Micro Motion Alarm Codes

Micro Motion alarms are classified into three severity levels


(Top) LED Color
Red
Amber
Green

Severity Level
Fault
Informational
Ignore

Fixable On Site
95% No
99% Yes
No Action Required

Based on experience, LSI has created a chart with the alarm codes and the likelihood of them being
"fixed" on site. Meters that cannot be fixed on site are typically discarded as it usually not cost effective
to have the factory repair a meter with worn flow tubes.
Alarm Code
A001
A002
A003
A004
A005
A006
A008
A009
A010
A011
A012
A013
A014
A016
A017
A018
A019
A020
A021
A022
A023

Problem

EEPROM Checksum Error (CP)


RAM Error (CP)
Sensor Failure
Temperature Sensor Failure
Input Overrange
Not Configured
Density Overrange
Transmitter Initializing/Warming Up
Calibration Failure
Zero Too Low
Zero Too High
Zero Too Noisy
Transmitter Failed
Line RTD Temperature Out-of-Range
Meter RTD Temperature Out-of-Range
E EPROM Checksum Error
RAM or ROM Test Error
Calibration Factors Unentered (FlowCal)
Incorrect Sensor Type (K1)
E EPROM Config. CB Corrupt (CP)
E EPROM Powerdown Totals Corrupt (CP)

LED
Red
Red
Red
Red
Red
Red
Red
Red
Red
Red
Red
Red
Red
Red
Red
Red
Red
Red
Red
Red
Red

Fixable On Site
No
No
Maybe
No
Maybe
Maybe
Maybe
Maybe
Maybe
Maybe
Maybe
Maybe
No
Maybe
Maybe
No
No
Maybe
No
No
No

Alarm Code
A024
A025
A026
A027
A028
A031
A032
A033
A034
A035
A100
A101
A102
A103
A104
A105
A106
A107
A108
A109
A110
A111
A112
A113
A114
A115
A116
A117
A118
A119
A120
A121
A131
A132

Problem

E EPROM Program Corrupt (CP)


Protected Boot Sector Fault (CP)
Sensor/Transmitter Communication Error
Security Breach
Core Processor Write Failure
Low Power
Meter Verification/Outputs in Fault/Outputs
Fixed
Sensor OK/Tubes Stopped by Process
Meter Verification failed
Meter Verification Aborted
Primary mA Output Standard
Primary mA Output Fixed
Drive Overrange
Data Loss Possible
Calibration in Progress
Slug Flow
Burst Mode Enabled
Power Reset Occurred
Event 1 Triggered
Event 2 Triggered
Frequency Output Saturated
Frequency Output Fixed
S/W Upgrade recommended
Secondary mA Output Saturated
Secondary mA Output Fixed
External Input Error
API: Temperature Outside Standard Range
API: Density Outside Standard Range
Discrete Output 1 Fixed
Discrete Output 2 Fixed
ED: Unable to Fit Curve Data
ED: Extrapolation Alarm
Meter Verification
Simulation Mode Active

LED

Fixable On Site

Red
Amber
Amber
Amber
Amber
Amber
Amber
Amber
Amber
Amber
Amber
Amber
Amber
Amber
Amber
Amber
Amber
Amber
Amber
Amber
Amber
Amber
Amber
Amber
Amber
Amber
Amber

Maybe
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
No
Yes
Yes
Yes

Red
Red
Red
Red
Red
Red
Red Or Amber

No
No
No
No
No
No
Maybe

This above information has not been evaluated by Micro Motion.

Call LSI for assistance with meter/computer calibration and inventory issues.

www.liquidsystemsinc.com

Você também pode gostar