A T
Applications water,
wastewater and more
Principle of operation
*Patent pending.
Method of analysis
The CL17 Chlorine Analyzer uses an aqueous
buffered colorimetric indicator N,N-diethyl-pphenylenediamine (DPD) to determine levels
of chlorine. DPD turns a magenta color in
response to the amount of free residual
chlorine (as hypochlorous acid or hypochlorite
ion). This reaction takes place at a buffered pH
of 6.3 to 6.6. By simply changing reagents, the
CL17 also measures total residual chlorine (free
residual chlorine plus mono-, di- and
trichloramines). This is done by adding
potassium iodide to the sample chloramines
Sample Inlet
Sample Valve
(Under Pressure)
Light Source
To Waste
Sample Cell
Buffer
Indicator
Peristaltic
Pump
Photo Cell
Magnetic
Stirrer
CL 17 Chlorine Analyzer
Specifications*
Installation
Range
Accuracy
Precision
0.035 mg/L
Cycle time
Sample Conditioning
Inlet Pressure
1 to 75 psig
Interferences
Relay/Alarm
Contact
Connections
MENU
Chlorine
mg/L
g/L
EXIT
Network and
4-20 mA
Connections
ENTER
Alarm
90% at 40C
4.02"
(102.17 mm)
2.66"
(67.55 mm)
5C to 40C
4.09"
(103.88 mm)
1.56"
(39.68 mm)
Operating Temperature
Power
Connections
Recorder Outputs
Alarms
Power
4.38"
(111.12 mm)
SIDE VIEW
FRONT VIEW
Drain Connection
19.95"
(506.78 mm)
Certification
CE approved
ETL listed to UL 1262
ETL certified to CSA 22.2 No. 142
R 14.89
(378.18 mm)
Enclosure
Overflow Drain
Recommended
.50 I.D. Tubing
Dimensions
Mounting
7.06"
(179.39 mm)
Wall mount
Shipping Weight
25 lbs. (11.3 kg)
BOTTOM VIEW
How to Order
Hach CL17 Chlorine Analyzers are shipped with
46436-00
OD tubing
44278-00
54400-01
54400-02
54400-03
Additional Reagents
One unit each of three reagents (DPD indicator
powder**, indicator and buffer solutions) is
sufficient for 30 days of operation. Reagent
sets include all three of the required reagents,
sufficient for a 30-day operating period.
Chlorine Analyzer
Catalog #
Reagent
25569-00
Chlorine Analyzer
25570-00
22972-55
HACH COMPANY
P.O. Box 389
Loveland, Colorado 80539-0389
U.S.A.
Telephone: 800-227-4224
Fax: 970-669-2932
E-mail: orders@hach.com
www.hach.com
23140-11
23141-11
22634-11
22635-11
473 mL
Accessory
46964-00
47439-00
54443-00
54443-01
473 mL
Solution, 473 mL
473 mL
* Kit includes tubing, caps, funnel and fittings.
**DPD indicator powder is added to indicator solution.
Hach Company is
ISO 9001 Certified
Other inquiries:
HACH COMPANY
P.O. Box 389
Loveland, Colorado 80539-0389
U.S.A.
Telephone: 970-669-3050
Fax: 970-669-2932
E-mail: intl@hach.com