Você está na página 1de 36

GE Healthcare

Chilin 19" Color LCD Monitor Option Installation


Service Documentation

5178234-100
Rev 2
2006-2007, General Electric Company,
All Rights Reserved.

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2

Page 2

CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2

CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

LEGAL NOTES
TRADEMARKS
Adobe, the Adobe logo, Acrobat, the Acrobat logo, Exchange, and PostScript are trademarks of
Adobe Systems Incorporated or its subsidiaries and may be registered in certain jurisdictions.
Microsoft is a registered trademark and Windows is a trademark of Microsoft Corporation.
All other products and their name brands are trademarks of their respective holders.

COPYRIGHTS
All Material, Copyright 2006-2007 by General Electric Company, Inc. All rights reserved.

Legal Notes

Page 3

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2

Page 4

CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

Legal Notes

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2

CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

IMPORTANT PRECAUTIONS
LANGUAGE

(ZH-CN)

TENTO PROVOZN NVOD EXISTUJE POUZE V ANGLICKM JAZYCE.

V PPAD, E EXTERN SLUBA ZKAZNKM POTEBUJE NVOD V


JINM JAZYCE, JE ZAJITN PEKLADU DO ODPOVDAJCHO JAZYKA
KOLEM ZKAZNKA.

NESNATE SE O DRBU TOHOTO ZAZEN, ANI BYSTE SI PEETLI


TENTO PROVOZN NVOD A POCHOPILI JEHO OBSAH.

V PPAD NEDODROVN TTO VSTRAHY ME DOJT K PORANN


PRACOVNKA PRODEJNHO SERVISU, OBSLUNHO PERSONLU NEBO
PACIENT VLIVEM ELEKTRICKHOP PROUDU, RESPEKTIVE VLIVEM
MECHANICKCH I JINCH RIZIK.

ADVARSEL

DENNE SERVICEMANUAL FINDES KUN P ENGELSK.

(DA)

HVIS EN KUNDES TEKNIKER HAR BRUG FOR ET ANDET SPROG END


ENGELSK, ER DET KUNDENS ANSVAR AT SRGE FOR OVERSTTELSE.

FORSG IKKE AT SERVICERE UDSTYRET MEDMINDRE


DENNE SERVICEMANUAL HAR VRET KONSULTERET OG ER FORSTET.

MANGLENDE OVERHOLDELSE AF DENNE ADVARSEL KAN MEDFRE


SKADE P GRUND AF ELEKTRISK, MEKANISK ELLER ANDEN FARE FOR
TEKNIKEREN, OPERATREN ELLER PATIENTEN.

VSTRAHA
(CS)

WAARSCHUWING
(NL)

DEZE ONDERHOUDSHANDLEIDING IS ENKEL IN HET ENGELS


VERKRIJGBAAR.
ALS HET ONDERHOUDSPERSONEEL EEN ANDERE TAAL VEREIST, DAN IS
DE KLANT VERANTWOORDELIJK VOOR DE VERTALING ERVAN.
PROBEER DE APPARATUUR NIET TE ONDERHOUDEN VOORDAT DEZE
ONDERHOUDSHANDLEIDING WERD GERAADPLEEGD EN BEGREPEN IS.
INDIEN DEZE WAARSCHUWING NIET WORDT OPGEVOLGD, ZOU HET
ONDERHOUDSPERSONEEL, DE OPERATOR OF EEN PATINT GEWOND
KUNNEN RAKEN ALS GEVOLG VAN EEN ELEKTRISCHE SCHOK,
MECHANISCHE OF ANDERE GEVAREN.

IMPORTANT PRECAUTIONS

Page 5

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2

WARNING
(EN)

HOIATUS
(ET)

CT/E DUAL, PROSPEED AII/FII SERIES


CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.


IF A CUSTOMERS SERVICE PROVIDER REQUIRES A LANGUAGE
OTHER THAN ENGLISH, IT IS THE CUSTOMERS RESPONSIBILITY TO
PROVIDE TRANSLATION SERVICES.
DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS
SERVICE MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE
SERVICE PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC
SHOCK, MECHANICAL OR OTHER HAZARDS.

KESOLEV TEENINDUSJUHEND ON SAADAVAL AINULT INGLISE KEELES.


KUI KLIENDITEENINDUSE OSUTAJA NUAB JUHENDIT INGLISE KEELEST
ERINEVAS KEELES, VASTUTAB KLIENT TLKETEENUSE OSUTAMISE EEST.
RGE RITAGE SEADMEID TEENINDADA ENNE EELNEVALT
KESOLEVA TEENINDUSJUHENDIGA TUTVUMIST JA SELLEST ARU
SAAMIST.
KESOLEVA HOIATUSE EIRAMINE VIB PHJUSTADA TEENUSEOSUTAJA,
OPERAATORI VI PATSIENDI VIGASTAMIST ELEKTRILGI, MEHAANILISE
VI MUU OHU TAGAJRJEL.

VAROITUS
(FI)

TM HUOLTO-OHJE ON SAATAVILLA VAIN ENGLANNIKSI.


JOS ASIAKKAAN HUOLTOHENKILST VAATII MUUTA KUIN
ENGLANNINKIELIST MATERIAALIA, TARVITTAVAN KNNKSEN
HANKKIMINEN ON ASIAKKAAN VASTUULLA.
L YRIT KORJATA LAITTEISTOA ENNEN KUIN OLET VARMASTI LUKENUT
JA YMMRTNYT TMN HUOLTO-OHJEEN.
MIKLI TT VAROITUSTA EI NOUDATETA, SEURAUKSENA VOI OLLA
HUOLTOHENKILSTN, LAITTEISTON KYTTJN TAI POTILAAN
VAHINGOITTUMINEN SHKISKUN, MEKAANISEN VIAN TAI MUUN
VAARATILANTEEN VUOKSI.

ATTENTION
(FR)

Page 6

CE MANUEL DE MAINTENANCE NEST DISPONIBLE QUEN ANGLAIS.


SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE
AUTRE LANGUE QUE LANGLAIS, CEST AU CLIENT QUIL INCOMBE
DE LE FAIRE TRADUIRE.
NE PAS TENTER DINTERVENTION SUR LES QUIPEMENTS TANT
QUE LE MANUEL SERVICE NA PAS T CONSULT ET COMPRIS.
LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRANER CHEZ
LE TECHNICIEN, LOPRATEUR OU LE PATIENT DES BLESSURES
DUES DES DANGERS LECTRIQUES, MCANIQUES OU AUTRES.

IMPORTANT PRECAUTIONS

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2

WARNUNG
(DE)

CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

DIESE SERVICEANLEITUNG EXISTIERT NUR IN ENGLISCHER SPRACHE.


FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENTIGT,
IST ES AUFGABE DES KUNDEN FR EINE ENTSPRECHENDE
BERSETZUNG ZU SORGEN.
VERSUCHEN SIE NICHT DIESE ANLAGE ZU WARTEN,
OHNE DIESE SERVICEANLEITUNG GELESEN UND VERSTANDEN ZU HABEN.
WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU
VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER
DES PATIENTEN DURCH STROMSCHLGE, MECHANISCHE ODER SONSTIGE
GEFAHREN KOMMEN.

(EL)

.

,
.


.
,
,
,
.

FIGYELMEZTETS
(HU)

EZEN KARBANTARTSI KZIKNYV KIZRLAG ANGOL NYELVEN RHET


EL.
HA A VEV SZOLGLTATJA ANGOLTL ELTR NYELVRE TART IGNYT,
AKKOR A VEV FELELSSGE A FORDTS ELKSZTTETSE.
NE PRBLJA ELKEZDENI HASZNLNI A BERENDEZST, AMG A
KARBANTARTSI KZIKNYVBEN LERTAKAT NEM RTELMEZTK.
EZEN FIGYELMEZTETS FIGYELMEN KVL HAGYSA A SZOLGLTAT,
MKDTET VAGY A BETEG RAMTS, MECHANIKAI VAGY EGYB
VESZLYHELYZET MIATTI SRLST EREDMNYEZHETI.

AVRUN
(IS)

ESSI JNUSTUHANDBK ER EINGNGU FANLEG ENSKU.


EF A JNUSTUVEITANDI VISKIPTAMANNS ARFNAST ANNAS
TUNGUMLS EN ENSKU, ER A SKYLDA VISKIPTAMANNS A SKAFFA
TUNGUMLAJNUSTU.
REYNI EKKI A AFGREIA TKI NEMA A ESSI JNUSTUHANDBK
HEFUR VERI SKOU OG SKILIN.
BROT SINNA ESSARI AVRUN GETUR LEITT TIL MEISLA
JNUSTUVEITANDA, STJRNANDA EA SJKLINGS FR RAFLOSTI,
VLRNU EA RUM HTTUM.

IMPORTANT PRECAUTIONS

Page 7

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2

AVVERTENZA

(IT)

CT/E DUAL, PROSPEED AII/FII SERIES


CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

IL PRESENTE MANUALE DI MANUTENZIONE DISPONIBILE


SOLTANTO IN INGLESE.
SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEMS
RICHIEDE IL MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE
TENUTO A PROVVEDERE DIRETTAMENTE ALLA TRADUZIONE.
SI PROCEDA ALLA MANUTENZIONE DELLAPPARECCHIATURA
SOLO DOPO AVER CONSULTATO IL PRESENTE MANUALE ED
AVERNE COMPRESO IL CONTENUTO.
IL NON RISPETTO DELLA PRESENTE AVVERTENZA POTREBBE FAR
COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALLADDETTO
ALLA MANUTENZIONE, ALLUTILIZZATORE ED AL PAZIENTE PER
FOLGORAZIONE ELETTRICA, PER URTI MECCANICI OD ALTRI
RISCHI.



(JA)

(KO)

,
.

,
, .

Page 8

IMPORTANT PRECAUTIONS

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2

BRDINJUMS
(LV)

CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

APKALPES ROKASGRMATA IR PIEEJAMA TIKAI ANGU VALOD.


JA KLIENTA APKALPES SNIEDZJAM NEPIECIEAMA INFORMCIJA CIT
VALOD, NEVIS ANGU, KLIENTA PIENKUMS IR NODROINT
TULKOANU.
NEVEICIET APRKOJUMA APKALPI BEZ APKALPES ROKASGRMATAS
IZLASANAS UN SAPRAANAS.
BRDINJUMA NEIEVROANA VAR RADT ELEKTRISKS STRVAS
TRIECIENA, MEHNISKU VAI CITU RISKU IZRAISTU TRAUMU APKALPES
SNIEDZJAM, OPERATORAM VAI PACIENTAM.

SPJIMAS
(LT)

IS EKSPLOATAVIMO VADOVAS YRA PRIEINAMAS TIK ANGL KALBA.


JEI KLIENTO PASLAUG TIEKJAS REIKALAUJA VADOVO KITA KALBA NE
ANGL, NUMATYTI VERTIMO PASLAUGAS YRA KLIENTO ATSAKOMYB.
NEMGINKITE ATLIKTI RANGOS TECHNINS PRIEIROS, NEBENT
ATSIVELGTE EKSPLOATAVIMO VADOV IR J SUPRATOTE.
JEI NEATKREIPSITE DMESIO PERSPJIM, GALIMI SUALOJIMAI DL
ELEKTROS OKO.
MECHANINI AR KIT PAVOJ PASLAUG TIEKJUI, OPERATORIUI AR
PACIENTUI.

ADVARSEL
(NO)

DENNE SERVICEHNDBOKEN FINNES BARE P ENGELSK.


HVIS KUNDENS SERVICELEVERANDR TRENGER ET ANNET SPRK, ER
DET KUNDENS ANSVAR SRGE FOR OVERSETTELSE.
IKKE FORSK REPARERE UTSTYRET UTEN AT DENNE
SERVICEHNDBOKEN ER LEST OG FORSTTT.
MANGLENDE HENSYN TIL DENNE ADVARSELEN KAN FRE TIL AT
SERVICELEVERANDREN, OPERATREN ELLER PASIENTEN SKADES P
GRUNN AV ELEKTRISK STT, MEKANISKE ELLER ANDRE FARER.

OSTRZEENIE
(PL)

NINIEJSZY PODRCZNIK SERWISOWY DOSTPNY JEST JEDYNIE W JZYKU


ANGIELSKIM.
JELI DOSTAWCA USUG KLIENTA WYMAGA JZYKA INNEGO NI
ANGIELSKI, ZAPEWNIENIE USUGI TUMACZENIA JEST OBOWIZKIEM
KLIENTA.
NIE PRBOWA SERWISOWA WYPOSAENIA BEZ ZAPOZNANIA SI I
ZROZUMIENIA NINIEJSZEGO PODRCZNIKA SERWISOWEGO.
NIEZASTOSOWANIE SI DO TEGO OSTRZEENIA MOE SPOWODOWA
URAZY DOSTAWCY USUG, OPERATORA LUB PACJENTA W WYNIKU
PORAENIA ELEKTRYCZNEGO, ZAGROENIA MECHANICZNEGO BD
INNEGO.

IMPORTANT PRECAUTIONS

Page 9

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2

CT/E DUAL, PROSPEED AII/FII SERIES


CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

ATENO

ESTE MANUAL DE ASSISTNCIA TCNICA S SE ENCONTRA DISPONVEL


EM INGLS.

(PT)

SE QUALQUER OUTRO SERVIO DE ASSISTNCIA TCNICA, QUE NO A


GEMS, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, DA
RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIOS DE
TRADUO.
NO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E
COMPREENDIDO ESTE MANUAL DE ASSISTNCIA TCNICA
O NO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANA
DO TCNICO, OPERADOR OU PACIENTE DEVIDO A CHOQUES ELTRICOS,
MECNICOS OU OUTROS.

ATENIE
(RO)

ACEST MANUAL DE SERVICE ESTE DISPONIBIL NUMAI N LIMBA ENGLEZ.


DAC UN FURNIZOR DE SERVICII PENTRU CLIENI NECESIT O ALT
LIMB DECT CEA ENGLEZ, ESTE DE DATORIA CLIENTULUI S
FURNIZEZE O TRADUCERE.
NU NCERCAI S REPARAI ECHIPAMENTUL DECT ULTERIOR
CONSULTRII I NELEGERII ACESTUI MANUAL DE SERVICE.
IGNORAREA ACESTUI AVERTISMENT AR PUTEA DUCE LA RNIREA
DEPANATORULUI, OPERATORULUI SAU PACIENTULUI N URMA
PERICOLELOR DE ELECTROCUTARE, MECANICE SAU DE ALT NATUR.

!
(RU)


.

, - ,
.


.

, ,
,
.

UPOZORNENIE
(SK)

TENTO NVOD NA OBSLUHU JE K DISPOZCII LEN V ANGLITINE.


AK ZKAZNKOV POSKYTOVATE SLUIEB VYADUJE IN JAZYK AKO
ANGLITINU, POSKYTNUTIE PREKLADATESKCH SLUIEB JE
ZODPOVEDNOSOU ZKAZNKA.
NEPOKAJTE SA O OBSLUHU ZARIADENIA SKR, AKO SI NEPRETATE
NVOD NA OBLUHU A NEPOROZUMIETE MU.
ZANEDBANIE TOHTO UPOZORNENIA ME VYSTI DO ZRANENIA
POSKYTOVATEA SLUIEB, OBSLUHUJCEJ OSOBY ALEBO PACIENTA
ELEKTRICKM PRDOM, DO MECHANICKHO ALEBO INHO
NEBEZPEENSTVA.

Page 10

IMPORTANT PRECAUTIONS

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2

ATENCION
(ES)

CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

ESTE MANUAL DE SERVICIO SOLO EXISTE EN INGLES.


SI ALGUN PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA
QUE NO SEA EL INGLES, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN
SERVICIO DE TRADUCCION.
NO SE DEBERA DAR SERVICIO TECNICO AL EQUIPO,
SIN HABER CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL
PROVEEDOR DE SERVICIOS, EL OPERADOR O EL PACIENTE SUFRAN
LESIONES PROVOCADAS POR CAUSAS ELCTRICAS, MECNICAS O DE
OTRA NATURALEZA.

VARNING
(SV)

DEN HR SERVICEHANDBOKEN FINNS BARA TILLGNGLIG P ENGELSKA.


OM EN KUNDS SERVICETEKNIKER HAR BEHOV AV ETT ANNAT SPRK N
ENGELSKA ANSVARAR KUNDEN FR ATT TILLHANDAHLLA
VERSTTNINGSTJNSTER.
FRSK INTE UTFRA SERVICE P UTRUSTNINGEN OM DU INTE HAR LST
OCH FRSTR DEN HR SERVICEHANDBOKEN.
OM DU INTE TAR HNSYN TILL DEN HR VARNINGEN KAN DET RESULTERA
I SKADOR P SERVICETEKNIKERN, OPERATREN ELLER PATIENTEN TILL
FLJD AV ELEKTRISKA STTAR, MEKANISKA FAROR ELLER ANDRA
FAROR.

DKKAT
(TR)

BU SERVS KILAVUZUNUN SADECE NGLZCES MEVCUTTUR.


EER MTER TEKNSYEN BU KILAVUZU NGLZCE DIINDA BR BAKA
LSANDAN TALEP EDERSE, BUNU TERCME ETTRMEK MTERYE DER.
SERVS KILAVUZUNU OKUYUP ANLAMADAN EKPMANLARA MDAHALE
ETMEYNZ.
BU UYARIYA UYULMAMASI, ELEKTRK, MEKANK VEYA DER
TEHLKELERDEN DOLAYI TEKNSYEN, OPERATR VEYA HASTANIN
YARALANMASINA YOL AABLR.

IMPORTANT PRECAUTIONS

Page 11

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2

CT/E DUAL, PROSPEED AII/FII SERIES


CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent write Damage
In Shipment on ALL copies of the freight or express bill BEFORE delivery is accepted or signed
for by a GE representative or hospital receiving agent. Whether noted or concealed, damage
MUST be reported to the carrier immediately upon discovery, or in any event, within 14 days after
receipt, and the contents and containers held for inspection by the carrier. A transportation
company will not pay a claim for damage if an inspection is not requested within this 14 day period.
To file a report:
Fill out a report on http://egems.med.ge.com/edq/home.jsp
Contact the local service coordinator.

CERTIFIED ELECTRICAL CONTRACTOR STATEMENT


All electrical Installations that are preliminary to positioning of the equipment at the site prepared
for the equipment shall be performed by licensed electrical contractors. In addition, electrical feeds
into the Power Distribution Unit shall be performed by licensed electrical contractors. Other
connections between pieces of electrical equipment, calibrations and testing shall be performed by
qualified GE Medical personnel. The products involved (and the accompanying electrical
installations) are highly sophisticated, and special engineering competence is required. In
performing all electrical work on these products, GE will use its own specially trained field
engineers. All of GEs electrical work on these products will comply with the requirements of the
applicable electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GEs field engineers,
personnel of third-party service companies with equivalent training, or licensed electricians) to
perform electrical servicing on the equipment.

IMPORTANT...X-RAY PROTECTION
X-ray equipment if not properly used may cause injury. Accordingly, the instructions herein
contained should be thoroughly read and understood by everyone who will use the equipment
before you attempt to place this equipment in operation. The General Electric Company, Medical
Systems Group, will be glad to assist and cooperate in placing this equipment in use.
Although this apparatus incorporates a high degree of protection against x-radiation other than the
useful beam, no practical design of equipment can provide complete protection. Nor can any
practical design compel the operator to take adequate precautions to prevent the possibility of any
persons carelessly exposing themselves or others to radiation.
It is important that anyone having anything to do with x-radiation be properly trained and fully
acquainted with the recommendations of the National Council on Radiation Protection and
Measurements as published in NCRP Reports available from NCRP Publications, 7910 Woodmont
Avenue, Room 1016, Bethesda, Maryland 20814, and of the International Commission on
Radiation Protection, and take adequate steps to protect against injury.
The equipment is sold with the understanding that the General Electric Company, Medical Systems
Group, its agents, and representatives have no responsibility for injury or damage which may result
from improper use of the equipment.
Various protective materials and devices are available. It is urged that such materials or devices be
used.

LITHIUM BATTERY CAUTIONARY STATEMENTS


CAUTION
Risk of
Explosion
Page 12

Danger of explosion if battery is incorrectly replaced. Replace only with the same or
equivalent type recommended by the manufacturer. Discard used batteries according to the
manufacturers instructions.
IMPORTANT PRECAUTIONS

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2

ATTENTION
Danger
dExplosion

CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

Il y a danger dexplosion sil y a replacement incorrect de la batterie. Remplacer uniquement


avec une batterie du mme type ou dun type recommand par le constructeur. Mettre au
rbut les batteries usages conformment aux instructions du fabricant.

OMISSIONS & ERRORS


Customers, please contact your GE Sales or Service representatives.
GE personnel, please use the GEMS CQA Process to report all omissions, errors, and defects in
this publication.

IMPORTANT PRECAUTIONS

Page 13

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2

Page 14

CT/E DUAL, PROSPEED AII/FII SERIES


CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

IMPORTANT PRECAUTIONS

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2

CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

Revision

Date

Reason for change

10/25/06

initial release.

06/07/06

Added Section 4 Disposal: warning for statement about Hg in device.

Revision History

Page 15

Revision History

Revision History

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2

Page 16

CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

Revision History

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2

CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

Table of Contents
Chapter 1
Introduction............................................................................................................ 19

Section 2.0
Material List ...................................................................................................... 19
Section 3.0
Flowchart .......................................................................................................... 20

Chapter 2
Installation Procedure........................................................................................... 21
Section 1.0
Overview ........................................................................................................... 21
Section 2.0
Installation Procedure...................................................................................... 21
2.1
2.2
2.3

EXISTING MONITOR REMOVAL (IF THERE IS A MONITOR TO BE REPLACED)......


Monitor Connection..........................................................................................................
General Procedures for Setup .........................................................................................
2.3.1 OSD (On-Screen-Display) Controls....................................................................
2.3.2 Main Menu ..........................................................................................................
2.3.3 OSD Function .....................................................................................................
2.3.4 Lock Mode Explain .............................................................................................
2.3.5 Hot Key Controls.................................................................................................
2.3.6 Contrast and Brightness Minor Adjustment ........................................................

21
21
26
26
27
27
29
30
31

Section 3.0
Function Check ................................................................................................ 32
Section 4.0
Disposal ............................................................................................................ 32

Chapter 3
Finish Installation .................................................................................................. 33
Section 1.0
Attach Rating Plate .......................................................................................... 33
Section 2.0
Return Installation Card................................................................................... 34

Table of Contents

Page 17

Table of Contents

Section 1.0
System Requirement........................................................................................ 19

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2

Page 18

CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

Table of Contents

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 1

CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

Chapter 1
Introduction
Section 1.0
System Requirement
This option is available for all Systems.

1 - Introduction

Section 2.0
Material List
Verify that the contents of this option kit are correct according to the Material List below.

Chilin 19" Color LCD Monitor Option Kit (5128434-2)


ITEM

PART NO.

DESCRIPTION

QTY.

5178602

Chilin 19" Color LCD Monitor

2111577

Rating Plate

5178234-100

Option Installation Manual (This Manual)

Chilin 19" Color LCD Monitor Option Kit_Black (5128434-3)


ITEM

PART NO.

DESCRIPTION

QTY.

5178602-2

Chilin 19" Color LCD Monitor - Black

2111577

Rating Plate

5178234-100

Option Installation Manual (This Manual)

Chapter 1 - Introduction

Page 19

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 1

CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

Section 3.0
Flowchart
The Illustration below shows the main flow of the installation procedures of the option.

START

Preparation
Monitor Installation
Functional Check

END

Page 20

Section 3.0 - Flowchart

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2

CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

Chapter 2
Installation Procedure
Section 1.0
Overview
This instruction describes the hardware installation procedure.

Section 2.0
Installation Procedure
2 INSTALLATION
PROCEDURE

2.1 EXISTING MONITOR REMOVAL (IF THERE IS A MONITOR TO BE


REPLACED)
Understand and follow the Safety Guidelines Manual.
1.) Turn OFF the Operator Console main power.
2.) Turn OFF the breaker in the Power Distribution Box (hospital supply).
3.) Remove the OC Rear Cover.
4.) Disconnect the Signal and Power cables connected from the monitor to the OC.
5.) Remove the monitor from the OC (or OC table).

2.2

Monitor Connection
1.) Unpack the LCD monitor and place it on the OC.
2.) Route and connect the cables as below illustrations.
Note:

For each LCD Monitor we supply two different kinds of Signal Cables. Check the ports of the
Display Board before connecting the Signal Cables. Select proper Signal Cable to connect
the proper port on the LCD according to different ports of the Display Board.
Figure 2-1 Signal Cable
DVI VGA

DVI

DVI

VGA
VGA

Chapter 2 - Installation Procedure

Page 21

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2

CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

For CT/e, CT ProSpeed AI/FI Series and CT/e Dual, ProSpeed AII/FII Series

For HiSpeed Dual (Big OC)


For ProSpeed FII (Small OC)

For BrightSpeed Series and RT16, Xtra; LS RT

For CT/e, CT ProSpeed AI/FI Series and CT/e Dual, ProSpeed AII/FII Series
a.) Connect the Power cable to the AC Power Supply Assy (or AC Box).
b.) Reinstall the Rear Cover.
c.) Connect the signal cable to the VSPL (Video Board).
d.) Reinstall the Rear Connector Cover.

Figure 2-2 AC Power Cable and Signal Cable Replacement


1

AC PS Assy
(rear view)

Connect the Monitor


Power cable to the AC
PS Assy

Video Board

Monitor
Slave Monitor

Rear
Connector
Cover

Rear
Reinstall the Cover
4
Rear Connector
Reinstall the
2
Rear Cover
Cover

AC PS Assy

Page 22

Chapter 2 - Installation Procedure

Connect the Monitor


signal cable connector
(VGA
VGA)

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2

CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

For HiSpeed Dual (Big OC)

Figure 2-3 Cables Connection

Rear View

OC Rear View

DVI VGA

Power Cable

REAR CN2
DASIFN2
VSPL

VGA
Monitor

VGA
Slave Monitor

Signal Cable
(VGA
VGA)

REAR CN1

2 INSTALLATION
PROCEDURE

HOST ASSY

Route the Power Cable through the hole


to connect the AC Box

Figure 2-4 Cables Connection (No VSPL)


Rear View

OC Rear View

DVI VGA

Power Cable

Signal Cable
(VGA
DVI)

REAR CN2
DASIFN2
REAR CN1

DVI

HOST ASSY

Route the Power Cable through the hole


to connect the AC Box

Chapter 2 - Installation Procedure

Page 23

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2

CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

For ProSpeed FII (Small OC)

Figure 2-5 Cables Connection


Rear View

OC Rear View

DVI VGA

Power Cable
Signal Cable
(VGA
VGA)

VGA

REAR CN1

VSPL

DASIFN2

REAR CN2

VGA

Figure 2-6 Cables Connection (No VSPL)


Rear View

OC Rear View

DVI VGA

Power Cable

Page 24

REAR CN1

DASIFN2

REAR CN2

Signal Cable
(VGA
DVI)

Chapter 2 - Installation Procedure

Host PC
DVI

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2

CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

For BrightSpeed Series and RT16, Xtra; LS RT


OC Rear View

Rear View

Power Cable

DVI VGA

Rear View

DVI VGA

Signal Cable
(DVI
DVI or
VGA
DVI)
Route the Power Cables thourgh
the hole to connect the Power Supply
HOST ASSY
DVI

2 INSTALLATION
PROCEDURE

DVI

Display
Signal Cable
Connectors

Scan

3.) Connect the signal cable and power cable to the LCD monitor.
Figure 2-7 Connect Cables to LCD Monitor
Note: Fix the Power Cable with metal clip

Power Cable

Signal Cable

4.) Reinstall the OC Rear Cover.


5.) Turn ON the main switch located at the back of the LCD monitor.
Chapter 2 - Installation Procedure

Page 25

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2

2.3
2.3.1

CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

General Procedures for Setup


OSD (On-Screen-Display) Controls

Main Function

Menu

[Power]

Power Indicator
Green: Normal
Orange: Power Saving, No Signal
Off: Power Off

[Enter]

Access and control the function of sub-menu


Quick input source selection. *

[>]

Press the ">" to selection sub-menu items and increase adjustment


Quick gamma item selection. *

[<]

Press the "<" to selection sub-menu items and decrease adjustment


Quick mode item selection. *

[Menu]

Start the OSD feature


Exits the OSD controls
Exits the OSD sub-menu

* Refer to the Hot Key Controls Section.


OSD control buttons on the front of the monitor function as follows:
1.) Power on the LCD Medical Display.
2.) Press the "MENU" button to start the OSD feature.
3.) Press the ">" or "<" button to selection sub-menu items.
4.) Press the ">" or "<" button to increase/decrease adjustment.
5.) Press the "ENTER" button to access the function to be adjusted.
6.) To exit the OSD menu or go back to the previous action by clicking the "MENU" button. It will
save the change automatically.
7.) To repeat above steps for changing the setting of other functions.

Page 26

Chapter 2 - Installation Procedure

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2

2.3.2

CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

Main Menu

Color Menu
Image Menu
Display Menu

Options Menu
OSD Menu

Display Menu

Utility Menu

Brightness

Contrast

Hue
Saturation
Operation
Explaining

2.3.3

2 INSTALLATION
PROCEDURE

OSD Control
Button State

OSD Function

Primary
Directory

Display

Secondary Directory

Brightness

Adjusts the background black level of the screen


image.

Contrast

Adjusts fore ground white level of the screen


image.

Hue

Adjusts the purity and depth of the red, blue and


green of color.

Saturation
Auto Adjust
Auto Color
Phase
Image

Description

Clock
H-Position
V-Position

Adjusts the saturation of the color.


Automatically adjusts the Image Position and H.
size (or V. size) setting and Fine Settings.
Moves the image up and down on the screen.
Adjusts image distortion appearing as horizontal
"noise" on the screen.
Adjusts the clock pulse of the image.
Moves the screen image left or right.
Moves the screen image up or down.

Chapter 2 - Installation Procedure

Page 27

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2

Primary
Directory

CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

Secondary Directory

Gamma

Color

Color Temp.

Image Size
Zoom Mode
Horizontal Pan
Option

Vertical Pan
Auto Mode
LED Mode
H-Position
V-Position
Rotation

OSD

Time Out
Language
All Reset
Key Lock

Utility

Information
Overlapped

Page 28

Description
Adjusts Gamma value.

Selects color temperature.

Adjust image size.


Change the screen size according to the image
aspect ratio.
Fill the Image left and right on the screen.
Fill the Image up and down on the screen.
On/Off auto adjust.
On/Off LED function.
Moves the OSD left and right on the screen.
Moves the OSD up and down on the screen.
Rotations the OSD.
Adjusts OSD display time setting.
Choose language of the display on the screen.
Resets the highlighted control menu to the factory
setting.
Lock OSD function.
Shows information of Firmware, Source and
Resolution.
Setting Value when the image to occur overlapped.

Chapter 2 - Installation Procedure

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2

Lock Mode Explain


MDC1900-1LG system default is Off mode.
LOCK

Off: Function keys are non-lock, press "Menu" key to start the OSD menu and you can
adjustment any functions.

Mode 1: All function keys are locked.

Mode 2: All function keys are locked unless brightness control. Press "Menu" key to start the
brightness control menu.
Operation step as follows:
1.) Press "Menu" key to start OSD freature.
2.) Select Setup Menu and press "Enter".

SetUp
M
All
Lock

Informa
ti
2 INSTALLATION
PROCEDURE

2.3.4

CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

Overlapp

3.) Press "<" or ">" to select Lock Mode and press "Enter".

Setup
M
All
Lock
Overlapp

Informa
ti

4.) Press "<" or ">" to select Lock Mode and press "Enter" will save the setting value
automatically.

Chapter 2 - Installation Procedure

Page 29

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2

CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

5.) When you press any buttons on the front of the display that will shows message "OSD
locked" after setting Lock Mode complete and exit OSD feature.

Unlock
Press left key "<" until display shows the message "Key unlocked", lock function will be remove.
Note:

2.3.5

When you in Mode 1 or Mode 2 and unlock after 15 minutes not adjustment any function, OSD
function will be lock automatically. Only setting in "Off" mode doesn't have this appearance.

Hot Key Controls


It can choose the best operation mode to offer to users. This product has offered the following
adjustment functions. The setting mode is not in OSD controls and single-click the hot key, can
start hot key function, as follows:
1.) Input Source Selection
Press "ENTER" button to select the input source.

DVI-A

VGA

DVI-D

VGA:Acquire analog signal from the D-sub connector.


DVI-A:Acquire analog signal from the DVI-I connector.
DVI-D:Acquire digital signal from the DVI-I connector.
2.) Mode item Selection
Press "<" (left) button, you can selecet color on the front of the screen of display.

User Mode

DICOM Mode

LB Mode

Page 30

CL Mode

BL Mode

User Mode: User identify the standard display operation.


DICOM Mode: System set the DICOM standard operation.
CL Mode: To imitate the display output color temperature of about 6500k.
BL Mode: To imitate the display output color temperature of about 9300k.
LB Mode: To imitate the operation for whitelight lampbox.

Chapter 2 - Installation Procedure

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2

CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

3.) Gamma item Selection

2.3.6

Native

Gamma 1.8

Gamma 2.0

User

DICOM

Gamma 2.2

Native: Default.

Gamma 1.8: The display state is 1.8 for gamma value.

Gamma 2.0: The display state is 2.0 for gamma value.


Gamma 2.2: The display state is 2.2 for gamma value.
DICOM: The display state is standard for DICOM.

User: The display state is by oneself set for gamma value.

Contrast and Brightness Minor Adjustment


1.) Select Service Image Quality Install SMPTE image (To install the SMPTE image on
the system.)
2.) Click on ImageWorks

3.) Select "SMPTE" image in the Examination list.


4.) Click on Viewer to display the SMPTE pattern.
5.) Enlarge the SMPTE image to full screen display, using Display Routine Display Zoom.
If necessary, drag the Zoom slider to display the pattern to the full screen.

Chapter 2 - Installation Procedure

Page 31

2 INSTALLATION
PROCEDURE

Press" >" (right) button, select Gamma value.

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2

CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

6.) Adjust Contrast and Brightness to the level an user desires so that a slight (not clear)
distinction should be made between 0% and 5% and between 95% and 100%.

5%

0%

100%

95%

7.) Also check if the patterns are properly displayed.

Section 3.0
Function Check
1.) Perform Power On Self Test:
a.) Switch ON the main switch of the OC.
The test progress information is displayed on the monitor.
b.) Verify that the system starts up normally.
2.) Perform the normal test scannings and check that the images are normal.

Section 4.0
Disposal
WARNING

Page 32

This product consists of devices that may contain mercury, which must be recycled
or disposed of in accordance with local, state, or country laws. (Within this system,
the backlight lamps in the monitor display contain mercury.)

Chapter 2 - Installation Procedure

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 1

CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

Chapter 3
Finish Installation
Section 1.0
Attach Rating Plate
1.) Attach the provided rating plate to the following location of OC.
Figure 3-1 Rating Plate for Big OC

Label area

Figure 3-2 Rating Plate for Small OC


Rear View

3 FINISH INSTALLATION

OC Rear Side

Rating plate

Option Rating Plate

Rating Plate

OC Rating Plate

Chapter 3 - Finish Installation

Page 33

GE HEALTHCARE
DIRECTION 5178234-100, REVISION 1

CHILIN 19" COLOR LCD MONITOR OPTION INSTALLATION

Section 2.0
Return Installation Card
1.) Complete and verify the site-specific information on the Red Product Locator Installation Card.
2.) Send the Product Locator Card to the appropriate address provided belowl:
- Address for GEMS-AM sites:
Send to the address printed on the Product Locator Card.
- Address for GEMS-E sites:
Send to the address Printed on the Product Locator Card.
- Address for GEMS-A sites:
Send to the local PLS or Service Administrator.

Page 34

Chapter 3 - Finish Installation

CT

3000 N. Grandview Blvd.,


Waukesha, WI 53188 U.S.A.

www.gehealthcare.com