Você está na página 1de 32

lz

o
h
a
V
s
n
m
ie
S

AXIOM Iconos R200


AX

Replacements of Parts
AXIOM Iconos R200
Replacement of the Lift-Tilting Base

Also applicable for:


- AXIOM Luminos dRF

Siemens, 2010

10094158
05902619
07152460
07030930
07030518
07030500

Siemens, 2010
All documents may only be used for rendering
services on Siemens Healthcare Products. Any
document in electronic form may be printed
once. Copy and distribution of electronic documents and hardcopies is prohibited. Offenders
will be liable for damages. All other rights are reserved.

Print No.:
Replaces:

AXD3-340.841.03.02.02
AXD3-340.841.03.01.02

English
Doc. Gen. Date: 04.10
CS PS AX R-F
n.a.

2010

Copyright / Version / Disclaimer

1Copyright / Version / Disclaimer

Copyright
Siemens, 2010 refers to the copyright of a Siemens entity such as Siemens Aktiengesellschaft - Germany, Siemens Shenzhen Magnetic Resonance Ltd. - China, Siemens
Shanghai Medical Equipment Ltd. - China, Siemens Medical Solutions USA Inc. - USA,
Siemens Healthcare Diagnostics Inc. - USA and/or Siemens Healthcare Diagnostics Products GmbH - Germany.
Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents correspond to the version at the time of system delivery and/or printout. Versions to hardcopy
documentation are not automatically distributed. Please contact your local Siemens office
to order current version or refer to our website http://www.healthcare.siemens.com.
Disclaimer
Siemens provides this documentation as is without the assumption of any liability under
any theory of law.
The installation and service of equipment described herein requires superior understanding of our equipment and may only be performed by qualified personnel who are specially
trained for such installation and/or service.

AXIOM Iconos R200

AXD3-340.841.03.02.02
04.10

Page 2 of 32
CS PS AX R-F

Siemens, 2010

Table of Contents

0Table of Contents

1 _______ Removing the Luminos dRF Unit Support ___________________________ 4


Preparatory Measures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Cover Panels on the Lifting Base . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Cover Panels on the Extender . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Cover Panels on the Unit Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Unit Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Cabling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Installing the Transport Carriage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

2 _______ Removing the Iconos R200 Unit Support ___________________________ 15


Cover Panels on the Lifting Base . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Cover Panels on the Extender . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Cover Panels on the Unit Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Unit Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Cabling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Installing the Transport Carriage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

3 _______ Removing the Lifting Base _______________________________________ 25


Removing the Extender . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Removing the Lifting Base . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

4 _______ Installation and Adjustment ______________________________________ 27


Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Adjusting the Safety Limit Switches: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Adjusting the Potentiometers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Adjustments with the Luminos dRF with Bucky Wall Stand . . . . . . . . . . . . . . . . . . .
Installing the Cover Panels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

28
28
28
28
29

5 _______ ______________________________________________________________ 30
6 _______ List of Hazard IDs ______________________________________________ 31

Siemens, 2010

AXD3-340.841.03.02.02
04.10

Page 3 of 32
CS PS AX R-F

AXIOM Iconos R200

Removing the Luminos dRF Unit Support


1-

Preparatory Measures

1Removing the Luminos dRF Unit Support

If the Luminos dRF system has a Bucky wall stand, the position of the lifting base must be
marked as exactly as possible on the floor.
The new lifting base must then be installed again in exactly (millimeter accuracy) the same
position so that center of the tube <-> detector center exactly match.
It is absolutely necessary to work very accurately here, because correction of a difference
takes a lot of time.

AXIOM Iconos R200

AXD3-340.841.03.02.02
04.10

Page 4 of 32
CS PS AX R-F

Siemens, 2010

Removing the Luminos dRF Unit Support

Cover Panels on the Lifting Base

1.1

1. Remove the top cover panel.


2. Remove the side cover panels.

Fig. 1:

Lifting base cover panel

Pos. 1

Front cover panel screws

Pos. 2

Cable cover panel

Pos. 3

Free holes

Siemens, 2010

AXD3-340.841.03.02.02
04.10

Page 5 of 32
CS PS AX R-F

AXIOM Iconos R200

Removing the Luminos dRF Unit Support

Cover Panels on the Extender

1.2

Remove the two cover panels on the extender (1/Fig. 2 / p. 6).


Remove the side cover panel on the extender (1/Fig. 3 / p. 6).
Remove the cover panel on the extender mount (2/Fig. 3 / p. 6).

Fig. 3:
Fig. 2:

AXIOM Iconos R200

Cover panels on the extender, 2

Cover panels on the extender, 1

AXD3-340.841.03.02.02
04.10

Page 6 of 32
CS PS AX R-F

Siemens, 2010

Removing the Luminos dRF Unit Support

Cover Panels on the Unit Support

1.3

1. Remove the footrest and the other parts on the patient table.
2. Remove the side cover panels on the unit support (1-3/Fig. 4 / p. 7).
3. Remove the fan cover panel at the head end on the unit support (1/Fig. 5 / p. 7).

Fig. 4:

Siemens, 2010

Cover panels at foot end

AXD3-340.841.03.02.02
04.10

Fig. 5:

Cover panels at head end

Page 7 of 32
CS PS AX R-F

AXIOM Iconos R200

Removing the Luminos dRF Unit Support

Unit Position

1.4

Move the unit into the following position:

Position the spotfilm device and column so they are centered.


Move the unit into the 0 position.
Position the tabletop so it is flush with the frame.
Move the lifting base with the SSW or mechanically.
Set a distance of 455 mm from the bottom edge of the extender to the top edge of
the floor plate.

Fig. 6:

AXIOM Iconos R200

Distance between the floor plate - extender

AXD3-340.841.03.02.02
04.10

Page 8 of 32
CS PS AX R-F

Siemens, 2010

Removing the Luminos dRF Unit Support

Cabling

1.5

NOTE

The illustrations in this section are for orientation only and


can differ from the actual status of the system.

Remove the cable clamp and flexible plastic chain on the extender.

Fig. 7:

Cable deflector

Remove the cable ties from the wiring harness on the side of the extender (Fig. 8 / p. 9).

Fig. 8:

Siemens, 2010

Laying the cables in the extender

AXD3-340.841.03.02.02
04.10

Page 9 of 32
CS PS AX R-F

AXIOM Iconos R200

10

Removing the Luminos dRF Unit Support


Unplug the connections between the lifting base and the M1 assembly (line power
assembly).

Fig. 9:

Remove the contact protector from the F1 trip breaker (1/Fig. 10 / p. 10).

Fig. 10: Power supply assembly shielding

Remove the 2 mounting nuts for the line power assembly.

AXIOM Iconos R200

AXD3-340.841.03.02.02
04.10

Page 10 of 32
CS PS AX R-F

Siemens, 2010

Removing the Luminos dRF Unit Support

11

Label the motor cable to the Micromasters (1/Fig. 11 / p. 11) and disconnect it.

Fig. 11:

Remove the transformer (2/Fig. 11 / p. 11). It will be used again.


Pull all cables to the right side of the base in the direction of the transformer out of the
base. Be careful of the connectors (Fig. 9 / p. 10).

Remove the support plate (3/Fig. 11 / p. 11) with the Micromasters. Take note of the
cabling! The support plate with the Micromasters will be used again.

Siemens, 2010

AXD3-340.841.03.02.02
04.10

Page 11 of 32
CS PS AX R-F

AXIOM Iconos R200

12

Removing the Luminos dRF Unit Support

Installing the Transport Carriage

1.6

1. Install the plates at the head and foot ends of the unit support and tighten all screws.

Fig. 12: Panel 1 at head end


Fig. 13: Panel 2 at head end

Fig. 14: Panel 1 at foot end


Fig. 15: Panel 2 at foot end

AXIOM Iconos R200

AXD3-340.841.03.02.02
04.10

Page 12 of 32
CS PS AX R-F

Siemens, 2010

Removing the Luminos dRF Unit Support

13

2. Install the transport carriages at the head and foot ends (thread the screws in only
loosely) and make sure that the safety screws to prevent tilting of the unit support are
inserted:

Fig. 16: Transport carriage


Fig. 17: Transport carriage, detail

3. Install the transverse struts (thread in the screws only loosely). Prior to installation, the
transverse struts must be connected together (tighten the screws).
4. First tighten all screws on the transport carriage and then the screws for the transverse
struts.
5. Install the transport castors at the head and foot ends.
6. Turn down the transport castors with the spindles so that the transport castors are
under light pressure.
7. Remove the screws between the unit support and the extender.
8. Evenly turn down the transport castors with the spindles so that the unit support and
extender come apart.

Siemens, 2010

AXD3-340.841.03.02.02
04.10

Page 13 of 32
CS PS AX R-F

AXIOM Iconos R200

14

Removing the Luminos dRF Unit Support


9. Move the unit support so that the base is accessible from the front and side. Pay careful attention to the cables and take them out of the extender!

Fig. 18: Unit supporet / extender

10. Carefully pivot the unit support at the foot end until the extender is accessible with the
lift truck and pallet. Be extremely careful of the cables!
11. Completely remove the M1 assembly (line power assembly) from the lifting base and
place it carefully to the side.

AXIOM Iconos R200

AXD3-340.841.03.02.02
04.10

Page 14 of 32
CS PS AX R-F

Siemens, 2010

Removing the Iconos R200 Unit Support


2-

Cover Panels on the Lifting Base

2Removing the Iconos R200 Unit Support

15
0

1. Remove the top cover panel.


2. Remove the side cover panels (if installed).

Fig. 19: Lifting base cover panel


Pos. 1

Front cover panel screws

Pos. 2

Cable cover panel

Pos. 3

Free holes

Siemens, 2010

AXD3-340.841.03.02.02
04.10

Page 15 of 32
CS PS AX R-F

AXIOM Iconos R200

16

Removing the Iconos R200 Unit Support

Cover Panels on the Extender

2.1

Remove the cover panels on the extender in the following sequence:


1. Top extender cover panels (2 pieces) (1/Fig. 20 / p. 16)
2. Left side cover panel (2/Fig. 20 / p. 16)
3. Bottom (3/Fig. 20 / p. 16)
4. Right side cover panel (4/Fig. 20 / p. 16)
5. Top cover (5/Fig. 20 / p. 16)

Fig. 20: Cover panels on the extender

AXIOM Iconos R200

AXD3-340.841.03.02.02
04.10

Page 16 of 32
CS PS AX R-F

Siemens, 2010

Removing the Iconos R200 Unit Support

17

Cover Panels on the Unit Support

2.2

2
3

2
1

Fig. 21: Unit support cover panels, left

Fig. 22: Unit support cover panels, right

Remove the side cover panel (1/Fig. 21 / p. 17).


Remove the two cover panel sections on the foot end of the unit support
(1/Fig. 22 / p. 17). Remove the bottom cover panel first.

Siemens, 2010

AXD3-340.841.03.02.02
04.10

Page 17 of 32
CS PS AX R-F

AXIOM Iconos R200

18

Removing the Iconos R200 Unit Support

Unit Position

2.3

Move the units into the following positions:

Position the spotfilm device and column so they are centered.


Move the unit into the 0 position.
Position the tabletop so it is flush with the frame.
Move the lifting base with the SSW or mechanically.
Set a distance of approx. 80 cm from the top edge of the extender to the top edge
of the floor plate (Fig. 23 / p. 18).

Fig. 23:

AXIOM Iconos R200

AXD3-340.841.03.02.02
04.10

Page 18 of 32
CS PS AX R-F

Siemens, 2010

Removing the Iconos R200 Unit Support

19

Cabling

2.4

Remove the cable deflector on the extender (1/Fig. 24 / p. 19).

1
2

Fig. 24:

Remove the cable ties from the wiring harness on the side of the extender
(Fig. 25 / p. 19).

Fig. 25: Laying the cables in the extender

Siemens, 2010

AXD3-340.841.03.02.02
04.10

Page 19 of 32
CS PS AX R-F

AXIOM Iconos R200

20

Removing the Iconos R200 Unit Support


Unplug the connections between the lifting base and the M1 assembly (line power
assembly).

Fig. 26:

Remove the contact protector from the F1 trip breaker (1/Fig. 27 / p. 20).

Fig. 27: Power supply assembly shielding

Remove the 2 mounting nuts for the line power assembly.

AXIOM Iconos R200

AXD3-340.841.03.02.02
04.10

Page 20 of 32
CS PS AX R-F

Siemens, 2010

Removing the Iconos R200 Unit Support

21

Label the motor cable to the Micromasters (1/Fig. 28 / p. 21) and disconnect it.

Fig. 28:

Remove the transformer (2/Fig. 28 / p. 21). It will be used again.


Pull all cables to the right side of the base in the direction of the transformer out of the
base. Be careful of the connectors (Fig. 26 / p. 20).

Remove the support plate (3/Fig. 28 / p. 21) with the Micromasters. Take note of the
cabling! The support plate and the Micromaster will be used again.

Siemens, 2010

AXD3-340.841.03.02.02
04.10

Page 21 of 32
CS PS AX R-F

AXIOM Iconos R200

22

Removing the Iconos R200 Unit Support

Installing the Transport Carriage

2.5

1. Install the plates at the head and foot ends of the unit support and tighten all screws.

Fig. 30: Panel, head end


Fig. 29: Panel, foot end

2. Install the transport carriages at the head and foot ends (thread the screws in only
loosely) and make sure that the safety screws to prevent tilting of the unit support are
inserted:

Fig. 31: Transport carriage 1

Fig. 32: Transport carriage 2

3. Install the transverse strut ((1/Fig. 33 / p. 23) and (1/Fig. 34 / p. 23)) and angled struts (
(2/Fig. 33 / p. 23) and (2/Fig. 34 / p. 23)) (thread in the screws only loosely). Prior to
installation, the transverse struts must be connected together (tighten the screws).
NOTE

AXIOM Iconos R200

If two transverse struts are installed in place of angled struts,


the two transverse struts must be installed.

AXD3-340.841.03.02.02
04.10

Page 22 of 32
CS PS AX R-F

Siemens, 2010

Removing the Iconos R200 Unit Support

Fig. 33: Transverse and angled strut, head


end

23

Fig. 34: Transverse and angled strut, foot end

4. First tighten all screws on the transport carriage and then the screws for the transverse
strut/struts and angled struts (if installed).
5. Install the transport castors at the head and foot ends.
6. Turn down the transport castors with the spindles so that the transport castors are
under light pressure.
7. Remove the screws between the unit support and the extender.
8. Evenly turn down the transport castors with the spindles so that the unit support and
extender come apart.
9. Move the unit support so that the base is accessible from the front and side. Pay careful attention to the cables and take them out of the extender!

Fig. 35: Unit supporet / extender

10. Carefully pivot the unit support at the foot end until the extender is accessible with the
lift truck and pallet. Be extremely careful of the cables!

Siemens, 2010

AXD3-340.841.03.02.02
04.10

Page 23 of 32
CS PS AX R-F

AXIOM Iconos R200

24

Removing the Iconos R200 Unit Support


11. Completely remove the M1 assembly (line power assembly) from the lifting base and
place it carefully to the side.

AXIOM Iconos R200

AXD3-340.841.03.02.02
04.10

Page 24 of 32
CS PS AX R-F

Siemens, 2010

Removing the Lifting Base

25

3-

Removing the Extender

3Removing the Lifting Base

Fig. 37:

Fig. 36:

Siemens, 2010

Install the transport carriage on the transport pallet (2/Fig. 36 / p. 25).


Install the transport castors (1/Fig. 36 / p. 25) and (1/Fig. 37 / p. 25).
Position the transport pallet under the extender.
Raise the transport castors with the spindles (Fig. 37 / p. 25).
Remove all Allen mounting screws from the extender (2/Fig. 37 / p. 25).
Pull down the extender from the rotation axis onto the transport pallet.
Move the extender away with the transport pallet and place it down on the floor.
Remove the transport carriage and transport castors.

AXD3-340.841.03.02.02
04.10

Page 25 of 32
CS PS AX R-F

AXIOM Iconos R200

26

Removing the Lifting Base

Removing the Lifting Base

3.1

Install the transport safety device (1/Fig. 38 / p. 26) from the new lifting base in the old
lifting base.

Fig. 38: Lifting base transport safety device

Remove all floor mounting screws from the lifting base.


Install the transport carriage on the lifting base (1/Fig. 39 / p. 26).

Fig. 39:

Install the transport castors.


Raise the lifting base and move it away.

AXIOM Iconos R200

AXD3-340.841.03.02.02
04.10

Page 26 of 32
CS PS AX R-F

Siemens, 2010

Installation and Adjustment

27

4-

Installation

4Installation and Adjustment

NOTE

Shipped status of the lifting base:


The new lifting base is shipped with floor mounting materials
and a complete set of cover panels.
The electronic assemblies should be taken from the old lifting base (transformer, Micromaster assembly and line power
assembly).
The safety limit switches are pre-set (for a unit with 40 cm I.I.)
(Iconos R200).

NOTE

With a Luminos dRF system with Bucky wall stand, the new
lifting base must be positioned exactly (millimeter accuracy)
in the same position as the old lifting base! This is prerequisite for centering the Bucky wall stand.

Installation of the lifting base must be performed per the Installation Instructions.
When doing this, the following points absolutely must be observed:
- Leveling
- Inserting shims
- Torque values for the mounting bolts.
Installation of the electronic assemblies in the lifting base is performed in the reverse
sequence.

NOTE

Shielding of the motor cables absolutely must be the same as


in the "old" lifting base.
It is possible that insulation must be removed at the particular location.

Installation of the lifting base and the unit support must be performed per the Installation
Instructions.

Siemens, 2010

AXD3-340.841.03.02.02
04.10

Page 27 of 32
CS PS AX R-F

AXIOM Iconos R200

28

Installation and Adjustment

Adjustment

4.1

Adjusting the Safety Limit Switches:

Luminos dRF: Perform adjustment of the safety limit switches per the chapter Lifting
Base > Replacing the Safety Limit Switches in the instructions Replacement of
Parts, Luminos dRF, Basic Unit with Tube Unit and Collimator, Print Number
AXD3-500.841.01 (for the latest version, see the CB-DOC).

Iconos R200: Perform adjustment of the safety limit switches per the chapter Lifting
Base > Adjustment of the Safety Limit Switches in the instructions Replacement of
Parts, Iconos R200, Basic Unit and Collimator, Print Number AXD3-340.841.01 (for
the latest version, see the CB-DOC).

Adjusting the Potentiometers

Adjust the potentiometers accordingly in the SSW.


- Adjustment/ Calibration/ Tilt Drive/ Tilt Drive: Potentiometer Adjustment
- Adjustment/ Calibration/ Tilt Drive/ Tilt Drive: Potentiometer Adjustment
- Adjustment/ Calibration/ Tilt Drive/ Tilt Drive: Limitation of + 90 Position
or
- Adjustment/ Calibration/ Tilt Drive/ Tilt Drive: Limitation of + 45 Position
- Adjustment/ Calibration/ Tilt Drive/ Tilt Drive: Limitation of - 90 Position
or
- Adjustment/ Calibration/ Tilt Drive/ Tilt Drive: Limitation of - 45 Position
- Adjustment/ Calibration/ Tabletop, longitudinal/ Patient mounting height: Lowered
footrest in + 90 or - 90 position

Adjustments with the Luminos dRF with Bucky Wall Stand

Place a ball phantom on the centering cross of the Bucky wall stand and trigger an
exposure with the tube unit of the Luminos dRF unit. If the ball in the exposure is outside of the center, a correction is necessary. This can be done only by moving the
Bucky wall stand or lifting base of the Luminos dRF (unit support and extender must be
removed again).

Corrections and ball phantom exposures must be repeated until the ball in the exposure is exactly in the center of the phantom.

AXIOM Iconos R200

AXD3-340.841.03.02.02
04.10

Page 28 of 32
CS PS AX R-F

Siemens, 2010

Installation and Adjustment

29

Installing the Cover Panels

Install all cover panels in the reverse sequence.


- Install all new lifting base cover panels included in the shipment.

Siemens, 2010

AXD3-340.841.03.02.02
04.10

Page 29 of 32
CS PS AX R-F

AXIOM Iconos R200

30
5

5-

No changes. Initial publication.

AXIOM Iconos R200

AXD3-340.841.03.02.02
04.10

Page 30 of 32
CS PS AX R-F

Siemens, 2010

List of Hazard IDs


6List of Hazard IDs

31

6-

There are no Hazard IDs in this document.

Siemens, 2010

AXD3-340.841.03.02.02
04.10

Page 31 of 32
CS PS AX R-F

AXIOM Iconos R200

32

AXIOM Iconos R200

List of Hazard IDs

AXD3-340.841.03.02.02
04.10

Page 32 of 32
CS PS AX R-F

Siemens, 2010

Você também pode gostar