Maintenance of measurement cell

Page 1 (30)

Maintenance of measurement cell


Code 10000006053e

Revision A

Page 2 (30)

Contents
1. ABOUT THIS DOCUMENT............................................................................................................................3
2. OVERVIEW ....................................................................................................................................................4
2.1 Analyzer probe with AWC ..................................................................................................................5
2.2 Analyzer probe with SCA ...................................................................................................................6
3. RADIATION SAFETY.....................................................................................................................................8
4. MAINTENANCE MODE ...............................................................................................................................10
4.1 Entering the maintenance mode ......................................................................................................10
4.2 Exiting the maintenance mode.........................................................................................................13
5. MAINTAINING THE MEASUREMENT CELL..............................................................................................14
5.1 Changing the discrete measurement cell windows in analyzer with SCA .......................................14
5.2 Periodical maintenance ....................................................................................................................15
Changing the window film reel for AWC ......................................................................................15
Resetting the window film consumption counter..........................................................................18
Changing the protective window ..................................................................................................19
5.3 Maintenance in fault situations .........................................................................................................20
Alarms and events in AWC fault situations ..................................................................................20
Alarms and events in SCA fault situations ...................................................................................25
5.4 Special situations .............................................................................................................................27
READER COMMENTS

Maintenance of measurement cell


Code 10000006053e

Revision A

Page 3 (30)

1. ABOUT THIS DOCUMENT


This document presents detailed information about the maintenance of the
measurement cell. The document gives a general description of the structure of
the device, and discusses radiation safety issues. The main part of the
document gives detailed instructions on how to get to the maintenance mode
and how to proceed when maintaining the measurement cell. The document is
aimed at the instrumentation and maintenance staff of the plant.
Note, Tip, and Warning signs used in this manual:
This sign stands for information especially worth noticing.

This sign stands for information including extra advice.

This sign stands for information of extreme importance.

Maintenance of measurement cell


Code 10000006053e

Revision A

Page 4 (30)

2. OVERVIEW
There are two types of measurement cell packages available for the Courier
analyzer:
Analyzer probe with Automatic Window Changer (AWC)
Analyzer probe with Sample Cell Assembly (SCA)
These measurement cell types are described in more detail in the following
section.

Maintenance of measurement cell


Code 10000006053e

Revision A

Page 5 (30)

2.1 Analyzer probe with AWC


The AWC is a mechanism that performs an automatic window change. When
changing the window, the AWC automatically releases the measurement cell
and the measurement cell window film, winds the window film forward
approximately 10 cm, and tightens the measurement cell window film and the
measurement cell.
The window film lasts up to 700 window changes, and therefore, the window
film storage reel needs to be changed manually at approx. six month intervals. If
there is a malfunction in the measurement cell mechanisms, e.g., the motor is
down, you can also change the window film manually.
The analyzer probe with AWC is comprised of:
Measurement cell
Window film reels
Window film material
Capstan motor
Cylinders and their valves
Moisture leakage detection
1
2
3
4
5
6
7
8
9

Figure 1. The analyzer probe with AWC.


1 = Maintenance button, 2 = Upper window film reel,
3 = Measurement cell window, 4 = Protective window,
5 = Moisture leakage detections, 6 = Protective window ring,
7 = Cylinders and their valves, 8 = Lower window film reel,
9 = Capstan motor

Maintenance of measurement cell


Code 10000006053e

Revision A

Page 6 (30)

Functionality
The AWC can either be used for a window change that is automatically
repeated in a certain interval or for a window change that is triggered by using
the commands on the Probe Control Set display. The default interval, which is
the actual time the sample has flown through the measurement cell, for the
automatic window change is 8 hours. The interval can be changed in the user
parameters. You can also set the AWC not to do automatic window changes at
all with an analyzer parameter.
Each window changing operation includes four different phases that are run
automatically. The phases are:
The measurement cell is flushed
The measurement cell is drained
The measurement cell window is changed
The water test is run
The measurement cell window change is based on a pulse mechanism that
stops the film winding when the proximity sensor has counted a certain amount
of pulses for new window film. The measurement cell window change in itself
includes five phases:
Releasing the measurement cell and the measurement cell window
Starting the window winding motor
Changing the window (winding the window forward)
Stopping the window winding motor
Locking the measurement cell window and the measurement cell
After the automatic window change and a successful water test, measurements
are continued normally.

2.2 Analyzer probe with SCA


In the SCA, the automatic window changing functions cannot be used and the
measurement cell maintenance tasks must be performed manually.
The measurement cell area is comprised of:
Measurement cell
Discrete measurement cell window
Measurement cell window clamping ring
Protective window ring
Moisture leakage detection

Maintenance of measurement cell


Code 10000006053e

Revision A

Page 7 (30)

Figure 2. The SCA with the measurement cell cover closed.

1
2
3
4
5
6

Figure 3. The SCA with the measurement cell cover open.


1 = Measurement cell window clamp, 2 = Moisture leakage detections,
3 = Protective window, 4 = Protective window ring,
5 = Measurement cell window, 6 = Measurement cell

Maintenance of measurement cell


Code 10000006053e

Revision A

Page 8 (30)

3. RADIATION SAFETY
The analyzer, and the measurement cell, meets all the radiation safety
regulations of the European Union. No radiation can be detected inside or
outside the probe covers. Radiation safety is guaranteed by using redundant
fail-safe sensors with hard-wired interlock circuits that shut down the X-ray
power.
Radiation in use indicator
The X-ray ON indicator lights are located on the roof of the analyzer probe and
on the control panel of the Probe Control Set. The X-ray ON light indicates that
the interlock circuit is closed and the high voltage power supply of the X-ray
tube is on. The X-ray OFF lights indicate when the interlock circuit is open and
the high voltage power supply is off.

Figure 4. The X-ray ON indicator light.

Only trained and authorized personnel are allowed to open the


measurement cell or the side covers of the analyzer probe.

Radiation safety interlocks


The analyzer includes safety interlocks that provide extra protection both for the
user and for the X-ray tube during maintenance procedures. The radiation
safety interlocks ensure that:
When you enter the measurement cell maintenance mode, the high voltage
of the X-ray tube is automatically switched off before you are able to open
the measurement cell lock. The high voltage remains in the OFF position
through the whole maintenance process.
If the high voltage of the X-ray tube is not switched to OFF within a given
time limit, an event message is triggered and you are not able to open the
measurement cell lock, nor can you enter the maintenance mode.
The maintenance mode function can be called only if the reference is in
front of the X-ray tube and blocking the radiation.
The mechanical design ensures that the analyzer can provide radiation safety.

Maintenance of measurement cell


Code 10000006053e

Revision A

Page 9 (30)

During the automatic window change, the measurement cell is not opened
and the high voltage of the X-ray tube is ON. If the window change fails,
the high voltage is switched OFF.

Maintenance of measurement cell


Code 10000006053e

Revision A

Page 10 (30)

4. MAINTENANCE MODE
Before you start the maintenance mode of the measurement cell, you must set
the standby switch on the Courier Local Interface to standby position. Using the
maintenance view in the local display starts the maintenance mode of the
Measurement Cell.

4.1 Entering the maintenance mode


Follow these instructions carefully when entering the maintenance mode.
To enter the maintenance mode:
1.

Turn the standby switch to the STANDBY position on the Courier Local
Interface control panel.

You can proceed when the Standby - switch in STANDBY position


indication appears in the local display.

Steps 2-5 are needed, if the analyzer has AWC. If the analyzer has no AWC
then after step 1 check that the reference is in the reference position (see that
the REFERENCE indication lamp is ON), and open the measurement cell. Now
the high voltage is shut down.
2.

Go to the Measurement cell maintenance view in the local display.

3.

Select the Select desired maintenance task button to activate a selection


menu.

Maintenance of measurement cell


Code 10000006053e

Revision A

Page 11 (30)

4.

From the menu, select the desired maintenance task.

5.

Press the Maintenance Button on the probe cover within 30 seconds to


activate the maintenance task.

If the Maintenance Button is not pressed within 30 seconds, the request to


enter maintenance mode is aborted.
6.

Press the Maintenance Button on the probe cover to open the


measurement cell cover locking latch.

You are now in the maintenance mode. The basic operating principles are
described in the following sections.

Maintenance of measurement cell


Code 10000006053e

Revision A

Page 12 (30)

Operating in maintenance mode


The local display and the Maintenance Button (MB) function as a user
interface for maintenance. You can perform a wide selection of different actions
in maintenance mode.
There are two types of actions:
Actions to be first selected with the commands in the local display and then
after executed by pushing the maintenance button (MB) on the probe cover
Actions to be executed by only using the commands in the local display
These actions are presented in more detail in the following sections.
Performing actions
To perform actions in the maintenance mode:
1.

Select the desired command by touching the corresponding activation


button. The bottom of the maintenance view shows the current action of
the button.

2.

Press the Maintenance Button (MB) on the probe cover to perform the
action.

The maintenance mode includes commands that are executed without


pressing the MB. These commands are:
Reset button commands:

Reset window film consumption counter to 100%.

Set window film consumption counter to a defined value.

Reset protective window lifetime to 60 days.

Set protective window lifetime to a defined value.


Finish button to end the maintenance mode.
To execute these commands select the command button in the local display.
Tools needed
When maintaining the measurement cell area, you need a water feed bottle, an
allen wrench, and screwdrivers.

Maintenance of measurement cell


Code 10000006053e

Revision A

Page 13 (30)

4.2 Exiting the maintenance mode


After maintenance of the measurement cell, you must exit the maintenance
mode.
To exit the maintenance mode:
1.

Close the measurement cell.

Steps 2-4 are needed, if the analyzer has AWC installed.


2.

Lock the measurement cell cover.

Select the action selection button in the local display.

Press the Maintenance Button on the probe cover to execute the


action.

3.

Select the Finish button in the maintenance view.

4.

Select the exit status of the maintenance operation.

If you selected AWC OK for use, the system automatically changes


the window, and tests it with water.

If the window change succeeds, and the window is detected to be in


OK condition, the analyzer starts the high voltage raise process, and
the X-ray On indication lamp comes ON.

5.

Turn the standby switch on the Courier Local Interface front panel to the
RUN position.

Maintenance of measurement cell


Code 10000006053e

Revision A

Page 14 (30)

5. MAINTAINING THE MEASUREMENT CELL


This section presents a detailed description of the maintenance procedures in
both AWC and SCA. The maintenance of the measurement cell can be divided
into four subcategories:
Film change, when the analyzer has no AWC installed (SCA).
Half-yearly maintenance
Maintenance in fault situations
Special situations
These maintenance types are described in more detail below.
Before you can start any of the maintenance tasks described below, you
need to enter the maintenance mode. See Maintenance mode for more
detailed information.

5.1 Changing the discrete measurement cell windows in


analyzer with SCA
If you have an analyzer probe with SCA, you need to change the measurement
cell window manually by replacing the old discrete window with a new one.
To change the measurement cell window for SCA:
1.

Enter the maintenance mode. See the chapter Entering the maintenance
mode for detailed instructions.

2.

Remove the clamp holding the measurement cell window.

3.

Remove the window.

4.

Place a new discrete window carefully in the frame.

5.

Place the clamp back in its place.

6.

Exit the maintenance mode. See the section Exiting the maintenance mode
for detailed instructions.

7.

The local display prompts you for confirmation with the status message
Waiting for confirmation of MC window status, if the system is
configured to do so.

8.

Confirm by selecting one of the following alternatives from the command


list depending on the situation:

MC window not ok for any use

MC window not changed but ok for test

MC window changed and ok for test

If you reply that the window is OK for test, or the system does not require the
response, the system automatically flushes the measurement cell with water to

Maintenance of measurement cell


Code 10000006053e

Revision A

Page 15 (30)

check that the window is tight and waterproof. If the window is OK, the analyzer
starts the high voltage raise process, and the X-ray On indication lamp comes
ON.

5.2 Periodical maintenance


The measurement cell requires basic maintenance periodically. Periodical
maintenance actions include:
Changing the window film reel for the AWC and resetting the window film
consumption counter
Changing the protective window
Checking the functionality
Checking the pressure
The performance of these maintenance tasks is described in the following step
lists.

Changing the window film reel for AWC


The window film reel lasts up to 700 window changes, and therefore, needs to
be changed at approximately six month intervals.
To change the window film material:
1.

Enter the maintenance mode. See the section Entering the maintenance
mode for detailed instructions.

2.

Activate the Release measurement cell and window film action to the
Maintenance Button.

3.

Press the Maintenance Button on the probe cover to execute the


action.

Remove the sample cell window frame (attached with 4 screws).

Check the condition of the sample cell and clean if required

4.

Remove the steel plate located under the upper film reel (attached with 2
screws).

5.

Cut the window film from the upper part of the film as shown in the picture
below.

6.

Release the window film reels.

Maintenance of measurement cell


Code 10000006053e

Revision A

Page 16 (30)

Release the upper film reel by lifting the winding mechanism as shown
in the picture below.

Take the upper reel from its place as shown in the picture below.

Maintenance of measurement cell


Code 10000006053e

Revision A

Page 17 (30)

7.

Put the upper film reel in its place as shown below.

8.

Pull the window film through its proper route.

9.

Remove the old (used film) film reel. See the figure of step 11 for the latch
(at right side) holding the reel. Turn the latch 180 degrees forward and it
pops out. Now the reel is released.

10. Fix the other end of the window film to the lower reel as shown in the
picture below.

11. Put the lower reel in its proper place.

Insert the reel in place so that the two pins go in the holes located in
the left end plate of the reel.

Lock the reel by pushing and turning the latch. Turn the latch until it
stops in the groove. (See the picture below.)

12. Wind the window film forward.

Maintenance of measurement cell


Code 10000006053e

Revision A

Page 18 (30)

Activate the Run MC window motor forward action to the command


button.

Press the Maintenance Button on the probe cover to execute the


action.

Hold the upper reel firmly to ensure that the window film starts winding
around the lower reel. (See the picture below.)

Check that the window film is in a straight position.

13. Put the sample cell window frame, and the steel plate back in their places.
Make sure that all parts are properly in place and screwed tightly.
14. Tighten the window film.

Activate the Release / lock MC window action to the command


button.

Press the Maintenance Button to execute the action. The window is


now being tightened.

15. Lock the measurement cell.

Activate the Release / lock Meas Cell action to the command button.

Press the Maintenance button to execute the action. The


measurement cell is now locked.

16. Check again that the window film is straight.


17. Perform complete window change, and check the operation

Activate the Perform MC window change action to the command


button.

Press the Maintenance button to start film change. See that the film
change runs properly.

After the film change is finished, the display shows the execution
status of the film change.

Next reset the film counter as told in the next chapter.

Resetting the window film consumption counter


To reset the film counter:
1.

Select the Reset button in the maintenance view.

2.

Choose alternative:

Reset film consumption cntr to 100%, if you installed a full film reel.

Maintenance of measurement cell


Code 10000006053e

Revision A

Page 19 (30)

Set film consumption cntr, if you installed a reel that is not full, and
give the amount of film left on the reel.

You have now finished changing the window film reel.

Changing the protective window


The protective window must be changed at 2-month intervals. The instructions
presented below apply both to AWC and SCA.

a
a

Before starting the protective window change make sure that there is no
risk of tripping water or sample getting inside the analyzer probe.

Also ensure that the new protective window material is the same the
original had.

To change the protective window:


1.

Enter the maintenance mode. See the chapter Entering the maintenance
mode for detailed instructions.

2.

Remove the protective window ring (attached with 6 screws) as shown


below.

3.

Carefully remove the protective window as demonstrated below.

Maintenance of measurement cell


Code 10000006053e

Revision A

Page 20 (30)

4.

Place the new protective window in the frame.

5.

Check the condition of the silicone-sealing ring of the protective window


frame. Replace if necessary.

6.

Put the frame back in its place. Make sure that the little hole on the edge of
the frame goes through the pin in the Measurement Head Assembly.

7.

Select the Reset Protective Window lifetime command from the reset
menu, to restart the window life time calculation.

8.

Close the measurement cell, and exit the maintenance process

You have now finished changing the protective window.

Do not touch the protective window as fingerprints on the window surface


can affect the measurement results.

5.3 Maintenance in fault situations


This section provides you with detailed instructions on how to proceed in a fault
situation. If there is a malfunction in the measurement cell, the system
generates an alarm indication that informs you about the situation. First check
the alarm in the local display and then check events for more detailed
information about the cause of the situation.

Alarms and events in AWC fault situations


This section describes the alarms that might occur when working with the AWC.
You are provided with a detailed account of the causes for different alarms and
with instructions on how to proceed when you receive a certain AWC related
alarm. The possible AWC related alarms include:
MC window rupture
MC window change failed
MC window change out of film
Measurement cell open/removed
Measurement cell blocked
The instructions on how to proceed when you receive any of these alarms are
presented in the following sections.

Maintenance of measurement cell


Code 10000006053e

Revision A

Page 21 (30)

MC window rupture
The MC window rupture alarm indicates either that the measurement cell
window is ruptured or that there is moisture between the measurement cell
window and the protective window. The alarm stops all measurements in the
analyzer system and shuts down the power to the X-ray tube.
There is a moisture detection hardware sensor that triggers this alarm. Note that
the alarm remains until the measurement cell is opened and the window is
replaced / dried.
Check and correct the situation by doing the following:
1.

Enter the maintenance mode. See the section Entering the maintenance
mode for detailed instructions.

2.

Check, and clean the window opening.

3.

Change the measurement cell window.

4.

5.

Activate the Perform MC window change action to the command


button.

Press the Maintenance Button on the probe cover to execute the


action. Wait for the measurement cell window change to complete.

Check the condition of the protective window.

If the protective window is dirty, wipe it with a dry cleaning tissue.

If the protective window is worn out or even ruptured, change the


protective window as described in section Changing the protective
window.

Test the moisture sensor by touching it with a moist tissue or fingers and
see whether an MC window rupture alarm is seen on the local display.

MC window rupture alarm should disappear after the moisture sensor


is dried.

If you encounter problems during or after this procedure, please contact the
maintenance service.

MC window change failed


The MC window change failed alarm indicates that an error occurred during
the automatic window change. As this alarm is triggered, the high voltage is
shut down. There can be several different reasons for this alarm: a malfunction
in the capstan motor, a cut in the window film, a loosely or wrong way turning
winding window film material around the lower reel, or a malfunction in the
proximity sensor.
If you encounter problems during or after this procedure, please contact the
maintenance service.
MC window change first pulse timeout
If you receive an MC window change first pulse timeout alarm, which
indicates that the window film does not move, check and tighten the window
moving mechanism.

Maintenance of measurement cell


Code 10000006053e

Revision A

Page 22 (30)

To check the measurement cell window film moving mechanism:


1.

Enter the maintenance mode. See the section Entering the maintenance
mode for detailed instructions.

2.

Check the measurement cell release mechanism.

3.

4.

5.

6.

Activate the Release / lock Meas Cell action to the command button.

Press the Maintenance button a few times to see that the


measurement cell release mechanism works. Leave the measurement
cell in released position.

Check the measurement cell window release mechanism.

Activate the Release / lock MC window action to the command


button.

Press the Maintenance button a few times to see that the


measurement cell window release mechanism works. Leave the
window in released position.

Wind the window film forward.

Activate the Run MC window motor forward action to the command


button.

Press the Maintenance button on the probe cover to execute the


action. Hold the upper reel tightly to ensure that the window film winds
tightly around the lower reel.

Lock the measurement cell window and the measurement cell.

Activate the Release / lock MC window action to the command


button.

Press the Maintenance button to execute the action. The


measurement cell window is now being tightened.

Activate the Release / lock Meas Cell action to the command button.

Press the Maintenance button to execute the action.

The measurement cell and the window is now being locked and
tightened.

Exit the maintenance mode.

MC window change pulse timeout


If you receive an MC window change pulse timeout alarm, it indicates that the
window film did not move enough.
There are two alternative procedures to correct this:
If the window film is cut or ruptured, follow the instructions in chapter MC
window rupture.
If the window film is finished, change the window film material. See chapter
Changing the window film reel for AWC for detailed instructions.
If you have completed the procedure successfully, the analyzer continues with
measuring the samples normally.
If you encounter problems during or after this procedure, please contact the
maintenance service.

Maintenance of measurement cell


Code 10000006053e

Revision A

Page 23 (30)

MC window change out of film warning / alarm


The AWC incorporates a pulse mechanism to check the movement and
measure the consumption of the window film. These alarms are generated,
when the pulse count reaches parameterized limit values.
The warning indicates that you should be prepared to change the window film
reel.
When you receive the alarm, check whether the window film material is finished
In case the window film material is finished, proceed as is described in
Changing the window film reel for AWC.
In case the window film material is not finished, the pulse counter has not
been reset or the pulse count limit parameters are not correct. Reset the
pulse counter, or set the limit parameters correctly

Measurement cell open/removed


The Measurement cell open/removed alarm indicates that the measurement
cell lock is open. The alarm stops all measurements in the analyzer system and
shuts down the power to the X-ray tube.
When you receive this alarm, do the following:
Enter maintenance mode.
Check the measurement cell cover mechanism.
Exit maintenance mode.
When the measurement cell cover is locked, the system automatically raises
the X-ray tube high voltage and continues operating normally.
In case the alarm still persists, contact the maintenance service.

Maintenance of measurement cell


Code 10000006053e

Revision A

Page 24 (30)

Measurement cell blocked


The Measurement cell blocked alarm indicates that there is foreign material
(rocks, wood chips, plastic) in the measurement cell and/or in the related hoses.
The alarm sets the measurement cell and the related hoses into a fault state
immediately, and no samples will flow through the measurement cell until the
alarm is cleared. Only reference measurements are performed.
If you receive an event message indicating that the measurement cell is
blocked, do the following:
1.

Enter the maintenance mode. See the section Entering the maintenance
mode for detailed instructions.

2.

Remove the sample cell frame.

3.

Check weather the measurement cell and/or related hoses are blocked.

4.

In case found, remove the blockage from the measurement cell and check
that the measurement cell is clear of foreign material.

5.

Place the sample cell frame back in its place.

6.

Close the measurement cell cover.

7.

Exit measurement cell maintenance.

If you have completed the procedure successfully, the analyzer continues with
measuring the samples normally.
If you encounter problems during or after this procedure, please contact the
maintenance service.

Maintenance of measurement cell


Code 10000006053e

Revision A

Page 25 (30)

Alarms and events in SCA fault situations


This section describes the alarms that might occur when working with the SCA.
You are provided with a detailed account of the causes for different alarms and
with instructions on how to proceed when you receive a certain SCA related
alarm. The possible SCA related alarms include:
MC window rupture
Measurement cell open / removed
Measurement cell blocked
The instructions on how to proceed when you receive any of these alarms are
presented in the following.

MC window rupture
The MC window rupture alarm indicates either that the measurement cell
window is ruptured or that there is moisture between the measurement cell
window and the protective window. The alarm stops all measurements in the
analyzer system and shuts down the X-ray tube power.
A moisture detection sensor triggers this alarm. Note that the alarm remains
until the measurement cell is opened and the window is replaced / dried.
Check and correct the situation by doing the following:
1.

Enter the maintenance mode. See the section Entering the maintenance
mode for detailed instructions.

2.

Remove the clamp holding the measurement cell window.

3.

Remove the window.

4.

Place a new discrete window carefully in the frame.

5.

Wipe and dry the protective window with a dry cleaning tissue.

6.

If the protective window is worn out or even ruptured, change the protective
window as described in section Changing the protective window.

7.

Place the clamp back in its place.

8.

Exit the maintenance mode. See the section Exiting the maintenance mode
for detailed instructions.

9.

The local display prompts you for confirmation with the status message
Waiting for confirmation of MC window status, if the system is
configured to do so.

10. Confirm by selecting one of the following alternatives from the command
list depending on the situation:

MC window not ok for any use

MC window not changed but ok for test

MC window changed and ok for test

If you reply that the window is OK for test, or the system does not require the
response, the system automatically flushes the measurement cell with water to
check that the window is tight and waterproof. If the window is OK, the analyzer

Maintenance of measurement cell


Code 10000006053e

Revision A

Page 26 (30)

starts the high voltage raise process, and the X-ray On indication lamp comes
ON.
If you encounter problems during or after this procedure, please contact the
maintenance service.

Measurement cell open/removed


The Measurement cell open / removed alarm indicates that the measurement
cell lock is open. The alarm stops all measurements in the analyzer system and
shuts down the power to the X-ray tube.
When you receive this alarm, do the following:
Check whether the measurement cell cover is actually open or not.

In case it is open, close it carefully.

When the measurement cell cover is locked, the system automatically raises
the X-ray tube high voltage and continues operating normally.
In case the alarm still persists, contact the maintenance service.

Measurement cell blocked


The Measurement cell blocked alarm indicates that there is foreign material
(rocks, wood chips, plastic) in the measurement cell and/or in the related hoses.
The alarm sets the measurement cell and the related hoses into a fault state
immediately, and no samples will flow through the measurement cell until the
alarm is cleared. Only reference measurements are performed.
If you receive an event message indicating that the measurement cell is
blocked, do the following:
1.

Enter the maintenance mode. See the section Entering the maintenance
mode for detailed instructions.

2.

Remove the measurement cell window ring.

3.

Check weather the measurement cell and/or related hoses are blocked.

4.

In case found remove the blockage from the measurement cell and check
that the measurement cell is clear of foreign material.

5.

Place the measurement cell protective window ring back in its place.

6.

Close the measurement cell cover.

7.

Flush and drain the measurement cell.

Maintenance of measurement cell


Code 10000006053e

Revision A

Page 27 (30)

If you have completed the procedure successfully, the analyzer continues with
measuring the samples normally.
If you encounter problems during or after this procedure, please contact the
maintenance service.

5.4 Special situations


There are certain special situations that you might encounter with AWC and/or
SCA. These situations include:
Manual AWC window change.
A separate window filmstrip used with the AWC.
Changing the front plate of the sample cell applying to both measurement
cell types.
These special situations are described in more detail in the following.

Changing the window for the measurement cell with AWC


If you have a measurement cell with AWC, you do not need to change the
measurement cell window under normal operating conditions. If there is a
malfunction in the measurement cell mechanisms, for example, the motor is
down, you can change the window manually.
Before you intend to change the window manually you must disable the
automatic window change by setting the appropriate user parameter at
the Options Manager. This parameter is only for temporary use.

To change the window for the measurement cell with AWC manually:
1.

Enter the maintenance mode. See the chapter Entering the maintenance
mode for detailed instructions.

2.

Release the measurement cell and the measurement cell window.

Activate the Release / lock Meas Cell action to the command button.

Press the Maintenance Button on the probe cover to release the


measurement cell.

Activate the Release / lock MC window action to the command


button.

Press the Maintenance Button on the probe cover to release the


window. The measurement cell and the window are now released.

3.

Check that the window is in proper condition.

4.

Wind the window film forward by turning the lower window reel manually.
See the section Changing the window film reel for AWC for releasing the
lower window reel.

5.

Lock the measurement cell window and the measurement cell.

Activate the Release / lock MC window action to the command


button.

Maintenance of measurement cell


Code 10000006053e

Revision A

6.

Page 28 (30)

Press the Maintenance Button to execute the action. The


measurement cell window is now being tightened.

Activate the Release / lock Meas Cell action to the command button.

Press the Maintenance Button to execute the action.

The measurement cell and the window is now being locked and
tightened.

Exit the maintenance mode.

The automatic window change will not start as the user parameter was
set to disable it.

You have now finished changing the measurement cell window.

AWC requires using a separate window film strip


With AWC, you might encounter situations in which you cannot use the normal
window film material. In this case you should use a separate window filmstrip as
a replacement. You can do so by running the piece of film through the normal
film path of the AWC
Before you intend to use a separate window filmstrip you must disable the
automatic window change by setting the appropriate user parameter at
the Options Manager. This parameter is only for temporary use.

Maintenance of measurement cell


Code 10000006053e

Revision A

Page 29 (30)

AWC or SCA requires changing the front plate of the sample cell
With both AWC and SCA, you might have to change the front plate of the
sample cell.

If you encounter this situation with AWC, do not try to change the front
plate by yourself. Contact Customer Support for assistance.

If you encounter this situation with SCA, do the following:


1.

Enter the maintenance mode. See the section Entering the maintenance
mode for detailed instructions.

2.

Remove the clamp holding the measurement cell window as shown in the
picture below.

Maintenance of measurement cell


Code 10000006053e

Revision A

3.

Page 30 (30)

Remove the front plate of the measurement cell (attached with 4 screws).

Figure 5. SCA parts.


1 = Front plate of the measurement cell, 2 = Measurement cell
4.

Attach a new front plate in place.

5.

Attach the clamp holding the measurement cell window in place.

6.

Close the measurement cell.

You have now finished changing the front plate of the measurement cell.

TELEFAX MESSAGE

DATE:

Page 1

READER COMMENTS
The design team of Outotec is committed to ensure the quality and usefulness of this manual
and the COURIER analyzer system. Your comments, recommendations and suggestions will
help make our continuous effort more effective.
Please take few moments to write your comments either directly on the form provided or on a
photocopy of it.
Your assistance is greatly appreciated.
Thank you for choosing the COURIER analyzer system.

Fax to:

Outotec, Analyzers
FAX: INT + 358 205 292998

or mail to:

P.O. Box 84, FI-02201 Espoo, Finland

or e-mail to:

analyzers@outotec.com

Attn:

Manager, Analyzer Sales Support

From:

Analyzer Serial Number:


Manual:

Courier 6i and Courier 5i


SL Slurry and HX Solution Analyzer
Maintenance of measurement cell
Code 10000006053e Revision A
( Please identify applicable software versions and individual
documents )

Comments /
recommendations:

Number of additional pages (

Você também pode gostar