NI 9236
8 AI, 29.4 mV/V, 24 Bit, 10 kS/s/ch
Simultaneous, 350 Quarter-Bridge Completion
Safety Guidelines
Operate the NI 9236 only as described in this document.
Caution Do not operate the NI 9236 in a manner not
| ni.com |
Safety Voltages
Connect only voltages that are within the following limits:
Between any two terminals
Isolation
Channel-to-channel
Channel-to-earth ground
Continuous
Withstand
30 V maximum
None
60 VDC,
Measurement Catergory I
1,000 Vrms, verified by a 5 s
dielectric withstand test
National Instruments |
| ni.com
0.2 F maximum
National Instruments |
| ni.com |
National Instruments |
ni.com
EXC0
AI0
RC0
EXC2
AI2
RC2
EXC4
AI4
RC4
EXC6
AI6
RC6
1
2
3
4
5
6
7
8
9
10
11
12
EXC1
AI1
RC1
EXC3
AI3
RC3
EXC5
AI5
RC5
EXC7
AI7
RC7
13
14
15
16
17
18
19
20
21
22
23
24
National Instruments |
Description
EXC
AI
RC
EXC
AI
RC
10
ni.com |
NI 9236
You must connect each EXC terminal to only one strain gage to
maintain the channel-to-channel crosstalk performance of the
module.
For the best system accuracy, observe the following conditions
when connecting to the NI 9236.
Set up the connections to EXC and RC with equal lengths of
an identical wire type and gauge.
Connect the AI terminal directly at the sensor instead of
shorting AI to RC directly at the terminals.
National Instruments |
11
1. Insert the screwdriver into a spring clamp activation slot to open the
corresponding connector terminal.
2. Press a wire into the open connector terminal.
3. Remove the screwdriver from the activation slot to clamp the wire into
place.
12
ni.com |
National Instruments |
13
Where to Go Next
CompactRIO
NI CompactDAQ
NI 9236 Datasheet
NI 9236 Datasheet
NI-RIO Help
NI-DAQmx Help
LabVIEW Help
RELATED INFORMATION
C Series Documentation
& Resources
ni.com/info cseriesdoc
Located at ni.com/manuals