Quantity Measured/
Instrument

Range / Frequency

*Calibration Measurement
Capability ()

Remarks

SOURCE#
1.

DC VOLTAGE

0.1 mV to 1 mV
1 mV to 300 mV
300 mV to 1000 V

3.685% to 0.36%
0.36% to 0.013%
0.013% to 0.011%

Using Fluke MultiProduct


calibrator
By Direct Method

2.

DC CURRENT

10 A to 3 mA
3 mA to 300 mA
300 mA to 11 A

0.817% to 0.022%
0.022% to 0.019%
0.019% to 0.089%

Using Fluke MultiProduct


calibrator
By Direct Method

10 A to 550 A

0.089% to 1.052%

Using current coil


By Direct Method

10 Hz to 500 kHz
1 mV to 10 mV
10 mV to 30 mV
30 mV to 3 V

2.76% to 1.86%
1.86% to 0.12%
0.12% to 0.73%

Using Fluke MultiProduct


calibrator
By Direct Method

10 Hz to 100 kHz
3.3 V to 33 V

0.069% to 0.90%

45 Hz to 20 kHz
33 V to 330 V

0.18% to 0.07%

50 Hz to 10 kHz
30 V to 1000 V

0.05% to 0.29%

10 Hz to 5 kHz
30 A to 300 mA
300 mA to 2.2 A

1.189% to 0.29%
0.29% to 1.73%

3.

4.

AC VOLTAGE

AC CURRENT

Using Fluke MultiProduct


calibrator
By Direct Method

Laboratory

National Centre for Quality Calibration, 902-903-904, SAMUDRA


Annexe, Sardar Patel Nagar, Off. C. G. Road, Ahmedabad, Gujarat

Accreditation Standard ISO/IEC 17025:2005


Discipline

Electro-Technical Calibration

Issue Date

11.04.2014

Certificate Number

C-0502

Valid Until

10.04.2016

Last Amended on

Page

2 of 6

Quantity Measured/
Instrument

AC CURRENT

5.

RESISTANCE
(2 wire and 4 wire)

6.

FREQUENCY

7.

CAPACITANCE

Range / Frequency

*Calibration Measurement
Capability ()

Remarks

10 Hz to 1 kHz
2.2 A to 11 A

1.73% to 0.40%

Using Fluke MultiProduct


calibrator
By Direct Method

50 Hz
10 A to 550 A

0.40% to 1.54%

Using current coil


By Direct Method

0.1 to 1
1 to 10
10 to 3 k
3 k to 3 M
3 M to 330 M

9.27% to 0.94%
0.94% to 0.107%
0.107% to 0.016%
0.016% to 0.195%
0.195% to 1.13%

Using Fluke MultiProduct


calibrator
By Direct Method

330 M to 999 M

2.31% to 3.74%

Using Resistance Decade


Box
By Direct Method

3 Hz to 10 Hz
10 Hz to 120 Hz
120 Hz to 2000 kHz

0.198% to 0.062%
0.062% to 0.012%
0.012% to 0.039%

Using Fluke MultiProduct


calibrator
By Direct Method

1 kHz
1 nF to 11 nF
11 nF to 1 mF

1.34% to 0.70%
0.70% to 1.335%

Using Fluke MultiProduct


calibrator
By Direct Method

Laboratory

National Centre for Quality Calibration, 902-903-904, SAMUDRA


Annexe, Sardar Patel Nagar, Off. C. G. Road, Ahmedabad, Gujarat

Accreditation Standard ISO/IEC 17025:2005


Discipline

Electro-Technical Calibration

Issue Date

11.04.2014

Certificate Number

C-0502

Valid Until

10.04.2016

Last Amended on

Page

3 of 6

Quantity Measured/
Instrument

8.

9.

Range / Frequency

*Calibration Measurement
Capability ()

TEMPERATURE
SIMULATION
Thermocouple
(Temperature Indicator /
Controller/ Recorder /
Test Kit / Universal
Calibrator/ Calibrators)
RTD
Pt-100

200C to 800C

0.10C to 0.29C

Thermocouple
B type

600C to 1800 C

0.55C to 0.47C

C type

0 to 2300 C

0.41C to 1.02C

E type

250C to 1000 C

0.62C to 0.34C

J type

210C to 1200 C

0.38C to 0.36C

K type

200C to 1350 C

0.44C to 0.52C

L type

200C to 900 C

0.48C to 0.30C

N type

200C to 1300 C

0.51C to 0.40C

R type

0 to 1750 C

0.69C to 0.53C

S type

0 to 1767 C

0.58C to 0.6C

T type

250C to 400 C

0.76C to 0.27C

U type

200C to 600 C

0.68C to 0.38C

50 Hz
250 V
5A
UPF to 0.5 PF
(lag and Lead)

0.8%

POWER/ ENERGY
(1 and 3)

Remarks

Using Fluke MultiProduct


calibrator
By Simulation method

Using Power Energy


Calibrator
By Direct Method

Laboratory

National Centre for Quality Calibration, 902-903-904, SAMUDRA


Annexe, Sardar Patel Nagar, Off. C. G. Road, Ahmedabad, Gujarat

Accreditation Standard ISO/IEC 17025:2005


Discipline

Electro-Technical Calibration

Issue Date

11.04.2014

Certificate Number

C-0502

Valid Until

10.04.2016

Last Amended on

Page

4 of 6

Quantity Measured/
Instrument

Range / Frequency

*Calibration Measurement
Capability ()

Remarks

MEASURE#
10.

DC VOLTAGE

1 mV to 10 mV
10 mV to 1000 V

0.413% to 0.075%
0.075% to 0.011%

Using 6 Digital
Multimeter
By Direct Method

11.

AC VOLTAGE

50 Hz
1 mV to 10 mV
10 mV to1000 V

4.726% to 0.532%
0.532% to 0.098%

Using 6 Digital
Multimeter
By Direct Method

12.

DC CURRENT

10 A to 100 mA
100 mA to 1 A
1 A to 10 A

0.361% to 0.065%
0.065% to 0.083%
0.083% to 0.192%

Using 6 Digital
Multimeter
By Direct Method

13.

AC CURRENT

50 Hz
10 A to 1 A
1 A to 10 A

0.95% to 0.169%
0.169% to 0.253%

Using 6 Digital
Multimeter
By Direct Method

14.

RESISTANCE

1 to 1 k
1 k to 10 M
10 M to 1 G

0.358% to 0.025%
0.025% to 0.059%
0.059% to 2.33%

Using 6 Digital
Multimeter
By Direct Method

15.

FREQUENCY

5 Hz to 1 MHz

0.116% to 0.015%

Using 6 Digital
Multimeter
By Direct Method

Laboratory

National Centre for Quality Calibration, 902-903-904, SAMUDRA


Annexe, Sardar Patel Nagar, Off. C. G. Road, Ahmedabad, Gujarat

Accreditation Standard ISO/IEC 17025:2005


Discipline

Electro-Technical Calibration

Issue Date

11.04.2014

Certificate Number

C-0502

Valid Until

10.04.2016

Last Amended on

Page

5 of 6

Quantity Measured/
Instrument

Range / Frequency

*Calibration Measurement
Capability ()

1 nF to 1 F
1 F to 4.7 mF

5.197% to 1.735%
1.735% to 2.39%

600C to 1820 C

0.55C to 0.46C

C type

0 to 2300 C

0.4C to 1.01C

E type

250C to 1000 C

0.61C to 0.33C

J type

210C to 1200 C

0.37C to 0.35C

K type

200C to 1300 C

0.43C to 0.52C

L type

200C to 900 C

0.47C to 0.29C

N type

200C to 1300 C

0.50C to 0.39C

R type

0 to 1750 C

0.69C to 0.53C

S type

0 to 1750 C

0.58C to 0.59C

T type

250C to 400 C

0.75C to 0.26C

U type

200C to 600 C

0.68C to 0.37C

RTD
Pt100

200C to 800 C

0.11C to 0.35 C

16.

CAPACITANCE

17.

TEMPERATURE
SIMULATION
Temperature Calibrators/
Test Kit / Universal
Calibrator
Thermocouple
B type

Remarks

Using 6 Digital
Multimeter
By Direct Method

Using Fluke MultiProduct


calibrator
By Simulation Method

Using 6 Digital
Multimeter & Multi
Function Calibrator

Laboratory

National Centre for Quality Calibration, 902-903-904, SAMUDRA


Annexe, Sardar Patel Nagar, Off. C. G. Road, Ahmedabad, Gujarat

Accreditation Standard ISO/IEC 17025:2005


Discipline

Electro-Technical Calibration

Issue Date

11.04.2014

Certificate Number

C-0502

Valid Until

10.04.2016

Last Amended on

Page

6 of 6

Quantity Measured/
Instrument

Range / Frequency

ENERGY
(1 and 3 )

50Hz
250 V
6A
0.5 to UPF (Lead/Lag)

*Calibration Measurement
Capability ()

Remarks

By simulation method
18.

19.

TIME
(Timer, Stop Watch etc.)

20 s to 60 s
60 s to 3600 s
3600s to 86400 s

0.31% to 0.35%

Using Three Phase Energy


meter
By Direct Method

0.44s to 1.29s
1.29s to 1.50s
1.50s to 17.34s

Using Master Stop watch


By comparison method

* Measurement Capability is expressed as an uncertainty () at a confidence probability of 95%


#

The laboratory is also capable for site calibration however, the uncertainty at site depends on the prevailing actual