Escolar Documentos
Profissional Documentos
Cultura Documentos
General Information
Service Notes
Checking Procedure
Trouble Shooting
Adjustment Procedure
Mechanical Parts
Preventative Maintenance
Theory of Operation
System Interconnections
10
General Information
Section 1
GENERAL INFORMATION
CONTENTS
PAGE
Safety Aspects...........................................................................................................................................2
1.1
1.2
System Overview.......................................................................................................................................4
2.1
Front View..............................................................................................................................................4
2.2
2.3
2.4
Models ...................................................................................................................................................6
2.5
Revision System......................................................................................................................................11
4.1
4.2
4.3
4.4
4.5
4.6
BI2201-E
1-1
General Information
Safety Aspects
The Pro Focus 2202 System complies with:
EN/IEC 60601-1
EN 60601-2-37
UL2601-1
CSAC22.2 No.601.1.
1.1
Type B (Body)
Type B (Body)
"CAUTION":
Symbol
IP57
Name
Description
ATTENTION
Potential
Equalisation
Protective
Earth
Type CF
Type BF
Type BF
BF, DEFIBRILLATOR-PROOF
Type B
SEALING
Stand-by
Push button for switching the scanner from stand-by to active. (The
power supply cord is the means of separation from the main power
supply.)
Off
On
Non-ionising
radiation
STERILE
1-2
BI2201-E
General Information
Any work done on the open instrument with power on must only be done by B-K
Medical or their authorised representatives, who are aware of the hazards
involved.
Any repair on the 2202 must be followed by an electrical safety test to verify a
continuous safe operation of the system.
Only the original mains cable must be used NEVER USE EXTENSION CABLES!!
The 2202 contains a Lithium battery. Under no circumstances must this battery be
removed or replaced by the user as there is danger of explosion.
Personal Safety:
Be aware that there may be a risk of infection due to contaminated equipment,
especially puncture guides/needles and transducers. The following precautions should
be taken:
At the hospital ask the staff to sterilise transducers and puncture guides before
receipt. Consoles must be disinfected as recommended in the User Guide before
any repair.
When working with possible infected equipment, use gloves especially if you have
open wounds or scratches.
Cautions:
BI2201-E
Ensure that the two mains voltage selector switches on the rear panel of the
power supply are set to match the actual mains voltage.
The power supply cord is the means of connecting the 2202 from the main power
supply.
1-3
General Information
2
2.1
1-4
System Overview
Front View
BI2201-E
2.2
General Information
Back View
VGA 1
VGA 2
VGA 1
L
VGA 2
RGBS
B
1
!
RGBS
RS232
5V
0.1 A
BI2201-E
1-5
General Information
2.3
System Configuration
The Diagnostic Ultrasound System 2202 is an ultrasound scanner optimised for use in:
2.4
B-mode
B-mode with Tissue Harmonic Imaging
B-mode + M-mode
Doppler mode including:
Spectral Doppler (PW & CW)
Power mode (including Directional Power Doppler)
Color mode (Color Doppler)
Triplex (B-mode plus Spectral Doppler and either Color Doppler or Power Doppler)
Contrast Imaging
Models
There are 3 versions of the Pro Focus 2202: Yellow, Green and, Blue. The available
features for each version can be found in the product catalogue. (BG0143-O)
1-6
BI2201-E
General Information
BI2201-E
B-K
Item
ZD0774
ZD0773 /
ZD0789
Core Board
ZD0784
Single Board PC
ZV0069 /
ZV0070
CRT Monitor
ZV0071
Flat Screen
UL0041
Picture
1-7
General Information
1-8
UL0036
ZE0797 /
ZE0817 /
ZE0821 /
ZE0823
ZE0798 /
ZE0815 /
ZE0822
ZE0820
ZG0346
Power Supply
ZH0757
Back Plane
ZH0758
Video IF board
ZH0759
BI2201-E
BI2201-E
General Information
ZN0046
ZN0047
ZN2220
ZN2230
ZN2231
ZN2221
ZN2226
ZN2241
ZN2242
1-9
General Information
ZN2201 /
ZN2223
ZN2202 /
ZN2222
1-10
BI2201-E
General Information
Revision System
All PC Boards in the 2202 are described by the board name, Type number, Revision
code (ID) and PCB Version.
The name of the board indicates the function of the board, for example the Core Board.
The name of the board is changed only if the function of the board changes.
Part number
Serial Number
| | | || | | ||| | | | | | |||
| ||| |
ZD0773
001 837
B-K Medical A/S
ABCDE FGHIJKLMNOPQRSTUVXY
Revision level
in this case: E
Bar Code Label
The Part number of the board, for example ZD0773, is the order number of the board.
The number is changed if the modification of the existing PCB is found to be too
extensive.
The serial number of the board is a unique number used for tracking purposes.
The revision code (ABCDEFGH.......) is related to the modifications made on the board.
The revision letter is marked when modifications are made. When boards are ordered
from the B-K Medical stock it is important that the serial number of the scanner is stated.
The serial number ensures that the correct revision is shipped.
The PCB version (number e.g. 3) is printed on the circuit board.
The Version and Revision code can be read electronically.
Note! Some modules (assemblies) that consist of a number of PCBs does only have a
Part No. and a serial number but no revision. E.g. Front-end assembly - ZN2226 consists
of ZE0815 and 16 ZE0817.
Information about Hardware/ Software compatibility can be found in the B-K Medical Hot
Line on the support section of www.bkmed.com (requires password)
BI2201-E
1-11
General Information
4.1
Transducer Phantom, model 254 (WQ 0973) for checking a 2202 system equipped
with Array- and Single-element Transducers.
Transducer Phantom, type 251 (WQ 0972) for checking a 2202 system equipped with
the Endosonic Probe 1850 or 2050.
Further it is recommended to bring a 150 ml bottle containing preserving fluid for the
phantom(s):
Prescription for preserving fluid:
94 ml Glycerol (85%)
50 ml Rodalon (10%)
1000 ml purified Water
1-12
BI2201-E
General Information
Troubleshooting (Section 4)
For accessing BIOS (new image on Harddisk drive)
4.3
4.4
4.5
Hex key 4 mm for adjusting monitor tilting base friction (CRT monitor only)
QA0229
Adjustment tool for Pro Focus
This tool is needed for adjusting the monitor and keyboard bracket mechanics.
4.6
Safety Tester
HV Tester
BI2201-E
WB 1275
WJ 0246
1-13
General Information
1-14
BI2201-E
2202
Electrical Safety Test Record
Equipment under test
Serial no:
Location:
Test Limits
Mains to Case
Console
5 M
0,2 ohm
Monitor shelf
0,2 ohm
Monitor
0,2 ohm
MCM connector
0,2 ohm
0,2 ohm
0,2 ohm
ECG
0,2 ohm
Earth
Leakage
Normal
500 A
Earth
Leakage
Reverse
500 A
Earth
Leakage
1000 A
Earth
Leakage
1000 A
2.2 kV (DC)
Cassette
No Flash over
2.2 kV (DC)
Monitor
No Flash over
Signature
1 of 2
The test complies with IEC 60601-1 regulations for medical equipment, safety class I, Type B (ECG type BF).
Important:
High Voltage testers deliver hazardous currents. Therefore these testers should only be operated by technicians who are aware of the
hazards involved.
Disconnect transducers and accessories from the equipment under test. Test limits are valid only for the equipment itself without any
applied part(s) connected.
The complete test must always follow immediately after a repair made on the 2202 Ultrasound Scanner, and always on the fully assembled
unit.
General remarks and settings for Protective GND test:
Cables:
Tester settings:
Connect the scanners Main cable from IUT POWER (on Tester) to power inlet on the scanner.
Connect the clip/probe lead to the test point and keep it firmly pressed against it as the current is 25 Amps when applying this test
Connect the mains cable for the scanner to the IUT POWER outlet (on Tester) to power inlet on the scanner
When activating the Leakage test then press the power button on the scanner and wait for it to start up and then record the Current before
closing down the scanner as normal turn off.
HV Insulation Tester
HV probe
HV test plug for mains WJ0246 can be used if the HV Tester cannot connect directly to scanner/monitor mains input
Test Cassette:
Connect HV tester plug for mains into socket for the scanner mains input. Insert tip of HV probe to the metal plate
at the right transducer connector slot. Start the test increasing the HV to 2.2 kV (DC) and apply this voltage for
maximum 5 sec. There must be no flash over nor breakdowns.
Test Monitor:
Connect the monitor power cable to the HV tester plug for mains. Insert tip of HV probe to screw next to monitor
power inlet. Start the test increasing the HV to 2.2 kV (DC) and apply this voltage for maximum 5 sec. There must
be no flash over nor breakdowns.
2 of 2
Service Notes
Section 2
SERVICE NOTES
Service Notes
To improve the performance of the Pro Focus Ultrasound System 2202, small changes
in the hardware and firmware may be made. All information about these changes, and
any other change or correction to this Service Manual will be released in the form of
Service Information.
Note
All existing and future Service Information concerning the Pro Focus 2202 system
should be placed in this section.
Edition Number
If any major changes are made to the Pro Focus 2202 system and/or the Service
Manual, a new edition will be released.
Any new editions can be recognized by the edition number given in the footer.
BI2201-E
2-1
Service Notes
2-2
BI2201-E
Checking Procedure
Section 3
CHECKING PROCEDURE
CONTENTS
3.1.
Introduction .................................................................................................................................................. 2
3.1.1.
3.2.
PAGE
3.2.1.
Visual Inspection..................................................................................................................................2
3.2.2.
Monitor ................................................................................................................................................2
3.2.3.
3.2.4.
Power Up.............................................................................................................................................3
3.3.
Functionality................................................................................................................................................. 3
3.3.1.
3.3.2.
3.3.3.
3.3.4.
Gain/TGC ............................................................................................................................................4
3.3.5.
Size .....................................................................................................................................................4
3.3.6.
Focusing..............................................................................................................................................4
3.3.7.
3.3.8.
3.3.9.
Persistence..........................................................................................................................................5
3.3.10.
Puncture ..............................................................................................................................................5
3.3.11.
Measure ..............................................................................................................................................5
3.3.12.
3.3.13.
M-mode ...............................................................................................................................................6
3.3.14.
3.3.15.
3.3.16.
CFM.....................................................................................................................................................7
BI2201-E
3-1
Checking Procedure
3.1.
Introduction
The Checking Procedure is to verify proper operation of the main functions of the 2202
System. The procedure should be used during installation, incoming inspection, regular
maintenance and before and after repair. The procedure is arranged in sections and
must be followed from the beginning when performed.
Notes
If the Checking Procedure cannot be performed successfully or if there is a specific fault
in the System, refer to Section 4, TROUBLESHOOTING or Section 5, ADJUSTMENT
PROCEDURE.
The Checking Procedure does not include a check of the different transducer types.
If in doubt about the functions of the 2202 you should consult the user guide.
Important!
Any inside repairs performed on the 2202 must be followed by an Electrical Safety Test.
3.1.1.
Transducer Phantom, model 254 (WQ 0973) for checking a 2400 system
equipped with Array- and Single-element Transducers.
Transducer Phantom, type 251 (WQ 0972) for checking a 2400 system
equipped with the Endosonic Probe 1850.
3.2.
Basic Checks
3.2.1.
Visual Inspection
Check the overall appearance of the 2202.
Especially check for scratches and stain's on the paint.
3.2.2.
Monitor
Check that the monitor is securely fastened and is easy to tilt/rotate.
3-2
BI2201-E
3.2.3.
Checking Procedure
Before Power Up
The 2202 System complies with the IEC 60601-1 safety regulations and as such it must
be connected to a mains outlet having a safety ground connection. Therefore: Check
that the 2202 mains cable is terminated with a three-pole plug that fits a three-pole
mains outlet. For safety reasons do not use extension cables!
3.2.4.
Power Up
1.
2.
Switch On the 2202. Follow the progress bar in the start-up screen. (It is
normal that the start up will take approx. 2 min and 40 sec.)
3.
Verify that the monitor opening layouts are shown in Fig. 3.3 -1. Note that the
menu shown (settings) differs from transducer type.
Check that the current date and time is displayed on the monitor:
Functionality
The chapter will cover some of the basic functions in the scanner. You can read more
about the user interface in the user's guide, as this is not the scope of this document.
3.3.1.
Monitor Checking
1. Press ALT+SHIFT+G
2. Press Enter to switch to next image.
3. If you have problems meeting the above requirements, consult chapter 5,
Adjustment Procedure.
BI2201-E
3-3
Checking Procedure
3.3.2.
Transducer Inputs
1.
Move the transducer connector to one of the other input module(s), Press
Freeze and confirm that the transducer can be identified.
2.
3.3.3.
Track Ball
The track ball should move smoothly. If it too tight or too loose you can adjust it using
the adjusting tool QA0228 (shipped with System 2202).
3.3.4.
3.3.5.
Gain/TGC
1.
2.
Adjust the gain using the gain potentiometer and observe the effects on the
B-image.
3.
Adjust the TGC curve using the TGC Potentiometers and observe the effect
on the B-image.
Size
Verify the function of the -/+ button (above the track ball). Press the left side of the
button (-) to decrease the size of the objects (higher scale units) and the right side of
the button (+) to enlarge the size of the objects (lower scale units).
3.3.6.
Focusing
Note: This section is only valid for array transducers.
3.3.7.
3-4
1.
Check that there is a focus zone mark on the left side of the ultrasound picture.
2.
There are two ways to use the focus functionality. You might check them both.
3.
Press Focus and move the track ball up and down and check that the arrow
moves accordingly on the monitor screen. Also check that max. lateral
(horizontal) resolution is at the position of the arrow - use the phantom, model
254, to check the focusing. The size of the focus zone can be regulated using
the -/+ button.
4.
Using the track ball, click on the middle of the focus zone mark. Now using the
track ball you can move the focus zone. If you click on the top or bottom piece
of the focus zone, you can regulate the size of the focus zone using the track
ball.
5.
Frame Rate
1.
2.
Connect a Array transducer, press the Res/Hz Up/Down and observe that
increased frame rate narrows the scanning field.
BI2201-E
Checking Procedure
Image Processing
Use phantom, model 254, to check the following:
3.3.9.
1.
Flip Dyn Range Up/down and observe the contrast effect on the image.
2.
Using the tracker ball, click in the Grey Scale bar in the right top of the US
image. Step trough the different gray scales. Observe the effect on the
image.
1.
2.
3.
Press Select to see all valid values. Click on desired value to select it.
Check the effect on the image.
4.
5.
Press the -/+ button to select the desired value. Check the effect on the
image.
Persistence
3.3.11. Measure
1.
2.
3.
4.
In some of the choices, you can see an arrow in the right side. By clicking
on it, you may find more valuable measurement types here.
1.
2.
Observe that a horizontal Track ball movement scrolls through the most
recently recorded B mode images.
BI2201-E
3-5
Checking Procedure
3.3.13. M-mode
Note:
1.
2.
Start the transducer and check that the echoes now are displayed both as a
normal image and as horizontal lines that reflect the echo information along
the M-mode line.
1.
2.
3.
Note:
Pressing Split will switch between scan "A" and scan "B". Holding Split
pressed for a while will switch of the Split Screen mode
3-6
Note:
1.
Press on the Doppler potentiometer and place the sample volume over a
vessel.
2.
Observe that sound can be heard in the speaker and a Doppler spectrum is
shown.
3.
Audio Volume - Press Volume up/down and check that the volume of the
audio signal can be adjusted.
4.
Gain - use the smaller topmost Gain Control and adjust the gain up and
down. Observe that spectrum noise and sound changes.
5.
Filter changes the setting of the wall filter. Observe that the low frequency
components close to the base-line disappear.
6.
Range press Dyn Range up/down and observe that the resolution of the
spectrum decreases when the range is increased.
7.
Baseline - press Baseline up/down and check that the spectrum is offset.
8.
Doppler Sample Volume - Press D/M Cursor to change the size of the
sample volume, using the trackball.
9.
Doppler Angle press Angle up/down and adjust the angle of the sample
volume with the trackball. Check that the scale is changed.
10.
Steered Doppler press Steer up/down to angle the Doppler beam. Press
steered to toggle between left angling, right angling and no angling.
Note:
BI2201-E
Checking Procedure
3.3.16. CFM
BI2201-E
1.
2.
Unfreeze and place the transducer over a vessel. Move the colour box
using the Track ball and verify flow in the colour box.
3.
Gain - Increase CFM gain using the smaller topmost Gain Control and
observe that the colour box is filled with coloured noise dots.
4.
Base line - move the base line up and down with the base line control.
Check that the colour scales is offset allowing measuring a larger range of
velocities in one direction.
5.
Colour resolution- press the resolution control Scale up/down. Increase the
resolution and check that the number of lines in the CFM image is increased
and frame rate is decreased.
3-7
Checking Procedure
3-8
BI2201-E
Troubleshooting
Section 4
TROUBLESHOOTING
CONTENTS
PAGE
Introduction ................................................................................................................................... 3
Replaceable Parts......................................................................................................................... 3
Keys to Troubleshoot Specific Problems (Hints)........................................................................... 4
Loss of time.............................................................................................................................. 4
Transducer or scanner............................................................................................................. 4
Service Mode ................................................................................................................................ 5
Enter Service Mode ................................................................................................................. 5
How to set scanner to Normal (console) Mode - exit Service Mode........................................ 6
How to start console application for service use...................................................................... 7
How to start Windows Explorer................................................................................................ 7
How to start the Control Panel ................................................................................................. 7
How to start the Device Manager............................................................................................. 7
How to shutdown the scanner.................................................................................................. 7
How to restart the scanner....................................................................................................... 7
How to delete patient-related files............................................................................................ 7
Update Hardware..................................................................................................................... 8
Event Viewer .......................................................................................................................... 11
Scandisk................................................................................................................................. 12
Disk Fragment........................................................................................................................ 13
Keyboard Test........................................................................................................................ 14
Installing a XP Embedded Image................................................................................................ 16
Preparing a clean hard disk for image installation ................................................................. 16
HIPAA (Health Insurance Portability and Accountability Act) ..................................................... 16
Background ............................................................................................................................ 16
Enable patient privacy............................................................................................................ 17
Create users........................................................................................................................... 17
Table of passwords ..................................................................................................................... 18
Dicom .......................................................................................................................................... 19
Host Setup ............................................................................................................................. 20
Worklist Setup ........................................................................................................................ 23
PACS Setup ........................................................................................................................... 24
Printers Setup ........................................................................................................................ 25
Troubleshooting ..................................................................................................................... 26
Testing errors.................................................................................................................... 26
Status................................................................................................................................ 27
Ping ................................................................................................................................... 28
How to recover from a crashed main database .......................................................................... 29
How to recover patient database from a crashed system........................................................... 29
Patient Database Recovery tool ............................................................................................ 29
Log files ....................................................................................................................................... 30
How to grab the log files ........................................................................................................ 30
How to send log files via email.................................................................................................... 31
Initialization ............................................................................................................................ 31
Verification (or how to use the feature).................................................................................. 32
Troubleshooting ..................................................................................................................... 32
How to configure scanner to use an external hard disk to save scans instead of the internal hard
disk .............................................................................................................................................. 33
4.13.1.
How to configure scanner to use a network drive.................................................................. 35
4.13.1.1.
Configure Microsoft Windows-based PC .......................................................................... 35
4.13.1.2.
Configure Network ............................................................................................................ 36
4.13.1.3.
Configure Scanner ............................................................................................................ 36
4.13.2.
How to transfer file from one hard disk to another................................................................. 37
4.13.3.
Using IDE (44 pins) to USB adapter ...................................................................................... 37
4.1.
4.2.
4.3.
4.3.1.
4.3.2.
4.4.
4.4.1.
4.4.2.
4.4.3.
4.4.4.
4.4.5.
4.4.6.
4.4.7.
4.4.8.
4.4.9.
4.4.10.
4.4.11.
4.4.12.
4.4.13.
4.4.14.
4.5.
4.5.1.
4.6.
4.6.1.
4.6.2.
4.6.3.
4.7.
4.8.
4.8.1.
4.8.2.
4.8.3.
4.8.4.
4.8.5.
4.8.5.1.
4.8.5.2.
4.8.5.3.
4.9.
4.10.
4.10.1.
4.11.
4.11.1.
4.12.
4.12.1.
4.12.2.
4.12.3.
4.13.
BI2201-E
4-1
Troubleshooting
4-2
BI2201-E
Troubleshooting
4.1. Introduction
This section is divided into separate subsections some of which cover specific areas within the
troubleshooting others describe the overall test facilities in the 2202.
But before you jump to any of the subsections you should first consult Section 4.3 which might
give you, if not the solution, then some ideas of how to troubleshoot this particular problem of
yours.
To prevent injury to persons and damage to the Pro Focus be aware of the following:
WARNINGs:
* Opening the Power Supply can expose live parts.
* Any work done on the open instrument with power On must only be done by B-K Medical or
their authorized representatives, who are aware of the hazards involved.
* Any repair work done on the 2202 system must be followed by an electrical safety test to verify
a continuous safe operation of the system.
* The Core board contains a lithium battery. The battery must only be replaced by a person
having special knowledge as described in sec. 4.2 Replaceable Parts
CAUTIONs:
* Switch Off the 2202 before connecting or disconnecting the Pro Focus to any peripheral units.
Failure to do so could damage the equipment.
* The PC Boards in the 2202 are sensitive for static discharge. Therefore, when handling PC
Boards always take steps to prevent static discharge - see Section 1.5, Special Tools and
Equipment.
BI2201-E
QB0041
Lithium battery
Type of unit
effected:
All
ZN2220
ZN2230
ZN2231
All
ZD0784
ZD0795
UL0041
ETX Module
All
Hard disk
All
ZD0757
Back Plane
From 1864900
4-3
Troubleshooting
4.3.1.
Loss of time
If you remove the ETX board from the Core board, the time settings will be lost. The same is
valid, if you remove the lithium battery on the Core board.
Check the lithium battery on the Core board (also check for bad contact).
4.3.2.
Transducer or scanner
If possible, connect a single-element transducer to bypass the FE64+ board and the beam
former board.
Connect the scanner to another mains outlet, preferably in another room. A change in the
noise pattern indicates noise on the mains.
Compare different transducer types and frequencies (MFI). External noise is often
frequency specific.
Note:
Before contacting the support group in Denmark for help with noise problems it is recommended
that you try out the suggestions above. Use the image storage facility of the 2202 to save a
couple of images showing the problems. The images can be attached to an e-mail.
4-4
BI2201-E
Troubleshooting
4.4.1.
Upgrade software
Check of Hard Disk
Access Windows Explorer
and other service related issues
BI2201-E
4-5
Troubleshooting
4.4.2.
4-6
2.
Click on Yes
3.
Click on OK
4.
BI2201-E
4.4.3.
Troubleshooting
4.4.4.
4.4.5.
4.4.6.
4.4.7.
4.4.8.
4.4.9.
BI2201-E
4-7
Troubleshooting
3. If the status field contains any needs update then click on Program otherwise quit the
Coreboard Flash Utility
Needs update
4-8
BI2201-E
Troubleshooting
Note! The screen will be blank for several minutes during this programming - this is quite normal
you just have to wait!
4. Wait for the Core board Flash Utility to finish programming i.e. until
Working field is no longer blue and the last lines in the Output/Progress window is :
Updating Flash Table in
AVR
Flash Table is updated
5. When the Coreboard Flash Utility has finished programming you should check for error
messages in the Output/Progress field. Errors will have been marked out by a line of
**************** - in that case you have to programme again. Note that you must scroll up to see
all text from the Output/Progress
BI2201-E
4-9
Troubleshooting
If an error has occurred then it will look like the below example in that case you should click on Clear
List and then Program as described in step 3.
Failed!!!
8. After it has turned off wait 30 seconds and then turn on the scanner again. This will complete
the Update Hardware function
4-10
BI2201-E
Troubleshooting
1. Left click on Event Viewer and select a log to view. The log will appear in the right hand
pane.
2. If you want to look at an event, double click on it. A dialog box will open showing information
about what caused the error.
3. You can navigate from error to error in the same log by clicking the up and down arrows.
4. If you need to record an error so you can email it to a technical support person, click on the
button right below the down arrow. This will copy the error to the clipboard. It is now
available to be pasted into a text file, you can send to Service at B-K Medical.
5. Close the dialog box when you are done reviewing the event log.
Event Viewer displays these types of events:
BI2201-E
4-11
Troubleshooting
Error: A significant problem, such as loss of data or loss of functionality. For example, if a
service fails to load during start-up, an error will be logged.
Warning: An event that is not necessarily significant, but may indicate a possible future
problem. For example, when disk space is low, a warning will be logged.
Information: An event that describes the successful operation of an application, driver, or
service. For example, when a network driver loads successfully, an Information event will be
logged.
Success Audit: An audited security access attempt that succeeds. For example, a user's
successful attempt to log on to the system will be logged as a Success Audit event.
Failure Audit: An audited security access attempt that fails. For example, if a user tries to
access a network drive and fails, the attempt will be logged as a Failure Audit event.
Note: Event Viewer is a part of Microsoft Windows XP. Further help can be gained using the
Event Viewers built-in help.
4.4.12. Scandisk
Scandisk is the program that verifies the logical integrity of a file system on Windows. When
Scandisk encounters logical inconsistencies it does not take actions to repair file system data,
but let you know that there is problem:
Errors found. CHKDSK cannot continue in read-only mode.
1. The scandisk will run automatically in a command prompt box. You are not supposed to
enter anything. When the scandisk has completed, you see this output.
C:\CasaEngine>C:\WINDOWS\system32\Chkdsk.exe
The type of the file system is NTFS.
Volume label is 2202 XP Embedded.
WARNING! F parameter not specified.
Running CHKDSK in read-only mode.
CHKDSK is verifying files (stage 1 of 3)...
File verification completed.
CHKDSK is verifying indexes (stage 2 of 3)...
Index verification completed.
CHKDSK is verifying security descriptors (stage 3 of 3)...
Security descriptor verification completed.
39070048
2922396
1060
0
71324
65536
36075268
KB
KB
KB
KB
KB
KB
KB
4-12
BI2201-E
Troubleshooting
BI2201-E
4-13
Troubleshooting
If you test everything, you should be able to get 0 remaining in the field in the upper left corner.
How to test the Sound
Test method:
Click on the various buttons one by one.
Expected Result: You should hear different sounds.
How to test the Legend LEDs
Test Method:
Click on Legend LEDs.
Expected Result: You should see light in all legend LEDs. The light should be uniform, i.e.
nearly same intensity in all LEDs. (You can alter the intensity using the slider
below the button)
How to test the Indicator LEDs
Test Method:
Click on Indicator LEDs.
Expected Result: You should see light in all 28 Indicator LEDs. The light should be uniform,
i.e. nearly same intensity in all LEDs. (You can alter the intensity using the
slider below the button)
How to do the Ping test
Test Method:
Click on Ping 1 sec.
Expected Result: You should see the ping result (d1) in the Received Key Codes field each
second. When done, click on Stop Ping.
How to test the Tracker ball
Test Method:
Turn tracker Ball up/down and left/right.
Expected Result: Check that the cursor (arrow) follows tracker ball.
4-14
BI2201-E
Troubleshooting
BI2201-E
Test Method:
Expected Result:
Test Method:
Expected Result:
Test Method:
Expected Result:
Test Method:
Expected Result:
Test Method:
Expected Result:
4-15
Troubleshooting
4.5.
4.5.1.
2.
3.
4.
5.
select disk 0
6.
clean
7.
8.
select partition 1
9.
active
10.
assign letter=C
11.
exit
12.
13.
14.
15.
Background
The American Health Insurance Portability and Accountability Act of 1996 (HIPAA) requires
security and privacy of health information. The Privacy Rule is a US federal regulation defining
administrative steps, policies, and procedures to safeguard individuals' personal, private health
information (protected health information or PHI). The Privacy Rule is designed to empower
patients by guaranteeing them access to their medical records, giving them more control over
how their PHI is used and disclosed, and providing a clear avenue of recourse if their medical
privacy is compromised. The rule is designed to protect medical records and other personal
health information maintained by certain health care providers, hospitals, health plans, health
insurers and health care clearinghouses.
The rule defines "protected health information" as health information that
1. identifies an individual and
2. is maintained or exchanged electronically or in hard copy.
4-16
BI2201-E
Troubleshooting
If the information has any components that could be used to identify a person, it would be
protected. The protection would stay with the information as long as the information is in the
hands of a covered entity or a business associate. The protections apply to individually
identifiable information in any form, electronic or non-electronic. The paper progeny of electronic
information is covered (i.e. the information would not lose its protections simply because it is
printed out of a computer), and oral communications are also covered.
4.6.2.
6. Click on Close.
4.6.3.
Create users
You have enabled password protection, but you must now create users to allow access.
1. Click on Login
2. Username is admininstrator
3. Password is superuser (This is the default password from the factory. If you reset it, the
password is blank, and you may have set it to something special.)
4. Click on OK
7. Click on Setup
8. Click on the System-tab
BI2201-E
4-17
Troubleshooting
Here you can to the user administration. Only the administrator user have rights to add new
users, remove existing users, reset the password for the existing user and enable/disable
password protection.
4-18
Purpose
Service Mode
Username
-
Password
viking
Reference
4.4
BKASTRA
4.5
administrator
superuser
4.6.2
Network Configuration
nw
4.8.1
BI2201-E
Troubleshooting
4.8. Dicom
DICOM is a global Information Technology standard that is used in virtually all hospitals
worldwide. Its current structure, which was developed in 1993, is designed to ensure the
interoperability of systems used to:
Produce
Store
Display
Process
Send
Retrieve
Query
Print
Medical images and derived structured documents as well as to manage related workflow.
To be able to use DICOM features on 2202, make sure that the DICOM license has been
entered. Then second try to get the Network Administrator from the hospital to fill out the
"DICOM network description" questionnaire.
In the following chapters we'll cover the DICOM features on 2202 and how to setup the scanner
to work in a DICOM network.
BI2201-E
4-19
Troubleshooting
4.8.1.
Host Setup
This is setup of the scanner.
1. Click on the DICOM tab
2. Enter the AE Title for the scanner and then click on Apply.
3. Click on Configure Network
4. Password is nw
4-20
BI2201-E
Troubleshooting
BI2201-E
4-21
Troubleshooting
8. Default setting is that the scanner will obtain an IP address from the DHCP server in the
network. If you want to use a fixed IP address, click on Use the following IP address and
fill out the fields for IP address, subnet mask and default gateway according to the data
collected from hospital.
9. Click OK until all windows are closed.
10. Finally close the Network Connections window.
4-22
BI2201-E
4.8.2.
Troubleshooting
Worklist Setup
Note! Scanner AE Title must be present before testing can be done.
2. Enter Name. Note, this can not be changed later on. Enter AE Title, IP Address and Port
Number. Select the desired value in the search filter and then click on Test.
Search filter can have these values:
This system only:
The AE Title of the scanner will then be used to differentiate
Ultrasound
(0008,0060)-tag must be US in the record in the worklist
Systems only:
All Systems:
No filtering
3. Click on OK
4. Click on OK
BI2201-E
4-23
Troubleshooting
4.8.3.
PACS Setup
Note! Scanner AE Title must be present before testing can be done.
2. Enter Name. Note, this can not be changed later on. Enter AE Title, IP Address and Port
Number. Select the desired archiving method and then click on Test. Each time a new
PACS server is added, it will be default. In case you want to modify that, change the
Default PACS setting.
Archiving Methods are:
Manual:
You have send the images manually.
Send complete
If you press and hold the patient id button for a couple of seconds, the
examination at end current examination will end and all images and clips (except 3D) will be
of examination:
sent. If you start on a new patient, the send behavior is the same.
3. Click on OK
4. Click on OK
4-24
BI2201-E
4.8.4.
Troubleshooting
Printers Setup
Note! Scanner AE Title must be present before testing can be done.
2. Enter Name. Note, this can not be changed later on. Enter AE Title, IP Address and Port
Number. Select Printer Type and click on Test. Each time a new Printer is added, it will be
default. In case you want to modify that, change the Default Printer setting. Other settings
can be modified on request.
3. Click on OK
4. Click on OK
BI2201-E
4-25
Troubleshooting
4.8.5.
Troubleshooting
When pinging is done successfully, the Dicom echo will be send towards the same target host.
If AE title is set and the Dicom echo still fails, you may get a message like this
Corrective actions can be:
Check target host AE Title and Port
Number on scanner
Check that the target host will reply on
an echo. Talk with hospital IT.
Check that the target host has granted
access to scanner. Talk with hospital
IT.
4-26
BI2201-E
Troubleshooting
4.8.5.2.
.8.5.2. Status
When Dicom is enabled and configured, you will
notice this:
Green LED
Yellow LED
Red LED
Everything is OK. Transferal of images was OK. The button will disappear.
Transfer is going on
Transfer failed. The button will stay.
In left you see the device list. These are the Dicom servers/printers the scanner has tried to
send something to. You also see the status indicated by the colored LED.
If you select one of the items, you can see more details in right area. These are the documents
that are either being send or could not be sent.
Clicking on Cancel will remove the document
from the list and thus stop sending of this
document. Clicking on Cancel All will clear the
list.
In case you regret, click on No otherwise click on
Yes to delete the item.
You also have the option to test each of the Dicom servers/printers in the device list.
Clicking on Details can give an indication on what went wrong.
BI2201-E
4-27
Troubleshooting
4.8.5.3. Ping
Click on Customize, Dicom and then Ping
Enter a target IP address and click on OK. In case you are running with a DNS in the network,
you can also try a name.
If you have success with pinging the target, you will get this message:
4-28
BI2201-E
4.9.
Troubleshooting
4.10.
2.
3.
Click on Start.
4.
Click on OK.
5.
6.
Click on View Archive and see if you can find the patients in the record.
4-29
Troubleshooting
4.11.
Log files
The log files are a collection of files which are saved when clicking Export in the Version tab
(Customize - System - Version)
4-30
1.
2.
Click Customize
3.
Click System
4.
Click Versions
5.
6.
7.
Enter fault description in the Comment / Description field and click Export.
8.
BI2201-E
4.12.
Troubleshooting
Now, the requirements can only be fulfilled with help from the hospital IT department, as we
would be using their SMTP server.
Before you continue make sure that these requirements have been fulfilled.
4.12.1. Initialization
BI2201-E
1.
2.
3.
Start Explorer
4.
Browse to c:\console.
5.
6.
7.
8.
9.
Remove the ; sign and enter the address to receive the log-files here. Fx:
SMTPToAddress=myemail@mycompany.com (This address is fictive.)
10.
11.
12.
13.
14.
4-31
Troubleshooting
Click on Setup
3.
4.
5.
Click on the down arrow next to Export and choose Send Log Files as E-mail
6.
Enter a comment in the field and click on Export. This comment will be written in the mail
and the log files with be attached to it. (If there is any log file of any kind it will be possible
to select the files in the two dropdown lists.)
7.
You will then receive an email on the address you provided as the receiver address.
Subject of this mail will be "Support call from scanner '<Identification number (MAC
address)>'".
4.12.3. Troubleshooting
Question: When I click on Send Log Files as E-mail, I get an error message saying: There is
no mail receiver configured (SMTP To-address). What should I do?
Answer:
Either you did not enter a valid email address, or you forgot to remove the ; sign
from the beginning of the line. Please check your console.ini file to correct the line.
Question: When I click on Send Log Files as E-mail, I get an error message saying: Failed to
connect to SMTP server "<name>" (). What should I do?
Answer:
This means the SMTP server is not valid. You could try to ping it as the first step.
Next step would be to make sure that the SMTP service on the specified server is
alive. This can be done by the hospital IT department only.
4-32
BI2201-E
4.13.
Troubleshooting
How to configure scanner to use an external hard disk to save scans instead of the
internal hard disk
Many situations demand that you use an external hard disk on the scanner. The support for
external hard disk exists already. You can use the external hard disk as an extended USB Flash
key to save scans on the hard disk.
On the hard disk scans (2D / 3D / cine) are stored in this directory: c:\console\documentation.
This folder also contains a patient database (PatientDB.mdb). The database is vital for the
archive system to work, as it contains information about patients that has been entered
manually or fetched from a Dicom worklist. This patient database also has the links for each of
the image-files from the documentation directory. This link is a hard link, which means that the
link consists of the partition data, directory data and file name data. A link could look like this:
c:\console\documentation\20051114_142222_NoID.bmp. The database is a Microsoft Access
database and it has been password protected, so unauthorized access can be prohibited. When
the console software is running, the patient database you will be able to see an additional file:
PatientDB.ldb. This file is a lock-file for the PatientDB.mdb-file. When the console software is
running, you will not be able to copy or view the PatientDB.mdb-file, because of the lock file.
When you examine the directory in service mode you may not find this lock-file anymore, but if
you do, you can delete it if needed. Next time the console application is started, a new lock file
will then be generated.
Finally there is a backup of the PatientDB.mdb in the directory as well, but at time being it is
useless. If your PatientDB.mdb has crashed, the backup has been damaged as well. (We are
looking into this).
The standard solution when you use the external hard disk as an extended USB flash key does
not save the patient database on the hard disk. The tweak below makes it possible to save the
patient database to the external hard disk as well. More precisely this tweak will make the
c:\console\documentation directory almost obsolete. Here is the procedure on to get it work. (If
you are not sure about what you are doing, please do not continue. A mistake could make the
scanner unstable.) Your external hard disk has been partitioned (NTFS) and formatted before
you plug it into the scanner. Only external hard disk types supported by the scanner will work.
BI2201-E
1.
2.
Connect you external hard disk and wait for approx 2 min.
3.
4.
5.
6.
4-33
Troubleshooting
7.
Now select your external hard disk and make sure that it has NOT any of these drive
letters: A, B, C, D, E. (E is reserved for the USB flash key, so even if the USB flash key is
not connected, we still want to keep the reservation.)
8.
9.
10.
11.
Make a directory (fx Documentation). (You can use the root of you external hard disk, as
this will not work).
12.
13.
14.
15.
16.
17.
Save console.ini
19.
20.
21.
Switch back to explorer and examine that you find the scans and the patient database on
your external hard disk.
Default behavior for console software is that it would look in the console.ini file for the
documentation paths. First it will look DocumentationDirPath1, if this is not available, it will look
for DocumentationDirPath2. Default values for DocumentationDirPath1 is q:\documentation and
for DocumentationDirPath2 is c:\console\documentation. This works out as a safety line for this
tweak. In case the external hard disk is not connected, console software will use the second
path. If the console software does not find a patient database, a new empty one will be created.
It can be confusing having two different patient databases and two different set of images etc.
4-34
BI2201-E
Troubleshooting
4.13.1.1.
BI2201-E
4-35
Troubleshooting
4.13.1.2.
Configure Network
You can connect the Microsoft Windows-based PC and the scanner different way. Its not the
scope of this document to cover that. Main idea is that you need to configure network for the
Microsoft Windows-based PC first. Then configure network on the scanner. This can be done by
entering Service mode, if Dicom is not available on the scanner. Just make sure the network
connection between scanner and PC is working.
4.13.1.3.
Configure Scanner
Click on Customize, System and then Network Drive
Enter the UNC path in the field. The UNC path is the path to the folder you created in the first
step. So fx. If your PC has this IP Address: 192.168.1.10 and the folder name is 2202 then you
enter this: \\192.168.1.10\2202
Now enter username and password for the user that is allowed to save files in the folder created
in the first step. If you are on a Domain you may have to write the username like this:
Domain\username. (eg bkmed\sla, where bkmed is our domain name and sla is my username.)
Click on Test Connection to test the connection.
A success full test looks like this:
When you have success fully tested the connection, click on Close. (Click on Yes to save, if
asked.)
4-36
BI2201-E
Troubleshooting
2.
Then you must start the scanner and see that everything is working perfect.
3.
4.
Now you take the old hard disk and mount it on the adapter and connect the adapter to
the USB connector (using the cable coming with the adapter).
5.
Wait until Windows has found the hardware and the hardware is ready to be used.
6.
Start Windows Explorer and use the hard disk (which now is your USB hard disk)
Cons
1. On the first core boards ZD0773 the USB
standard is 1.1. This gives a very limited
transfer speed.
2. Not all the adapters will be detected by
Windows. If your adapter is not detected,
only way out is to replace it.
BI2201-E
Pros
1. On the latest core boards ZD0773 the USB
standard is 2.0. This gives a very high
transfer speed.
2. Its more convenient to work with.
3. You can use the adapter to connect the hard
disk almost everywhere.
4-37
Troubleshooting
2.
3.
4.
5.
6.
7.
8.
9.
Then you must start the scanner and see that everything is working perfect.
10.
11.
Cons
1. Its not convenient to work with.
2. You must assemble and disassemble a
couple of time
4-38
Pros
1. Transfer speed is very high, as the IDE bus
is being used.
BI2201-E
4.14.
Troubleshooting
Software Structure
In this chapter we will look at the directory structure and then on some of the important files.
The hard disk has been partitioned into one single partition. This is mapped to drive letter C.
The CD drive has been mapped to drive letter D. Drive letter E is reserved for an optional USB
flash key.
The important directories have been highlighted in the image above. We will in the following
concentrate on these five directories.
4.14.1.1.
CasaEngine
All hardware configuration files are located in this directory. There are two sub directories.
4.14.1.2.
Console
This is the main software directory. There are 10 subdirectories which all require a short
presentation. In the console directory you find the important console.ini file, which is the
software configuration file.
Bodymarks
Each of the body marks is represented by an image file. The default naming policy is
bodym_<number>.bmp (fx. Bodym_001.bmp). If you find any files which do not follow the above
rule, they are custom made. In case of file rescuing these might be needed.
CasaWare
This directory includes supporting software for the console application.
BI2201-E
4-39
Troubleshooting
In this directory you can the find the Lagarith Codec. This codec is used in the Cine clips. First
time you want to view a clip on a PC, you must install the codec. This directory is copied to the
CD every time you copy exams (depending on your configuration.)
Database
This is a very
important
directory, as
(almost) all
databases are
stored here.
The main
database is
2202.mdb.
Then there is
the backup of
this file: 2202_backup.mdb. The Dicom setup is stored in DICOMSetupDB.mdb.
PatientDB.mdb is blank copy of the patient database.
The sub directory org contains an empty 2202.mdb
Documentation
This is where all exams are stored. All 3D scans, all US images and all clips. Furthermore you
find the actual patient database PatientDB.mdb and a backup of this here.
Report
Files needed to create a report printout are stored here.
StagingArea
This is the staging area directory. All exams that have been ordered to burn on a CD will be
here until the CD has been burned. Then this directory will be empty. If you have something in
this directory, each time the console application starts you will see a message about it.
Dicom
This directory stores both the Dicom program files as well as links to documents that are being
send to a Dicom device.
4-40
BI2201-E
Troubleshooting
XPE_Backup
This directory is important after software upgrade. If you make a clean installation, this directory
is not used. Inside this directory a collection of different files can be found. These are copied
from different places on the hard disk. Here is the directory listening:
File / Directory name
2202.mdb
Console.ini
Lbk.bin
PatientDB.mdb
Viking.ini
Bodymarks [DIR]
Documentation [DIR]
4.15.
Comment
Main setup database
Main configuration file
License File
Empty Template
Actual configuration
All bodymarks
All patient scans and
Patient database
Copied from
C:\Console\Database
C:\Console
C:\
C:\Console\Database
C:\
C:\Console
C:\Console
How to access the content of the CD in the CD-drive from Windows Explorer in service mode
Its not possible to read the content of the CD in the CD-drive in Windows Explorer, due to some
Microsoft setting. You can access the content using a Command Prompt.
4.16.
2.
3.
CD-Rom drive
BI2201-E
4.
5.
Click on the CD-rom device (without any brand name) and click on Action and then
Uninstall
6.
Click on OK.
7.
Troubleshooting
4.17.
8.
9.
Examine the CD-drive. It should be: fx. TEAC DW-224E-C USB Device
10.
4-42
BI2201-E
4.18.
Troubleshooting
License Codes
The software options on the scanner are controlled by a license enabling a specific option. The
license(s) are kept in a file on the hard disk. The license combined with the Scanner Identity
which is a unique number for each scanner opens up for the option so a specific license will
only work on one scanner. The license(s) and Scanner Identity can be found under Customize System - License.
Scanner Identification
Either serial no. or
MAC address
The scanner identification is the MAC address of the ETX board on scanners with serial no.
below 1864900. Scanners with a higher serial no. have the serial number programmed into the
Backplane board so the scanner identification is the serial no. in Mac address format.
This means that if changing the EXT board or Core board assembly in a scanner with a serial
no. below 1864900 then new licenses have to be entered. If replacing a backplane board in a
scanner with serial no higher than 1864900 then the MAC address will be the new Scanner
Identification unless the back plane board has been preprogrammed from the factory.
If installing a hard disk from another scanner then the license will be inactive because the
Scanner Identification does not match. In that case the licenses have to be changed using the
license CD that came with the system see 4.18.1 or by re-entering the license(s) manually. If
entering the license manually then the file C:\LBK.BIN must be deleted first. Licenses cannot be
entered on a system which has one or more invalid licenses in the LBK.BIN file.
If the LBK.BIN file contains an invalid license then all licenses will become invalid and the
scanner will behave as a Blue Pro Focus without any options enabled.
Note! The Delete License is only for demo licenses and not permanent licenses.
BI2201-E
4-43
Troubleshooting
1.
2.
3.
4.
Type d:
5.
Type install.bat
When the line: 1 file copied. appears the license(s) have been installed
4.19.
4.19.1. Introduction
The Power Supply On/Off control is placed on the Core board so if you have a problem starting
up the Power Supply it could be the Power Supply itself or the Core board. Note : A number of
things on the Core board Assembly could result in start up problems (Core board it self, Lithium
battery being completely flat, defective EXT board)
4-44
BI2201-E
Troubleshooting
5V Standby OK
BI2201-E
4-45
Troubleshooting
Tester/ATX
connector
Pin
1
3
4
8
9
10
16
Size AWG
Name
Color
16
16
16
20
20
16
20
+3.3V
COM (GND)
+5V
PWR_OK
+5Vsb
+12V
PS_ON
Orange
Black
Red
Gray
Violet
Yellow
Green
ZG0346
connector
Pin
5
12
8
2
3
4
13
When your adapter cable is ready, you can make the test:
1.
2.
Plug in your 20-pins 24-pins to 20-pins adapter connector to the power supply ZG0346
3.
4.
5.
Check LED lights on the tester: (+5V. +12V. +3.3V. +5VBS. PG)
If the power output is working, the LED will light and you will hear a sound
If the power output fails, you won't see the LED light & no sound
6.
7.
Warning: This tester is designed only for power supply output tests. Do not leave it connected
to the power supply unattended. It generates heat and could possibly damage your power
supply.
The pin layout for a standard ATX power connector looks like this:
4-46
BI2201-E
Adjustment Procedure
Section 5
ADJUSTMENT PROCEDURE
CONTENTS
PAGE
5.1. Introduction................................................................................................................................................2
5.2 Trackball Friction Adjustment .......................................................................................................................2
5.3 Adjusting the monitor friction ...................................................................................................................3
5.4 Adjusting the Brightness of the Screen...................................................................................................6
5.5 Adjusting the Reading Lights....................................................................................................................6
BI2201-E
5-1
Adjustment Procedure
5.1.
Introduction
All of the adjustable components in the 2202 are factory adjusted and need no further
adjustment. Do not attempt to make other adjustments than stated in this chapter.
5.2.
5-2
BI2201-E
5.3.
Adjustment Procedure
2. Eject the CD tray by inserting a paper clip or similar in the small hole on the front
and pull out the CD/DVD drive completely
3
2
1
BI2201-E
5-3
Adjustment Procedure
4.7- 5 Kg.
5-4
BI2201-E
Adjustment Procedure
4. Reinstall the CD/DVD drive by pushing it back into the monitor shelf. Make sure
that it is connected properly - i.e. you should hear a click when pushing it the last
cm.
1
2
BI2201-E
5-5
Adjustment Procedure
5.4.
5.5.
2.
3.
Use the contrast button on the monitor to adjust the bars so that the 100% bar
is bright white and the 90% to 10% bars seem to decrease progressively in
brightness.
4.
Use the brightness button on the monitor to increase the brightness until the
1%, 2% and 3% bars (above the 10% bar) become visible.
5.
Use the brightness button to decrease the brightness until the background
becomes black and the 1% bar just disappears.
Reading "lamps"
5-6
BI2201-E
Mechanical Parts
Section 6
Mechanical parts
CONTENTS
PAGE
6.1.
6.1.1.
6.1.2.
6.1.3.
6.1.4.
6.1.5.
6.1.6.
6.1.7.
6.1.8.
6.1.9.
6.1.10.
6.1.11.
6.1.12.
6.1.13.
6.1.14.
6.1.15.
6.1.16.
6.2.
6.2.1.
6.2.2.
6.2.3.
BI2201-E
6-1
Mechanical Parts
6.1.
6.1.1.
2. Remove all
transducers from the
transducer connectors
and disconnect the
keyboard.
6-2
BI2201-E
Mechanical Parts
BI2201-E
6-3
Mechanical Parts
6-4
BI2201-E
6.1.2.
Mechanical Parts
BI2201-E
6-5
Mechanical Parts
6.1.3.
3. Disconnect the 3
connectors on the flat
cable by pushing the
lock and then pulling the
cable
6-6
BI2201-E
Mechanical Parts
6.1.4.
1. Disconnect the 3
connectors on the flat
cable by pushing the
lock and then pulling the
cable
BI2201-E
6-7
Mechanical Parts
6.1.5.
6-8
BI2201-E
6.1.6.
Mechanical Parts
BI2201-E
6-9
Mechanical Parts
6-10
BI2201-E
6.1.7.
Mechanical Parts
BI2201-E
6-11
Mechanical Parts
6-12
BI2201-E
6.1.8.
Mechanical Parts
BI2201-E
6-13
Mechanical Parts
6-14
BI2201-E
Mechanical Parts
2. Remove all
transducers from the
transducer connectors
and disconnect the
keyboard.
BI2201-E
6-15
Mechanical Parts
6-16
BI2201-E
Mechanical Parts
1
2
BI2201-E
6-17
Mechanical Parts
6-18
BI2201-E
Mechanical Parts
BI2201-E
6-19
Mechanical Parts
6-20
BI2201-E
Mechanical Parts
BI2201-E
6-21
Mechanical Parts
1
1
2
2
6-22
BI2201-E
Mechanical Parts
BI2201-E
6-23
Mechanical Parts
6-24
BI2201-E
Mechanical Parts
2
1
BI2201-E
6-25
Mechanical Parts
2.4.
6-26
BI2201-E
Mechanical Parts
BI2201-E
6-27
Mechanical Parts
6-28
BI2201-E
Mechanical Parts
BI2201-E
6-29
Mechanical Parts
6.2.
Part list
6.2.1.
16
17
10
(12)
19
13
18
20
4
12
11
8
21
3
14
7
6-30
22
15
BI2201-E
Item no.
Part no.
ZN2242
ZN2243
ZE0798
Front-end board
ZE0815
Front-end board
ZE0822
Front-end board
ZE0820
Description
Quantity
ZE0797
16
ZE0817
16
ZE0823
16
ZE0821
16
ZD0774
ZN2220
ZN2231
Core board assembly (Core -, Video - , ETX board and HD) Green
ZN2230
Core board assembly (Core - , Video - , ETX board and HD) Blue
ZD0773
Core board
ZD0789
ZH0758
Video IF board
ZD0784
ZD0795
UL0041
ZG0346
Power Supply
ZN2201
0-1
ZN2223
0-1
10
ZN2202
MCM module
0-1
10
ZN2222
0-1
11
ZH0757
Back Plane
12
ZN2203
ECG Module
1
0-1
14
AQ1728
15
AQ1725
16
AQ1725
0-1
17
AQ1729
0-1
18
AT2314
SC3157
Cover no options
1
1
19
BI2201-E
Mechanical Parts
SC3158
SC3159
Cover 3D Mover
SC3160
SC3161
Cover ECG
20
DM0295
30
21
QB0041
22
VF0118
Fuse 10 A T
1
2
6-31
Mechanical Parts
6.2.2.
101
128
123
102
127
124
125 126
103
104
105
106
109
111
110
113
112
107
114
115
117
108
116
118
119
120
121
122
Item no.
Part no.
Description
101
ZV0065
101
ZV0069
15 CRT Monitor
Quant.
Item
no.
Part no.
Description
Quantity
N/A
116
GU1792
117
GU1799
101
ZV0070
15 CRT Monitor
118
YT7410
Screw M410 mm
102
ZV0071
17 Flat Screen
119
DP0811
Tilt Stop
103
DH0927
120
YO0687
Washer
104
YS1320
121
YO0800
Spring washer
105
YO0402
Washer
122
YQ2635
Screw M635 mm
106
DB3527
Bushing
123
HP0025
Loudspeaker
107
YB0007
124
UL0036
CD R/W - DVD
108
EA0201
125
ZH0759
109
YS9352
Screw M38 mm
126
GV2561
Plate
110
DZ9766
Tilting Base
127
YT4408
Screw M48 mm
111
YJ3133
O-ring
128
112
DP0822
Slide
113
YS1412
Screw M412 mm
114
DZ9767
Tilting Seat
115
YO1626
Sliding ring
6-32
DD0581
Cable cover
DH0459
BI2201-E
6.2.3.
Mechanical Parts
212
213
214
215
216
211
217
201
218
219
221
220
222
223
204
204
207
224
205
208
225
206
209
210
Item no.
Part no.
Description
Quant.
Item
no.
Part no.
Description
Quantity
201
UD0152
217
DH0912
204
DH0871
Thumbscrew
218
DS0967
Hand rest
205
YT1408
Screw M4 8 mm
219
DZ9001
Top plate
206
DZ9003
Back Cover
220
YS0823
Screw M8 12 mm
4
2
207
AO0462
USB cable
221
DZ9002
208
AO0455
222
DS0971
209
DF0257
222
DS0964
210
DF0256
Twin Wheel
222
DS0968
211
ZN0046
223
DK1300
211
ZN0047
223
DK1301
212
SN0220
Dial - small
1-2
223
DK1291
213
SN0221
Dial - Gain
224
DH0678
214
DP0912
Potentiometer button
225
DZ9751
Front cover
215
DZ9755
Transducer holder
216
DZ9754
Gel holder
BI2201-E
6-33
Mechanical Parts
6-34
BI2201-E
Preventative
Maintenance
Section 7
PREVENTATIVE MAINTENANCE
CONTENTS
PAGE
7.4.
7.5.
7.6.
BI2201-E
7-1
Preventative Maintenance
7.1.
Introduction
The purpose of the Preventative Maintenance is to ensure the performance and
stability over the years of use.
Note: The Preventative Maintenance should be performed once every year.
Preparation of the Preventative Maintenance is essential as it in most cases will take
place at the hospital as part of a Continuity Agreement.
The Preventative Maintenance consists of eight steps. Some of these steps refer to
specific sections in the Service Manual. If errors occur during the procedure go to
TROUBLESHOOTING, Section 4.
7.2.
Necessary Equipment
The equipment, necessary for a preventative visit is divided into 3 packages.
A: Parts necessary to ensure the performance and stability of the system.
Check in the Service Informations (Hot-Line) if there are any modifications to be done.
B: Parts necessary to meet customers complaints.
Before arranging the visit, check if the customer has any complaints and bring the
necessary spare parts and tools to solve the problem.
C: Tools and equipment necessary to perform the Preventative Maintenance.
The only equipment necessary to perform the Preventative Maintenance is the
equipment used in the Electrical Safety Test and phantoms for the Checking
Procedure:
For the Checking Procedure:
Transducer Phantom, model 254 (B-K no.WQ 0973) for checking a 2202
system equipped with Array- and Single Element Transducers.
Transducer Phantom, model 251 (B-K no.WQ 0972) for checking a 2202
system equipped with the Endosonic Probe 1850.
For the Electrical Safety Test:
Testers required:
Safety tester
HV Tester
Note:
7-2
It is recommended to bring additionally the Static Control Service Kit, type 3M 8501
(B-K no. WQ0969) or similar in case one of the static sensitive PC Boards is to be
removed from its location in the cassette.
BI2201-E
7.3.
Preventative
Maintenance
What to do
1. Users Comments and Corrections
Check with the user that the system is operating satisfactorily. A number of
malfunctions can be corrected by a set-up adjustment! If an error is obvious
correct it at this state.
2. Modifications
In accordance with the Service Informations (HOT-LINE) perform
necessary hw/sw modifications to ensure performance and stability of the
system.
3. Mobility
Check that the keyboard/monitor can be moved up/down
Check that the trackball moves smoothly
Caution - Never add oil or grease to any other movable parts.
4. Fans
Remove the rear cover from the 2202. Connect a mains cable and turn on
the unit. Check that both fans in the power supply rotates. Note that the fans
are controlled by electronics inside the power supply. Reassemble the
scanner.
5. Checking Procedure
To verify proper operation of the main functions perform the CHECKING
PROCEDURE, Section 3.
6. Peripherals
This is to verify the function of the video output and the peripherals used
with the 2202 System.
Make an image/print and check the quality. If necessary adjust the
peripheral using the original instruction manual.
7. Adjustment Procedure
If necessary, perform an adjustment of the Trackball using the
ADJUSTMENT PROCEDURE, Section 5.
BI2201-E
7-3
Preventative Maintenance
7-4
BI2201-E
Theory of Operation
Section 8
THEORY OF OPERATION
CONTENTS
PAGE
8.1.
Introduction................................................................................................................................................ 2
8.2.
8.2.1.
Block Diagram........................................................................................................................................... 3
8.3.
8.3.1.
8.3.2.
8.3.3.
8.3.4.
8.3.5.
8.3.6.
8.3.7.
8.3.8.
Optional Devices..................................................................................................................................... 11
BI2201-E
8-1
Theory of Operation
8.1.
Introduction
This section describes the theory of operation of the Diagnostic Ultrasound System
2202. The description is divided into two separate sub-sections:
General Description
Supported by a complete block diagram, this section aims to give you an
understanding of the interaction between modules, Boards and major function blocks.
It explains briefly the main signal flow through the 2202 system.
Board Description
This section provides an overall description of each Board, explaining the main
function of the function blocks on the board.
8.2.
General Description
The basic hardware architecture consists of three active PCBs connected to a passive
backplane. The active boards are:
1.
A 64-channel analog Front End (FE) board with transmitters, 3 array transducer
connectors, relays, receivers, Harmonic Imaging filters, and TGC amplifiers.
2.
A 64-channel Digital Beam former (DBF) with signal conditioning circuitry and
A/D converter per channel and 16 beam former ASICs plus control logic and
profile memory.
3.
A Core board with real-time scan controller, vector processor, Doppler system
and scan converter/display system. Further, an industry standard PC module
and a hard disk drive is attached to the Core board.
Power Supply
CD-RW Disk Drive*
Single Element Transducer Module** (field option)
Mover Control Module** for 3D application (field option)
Keyboard assembly
Foot switch (accessory)
Internal speakers*
8-2
BI2201-E
8.2.1.
Theory of Operation
Block Diagram
2202
Engine
FE64+ ZE0798
P2
64
HVMux
192 64
Relays
Bipolar
Bipolar
Xmitters
Xmitters
64
DBF ZD0774
XMIT RAM
64 Ch. Beamformer
PMEM 1
32
RAM Test
RAM Test
Filter
1
D/A
Element
Bus
192
conn.
P1
Dig Rotate
64
32
192
P3
Relays
16 x QCCM ZE0797
ATGC
Analog
Osc
192
conn.
Filter
64
Beamf. Control
ID,
Temp.
SEM
I/F
ZH0756
SGLM ZH0651
Mover
ASIC
# 16
A/D
64
ID,
LVDS
serializer
Temp.
Dig. Generator
Analog
Osc,
Tr code
ASIC
#8
D/A
8
PMEM
PMEM 82
RAM Test
Signal file
Single
Element
Probe
ASIC
#1
A/D
8
PMEM 1
RAM Test
64
Receivers
Receivers
192
conn.
A/D
1
Filter
8
PMEM
Backplane ZH0757
BEAM
6 pair
LVDS
lines
Analog TCG
ID
I2C
ID
MCM
SPI
ZH0752/0753/0754/0784
CORE ZD0773
CLOCK
GENERATOR
SHARC DSP
AC Output
Isolation
Transformer
SHARC Bus
HARD
DISK
HARD
DISK
ID
SDRAM
MEMORY
ATGC
BUS
BUFFER
48
AC Mains
Input
Amplitude
detector
LUT
RAM Test
Vector Processor
Power
Power
Fans
PC Back-end
BUS
CTRL
SHARED
RAM
RAM Test
Input
24
Buffer
SEQUENCER
FPGA
32
verter
16
Doppler
Integrator
2x16
16
ID,
Temp.
AVR uC
I C, WDT
IEEE
1394
VECP FPGA
2 x Firewire
I2C
Resampling
Contour
Enhance
Down
I2C
iLink
Sliding
16 ConTGC 16
Filter
Timing Bus
HOSTIF
FPGA
Contrast
LUT
RAM
Test
2x16
B.P. I/F
Local Bus
Spectral
Doppler
DSP
CFM
Input
Buffer
RAM
Test
Spectral
Doppler
output
buffer
8
CFM
LUT
RAM Test
CFM
32 CFM
DSP
DSP
Digital Audio
CFM
output
buffer
RAM
Test
8
IPC
Ethernet,
DICOM
IDE
PCI
Video/Merge
FPGA
Image
Format
Combiner
24
Serial
RS232
Scan Converter
Frame
Buffer
Host bus
18/24
SVGA/LCD
ETX
Card-PC
Audio
Codec
USB
Line in
16/24
Digital Audio
AVIB ZH0758
Triple
DAC
Line out
Video
encoder
Video
decoder
NTSC/PAL
NTSC/PAL
DSC
USB
Foot
switch
Keyb.
Control
Panel
Audio out
Audio in
VCR
VCR
VGA
Display
Memory
RAM Test
Review
Buffer
RAM Test
RGBS
Mouse
Trackball
Alphanumeric
Keyboard
SCON FPGA
ID
VGA
USB
USB
HUB
PCI to
host
bus
Serial Reset
LAN
USB
PCU
Parallel
On/Off
100Base-TX
8
8
ID
USB HUB
Doc.
Trig
Light
CVBS B/W
CVBS
Ext.
VGA
15" VGA
Monitor
RGBS
2
S-Video out
RGBS out
CVBS
S-Video
File:
Date:19 Jan, 2005
USB
ZV0065
USB
CD RW
NP0166
HUB
ZH0759
BI2201-E
8-3
Theory of Operation
8.3.
Board Description
8.3.1.
Front End, FE
Xducer
Conn
A
192
16 QCCM Modules
Rcv
Xducer
Conn
B
192
192
Relays
64:192
HV Mux
64
64
TGC
64
Beamformer
64
Xmit
Xducer
Conn
C
64
64
Digital
Rotate
32
Board
Control
FPGA
Backplane
192
Xmit
RAM
FE64Block.vsd
8-4
BI2201-E
8.3.2.
Theory of Operation
Input Channel 64
ADC
10
12
ASIC 16
Front End
64
FPGA
Input Channel 4
ADC
Input Channel 3
ADC
Input Channel 2
ADC
Input Channel 1
ADC
Control (LVDS)
Configuration
10
Backplane
10
12
10
ASIC 1
192
10
Profile
Memory
DBFblock.vsd
BI2201-E
8-5
Theory of Operation
8.3.3.
Core Board
AVIB
B/W Vectors
Overlay
VIME
FFT + CFM
Vectors
SCON
DOP
Host Bus
Graphics
Local Bus
Embedded
PC
RTSC
Misc I/O
VECP
Beam Sum
SEM
Backplane Interfaces
CoreBlock.vsd
RTSC
RTSC
DSP
RTSC
DSP
Memory
DSP
Memory
Memory
Sharc Bus
SEMIF
(ADSP21161N)
(MBGA225)
VECP,
SCON
(XC2S150E)
(FG456)
Bus
Buffers
Mosc
DBF
RTSC
Sequencer
FPGA
TGC
DAC
Local Bus
240MHz
RTSC
DSP
FE
Clock
Gen.
Host
Interface
FPGA
(XC2S150E)
(FG456)
Shared
Shared
Memory
Memory
Host Bus
CoreBlock_RTSC.vsd
8-6
The RTSC takes care of the hard real-time tasks in the scanner. A local processor
(Sharc DSP) is located here to perform a number of tasks:
Theory of Operation
A Host Interface provides a memory area, which is shared between the Embedded PC
and the Sharc DSP; it is through this memory only that they communicate and
exchange data.
The Sequencer part of the RTSC generates all timing signals for the acquisition of
ultrasound data to the FE and DBF and also internally to VECP, DOP and SCON. It
also contains the physical backplane interface to the front-end boards (FE, SEMIF and
DBF).
Finally, a master clock generator provides all system clocks (except for the Embedded
PC), both fixed and variable. The variable clock frequency is determined by the
Sequencer (ultimately controlled by the Sharc DSP).
RF Input
Buffer
DBF
CFM
Input Buffer
CFM
Output Buffer
CFM
DSP
Clusters
& LUT
VECP
FPGA
(XC3S1500)
(FG676)
B-mode
LUTs
Bus
Buffers
Local Bus
to SCON
FFT
DSP
(ADSP21161N)
(MBGA225)
to Audio Codec
CoreBlock_VPDOP.vsd
BI2201-E
8-7
Theory of Operation
Local Bus
DSC
ASIC
Review
Review
Memory
Memory
SCON
FPGA
Xform Coeff.
Memory
(XC3S1000)
(FG456)
Display
Memory
to VIME
LUT
CoreBlock_SCNV.vsd
Host Bus
Capture
Buffer
from SCNV
Video Merge
and Capture
FPGA
(XC3S1000)
(FG456)
merged graphics
digital video
Merge
Buffer
from PC
to/from AVIB
CoreBlock_VIME.vsd
8-8
BI2201-E
Theory of Operation
The VIME sub-system receives image streams from the Embedded PC, the SCON
and from external video equipment and merges the streams together, de-interlacing
the video input on the way, for display on the main monitor. This merging includes
scaling of either SCON or external video for picture-in-picture function. The Embedded
PC sets a window size and a transparency color for each of the images in order to
define the overlay priority. The VIME system also captures the merged frames, thus
allowing the Embedded PC to access screen snapshots or real-time sequences and
transfer them to documentation units such as disk drives, DICOM or digital VCRs. It
also captures SCON images for the benefit of the 3D application.
Embedded PC:
Host Bus
PCI Bridge
(PCI9056)
(PBGA256)
to VIME
from FFT
4 x I2C
PCI Bus
+ Backplane
graphics
Free PMC Slot
2"
HDD
Board ID
EEPROM
AVR
IEEE1394
Link & Phy
Embedded PC
(ETX Module)
E-IDE
analog audio
analog audio
digital audio
Audio
Codec
analog audio
analog audio
1394
Speaker
Backplane
Backplane
LAN
USB
COM
Backplane
CoreBlock_PC.vsd
BI2201-E
8-9
Theory of Operation
8.3.4.
VIME VIME
(VGA) (I2C)
VIME
(Video)
24
2
8/16
GPIO
Register
Video
DAC
Mux
8/16
Switch
VGA-to-TV
Digital Video
Encoder
AL128
Analog
Encoder
2 x VGA
Y/C + CVBS
Field
Store
AL422
Digital Video
Decoder
VPX3220A
24
Video
ADC
AD9884
Field
Store
AL422
Board ID
EEPROM
RGBS
Y/C
CVBS
RGB
AVIBblock.vsd
8.3.5.
8-10
BI2201-E
Theory of Operation
8.3.7.
External Devices
External devices are part of the scanner and can be mounted directly on the central
scanner unit (also known as the Engine) containing the main PCBs and Power Supply.
However, they may also be placed away from it at places convenient to the user.
Keyboard Assembly:
Assembly consisting of an alphanumeric keyboard, a pointing device (track ball), and
an ultrasound control panel with dedicated keys, gain dials and TGC slide
potentiometers. A microcontroller handles the key scanning, gain dial decoding and
A/D conversion of the TGC pots. A USB hub in the keyboard connects the various
modules and provides a single USB connection to the Embedded PC.
Monitor:
A standard 15" SVGA CRT or 17 flat panel monitor.
Mass Storage:
CD-RW with USB interface.
Power supply (non removable):
The power supply is a SMPS type. Four isolated outlets for external equipment are
provided. One is allocated for the monitor and one for a built-in video printer. The unit
supplies 3.3VD, 5VD, 12VD, 5VA, 12VA, HVMUX, HV. The HV output for the
transmitters is not programmable. A standard ATX handshake allows the Windows
powered PC to control start-up and shut-down. The unit also contains circuitry for
controlling the two fans below the card cage and its own fan.
8.3.8.
Optional Devices
Other devices may be connected. An accessory is a device, which the customer can
purchase and use without further requirements. It attaches to an external connector. A
field option is a device or module, which a customer purchases after the scanner, but
it may only be installed inside the scanner by service personnel.
Foot Switch (accessory):
Existing foot switch UA1215 or a similar one with three pedals (B-K Number TBD),
which is recommended for use with the 3D application. The foot switch is attached to
the Backplane (rear panel). Some pedal functions are user programmable.
Palm Control Unit (accessory):
Existing Palm Control Unit UA1270. Connects to the Backplane (COM1 on the rear
panel) using existing adapter UA1271. Requires a 5VDC supply of 100mA.
Magnetic Tracking Freehand (accessory):
Existing MTF Upgrade kit UA0576, can be connected to the Backplane (rear panel)
COM1 port.
BI2201-E
8-11
Theory of Operation
8-12
BI2201-E
System Interconnections
Section 9
SYSTEM INTERCONNECTIONS
CONTENTS
PAGE
9.1.
Introduction.............................................................................................................................................. 2
9.1.1.
9.2.
9.2.1.
Footswitch ...........................................................................................................................................4
9.2.2.
Monitor ................................................................................................................................................4
9.2.3.
USB .....................................................................................................................................................4
9.2.4.
LAN .....................................................................................................................................................4
9.2.5.
S-VHS .................................................................................................................................................4
9.3.
BI2201-E
9-1
System Interconnections
9.1.
Introduction
By showing the cables between PCB's, and external cable connections this section
aims to help you keep track of important signals internally in the 2202 and to other
equipment.
9-2
BI2201-E
System Interconnections
AVCON
BP
2 x FAN
2 x USB
10/100 LAN
2 x IEEE1394
COM1
Keyboard
Foot Switch
ECG
Fsw
2 x SVGA
RGBS
CD-RW
PSU Handshake
PSU I2C
Core
SVGA
Inputs
RGBS
2 x Audio
Light Pen
RS-232
IDE
Light Pen
AVIB I2C Bus
Audio
Video
Units:
SEM: Single Element Module
MCM: 3D Mover Control Module
SEMIF: Single Element Module Interface
FE: Front End Board
DBF: Digital Beamformer Board
CD-RW: CD-RW Disk Drive
HUB: USB Hub Board
ECG: ECG Module
BP: Backplane
Core: Digital Back End & RTC Board
PSU: Power Supply Unit
AVIB: Analog Video Interface Board
AVCON: AV Connector Board
Mover Unit
SE Xducer
MCM
SEM
SPI Bus
Control Bus
Real-time Ctrl
SEM Output
PWR
Button
SEMIF
PWRBTN
Embedded PC
Status
Array Xducer
FE
Light Pen
Doc Trig
Audio
Video
TAP
FE Control Bus
FE Real-time Control
I2C + SPI Bus
AVIB
FEout
Beam Sum
DBF Control Bus
DBF Real-time Control
DBF FPGA Configuration
DBF I2C Bus
DBF
TAP
S-video
CVBS
HUB USB
HUB Audio
Array Xducer
3 x CVBS
2 x Doc Trig
2 x USB
Array Xducer
S-video
2 x Audio
HUB
USB
2 x Speaker
PSU
Tach
Outputs
ECG
InterfaceDiagram.vsd
9-3
System Interconnections
9.2.
9.2.1.
Footswitch
9.2.2.
Monitor
This is a standard 15-pin DSUB socket.
9.2.3.
USB
This is a standard USB Type A socket.
9.2.4.
LAN
This is a standard RJ45 10/100 Mbps socket.
9.2.5.
S-VHS
This is a standard S-VHS socket.
9-4
BI2201-E
9.3.
System Interconnections
Pin Name
Colour
1
2
3
4
5
6
7
8
Red
Red
White
Yellow
Blue
Black
Red
Yellow
+5VA
+5VA
-5VA
+12VA
-12VA
AGND
+100VA
+80VA
Size
AWG
16
16
16
16
16
16
20
20
UL
Style
1007
1007
1007
1007
1007
1007
1007
1007
Colour
Name
Pin
Black
Black
Black
Black
Black
Black
White
Gray
AGND
AGND
AGND
AGND
AGND
AGND
-100VA
-80VA
9
10
11
12
13
14
15
16
Pin Name
Colour
1
2
3
4
5
6
7
8
9
10
Brown
Gray
Brown
Yellow
Orange
Orange
Orange
Red
Red
Red
BPPSUSCK
BPPWR_OK
BP5VSB
12VD
3V3D
3V3D
3V3D
5VD
5VD
5VD
BI2201-E
Size
AWG
20
20
20
16
16
16
16
16
16
16
UL
Style
1007
1007
1007
1007
1007
1007
1007
1007
1007
1007
Size
AWG
20
16
20
16
20
16
16
16
16
16
Colour
Name
Pin
Brown
Black
Brown
Black
Brown
Black
Black
Black
Black
Black
BPPSUSDA
GND
BPPS_ON_L
GND
BPSTOP_L
GND
GND
GND
GND
GND
11
12
13
14
15
16
17
18
19
20
9-5
System Interconnections
9-6
BI2201-E