INDEX
DESCRIPTION

Page #

Index

Evoked Potential main window

Heading Bar, Status Bar, Menus Bar

Menus, File Menu


Record Menu
Edit Menu
View Menu
Tools Menu
SetUp Menu
Help Menu
Start Collection

7-8
9
10 11
12 13
14 17
18 19
20
21

Quick Keys Help

23

Electrodes

24-26

Technical Specifications

27-29

Install the Software


PCMCIA Installation
Setting Up the equipment, external
connection
Front and Back Pannel, Headbox
Front and Back Pannel, 4 channel Headbox
Footstand Adapter

30
31-32
33-35

Troubleshooting

41

37
38
39

4 Potenciales Evocados V.10.0

5 Potenciales Evocados V10.0

Evoked Potential main window

Heading Bar

Menus Bar

Parameters

Information
Bar
Active Waveform

Status Bar

Context Menu
Display Area

Waveform Display area

6 Potenciales Evocados V.10.0

Heading Bar

Status Bar

Menus Bar

7 Potenciales Evocados V10.0

Menus

File
Record
Edit
View
Tools
SetUp
Help
File Menu
New
Open
Close
Close All
Save
Save All
Select Print Items
Print
Print Preview
Print Setup
Ultimos 8 estudios abiertos
recientemente
Exit

Ctrl + N
Ctrl + O

Ctrl + S

Ctrl + P

Ctrl + X

New
In this option you can start a new study. Select the File menu>new option>a window
will appear with the information that you must fill for the patient.
Surname, name, date of the study, clinical info.
Press OK.
Open
In this option you will open the file that you want to review.
The system automatically will open the folder where the files are saved, PUD.
You can select one or more waveforms to view on the screen, select all the waveform
that you wish to see in the same study and open all together.
You can select more than one waveform in the same time as you have the option in the
windows programs.
Select the File>Open> and select one waveform press the shift key and move the downup arrow or with a click of the mouse, the selectable waveform will appear on the
waveform display area.

8 Potenciales Evocados V.10.0

Close
In this option you can close one by one the active waveform, select the one that you
want to close with the arrows, the waveform turns active when it changes the color to
red.
Select the file menu>close> the active waveform will disappear.
Close All
Same as the last option, but closes all the waveform presented in the waveform display
area.
Before to close, the system will ask you to save the study and automatically will open
the window save in the correct folder, if you dont want to save changes exit of this
window.
Save
Save the active waveform.
Save all
Save all the waveforms that presented in the waveform display area
Select Print Items
From this option you can select the information that you want to be printed.
You can select Waveforms, Parameters and Patient Information. If you have the three
options with the mark 4three options will be printed. If you want to print just one of
these three options tip out the option you dont want to print.
4 Waveform
Parameters
4 Patient Information

Print
Prints the study
Print Preview
This option allows you to have a view of the study previously to print it and set the
position, weight etc.
Print Setup
Allows you to set the configuration of the printer, paper size etc.
Exit
Exits of the application

9 Potenciales Evocados V10.0

Record Menu

Continue
Overwrite
Start Next
Study Type
Input Channel
Impedance Test
Continue

Continue the active waveform increasing the existing average number.


Continuos the collection in the waveform that you where doing or in other that you
choose.
Overwrite
Overwrite active waveform and begin a new collection from average number = 0.
Start Next

Using this option you start a new waveform and it will be the new active.
The extension name of this waveform will be increased automatically. For example, if
you begin a study called Peter and you start more than one waveform the new
waveforms you will begin with this option will be called Peter.000, Peter.001, Peter.002
etc.
Change Study Type
Using this option you can change the Study Type to use during the study.
Record Menu > Change Study Type > a box will appear with a displayed list including
all the study type the program have as default and all the study type you have made
before. Select the option OK and the study type will change automatically, you are
ready to begin a new collection.
Input Channel
From this option, it allows you to define channels for the signal input.
Impedance Test
You can see the impedance value of each electrode in Kohms.
If you are sure that the reference and ground electrodes have good contact to the patient
and the system gives you values out of accepted you must chek:
1. Using same material electrodes (NO mixed electrodes)
2. Low quality of materials or electrodes
3. You are using good quality conductive paste

10 Potenciales Evocados V.10.0

Edit Menu
Patient Info
Patient Report
Report Template
Montage
Note
New Full Screen DIB
Open DIBs
Patient Info
Here it will appear a window to add the information of the patient.
-Surname and Name
-Date of study
-Study Number
-Clinical Info
Patient Report
Selecting this option the PNK software will open the Microsoft Word or WordPad
(what you have installed on your computer) and will generate a new word file called as
the study filename, for example Peter.rep.
The file will be saved on the PUD folder except you specify a new folder.
Microsoft is a registered trademark of Microsoft Company.
Report Templates
You have the possibility of using a template to write the report. Remember that you
have to have on view the study you will wite the epot of.
Once you select this option the open box will appea showing the DOT files.
You can change the template and save it wiht anothe name or with the same.
This is an useful tool to have the template achived with your clinic, institution and
addres. When you open it you only will fill it with the patient information and study
analysis. In thisd templates you have the possibility to add the BMP Graphic Screen.
Montage
Slecting this option will appear a box where you can type the montage name you used
for the collection. This information will appear in the Status Bar.

Edit Note
Each trace may have a different note. It is possible to write or change the note of the
active waveform.
You have two possibilities to edit or write a new note, one is using this menu and the
other one is using the contextual menu.
You must be sure you are inserted the Note to the active waveform and after you have
inserted the Note Save the active trace.

11 Potenciales Evocados V10.0

New Full screen DIB


Selecting this option you will create a New BMP graphic screen caption.
Select EDIT Menu > New Full screen DIB, the save file box will appear automatically.
Write down a name for the new file. It will be saved in the C:\AKODATA\EP folder.
Press the Save button.
Open DibS
From this option you will open the graphic screen captions that you have done before.
The open file box will be open in the corresponding folder automatically,
C:\AKODATA\EP in which the graphic screen are. The extension of the files is DIB.
Select one or more than one file. Use the mouse only with a click or by keeping pressed
the SHIFT key and clicking over the others.
Select the Open button.
The Microsoft Paint program will be open, this program cames in standard way with the
Window but if you have another kind of digital imaging graphic program you can use
them to view and reformat the image.
Microsoft Paint is a registered trademark of Microsoft Corp.

Other way to work with graphic screen images


Previous to work you have to select an area doing a click on the beginning point and
take the mouse up to the point you want it finished. It will be shown a line box outside
the area. When you left the button on the right of this box will appear a menu with two
options:
Copy to Clipboard
This option is for copiyng the area in a part of the computer memory.
When you select this option you have to open other program where you want or can
paste the copy you have done. It could be for example: Microsoft Word, Adobe
Photoshop etc.
You must copy and paste inmediatly because only one copy is keeped in the computer
memory.
New DIB
As in the last option you can generate a new BMP graphic file but just of the selected
area. The procedure is the same as in New Full Screen DIB.
Its no necesary to paste this file inmediatly you done it. It is achived in the same folder
as other graphic screen files are saved. You will be able to use it whenever you want.

12 Potenciales Evocados V.10.0

View Menu
Toolbar
Status bar
Select New Color
Change Trace Color
Toolbar
Keeping this option you have the possibility to have a quick acces to the tools we show
you in this menu.
New File
Open File
Print File
Save File
Waveform Superpoesition
Equal Spaced Waveforms
Moves the waveform UP in the waveform display area
You can also move the waveform UP by selecting the active waveform,
making a click over the number keeping the key mouse press and moving UP
to the point you want.
Moves the waveform DOWN in the waveform display area
You can also move the waveform DOWN by selecting the active waveform,
making a click over the number keeping the key mouse press and moving
DOWN to the point you want
Increase the amplitude of the active waveform
Decrease the amplitude of the active waveform
Inverts the wave of the active waveform
Continue the collection in the astarted register or in a new register
Overwrite the active waveform starting a new collection over the channel
Begin a new collection and add it automatically in the waveform display
area
Presents the marks previously setted in the active waveform
Edits the marks inserted before
Status Bar

13 Potenciales Evocados V10.0

Selecting this option you can see or not this information bar that is presented on the
bottom of the screen. Once selected this option it must have a mark for been visible 4

The color of the waveform are possible to be set from this menu.
Select New Color
Once the study has started you will see the trace on the screen in the color that you have
set the last time you have used the program.
From this option you can set the color of the new trace you will use and this will keep it
by default.
Change Trace Color
At the beginning of the study the trace will be shown in the color you have set before.
Selecting this option you can change the color in the same trace the times you want.

14 Potenciales Evocados V.10.0

Tools Menu
Waveform Superposition
Equal Spaced Waveforms
Waveform UP
Waveform DN
Display x UP
Display x DN
Inversion
Show Marks
Show Marking Values
Edit Marks
Digital Filter
Fourier

All the available options of the Tools Menu can be open and used by clicking on
Context Menu Area with the right key of the m.
Waveform Superposition
In this option you can make a waveform superposition of all the waveform presented in
the waveform display area
Equal Spaced Waveforms
If you did a waveform superposition of all the presented waveform in the display area,
with this option you returned the waveforms to their position in the display area.

Waveform UP
This option moves the selected waveform UP in the waveform display area.
You have this option available in this menu, in the Tool Bar using the corresponding
button and make a click over the contextual menu area keep the button preseed and take
the wavveform up to the position you want.
Waveform DN
This option moves the selected waveform DN in the waveform display area.
You have this option available in this menu, in the Tool Bar using the corresponding
button and make a click over the contextual menu area keep the button preseed and take
the wavveform up to the position you want.
Waveform Display x UP
Selecting this option you will increase the amplitude of the selected waveform in the
display area.
You have this option available in this menu, in the Tool Bar using the corresponding
button and make a click over the contextual menu area keep the button preseed and take
the wavveform up to the position you want.

15 Potenciales Evocados V10.0

Waveform Display x DN
Selecting this option you will decrease the amplitude of the selected waveform in the
display area.
Inversion
This allows you to invert the waveform, this will invert the active waveform. If you
want to return to the before position, select this option again
Show Marks
Selecting this option the software will present on the waveform display area and over
the active waveform the marks. Marks could be set up in the Study Type option from
the SetUp menu.
Marks will appear automatically and by default the label of each one will be saw in
roman numbers.
Show Marking Values
Selecting this option you will see the latency values over each mark.

Edit Marks
If the mark insertion isnt correct or you want to change it, you can select this option
and a box will appear with this different option:

From this box you are able to:


- Define the mark visible or not (Enabled 4)
- Show the line of the marks or not (Show Line)
- Increase or decrease the line length of the marks
- Shows the label and value of the selective mark above or below the line (Line UP)
- Change the LABEL Name of the selective mark. This name is presented in roman
numbers automatically.
- Change the differencies between marks.

To change marks position:


Having the Edit Marks box open in the screen, you can make a click over the mark you want to change
position, keep the button pressed and move the line up to the desired position.
Selecting the Tools Menu > the Edit Marks option > and changing the numeric values.

16 Potenciales Evocados V.10.0

Fourier
-Previous to select Fourier Option you have to set the desired waveform to make the
analysis active. The selected waveform will be highlight with a red box.
-Select the Tools Menu
-Select the Fourier option
-The program will present you
-The horizontal Axis: Hz.
-The Vertical Axis: uV2
The Power Spectrum

In the First Line appear the


Filename and the selected waveform
This is, for example:
Alfred.000 or Alfred.001
As it is shown in the graphic.

In the First Line appear the


Filename and the selected waveform
This is, for example:
Alfred.000 or Alfred.001
As it is shown in the graphic.

17 Potenciales Evocados V10.0

To select another waveform and make the FOURIER analysis you have to open the
waveform from the corresponding file, once the waveform is opened select the Tools
Menu > FOURIER option.
The software allows you to have more than one waveform opened and from differents
files, so you can have them open and turn active the one you want to make the
FOURIER analysis.

18 Potenciales Evocados V.10.0

Setup Menu
Notch Filter
Study Type
Warning (Save)
Technical

50Hz. 60 Hz.

Select Interface
Select Stimulator
Select Amplifiers
Calibrate Zero

Notch Filter
From this option you can select the Notch Filter between 50 or 60 Hz. According to the
country you are working in. The filter could be activated or not from the Information
Bar DATA.
Study Type
From this option the system allows you to edit or create a new study type and it is added
authomatically to the studies type list.
In this box appears:
Options:

NEW - It will create a new study type


APPLY It will apply the changes you have selected for the selected study
type or for the new study type you have created.
CLOSE Close the box

DELETE It will delete permanently the selected study type.

To create a new study type once you have selected the NEW option or to change the
parameters of one study you have this option:
- Analysis Time
- Channel #
- Amplifiers parameters
- Register parameter
- Stimulus parameter: Visual, Auditory, and Somatosensory
- Marks From this box you can set the marks to insert them automatically in the
study by selecting the Show Marks option or using the Tools bar button.

19 Potenciales Evocados V10.0

Warning (Save)
Keeping this option active with a tick on its side, the system will display a box with the
following message:
SAVE ACTIVE WAVEFORM
Save Changes in: C:\AKODATA\filename.000

You can select SAVE: To save the trace which name is shown in the box.
Option CLOSE, with this option the program will ask you if you have saved the
changes.
Close All option: If you select this option this box will appear with any study you have
open.
The CLOSE ALL option will save all the opened studies without asking with each one.
CANCEL: This option will cancel the changes and will not save anything.
If the Warning (SAVE) option wasnt activated, no message will appear asking to save
the changes made to the study.
In this case you have to be careful to remember if the changes were saved or not.
Technical
This option is for being used by technicians.
Select Interface It will select the Interface that the equipment will work with
Select Stimulator It will select the Stimulator type that the system will work with.
Select the Amplifier Type It will select the Amplifier type that the system will work
with
Calibrate Zero

20 Potenciales Evocados V.10.0

Help Menu
Help Topics
Manual

English
Spanish

Calendar
Calculator
Contact Akonic
About EP

Help Topics
You have in this option a complete help on line.
You are able to search in alphabetic order or item order.
Manual. Once you select this option, the program will open the Adobe Acrobat Reader.
If you dont have it installed on your PC it is easy to find on the internet in Adobe.com.
The manual is accessible in English or Spanish, just select the option you want for
reading or printing it.
Then you will find tools that will help you to calculate measurements know dates in the
calendar and take notes.

21 Potenciales Evocados V10.0

Start Collection
Once you have the Software installed, you add the electrodes to the patient:
Select the record menu > the Input channel option

Select the channel wich will be used in the study.


Select the Record menu > Impedance test option
This option will show the impedance of each electrode in Kohms.
If you are sure that the reference electrode and ground electrode are well connected and
values are out of the accepted values check the following items:
-Same material electrodes
-Low quality of electrodes
-Good adhesive paste quality
Select the File Menu > New option, it will appear a new box called Patient Information
where the patient information could be added
Name Surname
Study #
Study Date
Previous studies
After you write all the desired information press the OK button.

22 Potenciales Evocados V.10.0

It will appear a new window with a study types list, you must select one (The study type
must be defined before from the SetUP menu)
During the collection the Information Bar shows you the parameters that are being used
and are set according to the study type you have selected.
Those parameters are possible to be changed during the study. In this bar are shown 4
options:

DATA
Stimulus Frequency
Amplifiers
High Frequency Filter HZ
Low Frequency Filter HZ
Sensitivity

Average #

This is the number of average where the study stops automatically. You
can set this number in the Setup Menu.

Rejected #

This is the number of rejects that the collection had. The study dont
average them.

Display X

This is the information about the quantity of times that you increase the
amplitude of the selected waveform.

Cursor 1
Cursor 2
Diff
Amplitude
UV/Div

Latency 1 value
Latency 2 value
Value of the difference between cursor 1 and cursor 2
Amplitude between cursors 1 and 2
This is the information about the sensitivity of the selected waveform
per division

AUDIO
Stimulus Type:
Click Masking P300
Auditory stimulus parameters
Right
Left

VISUAL
Stimulus Type:
Photic - Pattern
Visual stimulus parameters

23 Potenciales Evocados V10.0

During the collection the Tools Bar change and the program shows you the quick
keys for changing the parameters:
F3: True/Average
F4: Notch
F5/F6: HFF
F7/F8: LFF
F9/F10: Gain
+ : DispxUP
- : DispxDN
DEL: Delete
R: Reject
I: Inv
RATE: P.Up/P.Dwn

Toggle between True / Average


Activate or deactivate the Notch Filter
Increase or Decrease High Frequency Filter
Increase or Decrease Low Frequency Filter
Increase or Decrease Gain
Increase the Display
Deacrease the Display
Delete the # of average and begin from 0
Activate or deactivate the rejection
Inverts the waveform
Increase or Decrease the stimulation RATE

SPACE Bar / ESC

Stops the collection

24 Potenciales Evocados V.10.0

Electrodes

25 Potenciales Evocados V10.0

26 Potenciales Evocados V.10.0

27 Potenciales Evocados V10.0

General Specifications
SEcurity
System Capabilities
Power Conection
System Requirements
Use, transportation and storage
Components
Setting Up the System
Troubleshooting

28 Potenciales Evocados V.10.0

Security
The Evoked Potential System and all their components have been designed to
meet the electrical security standards. (IEC 601-1)
System Capabilities
Channels # :

4 Channels

Impedance Input:

> 20Mohm

CMRR:

110dB a 50/60 Hz.

Noise Level:

0.7 uV.

Notch Filter:

50-60Hz

HFF:

10.000-5.000-3.000-2.000-800-300-100-300 Hz

LFF:

0.15-0.5-1.5-5-10-30-100-300 Hz

Sensitivity:

3-6-12-25-50-100-200-400-800-1600-3200-6400 uV

Stimulators
Auditory
Visual
Somatosensory
System Requirement
Pentium II o III, IBM
Windows 98 / Milenium
32 Mb RAM memory
Hard Disk 2 Gb
Drive 1.44 Mb
CD-ROM
1 slot ISA bus libre
Expanded keyboard of 101 keys alfa numeric and functions. Mouse
Color Monitor de 15 (1280 x 1024 pixels)
Akonic S.A. reserves the right to change any modifications without prior notice

Operative system requirement


Windows 98 / Millenium

29 Potenciales Evocados V10.0

ATTENTION!!!
Use always the functional ground connection not just to obtain a good
register and to avoid electrical risks to the technician/physician or the
patient in a bad equipment function.

System Components
All the system components are detailed in the Packing List.

30 Potenciales Evocados V.10.0

Installing EP Software
1 - INSTALLATION
The akonic Software runs under Windows 98 and Millenium
We provide you a CD with the Software
To install:
Insert the CD in the CD driver, go to the desktop, select Start, Run and select from the
CD unit SETUP.exe
The installation software will show you the Disk and Directory:
C:\AKONIC
It will create a folder named AKONIC and inside this folder will create another folder
for each program.
To create a direct access to the program from the desktop you must:
- Go to the windows explorer
- Select the program you want to create a direct access
- With the right mouse button select from the context menu the option Send to >
desktop as direct access
- To change the Icon that will appear in the desktop
3

DEMO STUDIES
All the DEMO studies provided by akonic s.a. are copied in the corresponding
folder.
The corresponding folder for each program are:
C:\Akodata\ eeg
C:\Akodata\ ep
C:\Akodata\ emg
C:\Akodata\ eeg16
Etc.

31 Potenciales Evocados V10.0

PCMCIA Installation under Windows


Once you connect the PCMCIA card, Windows recognize it and insert the draw of the Card in
the right side in the bottom of the screen, near the watch. This doesnt mean that the card is
correcting installed, only shows the zocalos used and not used. If you click on this Icon appear a
properties window of the PC Card (PCMCIA).
This Window has two options,
1. Connector Statement
2. Global Configuration those measurements are what you need for the PCMCIA card
Installation.
You also must select the phrase Automatic Selection, to be able the system to recognize the type
of card, if you dont do that you wont be able to install. If you dont select this option the
system wont put the correct name of the PCMCIA card when options are showed.
Those are procedures that Windows gives to you; the procedure for the Card installation is as
follows,
1. To install the driver of the PCMCIA card insert the CD or Diskettesof National called Data
Acquisition Driver Software, in the CD Driver and run the " install nidaq "
2. Only is needed Drivers Install (Driver Files) Accept the path of national.
3. Accept the propose folder
4. The installation starts automatically
5. Complete Installation is indicated
6. Restart the Computer
Once you restart your computer,
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.

When connected the PCMCIA card is recognized


Go to My PC
Inside the Control Panel
Go to System
Go to Device Administrator
In PCMCIA Drivers are indicated
Go to the National Icon (Blue Dove) data acquisition devices
daqcard dio-24 appear, double click
The properties of daqcard dio-24 appears
Seeing the resources dont touch anything only see the installed port (In and Out Interval)
take note the first value of the port for example 1000-101f, the necessary value for the
installation of the Akonic Program is the first = 1000.

For the correct function of the akonic system, this value 1000 must be written in the ako.ini
11. Close all
---------------------------------------------------------------------------ako.ini Edit,
The ako.ini file is inside the working path of the akonic programs, normally the akonic path
Using any Editor (Edit, et) edit the ako.ini
The ako.ini could have many paragraph of any modality, eeg, ep, emg, et.
1. Look for eeg, in this case,
In the line where says pcmcia :0000 (write the number that you found in the port after
the two pints, if the value is 240 you must write 0240, you must complete the four
digits

32 Potenciales Evocados V.10.0

2. Save ako.ini with these changes


3. Started the akonic software in the Setup indicates you the port being used by the akonic
software, if you did edit ok this in the ako.ini it must be showed in the setup. In this case the
240h number appear the letter h that is showed at the end indicates that the number is
hexadecimal.
---------------------------------------------------------------------------Properties of Akonic Software correction,
Once installed the akonic software, select the icon with a mouse click
With the right key of the mouse open the properties of the icon
In the Program folder,
- Select Close on exit
4. In the Memory folder,
- In Initial Environment put 4096
5. In the Screen folder
- Enable complete screen
6. In Miscellany folder
- Deselect the Screen saber
- Select the two properties of the mouse
- Deselect that windows use the print screen key
-------------------------------------------------------------------------------------------------------------Once,
1.
2.
3.

1.
2.
3.
4.
5.

Installed the akonic program


Corrected the icon of akonic properties
Installed the pcmcia card
Edited the ako.ini with the port value of the card in the computer
Getting ON and connected everything of the akonic system

You are able to proceed !!!!

33 Potenciales Evocados V10.0

SETTING UP YOUR BIO-PC SYSTEM: INTERNAL CONNECTION

To install Akonic Interface


Before installing the Akonic interface, please take a moment to review this information.
Failure to properlly install and integrate the Akonic Interface may impact your Warranty coverage.
To ensure a smooth installation, pay particular attention to the following steps:

1- With the power off, remove the computer cover.


2- Install the Akonic Interface card firmly into its des ignated expansion card slot.
Care should be taken to press it evenly and snugly into the slot.
3- Once you are certain the Akonic interface is installed properly, secure it with the bracket screw.
Note: The position of Akonic Interface is, generally, in the last slot.
4- Secure the secondary connector of interface withe the bracket screw close to Akonic Inteface.
5- Secure the computer cover, before to chek the earth connection.
Atention!!!: These installation instructions presume that you are familiar with your computer system.
You may need to seek assistance from your dealer.

34 Potenciales Evocados V.10.0

INTERFACE INSTALLATION

In the second groove cover of the interface board you can see:
3 female connectors
1 male connector
1234-

(female) 6.5mm and it is used for an external footswitch


(female) RCA external 2 input
(female) JACK PLUG audio output
(male) photo stimulator connection

In the same groove cover remains free 1 cable with 2 connectors:


- 1 female
- 1 male
Both are used for PC connections. These cables are: red, black and yellow.
Find one free connector of these and plug it to the cables in the groove cover.
It is possible one of these connectors remains free, in this case it will not be used.

35 Potenciales Evocados V10.0

THIRTH GROOVE COVER

Before you remove the computer cover, touch a grounded metal surface a dischange static
electrcity.
To install Stimulus card:
123456-

Turn off your computer, and diconnect the power cable


Disconnect the monitor and other devices thet are connected to the computer.
Remove the computer cover
Locate a 16-bit ISA slot to accommodate the stimulus card.
Remove the slot cover from your computer - save the screw to secure the stimulus card
Align the striped tabs on the bottom of the stimulus card with the 16 bit ISA slots, and firmly press
the card into place until it fits snugly.
7- Use the screw from the slot cover to secure the stimulus card to the frame of the computer-securing
the card assesures proper grounding.
8- Replace the computer cover.
9- Connect your earphones, pattern devices, or other devices to the back panel of the stimulus card.
10- Reconnect the power cable, monitor and other devices.

36 Potenciales Evocados V.10.0

SETTING UP YOUR BIO-PC SYSTEM: EXTERNAL CONNECTION


To ensure a smooth installation, pay particular attention to the following steps:

37 Potenciales Evocados V10.0

SETTING UP YOUR BIO-PC SYSTEM

OPCIONAL
FICHA
ELECTROCAP

INPUT
Evoked Potential

INPUT
EEG

INPUT
EMG

INPUT
POLYSOMNOGRAPHY

INTERFACE
DB37H

38 Potenciales Evocados V.10.0

4 Channel Headbox

AKONIC
1

4
Emg 1

AKONIC S.A.
Av Mosconi 2886 (1419) Buenos Aires
Fax : 541 573 2794
Model :
Tel : 541 572 2593
Serial :
Made in Argentina
: 541 574 2355

Emg 2

39 Potenciales Evocados V10.0

Stand for Headbox Adapter

Headbox to Stand connection

Side

To Headbox

To Headbox Stand

To stand

Stand to Headbox connection

40 Potenciales Evocados V.10.0

TROUBLESHOOTING
System
Trouble

Corrective Action

Software does not work


Bad Software function
You can not take the register

-Check the keylock in the Port.


-Check the Headbox connection
-Check the equipment ground
-Keep the headbox and system away
-Keep the headbox and monitor away
from the wall
-Ground near to metallic elements
-Check the patient ground.
-Install the isolation transformer
-Change the power line
Check that you have installed the
latest Driver Version

Printer is not ready


Windows is a registered trademark of Microsoft Corp.

Photic Stimulator
Trouble

Corrective Action

Photic Does not work

-Check the power line.


-Check the fuse
-Check the lamp connector
-Check the switch connection
(Internal/External)
-Check high frequency
-Check the temperature of the lamp
-Check the lamp connector
-Replace the photic lamp
-Check the external switch
-Check 110-220 Volts

Flash losed

Low flash power


Continuous flash
Overheating