Operation Manual
for
Dental Unit
MODEL: L1-660C & 670C
SYSM-660-1
SYSM-660-1
CONTENTS
Identification of Symbols
Classification of Equipment (EN-60601-1)
Environmental
Electromagnetic Compatibility
Flammable Gasses
About This Document
1.About L1-660 and L1-670 dental unit
1.1. External dimension of equipment
1.2 Equipment installation
1.3 parameters
1.4 Power On/Off Button
1.5 Activate Handpiece
1.6 Handpiece Lockout
1.7 Air/Water Coolant
1.8 Water System
1.9 Foot Control Touchpad
1.10 Touchpads
2. Operation
2.1 Status Icon
2.1.1 Water supply setting of rinsing water
2.1.2 Time setting of spittoon washin
2.1.3 Heating of rinsing water
2.1.4 Water supply of spittoon washing
3. Installation and Maintenance
3.1 Pipeline layout
3.2 Connection of air pipe and water pipe
3.3 Maintenance Cleaning and Asepsis
3.3.1 Handpiece
3.3.2 Water filter
3.3.3 Relief valve of air filtration
3.3.4 Saliva ejector and suction
3.3.5 Other
4. Disposal of equipmentSuction filter
5. Failure diagnosis
6. Warranty
7. Customer Service Handpiece
8. Water/Air schematicsOther
9. Electric wiring
10. Packing List for L1-660/L1-670
1
1
1
2
2
2
3
3
3
4
4
5
5
5
6
6
6
7
7
7
7
8
8
8
8
8
9
9
9
9
9
10
10
11
12
12
14
15
16
SYSM-660-1
Identification of Symbols
Table1 symbols
Symbols
Description
Type B applied part
Alternating current
Protective earth (ground)
SN
Serial Number
IPX7
Type/Mode
Types of shock protection
Degree of protection against
Water ingress
Mode of operation
Classification
CLASS I EQUIPMENT, TYPE B APPLIED PART
ORDINARY EQUIPMENT except the foot switch(IPX7)
CONTINUOUS OPERATION except the dental chair
(CONTINUOUS OPERATION WITH INTERMITTENT
LOADING: 5% duty cycle)
Environmental
Table 3
Environmental Specifications
SYSM-660-1
Temperature/Humidity
Specification
-40 to 55
93% maximum
Temperature:
Relative Humidity:
5 to 40
80% maximum
Electromagnetic Compatibility
This equipment has been tested and found to comply with the limits for medical devices in
EN60601-1-2. These limits are designed to provide reasonable protection against harmful
interference in a typical medical installation. Please contact Shanghai Linkgold Medical
Instrument Co., Ltd. Customer Service if you have any questions.
Flammable Gasses
Not suitable for use in the presence of a flammable anesthetic mixture with air, oxygen, or
nitrous oxide, where such gasses may accumulate in concentration (closed space).
Intended Audience
This guide is intended for doctors and dental staff.
SYSM-660-1
Figure.1
SYSM-660-1
Figure.2
1.3 Parameters
Table 4 Feature descriptions
Dental
chair
Treatment
unit
460mm
760mm
60
120mm
Maximum load
135kg
8000Lux15000Lux
3000K6500K
27kPa
35kPa
300000r/min, 0.06Ncm
14000r/min, 1Ncm
10Kg
023Mpa~0.26Mpa
0.30Mpa~0.32Mpa
For recommendation, please use handpiece made by W&H or NSK. Which are in
compliance with standards ISO 2001 or CE.
SYSM-660-1
Figure.3
NOTE
The power on/off button must be on for the dental unit to function.
To prolong the life of your equipment, turn the power off when it is not in use.
ELECTRICAL WARNING Switch the branch circuit breaker off prior to the servicing of
the
dental unit.
Figure 4
The water coolant button operates as a three-way switch with the toggle on the foot control.
The system defaults to the last setting selected for each individual handpiece position.
A
B
C
D
E
Figure 5
(A) turn drive air (B) drive air (C) barometer (D) water (E) air
The L1-660 and L1-670 dental unit and regulate air and water to operate dental handpieces,
SYSM-660-1
syringes and ancillary devices. You can adjust the delivery system water coolant flow, air
flow and drive air pressure (Figure 5).
B
Figure 6
(A) Chip Blower/Accessory Button; (B) Wet/Dry Toggle; (C) Foot Control Disc
To turn the water coolant off, use your foot to flip the toggle.
To turn the water coolant on, flip the toggle again.
NOTE On a deluxe touchpad the wet/dry toggle operates as a three-way switch with the
air
and water coolant buttons.
1.10 Touchpads
The touchpad centralizethe controls of the dental unit ( Figure 7). Some touchpad buttons
have indicators to alert you if the operation is functioning.
LINKGOLD
Figure 7
Touchpad Controls
SYSM-660-1
Icon
Description
Chair Rising
Chair Descending
Bend Backward
Bend Forward
Reset
Setting
Heating
Mouth Rinsing
Spittoon
2. Operation
2.1 Status Icon
The logo"shanghai linkgold" on the touchpad indicates the status of the system:
A solid blue logo indicates normal operation and power is on.
A blinking blue logo indicates the chair stop plate or the cuspidor limit switch has been
activated. Remove any obstructing objects. The icon should return to solid blue.
2.1.1 Water supply setting of rinsing water
a.
SYSM-660-1
b.
memorize and store, therefore, once spittoon washing button is pressed tenderly, it will
automatically wash till the set time when it is used in the future.
2.1.3 Heating of rinsing water
The unit has the automatic isothermal water heater. Please operate as the following to
get hot water. Press the heating button of water heater, the indicating light beside
button
will be on, and then the water heater will heat automatically till the
isothermal state. To close this system, please press this button again, and the indicating
light beside button
2.1.4
will be off.
Power supply
waterairpipe
Drainage pipe1-1/2
Figure.8
SYSM-660-1
Figure.9
SYSM-660-1
4. Disposal of equipment
Please follow the local environmental protection laws and regulations for the dispose of this
unit.
5. Failure diagnosis
NO
Symptom
Cause
Solution
After stopping for a while, It could
continue to operate properly.
No water in handpiece.
SYSM-660-1
2.The bearing of
handpiece is damaged.
3.The dental bur is worn
or isnt clamped tightly.
4.The pipeline of
handpiece is blocked.
5.The air pipe inside
handpiece leaks.
manual.
2.Replace the bearing of
Handpiece.
3.Repiace with a new dental
bur and then clamp the
dental bur tightly.
4.Dismount bearing assembly
and dredge the pipeline of
handpiece.
5.Replace the housing of
handpiece.
SYSM-660-1
6. Warranty
Shanghai linkgold all products against defects in materials or workmanship for one year
from time of delivery. shanghai linkgolds sole obligation under the warranty is to provide
parts for the repair, or at its option, to provide the replacement product (excluding labor).
The buyer shall have no other remedy. All special, incidental, and coincidental damages are
excluded.
Written notice of breach of warranty must be given to shanghai linkgold within the warranty
period. The warranty does not cover damage resulting from improper installation or
maintenance, accident or misuse. The warranty does not cover damage resulting from the
use of cleaning,
disinfecting or sterilization chemicals and processes. The warranty also does not cover light
bulbs. Failure to follow instructions provided in the shanghai linkgold Instructions For Use
(operation and maintenance instructions) may void the warranty.
No other warranties as to merchantability or otherwise are made. For service information,
contact your local authorized shanghai linkgold dealer.
SYSM-660-1
7. Customer Service
For service information, contact your local authorized dealer or manufacturer.
Any optional non shanghai linkgold attachments and/or accessories must comply with EN
60601-1 and EN 60601-1-2.
Specifications are subject to change without notice.
SYSM-660-1
8. Water/Air schematics
SYSM-660-1
9. Electric wiring
SYSM-660-1
Description
Quantity
01
Dental unit
02
3-way syringe
03
High Sunction
04
Saliva ejector
05
Dental light
06
Doctors chair
07
1/2*8mm connector
08
Fusion
Remarks
F5Ax250V
5mmx20mm