Escolar Documentos
Profissional Documentos
Cultura Documentos
1046-2023/99
logical structures. In LSCM, small apertures (pinholes) are used to eliminate the fluorescent light
from above and below the plane of focus, resulting in
increased resolution compared with that obtained
from conventional fluorescence microscopy. The addition of a powerful laser also allows the investigator
to image deeper inside a tissue than is possible with
a conventional mercury arc lamp. To achieve the
maximum performance from these sophisticated optical machines it is necessary that they are aligned
properly and tuned for efficiency (15). Although the
lasers are powerful and the optical systems are efficiently designed, photons should not be wasted.
From both scientific and quality assurance standpoints, optical performance delivered by the system
should be optimized (1). Without sufficient light being delivered by the optical system, it will be necessary to raise the PMT voltage to visualize the specimen. As the PMT is increased, the noise (CV) also
increases greatly. At higher PMT settings, it is necessary to use more averaging to reduce the noise
(CV) in the image. The need to raise the PMT settings should be minimized by operating the optical
system at maximum efficiency.
Although the human visual system is exquisitely
sensitive to contrast and resolution, it is less adept
at judging gradual luminance variations. Therefore,
it is beneficial to use beads or test slides to statistically quantify system performance. This review describes various methods that have been successfully
used in our laboratory to evaluate the functioning of
the Leica TCS-SP and TCS-4D confocal microscopes.
These tests should be valid for other point scanning
systems and should not only be confined to the
Leica equipment. Additional guidance on optimizing
LSCM performance has been published recently (1,
2).
447
448
MATERIALS
Fluorescent Slides
The field illumination test slides were three fluorescent plastic slides with excitation peak wavelengths of
408, 488, and 590 nm and emission peak wavelengths
of 440, 519, and 650 nm, respectively. The slides have
a chemical embedded within them and become fluorescent at the desired excitation wavelength. They were
obtained from Applied Precision Inc (Issaquah, WA).
Power Meter
The power meter used to measure light on the microscope stage was a Lasermate/Q (Coherent, Auburn,
CA) with visible (LN36) and UV (L818) detectors. A
second power meter (No. 0163-662-00, Coherent, Santa
Clara, CA) was used to measure and control the light
emitted from UV Enterprise laser (Coherent, Santa
Clara, CA).
Beads
The following beads used in this study were obtained
from Spherotech (Libertyville, Illinois): rainbow fluorescent
particles
(10
mm,
RFSP-100),
ex
365,488,568,647; yellow beads (4.2 mm, fps-4052), ex
488; Nile red beads (2.9 mm, fps-3056), ex 568; UV
beads (3.1 mm, fps-3040), ex 365; blue beads (3.2 mm,
fps-306,810), ex 647. Ten-micrometer Fluorospheres
(Fullbright green II, Coulter, Hialeah, FL, ex 488) and
1-mm 1 Tetraspeck beads (T7284 Molecular Probes,
Eugene, OR, ex 365,488,568,647) were used for standardization and spectral registration, respectively.
Biological Test Slides
Two slides of FluoCells (F-14781, and F-14780, Molecular Probes) stained with three fluorochomes were
used as biological test slides. Slide 1 (F-14780) was
stained with the following dyes: Mitotracker red
CMXRos, BODIPY FL phallacidin, DAPI.
Slide 2 (F-14781) was stained with the following
dyes: BODIPY Fl goat anti-mouse IgG, Texas Red-X
phalloidin, DAPI.
Additional slides were made with cells growing on
coverslips, fixed with paraformaldehyde or glutaraldehyde, and stained with DAPI and other suitable fluorochromes.
Axial (z) Resolution Test
To test the axial resolution of the laser scanning
confocal microscope it is necessary to use a reflecting
mirror. A single reflecting mirror, 21 mm square (No.
31008, Edmonds Scientific, Philadelphia, PA) was
glued onto a microscope slide. A coverglass 1.5 (Fischer, Pittsburgh, PA) was placed on top of the slide
with a drop of immersion oil (Leica immersion oil, n 5
LSCM OPTIMIZATION
449
FIG. 1. Field illumination of the following lenses on the Leica DMIRBE inverted microscope: (A) 6.23 (Zeiss 53 Fluotar, N.A. 5 0.25); (B)
103 (Plan Fluotar, N.A. 5 0.3); (C) 203 (Plan-Apo, N.A. 5 0.6); (D) 403 (Plan Fluotar, N.A. 5 0.51.0); (E) 633 (Plan Apo NA 1.2); (F) 1003
(Plan-Apo, N.A. 5 1.4). The Zeiss 53 lens yielded an effective magnification of 6.23 on the Leica system. The illumination pattern was
derived from a plastic fluorescent slide imaged 50 150 mm below the surface.
450
reduction in intensity. The lower-magnification objectives show highly concentrated illumination in the center of the field (bulls-eye). This problem is partially the
result of using the low-magnification objectives on a
machine designed for high-magnification work. The 53
(Fig. 1A) and 103 (Fig. 1B) objectives clearly illustrate
the problems with nonuniform illumination and bullseye patterns. One solution to this problem is to increase the zoom factor. This enlarges the illumination
center and pushes the lower intensities off the field of
view. However, increasing zoom also increases the
magnification and bleaching rate of the sample and
may defeat the purpose of using a low-magnification
objective to observe a large field of illumination. The
nonuniform pattern clearly illustrates a field illumination problem.
Some of the objectives demonstrate illumination intensities that are 40% less than the maximum illumination intensity (53 and 103). Usually the accepted
FIG. 2. Field illumination pattern of dirty (A) and clean (B) 203 (Plan Apo, N.A. 5 0.6) Leica lenses. The variation in intensity from the
left to right side of the field for both images is illustrated by the intensity profile (C). Note that the dirty lens yields a reduction in intensity
of 70% across the field.
LSCM OPTIMIZATION
clean and dirty x-direction intensity profiles of the images in Figs. 2A and 2B is shown in Fig. 2C. A similar
illustration of a dirty lens can be found in Ref. (1).
Power Meter Readings
In order not to waste photons (1) it is necessary to
evaluate the performance of the laser when it first
arrives in the laboratory and periodically thereafter.
Lasers usually need periodic care and both horizontal
and vertical controls have been installed to adjust the
mirrors. Since some of the manufacturers do not allow
users access to these adjustments, it is necessary to
make sure they are functioning at maximum performance. They do need alignment during their lifetime
and it is not always sufficient to turn the power up to
obtain additional power at the microscope stage. Running lasers at excessive power is not good for the life of
the laser and it wastes photons.
A power meter can be placed on the stage to evaluate
the power and stability of the different wavelength
lines. The Bio-Rad 1024 has an open configuration
allowing the laser light to be measured prior to arrival
at the scan head and at the microscope stage to determine laser attenuation in the LSCM system. We have
devised a special holder for a probe detector (Coherent)
to allow it to fit securely on the microscope stage for the
measurement of both UV and visible intensities.
The laser power should be measured with a lowmagnification lens. The test is done the in the following
manner: the lens is lowered to a minimum specified
height, the detector head of the power meter is placed
on the stage, the detector is centered using mercury arc
fluorescence light and laser light, and field is zoomed to
maximum (32 times). Initially the center of the detector can be located by using the mercury arc UV fluorescence light. The LSCM zoom factor is set to maximum to reduce the beam scan and to focus it into the
sweet spot of the detector. The scanner is set at slow
speed bidirectional. The detector position is slightly
adjusted to achieve maximum signal by the microscopes x/y joystick or positioning controls. Once this is
achieved the power can be evaluated with the laser
meter.
The values obtained over time will determine the
stability of the laser and the maximum power output of
the laser. Customarily, it is necessary to allow the laser
to run at the desired power for 30 min prior to testing
its stability. The stability test consists of repeated measurements of the power over time. In our experiments
the argon krypton laser varied less than 10% over 3 to
4 h, based on measurements made approximately every 15 min. The laser can be turned up briefly for a
short period to determine the maximum output at all
three lines. These two controls allow the user to determine the maximum laser output delivered and the
laser stability at the desired power used. It should be
emphasized that this test is performed at the micro-
451
452
ment will reduce the power in the system, necessitating an increase in PMT to compensate for lack of laser
light. When there is insufficient light coming though
the system, careful realignment of the laser beam is
required (a separate procedure done by qualified personnel) to increase the laser output. If this cannot be
done it may be time to replace the laser.
Reflectivity of Dichroic and Barrier Filters
Dichroic filters are made to reflect specific wavelengths of light and pass other wavelengths of light.
The efficiency of the dichroic filter in reflecting light
can be evaluated easily by placing the meter on the
stage and inserting various filters to bounce the light.
For example, with a 488-nm line, placing either a single, double, or triple dichroic in the light path should
reflect successively less light, if they were designed
properly. Photons should not be wasted in a LSCM
system (1). In our Leica TCS4D system, the triple
dichroic (TD) reflected more light than the double dichroic (DD) due to a possible ineffective DD coating.
Although this test can be done with a specific sample,
it is easier and more conclusive to do this with a laser
power meter positioned on the microscope stage.
Laser output can aid the user in determining the
relative efficiency of the dichroic filters. The maximum
power and half-maximum power should be measured
at each laser wavelength and each available dichroic
setting. For example, the 568-nm line output should be
quantified with both the DD 488/568 and TD 488/568/
647 dichroic filters. The relative differences in efficiency between dichroics will allow the user to choose
which filter setting to apply with a specific dye in a
given experiment.
FIG. 3. Axial resolution. Histograms of a front-surface, single reflecting mirror under a coverslip (0.170 mm). The test was done in xz
scanning mode, reflection mode, and a zoom magnification over 10
coupled with a 1003 lens. The resolution is given by the FWHM at
370 nm. Note that the peak intensity is 240 and the half-maximum
intensity is at 120.
LSCM OPTIMIZATION
453
Spherotech). In many systems, the 647-nm line, emitted by the argon krypton laser, begins to fail and the
visualization of the 647-nm excitation bead is a simple
test to determine the existence of this line. Beads may
also be moved to various regions of the field to test field
illumination. If the beads are spatially concentrated,
they can be used to test field illumination. The problem
with this test is that it is possible that the beads in a
given field are centered at different depths and thus
may produce different intensities. To ensure that the
beads are all centered in the same plane of focus it is
necessary to take a stack of images and perform a 3D
reconstruction (maximum projection).
Spectral Registration
To evaluate the registration of the 365-, 488-, 568-,
and 647-nm lines, we use 1-mm multiple-wavelength
fluorescent beads (Tetraspec, Molecular Probes; Rainbow, Spherotech). Figure 4 shows the visible spectral
registration of a lens (1003 Plan-Apo, N.A. 5 1.4)
using a 1-mm multicolored bead (Tetra Spec, excitation
and emission peaks). The acoustical optical transmission filter (AOTF) was used to minimize laser crossover
between detection channels. The image of the bead
showed that the emission of the 488-, 568-, and 647-nm
lines is not totally coherent (Fig. 4). The bead was
imaged (xy and xz scans) with a 243 zoom and a slow
scanning rate, and was averaged eight times. The xy
and xz outlines of the bead (Figs. 4A, 4C) were produced with Image-Pro Plus. A mask, eliminating intensities below half of the maximum intensity in each
color channel, was created for each channel. The resulting images were median filtered (7 3 7); edge detected (Roberts), and combined into a single three-color
image. These procedures result in the smoothed halfmaximum outlines in Figs. 4A and 4C. In addition to
this method of visualization, the conventional intensity
profiles through the center of the bead in the x direction (Fig. 4B) and the z direction (Fig. 4D) are also
shown. Only visible wavelengths were analyzed with
these tests.
One, however, can also compare UV with one or more
of the visible wavelengths (usually 568 nm) in a similar
manner. It should be noted that there could be significant registration misalignments between the UV and
visible wavelengths. The system should be checked in
this manner for each objective used in UV.
Daily Bead Tests
To test the system for stability, the power of a laser line
is set to a specific and constant level using the power
meter with either the 488- or 568-nm line. A large bead
(610 mm) from a population of uniform size and intensity is positioned in the center of the field. This large bead
eliminates the need for a zoom factor and reduces bleaching of the bead. Next, the machine variables (pinhole size,
454
PMT voltage, averaging, etc.) are set to reproducible values that will be used for all future studies. The image of
the bead should be obtained at a depth corresponding to
the maximum lateral diameter of the bead and with
maximum intensity slightly below saturation (5255).
This image is imported into Image-Pro Plus and its intensity values, lying inside a uniform region within the
circumference of the bead, are measured. These intensity
values should be relatively stable over a period. The CV of
the population was defined as CV 5 s/m (s 5 standard
deviation of intensity, m 5 mean intensity). This measure
can be thought of as a noise-to-signal ratio, where m
represents the true signal in the image and s represents
the noise in the image. In this case, the CV that we are
measuring is the variation of intensity within the bead,
as opposed to the variation of intensity among beads. A
reduction in the mean intensity of the population or an
increase in CV implies that there is either a laser power
or system alignment problem.
FIG. 4. The visible spectral registration of a lens (1003 Plan Apo, N.A. 5 1.4) was evaluated with a 1-mm multiple-wavelength fluorescent
bead (Tetra Spec). Image processing procedures were applied to the original, three-channel images to visualize spatial differences in spectral
emission among the channels. These procedures result in the smoothed half-maximum outlines shown in (A) and (C). In addition to this
method of visualization, the conventional intensity profiles through the center of the bead in the x direction (B) and the z direction (D) are
also shown. Each pair of figures demonstrates the lack of complete registration among the three channels.
LSCM OPTIMIZATION
a large 10-mm bead (Coulter) of nearly uniform intensity (small CV). The large bead permitted repeated measurements without bleaching the sample.
By varying the PMT settings, the number of averaged frames (exposure time), and laser power, the
noise associated with various settings was evaluated. The system was set up with a bandpass filter of
505555 nm (BP-FITC), a single dichroic, and laser
excitation wavelength of 488 nm. A fixed region of
interest (ROI) was determined to cover about half of
the beads area and the Leica analysis software permitted the mean and standard deviation (SD) of the
intensities in the ROI to be determined. Image-Pro
Plus can also be used to measure the intensities over
a ROI. The PMT was set at approximately 50-unit
intervals starting at a setting of 449 and ending at
1000. Images corresponding to averaged frame
amounts of 1, 4, and 32 scans are produced at each
PMT setting. The laser power was varied using a
combination of the AOTF and laser power knob. The
laser power was set at a value that allowed the mean
intensity level of the bead to be approximately 150
(out of 255) for each PMT setting. For each PMT
setting between 449 and 1000, the mean intensity
and standard deviation of intensity were measured
on the same 10-mm bead. Thus, the noise associated
with each PMT and averaging amount combination
was evaluated for a uniform particle at a constant
intensity level (5150). The relationship between
PMT value, number of frames averaged, and calculated CV is shown in Fig. 6.
FIG. 5. The intensity of a 10-mm bead was determined at a constant laser power and various PMT settings. The intensity was
measured at low PMT values followed by higher PMT values to
minimize any effects caused by bleaching. Note that the intensity
increases in a nonlinear, almost exponential, fashion as the PMT is
increased in value. Note that some systems may not be set up to yield
a linear relationship between PMT setting (user controlled) and
actual electronic PMT voltage. This test can be used to standardize
the system and evaluate stability over a long period. The intensity
and PMT values are measured in arbitrary units (a.u.).
455
456
It can be clearly shown that, at a given PMT setting, higher averaging values produce a greater reduction of CV (decreased noise). The use of lower
PMT settings allows for less averaging due to less
noise in the system. Thus increased scan exposure
FIG. 7. FluoCells (F-14780, Molecular Probes) were excited, with a 568-nm laser line and detected with a 50-nm bandpass filter (581 631
nm). The resolution was measured at 552 PMT using 1 and 4 averaging. The laser power was adjusted to yield images at the same intensity
level with PMT settings of 552 and 799. At the PMT setting of 552 the image was not averaged (A). At the PMT setting of 799, images were
obtained with no averaging (B), 4 frames averaged (C), and 32 frames averaged (D). The CVs of intensities within the same ROI in the cell
nuclei (lower right) were (A) 49%, (B) 212%, (C) 109%, and (D) 47%, respectively. A linear relationship between bead noise and noise in other
samples is implied by these data.
LSCM OPTIMIZATION
DISCUSSION
The functioning of a LSCM system is dependent on
a number of variables that have to be controlled and
checked periodically to ensure system performance.
We have described the following tests: field illumination for individual lenses, laser power indicators,
dichroic efficiency tests, lens chromatic aberration,
axial resolution, system stability using beads, PMT
stability, and noise analysis. It is important to conduct these tests when the system arrives and routinely thereafter to assess the functionality of the
machine.
To obtain an image, it is necessary to balance laser
power, PMT setting, frame averaging amount, pinhole aperture size, and zoom magnification. Normally, if there is insufficient light to visualize the
sample, it is best to increase the laser power, not the
PMT setting. We have demonstrated that the PMT
should be minimized and averaging should be used
to reduce noise (CV) and increase image quality.
Although the pinhole diameter determines the confocality of the system and should ideally be set at the
size of the first Airy disk, the diameter can be increased if insufficient light is obtained to visualize
the sample. If available, the detection barrier filter
can also be replaced with a Schott long-pass filter to
increase detected light. It is almost always better to
increase the laser power and lower the PMT voltage
for ideal image quality.
We demonstrated the existence of a relationship between PMT, laser power, and frame averaging using a
10-mm test bead. The noise (CV) in an image increases
in a quasi-linear fashion as the PMT voltage increases.
Optimum resolution of an image appears to be determined by choosing a low PMT setting, high frame averaging, and high laser power. These factors yield a low
CV, increasing image quality. The data in Figs. 5, 6,
and 7 are interrelated. Figure 6 can be used as a rough
guideline for generating different image qualities at
different PMT settings. It can also be used to determine how much averaging is necessary at a specific
PMT setting for adequate image quality. Frame averaging is useful to reduce noise, but at the increased
amount of averaging necessary to reduce noise, bleach-
457
458
performance of the machine (1, 2). We have developed some new tests in this study and perfected some
standard tests to assess the functionality of the laser
scanning confocal microscope. The major aim of this
study is to eliminate the subjective nature of assessing the confocal microscope by evaluating only a
user-defined sample. The ability to apply estimates
of noise (CV) and image resolution allows for a better
way of assessing the functionality of the confocal
system (11). It is anticipated that additional quality
assurance tests will be derived to assess this important scientific machine and ensure that it functions
at an optimum performance level.
ACKNOWLEDGMENTS
We thank the Leica Service EngineersJohn Ehli, Dennis Gil,
Paul Lilagan and Ulrich Fuchsfor their assistance. Without their
impute this investigation of LSCM system optimization would never
have been undertaken or accomplished.
REFERENCES
1. Centroze, V., and Pawley, J. (1995) in Handbook of Biological
Confocal Microscopy (Pawley, J., Ed.), 2nd ed., pp. 559 567,
Plenum, New York.
2. Centroze, V., and Pawley, J. (1998) in Cell Biology (Celis, J., Ed.),
2nd ed., Vol. 3, pp. 149 169, Academic Press, New York.
3. Sheppard, C. J. R., and Shotton, D. M. (1997) Confocal Laser
Scanning Microscopy, Bios Scientific, New York.
4. Pawley, J. (1995) in Handbook of Biological Confocal Microscopy,
(Pawley, J., Ed.), 2nd ed., pp. 19 36, Plenum, New York.
5. Roost, F. W. D. (1991) Quantitative Fluorescence Microscopy,
Cambridge Univ. Press, New York.
6. Cogwheel, C. J., and Lark, K. G. (1995) in Handbook of Biological
Confocal Microscopy (Pawley, J., Ed.), pp. 127137, Plenum,
New York.
7. Shapiro, H. (1995) Practical Flow Cytometry, 3rd ed., Wiley
Liss, New York.
8. Russ, J. C. (1995) Image Processing Handbook, 2nd ed., CRC
Press, Boca Raton, FL.
9. Art, J. (1995) in Handbook of Biological Confocal Microscopy
(Pawley, J., Ed.), 2nd ed., pp. 183195, Plenum, New York.
10. Teraskim, M., and Dailey, M. E. (1995) in Handbook of Biological
Confocal Microscopy (Pawley, J., Ed.), 2nd ed., pp. 327344,
Plenum, New York.
11. Watson, J. V. (1992) Flow Cytometry Data AnalysisBasic Concepts and Statistics, Cambridge Univ. Press, New York.