INFRARED SPECTROSCOPY
Activity No. 1

Submittted by:
John Mark C. Saburao
BS Chem- 3H1

February 25, 2016

OBJECTIVES
At the end of this experiment, the students should be able to:
1. Study the instrumentation of a Fourier Transform-Infrared (FT-IR) spectrometer
2. Identify and familiarize the components of a FT-IR spectrometer
3. Understand the theory and concept behind infrared spectroscopy as one of the analytical
methods used in chemistry

THEORETICAL BACKGROUND
One of the techniques used in determining the structure of an organic compound is
infrared spectroscopy. In infrared spectroscopy, the functional groups that are present in an
organic compound are determined. Infrared spectroscopy involves the interaction of molecules
with electromagnetic energy. Thus the concept of radiant energy and electromagnetic spectrum is
very important when dealing with spectroscopic techniques involving infrared spectroscopy.
Visible light, X rays, microwaves, radio waves, ultraviolet rays, gamma rays and so forth are all
different kinds of electromagnetic radiation. Collectively, they make up the electromagnetic
spectrum which is arbitrarily divided into regions, with the familiar visible region accounting for
only a small portion. The visible region is flanked by the infrared and ultraviolet regions. When
an organic compound is exposed to a beam of electromagnetic radiation, it absorbs energy of
some wavelengths but passes or transmits energy of other wavelengths. Irradiating the sample
with energy of many different wavelengths and determining which are absorbed and which are
transmitted; the absorption spectrum of a compound can be measured. The energy a molecule
gains when it absorbs radiation is distributed over the molecule in some way. With infrared
radiation, the absorbed energy causes bonds to stretch and bend more vigorously. The stretching
and bending of the molecules provide structural information when interpreted. The instrument
used in infrared spectroscopy is the Fourier transform infrared (FT-IR) spectrometer. FT-IR
provides useful information when identifying unknown substances, determining the quality or
consistency of a sample, and determining the amount of the components in a mixture.

METHODOLOGY
A. Liquid Sample Preparation
The IR cell is removed from the storage desiccator. Then, the crystal holder is carefully
disassembled by unscrewing the plastic (male) piece. The pieces are placed on a clean paper
towel strip. After which, the crystals are removed by inverting the metal part of the crystal holder.
The crystals are placed in a clean paper towel. Then, the teflon spacer in the metal holder may be
replaced if necessary. Following that, place about one gram of the salt crystals into the metal
holder. Then, using a clean pipet, a drop of the sample is placed in the center of the crystal in the
metal holder. The other crystal is squarely placed on top of the first crystal to spread the liquid
into a film. After that, the cell is reassembled by screwing the plastic holder piece. Next, the cell
mount is placed in the IR spectrometer. Then, the cell is placed on the cell mount. Lastly, the IR
spectrum of the sample is obtained.
B. Solid Sample Preparation
The IR cell is removed from the storage desiccator and placed on a clean paper towel
strip. Then, in a 1-dram vial, a very small volume (<0.5 mL) of a slightly concentrated solution
of the solid sample in a volatile solvent is prepared. After which, 1-2 drops of the solution is
placed in the center of one of the NaCl plates. The solvent is allowed to evaporate from the
crystal for several minutes. The solvent evaporation can be hastened by waving the hand back
and forth over the crystal. Following that, the salt crystal is placed into the metal holder with the
sample on top. The second crystal is not required any more. Then, the cell is reassembled by
screwing the plastic holder piece. After that, the IR cell is placed in the IR spectrometer and the
spectrum of the sample is obtained.
C. Gas Sample Preparation
A gas sample is introduced into the gas cells by connecting one of the taps to a vacuum
pump and opening it to evacuate the cell. After closing the tap, the other tap is connected to the
sample and opened to fill the gas cell with the gas sample. Then, the internal pressure is set from
several tens of Torr to 760 Torr. When filling the cell, the pressure is adjust by using a pressure
gauge or if there is no pressure gauge, a sampling system with a manometer is used. Then, the
gas cell, manometer, and vessel containing the sample is connected. The taps are opened to

evacuate the gas flow line, and then the third tap is closed, followed by the first tap. The fourth
tap is gradually opened to allow the sample to flow into the gas cell. The manometer is then
observed and the fourth tap is closed when the appropriate pressure is reached. Finally, the
second tap is closed and the gas cell is removed.
DISCUSSION
The basic components in an FT-IR system are sample, mirror, IR source, interferometer,
detector and computer. The same types of radiation sources are used for both dispersive and
Fourier transform spectrometers. Infrared energy is emitted from a glowing black-body source.
This beam passes through an aperture which controls the amount of energy presented to the
sample. The interferometer divides radiant beams, generates an optical path difference between
the beams, then recombines them in order to produce repetitive interference signals measured as
a function of optical path difference by a detector. The interferometer produces interference
signals, which contain infrared spectral information generated after passing through a sample.
The beam enters the interferometer where the spectral encoding takes place. The resulting
interferogram signal then exits the interferometer. An interferometer consists of three active
components: a moving mirror, a fixed mirror, and a beamsplitter. The two mirrors are
perpendicular to each other. The beamsplitter is a semi-reflecting device and is often made by
depositing a thin film of germanium onto a flat KBr substrate. Radiation from the broadband IR
source is collimated and directed into the interferometer, and impinges on the beamsplitter. At
the beamsplitter, half the IR beam is transmitted to the fixed mirror and the remaining half is
reflected to the moving mirror. After the divided beams are reflected from the two mirrors, they
are recombined at the beamsplitter. Due to changes in the relative position of the moving mirror
to the fixed mirror, an interference pattern is generated. The beam enters the sample
compartment where it is transmitted through or reflected off of the surface of the sample,
depending on the type of analysis being accomplished. This is where specific frequencies of
energy, which are uniquely characteristic of the sample, are absorbed.The resulting beam then
passes through the sample and is eventually focused on the detector. For an easier explanation,
the detector response for a single-frequency component from the IR source is first considered.
This simulates an idealized situation where the source a modulated monochromatic beam from a
helium neon (He-Ne) laser.The measured signal is digitized and sent to the computer where the

Fourier transformation takes place. The final infrared spectrum is then presented to the user for
interpretation and any further manipulation. The figure below shows a typical Fourier transform
infrared spectrometer.

Fig.1. Fourier transform infrared spectrometer

REFERENCES
Board of Trustees of Northern Illinois University. (2007). Sample preparation for FT-IR - NIU Chemistry Analytical Lab. Retrieved February from http://www.niu.edu/
ANALYTICALLAB/ftir/samplepreparation.shtml
D. L. Pavia, G. M. Lampman, and G. S. Kriz, in Introduction to Spectroscopy: A Guide for
Students of Organic Chemistry. Harcourt Brace College Publishers, Fort Worth (1996).
Derry, P., & Barron, A. R. (2000). OpenStax CNX. Retrieved from http://cnx.org/contents/uie
DnVBC@20.16:vdl5g77_@1/IR-Sample-Preparation-A-Practical Guide
Mitra, S. (Ed.). (2003). Sample Preparation Techniques in Analytical Chemistry. Hoboken, New
Jersey: John Wiley & Sons.
M. R. Derrick, D. Stulik, and J. M. Landry.(1999). Scientific Tools in Conservation: Infrared
Spectroscopy in Conservation Science. Getty Conservation Institute, Los Angeles.