Você está na página 1de 5

Comparison of Fuel Cell Electrolyte Resistance

Measurement Techniques

One of the key performance drivers for fuel cells is the


resistance of the electrolyte. The electrolyte is the
medium in which the ions produced at one electrode
migrate to the other electrode in order to complete the Fig.1 Simplified equivalent circuit for a fuel cell

electrical circuit. The passage of ions within a phase of


finite resistance gives rise to a voltage loss referred to Note that the values of these equivalent circuit
as ohmic polarization. Regardless of whether one components are a function of the cell's operating
considers the electrolyte to be the solid polymer in a current or voltage, making the fuel cell an electrically
proton exchange membrane (PEM) fuel cell, a non-linear device. The sum of the electrolyte and
high-temperature ionically-conductive oxide in a solid polarization resistances can be determined from the
oxide fuel cell (SOFC), or an acid-soaked matrix in slope of the linear portion of the DC voltage-current
phosphoric acid fuel cell (PAFC), electrolyte resistance (V-I) characteristic curve (polarization curve) of the
is an important metric for fuel cell performance. This cell. The electrolyte resistance is a particularly
article examines and compares presently available important measure of single fuel cell (or fuel cell stack)
methods for measuring the resistance of the electrolyte electrical performance since it quantifies internal cell
within a fuel cell. The fuel cell can be modeled by the losses. It is desirable to monitor electrolyte resistance
equivalent circuit shown in Figure 1. This is a during electrolyte development and subsequent
simplified version of the Randles equivalent circuit, a manufacture of stacks because ohmic losses generate
model commonly applied to electrochemical systems waste heat that must be removed from the fuel cell,
in which contact resistance and other effects are small resulting in a decrease in overall electrical efficiency.
enough to ignore. For simplicity, assume that the In addition, because fuel cell current densities are quite
polarization resistance of one electrode (say the high in comparison to other electrochemical processes,
cathode) is much larger than that of the other electrode even small amounts of ohmic resistance (milliohms)
(the anode), so that one can legitimately omit circuit have a significant effect on overall efficiency.
elements associated with one of the electrodes (the Unfortunately, electrolyte resistance cannot be directly
anode in this example). Polarization Resistance is the measured by conventional DC methods when installed
reaction equivalent, Double-Layer Capacitance is the in a fuel cell, nor can DC methods isolate electrolyte
interfacial capacitance of the cathode and Electrolyte resistance from polarization resistance. Furthermore,
Resistance is the resistive component of the fuel cell to because the resistivity of the electrolyte is a function of
be evaluated. The Voltage Source element is an ideal the operating conditions (temperature, current density,
DC voltage source (zero internal impedance, constant etc.) it is imperative that its resistance be determined
voltage) with a potential equal to the open-circuit under operating conditions.
voltage of the fuel cell. This element does not affect Fortunately, the electrode / solid electrolyte interface
AC analysis but allows the model to approximate the has a large capacitance associated with it (the
DC behavior of the fuel cell. Double-Layer Capacitance in Figure 1) that allows an
AC measurement to be used to determine the
Page 1 of 5
Comparison of Fuel Cell Electrolyte Resistance Measurement Techniques
Reedited from April/May 2005 issue of Fuel Cell Magazine
electrolyte resistance separately from the polarization
resistance. This may be performed several different
ways, but all have some traits in common:
1. All methods impose a changing electrical
condition on the fuel cell.
2. All methods measure current or voltage
waveforms resulting from that change.
3. All methods require an accurate voltage
measurement directly at the cell terminals using a Fig.2 Typical current interrupt waveform

four-terminal (Kelvin) method.


The four methods generally used for electrolyte AC Resistance Method
resistance measurement are current interrupt (iR), AC This method uses an AC resistance measurement
resistance, electrochemical impedance spectroscopy device, such as an external AC milliohm meter, to
(EIS) and high frequency resistance (HFR) methods. A apply a fixed, single high frequency sine wave
comparison of these methods follows a discussion of (typically ~1 kHz) to the fuel cell under test to measure
the measurement principle, and the pros and cons of the total impedance magnitude of the cell and the load
each. in parallel at that frequency. The set-up is shown in
Figure 3. The ohmic resistance of the electrolyte can be
Current Interrupt Method extracted after correcting for the impedance of the
This time-domain AC technique, developed over 50 load.
years ago, quickly interrupts the fuel cell current and
rapidly measures the terminal voltage before and
during the interruption. This is shown schematically in
Figure 2. Because the cell voltage is a combination of
the charged anode and cathode potentials less the
resistive drop of the electrolyte, the cell voltage rises
by the amount of the drop across the electrolyte. The Fig.3 Equipment set-up for electrolyte resistance by AC Resistance

cell ohmic resistance, dominated by the electrolyte, is measurement technique. Note that the electronic load is in

the ratio of the change in voltage (ΔV) due to the parallel with the fuel cell and therefore its impedance contributes

interrupt to the cell current. Advantages of this method to the observed AC resistance.

include a single data value which is easily interpreted.


Furthermore, there is no requirement for additional Like the current interrupt technique, this method only
equipment because the interrupt is brought about by provides a single data point. An advantage is that the
the load. The primary disadvantage of the iR method is cell is minimally disturbed electrochemically by the
that it imposes a significant perturbation on the cell, if measurement, but unfortunately accurate results cannot
only for a short duration (~ 10s μsec). Users of this be obtained without exact gain-phase characterization
method are also cautioned that iR data is degraded of the impedance of the electronic load at the operating
when long cell cables are used due to 'ringing' caused conditions of the fuel cell during the AC measurement
by cable inductance. (DC voltage, DC current and frequency) using
Page 2 of 5
Comparison of Fuel Cell Electrolyte Resistance Measurement Techniques
Reedited from April/May 2005 issue of Fuel Cell Magazine
additional external frequency analysis equipment.
Additionally, the zero-phase condition of the parallel
load-fuel cell combination measured by the milliohm
meter is not meaningful since it is not the zero-phase
impedance of the cell (the load has a complex
impedance in parallel with the complex impedance of
the cell). The only way to separate the actual cell Fig.4 Equipment set-up for HFR and EIS measurement

impedance is to use the measured load impedance and techniques. The AC signal imposed on the cell is done by

reported magnitude and phase data of the milliohm modulating the load

meter to calculate the impedance of the cell itself. The


difficulties of this technique stem from the millohm The optimum measurement frequency can vary with
meter not being intended to measure energy sources the electrochemical system under study and its
under load. selection must be made with care. Selection of the
proper frequency is best accomplished by examining
High Frequency Resistance Method (HFR) the phase difference between the applied current and
In the HFR method to determine electrolyte resistance, measured voltage. When this difference is zero, the cell
a small AC signal is applied to the electronic load to is functioning in purely resistive manner. It should
modulate the DC load current, as illustrated in Figure 4. ideally be the frequency at which imaginary
The resulting magnitude and phase of the AC voltage component of the impedance is zero (Z'' = 0). In terms
and current response are measured by a frequency of a Nyquist plot, this condition exists when the
response analyzer. A single, high frequency is used, impedance data cross the real axis (Figure 5) at high
typically on the order of 1 kHz. This method is actually frequency (small Z'). If the Nyquist plot does not cross
a subset of the method described below wherein a the Real axis at high frequency, the highest frequency
broad range of frequencies are employed. It is the Real before the plot deviates from the semicircular shape
(Z') component of the impedance that is of interest, may be used or the intercept of the data fit to a
usually displayed in milliohms. Looking at the semicircle or other model may be used. Typical HFR
equivalent circuit in Figure 1, the double-layer measurement frequencies range from 500 Hz to 3 kHz.
capacitance will act as a very low resistance for high In any case, the same frequency must be used for valid
frequency signals. If a high frequency current is data comparison.
applied and the corresponding voltage measured, the Note that the method for choosing the HFR frequency
resistance of the electrolyte is the ratio of these values. requires that the test system also have Electrochemical
The HFR measurement minimally disturbs the cell Impedance Spectroscopy capability. This is generally
from its operating condition, both in magnitude and not a problem because a true frequency response
duration, and therefore it is suitable for routine, analyzer can measure over a wide range of frequencies,
periodic application during normal fuel cell operation so a test system capable of true HFR measurement will
with the result displayed in real time. also be capable of performing EIS measurements.

Electrochemical Impedance Spectroscopy (EIS)


Method
Page 3 of 5
Comparison of Fuel Cell Electrolyte Resistance Measurement Techniques
Reedited from April/May 2005 issue of Fuel Cell Magazine
The EIS technique is an extension of the HFR method processes within the cell. Disadvantages include the
previously described and differs in two ways. Whereas need to interpret the spectra produced and the cost of
HFR employed measurement at a single frequency and the equipment required to make the measurement.
generally only examined the real component of the
impedance, EIS involves imposing the AC perturbation Comparison of Techniques
over a broad range of frequencies - typically 10 kHz to The objective for each method is to determine the
1 Hz or lower - and monitoring the resulting variations electrolyte resistance of the fuel cell. The two
in magnitude and phase of the cell voltage and current techniques most easily compared are the Current
with a frequency response analyzer in order to Interrupt and HFR methods. In practice, the results
determine the complex impedance (Z' and Z'') of the from these two techniques usually correlate well if the
fuel cell. This results in a rich data set from which HFR measurement frequency is properly chosen. There
several parameters may be extracted. The amplitude are, however, inherent differences in the two methods:
and phase of the signals may be plotted in Bode and the current interrupt method introduces a large
Nyquist formats for analysis and modeling. Figure 5 perturbation to the fuel cell and looks at its
shows an idealized Nyquist plot of the equivalent time-domain response where as the HFR method
circuit shown in Figure 1. This method may be used to applies a small signal and uses the frequency domain
determine electrolyte resistance as well as provide response of the cell. Discrepancies between results
information about kinetics and mass transport within obtained form these two methods derive from whether
the fuel cell. the current distribution present during the current
interruption or impedance measurement is the same as
the current distribution during standard DC operation
of the cell. In the current interrupt technique, after the
interrupt, the "true" cell voltage is only measured if the
current is zero everywhere within the cell and on the
surface of the electrodes. This condition exists for a
uniform double layer charge. If however, the electrode
double layer was non-uniformly charged when the
Fig.5 Idealized Nyquist plot of the impedance of a fuel cell current was flowing, then after the interruption, current
will flow through the electrolyte from one part of the
Unlike the previously mention techniques, EIS electrode surface to the other to redistribute the surface
provides the advantage of a large amount of potentially charge. This current flow just after the interruption will
useful information about the cell. Like the AC and create an ohmic voltage drop within the cell which will
HFR resistance method, the cell under test is introduce an error in the resistance measurement. That
minimally disturbed from its operating condition and is, the measured
therefore not electrochemically changed by the ΔV will not equal the true voltage drop due to ohmic
measurement. Modeling software is available to allow resistance when the cell is operating under DC current.
accurate analysis of the electrical and reaction This artifact is most likely to occur for primary current
characteristics of the fuel cell which can provide distributions. If the polarization resistance of the
insight into fundamental kinetic and transport electrodes is large relative to the electrolyte resistance,
Page 4 of 5
Comparison of Fuel Cell Electrolyte Resistance Measurement Techniques
Reedited from April/May 2005 issue of Fuel Cell Magazine
the current distribution is more uniform and the of the Electrochemical Society, 117, 507 (1970).

measured voltage error will be fairly accurate. 2. C. Lagergren, G. Lindbergh and D. Simonsson, "Investigation of Porous

Knowledge of the relative magnitude of the Electrodes by Current Interruption," Journal of the Electrochemical Society,

polarization resistance and electrolyte resistance for a 142, 787 (1995).

given fuel cell system is beneficial when evaluating the 3. F.N. Buchi, A. Marek and G.G. Scherer, "In Situ Membrane Resistance

applicability of the interrupt method. Measurements in Polymer Electrolyte Fuel Cells by Fast Auxiliary Current

In the HFR method, the electrolyte resistance is Pulses," Journal of the Electrochemical Society, 142, 1895 (1995).

determined from the real component of the impedance 4. C.-G. Lee, H. Nakano, T. Nishina, I. Uchida and S. Kuroe, "Characterization

measured at high frequency. However, total electrolyte of a 100 cm2 Class Molten Carbonate Fuel Cell with Current Interruption,"

resistance may change with current redistribution. Journal of the Electrochemical Society, 145, 2747 (1998).

Therefore, the AC signal should be small relative to the 5. Springer, Zawodzinski, Wilson and Gottesfeld, "Characterization of Polymer

DC current such that imposition of the AC signal on electrolyte Fuel Cells using AC Impedance Spectroscopy," Journal of the

the cell does not itself significantly influence the Electrochemical Society, 143, 587 (1996).

current distribution. The resistance measured in this 6. M. Lefebvre, R. Martin and P. Pickup "Characterization of Ionic Conductivity

instance will be similar to that experienced under pure Profiles within Proton Exchange Membrane Fuel Cell Diffusion Electrodes by

DC conditions and the electrolyte resistance Impedance Spectroscopy," Electrochemical and Solid State Letters, 2, 259

determined with the impedance method will be (1999).

accurate. The Electrochemical Impedance Method 7. The Measurement and Correction of Electrolyte Resistance in Electrochemical

produces a spectrum of data, a superset of that from the Tests, ASTM STP 1056, L. Scribner, S.R. Taylor, Eds., 1990.

HFR method. The Real (Z') impedance measured using 8. H.P. Hack, P.J. Moran, and J.R. Scully, "Influence of Electrolyte Resistance on

EIS at the frequency used for an HFR measurement Electrochemical Measurements and Procedures to Minimize or Compensate

should be identical to that for the latter. The data from for Resistance Errors," in The Measurement and Correction of Electrolyte

the AC Resistance method is not directly comparable Resistance in Electrochemical Tests, ASTM STP 1056, L. Scribner, S.R. Taylor,

to that produced by the other methods due to the Eds., ASTM International: Philadelphia, PA (1990).

artifacts imposed by the parallel-connected electronic


load.

Conclusion
Each of the four methods described may be used to
determine the ohmic resistance of the fuel cell
electrolyte, and in case of EIS, valuable information
about other cell parameters. However, users of these
techniques should be cognizant of differences in these
methods in order to properly apply and interpret the
results if accurate and useful measurement of cell
resistance is to be obtained.

References

1. J. Newman, "Ohmic Potential Measured by Interrupter Techniques" Journal


Page 5 of 5
Comparison of Fuel Cell Electrolyte Resistance Measurement Techniques
Reedited from April/May 2005 issue of Fuel Cell Magazine

Você também pode gostar