SYSM-660-1

Operation Manual
for
Dental Unit
MODEL: L1-660C & 670C

Shanghai LinkGold Medical Instrument Co., Ltd


Release Date:October ,2007

SYSM-660-1

Comments and Feedback


Thank you for taking the time to use the Operation Manual for Dental
Unit L1-660C and L1-670C . We would appreciate any feedback or
comments you have about this document. Please mail, email, or
phone us with your comments. You can reach us at:

Shanghai Linkgold Medical Instrument Co., Ltd.


Address: No. 26, Chngkang Road, Shanghai, China
Zip Code: 200443
Tel: +86-21-29422163
Fax: +86-21-51561729
E-mail: Terrytx@gmail.com
European representative:
Lotus Global Co., Ltd.
47 Spenlow House Bermondsey London SE16 4SJ

SYSM-660-1

CONTENTS
Identification of Symbols
Classification of Equipment (EN-60601-1)
Environmental
Electromagnetic Compatibility
Flammable Gasses
About This Document
1.About L1-660 and L1-670 dental unit
1.1. External dimension of equipment
1.2 Equipment installation
1.3 parameters
1.4 Power On/Off Button
1.5 Activate Handpiece
1.6 Handpiece Lockout
1.7 Air/Water Coolant
1.8 Water System
1.9 Foot Control Touchpad
1.10 Touchpads
2. Operation
2.1 Status Icon
2.1.1 Water supply setting of rinsing water
2.1.2 Time setting of spittoon washin
2.1.3 Heating of rinsing water
2.1.4 Water supply of spittoon washing
3. Installation and Maintenance
3.1 Pipeline layout
3.2 Connection of air pipe and water pipe
3.3 Maintenance Cleaning and Asepsis
3.3.1 Handpiece
3.3.2 Water filter
3.3.3 Relief valve of air filtration
3.3.4 Saliva ejector and suction
3.3.5 Other
4. Disposal of equipmentSuction filter
5. Failure diagnosis
6. Warranty
7. Customer Service Handpiece
8. Water/Air schematicsOther
9. Electric wiring
10. Packing List for L1-660/L1-670

1
1
1
2
2
2
3
3
3
4
4
5
5
5
6
6
6
7
7
7
7
8
8
8
8
8
9
9
9
9
9
10
10
11
12
12
14
15
16

SYSM-660-1

Identification of Symbols
Table1 symbols

Symbols

Description
Type B applied part

Attention, consult accompanying document

Alternating current
Protective earth (ground)

SN

Serial Number

IPX7

Protected against the effects of temporary


immersion in water
compliance with the Medical device directive
93/42/EEC
Manufacturer Name and address
European representative
Manufacture date

Classification of Equipment (EN-60601-1)


Table2 Classification of Equipment

Type/Mode
Types of shock protection
Degree of protection against
Water ingress
Mode of operation

Classification
CLASS I EQUIPMENT, TYPE B APPLIED PART
ORDINARY EQUIPMENT except the foot switch(IPX7)
CONTINUOUS OPERATION except the dental chair
(CONTINUOUS OPERATION WITH INTERMITTENT
LOADING: 5% duty cycle)

Environmental
Table 3

Environmental Specifications

SYSM-660-1

Temperature/Humidity

Specification

Storage and Transportation :


Temperature:
Relative Humidity:

-40 to 55
93% maximum

Range of the atmospheric pressure


Working conditions:

500 hPa1060 hPa

Temperature:
Relative Humidity:

5 to 40
80% maximum

Range of the atmospheric pressure

860 hPa1060 hPa

Electromagnetic Compatibility
This equipment has been tested and found to comply with the limits for medical devices in
EN60601-1-2. These limits are designed to provide reasonable protection against harmful
interference in a typical medical installation. Please contact Shanghai Linkgold Medical
Instrument Co., Ltd. Customer Service if you have any questions.

Flammable Gasses
Not suitable for use in the presence of a flammable anesthetic mixture with air, oxygen, or
nitrous oxide, where such gasses may accumulate in concentration (closed space).

About This Document


Welcome to read the Operation Manual for the operation of Dental Unit (L1-660B/C and
L1-670B/C). This guide provides an easy-to-use source of technical information for operating
and maintaining your L1-660B/C and L1-670B/C Dental Unit .
Before using the L1-660B/C and L1-670B/C Dental Unit , make note of all cautions and
warnings. Read the component descriptions, and become familiar with the locations of
features and controls.
This document contains the following information:
Module descriptions
Feature descriptions
Adjustments
Maintenance tasks

Intended Audience
This guide is intended for doctors and dental staff.

SYSM-660-1

1. About L1-660C and L1-670C dental unit


Model: L1-660C and Model: L1-670C dental unit are suitable for the diagnosis,
treatment and operation on oral disease. Model: L1-660C/L1-670C are both fully computer
controlled system treatment equipments designed by our company according to
ergonomics principle with complete function. As for its operation of complete machine, its
fully automatic operation is all controlled by the function key of microcomputer. It is
composed of multifunction dental complex treatment unit, electric dental chair and dentist
chair. It adopts low voltage DC motor and features with simple structure and convenient
operation.
Water input hose (ID): 10mm hose fitting 1/2"
Input water pressure: 200 400Kpa (Tap water)
Air input hose (ID): 10mm hose fitting 1/2"
Input water pressure: 550 800Kpa (Oil-free Anhydrous and Clean)
Drain pipe (ID): 4mm
Input power: 230v 50Hz 800VA

1.1 External dimension of equipment


To ensure its normal operation and safe movement, please refer to Figure.1 and keep
enough space for clinical use.

Figure.1

1.2 Equipment installation


Preparation before installation ( Figure.2)
According to the layout of consulting room, the lighting conditions and the convenience for
use, confirm the position of the ground box and the position of the whole equipment in
accordance with Figure.4and the form of ground box.

SYSM-660-1

Figure.2

1.3 Parameters
Table 4 Feature descriptions
Dental
chair

Treatment
unit

Minimum height between seat cushion and floor

460mm

Maximum height between seat cushion and floor

760mm

Backward angle of backrest

60

Up and down sliding stroke of Head rest

120mm

Maximum load

135kg

Operating light Illumination


Color temperature

8000Lux15000Lux
3000K6500K

Negative pressure of Saliva ejector

27kPa

Negative pressure of high vacuum aspirator

35kPa

High handpiece speed, torque

300000r/min, 0.06Ncm

Low handpiece speed, torque

14000r/min, 1Ncm

Delivery tray maximum

10Kg

handpiece High handpiece air

023Mpa~0.26Mpa

Low handpiece air

0.30Mpa~0.32Mpa

For recommendation, please use handpiece made by W&H or NSK. Which are in
compliance with standards ISO 2001 or CE.

1.4 Power On/Off Button (Figure.3)


The delivery system have the power on/off button, on the chair.
Press the power on/off button once to turn the power on. Press the button again to turn the
power off.
NOTE Please shut off the power before replacing the safety fuse: F5Ax250V
5mmx20mm

SYSM-660-1

Figure.3
NOTE

The power on/off button must be on for the dental unit to function.
To prolong the life of your equipment, turn the power off when it is not in use.
ELECTRICAL WARNING Switch the branch circuit breaker off prior to the servicing of

the
dental unit.

1.5 Activate Handpiece


The handpieces automatically activate when you lift a handpiece from its holder or pull the
Continental whip forward. Press the foot control to operate the activated handpiece.

1.6 Handpiece Lockout


When a handpiece is removed from the holder and the foot control is pressed, the chair
does not operate.

1.7 Air/Water Coolant


Each handpiece is set for independent on/off operation. When you withdraw a handpiece
from its resting position, air and water coolant will be activated. An icon displays on the
screen to indicate ctivation.(Figure 4)

Figure 4

The water coolant button operates as a three-way switch with the toggle on the foot control.
The system defaults to the last setting selected for each individual handpiece position.
A

B
C
D
E
Figure 5
(A) turn drive air (B) drive air (C) barometer (D) water (E) air
The L1-660 and L1-670 dental unit and regulate air and water to operate dental handpieces,

SYSM-660-1

syringes and ancillary devices. You can adjust the delivery system water coolant flow, air
flow and drive air pressure (Figure 5).

1.8 Water System


The L1-660 or L1-670 handpiece water system sed of distilled water.

1.9 Foot Control

B
Figure 6
(A) Chip Blower/Accessory Button; (B) Wet/Dry Toggle; (C) Foot Control Disc
To turn the water coolant off, use your foot to flip the toggle.
To turn the water coolant on, flip the toggle again.
NOTE On a deluxe touchpad the wet/dry toggle operates as a three-way switch with the
air
and water coolant buttons.

1.10 Touchpads
The touchpad centralizethe controls of the dental unit ( Figure 7). Some touchpad buttons
have indicators to alert you if the operation is functioning.

LINKGOLD

Figure 7

Touchpad Controls

SYSM-660-1

The following table describes the function of each button:


Table 5 Touchpad Button Identification.

Icon

Description
Chair Rising
Chair Descending
Bend Backward
Bend Forward
Reset
Setting
Heating
Mouth Rinsing
Spittoon

2. Operation
2.1 Status Icon
The logo"shanghai linkgold" on the touchpad indicates the status of the system:
A solid blue logo indicates normal operation and power is on.
A blinking blue logo indicates the chair stop plate or the cuspidor limit switch has been
activated. Remove any obstructing objects. The icon should return to solid blue.
2.1.1 Water supply setting of rinsing water
a.

Press down set button

, gently press down SET, the setting state starts when

the indicating light is on.


b.

Press rinsing water button

, starts to supply water, stop pressing when the water

in measuring cup reaches anticipated water level, stops supplying water.


c. Press set button again, setting is over, computer will memorize and store, therefore,
once press the rinsing water button, it will automatically supply water with such amount
when it is used in the future.
2.1.2 Time setting of spittoon washing
a. Press down set button

, gently press down SET and the indicating light is on.

SYSM-660-1

b.

Press spittoon washing button

, it starts to fill water into spittoon. When it reaches

the required and set time, press

, again and the setting is over. Computer will

memorize and store, therefore, once spittoon washing button is pressed tenderly, it will
automatically wash till the set time when it is used in the future.
2.1.3 Heating of rinsing water
The unit has the automatic isothermal water heater. Please operate as the following to
get hot water. Press the heating button of water heater, the indicating light beside
button

will be on, and then the water heater will heat automatically till the

isothermal state. To close this system, please press this button again, and the indicating
light beside button
2.1.4

will be off.

Water supply of rinsing


to supply water and it will be closed automatically

Press rinsing water supply button


when it reaches preset water amount.
2.1.5 Water supply of spittoon washing
Press water supply button

of spittoon washing to supply water automatically and it

will be closed automatically when ti reaches preset time of water supply.

3. Installation and Maintenance


The installation of officers for professionals.

3.1 Pipeline layout (Figure.8

Power supply

waterairpipe

Drainage pipe1-1/2

Figure.8

3.2 Connection of air pipe and water pipe (Figure.9)

SYSM-660-1

Water input(8mm1) Blue)

Air input (8mm1) transparent

Figure.9

3.3 Maintenance Cleaning and Asepsis


3.3.1 Handpiece
Handpiece is a delicate device, so it shall not be pounded or dropped. It shall be cleaned
and lubricated with special detergent and lubricant every day. As for the specific method,
pleace refer to the instruction manual of handpiece provided by they manufacturer.
z Clean the surface of dental drill with alcohol cotton.
z Blow it with clean lubricating oil for 2-3s.
z Wrap it with sterilization package.
z High-temperature steam sterilization(135,15min,220KPa)
3.3.2 Water filter
The unit has water filter to ensure its normal use.After serving a certain time,the dirt will block
the filter core and affect the water flow,therefore,it is necessary to clean and replace the filter
core. The filter core shall be cleaned and replaced under any of the following conditions:
z It has been used for over half a year.(Depend on the local water quality).
z The extracted water becomes turbid suddenly.
The specific cleaning method is as follows: open the ground box cover first, counterclockwise
rotate and dismount water filter cover to remove filter core. Clean it with the ultrasonic or
neutral detergent, wash it with tap water thoroughly after cleaning and then mount water filter
cover. The water filter can stop the inpurity of water, clean the filter core for every three months.
3.3.3 Relief valve of air filtration
Drain the water when:
z It is used for over one week.
z The water in filter cup accounts for three-quarter of volume.
z The color of water in filter cup changes turbid .
3.3.4 Saliva ejector and suction
Besides the maintenance of their exterior cleanness, the cleanness of interior pipeline
system of saliva ejector and suction still need to be maintained. The method of cleaning
pipe system is as follows: regularly pump clear water or special sterilized water about 30s

SYSM-660-1

with recommended 1 or 2 times every day.


3.3.5 Other
a) 3-way syringe tip: high-temperature steam sterilization(135C,15min,220KPa) .
b) Dental chair: clean and sterilize it with leather detergent and medical alcohol.

4. Disposal of equipment
Please follow the local environmental protection laws and regulations for the dispose of this
unit.

5. Failure diagnosis
NO

Symptom

Cause

When it is used frequently, the


The motor is overheated
dental chair doesnt operate
and stops operating.
suddenly,while the control system
under normal operation.

Solution
After stopping for a while, It could
continue to operate properly.

The saliva ejector or high


vacuum aspirator leak
automatically.

The mounting position


of rack valve changes.

The suction of saliva


ejector and high
vacuum aspirator is small
or without suction.

1.The pressure of tap


water is too low.
2.The water filter is
blocked.
3.Pressure of supply gas
Is too low.
4.Negative pressure
Generator is blocked.
5.The pipeline and joint
leak.

1. Install supercharge water


pump if the water pressure
is too low.
. Clean the water filter and
Replace filter core.
. Adjust supply gas pressure
To 0.5MPa.
. Dismount and wash
negative pressure
generator ,remove the
blocking.
5.Check the pipeline and joint.

No water in handpiece.

1.The jet hole of


handpiece is blocked.
2.The water filling switch
of foot control air
switch isnt open.
3.The water regulating
valve of handpiece
isnt open.

1.Dredge the jet hole of


handpiece.
2.Shift the water filling switch
of foot control air switch to
the right.
3.Open the water regulating
Valve of handpiece and
adjust it properly.

The handpiece isnt


powerful or doesnt operate.

1.The operation air


pressure of handpiece
Is too low.

1.Adjust the operation air


pressure of handpiece
according to instruction

Open the rear cover of rack,adjust


the front and back mounting
position of rack valve until it stops
When putting down the instrument
and operates when taking up the
instrument.

SYSM-660-1

2.The bearing of
handpiece is damaged.
3.The dental bur is worn
or isnt clamped tightly.
4.The pipeline of
handpiece is blocked.
5.The air pipe inside
handpiece leaks.

manual.
2.Replace the bearing of
Handpiece.
3.Repiace with a new dental
bur and then clamp the
dental bur tightly.
4.Dismount bearing assembly
and dredge the pipeline of
handpiece.
5.Replace the housing of
handpiece.

The button of three-way


syringe leaks.

O-ring is worn or aging.

Replace O-ring and add


Silicon grease oil to lubricate.

The three-way syringe


doesnt sufficiently spray
vapor.

1.The water-pipe and


air-pipe of three-way
ssyringe is connected
oppositely.
.The water of three-way
syringe is too strong or
the air is too weak.

1. Exchange the water-pipe


with the air-pipe of three-way
syringe.
. Adjiust the water and air of
three-way syringe.

Rinsing water and spittoon


washing water are too few
or without water.

1.The water pressure is


too low.
2.The water filter is
blocked.
3.Pressure of supply gas
is too low.
4.Electromagnetic valve
doesnt work.
5.The connecting wire
and connector dont
contact properly.
6.The circuit board
breaks down.
7.The control keyboard
doesnt contact properly.

1.Install supercharge water


pump if the water pressure
is too low.
2.Clean the water filter and
replace filter core.
3.Adjust supply gas pressure
to 0.5MPa.
4.Replace the damaged
electromagnetic valve.
5.Find out the contact failure
position of connecting wire
and reconnect it.
6.Replace defective circuit
board.
7.Replace control keyboard.

The treatment unit doesnt


operate.

1. Thepower plug isnt


properly plugged or
doesnt properly contact.
. The power fuse is
blow-out.
. The transformer of
ground box is damaged.
. The power switch is
broken.

1.Replace power plug or


outlet.
2.Find out the reason of
blow-out first and replace
power fuse.
3.Replace the transformer of
Ground box.
4.Replace power switch.

SYSM-660-1

10 No water in treatment unit

The water filter is blocked.


Water pipe is pinched.

Clean filter core.


Remove the pinching source.

11 There is water spraying out


from the air.

1.The joint of handpiece


Isnt screwed tightly.
2.The filter relief valve is
full of water.

1.Screw the joint of


handpiece tightly.
2.Discharge the accumulated
water inside filter relief valve.

12 The operating light isnt on.

1.The bulb of operating


Light is broken.
2.The connecting wire of
operating light doesnt
properly contact.
3.The switch of operating
light doesnt properly
contact.
4.The lead wire of ground
box transformer doesnt
properly contact.

1.Replace a new bulb.


2.Find out the defective
connection position.
3.Replace the switch of
operating light.
4.Reconnect the lead wire of
ground box transformer or
replace the transformer of
ground box.

13 The chair doesnt operate,


while its sound indication is
normal and other buttons
are under normal operation.

1.The relay contact


doesnt contact properly
or the coil is broken.
2.The motor is broken or
the connecting wire is
disconnected.

1.Clean the relay contact or


replace a new relay.
2.Find out the disconnect
position,reconnect or replace
the motor.

6. Warranty
Shanghai linkgold all products against defects in materials or workmanship for one year
from time of delivery. shanghai linkgolds sole obligation under the warranty is to provide
parts for the repair, or at its option, to provide the replacement product (excluding labor).
The buyer shall have no other remedy. All special, incidental, and coincidental damages are
excluded.
Written notice of breach of warranty must be given to shanghai linkgold within the warranty
period. The warranty does not cover damage resulting from improper installation or
maintenance, accident or misuse. The warranty does not cover damage resulting from the
use of cleaning,
disinfecting or sterilization chemicals and processes. The warranty also does not cover light
bulbs. Failure to follow instructions provided in the shanghai linkgold Instructions For Use
(operation and maintenance instructions) may void the warranty.
No other warranties as to merchantability or otherwise are made. For service information,
contact your local authorized shanghai linkgold dealer.

CAUTION Restricts this device to sale by or on the order of a dentist, physician, or


any other practitioner licensed by the law of the state in which he or she practices to use or
order the use of this device

SYSM-660-1

7. Customer Service
For service information, contact your local authorized dealer or manufacturer.
Any optional non shanghai linkgold attachments and/or accessories must comply with EN
60601-1 and EN 60601-1-2.
Specifications are subject to change without notice.

SYSM-660-1

8. Water/Air schematics

SYSM-660-1

9. Electric wiring

SYSM-660-1

10. Packing List for L1-660C/L1-670C


No.

Description

Quantity

01

Dental unit

02

3-way syringe

03

High Sunction

04

Saliva ejector

05

Dental light

06

Doctors chair

07

1/2*8mm connector

08

Fusion

Remarks

F5Ax250V