Escolar Documentos
Profissional Documentos
Cultura Documentos
b Dispersionof chromatographiczones
in packed columns caused by uneven
flow of eluent through the packing
is relaxed by lateral mass transfer,
which consists of flow-independent
molecular diffusion and flow-dependent
anastomosis resulting from lateral random walk of eluent gas through the
interstices of packing. The relative
magnitude of the two mass transfers
is discussed and the conclusion is
that anastomosis is appreciable at
flow rates faster than about 20 cm.
per second in hydrogen or 5 cm.
per second in nitrogen. By using
flow rates faster than these, the extra
mass transfer provided by anastomosis enables columns to be operated
at more plates per second and thus
enables separations to be achieved
more rapidly.
rn
ANALYTICAL CHEMISTRY
D, = aud,
(2)
where is a constant. The main point
of interest is the value of the constant.
We suggest below that a t fast flows
D, is a major contributor to radial diffusion in cylindrical columns of packed
beds, and experimental evidence (10)
shows that in these conditions the
coefficient of radial diffusion is altered
by small alterations in the packing
technique. It is apparent, therefore,
(Y
D, f D ,
(3)
H
0.9
m m.
0.80.7*
0.6
0.5 .
0.40.3.
0.2
'
20
Figure 1.
EXPERIMENTAL
Carrier gas, either nitrogen or hydrogen, passed through a pressure controller (V.P.C. 1, Edwards and Co.,
Ltd.) , and through a simple introducing
port consisting of a compressed silicone
rubber disk in the side arm of a brass
T-piece such as is used for plumbing
1/4-inch pipe to the column. The
inlet pressure was measured with a
mercury manometer. At the column
outlet, the gas passed first through a
T-piece, the side arm of which was
fitted with an adjustable tap, and then
to a flame detector (Gas Chromatography, Ltd.) . For most of the runs,
the carrier gas flow was too great for
the flame detector, and thus most of the
gas was vented to waste through the
tap. Recording was done on a 1-second
full scale travel recorder (KoneywellBrown). The time constant of the
recorder was much less than 1 second for
short pen travels. For rapid peaks,
therefore, the peak height was restricted by reducing the detector sensitivity so that the pen could always
keep pace.
KOflowmeter was incorporated. The
linear flow rate a t a few suitable slow
flows was measured from the column
length and the time of passage through
the column of a nonabsorbed gas.
Values of the flow speed a t faster flows
were calculated from the retention
times of solutes compared to those a t
the measured slow flows and the appropriate pressure correction factors, j
(see below).
Columns not longer than about 2 feet
were used, so that flows of up to 50 crn.
Ref.
No.
I
Internal
diameter,
4
Support
Celite
I1
Celite
I11
Celite
IV
Celite
mm.
40
80
uo
100 cm/s
3-Methylhexane in nitrogen
n-Heptane in nitrogen
2-Methyihexane in hydrogen
n-Heptane in hydrogen
fB
=jH*+
60
G+
Column
length,
cm.
59.5
60.0
59.5
71
Solutes used
3-Methylhexane
+Heptane
2-Methylpentane
n-Hexane
3-Methylhexane
n-Heptane
2-Methylpentane
%-Hexane
62
86
18
26
62
86
14
19
llr2
c, = 240,
~
Figure 2.
Typical plot of H
1.
2.
3.
4.
2-Methylpentane in nitrogen
n-Hexane in nitrogen
2-Methylpentane in hydrogen
n-Hexane in hydrogen
Table 11.
2-Methylpentane
B
33
41
9
11
D,
H2
N2
ANALYTICAL CHEMISTRY
06
rnrn.
1.
05
04
20
40
60
80
crnl s
Figure 3.
IO0
a
.
D,
27
7.2
n-Heptane
B
34
9
rnm.
Hp8
Table 111.
Column
I
Solute
3-Methylhexane
*Heptane
2-Methylpentane
n-Hexane
n-Methylheptane
+Heptane
2-Methylpentane
n-Hexane
I1
I11
IV
Figure 4.
H*
c = C,
0.68
0.50
1.33
0.83
0.7
0.2
1.3
0.8
Experimental points
2-Methylpentane
n-Hexane
Higher. Nitrogen carrier gas
lower. Hydrogen carrier gas
Theoretical lines, a = 0.24
1 2-Methylpentane in nitrogen
2, n-Hexane in nitrogen
3. n-Hexane In hydrogen
4. 2-Methylpentane in hydrogen
Ci, msec.
HX4
c1
mm.
(7)
H*.
The above procedure for isolating
H * from H was followed for solutes
given in Tables I1 and I11 on the
columns described in Table I. The results are presented as the points in the
plots of H * us. uoin Figures 3, 4, 5, and
6: one figure for each column. In all
cases, H* is a monotonically increasing
function of u,, but except a t the slowest
flows it increases less than linearly, as
predicted by Equation 4.
The full lines in the figures are calculated from Equation 4,. putting u =
uoand with CY as specified in the legends.
For these plots, K was obtained from
30
Figure 5.
60
H*
cm/s
I50
120
0
Table IV.
Column
No.
I
Solute
I1
n-Cl
MeH)
2 MeP
n-Ce
I11
Z-FH>
Value of D , used,
sq. mm/sec.
H2
Nz
19.6
22.8
30
8.1
35
33
30.3
25.5
33
31
30
8.1
140
142
8.5
IV
75
75
30
8.1
n-Cs
Except for column 11,plots given by different solutes are not significantly different.
VOL. 38,
NO. 1,
JANUARY 1966
Figure 6.
ANALYTICAL CHEMISTRY
curvature of plots of H* us. uo,and present their data in graphs which are
essentially similar to those of Figures 3
to 6. Calculations from measurements
from their graphs, and also calculations
made from the parameters used in their
discussion (which differ from those
used here) lead to the result
= 1.
For example, simple calculation shows
that l / a = 4h/p2, where a is as in this
paper, and h and p are as in Harper and
Hammonds; Harper and Hammond
propose p = 2 h and h = 1; whence
a = 1. This value is very much higher
than is found here, and implies appreciable curvature of plots of H* us.
uo a t flows as small as 1 or 2 cm. per
second. This is contrary to our
experience.
Bernard and Wilhelm (1,12) describe
experiments performed using much
larger particles in a much larger tube
than here, with very fast flows indeed,
so that flow is turbulent and molecular
diffusion is negligible. They have
measured the coefficient of radial diffusion, which in their circumstances
1342 (1963).
H. W. HABGOOD
Research Council o f Alberta, Edmonfon, Alberta, Canada
The temperature dependence of
carrier-gas flow rate under conditions of constant inlet and outlet
pressures was measured for packed
and capillary columns of metal and
of glass using helium, hydrogen, and
nitrogen as carrier gases. With all
three gases over the temperature
range 25 to 200 C. the flow rate
as measured at a constant temperature may be taken as proportional
to ( l / T ) 1 . 7 for packed metal columns
and for capillary columns of either
metal or glass, and as proportional
to ( 1 /T)1.8 for packed glass columns.
The flow rates expressed at column
temperature will be proportional to
(1 / T ) O . 7 and ( 1 /T)O.*, respectivey. The
temperature coefficients for packed
columns are about 0.1 higher than
would be predicted theoretically.
outlet pressures,
the decreasing
volumetric flow rate as measured a t the
column outlet and a t a constant temperature results in increasing peak areas
(using a thermal conductivity detector).
In general, the time taken for analysis
and the efficiency of a separation can be
influenced by the changes in flow resulting from temperature changes. I n
this study the effect of temperature on
flow rate is investigated for several
gases of interest in chromatography.
THEORY
F_, -- - BdP
A
t dz
(1)