Escolar Documentos
Profissional Documentos
Cultura Documentos
SERVICE
MANUAL
Systems:
Color Therapy System
(Serial Numbers- 1000 and above)
Optional Accessories:
Channel 3/4 Electrotherapy Module
(Serial Numbers- 1000 and above)
sEMG Module
(Serial Numbers- 1000 and above)
Laser Module
(Serial Numbers- 1000 and above)
TABLE of CONTENTS
Foreword . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1- Safety Precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2- Theory of Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4
3- Nomenclature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-16
Intelect Advanced Therapy System. . . . . . . . . . . . . . . . . . . . . . . . . . 5
Intelect Advanced 2 Channel Combination Therapy System . . . . . . 6
Intelect Advanced 2 Channel Electrotherapy System . . . . . . . . . . . . 7
Channel 3/4 Electrotherapy Module . . . . . . . . . . . . . . . . . . . . . . . . . 8
NiMH Battery Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Laser Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Laser Applicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
sEMG and sEMG + Electrical Stimulation Module . . . . . . . . . . . . . 12
Therapy System Cart. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Vacuum Electrode Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Operator Remote Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Hardware and Software Symbol Definitions . . . . . . . . . . . . . . . . . . 16
4- Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17-26
Intelect Advanced Therapy System. . . . . . . . . . . . . . . . . . . . . . . . . 17
Electrotherapy Waveform Specifications. . . . . . . . . . . . . . . . . . .18-25
IFC Traditional (4p) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
TENS- Asymmetrical Biphasic. . . . . . . . . . . . . . . . . . . . . . . . . 18
TENS- Symmetrical Biphasic . . . . . . . . . . . . . . . . . . . . . . . . . 19
TENS- Alternating Rectangular . . . . . . . . . . . . . . . . . . . . . . . . 19
TENS- Monophasic Rectangular . . . . . . . . . . . . . . . . . . . . . . . 19
High Voltage Pulsed Current (HVPC) . . . . . . . . . . . . . . . . . . . 20
VMS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Diadynamic Waveforms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
IFC Premodulated (2p) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Russian . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Microcurrent. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
VMS Burst . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Monophasic Rectangular Pulsed. . . . . . . . . . . . . . . . . . . . . . . 23
Monophasic Triangular Pulsed . . . . . . . . . . . . . . . . . . . . . . . . 23
Galvanic Continuous . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Galvanic Interrupted . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Trbert (Ultrareiz) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Surged Monophasic Rectangular . . . . . . . . . . . . . . . . . . . . . . 25
Surged Monophasic Triangular . . . . . . . . . . . . . . . . . . . . . . . . 25
Intelect Advanced Therapy System Ultrasound . . . . . . . . . . . . . . . 26
Intelect Advanced Therapy System Laser. . . . . . . . . . . . . . . . . . . . 26
5- Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27-57
Intelect Advanced Software Error Messages . . . . . . . . . . . . . . .27-35
Intelect Advanced Therapy System Testing. . . . . . . . . . . . . . . . . . . 36
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Special Tools, Fixture & Materials Required . . . . . . . . . . . . . . 36
Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Visual Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Leakage Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Unit Startup and Fan Testing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Electrical Stimulator Tests- System Setup. . . . . . . . . . . . . . . . . . . .
VMS Mode Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Interferential Mode Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Premodulated Mode Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Russian Mode Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Microcurrent Mode Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
High Voltage Pulsed Current (HVPC) Mode Test . . . . . . . . . .
Microcurrent Probe Mode Test . . . . . . . . . . . . . . . . . . . . . . . .
Ultrasound Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Ultrasound Applicator Output Test. . . . . . . . . . . . . . . . . . . . . . . . . .
38
38
39
39
40
41
42
43
44
45
2005 Encore Medical Corporation or its affiliates, Austin, Texas, USA. Any use of editorial, pictorial or layout composition of this publication without express written consent from the
Chattanooga Group of Encore Medical, L.P. is strictly prohibited. This publication was written, illustrated and prepared for print by the Chattanooga Group of Encore Medical, L.P.
FOREWORD
Read, understand, and follow the Safety Precautions and all other information contained in this manual.
This manual contains the necessary safety and field service information for those field service technicians,
certified by Chattanooga Group, to perform field service on the Intelect Advanced Therapy Systems, modules,
and accessories.
At the time of publication, the information contained herein was current and up-to-date. However, due to
continual technological improvements and increased clinical knowledge in the field of electrotherapy, as well
as Chattanooga Groups policy of continual improvement, Chattanooga Group reserves the right to make
periodic changes and improvements to their equipment and documentation without any obligation on the part
of Chattanooga Group.
It is the sole responsibility for certified field technicians to stay informed and trained in the latest technology
utilized in the Intelect Advanced Therapy Systems by Chattanooga Group. From time to time, as significant
improvements are incorporated, service bulletins will be produced and made available on our web site (www.
chattgroup.com) in lieu of reprinting a complete manual prematurely. These service bulletins will provide
updated service information and technological improvements to the Intelect Advanced Therapy Systems for
use by certified service technicians.
Certified Service Technician Definitions;
1. Level I- Those certified field service technicians that have successfully completed the minimal training
required by Chattanooga Group in basic service techniques.
2. Level II- Those certified field service technicians that have successfully completed Level I Training as
well as Level II training as required to perform specific troubleshooting and repair techniques and
procedures.
3. Level III- Those certified field service technicians that have successfully completed Levels I & II Training
as well as Level III Advanced Training as required to perform all necessary troubleshooting and
repair techniques. The technician having successfully completed the three levels of training and
coupled with experience should have the ability to train other technicians in Level I and Level II
training with the necessary training materials from Chattanooga Group.
4. Temporary- Chattanooga Group, at its discretion and based on known experience of the technician, may
grant a Temporary Certification to a field technician for particular troubleshooting and repair
of a specific system requiring immediate attention. This Temporary Certification in no fashion
acknowledges the training level of a technician as defined above. This Temporary
Certification is utilized only in unique situations for a specific unit for a specific service
technique only and is documented as such.
Due to the complex nature of the technology utilized by Chattanooga Group, the recommended
troubleshooting techniques are to determine Bad Board and board replacement only. No board component
level troubleshooting is recommended, nor will information or parts be supplied by Chattanooga Group.
Any board component level troubleshooting performed will be at the sole risk and liability of the certified field
service technician performing such troubleshooting techniques. Performance of such techniques may render
the warranty null and void.
This equipment is to be used only under the prescription and supervision of a licensed medical practitioner.
1- SAFETY PRECAUTIONS
A. CAUTION
CAUTION
Text with a CAUTION indicator will explain
possible safety infractions that have the potential to
cause minor to moderate injury or damage to
equipment.
B. WARNING
WARNING
Text with a WARNING indicator will explain
possible safety infractions that will potentially cause
serious injury and equipment damage.
CAUTION
Read, understand, and practice the precautionary and
operating instructions. Know the limitations and hazards
associated with using any electrical stimulation or ultrasound
device. Observe the precautionary and operational decals
placed on the unit.
C. DANGER
DANGER
Text with a DANGER indicator will explain
possible safety infractions that are imminently
hazardous situations that would result in death or
serious injury.
D. DANGEROUS VOLTAGE
Text with a Dangerous Voltage indicator serves to
inform the user of possible hazards resulting in the
electrical charge delivered to the patient in certain
treatment configurations of TENS waveforms.
F. BIOHAZARDOUS MATERIAL
Areas marked with the Biohazard Symbol
indicate components of the system that
require personal care and proper disposal of
drainage material according to national, state,
and local rules and regulations.
G. NOTE:
2- THEORY of OPERATION
2.1 Overview
The Intelect Advanced Therapy Systems are comprised of several PC board assemblies housed within a common enclosure. These
assemblies each support a distinct function in the product. The basic elements are User Interface, Control Board, Stim Board,
Ultrasound Board, Ultrasound Applicator, and Power Supply Circuits.
When a Module (Channel 3/4 Electrotherapy, NiMH Battery, Laser, or sEMG) is installed, the Control Board software automatically
recognizes that a Module has been installed and prompts the installer to perform certain tasks, for verification of Module installed, to
make the respective Module fully functional. No additional software installation is required as the Therapy System contains all
necessary software to accommodate any Module installation.
2.2 Power Supply Circuits
A universal input 100 Watt power supply provides the Control Board and Stim board of the system with 24 volts DC. The supply is
connected to the mains at all times when the cord is attached. The 24V supply is regulated locally at each PC board as required. On
Combination Systems, a separate universal 75 Watt Power Supply provides 24 volts DC to the Ultrasound PC Board. The 24 volt
DC power is regulated at the board, as required.
2.3 Control Board
The Control Board serves just as its name implies. It controls the operation of the stim board, ultrasound board, user interface,
optional modules, and accessories. The control board communicates to the stim boards and ultrasound board through a proprietary
bus. The control board drives the display. The control board reads the menu buttons. The control board also reads the amplitude
and the contrast control on the monochromatic systems. The control board reads and manages the Multimedia (MMC) Card, Patient
Data Card, and sEMG Data Card. Sound output is generated by the control board and routed to an internal speaker. The control
board reads the Patient Interrupt Switch and the Operator Remote Control (used to administer Manual Stimulation Therapy).
2.4 Stim Board
The Stim Board creates all muscle stimulation output. Communications to the Stim Board is via a proprietary bus. A
Processor on the Stim Board acts on messages passed to it by the Control Board to set up waveforms and adjust
output amplitude. Information can likewise be passed from the Stim Board back to the Control Board for monitoring Current,
Microcurrent Probe Contact Quality indication, etc. If the Stim Board does not respond as expected to a command from the
Control Board, output is stopped and an Error Message is generated.
2.5 Ultrasound Board and Applicator
The Ultrasound Board generates the 1 or 3.3 MHz output to drive the Sound Head of the Applicator. The Ultrasound Board is
accessed through the proprietary bus by the Control Board. It can provide current and voltage information about the ultrasound
output of the board. The calibration data for the Sound Head is passed through the Ultrasound Board from the Applicator to the
Control Board. By storing the calibration data in the Applicator, there is no calibration necessary for the Ultrasound Board and any
calibrated Chattanooga Group Intelect Advanced or Intelect Mobile Ultrasound Applicator can be connected and operated to
provide accurate coupling and output.
2.6 User Interface and Accessories
The LCD display panel provides the operator visible feedback in the way of menu choices. Pressing of the menu buttons makes
selections from the menus. The control board interprets these user inputs and responds accordingly. Audible feedback is given as
well for events such as key presses and end of treatment.
The control board accesses the Patient Data Card, sEMG Data Card and MMC Card via an on board Reader/Writer Interface. The
voltage necessary to operate the reader is provided by the 100 Watt Power Supply and is regulated by the Control Board.
Channel 3/4 Electrotherapy Module
The Channel 3/4 Electrotherapy Module creates all muscle stimulation output for Channels 3 and 4. The Channel 3/4
Electrotherapy Module is interfaced with the System via a ribbon cable which supplies power and facilitates communication
between the stim board and control board of the system. All waveforms available to channels 1 and 2 are available to
channels 3 and 4 via the system software. No additional software is required for full functionality of the module.
2- THEORY of OPERATION
3- NOMENCLATURE
FIGURE 3.1
3- NOMENCLATURE
1
7
2
6
3
8
5
4
9
12
11
10
20
18
16
14
15
17
19
13
FIGURE 3.2
1. Screen Contrast Control (Not functional on Color Systems)
2. System Power On/Off Switch
3. Technical Maintenance Port
4. Fuses
5. Main Power Cord
6. Rear Access Panel
7. Two Channel Combo System
8. Ultrasound Applicator (5cm2 shown) Combo Systems Only
9. User Interface (Screen and Buttons)
10. Front Access Panel
11. Patient Data Card and sEMG Data Card access port.
3- NOMENCLATURE
1
7
2
6
3
5
4
8
11
10
9
18
17
15
13
14
16
12
FIGURE 3.3
1. Screen Contrast Control (Not functional on Color Systems)
2. System Power On/Off Switch
3. Technical Maintenance Port
4. Fuses
5. Main Power Cord
6. Rear Access Panel
7. Two Channel Electrotherapy System
8. User Interface (Screen and Buttons)
9. Front Access Panel
10. Patient Data Card and sEMG Data Card access port.
11. Multimedia Card (MMC) access port.
3- NOMENCLATURE
11
1
5
10
FIGURE 3.4
1. Channel 3/4 Electrotherapy Module
2. Extended Front Access Panel
3. Module to System Mounting Holes
4. Module to System Feet Alignment Indents
5. Power Cord Routing Port
6. Optional Channel 3/4 Operator Remote Control Connector
7. Patient Interrupt Switch Connector
8. Channel 3 Lead Wire Connector
9. Channel 4 Lead Wire Connector
10. Microcurrent Probe Connector
11. Module to System Header
NOTE:
The Intelect Advanced Channel 3/4 Electrotherapy Module is not
operable unless it is properly connected to the Intelect Advanced
Therapy System.
3- NOMENCLATURE
6
5
FIGURE 3.5
1.
2.
3.
4.
5.
6.
NOTE:
The Intelect Advanced NiMH Battery Module is not operable
unless it is properly connected to the Intelect Advanced Therapy
System.
3- NOMENCLATURE
6
5
10
FIGURE 3.6
1. Laser Module
2. Extended Front Access Panel
3. Module to System Mounting Holes
4. Module to System Feet Alignment Indents
5. Power Cord Routing Port
6. Module to System Header
7. Patient Interrupt Switch
8. Therapy Room Door Lockout Jack
9. Point Locator (for use with Single Probe Laser Applicators)
10. Laser Applicator
NOTE:
The Intelect Advanced Laser Module is not operable unless it is
properly connected to the Intelect Advanced Therapy System.
DANGER
10
3- NOMENCLATURE
1
2
6
6
1
4
2
1
FIGURE 3.7
1. Laser On LED
2. Laser Applicator On/Off Button
3. Single Diode Applicator Housing
4. LED Cluster Applicator Housing
5. Laser Cluster Applicator Housing
6. Laser Aperature Lens
7. Laser Aperature
DANGER
NOTE:
The Intelect Advanced Laser Applicators are not operable unless
they are connected to the Intelect Advanced Therapy System via
the Laser Module.
11
3- NOMENCLATURE
4
3
FIGURE 3.8
1.
2.
3.
4.
5.
6.
NOTE:
The Intelect Advanced sEMG Module is not operable unless it is
connected to the Intelect Advanced Therapy System.
12
3- NOMENCLATURE
FIGURE 3.9
1. Cart Top
2. System to Cart Retaining Screw (4)
3. Storage Bins (6)
(5 Bins with the Intelect Advanced Vacuum Electrode
Module installed)
4. Cart Rear Swivel Casters
5. Cart Base
6. Cart Front Swivel, Locking Casters
7. Cart Bottom Access Plate
8. Front and Rear Cart Extrusions
13
3- NOMENCLATURE
13
15
12
11
9
10
FIGURE 3.10
1. Reservoir Exhaust/Overflow Hose (Routes to bottom of
Therapy System Cart)
2. Vacuum Module to Therapy System Lead Wires
3. Vacuum LED Indicators
4. Vacuum Mode Selector Button
5. Vacuum Intensity Knob
6. Channels 1 and 2 Lead Wire Connectors
7. Reservoir Drain Hose (Routes into Cart Storage Bin below
Vacuum)
8. Channel 1 and 2 Vacuum Lead Hose Connectors
9. Vacuum Main Power Switch
14
3- NOMENCLATURE
7
1
6
2*
15
Pause
Treatment
M Manual
Stimulation
3- NOMENCLATURE
Clinical
Resources
Increase
Intensity
Back
On/Off
Switch
Data
Port
Multi-Media and
Patient Card
Stop
Treatment
Pause
Treatment
Start
Treatment
Decrease
Intensity
Channel 1/2
Operator
Remote
Control
(Optional)
Charge Level
Patient|
Interrupt
Switch
Patient
Interrupt
Switch
Channel 1
Lead Wires
Channel 3
Lead Wires
Channel 2
Lead Wires
Channel 4
Lead Wires
Ultrasound
Applicator
Move DOWN
Move RIGHT
Microcurrent
Probe
Channel 3/4
Operator
Remote|
Control
(Optional)
Therapy
Intensity
Control
Move UP
Battery
Charging
Microcurrent
Probe
Home
Pause
Treatment
Manual
Stimulation
Vacuum
Hoses
& Cups
Continuous
Vacuum
Lead
Wires
1.65 Sec On
0.35 Sec Off
Vacuum
Vacuum
On/Off
Switch
0.75 Sec On
0.25 Off
Vacuum
Move LEFT
Patient
Interrupt
Switch
Accept and
Return
Treatment
Room Interlock
Connector
Do not accept
and Return
16
Point
Locator
Laser
Applicator
4- SPECIFICATIONS
HEIGHT
WITH SYSTEM & SYSTEM WITH MODULE
HEIGHT
CART ONLY
DEPTH
WIDTH
FIGURE 4.1
Dimensions
Height
Cart Only . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85.7 cm (33.75)
With System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .108 cm (42.50)
With System and Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112.4 cm (44.25)
Width . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43.2 cm (17)
Depth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41.3 cm (16.25)
Power (Combo and Electrotherapy Systems)
Input. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 - 240V - 175VA, 50/60 Hz
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . +24, 8.3A
Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Two 6.3A Time Lag (Part Number 71772)
Electrical Class . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .CLASS I
Electrical Type
Ultrasound (Combination Systems Only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TYPE B
Electrotherapy, sEMG, Vacuum, Channel 3/4 Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . TYPE BF
17
4- SPECIFICATIONS
NOTE:
All waveforms except High Voltage Pulsed Current (HVPC) of the
Intelect Advanced Therapy System have been designed with a
200mA current limit.
VMS, VMS Burst, and all TENS waveform output intensities
are measured, specified, and listed to peak, not peak to peak.
All Waveforms are available on all channels.
FIGURE 4.2
FIGURE 4.3
DANGER
Stimulus delivered by the TENS waveforms of this
device, in certain configurations, will deliver a charge of
25 microcoulombs (C) or greater per pulse and may be
sufficient to cause electrocution. Electrical current of this
magnitude must not flow through the thorax because it may
cause a cardiac arrhythmia.
18
4- SPECIFICATIONS
FIGURE 4.4
FIGURE 4.5
FIGURE 4.6
*CC= Constant Current
CV= Constant Voltage
19
4- SPECIFICATIONS
FIGURE 4.7
FIGURE 4.8
20
4- SPECIFICATIONS
FIGURE 4.10
FIGURE 4.11
FIGURE 4.12
FIGURE 4.9
FIGURE 4.13
21
4- SPECIFICATIONS
FIGURE 4.14
FIGURE 4.15
FIGURE 4.16
22
4- SPECIFICATIONS
FIGURE 4.17
FIGURE 4.18
FIGURE 4.19
23
4- SPECIFICATIONS
FIGURE 4.20
FIGURE 4.22
24
4- SPECIFICATIONS
25
4- SPECIFICATIONS
Ultrasound
Frequency . . . . . . . . . . . . . 1 MHz, 5%; 3.3 MHz, 5%
Duty Cycles . . . . . . . . . 10%, 20%, 50%, and Continuous
Pulse Frequency . . . . . . . . . . . . . . 16 Hz, 48 Hz, 100 Hz
Pulse Duration . . . . . . . . . 1 mSec, 20%; 2 mSec, 20%
5 mSec, 20%
Output Power
10 cm2 Crystal . . . . . . . . . . . . . . . . . 0-20 Watts at 1 MHz,
0-10 Watts at 3.3 MHz
5 cm2 Crystal . . . . . . . . . . . . . 0-10 Watts, 1 and 3.3 MHz
2 cm2 Crystal . . . . . . . . . . . . . . 0-4 Watts, 1 and 3.3 MHz
1 cm2 Crystal . . . . . . . . . . . . . . . .0-2 Watts 3.3 MHz Only
Amplitude . . . . . . . . .0 to 2.5 w/cm2 in continuous mode,
0-3 w/cm2 in pulsed modes
Output accuracy . . . . . . 20% above 10% of maximum
Temporal Peak to Average Ratios:
2:1, 20%, at 50% Duty Cycle
5:1, 20%, at 20% Duty Cycle
9:1, 20%, at 10% Duty Cycle
Beam Nonuniformity Ratio. . . . . . . . . . 5.0 : 1 maximum
Beam Type . . . . . . . . . . . . . . . . . . . . . . . . . . . Collimating
Effective Radiating Areas
10 cm2 Crystal - 8.5 cm2, 1.5
5 cm2 Crystal - 4.0 cm2, 1.0
2 cm2 Crystal - 1.8 cm2, +0.2/-0.4
1 cm2 Crystal - 0.8 cm2, +0.2/-0.4
Treatment Time . . . . . . . . . . . . . . . . . . . . . . 1-30 Minutes
Laser Applicators
Single Diode GaAs Laser Applicators
670 nm, 10 mW LED
Patient Contact Spot Size . . . . . . . . . . . . . . . . .0.16 cm2
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 400 g
850 nm, 100 mW
Patient Contact Spot Size . . . . . . . . . . . . . . . . .0.07 cm2
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 400 g
850 nm, 200 mw
Patient Contact Spot Size . . . . . . . . . . . . . . . . .0.16 cm2
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 400 g
Cluster GaAs Laser Applicators
690 mW
Elements . . . . . . . . . . . . 5- 50mW, 850nm GaAs Lasers
12- 10 mW, 670 nm LEDs
8- 15 mW, 950 nm LEDs
8- 25 mW, 880 nm LEDs
26
5- TROUBLESHOOTING
TYPE
MESSAGE
PROBABLE CAUSES
POSSIBLE REMEDIES
100
Warning
Overcurrent
A. Check Electrodes and Lead Wires. Make certain Lead Wires are
not damaged and are properly connected to the system. Make
certain Lead Wires are properly connected to the Electrodes and
that electrodes are not damaged and are making proper contact
with treatment area.
B. Replace Lead Wires and Electrodes
101
Warning
A. Check Electrodes and Lead Wires. Make certain Lead Wires are
not damaged and are properly connected to the system. Make
certain Lead Wires are properly connected to the Electrodes and
that electrodes are not damaged and are making proper contact
with treatment area.
B. Replace Lead Wires and Electrodes
A. Make certain Electrodes are making proper contact with the
treatment area.
102
Warning
103
Warning
Blank Patient ID
Properly enter Patient ID. Refer to Patient Data Card section (page
45-49) of Therapy System User Manual.
104
Warning
106
Warning
107
Warning
108
Warning
109
110
111
Warning
Warning
Warning
Warning
Ultrasound Applicator
disconnected from system during
treatment session
Warning
112
113
27
5- TROUBLESHOOTING
TYPE
MESSAGE
PROBABLE CAUSES
POSSIBLE REMEDIES
114
Warning
115
Warning
116
117
Warning
Warning
A. Properly insert the Patient Data Card into the system port. Refer
to Therapy System User Manual, page 45, for new Patient Data
Card or page 50 for existing Patient Data Card.
B. Attempt to use a known good Patient Data Card.
C. Make certain a Patient Data Card and not an sEMG Data Card is
being used,
D. If problem persists, replace Control Board
118
Warning
119
120
121
Warning
Warning
Warning
A. Use a Patient Data Card with proper treatment data for the system.
B. Properly insert a Patient Data Card.
C. Insert a known good Patient Data Card.
D. If problem persists, insert a known good Patient Data Card. If
problem continues, replace Control Board.
E. Insert a known good Patient Data Card.
123
Warning
124
Warning
125
Warning
126
Warning
127
Warning
128
Warning
129
Warning
A. sEMG Data Card faulty. Insert a known good sEMG Data Card.
B. If problem persists, replace Control Board.
28
5- TROUBLESHOOTING
130
TYPE
MESSAGE
Warning
PROBABLE CAUSES
Another treatment is running
while attempting to set up
and perform a Laser Therapy
treatment.
POSSIBLE REMEDIES
A. Allow existing treatment to complete before starting Laser Therapy.
B. If no other treatment is running, reset Therapy System by turning
main power switch Off and On.
A. Make certain Treatment Room Door is completely closed.
B. Make certain the Lockout cable is connected to the system.
131
Warning
132
Warning
133
Warning
134
Warning
135
Warning
136
Warning
137
Warning
138
Warning
139
Warning
140
Warning
141
Warning
142
Warning
Warning
29
5- TROUBLESHOOTING
TYPE
MESSAGE
PROBABLE CAUSES
POSSIBLE REMEDIES
A. Connect correct Laser Applicator to the system.
144
Warning
145
Warning
Properly insert a Patient Data Card, set up and perform the treatment
and, save data to Patient Data Card.
Error
201
Error
202
Error
203
Error
204
Error
205
Error
206
Error
207
Error
208
Error
209
Error
Error
211
Error
212
Error
213
Error
30
5- TROUBLESHOOTING
TYPE
MESSAGE
PROBABLE CAUSES
POSSIBLE REMEDIES
Error
215
Error
216
Error
217
Error
218
Error
219
Error
220
Error
221
Error
222
Error
223
Error
224
Error
225
Error
226
Error
227
Error
228
Error
229
Error
230
Error
231
Error
NOTE:
Errors 219-231: after replacing the PC Board, and if the problem persists, send the Therapy System to the Factory for Service.
31
5- TROUBLESHOOTING
TYPE
MESSAGE
PROBABLE CAUSES
POSSIBLE REMEDIES
233
Error
234
Error
235
Error
236
Error
237
Error
238
Error
239
Error
240
Error
NOTE:
Errors 233-240: after replacing the PC Board, and if the problem persists, send the Therapy System to the Factory for Service.
CRITICAL ERRORS DEMANDING TECHNICAL SERVICE (300-399)
A. On Therapy System, make certain internal Ribbon Cable is seated
on Stim PC Board and Control Board.
300
Critical Error
301
Critical Error
302
Critical Error
32
5- TROUBLESHOOTING
TYPE
MESSAGE
PROBABLE CAUSES
POSSIBLE REMEDIES
A. On Therapy System, make certain internal Ribbon Cable is seated
on Stim PC Board and Control Board.
303
Critical Error
304
Critical Error
305
Critical Error
306
Critical Error
307
Critical Error
308
Critical Error
309
Critical Error
33
5- TROUBLESHOOTING
310
TYPE
MESSAGE
Critical Error
PROBABLE CAUSES
POSSIBLE REMEDIES
A. Make certain Ultrasound PC Board is completely seated on internal
Header and Ribbon Cable is seated on Stim PC Board and Control
Board.
B. If problem persists, replace Ultrasound PC Board.
C. Replace Control Board
311
Critical Error
312
Critical Error
313
Critical Error
314
Critical Error
315
Critical Error
316
Critical Error
317
Critical Error
318
Critical Error
319
Critical Error
34
5- TROUBLESHOOTING
PROBABLE CAUSES
POSSIBLE REMEDIES
2. Bad Module.
35
5- TROUBLESHOOTING
ppm OF Oxygen
Less than 5 ppm
Tap Water
Up to 35 ppm
Distilled Water
Up to 20 ppm
36
5- TROUBLESHOOTING
WARNING
UNIT FAILING DIELECTRIC WITHSTAND OR LEAKAGE TESTS
COULD INDICATE SERIOUS INTERNAL PROBLEMS.
DO NOT PLACE UNIT BACK INTO SERVICE! SEND UNIT TO
FACTORY FOR REPAIR! DO NOT ATTEMPT TO REPAIR!
5.5 Unit Startup and Fan Testing
A. Test
1. Place unit face up on work surface.
2. Connect power cord to unit and plug into proper
power receptacle.
3. Turn system on. Press the Electrotherapy Button.
Press the IFC Traditional (4p) Button.
4. Place hand at the back of system, at Mains Power
Switch, to verify fan is blowing out. See Figure 5.2.
B. Test Results
1. Unit will not Start= Unit Failed Test
a) Bad Fuse
b) Possible bad Main Power Switch.
c) Possible bad Power Supply.
d) Possible bad power outlet or Power Cord.
2. Home Screen does not display= Unit Failed Test.
a) Possible bad display.
b) Possible bad Control Board.
c) Possible bad Power Supply.
Visually check power LED. LED Should
illuminate Blue. Turn system off with power
switch. Power LED should illuminate Green. If
Power LED illuminates Blue with system On and
Green with system Off, the Power Supply is
good. Replace Control Board.
3. Fan not blowing outward= Unit Failed Test
a) Fan Blowing Inward.
Fan wired wrong. Rewire or replace fan.
b) Fan not blowing.
1) Possible bad Fan.
2) Possible bad Power Supply.
3) Possible bad Control Board.
FIGURE 5.1
FIGURE 5.2
37
5- TROUBLESHOOTING
FIGURE 5.3
SCOPE TO
ESTI-2
LOAD SWITCH
TO 1K
THERAPY SYSTEM TO
ESTI-2
FIGURE 5.4
FIGURE 5.5
38
5- TROUBLESHOOTING
FIGURE 5.6
FIGURE 5.7
39
5- TROUBLESHOOTING
FIGURE 5.8
FIGURE 5.9
FIGURE 5.9A
40
5- TROUBLESHOOTING
LOAD SWITCH
TO 10K Micro
FIGURE 5.10
PROPER POSITIVE (+) Microcurrent WAVEFORM
FIGURE 5.11
PROPER NEGATIVE (-) Microcurrent WAVEFORM
FIGURE 5.11A
41
5- TROUBLESHOOTING
FIGURE 5.12
PROPER POSITIVE High Volt WAVEFORM
FIGURE 5.13
AMPS MUST NOT
EXCEED 1.5
FIGURE 5.14
42
5- TROUBLESHOOTING
FIGURE 5.15
FIGURE 5.16
FIGURE 5.17
43
5- TROUBLESHOOTING
FIGURE 5.18
FIGURE 5.19
FIGURE 5.20
44
5- TROUBLESHOOTING
WARNING
USE ONLY DEGASSED WATER IN POWER METER FOR
TESTING ULTRASOUND APPLICATORS. USE OF OTHER
TYPES OF WATER WILL CAUSE FALSE TEST RESULTS.
SEE PAGE 32 FOR DEGASSED WATER RECIPES.
DO NOT AERATE WATER WHEN FILLING POWER METER
2. Place an Applicator into the Power Meter retainer.
Make certain the Sound Head is completely
submerged in the degassed water and centered
directly over the Stainless Steel Cone. See Figure
5.21.
3. Zero meter.
4. Press Ultrasound Button. Press Edit Button.
5. Press Duty Cycle Button until 100% is displayed
within the Duty Cycle icon.
6. Press Display Button until Watts appears within the
Display icon.
7. Press Start.
8. Rotate Therapy Intensity Control clockwise until the
appropriate Watts is displayed per Figure 5.22.
9. Compare Power Meter readings to Figure 5.22 to all
settings for the respective Applicator being tested as
shown in Figure 5.22.
10. Press Frequency Button until 3.3 MHz is displayed
within the Frequency icon. Repeat test and compare
readings to Figure 5.22.
FIGURE 5.21
2 cm2
5 cm2
NOTE:
The Applicator LED should constantly illuminate green during the
Applicator Output tests.
B. Ultrasound Applicator Output Test Results
1. Output ranges fall within the specified ranges as
listed in Figure 5.22.
Unit passed test.
2. Readings fall outside specified ranges of Figure 5.22.
a) Possible bad Degassed Water in Power Meter.
b) Possible use of Power Meter other than Ohmic
Instruments UPM DT 10 or DT 100 Ultrasound
Power Meter.
c) Possible bad or out of calibration Applicator.
d) Possible bad internal connection at Ultrasound
Board.
e) Check Ultrasound Board internal connections.
f) Replace Ultrasound Board.
g) Replace Control Board
10 cm2
POWER SETTING
OUTPUT RANGE
0.8 - 1.2
2
1
1.6 - 2.4
0.8 - 1.2
1.6 - 2.4
3.2 - 4.8
0.8 - 1.2
1.6 - 2.4
4.0 - 6.0
10
1
8.0 - 12.0
0.8 - 1.2
4.0 - 6.0
10
8.0 - 12.0
15*
12.0 - 18.0
20*
16.0 - 24.0
* 1 MHz Only
FIGURE 5.22
45
5- TROUBLESHOOTING
WARNING
USE ONLY DEGASSED WATER IN POWER METER FOR
TESTING ULTRASOUND APPLICATORS. USE OF OTHER
TYPES OF WATER WILL CAUSE FALSE TEST RESULTS.
SEE PAGE 32 FOR DEGASSED WATER RECIPES.
DO NOT AERATE WATER WHEN FILLING POWER METER.
2. Place an Applicator into the Power Meter retainer.
Make certain the Sound Head is completely
submerged in the degassed water and centered
directly over the Stainless Steel Cone. See
Figure 5.23.
3. Zero meter.
4. Press Ultrasound Button. Press Edit Button.
5. Press Duty Cycle Button until 100% is displayed
within the Duty Cycle icon.
6. Press Display Button until Watts appears within the
Display icon.
7. Press Start.
8. Rotate Therapy Intensity Control clockwise until the
appropriate Watts is displayed. See Figure 5.24.
9. Compare Power Meter readings to Figure 5.24 to all
settings for the respective Applicator being tested as
shown in Figure 5.24.
10. Press Frequency Button until 3.3 MHz is displayed
within the Frequency icon. Repeat test and compare
readings to Figure 5.22.
B. Ultrasound Duty Cycle Test Results
1. Duty Cycles fall within the specified ranges as listed in
Figure 5.24.
Unit passed test.
2. Readings fall outside specified ranges of Figure 5.24.
a) Possible bad degassed water in Power Meter.
b) Possible use of Power Meter other than Ohmic
Instruments UPM DT 10 or DT 100 Ultrasound
Power Meter.
c) Possible bad or out of calibration Applicator. Re-test
with known good Intelect Advanced Applicator.
d) Possible bad internal connection at Ultrasound
Board.
e) Check Ultrasound Board internal connections.
f) Replace Ultrasound Board.
g) Replace Control Board.
FIGURE 5.23
5 cm2
DUTY CYCLE
OUTPUT RANGE
10%
0.8 - 1.2
20%
1.6 - 2.4
50%
4.0 - 6.0
100% (Continuous)
8.0 - 12.0
FIGURE 5.24
46
5- TROUBLESHOOTING
FIGURE 5.25
COMBO INDICATOR
ILLUMINATED
FIGURE 5.26
FIGURE 5.27
47
5- TROUBLESHOOTING
ATTENUATOR SCHEMATIC
RED
100K 1%
RED
49.9 Ohm 1%
TO AUDIO WAVEFORM
GENERATOR
BLACK
49.9 Ohm 1%
GREEN
BLACK
100K 1%
FIGURE 5.28
SET FREQUENCY
RANGE Hz TO X1
AMPLITUDE TO
MAXIMUM
FREQUENCY
DIAL TO 100
TEST LEADS TO
SYNC PORTS
WAVEFORM
TO SINE
ATTEN TO 0
FIGURE 5.29
sEMG AND sEMG + STIM ICONS
FIGURE 5.30
48
5- TROUBLESHOOTING
FIGURE 5.31
4. Connect the Channel 1sEMG lead wires into the
Attenuator. Make certain each sEMG Lead is
connected to its respective color on the Attenuator.
See Figure 5.32.
FIGURE 5.32
5. Press the sEMG Button on Home Screen.
Channel 1 should read no more than 7. See Figure
5.33.
6. If Channel 1 reads less than 7, repeat steps 2 through
5 on all other sEMG Channels.
If any Channel being tested reads greater than 7,
replace the respective sEMG Module and re-test.
NOTE:
The reading on the Channel not being tested may vary in its
reading. This is insignificant as it is not under load.
FIGURE 5.33
49
5- TROUBLESHOOTING
FIGURE 5.34
FIGURE 5.35
FIGURE 5.35A
50
5- TROUBLESHOOTING
FIGURE 5.36
FIGURE 5.37
FIGURE 5.38
51
5- TROUBLESHOOTING
FIGURE 5.39
12. Visually Inspect individual cells of both battery packs
for leaks or ruptures. See Figure 5.40.
DANGER
NiMH batteries contain Class E corrosive materials. In the event
of battery cell rupture or leakage, handle battery module wearing
neoprene or natural rubber gloves. Contents of a ruptured or
leaking battery can cause respiratory irritation. Hypersensitivity
to nickel can cause allergic pulmonary asthma. Contents of cell
coming in contact with skin can cause skin irritation and chemical
burns.
Never, under any circumstances, open the battery cells.
Should an individual cell from a battery become disassembled,
spontaneous combustion of the negative electrode is possible.
There can be a delay between exposure to air and spontaneous
combustion.
FIGURE 5.40
52
5- TROUBLESHOOTING
FIGURE 5.41
FIGURE 5.42
5. Rotate Vacuum Intensity Knob to 10. See Figure 5.43.
6. Allow Vacuum to run approximately 30 seconds and
take reading from Vacuum Gauge.
7. Rotate Vacuum Intensity Knob to 5. Allow Vacuum to
run approximately 1 minute. Take reading from
Vacuum Gauge.
8. Compare Vacuum Gauge readings to the chart below.
VACUUM
SETTING
mbar
READING
in. Hg
READING
10
538.4 - 660.3
15.9 - 19.5
267.5 - 328.4
7.9 - 9.7
CONTINUOUS
VACUUM LED
FIGURE 5.43
53
5- TROUBLESHOOTING
mbar
READING
in. Hg
READING
152.4 - 271.0
4.5 - 8.0
169.3 - 271.0
5.0 - 8.0
FIGURE 5.44
FIGURE 5.45
INTERMITTENT
0.75 Sec On 0.25 Off LED
INTERMITTENT
1.65 Sec On 0.35 Sec Off LED
VACUUM MODE
BUTTON
FIGURE 5.46
54
5- TROUBLESHOOTING
FIGURE 5.47
FIGURE 5.48
FIGURE 5.49
55
5- TROUBLESHOOTING
FIGURE 5.50
FIGURE 5.51
FIGURE 5.52
56
5- TROUBLESHOOTING
FIGURE 5.53
FIGURE 5.54
57
REAR PANEL
WARNING
DISCONNECT THE SYSTEM FROM THE POWER SOURCE
(OUTLET OR REMOVE BATTERY MODULE IF INSTALLED)
BEFORE ATTEMPTING ANY MAINTENANCE, INSTALLATION,
REMOVAL, OR REPLACEMENT PROCEDURES TO PREVENT
ELECTRICAL SHOCK AND POSSIBLE DAMAGE TO SYSTEM.
DISCONNECT
POWER CORD
FIGURE 6.1
REMOVE FRONT
ACCESS PANEL AND
DISCONNECT ALL
FIGURE 6.2
REMOVE
LANYARD
CAUTION
Be careful not to disconnect the Ribbon Cable from the
Intelect Advanced Therapy System.
NOTE:
When turning over Therapy System onto its face, place a clean,
soft cloth under the lens to prevent scratching or lens damage.
If the system is equipped with an sEMG Module, leave it in place
to maintain sEMG functions for Channels 1 and 2. The sEMG
Module will not interfere with installation of a Module to the
Therapy System.
FIGURE 6.3
RIBBON
CABLE
REMOVE
LABEL
FIGURE 6.4
58
LEAVE sEMG
IN PLACE
WARNING
BLUE STRIP
CAUTION
DO NOT TWIST RIBBON CABLE!
If Ribbon Cable is twisted, the pins will not properly align. If power is
applied to the system with misalignment of pins or a twisted ribbon
cable, the controlling electronics, in the Module will be destroyed and
possible damage to the System internal components could occur.
POSITION THERAPY
SYSTEM AND MODULE AS
SHOWN
FIGURE 6.5
8. Position Therapy System over Module as shown in
Figure 6.6. Align Therapy System Feet with Module
Indentions. RIBBON CABLE MUST BE POSITIONED
AS SHOWN!
9. Place System onto Module making certain the System
Feet are within the Module Identions.
RIBBON CABLE MUST
BE POSITIONED AS
SHOWN
FEET ALIGNED
WITH INDENTIONS
FIGURE 6.6
59
WARNING
DISCONNECT THE SYSTEM FROM THE POWER SOURCE
(OUTLET OR REMOVE BATTERY MODULE IF INSTALLED)
BEFORE ATTEMPTING ANY MAINTENANCE, INSTALLATION,
REMOVAL, OR REPLACEMENT PROCEDURES TO PREVENT
ELECTRICAL SHOCK AND POSSIBLE DAMAGE TO SYSTEM.
FIGURE 6.7
11. Set the assembly upright on the work surface and
install the new Extended Front Access Panel to the
Lanyard. See Figure 6.8.
EXTENDED FRONT
NOTE:
When mounting the Front Access Panel to the Therapy System,
make certain the Lanyard does not become kinked.
LANYARD
FIGURE 6.8
FIGURE 6.9
60
FIGURE 6.10
14. Turn the System On using the On/Off Switch. The
system will automatically recognize the added
module and display a configuration change message.
See Figure 6.11.
15. Read and carefully follow the instructions on the
Screen.
WARNING
Verify that the module installed is the module displayed in the
message BEFORE pressing the START Button. If it is not,
DO NOT press the START Button. Turn the system OFF and back
ON. If the problem persists, call the dealer or Chattanooga Group
Technical Support immediately. DO NOT USE THE SYSTEM until all
necessary repairs are made by a Technician certified by Chattanooga
Group. If use is attempted before repairs are made, the system may
operate unpredictably and has the potential of causing injury to the
patient or damage to the system internal components.
B. Module Removal
1. Remove Module in reverse order of installation
beginning with step A, 13.
2. After module is removed, follow instructions in
steps A, 14 and A, 15.
FIGURE 6.11
61
WARNING
DISCONNECT THE SYSTEM FROM THE POWER SOURCE
(OUTLET OR REMOVE BATTERY MODULE IF INSTALLED)
BEFORE ATTEMPTING ANY MAINTENANCE, INSTALLATION,
REMOVAL, OR REPLACEMENT PROCEDURES TO PREVENT
ELECTRICAL SHOCK AND POSSIBLE DAMAGE TO SYSTEM.
FIGURE 6.12
FIGURE 6.13
WARNING
Be careful not to damage the contacts of the sEMG Module,
Therapy System, or Module Stim Board contacts.
FIGURE 6.14
62
WARNING
DISCONNECT THE SYSTEM FROM THE POWER SOURCE
(OUTLET OR REMOVE BATTERY MODULE IF INSTALLED)
BEFORE ATTEMPTING ANY MAINTENANCE, INSTALLATION,
REMOVAL, OR REPLACEMENT PROCEDURES TO PREVENT
ELECTRICAL SHOCK AND POSSIBLE DAMAGE TO SYSTEM.
FIGURE 6.15
7. Turn the System On using the On/Off Switch. The
system will automatically recognize the added
module and display a configuration change message.
See Figure 6.16.
8. Read and carefully follow the instructions on the
Screen.
WARNING
Verify that the module installed is the module displayed in the
message BEFORE pressing the START Button. If it is not,
DO NOT press the START Button. Turn the system OFF and back
ON. If the problem persists, call the dealer or Chattanooga Group
Technical Support immediately. DO NOT USE THE SYSTEM until all
necessary repairs are made by a Technician certified by Chattanooga
Group. If use is attempted before repairs are made, the system may
operate unpredictably and has the potential of causing injury to the
patient or damage to the system's internal components.
FIGURE 6.16
C. Module Removal
1. Remove Module, if necessary, from Therapy System.
Refer to 6.2, B for instructions.
2. Place a Flat Blade Screwdriver under the locking tab
of the sEMG Module. Firmly push in and twist
Screwdriver. The Module should release from the
Therapy System. See Figure 6.17.
NOTE:
If no sEMG Module will be re-installed on the system. Install
sEMG Fill Kit to protect and cover the Stim Board contacts.
3. Refer to steps 7 and 8 to finalize removal.
FIGURE 6.17
63
WARNING
FIGURE 6.18
FIGURE 6.19
3. Select the side of the Therapy System Cart the
Vacuum Module is to be installed. Remove the Top
Storage Bin Mounting Rails from the side of the cart
chosen for Vacuum Electrode Module installation.
Store these rails for future Top Bin re-installation in the
event the Vacuum Electrode Module is removed from
the Therapy System Cart. See Figure 6.20.
FIGURE 6.20
64
WARNING
DISCONNECT THE SYSTEM FROM THE POWER SOURCE
(OUTLET OR REMOVE BATTERY MODULE IF INSTALLED)
BEFORE ATTEMPTING ANY MAINTENANCE, INSTALLATION,
REMOVAL, OR REPLACEMENT PROCEDURES TO PREVENT
ELECTRICAL SHOCK AND POSSIBLE DAMAGE TO SYSTEM.
FIGURE 6.21
5. Disconnect the Power Cables. See Figure 6.22.
FIGURE 6.22
6. Install the Power Distribution Strip into the housing in
the Cart Bottom. See Figure 6.23.
NOTE:
Make certain the Power Distribution Strip is seated completely in
housing.
FIGURE 6.23
65
WARNING
DISCONNECT THE SYSTEM FROM THE POWER SOURCE (WALL
OUTLET OR REMOVE BATTERY MODULE IF INSTALLED) BEFORE
ATTEMPTING ANY MAINTENANCE, INSTALLATION, REMOVAL
OR REPLACEMENT PROCEDURES TO PREVENT ELECTRICAL
SHOCK AND POSSIBLE DAMAGE TO SYSTEM.
MAINS
POWER CORD
VACUUM
MODULE
POWER
CORD
NOTE:
If a second Vacuum Electrode Module is installed, the power
cord will connect to the Power Distribution Strip below the
Therapy System Power Cord.
THERAPY
SYSTEM
POWER
CORD
FIGURE 6.24
8. Connect the Vacuum Module Power Cord (short cord)
to the Vacuum Electrode Module. See Figure 6.25.
NOTE:
Make certain the Power Cord is completely seated in the
Vacuum Electrode Module Mains Disconnect.
FIGURE 6.25
FIGURE 6.26
66
WARNING
DISCONNECT THE SYSTEM FROM THE POWER SOURCE
(OUTLET OR REMOVE BATTERY MODULE IF INSTALLED)
BEFORE ATTEMPTING ANY MAINTENANCE, INSTALLATION,
REMOVAL, OR REPLACEMENT PROCEDURES TO PREVENT
ELECTRICAL SHOCK AND POSSIBLE DAMAGE TO SYSTEM.
FIGURE 6.27
11. Route the Exhaust/Overflow Hose through the
Extrusion track opposite the Power cord and into the
Therapy System Cart Base. See Figure 6.28.
FIGURE 6.28
12. Mount the Bottom Plate to the Cart Base and
secure with the seven screws removed earlier. See
Figure 6.29.
NOTE:
Make certain the Mains Power Cord is routed properly through
the Cart Base.
Stand Therapy System Cart upright on the casters.
FIGURE 6.29
67
WARNING
DISCONNECT THE SYSTEM FROM THE POWER SOURCE
(OUTLET OR REMOVE BATTERY MODULE IF INSTALLED)
BEFORE ATTEMPTING ANY MAINTENANCE, INSTALLATION,
REMOVAL, OR REPLACEMENT PROCEDURES TO PREVENT
ELECTRICAL SHOCK AND POSSIBLE DAMAGE TO SYSTEM.
FIGURE 6.30
14. Install Cart Top to cart and secure with the four
screws removed earlier. See Figure 6.31.
FIGURE 6.31
FIGURE 6.32
68
WARNING
DISCONNECT THE SYSTEM FROM THE POWER SOURCE
(OUTLET OR REMOVE BATTERY MODULE IF INSTALLED)
BEFORE ATTEMPTING ANY MAINTENANCE, INSTALLATION,
REMOVAL, OR REPLACEMENT PROCEDURES TO PREVENT
ELECTRICAL SHOCK AND POSSIBLE DAMAGE TO SYSTEM.
FIGURE 6.33
18. Raise the Vacuum Electrode Module by hand until it
touches the Cart Top. Place the Positioning Tabs into
the Bin Guides as shown in Figure 6.34.
19. Secure each Positioning Tab in place with the screw
provided.
FIGURE 6.34
FIGURE 6.35
69
WARNING
DISCONNECT THE SYSTEM FROM THE POWER SOURCE
(OUTLET OR REMOVE BATTERY MODULE IF INSTALLED)
BEFORE ATTEMPTING ANY MAINTENANCE, INSTALLATION,
REMOVAL, OR REPLACEMENT PROCEDURES TO PREVENT
ELECTRICAL SHOCK AND POSSIBLE DAMAGE TO SYSTEM.
22. Install the Drain Hose Stop onto the center back of
the Storage Bin that is to go beneath the Vacuum
Module See Figure 6.36.
FIGURE 6.36
23. Install the Drain Hose from the Vacuum Electrode
Module onto the Drain Hose Stop. See Figure 6.37.
NOTE:
The Drain Hose must be installed onto the Drain Hose Stop for
the Vacuum to properly operate.
24. Conduct Tests as described in 5.21 through 5.24.
C. Vacuum Electrode Module Removal
Remove the Vacuum Electrode Module by following the
installation instructions.
FIGURE 6.37
70
WARNING
DISCONNECT THE SYSTEM FROM THE POWER SOURCE
(OUTLET OR REMOVE BATTERY MODULE IF INSTALLED)
BEFORE ATTEMPTING ANY MAINTENANCE, INSTALLATION,
REMOVAL, OR REPLACEMENT PROCEDURES TO PREVENT
ELECTRICAL SHOCK AND POSSIBLE DAMAGE TO SYSTEM.
FIGURE 6.38
FIGURE 6.39
WARNING
Use part number 71772 fuses only.
Using any other fuse may cause damage to internal components of
the Intelect Advanced Therapy System.
FIGURE 6.40
71
REMOVE 4 SCREWS
WARNING
DISCONNECT THE SYSTEM FROM THE POWER SOURCE
(OUTLET OR REMOVE BATTERY MODULE IF INSTALLED)
BEFORE ATTEMPTING ANY MAINTENANCE, INSTALLATION,
REMOVAL, OR REPLACEMENT PROCEDURES TO PREVENT
ELECTRICAL SHOCK AND POSSIBLE DAMAGE TO SYSTEM.
FIGURE 6.41
FIGURE 6.42
CONTROL BOARD
RIBBON CABLE
FIGURE 6.43
72
WARNING
DISCONNECT THE SYSTEM FROM THE POWER SOURCE
(OUTLET OR REMOVE BATTERY MODULE IF INSTALLED)
BEFORE ATTEMPTING ANY MAINTENANCE, INSTALLATION,
REMOVAL, OR REPLACEMENT PROCEDURES TO PREVENT
ELECTRICAL SHOCK AND POSSIBLE DAMAGE TO SYSTEM.
REMOVE
FIGURE 6.44
73
WARNING
DISCONNECT THE SYSTEM FROM THE POWER SOURCE
(OUTLET OR REMOVE BATTERY MODULE IF INSTALLED)
BEFORE ATTEMPTING ANY MAINTENANCE, INSTALLATION,
REMOVAL, OR REPLACEMENT PROCEDURES TO PREVENT
ELECTRICAL SHOCK AND POSSIBLE DAMAGE TO SYSTEM.
REMOVE
SCREWS
FIGURE 6.45
3. Remove the Fan Harness from the Control Board. See
Figure 6.46.
REMOVE
FAN HARNESS
FIGURE 6.46
4. Remove the Fan Baffle from the Fan Housing. See
Figure 6.47.
C. Replacing Fan
1. Replace new Fan, part number 27158 by reversing the
steps in section 6.7.
NOTE:
Do not over tighten the screws. Over tightening will damage the
threads of the brass standoffs.
2. Re-assemble system. Refer to 6.6, part C for proper
instructions.
REMOVE
FIGURE 6.47
74
POWER
DISTRIBUTION
BOARD
WARNING
DISCONNECT THE SYSTEM FROM THE POWER SOURCE
(OUTLET OR REMOVE BATTERY MODULE IF INSTALLED)
BEFORE ATTEMPTING ANY MAINTENANCE, INSTALLATION,
REMOVAL, OR REPLACEMENT PROCEDURES TO PREVENT
ELECTRICAL SHOCK AND POSSIBLE DAMAGE TO SYSTEM.
FIGURE 6.48
REMOVE
RIBBON CABLE
REMOVE
RETAINING
SCREWS
WARNING
MAKE CERTAIN THE CONTROL BOARD RIBBON CABLE IS
INSTALLED WITH THE BLACK STRIP AT THE NUMBER 1 ON THE
CONTROL BOARD. FAILURE TO PROPERLY INSTALL RIBBON
CABLE COULD CAUSE EXTENSIVE DAMAGE TO THE INTERNAL
COMPONENTS OF THE SYSTEM WHEN TURNED ON.
FIGURE 6.49
NOTE:
Do not over tighten the screws. Over tightening will damage the
threads of the brass inserts.
2. Install Therapy Intensity Control Knob on the front side
of Therapy System Top.
3. Re-assemble Therapy System referring to the
appropriate sections of this manual for proper
instructions.
FIGURE 6.50
75
WARNING
DISCONNECT THE SYSTEM FROM THE POWER SOURCE (WALL
OUTLET OR REMOVE BATTERY MODULE IF INSTALLED) BEFORE
ATTEMPTING ANY MAINTENANCE, INSTALLATION, REMOVAL
OR REPLACEMENT PROCEDURES TO PREVENT ELECTRICAL
SHOCK AND POSSIBLE DAMAGE TO SYSTEM.
REMOVE
RETAINING
SCREWS
FIGURE 6.51
WARNING
MAKE CERTAIN WHEN INSTALLING CONTROL BOARD
ASSEMBLY TO KEYMAT PC BOARD PINS FROM THE
CONTROL BOARD ALIGN WITH THE CONNECTOR ON
THE KEYMAT PC BOARD.
NOTE:
Do not over tighten the screws. Over tightening will damage the
threads of the brass inserts.
2. Re-assemble Therapy System referring to the
appropriate sections of this manual for proper
instructions.
FIGURE 6.52
FIGURE 6.53
76
WARNING
DISCONNECT THE SYSTEM FROM THE POWER SOURCE
(OUTLET OR REMOVE BATTERY MODULE IF INSTALLED)
BEFORE ATTEMPTING ANY MAINTENANCE, INSTALLATION,
REMOVAL, OR REPLACEMENT PROCEDURES TO PREVENT
ELECTRICAL SHOCK AND POSSIBLE DAMAGE TO SYSTEM.
FIGURE 6.54
PULL CONNECTOR
BOARD TOWARD
SYSTEM FRONT
FIGURE 6.55
C. Replacing Connector Board
1. Replace new Connector Board, part number 27059
in reverse order of steps 1-3 of section 6.10. Make
Certain the words THIS SIDE UP on Connector
Board are facing up. Make certain Connector Board is
completely seated in Stim Board Connector. See
Figure 6.56.
2. Re-install the Connector Infill and Lanyard. Refer to
Figure 6.54.
3. Re-assemble Therapy System. Refer to 6.6, part C for
proper instructions.
NOTE:
Do not over tighten the screws. Over tightening will damage the
threads of the brass inserts.
FIGURE 6.56
77
WARNING
DISCONNECT THE SYSTEM FROM THE POWER SOURCE
(OUTLET OR REMOVE BATTERY MODULE IF INSTALLED)
BEFORE ATTEMPTING ANY MAINTENANCE, INSTALLATION,
REMOVAL, OR REPLACEMENT PROCEDURES TO PREVENT
ELECTRICAL SHOCK AND POSSIBLE DAMAGE TO SYSTEM.
FIGURE 6.57
FIRMLY PULL EACH SIDE OF
THE BOARD UP
FIGURE 6.58
C. Replacing Ultrasound Board
1. Replace new Ultrasound Board, part number 27055 in
reverse order of steps 1-5 above.
WARNING
MAKE CERTAIN THE STIM BOARD HEADER PINS ARE PROPERLY
ALIGNED WITH THE ULTRASOUND BOARD WHEN INSTALLING
ULTRASOUND BOARD. SEE FIGURE 6.59.
FAILURE TO PROPERLY ALIGN HEADER PINS WILL RESULT IN
SEVERE DAMAGE TO THE SYSTEM.
FIGURE 6.59
78
WARNING
DISCONNECT THE SYSTEM FROM THE POWER SOURCE
(OUTLET OR REMOVE BATTERY MODULE IF INSTALLED)
BEFORE ATTEMPTING ANY MAINTENANCE, INSTALLATION,
REMOVAL, OR REPLACEMENT PROCEDURES TO PREVENT
ELECTRICAL SHOCK AND POSSIBLE DAMAGE TO SYSTEM.
LOCKED
INDICATOR
FIGURE 6.60
REMOVE
RIBBON CABLE
BLUE STRIP
FIGURE 6.60A
RELEASE LOCKING TABS
ON EACH SIDE OF STIM
BOARD
FIGURE 6.61
CHANNEL 1/2
POSITION
CHANNEL 3/4
POSITION
FIGURE 6.61A
79
WARNING
DISCONNECT THE SYSTEM FROM THE POWER SOURCE
(OUTLET OR REMOVE BATTERY MODULE IF INSTALLED)
BEFORE ATTEMPTING ANY MAINTENANCE, INSTALLATION,
REMOVAL, OR REPLACEMENT PROCEDURES TO PREVENT
ELECTRICAL SHOCK AND POSSIBLE DAMAGE TO SYSTEM.
REMOVE
SCREWS
FIGURE 6.62
REAR
MOUNTING
TABS
BRACKETS
POWER
SUPPLY
ASSEMBLY
DANGER
POWER SUPPLIES RETAIN HIGH VOLTAGE!
WHEN REMOVING FROM SYSTEM, HANDLE POWER
SUPPLIES BY MOUNTING BRACKETS ONLY.
FIGURE 6.63
100 WATT
POWER SUPPLY
75 WATT
POWER SUPPLY
(COMBINATION SYSTEMS ONLY)
FIGURE 6.64
80
WARNING
DISCONNECT THE SYSTEM FROM THE POWER SOURCE
(OUTLET OR REMOVE BATTERY MODULE IF INSTALLED)
BEFORE ATTEMPTING ANY MAINTENANCE, INSTALLATION,
REMOVAL, OR REPLACEMENT PROCEDURES TO PREVENT
ELECTRICAL SHOCK AND POSSIBLE DAMAGE TO SYSTEM.
MAINS WIRING
HARNESS
PLIERS NOT
SHOWN
FOR CLARITY
NOTE:
Steps 10 and 11 below applies only to Combination Systems.
10. Remove 75 Watt Power Supply from Mounting
Bracket by removing the two mounting screws on the
back of Mounting Bracket securing the 75 Watt
Power Supply. See inset at Figure 6.66
11. Using Insulated Needle Nose Pliers, remove the 100
Watt Power Supply Harness from the 75 Watt Power
Supply. See Figure 6.67.
FIGURE 6.65
100 WATT
FIGURE 6.66
FIGURE 6.67
81
WARNING
DISCONNECT THE SYSTEM FROM THE POWER SOURCE
(OUTLET OR REMOVE BATTERY MODULE IF INSTALLED)
BEFORE ATTEMPTING ANY MAINTENANCE, INSTALLATION,
REMOVAL, OR REPLACEMENT PROCEDURES TO PREVENT
ELECTRICAL SHOCK AND POSSIBLE DAMAGE TO SYSTEM.
RELEASE
TABS
FIGURE 6.68
LIFT CONNECTOR
INFILL OUT OF MODULE
HOUSING
FIGURE 6.69
FIGURE 6.70
82
WARNING
DISCONNECT THE SYSTEM FROM THE POWER SOURCE
(OUTLET OR REMOVE BATTERY MODULE IF INSTALLED)
BEFORE ATTEMPTING ANY MAINTENANCE, INSTALLATION,
REMOVAL, OR REPLACEMENT PROCEDURES TO PREVENT
ELECTRICAL SHOCK AND POSSIBLE DAMAGE TO SYSTEM.
CONNECTOR
BOARD
FIGURE 6.71
LIFT OUT
STIM BOARD
FIGURE 6.71a
83
WARNING
DISCONNECT THE SYSTEM AND VACUUM MODULE FROM
THE POWER SOURCE (OUTLET OR REMOVE BATTERY MODULE
IF INSTALLED) AND REMOVE VACUUM MODULE FROM THE
SYSTEM CART BEFORE ATTEMPTING ANY MAINTENANCE,
INSTALLATION, REMOVAL OR REPLACEMENT PROCEDURES
TO PREVENT ELECTRICAL SHOCK AND POSSIBLE DAMAGE TO
SYSTEM OR VACUUM MODULE.
REMOVE
PANEL SCREWS
FIGURE 6.72
REMOVE TIE WRAP
WIRING HARNESS
FIGURE 6.73
RETAINING
SCREWS
VACUUM
HOSES
CHECK
VALVE
FIGURE 6.74
84
WARNING
DISCONNECT THE SYSTEM AND VACUUM MODULE FROM
THE POWER SOURCE (OUTLET OR REMOVE BATTERY MODULE
IF INSTALLED) AND REMOVE VACUUM MODULE
FROM THE SYSTEM CART BEFORE ATTEMPTING ANY
MAINTENANCE, INSTALLATION, REMOVAL, OR REPLACEMENT
PROCEDURES TO PREVENT ELECTRICAL SHOCK AND
POSSIBLE DAMAGE TO SYSTEM OR VACUUM MODULE.
REMOVE
PANEL SCREWS
FIGURE 6.75
REMOVE TIE WRAP
WIRING HARNESSES
FIGURE 6.76
VACUUM
HOSES
RETAINING
SCREWS
FIGURE 6.77
85
WARNING
REMOVE
NUTS AND
WASHERS
REMOVE
HOSE
FIGURE 6.78
86
WARNING
DISCONNECT THE SYSTEM AND VACUUM MODULE FROM
THE POWER SOURCE (OUTLET OR REMOVE BATTERY MODULE
IF INSTALLED) AND REMOVE VACUUM MODULE
FROM THE SYSTEM CART BEFORE ATTEMPTING ANY
MAINTENANCE, INSTALLATION, REMOVAL, OR REPLACEMENT
PROCEDURES TO PREVENT ELECTRICAL SHOCK AND
POSSIBLE DAMAGE TO SYSTEM OR VACUUM MODULE.
REMOVE
PANEL SCREWS
FIGURE 6.79
1
2
QTY
1
8
1
1
1
1
1
1
1
1
2
4
5
8
7
9
10
FIGURE 6.80
87
WARNING
DISCONNECT THE SYSTEM AND VACUUM MODULE FROM
THE POWER SOURCE (OUTLET OR REMOVE BATTERY MODULE
IF INSTALLED) AND REMOVE VACUUM MODULE
FROM THE SYSTEM CART BEFORE ATTEMPTING ANY
MAINTENANCE, INSTALLATION, REMOVAL, OR REPLACEMENT
PROCEDURES TO PREVENT ELECTRICAL SHOCK AND
POSSIBLE DAMAGE TO SYSTEM OR VACUUM MODULE.
REMOVE
HARNESS
REMOVE
VACUUM HOSE
FIGURE 6.81
REMOVE SCREWS
TABS
TABS
FIGURE 6.82
88
WARNING
REMOVE
VACUUM HOSE
ASSEMBLY
REMOVE
HARNESS
REMOVE
LEAD WIRES
FIGURE 6.83
DO NOT OVER
TORQUE
FIGURE 6.84
89
REMOVE SCREWS
WARNING
DISCONNECT THE SYSTEM AND VACUUM MODULE FROM
THE POWER SOURCE (OUTLET OR REMOVE BATTERY MODULE
IF INSTALLED) AND REMOVE VACUUM MODULE
FROM THE SYSTEM CART BEFORE ATTEMPTING ANY
MAINTENANCE, INSTALLATION, REMOVAL, OR REPLACEMENT
PROCEDURES TO PREVENT ELECTRICAL SHOCK AND
POSSIBLE DAMAGE TO SYSTEM OR VACUUM MODULE.
REMOVE
SCREWS
FIGURE 6.85
WIRING
HARNESS
SCREWS
FIGURE 6.86
90
REMOVE SCREWS
WARNING
DISCONNECT THE SYSTEM AND VACUUM MODULE FROM
THE POWER SOURCE (OUTLET OR REMOVE BATTERY MODULE
IF INSTALLED) AND REMOVE VACUUM MODULE
FROM THE SYSTEM CART BEFORE ATTEMPTING ANY
MAINTENANCE, INSTALLATION, REMOVAL, OR REPLACEMENT
PROCEDURES TO PREVENT ELECTRICAL SHOCK AND
POSSIBLE DAMAGE TO SYSTEM OR VACUUM MODULE.
REMOVE
SCREWS
FIGURE 6.87
CAPACITOR
(C4)
DANGER
POWER SUPPLIES RETAIN HIGH VOLTAGE!
WHEN REMOVING FROM SYSTEM, HANDLE POWER
SUPPLIES BY MOUNTING BRACKETS ONLY.
FIGURE 6.88
SCREWS
FIGURE 6.89
91
WARNING
DISCONNECT THE SYSTEM AND VACUUM MODULE FROM
THE POWER SOURCE (OUTLET OR REMOVE BATTERY MODULE
IF INSTALLED) AND REMOVE VACUUM MODULE
FROM THE SYSTEM CART BEFORE ATTEMPTING ANY
MAINTENANCE, INSTALLATION, REMOVAL, OR REPLACEMENT
PROCEDURES TO PREVENT ELECTRICAL SHOCK AND
POSSIBLE DAMAGE TO SYSTEM OR VACUUM MODULE.
REMOVE
ALL BINS
FIGURE 6.90
11.5 cm (4.5)
SYSTEM ONLY
16.5 cm (6.5)
SYSTEM W/MODULE
FIGURE 6.91
FIGURE 6.92
92
7- GENERAL MAINTENANCE
WARNING
DISCONNECT THE SYSTEM AND VACUUM MODULE FROM
THE POWER SOURCE (OUTLET OR REMOVE BATTERY MODULE
IF INSTALLED) AND REMOVE VACUUM MODULE
FROM THE SYSTEM CART BEFORE ATTEMPTING ANY
MAINTENANCE, INSTALLATION, REMOVAL, OR REPLACEMENT
PROCEDURES TO PREVENT ELECTRICAL SHOCK AND
POSSIBLE DAMAGE TO SYSTEM OR VACUUM MODULE.
93
WARNING
FIGURE 8.1
ULTRASOUND CALIBRATION
BUTTON
FIGURE 8.2
START
BUTTON
HEAD SIZE
BUTTON
FIGURE 8.3
94
9- PARTS
Bottom Assembly
See Page 90
2
5
3
ITEM
NUMBER
1
2
3
4
5
6
7
PART
NUMBER
27159
27306
27138
27029
27020
21188
27007
DESCRIPTION
Bottom Assembly to Top Assembly Ribbon Cable
Rear Access Panel
Screw, M3 x 16 mm
Front Access Panel
Lanyard
Screw, #4-40 x .375"
Front Infill
95
QTY
REQD
1
1
4
1
1
1
1
9- PARTS
11
1
12
13
3, 4, 15
10
14
6
9
ITEM
NUMBER
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
PART
NUMBER
27022
27276
27153
27154
27277
27010
27274
27436
27059
27142
27006
27048
27049
27000
71772
7
8
DESCRIPTION
Mains Harness Clip
Main Power Switch
Fuse Holder
Fuse Carrier
Mains Power Input Connector
Applicator Holder
Feet
Stim Board 1/4 Turn Pin
Stim Connector Board
Screws, M3 x 6mm
Base Housing
100 Watt Power Supply
75 Watt Power Supply (Combination Systems Only)
Power Supply Mounting Bracket
Fuse, 6.3A
96
QTY
REQD
1
1
2
2
1
1
4
1
1
2
1
1
1
1
2
9- PARTS
3
3
3
4
ITEM
NUMBER
1
2
3
4
PART
NUMBER
27055
27161
27160
27057
DESCRIPTION
Ultrasound PC Board
Ultrasound to Stim Header
PC Board Stand Off
Stim PC Board
97
QTY
REQD
1
1
4
1
9- PARTS
1
7
4
6
ITEM
NUMBER
1
2
3
4
5
6
7
PART
NUMBER
27107
27026
27027
27051
27142
27028
27305
DESCRIPTION
Top Housing
Left Keymat
Right Keymat
Keymat PC Board
Screw, M3 x 6 mm
Bottom Keymat
Intensity Control Knob
98
QTY
REQD
1
1
1
1
10
1
1
9- PARTS
11
3
1
6
6
10
ITEM
NUMBER
1
2
3
4
5
6
7
8
9
10
11
12
PART
NUMBER
27093
27162
27012
27053
27137
21188
27021
27158
27136
27144
27024
27009
DESCRIPTION
Color Display Mounting Bracket/Spacer
Color Display
Contrast Knob
Color Control PC Board
Screw, M4 x 8 mm
Screw, PSM PT Type
Air Exhaust Funnel
Cooling Fan
Screw, #4-40 x .375
Fan Mounting Stand Off, M4 x 16 mm
MMC Card Top
Patient Data Card and MMC Card System Port Housing
99
QTY
REQD
2
1
1
1
2
4
1
1
4
2
1
1
9- PARTS
2
14
3
4
13
5
7
12
8
8
9
10
11
ITEM
NUMBER
1
2
3
4
5
6
7
8
9
10
11
12
13
14
PART
NUMBER
27512
27180
27145
27012
27053
27137
27021
27144
27142
27158
27136
27756
27009
27085
DESCRIPTION
Screw, M3 x 4 mm
Monochromatic LCD
Standoff, M3 x 10 mm
Contrast Knob
Control PCB
Screw, M4 x 8 mm
Air Exhaust Funnel
Standoff, M4 x 16 mm
Screw, M3 x 6 mm
Cooling Fan
Screw, M4 x 35 mm
Inverter
Patient Data Card and MMC Card System Port Housing
MMC Card Top
100
QTY
REQD
4
1
4
1
1
2
1
2
4
1
2
1
1
1
9- PARTS
7
3
ITEM
NUMBER
1
2
3
4
5
6
7
8
PART
NUMBER
27015
27353
27057
27016
27150
27061
27017
27136
DESCRIPTION
Module Top
Stim Board to Ribbon Cable Header
Stim PC Board
Module Bottom Housing
Feet
Channel 3/4 Stim Connector PC Board
Connector Infill
Module to Therapy System Ribbon Cable
101
QTY
REQD
1
1
1
1
4
1
1
1
9- PARTS
PNEUMATIC ASSEMBLY
See Pages 97-98
1
2
3
4
5
6
10
11
12
13
27
14
15
26
9
9
16
25
24
17
23
22
18
9
21
19
17
18
20
102
20
19
20
9- PARTS
PART
NUMBER
27667
27726
27438
27317
27318
27442
27151
27637
27140
27741
27265
27142
27732
27737
27152
27327
27738
73168
27743
27137
27722
27736
27623
27620
27751
27440
27632
DESCRIPTION
Vacuum Intensity Knob
Vacuum Fascia with Lens
Vacuum Control PC Board
Channel 1 Vacuum to Therapy System Lead Wire
Channel 2 Vacuum to Therapy System Lead Wire
Vacuum Connector PC Board
Power Switch
Vacuum Motor/Pump Assembly
Screw, M3 x 8 mm
Grommet
Power Supply
Screw, M3 x 6 mm
Screw, #6 x 1/4
Rear Cover
Mains Connector Socket
Vacuum to Cart Power Cable
Vacuum Positioning Tab
Flat Washer, Plated
Lock Washer, #8 Split
Screw, M4 x 8 mm
Screw, PSM K40 x 8 mm
Vacuum Bracket
Reservoir
Reservoir PC Board Assembly
Nut, Keps
Stim Connector PC Board PC BOARD
Keymat/Light Pipe Assembly
103
QTY
REQD
1
1
1
1
1
1
1
1
13
1
1
4
2
1
1
1
2
2
2
4
3
1
1
1
3
1
1
9- PARTS
5
3
5
6
21
5
6
7
5
6
7
17
20
10
11
16
18
12
19
13
15
14
104
9- PARTS
PART
NUMBER
27728
27730
27733
27734
27730
27727
60419
79557
27637
60075
27735
27730
27730
27631
27623
27728
27730
27628
27753
27633
27730
DESCRIPTION
Tube, Manifold to Control PC Board Sensor- 12.7 cm (5.00) Long
Tube, Manifold- 2.3 cm (0.875) Long
Manifold Tee
Manifold Cap
Tube, Manifold to Connector and Compressor to Check Valve- 2.5 cm (1.00) Long
Nut, M6 Hex Zinc Plated
Washer, 1/4 Internal Tooth Lock
Harness Clip, 6.3 mm x 6.3 mm (1/4 x 1/4)
Compressor Assembly
Tyrap
Check Valve
Tube, Compressor to Top Right Side of Reservoir- 6.3 cm (2.50) Long
Tube, Reservoir Drain- 28.6 cm (11.250) Long
Drain Bin Clip
Reservoir
Tube, Reservoir to Reservoir PC Board- 12.7 cm (5.00) Long
Tube, Exhaust/Overflow- 122 cm (48.00) Long
Vacuum Lead Hose Connector Pin
Connector Pin Cap
Connector Pin Manifold
Tube, Manifold to Top Left Side of Reservoir- 20.3 cm (8.00) Long
105
QTY
REQD
1
4
5
1
5
4
4
1
1
1
1
1
1
1
1
1
1
4
4
1
1
10- SCHEMATICS
,QWHOHFW$GYDQFHG7KHUDS\6\VWHP
10- SCHEMATICS
,QWHOHFW$GYDQFHG7KHUDS\6\VWHP
10- SCHEMATICS
,QWHOHFW$GYDQFHG7KHUDS\6\VWHP
10- SCHEMATICS
,QWHOHFW$GYDQFHG7KHUDS\6\VWHP
10- SCHEMATICS
,QWHOHFW$GYDQFHG7KHUDS\6\VWHP
10- SCHEMATICS
,QWHOHFW$GYDQFHG7KHUDS\6\VWHP
10- SCHEMATICS
,QWHOHFW$GYDQFHG7KHUDS\6\VWHP
10- SCHEMATICS
,QWHOHFW$GYDQFHG7KHUDS\6\VWHP
10- SCHEMATICS
,QWHOHFW$GYDQFHG7KHUDS\6\VWHP
10- SCHEMATICS
,QWHOHFW$GYDQFHG7KHUDS\6\VWHP
10- SCHEMATICS
,QWHOHFW$GYDQFHG7KHUDS\6\VWHP
10- SCHEMATICS
,QWHOHFW$GYDQFHG7KHUDS\6\VWHP
10- SCHEMATICS
,QWHOHFW$GYDQFHG7KHUDS\6\VWHP
10- SCHEMATICS
,QWHOHFW$GYDQFHG7KHUDS\6\VWHP
10- SCHEMATICS
,QWHOHFW$GYDQFHG7KHUDS\6\VWHP
10- SCHEMATICS
,QWHOHFW$GYDQFHG7KHUDS\6\VWHP
10- SCHEMATICS
,QWHOHFW$GYDQFHG7KHUDS\6\VWHP
10- SCHEMATICS
,QWHOHFW$GYDQFHG7KHUDS\6\VWHP
10- SCHEMATICS
,QWHOHFW$GYDQFHG7KHUDS\6\VWHP
10- SCHEMATICS
,QWHOHFW$GYDQFHG7KHUDS\6\VWHP
11- WARRANTY
Chattanooga Group ("Company") warrants that the Intelect Advanced Therapy System, Channel 3/4 Electrotherapy Module, Laser
Module, and sEMG Module ("Products") are free of defects in material and workmanship. This warranty shall remain in effect for two
years (24 months) from the date of original consumer purchase. If these Products fail to function during the two year warranty period due
to a defect in material or workmanship, Company or the selling dealer will repair or replace the respective Product without charge within
a period of thirty (30) days from the date on which the Product is returned to the Company or the dealer.
All repairs to the Product must be performed by a service center authorized by the Company. Any modifications or repairs performed by
unauthorized centers or groups will void this warranty.
The warranty period for certain accessories is 180 days. Accessories consist of Lead Wires, Operator Remote, Electrodes, Patient Data
Cards, sEMG Data Cards, and Nylatex.
The warranty period for the Therapy System Cart, Battery Module, Ultrasound Applicators, and Laser Applicators is one year (12
Months).
This Warranty Does Not Cover:
Replacement parts or labor furnished by anyone other than the Company, the selling dealer, or a certified Company service technician.
Defects or damage caused by labor furnished by someone other than Company, the selling dealer or a certified Company service
technician.
Any malfunction or failure in the Product caused by product misuse, including, but not limited to, the failure to provide reasonable and
required maintenance or any use that is inconsistent with the Product User's Manual.
COMPANY SHALL NOT BE LIABLE IN ANY EVENT FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES.
Some locations do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation or exclusion may
not apply to you.
To obtain service from Company or the selling dealer under this warranty:
1. A written claim must be made within the warranty period to the Company or the selling dealer. Written claims made to the
Company should be sent to:
Chattanooga Group
4717 Adams Road
Hixson, TN 37343 USA
Phone: +1-423-870-7200
FAX: +1-423-870-2046
and
2. The Product must be returned to the Company or the selling dealer by the owner. A Return Authorization (RA) Number must be
obtained before returning any product to the Company.
This warranty gives you specific legal rights and you may also have other rights which vary from location to location.
The Company does not authorize any person or representative to create for it any other obligation or liability in connection with the sale
of the Product.
Any representative or agreement not contained in the warranty shall be void and of no effect.
THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED,
INCLUDING ANY WARRANTY OR MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
126
27833A
www.chattgroup.com
2005 Encore Medical, L.P.