GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2
Page 2
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2
LEGAL NOTES
TRADEMARKS
Adobe, the Adobe logo, Acrobat, the Acrobat logo, Exchange, and PostScript are trademarks of
Adobe Systems Incorporated or its subsidiaries and may be registered in certain jurisdictions.
Microsoft is a registered trademark and Windows is a trademark of Microsoft Corporation.
All other products and their name brands are trademarks of their respective holders.
COPYRIGHTS
All Material, Copyright 2006-2007 by General Electric Company, Inc. All rights reserved.
Legal Notes
Page 3
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2
Page 4
Legal Notes
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2
IMPORTANT PRECAUTIONS
LANGUAGE
(ZH-CN)
ADVARSEL
(DA)
VSTRAHA
(CS)
WAARSCHUWING
(NL)
IMPORTANT PRECAUTIONS
Page 5
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2
WARNING
(EN)
HOIATUS
(ET)
VAROITUS
(FI)
ATTENTION
(FR)
Page 6
IMPORTANT PRECAUTIONS
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2
WARNUNG
(DE)
(EL)
.
,
.
.
,
,
,
.
FIGYELMEZTETS
(HU)
AVRUN
(IS)
IMPORTANT PRECAUTIONS
Page 7
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2
AVVERTENZA
(IT)
(JA)
(KO)
,
.
,
, .
Page 8
IMPORTANT PRECAUTIONS
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2
BRDINJUMS
(LV)
SPJIMAS
(LT)
ADVARSEL
(NO)
OSTRZEENIE
(PL)
IMPORTANT PRECAUTIONS
Page 9
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2
ATENO
(PT)
ATENIE
(RO)
!
(RU)
.
, - ,
.
.
, ,
,
.
UPOZORNENIE
(SK)
Page 10
IMPORTANT PRECAUTIONS
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2
ATENCION
(ES)
VARNING
(SV)
DKKAT
(TR)
IMPORTANT PRECAUTIONS
Page 11
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent write Damage
In Shipment on ALL copies of the freight or express bill BEFORE delivery is accepted or signed
for by a GE representative or hospital receiving agent. Whether noted or concealed, damage
MUST be reported to the carrier immediately upon discovery, or in any event, within 14 days after
receipt, and the contents and containers held for inspection by the carrier. A transportation
company will not pay a claim for damage if an inspection is not requested within this 14 day period.
To file a report:
Fill out a report on http://egems.med.ge.com/edq/home.jsp
Contact the local service coordinator.
IMPORTANT...X-RAY PROTECTION
X-ray equipment if not properly used may cause injury. Accordingly, the instructions herein
contained should be thoroughly read and understood by everyone who will use the equipment
before you attempt to place this equipment in operation. The General Electric Company, Medical
Systems Group, will be glad to assist and cooperate in placing this equipment in use.
Although this apparatus incorporates a high degree of protection against x-radiation other than the
useful beam, no practical design of equipment can provide complete protection. Nor can any
practical design compel the operator to take adequate precautions to prevent the possibility of any
persons carelessly exposing themselves or others to radiation.
It is important that anyone having anything to do with x-radiation be properly trained and fully
acquainted with the recommendations of the National Council on Radiation Protection and
Measurements as published in NCRP Reports available from NCRP Publications, 7910 Woodmont
Avenue, Room 1016, Bethesda, Maryland 20814, and of the International Commission on
Radiation Protection, and take adequate steps to protect against injury.
The equipment is sold with the understanding that the General Electric Company, Medical Systems
Group, its agents, and representatives have no responsibility for injury or damage which may result
from improper use of the equipment.
Various protective materials and devices are available. It is urged that such materials or devices be
used.
Danger of explosion if battery is incorrectly replaced. Replace only with the same or
equivalent type recommended by the manufacturer. Discard used batteries according to the
manufacturers instructions.
IMPORTANT PRECAUTIONS
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2
ATTENTION
Danger
dExplosion
IMPORTANT PRECAUTIONS
Page 13
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2
Page 14
IMPORTANT PRECAUTIONS
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2
Revision
Date
10/25/06
initial release.
06/07/06
Revision History
Page 15
Revision History
Revision History
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2
Page 16
Revision History
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2
Table of Contents
Chapter 1
Introduction............................................................................................................ 19
Section 2.0
Material List ...................................................................................................... 19
Section 3.0
Flowchart .......................................................................................................... 20
Chapter 2
Installation Procedure........................................................................................... 21
Section 1.0
Overview ........................................................................................................... 21
Section 2.0
Installation Procedure...................................................................................... 21
2.1
2.2
2.3
21
21
26
26
27
27
29
30
31
Section 3.0
Function Check ................................................................................................ 32
Section 4.0
Disposal ............................................................................................................ 32
Chapter 3
Finish Installation .................................................................................................. 33
Section 1.0
Attach Rating Plate .......................................................................................... 33
Section 2.0
Return Installation Card................................................................................... 34
Table of Contents
Page 17
Table of Contents
Section 1.0
System Requirement........................................................................................ 19
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2
Page 18
Table of Contents
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 1
Chapter 1
Introduction
Section 1.0
System Requirement
This option is available for all Systems.
1 - Introduction
Section 2.0
Material List
Verify that the contents of this option kit are correct according to the Material List below.
PART NO.
DESCRIPTION
QTY.
5178602
2111577
Rating Plate
5178234-100
PART NO.
DESCRIPTION
QTY.
5178602-2
2111577
Rating Plate
5178234-100
Chapter 1 - Introduction
Page 19
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 1
Section 3.0
Flowchart
The Illustration below shows the main flow of the installation procedures of the option.
START
Preparation
Monitor Installation
Functional Check
END
Page 20
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2
Chapter 2
Installation Procedure
Section 1.0
Overview
This instruction describes the hardware installation procedure.
Section 2.0
Installation Procedure
2 INSTALLATION
PROCEDURE
2.2
Monitor Connection
1.) Unpack the LCD monitor and place it on the OC.
2.) Route and connect the cables as below illustrations.
Note:
For each LCD Monitor we supply two different kinds of Signal Cables. Check the ports of the
Display Board before connecting the Signal Cables. Select proper Signal Cable to connect
the proper port on the LCD according to different ports of the Display Board.
Figure 2-1 Signal Cable
DVI VGA
DVI
DVI
VGA
VGA
Page 21
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2
For CT/e, CT ProSpeed AI/FI Series and CT/e Dual, ProSpeed AII/FII Series
For CT/e, CT ProSpeed AI/FI Series and CT/e Dual, ProSpeed AII/FII Series
a.) Connect the Power cable to the AC Power Supply Assy (or AC Box).
b.) Reinstall the Rear Cover.
c.) Connect the signal cable to the VSPL (Video Board).
d.) Reinstall the Rear Connector Cover.
AC PS Assy
(rear view)
Video Board
Monitor
Slave Monitor
Rear
Connector
Cover
Rear
Reinstall the Cover
4
Rear Connector
Reinstall the
2
Rear Cover
Cover
AC PS Assy
Page 22
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2
Rear View
OC Rear View
DVI VGA
Power Cable
REAR CN2
DASIFN2
VSPL
VGA
Monitor
VGA
Slave Monitor
Signal Cable
(VGA
VGA)
REAR CN1
2 INSTALLATION
PROCEDURE
HOST ASSY
OC Rear View
DVI VGA
Power Cable
Signal Cable
(VGA
DVI)
REAR CN2
DASIFN2
REAR CN1
DVI
HOST ASSY
Page 23
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2
OC Rear View
DVI VGA
Power Cable
Signal Cable
(VGA
VGA)
VGA
REAR CN1
VSPL
DASIFN2
REAR CN2
VGA
OC Rear View
DVI VGA
Power Cable
Page 24
REAR CN1
DASIFN2
REAR CN2
Signal Cable
(VGA
DVI)
Host PC
DVI
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2
Rear View
Power Cable
DVI VGA
Rear View
DVI VGA
Signal Cable
(DVI
DVI or
VGA
DVI)
Route the Power Cables thourgh
the hole to connect the Power Supply
HOST ASSY
DVI
2 INSTALLATION
PROCEDURE
DVI
Display
Signal Cable
Connectors
Scan
3.) Connect the signal cable and power cable to the LCD monitor.
Figure 2-7 Connect Cables to LCD Monitor
Note: Fix the Power Cable with metal clip
Power Cable
Signal Cable
Page 25
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2
2.3
2.3.1
Main Function
Menu
[Power]
Power Indicator
Green: Normal
Orange: Power Saving, No Signal
Off: Power Off
[Enter]
[>]
[<]
[Menu]
Page 26
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2
2.3.2
Main Menu
Color Menu
Image Menu
Display Menu
Options Menu
OSD Menu
Display Menu
Utility Menu
Brightness
Contrast
Hue
Saturation
Operation
Explaining
2.3.3
2 INSTALLATION
PROCEDURE
OSD Control
Button State
OSD Function
Primary
Directory
Display
Secondary Directory
Brightness
Contrast
Hue
Saturation
Auto Adjust
Auto Color
Phase
Image
Description
Clock
H-Position
V-Position
Page 27
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2
Primary
Directory
Secondary Directory
Gamma
Color
Color Temp.
Image Size
Zoom Mode
Horizontal Pan
Option
Vertical Pan
Auto Mode
LED Mode
H-Position
V-Position
Rotation
OSD
Time Out
Language
All Reset
Key Lock
Utility
Information
Overlapped
Page 28
Description
Adjusts Gamma value.
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2
Off: Function keys are non-lock, press "Menu" key to start the OSD menu and you can
adjustment any functions.
Mode 2: All function keys are locked unless brightness control. Press "Menu" key to start the
brightness control menu.
Operation step as follows:
1.) Press "Menu" key to start OSD freature.
2.) Select Setup Menu and press "Enter".
SetUp
M
All
Lock
Informa
ti
2 INSTALLATION
PROCEDURE
2.3.4
Overlapp
3.) Press "<" or ">" to select Lock Mode and press "Enter".
Setup
M
All
Lock
Overlapp
Informa
ti
4.) Press "<" or ">" to select Lock Mode and press "Enter" will save the setting value
automatically.
Page 29
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2
5.) When you press any buttons on the front of the display that will shows message "OSD
locked" after setting Lock Mode complete and exit OSD feature.
Unlock
Press left key "<" until display shows the message "Key unlocked", lock function will be remove.
Note:
2.3.5
When you in Mode 1 or Mode 2 and unlock after 15 minutes not adjustment any function, OSD
function will be lock automatically. Only setting in "Off" mode doesn't have this appearance.
DVI-A
VGA
DVI-D
User Mode
DICOM Mode
LB Mode
Page 30
CL Mode
BL Mode
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2
2.3.6
Native
Gamma 1.8
Gamma 2.0
User
DICOM
Gamma 2.2
Native: Default.
Page 31
2 INSTALLATION
PROCEDURE
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 2
6.) Adjust Contrast and Brightness to the level an user desires so that a slight (not clear)
distinction should be made between 0% and 5% and between 95% and 100%.
5%
0%
100%
95%
Section 3.0
Function Check
1.) Perform Power On Self Test:
a.) Switch ON the main switch of the OC.
The test progress information is displayed on the monitor.
b.) Verify that the system starts up normally.
2.) Perform the normal test scannings and check that the images are normal.
Section 4.0
Disposal
WARNING
Page 32
This product consists of devices that may contain mercury, which must be recycled
or disposed of in accordance with local, state, or country laws. (Within this system,
the backlight lamps in the monitor display contain mercury.)
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 1
Chapter 3
Finish Installation
Section 1.0
Attach Rating Plate
1.) Attach the provided rating plate to the following location of OC.
Figure 3-1 Rating Plate for Big OC
Label area
3 FINISH INSTALLATION
OC Rear Side
Rating plate
Rating Plate
OC Rating Plate
Page 33
GE HEALTHCARE
DIRECTION 5178234-100, REVISION 1
Section 2.0
Return Installation Card
1.) Complete and verify the site-specific information on the Red Product Locator Installation Card.
2.) Send the Product Locator Card to the appropriate address provided belowl:
- Address for GEMS-AM sites:
Send to the address printed on the Product Locator Card.
- Address for GEMS-E sites:
Send to the address Printed on the Product Locator Card.
- Address for GEMS-A sites:
Send to the local PLS or Service Administrator.