ALUMNO: Julio Miguel Naranjo Fiallos

PARALELO:103

APARATO MEDIDOR DE FLUJO

PRCTICA #3

Laboratorio de Instrumentacin Bsica


Facultad de Ingeniera Mecnica y Ciencias de la Produccin (FIMCP)
ESPOL, Guayaquil
25/11/2015

ABSTRACT
At the beginning a short evaluation of the learned to the previous practice (deadweight), in practice they discussed the
instruments flow measurement and what they then proceeded to get the data using the manual was made to calculate
discharge coefficients using the equation obtained from the conservation of energy and flow in order to obtain an equation
that helps us calculate the speed, which will help us calculate the volumetric flow and mass flow later. The test bench allows
familiar with some typical methods of measuring flow of an incompressible fluid, the same applications showing the
Bernoulli equation. Measuring the flow is done using a direct method; in this case the weighing tank and timer, and an
indirect method through a venturi, nozzle, orifice and a rotameter connected in series.

KEYWORDS

Nozzle

Venturi

Plate hole

Flow

discharge coefficient

Rotameter

Energy conservation

RESUMEN
Al inicio se tom una corta evaluacin de los aprendido hasta la prctica anterior(pesos muertos), En la prctica se convers
de los instrumentos de medicin de caudal y para qu sirven, luego se procedi a obtener los datos siguiendo las

ALUMNO: Julio Miguel Naranjo Fiallos

PARALELO:103

instrucciones del manual para calcular los coeficientes de descarga usando las ecuacin obtenidos a partir de la
conservacin de la energa y caudal para poder obtener una ecuacin que nos ayuda a calcular la velocidad, la cual nos
ayudara calculando el flujo volumtrico y posteriormente el flujo msico. El banco de prueba permite familiarizarse con
algunos de los mtodos tpicos de medicin de flujo de un fluido incompresible, al mismo que muestra las aplicaciones de la
ecuacin de Bernoulli. La medicin del flujo se la hace utilizando un mtodo directo; en este caso tanque pesaje y
cronmetro, y un mtodo indirecto a travs de un Venturi, Tobera, Orificio y un Rotmetro conectados en serie.

PALABRAS CLAVES

Tobera

Venturi

Placa orificio

Caudal

Rotmetro

Coeficiente descarga

Conservacin de la energia

ndice:

I.

INTRODUCCIN

II.

OBJETIVOS

ALUMNO: Julio Miguel Naranjo Fiallos


III.

MATERIALES Y EQUIPOS

IV.

PROCEDIMIENTO

V.

TABLA DE DATOS

VI.

CALCULOS REPRESENTATIVOS

VII.

TABLA DE RESULTADOS

VIII.

GRFICAS

IX.

ANALISIS DE RESULTADOS

X.

CONCLUSIONES

XI.

RECOMENDACIONES

XII.

BIBLIOGRAFIA

PARALELO:103

Introduccin
En la practica se estudio los medidores volumtricos los cuales miden presiones diferenciales, de acuerdo a este principio de
funcionamiento se aplico la ecuacin de bernoulli para poder obtener el coeficiente de descarga a partir de los datos
recolectados de 10 mediciones, que se obtuvieron de utilizando el mtodo directo que consiste en un tanque pesaje y
cronometro y mtodo indirecto a travs de un venturi, tobera, orificio y un rotmetro conectados en serie, en objetico es
calcular las perdidas en cada uno de estos elementos conectados en serie.
Objetivos

Aplicacin de la ecuacin de bernoulli a fluido incompresible

ALUMNO: Julio Miguel Naranjo Fiallos

PARALELO:103

Comparacin de la medicin de flujo empleado:


o Un tanque de pesaje y cronometro
o Un medidor venturi
o Una tobera(difusor)
o Un medidor de placa de orificio
o Un rotmetro
Obtener el coeficiente de descarga del venturi, tobera y placa orificio

Obtener la curva de calibracin del venturi, tobera y placa orificio, tobera y rotmetro. En el cual se grafica la masa
del tanque vs la masa de medidor, para los medidores de obstruccin, y en el caso del rotmetro graficar la masa
del tanque vs la altura Y del rotmetro.

Materiales y equipos

banco de prueba
medidor venturimulti-tubular

cronometro

ALUMNO: Julio Miguel Naranjo Fiallos

PARALELO:103

Figura 1: Diagrama del banco de prueba o el aparato medidor de flujo.

ALUMNO: Julio Miguel Naranjo Fiallos

Figura 2: banco de prueba o banco hidraulico

PARALELO:103

ALUMNO: Julio Miguel Naranjo Fiallos

PARALELO:103

Figura 3: instrumento de medicin rotametro

Figura 4: estanque de llenado con su respectivo mecanismo de palanca y peso muerto


Procedimiento

Conectar el equipo y accionar el pulsador para prender la bomba hidrulica

ALUMNO: Julio Miguel Naranjo Fiallos

PARALELO:103

Extraer el aire de los tubos manomtricos por flexin de la tubera y presurizar el banco de tubos manomtricos

Ajuste la vlvula de suministro hasta que el flotador del Rotmetro se encuentre en la parte superior
aproximadamente unos 24 cm en la escala

Tomar las lecturas de los niveles manomtricos Ha, Hb,,Hf respectivamente; nivel del Rotmetro; cantidad de
agua almacenada en el tanque de pesaje para un cierto tiempo

Repita el paso 3 para diferentes razones de flujo (diferentes nivel del rotmetro)

Aplicar las formulas para obtener el coeficiente de descarga

m ideal = A2 V2

m actual KA2 2( p1 p 2 ) /

m actual = K A2 [2 (p1 p2)/](1/2)


K= C / (1- 4)(1/2)

p1 V1 2
p2 V2 2

h1

h2
1 2g
2 2g

A1V1 A2V2

Qideal k h1 h2

Cd

Qreal
Qideal

C= m actual / m ideal
= d / D = (A2 / A1)(1/2)

Donde:

m ideal, actual:Flujo msico ideal y actual. [ Kg/s]


A2 y V2: Condiciones en la salida de los medidores. [m2], [m/s]
: Densidad del fluido. [Kg/m3]
K: Coeficiente de flujo.
:Relacin de dimetros.
C:Coeficiente de descarga.
D y d: Dimetro de entrada y salida respectivamente. [m]
p1, p2: Presionesde entrada y salida. [ Pa]

Tabla datos
ha

hb

hc

hd

he

hf

Y (cm)

Cantidad de

(mm)
360
345

(mm)
100
123

(mm)
310
310

(mm)
320
320

(mm)
335
335

(mm)
60
85

22
21

Agua (Kg.)
15
15

(seg.)
40.7
43.3

ALUMNO: Julio Miguel Naranjo Fiallos

PARALELO:103

340
140
323
334
352
105
20
15
44.4
362
173
329
340
350
142
19
15
47.4
362
190
334
341
351
162
18
15
50.6
356
203
324
335
348
180
17
15
54.21
355
215
330
335
340
190
16
15
56.8
349
225
330
335
340
205
15
15
60.4
343
235
327
332
339
218
14
15
64.1
342
245
327
330
336
231
13
15
67.9
337
252
324
327
334
242
12
15
73
Tabla1: datos observados en el medidor de venturimulti-tubular, ratametro y tiempo de llenado del estanque.
Clculos representativos
Calculo de CD de venturi: se va a calcular el primer coeficiente y de ah solo se registraran los resultados en la tabla de
resultados correspondiente.

D2=16 mm
D1=26 mm
Qteorico2=V 2A 2=

(0.0262)
Qteorico2=V 2A 2=

Qteorico2=4.09 104 m/ s
real= 15
M
40.7
real=0.368 kg/s
M
teorico=4.09 104 (1000)
M
teorico=0.409 kg /s
M
CD=

0.368
0.409

CD=0.75
Calculo para el CD de la placa orificio:

( D2)

2( g)
(h2h1)
D1 4
( ) 1
D2

2(9.81)
(0.10.36)
0.016 4
(
) 1
0.026

ALUMNO: Julio Miguel Naranjo Fiallos

PARALELO:103
D e =51 mm
D f =20 mm

Qteorico2=V 2A 2=

( Df )

(0.022)
Qteorico2=V FA F =

2(g)
(hf he )
De 4
( ) 1
Df

2(9.81)
(0.3350.06)
0.02 4
(
) 1
0.051

Qteorico2=7.38 10 m/ s
real= 15
M
40.7
real=0.368 kg/s
M
teorico=7.38 104 (1000)
M
teorico=0.7385 kg /s
M
CD=

0.368
0.7385

CD=0. 5054
Calculo para el CD de la tobera
El coeficiente de descarga por efectos bidimensionales de la tobera se aplica la siguiente ecuacin obtenidas mediante
ensayos:

CD=0.998440.0996 0.2113564
Re se puede obtener a partir de

Donde:

ALUMNO: Julio Miguel Naranjo Fiallos

PARALELO:103

: Densidad del fluido


Ds : Velocidad caracterstica del fluido

D: Dimetro de la tubera a travs de la cual circula el fluido o longitud caracterstica del sistema
: Viscosidad dinmica del fluido
V: Viscosidad cinemtica del fluido (m/s)

Entonces el

coeficiente de descarga se obtiene a partir de la siguiente ecuacin:

2g

(hf ho)
Df 4
( ) 1
Do
V

CD=0.998440.0996

Se asume una temperatura de 20 grados para el coeficiente de viscosidad dinmica del agua ( =0.001003)

2(9.81)
(0.320.31)
0.051 4
(
) 1
0.026
0.001003

CD=0.998440.0996

0.026

CD=0.9199

ALUMNO: Julio Miguel Naranjo Fiallos

PARALELO:103

CALCULOS EN EXCEL

0,36
0,345
0,34
0,362
0,362
0,356
0,355
0,349
0,343
0,342
0,337

MREAL
0,368550
37
0,346420
32
0,337837
84
0,316455
7
0,296442
69
0,276701
72
0,264084
51
0,248344
37
0,234009
36
0,220913
11
0,205479
45
0,1
0,123
0,14
0,173
0,19
0,203
0,215
0,225
0,235
0,245
0,252

77,17031 0,040971
45
97

0,31
0,31
0,323
0,329
0,334
0,324
0,33
0,33
0,327
0,327
0,324

0,32
0,32
0,334
0,34
0,341
0,335
0,335
0,335
0,332
0,33
0,327

0,335
0,335
0,352
0,35
0,351
0,348
0,34
0,34
0,339
0,336
0,334

0,06
0,085
0,105
0,142
0,162
0,18
0,19
0,205
0,218
0,231
0,242

360

100

310

320

335

60

22

15

40,7

345

123

310

320

335

85

21

15

43,3

340

140

323

334

352

105

20

15

44,4

362

173

329

340

350

142

19

15

47,4

362

190

334

341

351

162

18

15

50,6

ALUMNO: Julio Miguel Naranjo Fiallos

PARALELO:103

356

203

324

335

348

180

17

15

54,21

355

215

330

335

340

190

16

15

56,8

349

225

330

335

340

205

15

15

60,4

343

235

327

332

339

218

14

15

64,1

342

245

327

330

336

231

13

15

67,9

337

252

324

327

334

242

12

15

73

diametro

CONSTANTE
RAIZ
-22,9048363

area

0,026

0,016

0,026

0,051

0,051

0,02

DIFRENCIA
VENTURI
-0,26
-0,222
-0,2
-0,189
-0,172
-0,153
-0,14
-0,124
-0,108
-0,097

0,00053092
9
0,00020106
2
0,00053092
9
0,00204282
1
0,00204282
1
0,00031415
9
RAIZ
CUADRADA
2,44033961
5
2,25496644
3
2,14031942
9
2,08062828
5
1,98485058
5
1,87201494
5
1,79071971
1,68528920
4
1,57280714
6
1,49056000

0,856587
65
5,972900
39
13,80428
07

3,2848363
1,42129824

0 -20,0952618
0,976349
56
-19,62
-1
Q
TEORICO MTEORICO
0,000490
66 0,49065939
0,000453
39
0,4533879
0,000430
34 0,43033675
0,000418
34 0,41833514
0,000399
08 0,39907789
0,000376
39 0,37639094
0,000360
05 0,36004556
0,000338
85
0,3388475
0,000316
23 0,31623164
0,000299 0,29969487

CD
0,751132
81
0,764070
5
0,785054
57
0,756464
54
0,742819
12
0,735144
47
0,733475
25
0,732908
97
0,739993
51
0,737126

ALUMNO: Julio Miguel Naranjo Fiallos

-0,085

DIFERENCIA PO
-0,275

-0,247
-0,208
-0,189
-0,168
-0,15
-0,135
-0,121
-0,105
-0,092

UN
DIFERENCIA
TOBERA
0,01
0,01
0,011
0,011
0,007
0,011
0,005

2
1,39531755
7

69
0,000280
55

RAIZ
CUADRADA

1,81553298
1,71551741
5
1,62748271
9
1,54078551
4
1,43530484
6
1,34351777
1

Q
TEORICO MTEORICO
0,000729
74 0,72973591
0,000695
78 0,69577589
0,000691
59 0,69158864
0,000634
65 0,63464526
0,000604
97 0,60496504
0,000570
37 0,57036651
0,000538
95 0,53894569
0,000511
29 0,51128878
0,000484
05 0,48405205
0,000450
91 0,45091432
0,000422
08 0,42207856

CD
0,505046
22
0,497890
67
0,488495
35
0,498633
99
0,490016
23
0,485129
67
0,490002
08
0,485722
32
0,483438
43
0,489922
59
0,486827
51

RAIZ
CUADRADA
0,11921821
4
0,11921821
4
0,12503711
7
0,12503711
7
0,09974511
4
0,12503711
7
0,08430000

RE
POTENCIA
3,090402
35 0,78782941
3,090402
35 0,78782941
3,241241
32 0,77993406
3,241241
32 0,77993406
2,585616
11 0,81809163
3,241241
32 0,77993406
2,185244 0,84770452

CD
0,919972
19
0,919972
19
0,920758
57
0,920758
57
0,916958
07
0,920758
57
0,914008

2,32282156
2,21472345
9
2,20139501
2
2,02013860
9
1,92566352
2

-0,25

PARALELO:103
76
0,732428
0,28054524
94

0,0010
03

ALUMNO: Julio Miguel Naranjo Fiallos

PARALELO:103

7
0,08430000
7
0,08430000
7
0,06529850
5
0,06529850
5

0,005
0,005
0,003
0,003

46
2,185244
46
2,185244
46
1,692683
08
1,692683
08

0,84770452
0,84770452
0,89472397
0,89472397

Tabla de resultados
CD venturi

CD tobera

CD placa
orificio

0,75113281

0,50504622
0,497890

0,91997219
0,9199721

0,7640705
0,7850545

67
0,488495

9
0,9207585

7
0,7564645

35
0,498633

7
0,9207585

99

63
0,914008
63
0,914008
63
0,909325
49
0,909325
49

ALUMNO: Julio Miguel Naranjo Fiallos

PARALELO:103

0,7428191

0,490016

0,9169580

2
0,7351444

23
0,485129

7
0,9207585

7
0,7334752

67
0,490002

7
0,9140086

5
0,7329089

08
0,485722

3
0,9140086

7
0,7399935

32
0,483438

3
0,9140086

1
0,7371267

43
0,489922

3
0,9093254

6
0,7324289

59
0,486827

9
0,9093254

51

Grfica

CURVA DE CALIBRACION PLACA ORIFICIO


0,50

MTEORICO

0,45

0,40

0,35

0,30

0,20

0,22

0,24

0,26

0,28

0,30

0,32

0,34

0,36

0,38

MREAL
FIGURA 5: Curva de calibracinVenturi, pendiente representa el coeficiente de descarga

ALUMNO: Julio Miguel Naranjo Fiallos

PARALELO:103

CURVA DE CALIBRACION VENTURI


0,75
0,70

MTEORICO_1

0,65
0,60
0,55
0,50
0,45
0,40
0,20

0,22

0,24

0,26

0,28

0,30

0,32

0,34

0,36

0,38

MREAL

FIGURA 6: Curva de calibracin Placa Orificio, pendiente representa el coeficiente de descarga

Empirical CDF of C1
Normal
Mean
0,9164
StDev 0,004473
N
11

100

Percent

80

60

40

20

0
0,905

0,910

0,915

0,920

0,925

C1

FIGURA 7: Curva emprica del coeficiente descarga de la tobera, pendiente representa el coeficiente de descarga

ALUMNO: Julio Miguel Naranjo Fiallos

PARALELO:103

curva de calibracion del llenado del tanque


0,38
0,36
0,34

MREAL

0,32
0,30
0,28
0,26
0,24
0,22
0,20
12

14

16

18

20

22

altura

FIGURA 4: Curva de calibracin del rotmetro


Anlisis de resultados
Como se pudo observar en la graficas de calibracin el coeficiente de descarga se comporta lineal en el Venturi, placa
orificio y el rotmetro, mientras que en la tobera sucede un fenmeno fsico debido a los efectos bidimensionales que
ocurren dentro de la tobera, su coeficiente de descarga se comporta como una exponencial, se puede inferir que los datos
que se obtuvieron muestran un nivel de confianza de ms del 95% y el experimento sali como lo previsto antes de
comenzar la prctica.
Conclusiones
El medidor de Venturi muli-tubular nos ayuda a calcular el coeficiente de descarga o la relacin que existe entre el flujo
msico del medidor (instrumento de medicin) y el flujo msico del tanque el cual va variando a medida que se disminuye
la altitud del rotmetro.

Recomendaciones

accionar el mecanismo de palanca de manera muy cuidadosa se puede averiar


tener en cuenta el aire que existe dentro de las mangueras de medidor de Venturi multi-tubular
tomar el tiempo de manera precisa para evitar errores en el calcula de la curva de calibracin de rotmetro

Bibliografa

Roca, Jorge. 16/11/2015 Medicin de caudal. Diapositivas.

ALUMNO: Julio Miguel Naranjo Fiallos

PARALELO:103

J. A. Cruz Maya, F. Snchez Silva, G. Tolentino Eslava, A. Gmez Mercado, I. Carvajal


Mariscal. 2015. ESTUDIO ANALTICO DEL COEFICIENTE DE DESCARGA EN
TOBERAS SNICAS TOROIDALES ISO-9300.. En Laboratorio de Ingeniera
Trmica e Hidrulica Aplicada SEPI-ESIME-IPN-COFAA (1, 6) .: Av. IPN S/N Edificio
5 3er. Piso. 07738 Mxico, D.F. Tel. (5)729-6000 ext. 54783.

Você também pode gostar