Escolar Documentos
Profissional Documentos
Cultura Documentos
Flowmeter series FM
Version FM1, FM10
FM10
Application
Description
The M&C flowmeters FM 1 and FM 10 consist of a glass tube (the inside is conical
and becomes enlarged in the upper sector) in which the floating ball can move.
The substance flows through the tube from the below part to the upper part so that
the floating ball is raised. It arises a gap between the tube wall and the substance
and creates an equilibrium. Each level of the floating ball corresponds to a certain
flow rate which is shown on the scale.
The sealing and connecting parts are constructively reduced and consist only of
in- and outlet connection. All medium contacted parts are made out of PTFE and
glass. The flowmeter is configurated in such a way that it can easily be dismantled
and cleaned.
The flowmeter FM 10 is equipped with a needle valve in the inlet which enables you
to adjust the flow rate with high precision.
1/3
For automatic flow control the control unit types FA... are used.
Please see separate data sheets 5-6.10.1 / 5-6.10.2 / 5-6.10.3 / 5-6.15.
5-6.1.1
M&C 09.96 / 06.01 Technical details are subject to change. Copying, including extracts, not permitted unless authorized by M&C.
M&C Products Analysentechnik GmbH Rehhecke 79 40885 Ratingen Germany Tel.: +49 (0) 21 02 / 9 35-0 Fax: / 9 35-111 E-mail: info@muc-products.de www.muc-products.de
5
Dimensions
Flowmeter FM 1
GL18
4,5
4,5
7,5
25
100
100
250
7,5
320
25
145
75
GL1/4"
8
GL18
40
40
8
26
26
43
max. 87
Dimensions in mm
2/3
5-6.1.1
M&C 09.96 / 06.01 Technical details are subject to change. Copying, including extracts, not permitted unless authorized by M&C.
M&C Products Analysentechnik GmbH Rehhecke 79 40885 Ratingen Germany Tel.: +49 (0) 21 02 / 9 35-0 Fax: / 9 35-111 E-mail: info@muc-products.de www.muc-products.de
5
Technical Data
Flowmeter
Needle valve in the inlet
Standard- maximum measures
adjusted on 1,2 bar; 20 C ..l/hr air:
..l/hr water:
Measures ....l/h air at 20 C; 1,2 bar
Scale of measurement
Status of precision
Scale
Pressure
Sample temperature, max.
Ambient temperature, max.
Storage temperature
Connections, INLET
Connections, OUTLET
Method and position of mounting
Materials of sample contacting parts
Dimensions, h x w x d
Weight
* Standard (alternatively)
FM1
FM1-H
NO
NO
types on stock = ___
16 40 60 100 250 500 800
2,5 5 12 25 40 60 100r
min. 0,8-8 l/hr; max. 120-1200 l/hr is possible
10 : 1
2,5 %
Length of the scale 100 mm, adjusted in l/hr
max. 4 bar g
+80 C
+150 C
+60 C
+150 C
-25 C to +80 C
GL18-6* (8, 10)
GL18-6* (8, 10)
Wall mounting / vertical
Borosilicate glass, PTFE
250 x 40 x 40 mm
130 g
280 g
FM10
YES
FM10-H
YES
+80 C
+60 C
+150 C
+150 C
440 g
FM1
Part-No.
09F1000
09F1010
09F1020
09F1030
FM1-H
Part-No.
09F1100
FM10
Part-No.
09F1500
09F1510
09F1520
09F1530
FM10-H
Part-No.
09F1550
3/3
5-6.1.1
M&C 09.96 / 06.01 Technical details are subject to change. Copying, including extracts, not permitted unless authorized by M&C.
M&C Products Analysentechnik GmbH Rehhecke 79 40885 Ratingen Germany Tel.: +49 (0) 21 02 / 9 35-0 Fax: / 9 35-111 E-mail: info@muc-products.de www.muc-products.de