KA1/S8-Check
Sb KA1/S8-Check
144.661
113.037
57.489
25.007
109.026
13.415
144.687
113.127
57.570
25.008
109.009
13.457
18.203
157.61
7.2848
249.56
244.73
40.92
Adjusted
Adjusted
Adjusted
Not Adjusted
Adjusted
Adjusted
Operators Manual
Sample_ID
Xtal
Soller
Det.
ZI1
ZI2
Int
Fe KA1
QuantExpress-GRA
LiF200
0.23
Zn KB1
QuantExpress-SQ1
LiF200
0.23
Zn KB1
QuantExpress-SQ1
LiF200
Zn KB1
QuantExpress-SQ1
LiF200
FWHM HV
Res.
ZI7
ZIMask
Gas
---
---
---
---
---
Scint
---
---
90.376
---
---
---
---
30.70
---
79.750
0.23
Gas
-0.961
10.194
77.645
0.346
---
---
0.23
Gas
---
---
---
---
1493
14
Zn KB1
QuantExpress-SQ1
LiF200
0.23
Scint
-0.958
-24.807
104.138
0.331
---
---
Zn KB1
QuantExpress-SQ1
LiF200
0.23
Scint
---
---
---
---
677
35
Zn KB1
QuantExpress-SQ1
LiF200
0.46
Gas
-0.938
10.216
122.697
0.589
---
---
Zn KB1
QuantExpress-SQ1
LiF200
0.46
Scint
-0.929
-24.787
93.377
0.563
---
---
Zn KB1
QuantExpress-SQ1
LiF220
0.23
Gas
-0.915
10.166
35.355
0.355
---
---
Zn KB1
QuantExpress-SQ1
LiF220
0.23
Gas
---
---
---
---
1454
15
Zn KB1
QuantExpress-SQ1
LiF220
0.23
Scint
-0.909
-24.824
114.593
0.339
---
---
Zn KB1
QuantExpress-SQ1
LiF220
0.23
Scint
---
---
---
---
650
36
Zn KB1
QuantExpress-SQ1
LiF220
0.46
Gas
-0.891
10.188
56.232
0.599
---
---
Zn KB1
QuantExpress-SQ1
LiF220
0.46
Scint
-0.886
-24.802
118.784
0.574
---
---
Na KA1
QuantExpress-STG
XS-55
0.23
Gas
-0.810
10.585
128.086
0.557
---
---
Na KA1
QuantExpress-STG
XS-55
0.23
Gas
---
---
---
---
1750
39
Na KA1
QuantExpress-STG
XS-55
0.46
Gas
-0.793
10.596
245.616
0.736
---
---
Na KA1
QuantExpress-STG
XS-55
1.00
Gas
-0.799
10.596
172.858
1.135
---
---
N KA1
QuantExpress-BN
XS-N
1.00
Gas
-0.848
10.538
8.503
2.196
---
---
N KA1
QuantExpress-BN
XS-N
1.00
Gas
---
---
---
---
1737
80
N KA1
QuantExpress-BN
XS-N
0.46
Gas
-0.848
10.538
4.534
2.013
---
---
C KA1
QuantExpress-GRA
XS_C
1.00
Gas
-0.903
10.393
30.081
2.428
---
---
C KA1
QuantExpress-GRA
XS_C
1.00
Gas
---
---
---
---
1746
91
7-21
Operators Manual
C KA1
QuantExpress-GRA
XS_C
0.46
Gas
-0.903
10.393
60.324
2.303
---
---
P KA1
QuantExpress-SQ2
PET
0.23
Gas
-0.987
10.213
72.728
0.377
---
---
P KA1
QuantExpress-SQ2
PET
0.23
Gas
---
---
---
---
1580
28
P KA1
QuantExpress-SQ2
PET
0.46
Gas
-0.970
10.220
116.644
0.586
---
---
Please note down the the HVs and Resolutions (Res.) in the LOG spreadsheet.
The changes in the example are small and show that all is ok. Drift correction is not necessary.
If you see the resolution go up there will be a problem with a contaminated FC wire and the foil.
If the HV goes down the detector is leaking and the foil needs to be replaced.
In order to replace a contaminated wire please contact the Bruker AXS service organisation or representative or refer to the FC Detector Service Procedure.
Run the check samples, which are defined for your methods and verify that the concentration
values stay within their limits. If they are out of bounds perform the drift correction for those methods as usual and rerun the respective check sample(s).
To check the instrument as compared to the installation you also can use the Instrument check application S8 CHECK.
7-22
Operators Manual
Start the samples by clicking the SEND CURRENT SAMPLE or SEND SELECTED SAMPLES or
SEND ALL SAMPLES.
Alternatively you can use TouchControl to start the sample from the window LOADER:
7-23
Operators Manual
Fig. 7-13:
7-24
Print the results and compare with the values from the acceptance test, or in lieu of this with results obtained at a day close to the installation of the software, If the values are within the prescribed limits no further actions are necessary, else it is necessary to run the drift correction/recalibration.
Operators Manual
7-25
Operators Manual
Consumables
Tab. 7-1:
List of consumables
Order number
C79298A3228C10
M34055A1851
K280C6
Disposable Resin for ion exchanger, 2.7 litre bag to refill K280C5
C71428A4D15
C70144A952B40
7KP19018CF
K120C1
Order number
S8 1kW
S8 3kW
S8 4kW
C79298-A3246-B99
C79298-A3204-C10
Vacuum Seal-O-ring
C71121-Z100-A8
Vacuum Seal-O-ring
C79298-A3204-C17
Sealing-Sample Input
C79298-A3204-C16
Sealing-Sample Chamber
C71121-Z100-A307
C71121-Z100-A307
7-26
Operators Manual
Order number
S8 1kW
S8 3kW
S8 4kW
C79298-A3248-C45
C79121-Z101-A21
Slot Ring
A12D2
K140-C19
7KP19018CF
K280-C6
Service Contracts
There are several service contract levels available to meet your specific needs, covering basic maintenance up to full service support with immediate response.
Please contact your local Bruker AXS representative for the options included in those packages and
prices. Availability may be subject to instrument location.
7-27
Operators Manual
7-28
To troubleshoot a problem concerning your spectrometer system you have the following different
ways:
Run a system check and comply with the trouble shooting flowchart in this section.
8-1
The first aid function is a component of the S8Tools. You can find the first aid button on the S8Tools
screen of the touchscreen.
Fig. 8-1:
8-2
Connecting to WebEx
There are two ways to connect to a Bruker / WebEx support session. The first is to use the link in the
email that the Bruker service representative will send. The second is to go to the Web address:
https://bruker-axs.webex.com/bruker-axs/
Both will be described in this document.
Connecting to an Emailed Web Address
The Bruker service representative (host) will send an email to one person, or several when starting a
Support Session. This email will look similar to Fig. 8-2.
8-3
Fig. 8-2:
Click the Web address (see Fig. 8-2) to open the WebEx Pre-Session Form in your Web browser
(i.e., Internet Explorer).
Note
If your browser does not automatically open the Web address after clicking the link, cut
and paste the Web address from the email directly into you browsers Address Bar.
8-4
Fig. 8-3:
8-5
Fig. 8-4:
8-6
Fill in the Support session number that the Host will provide, along with the information for the
primary user of the computer: name, email, and company. If you connected via an email link, the
Support session number will already be filled in.
Note
Your name, email, and company information will be saved for the next session. The Support Session number is unique for every session.
In your Web browser, save this page as a favourite, and name it Bruker AXS WebEx Support.
When starting WebEx, the Windows ActiveX Control WebEx Client may need to be loaded. In
this case, the screen shown in Fig. 8-5 will appear.
If the information bar (see Fig. 8-5) appears, click on it and select Install ActiveX Control.
8-7
Fig. 8-5:
8-8
Fig. 8-6:
Note
The WebEx client will take about two minutes to install, depending on your computer.
8-9
When installation is finished, the Welcome to Support Center window (Fig. 8-7) will appear. At
this point, you have joined the session.
Fig. 8-7:
8-10
When the WebEx host selects certain actions, dialogs similar to those in Fig. 8-8 may appear.
Fig. 8-8:
While sharing the desktop, the desktop will look similar to Fig. 8-9.
8-11
Fig. 8-9:
Desktop sharing
If you wish to disconnect from any action, click the green Sharing button at the top of the screen
(see Fig. 8-9) and select the appropriate Stop action. This will bring you back to the Welcome to
Support Center window.
8-12
When the session is over, select Leave Session from the Welcome window.
2.
Therefore Spectra Mail consists of different tabs that correspond to the different file types: the Unknown Samples tab, the Calibration Methods tab, the Evaluation Model tab (for EVM files). The
Output Results tab shows the compression process.
Here is how to create a Spectra Mail file:
1.
8-13
2.
3.
4.
5.
6.
7.
8.
Select the SSD files corresponding to the measurements that you want to include into the Spectra
Mail file.
9.
The resulting SPM file can easily be sent by e-mail to the Bruker AXS Hotline.
8-14
Fig. 8-10:
8-15
Menu
Command
Description
File
Save As
File
Exit
Edit
Select All
Edit
Unselect All
Edit
Copy to Clipboard
Options
Options
If checked, saves the full Measure.mdb into the stored Spectra Mail file. If not checked, the stored meaured.mdb contains
only the records corresponding to the selected SSD files.
Use this option if you want our Hotline to get a view of all
measured data.
Options
Options
8-16
Check the scans using the graphical display in EVAL2 and re-evaluate the sample.
(see SPECTRAplus V2 User's Manual)
Check all necessary supplies, such as counter gas for flow counters, helium gas and cooling water.
(see chapter "Maintenance" in the S8 TIGER Operator's Manual)
Run S8 CHECK / QE CHECK and comply with the following trouble shooting flowchart.
8-17
S8 CHECK
QE CHECK
All results OK
YES
Start measuring
your samples
NO
NO
Spectrometer system OK:
- no alarm messages
- right sample, sample clean
- right sample cup mask
- helium gas / counter gas supply
- cooling water supply
YES
S8 CHECK
QE CHECK
YES
NO
S8 CHECK
QE CHECK
All results OK
All results OK
YES
Start measuring
your samples
NO
Fig. 8-11:
8-18
9-1
Bruker AXS can also provide fully pre-calibrated analyzers matching the needs of your laboratory. You
select the norms and we set-up the specific calibrations. After on-site installation, the analyzers can
perform immediate routine analysis.
PETRO-QUANT BASIC is a unique solution containing:
256 ready to use cups with 4m Prolene film, samples just need to be poured into the cups
256 ready to use cups with 3,6m Mylar film, samples just need to be poured into the cups
100 pre-cut foils on carrier frames for liquid cups, 2.5m Mylar
100 pre-cut foils on carrier frames for liquid cups, 3.6m Mylar
Note
Please make sure that your Bruker AXS WDXRF spectrometer is ordered and equipped
with a Germanium analyzer crystal.
Order numbers:
A1001A1 - 1kW
A1003A1 - 3kW
A1004A1 - 4kW
9-2
the GEO-QUANT calibration for accurate and precise analysis of 27 trace elements in pressed
pellets. The calibration is done in the factory and is ready for use after installation of the
S8 TIGER 3 or 4 kW.
a simple, optimized sample preparation procedure and 3,900 grinding aid pellets
9-3
Note
For sample preparation a fine mill and a press are required. This package requires a LiF
220 anaylzer crystal.
Order numbers:
A1003A2 - 3 kW
A1004A2 - 4 kW
9-4
CEMENT-QUANT
CEMENT-QUANT for the S8 TIGER The XRF Solution for the Cement Industry
CEMENT-QUANT is a powerful analytical solution developed specifically for the cement industry. It
provides quick and easy quantitative elemental analysis for process and quality control laboratories
with either the world class S8 TIGER spectrometers.
CEMENT-QUANT produces both accurate and precise analytical results for analysing of cement,
clinker and raw mix. The measurement parameters for the reference calibration were modified to get
the best results and fulfill the requirements of ASTM C114 and ISO 29581.
Following elements are calibrated in reference calibration as oxide: Na2O, MgO, Al2O3, SiO2, P2O5,
SO3, K2O, CaO, TiO2, MnO, Fe2O3, ZnO and SrO. The calibration parameters are fixed and the
quality of the calibration will be checked with a control sample.
A stable glass sample is used to control the stability of spectrometer. This sample will measure every
day before starting the routine measurement.
The CEMENT-QUANT solution contains:
a suitcase with 21 international certified reference cement standards 10g each from NIST and a
set of Japanese cement standards (for establishing the reference calibration) and one quality control standard, two drift correction glass samples and one control sample for spectrometer stability
an installation CD
a user manual
Note
Please make sure that your Bruker AXS WDXRF spectrometer is ordered and equipped
with a curved Germanium (XS-Ge-C) and a XS-CEM analyzer crystal.
9-5
Order numbers:
A1001A6 1 kW
A1003A6 3 kW
A1004A6 4 kW
9-6
MLplus
MLplus Multilayer and Thin Film Software
MLplus is an optional software module extending the possibilities of SPECTRAplus for the analysis of
single layer and multilayer samples.
MLplus allows determining thickness and composition of layers in multilayer samples down to several
atomic layers (less than 1nm) and up to the m or even mm-range.
The number of layers which can be determined and the actual thickness range is determined by the
elements present in bulk and layers and not by software features.
Therefore this can be specified only for specific applications.
MLplus is using a full fundamental parameter approach for all calculations.
The determination of thickness and composition in MLplus is based on the standardless calibration of
SPECTRAplus, i.e. no specific multilayer standards are required. Nevertheless any available multilayer
standard can be used to optimize the results.
An interactive mode allows to define the sample structure (which element in which layer, typical values
for thickness and composition) and helps to find the optimum element lines for the measurement.
Once an interactive evaluation has been set up successfully, the parameters can be stored for automatic evaluation of similar samples.
Order number:
7KP78048AF
9-7
QUANT-EXPRESS
Standardless Analysis Package
QUANT-EXPRESS allows the automatic analysis of totally unknown samples without performing a
calibration. QUANT-EXPRESS provides unrivalled analytical flexibility. It can either perform a complete standardless analysis as fast scan in less than 2 minutes or an accurate analysis of samples to
measure even trace levels. According to users. requirements, QUANT-EXPRESS performs scan
measurements to provide full performance for interactive evaluation or can be set to peak and background mode for a faster time-to-result.
QUANT-EXPRESS automatically performs peak fitting, element and peak identification and accurate
determination of net intensities with background and line overlay corrections. The powerful variable
alpha matrix correction, unique to SPECTRAplus, automatically corrects for matrix effects as well as for
all different kind of samples (thin films, infinitely or non-infinitely thick samples). User input of sample
properties, known concentrations of elements and matrix components can be integrated to further
improve the analytical results.
QUANT-EXPRESS can be fully utilized for user-specific calibrations, based on the variable alpha
model, to allow the determination of elements where no standards are available. It therefore can be
integrated into SPECTRAplus to provide an XRF- expert system with the best line settings for any application.
Order number:
A1000A3
9-8
10
S8 TIGER Parts
Object description
A14B1
Filter / water
A23B277
A23B278
A15B62
C71121Z100A22
C71121Z100A307
O-ring
C71121Z100A32
Gasket
C71121Z100A68
Gasket
C71121Z100A69
C71121Z100A7
Gasket
C71121Z100A8
Gasket
C71121Z100A97
C79121Z101A21
Lip seal
C79121Z101A24
C79121Z101A25
10-1
S8 TIGER Parts
C79123Z1511C1
C79298A3194C8
K150C32
C79298A3196C28
Valve / proportional
A15B8
Stepper motor
C79298A3196C41
Valve / electro
C79298A3196C42
Valve / electro
C79298A3196C44
Valve / proportional
C79298A3196C45
Valve / solenoid
C79298A3200C1
C79298A3200C2
C79298A3200C26
SW/2 element
C79298A3200C28
C79298A3200C4
C79298A3200C53
Contactor
C79298A3200C54
Rectifier
C79298A3200C55
Relay / semiconductor
C79298A3200C61
Semiconductor relay; 25 A
C79298A3204C10
C79298A3204C12
C79298A3204C13
O-ring
C79298A3204C14
O-ring
C79298A3204C15
C79298A3204C16
C79298A3204C17
C79298A3204C18
O-ring
10-2
S8 TIGER Parts
C79298A3204C19
C79298A3204C20
O-ring
C79298A3204C41
O-ring
C79298A3204C60
O-ring
C79298A3220B220
Display / status
C79298A3230A35
Axial fan
C79298A3230A36
A15B215
Position sensor
C79298A3246B45
C79298A3246B99
A15B214
C79298A3248C45
Filter mat
E128C1
WI-NTC K45/6.8K/10%
K140C4
K150C25
K180C160
Magnetic sensor
K180C72
K180C81
K180C82
K280C15
K280C18
Axial fan
W75050U1104A102
A14D2
10-3
S8 TIGER Parts
Object description
A14B1
Filter / water
A23B277
A23B278
A15B62
C71121Z100A22
C71121Z100A307
O-ring
C71121Z100A32
Gasket
C71121Z100A68
Gasket
C71121Z100A69
C71121Z100A7
Gasket
C71121Z100A8
Gasket
C71121Z100A97
C79121Z101A21
Lip seal
C79121Z101A24
C79121Z101A25
C79123Z1511C1
C79298A3194C8
K150C32
C79298A3196C28
Valve / proportional
A15B8
Motor
C79298A3196C41
Valve / electro
C79298A3196C42
Valve / electro
C79298A3196C44
Valve / proportional
10-4
S8 TIGER Parts
C79298A3196C45
Valve / solenoid
C79298A3200C1
C79298A3200C2
C79298A3200C26
SW/2 element
C79298A3200C28
C79298A3200C4
C79298A3200C53
Contactor
C79298A3200C54
Rectifier
C79298A3200C55
Relay / semiconductor
C79298A3200C61
Semiconductor relay; 25 A
C79298A3204C10
C79298A3204C12
C79298A3204C13
O-ring
C79298A3204C14
O-ring
C79298A3204C15
C79298A3204C16
C79298A3204C18
O-ring
C79298A3204C19
C79298A3204C20
O-ring
C79298A3204C60
O-ring
C79298A3220B220
Display / status
C79298A3230A36
A15B215
Position sensor
C79298A3246B45
C79298A3246B99
A15B214
10-5
S8 TIGER Parts
E128C1
WI-NTC K45/6.8K/10%
K140C4
K150C25
K180C160
Magnetic sensor
K180C72
K180C81
K180C82
K200C9
K280C15
K280C26
W75050U1104A102
W79054L1011T160
Fuse 1.6A
A14D2
10-6
S8 TIGER Parts
Object description
A23B253
Safety controller
A23B232
K300C38
K300C5
A23B102-S
A23B221-S
A23B216
UMC board
A23B217
Power IO board
K300C39
A15B57
C79298A3220B211
A12D410
A12B56
Inverter
A12D442
10-7
S8 TIGER Parts
A15C250
A15C251
C79298A3246B59
C79298A3246B58
C79298A3246B49
C79298A3246B57
C79298A3246B56
C79298A3246B55
C79298A3246B54
C79298A3246B53
C79298A3246B52
C79298A3246B51
C79298A3246B50
10-8
Order number
C79298A3246D100
C79298A3246D101
C79298A3246D102
C79298A3246D103
C79298A3246D104
C79298A3246D105
C79298A3246D106
C79298A3246D107
C79298A3246D108
S8 TIGER Parts
C79298A3246C95
34 mm diameter, steel
C79298A3246C96
28 mm diameter, steel
C79298A3246C185
23 mm diameter, steel
C79298A3246C181
18 mm diameter, steel
C79298A3246C186
8 mm diameter, steel
A15B72
5 mm diameter, graphite
10-9
S8 TIGER Parts
Order number
C79298A3246B69
C79298A3177B75
C79298A3246B60
C79298A3114B71
A15B49
C79298A3114D7
Distance plastic rings for sample centering (20 rings per unit)
C79298A3173D30
Order number
7KP90018BS
7KP90018BE
C79298A3173B23
C79298A3173B22
Process sample cassette for fused beads, (51.5 mm x 8.5 mm), 1 off
A15B20
51.5 mm sample cassette for 40mm samples and liquid cups, 1 off
10-10
S8 TIGER Parts
C79298A3246B22
C79298A3246B23
C79298A3246B62
C79298A3246B61
C79298A3246B63
C79298A3246C61
C79298A3246C60
C79298A3246C87
C79298A3246C180
C79298A3246C93
C79298A3246C94
Order number
K230C201
K230C202
K230C203
K230C204
K230C205
K230C206
K230C207
K230C208
10-11
S8 TIGER Parts
C79213A3000C100
C79213A3000C101
C79213A3000C102
C79213A3000C104
C79213A3000C105
C79213A3000C106
C79213A3000C107
Liquid cups
7KP19018FA
P40-A10
RDL-1000-My
Liquid cups for S8 with Mylar 2,5 mm, 35/40mm I.D/O.D, 256 off
RDL-1000-Pro
Liquid cups for S8 with Prolene 4 mm, 35/40mm I.D/O.D, 256 off
10-12
Order number
C71298A18C286
7KP19018BR
7KP78048CD
7KP19018BZ
BRE-442
Kapton, 7.5m
C71298A18C262
C71298A18C264
7KP19018BX
7KP19018BS
C71298A18C288
7KP19018BU
7KP19018BT
7KP19018BG
7KP78048CB
BRE-CH-3024-10+
S8 TIGER Parts
10-13
S8 TIGER Parts
7KP78048AW-S8
7KP78048AK
Order number
A15D112
A15D113
A15D100
A15D110
A15D111
Order number
7KP18018AB
7KP18018EA
7KP18018CA
7KP18018CB
7KP18018CC
7KP18018CD
10-14
11
S8 TIGER Specifications
Analytical Performance
The S8 TIGER combines high analytical performance, space saving and cost efficiency. Its outstanding light element analysis capacity is based on innovative multilayer analyzer crystals, appropriate collimators and the optimized close sample/X-ray tube coupling. Besides this, the S8 TIGER is
optimized by the 75 m ultra thin X-ray tube window, up to 170 mA excitation and the high transmission thin detector window technology.
Features:
Sample form: powder, solid, liquid, paste, coating, slurry, film, filter deposit, etc.
11-1
Specifications
Instrument Specification
Tab. 11-1:
Instrument specification
Technical Data
Sample size
Samples
Analysis range
Beryllium to Uran
Concentration range
Precision
TouchControlTM
SampleCareTM
Filter
Collimator
Masks
Analyzer crystals
11-2
Specifications
Technical Data
Flow proportional
counter
Scintillation counter
NaI (Ti)
Energy range: > 4 keV (Sc U)
PHA channel: 512
Counting rate: > 2000 cps
DynaMatch: more than 10 Mio cps
Detector gas
Goniometer
Vacuum pump
Integrated
S8 TIGER 1 kW
50 kV
50 mA
S8 TIGER 3 kW
60 kV
150 mA
Power requirements
208 - 240 V
(1P/3P)
50/60 Hz
208 V, 60 Hz (1P/3P)
230 V, 50/60 Hz (3P)
Compressed air
Not required
No cooling water
S8 TIGER 4 kW
60 kV
170 mA
Cooling water
Water consumption automatically regulated and minimized,
short term interruptions are compensated
11-3
Specifications
Technical Data
Dimensions
(height x width x depth)
135 cm x 84 cm x
90 cm;
53.1 x 33.1 x
35.4
11-4
476 kg / 1047 lb
Specifications
Certified Compliance
Tab. 11-2:
Certified compliance
Certified Compliance
Electrical Safety
Electromagnetic Compatibility
Radiation Safety
Class I
IP 2 0
Machinery Safety
11-5
Specifications
Electrical Specification
Tab. 11-3:
Electrical specification
Three-phase supply:
3 x 120 VAC (+/- 10 %)
3 x 230 VAC (+/- 10 %)
3 x 240 VAC (+6 % / - 10 %)
One-phase supply (on request only):
208 VAC (-10% / +10 %)
230 VAC (+/- 10 %)
240 VAC (+ 6 % / - 10 %)
Frequency Range
47Hz 63Hz
Power Consumption
S8-1kW
S8-3kW
S8-4kW
2.8 kVA
5.2 kVA (3-phases mains supply)
7 kVA (3-phases mains supply)
II
11-6
Specifications
High-voltage Generator
Tab. 11-4:
1 kW
Output voltage
Output current
3 kW
Output voltage
Output current
4 kW
Output voltage
Output current
11-7
Specifications
X-ray Tubes
Below is a list of X-ray tubes recommended for use with your S8 TIGER. There are more options
available (e.g. Mo as target material for 3 or 4 kW operation). For availability of a specific tube types
target materials, please contact your Bruker AXS Service or Sales Representative. In most countries
and states, you will need an individual on-site X-ray test if you choose to use a target material heavier
than rhodium. The table lists the recommended operation parameters.
Tab. 11-5:
1 kW operation
Type
AG22
50 kVDC
Maximum current
50 mADC
Maximum power
1 kW
Anode material
Rhodium
3 and 4 kW operation
Type
OEG95LT
60 kVDC
Maximum current
Maximum power
4 kW
Anode material
Rhodium
11-8
Specifications
Laboratory Requirements
S8 TIGER Mechanical Specifications
Tab. 11-6:
Dimensions
S8-1kW
Height
S8-3kW/4kW
135 cm / 53.1''
Depth
90 cm / 35.4"
104 cm / 41"
Width
84 cm / 33.1"
Screen
49 cm / 19.3"
Weight
~ 446 kg / 981 lb
~ 476 kg / 1047 lb
Environmental Specifications
Tab. 11-7:
Environmental specifications
Specifications
Data
24 C / 75F
Relative humidity
Atmospheric pressure
Spectrometry system applicable at all terrestrial locations under atmospheric air pressure conditions up to an altitude of 3000 meters (9800
feet).
11-9
Specifications
Gas Consumption
Helium/Nitrogen Operation
Tab. 11-8:
Gas specifications
Helium or Nitrogen
~ 0.5 l/min
~ 5 - 10 l
Helium or Nitrogen
~ 1 - 4 l/min
~ 10 - 30 l
atmospheric
~ 2l/hr
Accessories
Below is a list of additional equipment that may enhance the systems reliability or increase its uptime
if the laboratory infrastructure is not reliable enough.
11-10
Specifications
10 kVA / 8 kW
Backup time
Dimensions
Weight
250 kg
Input
Output
3 x 32 A, 50/60 Hz
8 kVA / 6.8 kW
Backup time
Dimensions
Weight
260 kg
Input
Output
3 x 32 A, 50/60 Hz
Note
Configure the output voltage of the Toshiba UPS to 240V!
208V are preset by the manufacturer.
11-11
Specifications
15 kVA / 12 kW
Backup time
Dimensions
Weight
260 kg
Input
Output
2 x 40 A, 50/60 Hz
14 kVA / 11.9 kW
Backup time
Dimensions
Weight
281 kg
Input
Output
2 x 40 A, 240V, 50/60 Hz
Note
Configure the output voltage of the Toshiba UPS to 240V!
208V are preset by the manufacturer.
11-12
Specifications
15 to 25C (adjustable)
Temperature stability
1 C
Tank capacity
40 l
Water flow
300 l/h
Required electricity
400V, 3-phase, 50 Hz
Power consumption
3.52 kW
Dimensions
Weight (empty)
220 kg
Compliance
CE
Degree of protection
IP44
The unit is delivered with a 10 m power cable, a 10 m controller cable, and two 10 m water hoses. For
internal pressure compensation, it is equipped with a bypass to improve operation reliability. Technical
specifications can be changed with additional parts.
11-13
Specifications
11-14
12
References
Description
Order number
Manuals:
Introduction to X-ray Fluorescence Analysis, English
DOC-M84-EXX001
DOC-M84-DXX001
DOC-M90-ZXX006
DOC-M80-ZXX031
DOC-M80-EXX032
DOC-M80-EXX042
DOC-M80-DXX042
DOC-M80-FXX042
plus
SPECTRA
DOC-M80-EXX038
Brochures:
S8 TIGER Spectrometry Solutions Brochure
DOC-B80-EXS009
12-1
References
Lab Reports:
XRF 79 - S8 TIGER: Process and Quality Control in Cement Production
DOC-L80-E00079
DOC-L80-E00080
DOC-L80-E00090
Posters:
Periodic table, DIN A3
DOC-P84-E00011
DOC-P84-E00010
S8 TIGER
DOC-P80-EXS025
12-2
13
Index
Index
A
Accessories
Advanced search
Alarm
Application
Application chooser
Application support
Auto start up
D
11-10
5-14
5-29
5-8
5-9
10-14
4-1
5-2
Elements
Environmental specifications
3-6
11-9
C
Calibration
Certified compliance
Changing detector gas
Changing gas
Changing password
Collimator
Consumables
Cooling unit
internal
Crystal
5-5
7-3
7-12
11-9
12-1
B
Button
Date selector
Deionized water
Detector gas
Dimensions
Documentation
6-26, 10-14
11-5
7-12
7-25
5-24
10-11
7-26
7-3
7-3
10-11
File formats
First aid
First aid button
Front view
6-26
8-1
8-1
3-9
G
Gas change
Gas consumption
General S8 TIGER safety precautions
7-25
11-10
1-5
13-1
Index
H
Height
O
11-9
I
Instrument components
Instrument Performance Verification
Instrument specification
10-12
3-10
5-8, 6-10
5-6
5-27, 10-11
6-20
6-16
6-9
11-9
13-2
7-12
7-1
5-24
4-6
Q
QUANT-EXPRESS
6-1
Radiation safety
Recommended X-ray tubes
References
Result database
Results
11-5
11-8
11-5
5-13
5-8
N
Network connection
M
Mask
Measurement method
Measurement mode
Measurement with QUANT-EXPRESS
Mechanical specifications
3-3
3-2, 10-8
P
3-8
6-4
11-2
L
Liquid sample
Loader
Loader screen
Login information
Operation principle
Optional parts
3-21
S8Tools
Safety precautions
Sample
Diameter
Liquid sample
Material
5-26, 8-2
1-5
6-1
6-18
10-12
6-12
Index
Preparation
6-14
Sample mask diameter
6-19
Unknown sample
6-9
Sample Cups
7-1
Sample holder
10-8
Sample holder mask
10-9
Sample Holders
7-1
Sample magazine
5-8
Schedule
5-18
Screen keyboard
5-4
Screen selector
5-3
Selection list
5-3
Shutdown
4-7
Spare parts
10-1
Maintenance spare parts package
7-26
Optional parts
10-8
Spare parts packages
10-1
Spectra Mail
8-13
Commands
8-16
Menus
8-16
plus
SPECTRA
File formats
6-26
SPECTRAplus
3-21, 6-3, 6-26
Status
5-16
Status bar
5-5
Status screen
5-11, 5-16
Switching on/off
4-1
5-4
5-2
5-19
5-26, 8-2
5-12
3-3, 6-3
5-1, 5-8, 7-2
10-14
8-1, 8-17
4-1
U
Unknown sample
UPS
User editor
6-9
11-11
5-23
W
Warning symbols
Warnings
Water chiller
WebEx
1-1
1-1
11-13
8-3
T
Technical specifications
Electrical specifications
Helium/Nitrogen operation
High-voltage generator
Textbox
Thumbwheel
Tools
S8Tools
Tools screen
TouchControl
Touchscreen
Training courses
Troubleshooting
Turning on/off
11-6
11-10
11-7
11-5
11-8
13-3
Index
13-4
Atomic number
1.01
H
Hydrogen
6.94
Li
0.534
22.99
Beryllium
1.85
KA 1.2 0.108
74.34
Na
Sodium
KA 1.2
1.0
KA 1.2
1.0
KB 1
1.1
25.00
55.08
24.40
19
39.10
Magnesium
KA 1.2
1.3
KA 1.2
1.3
KB 1
1.3
3.3
KA 1.2
3.3
KB 1
3.6
37
85.47
KA 1.2
3.7
KA 1.2
3.7
KB 1
4.0
LA 1
0.3
LB 1
0.3
1.53
13.4
KA 1.2
13.4
KB 1
15.0
LA 1
1.7
LB 1
1.8
KA 1.2
13.4
KB 1
15.0
26.61
37.99
23.75
113.68
108.08
62.00
54.79
132.91
56
LA 1
4.3
LB 1
4.6
(223)
KA 1.2
4.1
KB 1
4.5
LA 1
0.4
LB 1
0.4
39
2.64
14.9
KA 1.2
14.9
KB 1
15.8
KB 1
16.7
LA 1
1.8
LB 1
1.9
KA 1.2
14.1
KB 1
15.8
25.15
35.85
22.42
103.45
98.53
58.31
51.54
KA 1.2
14.1
137.33
57
KA 1
32.1
LA 1
4.5
LB 1
4.8
88
12.0
LB 1
14.8
29.65
24.06
LA 1
12.3
LB 1
15.2
4.5
KA 1.2
4.5
KB 1
4.9
LA 1
0.5
LB 1
0.5
86.12
36.67
77.25
59.83
58.97
40
1.9
LB 1
2.0
KA 1.2
14.9
KB 1
16.7
KA 1
33.4
LA 1
4.7
LB 1
5.0
KA 1.2
5.0
KB 1
5.4
LA 1
0.5
LB 1
0.5
41
92.91
6.51
15.8
16.6
KA 1.2
16.6
KB 1
17.7
KB 1
18.6
LA 1
2.0
LA 1
2.2
LB 1
2.1
LB 1
2.3
KA 1.2
15.8
KA 1.2
16.6
KB 1
17.7
22.50
32.04
20.07
87.96
83.76
51.75
45.28
KA 1.2
15.8
KB 1
18.6
178.49
73
Hafnium
KA 1
55.8
LA 1
7.9
LB 1
9.0
MA 1
1.6
LA 1
7.9
LB 1
9.0
6.11
76.92
123.17
69.12
52.35
51.49
13.31
8.95
45.87
39.90
119.17
121.27
99.46
Niobium
KA 1.2
5.4
KA 1.2
5.4
KB 1
6.0
LA 1
0.6
LB 1
0.6
42
8.57
Molybdenum 10.22
17.5
KA 1.2
17.5
KB 1
19.6
LA 1
2.3
LB 1
2.4
KA 1.2
17.5
KB 1
19.6
180.95
74
Ta
LA 1
8.2
LB 1
9.3
8.68
44.41
38.47
112.11
115.36
94.92
1.7
8.1
LB 1
9.3
Manganese
KA 1.2
5.9
KA 1.2
5.9
KB 1
6.5
LA 1
0.6
LB 1
0.6
18.4
KA 1.2
18.4
KB 1
20.6
LA 1
2.4
LB 1
2.5
KA 1.2
18.4
KB 1
20.6
183.84
75
19.25
LA 1
8.4
LB 1
9.7
43.01
37.12
106.03
110.13
90.75
1.8
LA 1
8.4
LB 1
9.7
KA 1.2
6.4
KA 1.2
6.4
Technetium
KB 1
7.1
LA 1
0.7
LB 1
0.7
44
11.5
Ruthenium
KA 1.2
19.3
KA 1.2
19.3
KB 1
21.7
LA 1
2.6
LB 1
2.7
KA 1.2
19.3
KB 1
21.7
186.21
76
Re
LA 1
8.7
LB 1
10.0
41.68
35.82
100.66
105.43
86.90
1.8
8.7
LB 1
10.0
Cobalt
Osmium
LA 1
8.9
LB 1
10.4
MA 1
1.9
LA 1
8.9
LB 1
10.4
28
6.9
KA 1.2
6.9
KB 1
7.7
LA 1
0.8
LB 1
0.8
45
58.69
29
Ni
8.56
KA 1.2
52.79
77.89
47.46
33.77
33.12
102.91
Nickel
42.51
7.5
KA 1.2
7.5
KB 1
8.3
LA 1
0.9
LB 1
0.9
KA 1.2
7.5
KB 1
8.3
46
63.55
Copper
48.67
71.26
43.72
30.70
30.08
134.33
112.81
KA 1.2
8.0
KA 1.2
8.0
KB 1
8.9
LA 1
0.9
LB 1
1.0
KA 1.2
8.0
KB 1
8.9
106.42
47
Pd
12.37
Rhodium
12.41
Palladium
12.02
Silver
KA 1.2
20.2
21.2
22.2
KA 1.2
21.2
KA 1.2
22.2
KB 1
22.7
KB 1
23.8
KB 1
24.9
LA 1
2.7
LA 1
2.8
LA 1
3.0
LB 1
2.8
LB 1
3.0
LB 1
3.2
KA 1.2
20.2
KA 1.2
21.2
KA 1.2
22.2
KB 1
22.7
KB 1
23.8
16.71
23.73
14.85
59.95
56.63
37.94
33.60
KA 1.2
20.2
17.51
24.87
15.57
63.46
60.05
39.82
35.51
KA 1.2
KA 1.2
KB 1
24.9
190.23
77
192.22
78
195.08
79
Ir
22.61
40.42
34.59
95.64
101.16
83.33
Iridium
LA 1
9.2
LB 1
10.7
MA 1
2.0
LA 1
9.2
LB 1
10.7
30
Pt
Platinum
21.46
Zinc
KA 1.2
8.6
KA 1.2
8.6
KB 1
9.6
LA 1
1.0
LB 1
1.0
KA 1.2
8.6
KB 1
9.6
107.87
48
39.21
33.42
91.50
97.24
80.01
LA 1
9.4
LB 1
11.1
MA 1
2.0
LA 1
9.4
LB 1
11.1
38.06
32.29
87.53
93.62
76.91
LA 1
9.7
LB 1
11.4
MA 1
2.1
LA 1
9.7
LB 1
11.4
1.5
KA 1.2
1.5
KB 1
1.5
KA 1
1.5
KA 1.2
0.28
KA 1.2
0.28
14
9.2
KA 1.2
9.2
KB 1
10.3
LA 1
1.1
LB 1
1.1
KA 1.2
9.2
KB 1
10.3
112.41
49
8.69
KA 1.2
23.1
24.2
23.1
KA 1.2
24.2
KB 1
26.1
KB 1
27.3
LA 1
3.1
LA 1
3.3
LB 1
3.3
LB 1
3.5
KA 1.2
23.1
KA 1.2
24.2
KB 1
26.1
15.27
21.66
13.55
53.82
50.63
34.65
30.59
KA 1.2
KA 1.2
KB 1
27.3
200.59
81
196.97
80
Mercury
36.96
31.22
83.83
90.27
73.98
LA 1
10.0
11.8
LB 1
MA 1
2.1
LA 1
10.0
LB 1
11.8
1.7
KB 1
1.8
KA 1
1.7
69.72
32
5.91
9.9
KA 1.2
9.9
KB 1
11.0
LA 1
1.2
LB 1
1.2
KA 1.2
9.9
KB 1
11.0
114.82
50
35.90
30.19
80.49
87.13
71.22
LA 1
10.3
LB 1
12.2
MA 1
2.3
LA 1
10.3
LB 1
12.2
28.09
15
2.0
KA 1.2
2.0
KB 1
2.1
KA 1
2.0
72.64
33
5.32
10.5
KA 1.2
10.5
KB 1
11.7
LA 1
1.3
LB 1
1.3
KA 1.2
10.5
KB 1
11.7
118.71
51
25.2
KA 1.2
25.2
KB 1
28.5
LA 1
3.4
LB 1
3.7
KA 1.2
25.2
KB 1
28.5
204.37
82
34.92
29.20
77.30
84.20
68.62
Lead
10.6
LB 1
12.6
MA 1
2.3
LA 1
10.6
LB 1
12.6
16
KA 1.2
2.3
KA 1.2
2.3
KB 1
2.5
KA 1
2.3
74.92
34
35
12.5
KB 1
13.3
LA 1
1.4
LA 1
1.5
LB 1
1.4
LB 1
1.5
KA 1.2
11.2
KA 1.2
11.9
KB 1
12.5
KB 1
13.3
127.60
53
26.4
KB 1
29.7
208.98
84
33.00
27.34
71.67
78.86
63.82
KA 1.2
0.85
KA 1.2
0.85
35.45
18
6.23
KA 1.2
27.4
28.6
27.4
KA 1.2
28.6
KB 1
31.0
KB 1
32.3
LA 1
3.8
LA 1
3.9
LB 1
4.0
LB 1
4.2
KA 1.2
27.5
KA 1.2
28.6
KB 1
31.0
12.87
18.23
11.40
44.20
41.21
29.02
25.67
KA 1.2
KA 1.2
KB 1
32.3
(209)
85
Ar
Argon
KA 1.2
3.0
KA 1.2
3.0
KB 1
3.2
79.90
36
9.32
LA 1
11.1
LB 1
13.4
32.11
26.47
Kr
Krypton
12.6
KA 1.2
12.6
KB 1
14.1
LA 1
1.6
LB 1
1.6
126.90
54
LA 1
11.1
LB 1
13.4
13.9
131.29
Xenon
0.005
12.35
17.50
10.94
42.22
39.27
27.84
24.62
KA 1.2
29.7
KA 1.2
29.7
KB 1
33.6
LA 1
4.1
LB 1
4.4
(210)
86
11.85
16.81
10.50
40.37
37.42
(222)
Rn
Astatine
LB 1
0.003
28.21
40.32
25.20
126.81
120.14
Xe
4.93
11.4
57.32
79.86
52.79
83.80
KA 1.2
Iodine
LA 1
0.002
29.96
42.88
26.79
17.52
16.99
70.69
62.39
At
Polonium
30.8
69.1
39.95
0.003
3.12
Tellurium
0.0009
65.49
92.76
60.49
Po
9.81
Neon
Bromine
Te
KA 1.2
0.001
38.91
90.60
Br
KB 1
3.8
13.0
78.96
11.9
3.6
LB 1
2.8
KA 1.2
29.7
2.4
KB 1
11.2
LB 1
10.8
2.6
KA 1.2
LA 1
13.0
KA 1.2
11.9
KB 1
LA 1
2.6
KA 1.2
6.64
MA 1
KA 1.2
4.81
52
Bismuth
Chlorine
31.89
45.71
28.52
18.82
18.28
75.84
66.86
121.76
20.18
Cl
2.07
75.85
110.69
70.28
22.53
11.2
13.41
19.01
11.89
46.34
43.31
30.28
26.78
LB 1
17
KA 1.2
26.4
10.8
32.07
Selenium
26.4
LA 1
0.68
Se
5.78
33.99
48.83
30.44
20.25
19.71
81.69
71.90
Bi
33.93
28.25
74.41
81.45
66.16
0.68
KA 1.2
0.0002
10
Ne
Fluorine
KA 1.2
Sulphur
KA 1.2
83
0.001
50.90
25.10
S
1.82
89.56
141.03
83.20
25.88
KA 1.2
207.20
11.34
LA 1
30.97
Antimony
13.99
19.84
12.41
48.64
45.56
31.62
27.97
Pb
11.86
0.53
Sb
7.29
KA 1.2
0.53
KA 1.2
As
KA 1.2
19.00
Oxygen
KA 1.2
Arsenic
36.32
52.31
32.56
21.88
21.33
88.44
77.64
Tin
14.61
20.72
12.96
51.12
47.99
33.04
29.24
0.001
P
KA 1.2
16.00
70.14
33.60
Phosphorus
109.21
144.58
101.67
30.04
Sn
Tl
Thallium
0.39
Ge
KA 1.2
7.31
13.53
0.39
KA 1.2
Germanium
38.91
56.20
34.90
23.69
23.12
96.35
84.24
Indium
Hg
19.28
KA 1.2
Nitrogen
KA 1.2
2.33
1.7
In
Cadmium
15.96
22.66
14.18
56.75
53.50
36.19
32.04
2.27
107.24
43.31
Silicon
KA 1.2
Gallium
KA 1.2
14.01
Si
2.70
Ga
7.13
41.80
60.58
37.52
25.74
25.18
105.87
91.98
Carbon
145.12
17.45
131.86
35.31
31
Cd
10.55
Gold
KA 1.2
Zn
8.93
Au
22.65
65.41
45.02
65.55
40.45
28.06
27.45
117.85
101.26
Ag
18.37
26.10
16.34
67.33
63.82
41.84
37.07
26.98
Aluminium
Cu
8.91
KA 1.2
Rh
Os
21.02
LA 1
101.07
19.29
27.42
17.17
71.62
67.98
44.03
39.01
58.93
Co
7.87
57.51
85.73
51.72
37.30
36.60
Ru
Rhenium
MA 1
27
12.01
Al
nL = 2d sin Q
Fe
Iron
(98)
KA 1.2
55.85
7.44
Tc
20.28
28.84
18.06
76.41
72.63
46.39
41.11
Tungsten
MA 1
26
62.96
95.20
56.63
41.45
40.65
43
16.65
57.6
54.94
Mn
7.15
69.34
107.11
62.35
46.33
45.49
95.94
KA 1.2
KA 1
LA 1
25
Mo
21.35
30.38
19.03
81.81
77.84
48.95
43.39
Tantalum
MA 1
52.00
Chromium
Nb
Hf
6.15
14.96
82.90
75.27
89
7.00
31.26
25.64
Radon
0.01
LA 1
11.7
LB 1
14.3
30.44
24.84
76.41
61.59
(227)
Actinium
10.07
LA 1
12.7
LB 1
15.7
Name, material
5.0
KA 1.2
72
Ac
5.5
28.89
23.31
91.22
KA 1.2
KA 1.2
138.91
24
Cr
Vanadium
Zirconium
23.79
33.90
21.19
95.07
90.57
54.95
48.57
Lanthanium
50.94
V
4.54
KA 1.2
La
3.59
15.54
87.16
79.24
(226)
Radium
4.47
LA 1
23
Zr
Yttrium
Ra
1.87
LA 1
88.91
14.1
47.87
Titanium
KA 1.2
Barium
22
Ti
2.99
97.68
40.59
87.28
69.50
68.67
KA 1.2
Fr
Francium
Analyser crystal
4.1
Ba
1.87
16.16
91.80
83.58
87
87.62
KA 1.2
Strontium
Cs
Caesium
44.96
Scandium
Sr
KA 1.2
30.9
21
2.34
0.18
KA 1.2
13
Bragg angle (2 Q)
Boron
Energy (keV)
Sc
1.54
113.07
45.20
100.21
82.68
81.61
38
Rubidium
KA 1
40.08
10.81
1.24
E (keV)
L (nm) =
29.96
11.9
KA 1.2
20.70
136.72
20.00
Calcium
Rb
55
Spectral line
Ca
0.86
136.67
50.69
118.10
Density (g/cm3)
1.74
20
K
Potassium
KA 1.2
24.31
Mg
0.97
Bromine 3.12
12
4.00
Helium
79.90
Br
Be
Lithium
11
9.01
He
35
0.0007
Atomic weight
28.17
22.59
2d-value (nm)
XS-B
La/B4C multilayer
19.0
XS-C
Ti02/C multilayer
XS-N
XS-55
XS-CEM
Lanthanides, Actinides:
58
140.12
Ce
Cerium
LA 1
4.8
12.0
LA 1
4.8
Ni/BN multilayer
11.0
LB 1
5.3
W/Si multilayer
5.5
Specific structure
2.75
2.576
59
90
Thallium biphthalate
ADP
1.064
PET
Pentaerythrite
0.874
InSb
Indium antimonide
0.748
Thorium
Ge
Germanium
0.653
LA 1
13.0
LA 1
13.0
Praseodymium 6.77
LA 1
5.0
LA 1
5.0
LB 1
5.5
91
Th
LiF (200)
Lithium fluoride
0.403
LB 1
16.2
LiF (220)
Lithium fluoride
0.285
LA 1
13.0
LiF (420)
Lithium fluoride
0.180
LB 1
16.2
60
Pr
6.77
128.13
79.00
71.61
232.04
TIAP
140.91
11.72
39.22
27.46
21.90
64.12
50.29
119.69
75.41
68.23
231.04
LA 1
13.3
LA 1
13.3
LB 1
16.7
LA 1
13.3
LB 1
16.7
38.23
26.79
21.24
62.39
48.68
61
Nd
Neodymium
LA 1
5.2
LA 1
5.2
LB 1
5.7
92
Pa
Protactinium 15.37
144.24
7.01
112.66
72.12
65.10
238.03
LA 1
13.6
LA 1
13.6
LB 1
17.2
LA 1
13.6
LB 1
17.2
62
150.36
63
Pm Sm
Promethium
LA 1
5.4
LA 1
5.4
LB 1
6.0
93
U
Uranium
(145)
7.26
106.48
69.03
62.19
(237)
Samarium
LA 1
5.6
LA 1
5.6
LB 1
6.2
94
Np
18.45
Neptunium
37.29
26.14
20.59
60.75
47.13
LA 1
14.0
LA 1
14.0
LB 1
17.8
7.52
101.12
66.22
59.49
(244)
Pu
20.25
Plutonium
36.38
25.51
19.96
LA 1
14.3
LA 1
14.3
LB 1
18.3
151.96
64
Eu
65
Gd
Europium
5.24
LA 1
5.9
LA 1
5.9
LB 1
6.5
96.23
63.55
56.96
95
157.25
(243)
Gadolinium
LA 1
6.1
LA 1
6.1
LB 1
6.7
96
7.90
91.87
61.09
54.59
(247)
19.84
Americium
LA 1
14.6
LA 1
14.6
LB 1
18.8
13.69
Curium
34.65
24.32
18.81
LA 1
15.0
LA 1
15.0
LB 1
19.4
66
Tb
Terbium
LA 1
6.3
LA 1
6.3
LB 1
7.0
97
Am Cm
35.50
24.90
19.39
158.93
162.50
67
Dy
8.23
87.87
58.78
52.36
(247)
Dysprosium
LA 1
6.5
LA 1
6.5
LB 1
7.2
8.55
84.15
56.59
50.27
98
Bk
(251)
Cf
13.51
Berkelium
14.79
Californium
33.84
23.76
18.26
LA 1
15.3
15.7
15.3
LA 1
15.7
LB 1
20.0
33.04
23.20
17.74
LA 1
LA 1
LB 1
20.6
164.93
Ho
Holmium
8.80
LA 1
6.7
LA 1
6.7
LB 1
7.5
80.73
54.53
48.30
99
(252)
Es
15.1
32.38
22.68
17. 22
68
Einsteinium
167.26
69
Er
Erbium
LA 1
7.0
LA 1
7.0
LB 1
7.8
100
168.93
Tm
9.07
77.58
52.60
46.43
(257)
Thulium
LA 1
7.2
LA 1
7.2
LB 1
8.1
9.32
74.63
50.78
44.67
101
(258)
Fm Md
13.5
Fermium
70
Mendelevium
173.04
71
Yb
Ytterbium
LA 1
7.4
LA 1
7.4
LB 1
8.4
LA 1
7.4
LB1
8.4
102
6.97
71.89
49.06
42.99
136.36
110.04
(259)
No
Nobelium
174.47
Lu
Lutetium
9.84
LA 1
7.7
LA 1
7.7
LB 1
8.7
69.31
47.42
41.40
128.12
104.46
LA 1
7.7
LB1
8.7
103
Lr
Lawrencium
(262)