Type of Equipment
Model Name
Name of the Manufacturer
POPULAR MEDICARE
LIMITED,POPULAR HOSPITAL (MOHIT
DIAGNOSTIC NO. 10/60,A-2
KAKARMATTA,VARANASI
RADIOGRAPHY
BRIVO XR 115
M/S WIPRO GE HEALTHCARE PVT
LTD, BANGLORE
NEELU DIXIT, GENWORKS HEALTH
PVT LTD
26.05.2016
4
5
6
Parameters
Tested
Congruence
of radiation
and optical
field
Central Beam
Alignment
Effective focal
spot
measurement
FFD= 60 cm
Specified
Values
Measurement
at 60 kV
Accuracy of
Operating
Potential (kV)
Accuracy of
Irradiation
Time(sec.)
Total filtration
(last value
maximum kV)
Measured
Values
1.2%
0.8%
Measurement
at 60 kV
Single Focus
1.50
1.8mm X
1.8mm
104
N/A
Measurement
at maximum
kVp 100
4.4 mmAl
Tolerance
Tolerance :
X + X 2% of FFD
Y + Y 2% of FFD
Central Beam Alignment
1.50
Tolerance :
1. + 0.5 f for f 0.8 mm
2. + 0.4 f for 0.8 f 1.5
mm
3. + 0.3 f for f 1.5 mm
5 kV
% Error < 10 %
Tolerance :
1.5 mm Al for kV 70
2.0 mm Al for 70 < kV 100
2.5 mm Al for kV 100
Remarks
PASS
PASS
PASS
PASS
N/A
PASS
8
9
10
11
12
Linearity of
mA/mAs
loading
Stations
Consistency
of radiation
output
Low contrast
resolution
0.023228
0.009108
Measurement
at Maximum
kVp and
corresponding
tube current
PASS
80 KVp
Measurement
shall be done
on each mode
of Image
Intensifier
High contrast
Measurement
resolution
shall be done
on each mode
of Image
Intensifier
Exposure Rate Applicable
at Tabletop
Fluoroscopy
Mode
Radiation
leakage level
at 1m from
tube housing
and
Collimator
(Pl. mention
values
separately)
CoL<0.1
PASS
3.0 mm hole pattern must be
resolved
N/A
N/A
1.5 lp/mm pattern must be
resolved
N/A
N/A
N/A
Tolerance :
1. Exposure Rate without AEC
mode 5 cGy/Min
2. Exposure Rate with AEC
mode 10 cGy/Min
Tube Leakage < 1 mGy in one
hour
N/A
PASS
.26 mR
Or
0.00221
mGy
I hereby undertake that all the information provided above is correct and in accordance with the detailed
Quality Assurance Report enclosed herewith.
Place:Mughal Sarai
Date:27.05.2016
Signature:
Name of the Service Engineer: Mr. Neelu Dixit
Name of Supplier/Service Agency: Genworks Health
Pvt Ltd
Seal of Supplier/Service Agency:
100
kV
60
% of FFD
1.2
0.8
mAs
Tolerance
2 % of FFD
2 % of FFD
Remark
PASS
PASS
100
kV
60
mAs
Observe the images of the two steel balls on the radiograph and evaluate tilt in the central beam.
Observed tilt
PASS
1.50
0
Tolerance: Central Beam Alignment 1.5
3. EFFECTIVE FOCAL SPOT MEASUREMENT
Large focus
Small focus
FFD (cm)
60
FFD (cm)
100
Applied
Set
(s)
Measured values
mA stations
Routinely used
Maximum mA
mA station
station
kVp
Time (s)
kVp Time (s)
60.5
NA
60.4
NA
70.5
NA
70.2
NA
79.2
NA
80.4
NA
98.5
NA
98.4
NA
Minimum mA
station
kVp Time (s)
60.5
NA
70.4
NA
80.5
NA
98.2
NA
Average
kVp
Average
Time (s)
60
NA
60.4667 NA
70
NA
70.37
NA
80
NA
80.03
NA
Max.
NA
98.37
NA
100kVp
Tolerance for kVp : 5 kV
Tolerance for Timer : % Error: 10 %
Total Filtration (measurement at maximum kVp) : 4.5 mm of Al
Tolerance for Total Filtration: 1.5 mm Al for kV 70, 2.0 mm Al for kV 100, 2.5 mm Al for kV 100
Remarks
PASS
PASS
PASS
PASS
100
mA
applied
kV
60
Reading 1 Reading 2
25
42.3
42.6
50
87.3
86.3
100
166.5
166.23
Tolerance: COL 0.1
Average
Output
Reading 3
42.72
87.8
165.561
(Gy)
42.54
87.13
166.097
Time (s)
1 Sec
Gy /mAs
( X)
Coefficient
of Linearity
(CoL)
Remarks
0.023228
PASS
FFD (cm)
100
1.7
1.74
1.661
8. OUTPUT CONSISTANCY
Applied
kV
mAs
70
160
403.23
402.87
80
120
157.3
100
60
253.4
404.12
403.23
403.35
158.2
160.1
160.2
157.3
254.7
253.3
253.1
252.2
N/A
Avera
ge
(X)
403.1
7
158.6
2
Coefficient of Remarks
Variation
(CoV)
0.001144
PASS
0.009108
PASS
253.3
4
0.003537
PASS
N/A
1.5 lp/mm pattern must be resolved
Remarks
N/A
100
kVp
(Max)
mA
60
Time(s)
1sec
Location
Exposure level (mR/hr)
Result
(at 1.0 m from
Left
Right
Front
Back
Top
the focus)
Tube
3.7
3.1
1.54
3.87
1.2
PASS
Collimator
5.2
5.7
2.1
2.6
1.4
PASS
Tolerance: Maximum leakage radiation level at 1 meter from the focus should be 1 mGy (114 mR)
in one hour.
Work load 180 mAmin in one hr
Max leakage = 180 mAmin in 1 hr X 5.2mR/hrMax Exposure level (mR/hr)
60 X 60 mA used for measurement
Maximum radiation leakage from tube housing = 0.26 mR in one hour 0.002281 mGy
Maximum radiation leakage from tube collimator = 0.26 mR in one hour
Details of Radiation Protection Survey of the installation
Date of radiation protection survey:
Whether radiation survey meter used for the survey has valid calibration certificate: Yes
Equipment Setting:Applied Current (mA): 60
Applied Voltage (kV): 70
Exposure time(s): 1 Sec