Calibrations
CBL Calibrations
Prior to running the tool in to the well calibrations
must be performed!
Calibrations are divided into two major categories
Surface
Master Calibrations
Pre-Job
Post Job
In Hole
Normalization
CBL Calibrations
Gamma Ray
Calibrate gamma ray using standard Thorium Blanket.
Acoustics
Each tool will have a series of calibrations that need to be
performed prior to use.
Some tools have more calibration sets than others
Minimum calibration should be;
Pipe Zero
Pipe Amplitude
CBL Calibrations
Acoustic Calibrations cont.
Calibrations should be performed in a calibration fixture or in
a test well.
Calibrations performed in a clients well are normalizations.
Perform CBL Air calibrations with tool suspended in air and
not making contact with any object.
CBL GR Calibrations
Gamma Ray
Calibrate gamma ray using standard Thorium Blanket
Same calibration standards are required as for all NGRT
CBL Calibrations
Log down with gates set at expected values
Look for calibration location (should be highest
amplitude on log)
Preferably with 500 PSI hydrostatic pressure and
single string of casing
CBL CALIBRATIONS
CBL CALIBRATIONS
Once location is found take note of raw amplitude
value
Turn around, pull in to position, watching raw sensor
outputs to ensure correct calibration location
After Calibration
Once calibrations is performed Gate and gain settings
should not be adjusted
Log 200 of all sensors through calibration location
Perform proper LQA for free pipe